Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242528

    Validate with FDA (Live)

    Date Cleared
    2025-01-10

    (137 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.

    Device Description

    The Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa. The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low. The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients. The product is provided non-sterile, and not to be sterilized by the user prior to use.

    AI/ML Overview

    Acceptance Criteria and Study for Blood Pressure Monitor (RAK series)

    The provided document describes the acceptance criteria and the study that proves the Blood Pressure Monitor (RAK series) meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Blood Pressure Monitor are based on various international standards, primarily IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers, and ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. These standards define the accuracy requirements for blood pressure measurement devices.

    Acceptance Criteria (Standard Reference)Specific RequirementReported Device Performance
    Blood Pressure AccuracyMean error of differences (device vs. reference)Not over the limits of ISO 81060-2:2018
    (ISO 81060-2:2018)Standard deviation of differences (device vs. reference)Not over the limits of ISO 81060-2:2018
    Absolute difference for individual subject (within 5 mmHg)Implicitly met if mean error & SD are within limits
    Blood Pressure Measurement Range25-255 mmHg25-255 mmHg (verified by IEC 60601-1 and IEC 80601-2-30)
    Heart Rate Measurement Range40-199 bpm40-199 bpm
    Heart Rate Accuracy± 5%± 5%
    Electrical SafetyCompliance with IEC 60601-1: 2005+AMD1 (2012) +AMD2 (2020)Device complies
    EMCCompliance with IEC 60601-1-2: 2014/AMD1:2020Device complies
    BiocompatibilityCompliance with ISO 10993-1, ISO 10993-5, ISO 10993-10Device complies (Cytotoxicity, Skin Sensitization, Irritation)
    UsabilityCompliance with IEC 60601-1-6:2010+A1:2013+A2:2020Device complies
    Home Healthcare EnvironmentCompliance with IEC 60601-1-11:2015 +A1:2020Device complies

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 100 adult subjects were used for the clinical validation.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study. It also does not specify if the study was retrospective or prospective, but clinical validation studies for device clearance are typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document refers to a "clinical test" for validation in accordance with ISO 81060-2:2018. This standard typically requires concurrent or sequential measurements by a trained observer using a standard reference sphygmomanometer (e.g., mercury sphygmomanometer). However, the specific number of experts (e.g., clinicians, technicians) involved in establishing the ground truth measurements for the test set and their qualifications are not explicitly stated in the provided text.

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method used for the clinical test set. ISO 81060-2:2018 outlines specific protocols for obtaining reference blood pressure measurements, often involving multiple observers and an agreed-upon method for handling discrepancies, but the specific details for this study are not provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size of how much human readers improve with AI vs without AI assistance. The device in question is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The clinical validation, in accordance with ISO 81060-2:2018, evaluates the accuracy of the algorithm and device against a reference standard without human-in-the-loop assistance. The results state that "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018," indicating direct measurement of the device's accuracy.

    7. Type of Ground Truth Used

    The ground truth used for the clinical validation was established by concurrent or sequential measurements using a reference manometer (as per ISO 81060-2:2018 methodology). This is a gold standard for blood pressure validation studies, effectively functioning as an "expert consensus" on the true blood pressure readings, established by highly trained personnel following a strict protocol.

    8. Sample Size for the Training Set

    The document does not provide information regarding a "training set" or its sample size. This is typical for medical devices that are validated directly against established performance standards using a test set, rather than being developed using machine learning models that require distinct training and testing datasets. The device's underlying algorithm is likely based on established oscillometric principles and refined through engineering, not trained on a large dataset in the way an AI model would be.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, information on how its ground truth was established is not applicable or provided.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1