Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K211003

    Validate with FDA (Live)

    Date Cleared
    2022-01-19

    (292 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by chemotherapy drugs.

    Device Description

    The SHOWA® Blue Nitrile Powder Free Medical Examination Glove is a single use, disposable device made from a Nitrile Butadiene Rubber, blue in color, powder free and non-sterile (per 21 CFR Part 880.6250, class I). The device meets all the specifications in ASTM D6319-10. Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs.

    The SHOWA® Blue Nitrile Powder Free Medical Examination Glove is designed to be used for medical purposes to be worn on the examiner's hands or fingers to prevent contamination between the patient and the examiner (product code LZA). In addition, these gloves were tested for use with chemotherapy drugs (product code LZC) in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by chemotherapy drugs.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device, specifically the "SHOWA® Blue Nitrile Powder Free Medical Examination Glove." It outlines the acceptance criteria and the studies performed to demonstrate the device's substantial equivalence to a predicate device.

    Here's the requested information:

    1. A table of acceptance criteria and the reported device performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-10 (Dimensions)Demonstrate accurate sizingOverall Length (mm): 220 mm (sizes XS – S) and 230 mm (sizes M - XL) Width (± 10 mm): Size XS = 70 mm, Size S = 80 mm, Size M = 95 mm, Size L = 110 mm, Size XL = 120 mm Thickness at Palm (mm): 0.05 min. Thickness at Finger Tip (mm): 0.05 min.Pass
    ASTM D6319-10 (Tensile Properties)Demonstrate tensile conforming propertiesBefore Aging (D 412): Tensile Strength (MPa) = 14 min Ultimate Elongation (%) = 500 min. After Aging (D 573) (70°C ± 2°C for 166 hrs ± 2 hrs.): Tensile Strength (MPa) = 14 min Ultimate Elongation (%) = 400 minPass
    ASTM D6319-10 (Freedom from Holes)Demonstrate glove integrityFreedom from Holes (D 5151): AQL 2.5 Inspection Level G-1Pass
    ASTM D6319-10 (Residual Powder)Demonstrate biocompatibility: Residual powderResidual Powder (D 6124): ≤ 2.0 mg/pcPass
    Primary Skin Irritation Test ISO 10993-10Demonstrate biocompatibility: Skin irritationUnder the conditions of the study, Not a primary skin irritant.Pass
    Dermal Sensitization Assay ISO 10993-10Demonstrate biocompatibility: Skin sensitivityUnder the conditions of the study, Not a contact sensitizer.Pass
    Acute Systemic toxicity Study ISO 10993-11Demonstrate biocompatibility: acute systemic toxicityUnder conditions of the ISO Acute Systemic Injection test, not toxicPass
    Open box testing AAMI 11737-1Demonstrate acceptable open box bioburdenAcceptable bioburden levels after 30 days open box.Pass
    ASTM D6978-05 (Chemo Permeation)Resistance against Chemotherapy Drugs (permeation time)Breakthrough greater than 240 min. for: Blenoxane, Busulfan, Carboplatin, Cisplatin, Cyclophosphamide, Cytarabine, Dacarbazine, Daunorubicin HCl, Docetaxel, Doxorubicin HCl, Epirubicin HCl, Etoposide, Fludarabine, Fluorouracil, Gemcitabine, Idarubicin HCl, Ifosfamide, Irinotecan, Mechlorethamine HCl, Melphalan, Methotrexate, Mitomycin C, Mitoxantrone, Paclitaxel, Rituximab, Trisenox, Vincristine Sulfate. Breakthrough at 73.7 min. for: Carmustine (Warning: Do not use) Breakthrough at 25.4 min. for: ThioTepa (Warning: Do not use)All tested drugs met or exceeded the specified breakthrough times. (Carmustine and ThioTepa had specific breakthrough times below 240 mins, with warnings not to use).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each specific test (e.g., number of gloves tested for dimensions, tensile strength, or chemotherapy permeation). It references specific ASTM and ISO standards for testing, which typically define the appropriate sample sizes and testing methodologies. The data provenance (country of origin, retrospective/prospective) is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to this type of device submission. The "ground truth" for medical gloves is established by quantitative laboratory testing against defined material and performance standards (ASTM, ISO). There are no human experts judging images or clinical outcomes in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., radiology reads) to resolve disagreements. Here, the "truth" is determined by objective physical and chemical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. MRMC studies are typically performed for AI-powered diagnostic devices that assist human readers. This submission is for a physical medical device (examination glove).

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. There is no algorithm involved; it is a physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is based on established industry standards and laboratory testing results. Specifically:

    • Dimensions, Physical Properties (Tensile Strength, Elongation), Freedom from Holes, Residual Powder: Defined by the ASTM D6319-10 standard.
    • Biocompatibility (Skin Irritation, Dermal Sensitization, Acute Systemic Toxicity): Defined by ISO 10993-10 and ISO 10993-11 standards.
    • Resistance against Chemotherapy Drugs: Defined by ASTM D6978-05 standard.
    • Open box bioburden: Defined by AAMI 11737-1.

    8. The sample size for the training set

    This is not applicable. There is no "training set" as this is not an AI/machine learning device. The performance is assessed through direct testing of the product.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190159

    Validate with FDA (Live)

    Date Cleared
    2019-05-06

    (96 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Showa® Medical Exam Glove (Green) Powder-free, Disposable Nitrile Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    The Showa® Medical Exam Glove (Green) Powder-free, Disposable Nitrile Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. The gloves are single use only, supplied in a nonsterile state, and come in six (6) sizes: XS, S, M, L, XL, and XXL. The gloves are made of Nitrile Butadiene Rubber, designed, and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. Their physical and performance characteristics meet all requirements of ASTM D6319-10. The gloves are powder-free and meet the requirements of ASTM D6124-06. No powder is used during any stage of manufacturing. Rather, the gloves are manufactured using online chlorination - a process wherein the examination gloves are dipped in a dilute chlorine solution to reduce surface friction. More specifically, through the chlorination reaction, the compound is stabilized and further cross-linked and the gloves' surface becomes hardened and smooth, thereby providing the gloves with good donning properties. The gloves then are removed from the chlorination solution and neutralized in an aqueous ammonia solution and washed until any residual chlorine is removed, resulting in the gloves' surface pH being similar to that of water. Once the wash process is completed, the reactive material is removed, thus ceasing the chlorination process.

    AI/ML Overview

    The document describes the non-clinical testing and acceptance criteria for the "Showa® Medical Exam Glove (Green) Powder-free, Disposable Nitrile Glove" (K190159). The study aimed to demonstrate that the proposed device is substantially equivalent to a legally marketed predicate device (K082125).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / SpecificationStandard / Test MethodReported Device Performance
    Performance StandardsASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical ApplicationConforms
    ASTM D6124-06 Standard Test Method for Residual Powder on Medical GlovesConforms
    ASTM D5151-06 Standard Test Method for Detection of Holes in Medical GlovesConforms
    ASTM D7160-16 Standard Practice for Determination of Expiration Dating for Medical GlovesConforms (Shelf Life: 3 years, as noted in the technological comparison table, and this standard supports that determination).
    ASTM D412-15a Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—TensionConforms (Indicated by compliance with ASTM D6319-10, which includes physical property requirements like tension).
    Water TightnessASTM D5151-06 Standard Test Method for Detection of Holes in Medical GlovesConforms
    ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical ApplicationConforms
    ASTM D7160-16 Standard Practice for Determination of Expiration Dating for Medical GlovesConforms
    BiocompatibilityISO 10993-10 Skin Irritation Study"Under the conditions of the study, the non-polar and polar device extracts did not elicit a response in the animal model. Therefore, the device is not a skin irritant."
    ISO 10993-10 Closed Patch Sensitization Study"Under the conditions of the study, the non-polar and polar device extracts did not elicit a response in the animal model. Therefore, the device is not a contact sensitizer."
    ISO 10993-11 Acute Systemic Toxicity Study"Under conditions of the ISO Acute Systemic Injection Test, the non-polar and polar device extracts did not elicit a response in the animal model. Therefore, the device does not present an acute toxicity potential."
    Powder ResidualASTM D6124-06 (acceptance criterion: <2mg/glove)Meets ASTM D6124 of <2mg/glove
    Physical PropertyASTM D6319-00 (Note: document mentions D6319-10 in other sections, but D6319-00 here)Meets ASTM D6319-00 (This is from the comparison table, it seems to be an older version listed for the predicate, but the proposed device explicitly states it meets D6319-10 in the Summary of Non-Clinical Testing section, which encompasses physical properties). For consistency, the primary standard for the proposed device is D6319-10.

    Note on ASTM D6319 versions: The "Technological Characteristic Comparison Table" lists "Meets ASTM D6319-00" for both predicate and proposed devices under "Physical Property." However, the "Summary of Non-Clinical Testing" for the proposed device explicitly lists "ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application" as a standard it conforms to. It's likely that compliance with D6319-10 (a newer version) supersedes or includes requirements from D6319-00.

    2. Sample size used for the test set and the data provenance

    The document refers to non-clinical testing based on various ASTM and ISO standards for medical gloves. These standards typically involve specific sampling plans and test methods for performance, water tightness, and biocompatibility.

    • Sample Size: The exact sample sizes for each test within the ASTM and ISO standards are not explicitly stated in this summary. These standards define the required sample sizes for tests such as water tightness (e.g., AQL levels), tensile properties, and force at break. For biocompatibility studies (ISO 10993), specific numbers of animal models are used per test, but are not detailed in this summary.
    • Data Provenance: The studies are described as "non-clinical testing" conducted to demonstrate conformity with recognized international standards (ASTM and ISO). This implies laboratory testing of the manufactured gloves. There is no information regarding the country of origin of the data or whether it was retrospective or prospective, beyond it being a submission for premarket notification K190159 for a newly manufactured device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable to this device submission. This document pertains to the physical and biological performance testing of a medical exam glove, not a diagnostic device requiring expert interpretation of results to establish ground truth. The "ground truth" here is the objective measurement of the glove's properties against established standard specifications (e.g., tensile strength, freedom from holes, biocompatibility).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this involves objective physical and chemical testing against established standards, not interpretation by human adjudicators.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-enabled diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth is established by objective measurements and chemical/biological testing against pre-defined, standardized acceptance criteria outlined in the referenced ASTM and ISO standards (e.g., ASTM D6319-10, ASTM D6124-06, ASTM D5151-06, ISO 10993-10, ISO 10993-11). For example, the "ground truth" for water tightness is the absence of holes when tested using the specified method; for biocompatibility, it's the lack of specified adverse biological responses in animal models.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-based device that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. No training set is used for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180467

    Validate with FDA (Live)

    Date Cleared
    2018-10-10

    (231 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

    Device Description

    The SBG Blue Nitrile Powder Free Medical Examination Glove is a single use, disposable device made from a nitrile latex compound. Blue in color, powder free and non-sterile.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "SBG Blue Nitrile Powder Free Medical Examination Glove." This document outlines the device's characteristics, its comparison to a predicate device, and the results of various non-clinical tests to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicAcceptance CriteriaProposed Device Performance (SBG Blue Nitrile Powder Free Medical Examination Glove)Predicate Device Performance (EMG Blue Nitrile Medical Examination Glove Powder Free)Comparison
    Product CodeLZALZALZASame
    Intended UseDisposable device for medical purposes, worn on hands/fingers to prevent contamination, for over-the-counter use.Matches acceptance criteria description.Matches acceptance criteria description.Same
    Material UseNot made from Natural Rubber LatexNitrile latex compoundNitrile latex compoundSame
    ColorBlueBlueBlueSame
    SterilityNon-SterileNon-SterileNon-SterileSame
    Dimensions: Overall Length (mm)220 mm (sizes XS-S), 230 mm (sizes M-XL)Size Medium: 234 mmSize Medium: 245 mmSame
    Dimensions: Width (± 10 mm)Size S = 80 mm, Size M = 95 mm, Size L = 110 mm, Size XL = 120 mmSize Medium: 97 mmSize Medium: 94 mmSame
    Dimensions: Thickness at Finger (mm)All Sizes = 0.05 mm0.11 mm0.17 mmSame
    Dimensions: Thickness at Palm (mm)All Sizes = 0.05 mm0.10 mm0.10 mmSame
    Physical Properties
    Tensile Strength (MPa)14 min.3734Same
    Ultimate Elongation (%)500 min.590600Same
    After Aging (ASTM 6319-10 (D573) - 70°C ± 2 °C for 166 hrs ± 2 hrs.): Tensile Strength (MPa)14 min.3830Same
    After Aging: Ultimate Elongation (%)500 min.533560Same
    Freedom from Holes (D 5151)AQL 2.5 Inspection Level G-1, Accept at 5 failures, Reject at 6 failuresMeets ASTM D5151-06Meets ASTM D5151-06Same
    Residual Powder (D 6124)≤ 2.0 mg/pc0.18 mg0.20 mgSame
    Biocompatibility
    Primary Skin Irritation Test (ISO 10993-10:2010 (E))(Implicitly: not a primary skin irritant)Under the conditions of the study, not a primary skin irritant.Under the conditions of the study, not a primary skin irritant.Same
    Dermal Sensitization Assay (ISO 10993-10:2010 (E))(Implicitly: not a contact sensitizer)Under conditions of the study, not a contact sensitizer.Under conditions of the study, not a contact sensitizer.Same
    Safety Assessment (Systemic Toxicity)(Implicitly: low potential health risk)The results from the safety assessment demonstrated the device presents a low potential health risk./ (Not explicitly stated for predicate in table, but implied by "Same" in other biocompatibility tests)Different (Table shows a "/" for predicate implying direct data not provided in this row)

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document does not explicitly state the specific sample sizes used for each test (e.g., freedom from holes, physical properties, biocompatibility) or their data provenance (country of origin, retrospective/prospective). However, for tests like "Freedom from Holes (D 5151)," it mentions an "AQL 2.5 Inspection Level G-1," which implies a statistically defined sampling plan as per ASTM D5151-06. The standard for biocompatibility tests (ISO 10993-10) also specifies testing guidelines.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not provided in the document. The document describes non-clinical tests (physical properties, biocompatibility) of a medical examination glove, which typically rely on standardized testing protocols rather than expert consensus on a test set (as would be the case for AI/imaging devices).

    4. Adjudication Method for the Test Set:

    This information is not applicable and therefore not provided in the document. Adjudication methods (like 2+1, 3+1) are common in studies involving human interpretation or subjective assessments, particularly in AI models for medical imaging. The tests described here are objective, quantitative measurements of material properties and biological responses.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for AI-powered diagnostic or interpretive tools to assess how human readers' performance changes with AI assistance. The device in question is a physical glove, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No, a standalone study as defined for AI algorithms was not done. The device is a physical product, and its performance is evaluated based on its material properties and safety characteristics, not an algorithm's output.

    7. The Type of Ground Truth Used:

    The "ground truth" for the tests described are objective physical and chemical measurements and biological responses as defined by specific ASTM and ISO standards. For example:

    • Freedom from Holes: Defined by the ASTM D5151-06 standard and its specified AQL (Acceptance Quality Limit).
    • Tensile Strength & Ultimate Elongation: Measured according to material testing standards.
    • Residual Powder: Measured according to ASTM D6124.
    • Biocompatibility (Primary Skin Irritation, Dermal Sensitization): Determined by standardized animal or in-vitro tests following ISO 10993-10:2010 (E) protocols. The outcomes (e.g., "not a primary skin irritant") serve as the ground truth based on the established biological criteria of these tests.

    8. The Sample Size for the Training Set:

    This concept is not applicable to this device. A "training set" is used for machine learning algorithms. This device is a physical glove, not an AI model, and therefore does not have a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180468

    Validate with FDA (Live)

    Date Cleared
    2018-10-10

    (231 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

    Device Description

    The SBG Black Nitrile Powder Free Medical Examination Glove is a single use, disposable device made from a nitrile latex compound, black in color, powder free and non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Pre-Market Notification for a medical device, specifically the "SBG Black Nitrile Powder Free Medical Examination Glove" (K180468).

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comparison table (page 5) between the proposed device (SBG Black Nitrile Powder Free Medical Examination Glove, K180468) and a predicate device (EMG Black Nitrile Medical Examination Glove Powder Free, K141579). This table serves as the primary source for acceptance criteria and reported performance for the proposed device, demonstrating equivalence to the predicate.

    CharacteristicsAcceptance Criteria (from Predicate/Standards)Proposed Device Performance (SBG Black Nitrile Powder Free Medical Examination Glove K180468)Comparison Status
    Product CodeLZALZASame
    Intended UseA patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The device is for over-the-counter use.A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.Same
    Material UseNot made from Natural Rubber LatexNitrile latex compoundSame
    ColorBlackBlackSame
    SterilityNon-SterileNon-SterileSame
    Dimensions
    Overall Length (mm)220 mm = (sizes XS-S), 230 mm = (sizes M-XL) (Standard)272 mm Minimum All SizesSimilar (Proposed device exceeds minimums)
    Width (± 10 mm)Size S = 80 mm, Size M = 95 mm, Size L = 110 mm, Size XL = 120 mm (Standard)Size S = 70-90 mm, Size M = 85-105 mm, Size L = 100-120 mm, Size XL = 110-130 mmSimilar (Proposed device falls within ranges)
    Thickness at Finger (mm)All Sizes = 0.05 mm (Standard)0.16 mm Minimum All SizesSimilar (Proposed device exceeds minimum)
    Thickness at Palm (mm)All Sizes = 0.05 mm (Standard)0.14 mm Minimum All SizesSimilar (Proposed device exceeds minimum)
    Physical Properties Before Aging (ASTM 6319-10 (D 412))
    Tensile Strength (MPa)≥ 14 min.≥ 14Same
    Ultimate Elongation (%)≥ 500 min.500 min.Same
    Physical Properties After Aging (ASTM 6319-10 (D573) - 70°C ± 2 °C for 166 hrs ± 2 hrs.)
    Tensile Strength (MPa)≥ 14 min.≥ 14Same
    Ultimate Elongation (%)≥ 500 min.500 min.Similar
    Freedom from Holes (D 5151)AQL 2.5 Inspection Level G-1, Accept at 5 failures, Reject at 6 failuresAQL 2.5 Inspection Level G-1, PassSame
    Residual Powder (D 6124)≤ 2.0 mg/pcContains less than 2.0 mgSame
    Biocompatibility Test - Primary Skin Irritation Test (ISO 10993-10:2010 (E))(Implicit: not a primary skin irritant)Under the conditions of the study, not a primary skin irritant.Same
    Biocompatibility Test - Dermal Sensitization Assay (ISO 10993-10:2010 (E))(Implicit: not a contact sensitizer)Under conditions of the study, not a contact sensitizerSame
    Safety Assessment (Systemic Toxicity)(Implicit: acceptable health risk)The results from the safety assessment demonstrated the device presents a low potential health riskDifferent (Predicate not shown, but implies satisfactory result)

    Note on "Acceptance Criteria": For characteristics where a numerical value is given (e.g., Tensile Strength), that is the explicit acceptance criteria. For qualitative characteristics (e.g., Material Use, Color), the "Acceptance Criteria" column reflects the shared characteristic with the predicate or a general industry expectation. For biocompatibility and safety assessments, the criteria are implied to be "not a primary skin irritant," "not a contact sensitizer," and "low potential health risk," respectively.

    2. Sample size used for the test set and the data provenance

    The document refers to various ASTM and ISO standards for testing (e.g., D 412, D 573, D 5151, D 6124, ISO 10993-10:2010 (E)). These standards often specify sample sizes for their respective tests. However, the exact sample sizes used for each test performed on the proposed device are not explicitly stated in this summary.

    The data provenance is not specified regarding country of origin or whether it was retrospective or prospective. It is assumed these were prospective tests performed by the manufacturer or contracted labs to verify the device's conformance to standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is not applicable to the evaluation of this device. The device is a "Patient Examination Glove," which is a Class I device and its performance is assessed through physical, chemical, and biological testing against established standards, not through expert consensus on diagnostic images or clinical outcomes.

    4. Adjudication method for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials or studies involving human interpretation or subjective assessments. The tests for a medical examination glove are objective measurements against predefined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is an examination glove, not an AI-powered diagnostic tool. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. As mentioned above, the device is a physical examination glove, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is based on established national and international standards for medical examination gloves (e.g., ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993-10). These standards define the acceptable physical properties (tensile strength, elongation, dimensions), freedom from holes, residual powder levels, and biocompatibility.

    8. The sample size for the training set

    This information is not applicable. The concept of a "training set" refers to data used to train machine learning models. This device is a physical product and does not involve machine learning.

    9. How the ground truth for the training set was established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1