Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K232587

    Validate with FDA (Live)

    Date Cleared
    2024-04-12

    (231 days)

    Product Code
    Regulation Number
    862.1825
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGLUMI X3 Fully-auto chemiluminescence immunoassay analyzer is an automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens.

    The MAGLUMI 25-OH Vitamin D is an in vitro chemiluminescence immunoassay for the quantitative determination of 25-OH Vitamin D (25-OH VD) in human serum and plasma using the MAGLUMI series Fully-auto chemiluminescence immunoassay analyzer, and the assay is used for an aid in assessment of vitamin D sufficiency.

    Device Description

    MAGLUMI X3 Fully-auto chemiluminescence immunoassay analyzer:

    The MAGLUMI X3 Fully-auto chemiluminescence immunoassay analyzer is a fully automated instrument system designed to perform in vitro diagnostic tests on clinical specimens. The system utilizes chemiluminescent technology and uses pre-packaged reagent packs for qualitative or quantitative analysis of the analytes in human samples. The analyzer performs automatic sample pipetting, reagent loading, incubation, washing, measurements, and result calculations.

    MAGLUMI 25-OH Vitamin D assay:

    MAGLUMI 25-OH Vitamin D kit consists of the following reagents:

    Magnetic Microbeads- coated with anti-25-OH VD antibody in PBS buffer, NaN3 (<0.1%). Calibrator Low- A low concentration of 25-OH VD antigen in Carbonate buffer, NaN3 (<0.1%). Calibrator High- A high concentration of 25-OH VD antigen in Carbonate buffer, NaN3 (<0.1%). Buffer- Acidic buffer ABEI Label- ABEI labeled with anti-25-OH VD antibody (~0.500 ug/mL) in PBS buffer, NaN3 (<0.1%). Control 1- A low concentration of 25-OH VD antigen in Carbonate buffer, NaN3 (<0.1%). Control 2- A high concentration of 25-OH VD antigen in Carbonate buffer, NaN3 (<0.1%).

    AI/ML Overview

    The provided text describes the performance characteristics of the MAGLUMI 25-OH Vitamin D assay and MAGLUMI X3 Fully-auto chemiluminescence immunoassay analyzer as part of a 510(k) summary for FDA clearance. The information focuses on analytical performance rather than clinical validation with human-in-the-loop studies.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of predefined acceptance criteria. Instead, it reports the analytical performance of the device, which implicitly demonstrates that the device meets some internal or expected performance metrics for an in vitro diagnostic device. The performance data is presented in the "11. Performance Characteristics" section.

    Here's an approximation of an acceptance criteria table based on the reported performance, assuming the reported values are the achieved performance that meets internal criteria for release:

    Performance MetricAcceptance Criteria (Implicit from Reported Performance)Reported Device Performance (MAGLUMI 25-OH Vitamin D)
    PrecisionCalibrator Low (N=240)
    Repeatability (%CV)≤ 3.86%3.86%
    Within Instrument Total (%CV)≤ 7.53%7.53%
    Reproducibility (%CV)≤ 8.07%8.07%
    Calibrator High (N=240)
    Repeatability (%CV)≤ 1.58%1.58%
    Within Instrument Total (%CV)≤ 2.73%2.73%
    Reproducibility (%CV)≤ 2.76%2.76%
    (Similar detail for Controls and Serum Pools as reported in the text)(Specific values for each level and component)(Specific values for each level and component)
    LinearityCorrelation coefficient R² ≥ 0.9974 (for 1.8-195.0 ng/mL)R² = 0.9974
    Relationship: Observed ≈ 1.0016 (Expected) - 0.4938Observed = 1.0016 (Expected) - 0.4938
    Stability (Real-time)Stable for 18 months at 2-8°C18 months @ 2-8°C
    Detection Limit
    Limit of Blank (LOB)≤ 0.95 ng/mL0.95 ng/mL
    Limit of Detection (LOD)≤ 1.4 ng/mL1.4 ng/mL
    Limit of Quantitation (LOQ)≤ 2.289 ng/mL (CV ≤ 20%)2.289 ng/mL
    InterferenceNo significant interference (recovery ± 10% of initial value) at tested concentrations for:Achieved for all tested substances at specified concentrations.
    Cross-reactants(Specific % cross-reactivity for listed substances)Reported for 25-OH Vitamin D2, D3, etc.
    Endogenous substances(Specific highest concentrations for bilirubin, hemoglobin, etc.)Reported for bilirubin, hemoglobin, etc.
    Common drugs & substances(Specific highest concentrations for Cefoxitin, Biotin, etc.)Reported for Cefoxitin, Biotin, etc.
    HAMA, RF, Total protein(Specific highest concentrations for HAMA, RF, Total protein)Reported for HAMA, RF, Total protein
    Method ComparisonPassing-Bablok with predicate: Slope near 1, Intercept near 0, R near 1y=0.989x+0.249, R=0.997
    Matrix ComparisonPassing-Bablok for serum vs. plasma: Slope near 1, Intercept near 0, R² near 1y=0.977x-0.0256, R²=0.9938
    Reference RangeEstablished7.4 - 45.1 ng/mL

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Study:

      • Sample Size: 240 measurements per material (two controls, two calibrators, one spiked patient serum pool, three native patient sample pools) across three instruments. Each measurement was taken in duplicate, with 2 runs per day over 20 days.
      • Data Provenance: Not explicitly stated, but common practice for IVD analytical studies focuses on sample types and analytical conditions, not patient demographics or geo-location beyond what's relevant to endogenous interferents. The samples include "patient serum pool" and "native serum pool," implying human biological samples.
    • Linearity Study:

      • Sample Size: Eleven linearity samples, each measured in quadruplicate on 3 lots of reagent.
      • Data Provenance: Samples prepared by blending "a low serum sample pool and a high serum sample pool," implying human serum. No country of origin specified.
    • Detection Limit Study:

      • LOB: 60 measurements of 25-OH Vitamin D depleted serum samples using 3 different lots of reagents over 3 days.
      • LOD: Four levels of low samples, each measured in 60 replicates over 3 days per sample using 3 lots of reagents.
      • LOQ: Six low serum samples, in five replicates per run, one run per day, over 3 days, using 3 lots of reagents.
      • Data Provenance: "Serum samples," "low serum samples," implying human serum. No country of origin specified.
    • Interference Study:

      • Sample Size: Three base serum samples (10, 50, 100 ng/mL 25-OH VD) for endogenous substances and common drugs; human serum samples for HAMA, RF, total protein.
      • Data Provenance: "Human serum pools" and "human serum samples." No country of origin specified.
    • Method Comparison Study:

      • Sample Size: 329 native, single donor patient serum samples (121 male, 208 female, age 7 to 98 years).
      • Data Provenance: "Human serum samples." No country of origin specified explicitly, but this is typically retrospective collection of de-identified clinical samples.
    • Reference Range Study:

      • Sample Size: 312 serum samples from normal, apparently healthy adult (21 years and older) individuals (181 males, 131 females).
      • Data Provenance: Samples collected from three regions (Central, Southeast and Northeast) in the US. This specifies the country of origin (USA) and regional distribution. This would be considered retrospective.
    • Matrix Comparison Study:

      • Sample Size: 78 serum/plasma pairs from the same donor.
      • Data Provenance: "Samples drawn into serum and plasma collection tubes," "from the same donor." No country of origin specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This document describes an In Vitro Diagnostic (IVD) device, specifically a quantitative immunoassay. For such devices, "ground truth" is typically established by:

    • Reference methods (e.g., LC-MS/MS for Vitamin D, which is considered a gold standard).
    • Master Lot/Calibrator values.
    • Comparison to a legally marketed predicate device (as done in the "Comparison Studies" section).

    There is no mention of human experts (like radiologists) establishing ground truth for individual samples in this context. The "ground truth" in this analytical study is the quantitative concentration of 25-OH Vitamin D as determined by reference materials, calibrators, or the established predicate method.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving qualitative or semi-quantitative assessments (e.g., image-based diagnosis) where multiple human readers interpret data, and discrepancies need to be resolved. This document pertains to the analytical performance of a quantitative immunoassay analyzer, where the output is a numerical concentration.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated immunoassay analyzer for a quantitative biomarker (25-OH Vitamin D). It is not an AI-assisted diagnostic tool that aids human readers in interpreting complex data like medical images. Therefore, an MRMC study or AI assistance is not relevant to its stated purpose or performance evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is inherently a "standalone" system in an analytical sense. The MAGLUMI X3 analyzer and the MAGLUMI 25-OH Vitamin D assay perform the quantitative determination automatically. The performance characteristics (precision, linearity, detection limit, interference, method comparison) are all tests of the algorithm's performance (i.e., the instrument and reagent system) without human interpretation affecting the result generation process itself. Clinical interpretation of the numerical results (e.g., patient has vitamin D deficiency based on the number) happens downstream by a clinician, but the device performance itself is standalone.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the analytical performance:

    • Precision, Linearity, Detection Limit, Interference: Ground truth is established by the known concentrations of controls, calibrators, spiked samples, and highly characterized depleted/low concentration samples. The "truth" is the intended/expected concentration as determined by a highly accurate measurement or formulation.
    • Method Comparison: The predicate device, MAGLUMI 2000 25-OH Vitamin D assay manufactured by SNIBE, served as the comparative "truth" or reference. The study measures agreement between the new device and the predicate. It states "Comparison of the MAGLUMI 25-OH Vitamin D assay (y) with the predicate device, MAGLUMI 2000 25-OH Vitamin D assay (x)."
    • Reference Range: Established from empirically tested healthy individuals.

    8. The sample size for the training set

    This document describes the validation of an immunoassay kit and analyzer, not a machine learning or AI model that requires a "training set" in the computational sense. The "training" for such a system refers to the development and optimization of the chemical reagents, assay protocol, and instrument calibration algorithms by the manufacturer during product development, prior to the validation studies described here. The document does not provide details on the sample sizes or data used during this internal "training" or development phase.

    9. How the ground truth for the training set was established

    As noted above, "training set" is not applicable in the context of a traditional immunoassay system validation as described here. The "ground truth" for the development and optimization of the assay would typically involve:

    • Certified Reference Materials (CRMs): Substances with accurately known concentrations of the analyte, often traceable to international standards.
    • Internal Reference Materials: Carefully prepared and validated in-house samples.
    • Established Reference Methods: Highly accurate and precise methods (e.g., LC-MS/MS) used to characterize samples during development.
    • Cross-validation with existing, well-characterized methods/platforms.

    These elements guide the chemical formulation, antibody selection, and instrument parameter setting during the development phase.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192547

    Validate with FDA (Live)

    Date Cleared
    2020-01-17

    (122 days)

    Product Code
    Regulation Number
    862.1155
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MAGLUMI 2000 HCG/S-HCG is an in vitro chemiluminescence immunoassay for the quantitative determination of total beta human chorionic gonadotropin (total ß-hCG) in human serum. The measurement of total ß-hCG is used as an aid in the early detection of pregnancy.

    Device Description

    MAGLUMI 2000 HCG/B-HCG kit consists of the following reagents: Magnetic Microbeads- coated with anti-HCG monoclonal antibody, containing BSA, NaN3 (<0.1%) Calibrator Low-Containing BSA and HCG antigen, NaN3(<0.1%) Calibrator High- Containing BSA and HCG antigen, NaN3(<0.1%) Buffer- containing BSA, NaN3 (<0.1%) ABEI Label- Anti-HCG monoclonal antibody labeled with ABEI, containing BSA, NaN3 (<0.1%) Control 1- Containing BSA and HCG antigen, NaN3 (<0.1%) Control 2- Containing BSA and HCG antigen, NaN3 (<0.1%)

    AI/ML Overview

    The provided text describes the performance characteristics of the MAGLUMI 2000 HCG/ß-HCG device, an in vitro chemiluminescence immunoassay for the quantitative determination of total beta human chorionic gonadotropin (total ß-hCG) in human serum. This information is presented as part of a 510(k) summary submitted to the FDA. While the document outlines various analytical performance studies, it does not explicitly define acceptance criteria in a quantitative table or refer to a multi-reader multi-case (MRMC) comparative effectiveness study.

    However, we can infer acceptance based on the reported performance results aligning with standard clinical laboratory expectations and the successful substantial equivalence determination by the FDA. The study focuses on analytical performance characteristics rather than diagnostic accuracy involving human interpretation of results.

    Here's an attempt to structure the information based on your request, identifying what is and isn't available in the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria. Instead, it describes performance characteristics and indicates compliance with CLSI guidelines. We can infer the "acceptance" derived from the presented data and the overall conclusion of substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Inferred from CLSI/Clinical Practice)Reported Device Performance
    PrecisionWithin-Run CV%: Generally < 10-15% for immunoassay, lower for higher concentrations.Control 1 (5.023 mIU/mL): SD=0.226, CV%=4.50%
    Control 2 (37.135 mIU/mL): SD=1.529, CV%=4.12%
    Control 3 (300.135 mIU/mL): SD=10.737, CV%=3.58%
    Calibrator Low (2.876 mIU/mL): SD=0.199, CV%=6.92%
    Calibrator High (1722.515 mIU/mL): SD=47.252, CV%=2.74%
    Total (within instrument) CV%: Generally < 15-20%, lower for higher concentrations.Control 1: SD=0.321, CV%=6.39%
    Control 2: SD=2.155, CV%=5.80%
    Control 3: SD=16.757, CV%=5.58%
    Calibrator Low: SD=0.257, CV%=8.94%
    Calibrator High: SD=83.853, CV%=4.87%
    Reproducibility (across instruments) CV%: Similar to Total CV%, reflecting inter-instrument consistency.Control 1: SD=0.326, CV%=6.49%
    Control 2: SD=2.250, CV%=6.06%
    Control 3: SD=17.665, CV%=5.89%
    Calibrator Low: SD=0.277, CV%=9.65%
    Calibrator High: SD=92.799, CV%=5.39%
    LinearityR² > 0.99 for quantitative assays over measurement range.R² = 0.9932 (between 1.134 and 4680 mIU/mL)
    Detection Limit (LOB)Determined by 95th percentile, should be clinically appropriate.0.302 mIU/mL (highest of 3 lots)
    Detection Limit (LOD)Clinically appropriate for early detection of pregnancy.0.471 mIU/mL (highest of 3 lots)
    Limit of Quantitation (LOQ)CV% no more than 20%, Bias no more than 15%.1.134 mIU/mL (highest of 3 lots)
    InterferenceRecovery ±10% of initial value for various interferents.No significant interference observed at tested concentrations (details in text for TSH, LH, FSH, hGH, hCG α-subunit, bilirubin, hemoglobin, triglyceride, common drugs, HAMA, RF, total protein).
    Hook EffectNo hook effect within clinically relevant high concentrations.No HOOK effect observed within 1,000,000 mIU/mL.
    Dilution RecoveryPercent differences for diluted specimen versus expected concentration within 10%.Percent differences for diluted specimen versus expected concentration were within 10%.
    Method ComparisonStrong correlation (high R²) and acceptable bias (slope near 1, intercept near 0) when compared to predicate device.Y = 0.988X + 1.995, R² = 0.993 (All three sites)
    StabilityReagents/controls stable for specified period at specified conditions.Accelerated stability showed 12 months at 2-8°C. Real-time stability is ongoing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Study: 80 samples analyzed per level (Control 1, Control 2, Control 3, Calibrator Low, Calibrator High, and 6 Native Serum Pools) on each of three instruments. Total N=240 samples per level across three instruments.
    • Linearity Study: Linearity samples prepared by mixing high and low level samples. Each sample was measured in quadruple on 3 lots of reagent. The exact number of distinct linearity samples (different concentrations) is not specified, but the range is 0.3 to 4680 mIU/mL.
    • Detection Limit Studies:
      • LOB: 80 measurements of HCG/ß-HCG negative serum samples using 3 different lots over 5 days.
      • LOD: Four levels of low samples, measured in 80 replicates over 5 days per sample using 3 lots of reagents.
      • LOQ: Six low serum samples, in six replicates per run, one run per day, over 5 days, using 3 lots of reagents.
    • Interference Study: Two base serum samples (6.0 mIU/mL and 100 mIU/mL HCG/ß-HCG) spiked with various cross-reactants. Also, human serum pools with 6.0 mIU/mL, 100 mIU/mL, and 2000 mIU/mL HCG/ß-HCG for endogenous substances and common drugs. Each tested using 3 lots of reagents. Exact number of distinct interferent samples not specified.
    • Hook Effect: Six samples with HCG/ß-HCG concentrations from 5000 to 1,000,000 mIU/mL prepared by spiking. Serial dilutions tested using 3 different lots.
    • On-board Dilution Recovery: Twelve serum samples with HCG/ß-HCG concentrations from 4475 to 223750 mIU/mL tested using three reagent lots and three instruments.
    • Method Comparison Study: 201 human serum samples with concentrations ranging from 1.1 to 4934 mIU/mL (as determined by the predicate device).
    • Expected Values/Reference Range: 431 serum samples from non-pregnant, apparently healthy females (20 years and older).

    Data Provenance: The document does not explicitly state the country of origin for the data. Given the submitter's location (Shenzhen, China) and the FDA submission, it's likely the studies were conducted in China or involved samples from that region, but this is not explicitly confirmed. The studies are described in a manner typical of prospective performance evaluation studies for an in vitro diagnostic device, rather than retrospective analysis of pre-existing data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The device is an in vitro diagnostic immunoassay for quantitative measurement of a biomarker (hCG). Ground truth is established by the reference measurement procedure of the MAGLUMI 2000 instrument itself and validated through analytical performance studies (precision, linearity, detection limits, method comparison to a legally marketed predicate device, etc.) rather than through expert human interpretation of images or clinical findings.

    4. Adjudication Method for the Test Set

    Not applicable. As the device is an in vitro diagnostic assay providing a quantitative numerical result, there is no human interpretation or subjective assessment that would require an adjudication method. The "ground truth" for method comparison is the value obtained from the predicate device. For analytical performance studies, it's the objectively measured values and their statistical distributions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC study is typically conducted for image-based diagnostic aids (e.g., AI for radiology) where human readers interpret medical images. This device is a lab-based immunoassay that provides a quantitative numerical result. Therefore, an MRMC study is not relevant or applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the entire study is essentially a standalone performance evaluation. The MAGLUMI 2000 HCG/ß-HCG is an automated chemiluminescence immunoassay system. Its performance characteristics (precision, linearity, detection limits, interference, hook effect, dilution recovery, and method comparison) are evaluated as the direct output of the instrument and its reagents, without explicit human interpretation being part of the device's function or the primary subject of these studies. The quantitative output of the device itself constitutes its "performance."

    7. The Type of Ground Truth Used

    The ground truth for this device's performance evaluation is established through:

    • Reference Measurement Procedures/Known Concentrations: For studies like linearity, detection limits, and interference, known concentrations or "spiked" samples are used as a reference.
    • Measurement against a Predicate Device: For the method comparison study, the results from the Beckman Access Total B-HCG (5th IS) Assay (the predicate device) served as the comparative "ground truth" to establish substantial equivalence.
    • Statistical Analysis of Replicate Measurements: For precision, the statistical variation around the mean measured value for controls and patient samples provides the "ground truth" of the device's reproducibility.
    • Physiological/Clinical Samples: For determining expected values/reference ranges, samples from apparently healthy individuals are used.

    8. The Sample Size for the Training Set

    The document does not explicitly describe a "training set" in the context of an algorithm or machine learning model. This is an in vitro diagnostic assay, not an AI/ML-based device that typically undergoes a distinct training/validation/test split of data. The studies described are analytical verification and validation studies in a traditional medical device development sense.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no defined "training set" for an AI/ML algorithm, this question is not relevant to the described device and its evaluation. The "training" of such a system would involve the manufacturer's internal assay development and optimization, which isn't part of a regulatory submission summary like this.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191499

    Validate with FDA (Live)

    Date Cleared
    2019-08-01

    (56 days)

    Product Code
    Regulation Number
    862.1825
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MAGLUMI 2000 25-OH Vitamin D is an in vitro chemiluminescence immunoassay for the quantitative determination of 25-OH Vitamin D in human serum using Maglumi 2000 Fully-auto chemiluminescence immunoassay analyzer. The measurement of 25-OH Vitamin D is to be used as an aid in the assessment of vitamin D sufficiency.

    Device Description

    MAGLUMI 2000 25-OH VITAMIN D kit consists of the following reagents: Magnetic Microbeads- coated with 25-OH Vitamin D monoclonal antibody, containing BSA, NaN3 (<0.1%) Calibrator Low-Containing BSA and 25-OH Vitamin D antigen, NaN3(<0.1%) Calibrator High- Containing BSA and 25-0H Vitamin D antigen, NaN3(<0.1%) Displacing Reagent- Acidic buffer ABEI Label- 25-OH Vitamin D antigen labeled with ABEI Control 1- Containing BSA and 25-OH Vitamin D antigen, NaN3 (<0.1%) Control 2- Containing BSA and 25-OH Vitamin D antigen, NaN3 (<0.1%)

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance & Compliance
    PrecisionCV% within acceptable limits for various levels (Controls, Calibrators, Serum Pools) across repeatability, within-run, between-day, and total reproducibility.All reported CV% values are within typical acceptable ranges for clinical assays, indicating good precision. Example: Control 1 Total CV of 6.33% and Reproducibility CV of 6.46%.
    LinearityDemonstrate linearity across the claimed measuring range with a strong correlation coefficient (R²).Linear between 4.6 and 145.8 ng/mL with R² = 0.9986. (Meets)
    StabilityReagents, calibrators, and controls maintain stability over a reasonable period (e.g., 12 months at 2-8°C).Accelerated studies show stability for 12 months at 2-8°C for controls, calibrators, and reagents. Real-time stability is ongoing. (Meets based on accelerated data)
    Detection Limit (LOB)Limit of blank should be low, indicating the ability to differentiate from zero.LOB = 1.990 ng/mL. (Meets)
    Detection Limit (LOD)Limit of detection should be low, indicating the lowest concentration at which analyte can be detected.LOD = 3.8 ng/mL. (Meets)
    Limit of Quantitation (LOQ)LOQ should be the lowest concentration reproducibly measurable with an intermediate precision CV of ≤ 20%.LOQ = 5.371 ng/mL with an intermediate precision CV of ≤ 20%. (Meets)
    Interference (Cross-reactivity)Minimal cross-reactivity with structurally similar compounds.High cross-reactivity with 25-OH Vitamin D2 (98.10%) and D3 (96.13%), which is expected as the assay measures total 25-OH Vitamin D. Low cross-reactivity with other D metabolites (e.g., Vitamin D2, Vitamin D3, 1,25-(OH)2-Vitamin D3). (Meets, as intended for total 25-OH Vitamin D)
    Interference (Endogenous Substances)No significant interference from common endogenous substances (bilirubin, hemoglobin, triglycerides, etc.) at high concentrations.No significant interference observed (defined as recovery ± 10% of initial value) at high concentrations listed for each substance. (Meets)
    Interference (Common Drugs/Substances)No significant interference from common drugs and other substances (e.g., Ascorbic Acid, Acetaminophen, Biotin) at typical therapeutic/high concentrations.No significant interference observed (defined as recovery ± 10% of initial value) at high concentrations listed for each substance. (Meets)
    Interference (HAMA, RF, Total Protein)No significant interference from HAMA, RF, and total protein at high concentrations.No significant interference observed (defined as recovery ± 10% of initial value) at high concentrations listed for each substance. (Meets)
    Method ComparisonStrong correlation and agreement with a legally marketed predicate device.Regression equation: Y = 1.013X - 0.504, R² = 0.9739, indicating strong correlation and agreement with the predicate. (Meets)

    Study Details:

    The provided document describes a series of analytical performance studies and a method comparison study to demonstrate the performance characteristics of the MAGLUMI 2000 25-OH Vitamin D assay.

    2. Sample Sizes and Data Provenance

    • Test Set (for specific performance characteristics):

      • Precision: 240 samples per level across 9 sample types (3 controls, 2 calibrators, 3 spiked serum pools, 3 native serum pools). Total N = 240 * 9 = 2160 individual measurements (though some are duplicates/runs, the total 'data points' are numerous).
      • Linearity: 11 levels of linearity samples, each measured in quadruplicate, on 3 lots of reagent.
      • Detection Limit:
        • LOB: 60 measurements of 25-OH VITAMIN D depleted serum samples using 3 different lots of reagents over 5 days.
        • LOD: 4 levels of low samples measured in 60 replicates over 5 days per sample using 3 lots of reagents.
        • LOQ: Six low serum samples, in six replicates per run, one run per day, over 5 days, using 3 lots of reagents.
      • Interference (Cross-reactivity): Used two base serum samples (30 ng/mL and 60 ng/mL total 25-OH VD) spiked with various cross-reactants, measured using 3 lots of reagents.
      • Interference (Endogenous Substances & Common Drugs): Three serum samples (20, 30, and 60 ng/mL 25-OH VITAMIN D) analyzed for each substance.
      • Interference (HAMA, RF, Total Protein): Human serum samples supplemented with potential interferents, tested using 3 lots of reagents.
      • Method Comparison: 241 human serum samples.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical samples. It uses "human serum samples" and "patient serum pools." The studies appear to be retrospective in nature, using collected serum samples.

    3. Number of Experts and Qualifications for Ground Truth

    • This device is an in vitro diagnostic immunoassay. The concept of "ground truth" established by experts (like radiologists for imaging) is not applicable in the same way. The ground truth for such assays is typically established by:

      • Known concentrations for controls and calibrators, often traceable to reference materials (e.g., NIST RMP 2972 for traceability as mentioned).
      • Spiking studies where a known amount of analyte is added to a sample.
      • Comparison to a legally marketed predicate device, which itself has an established ground truth.
      • Pathology/biochemical analysis where the exact concentration of the analyte is determined by highly accurate reference methods (e.g., ID-LC-MS/MS).

      Therefore, there were no human experts establishing the ground truth in the context of clinical interpretation, but rather a robust analytical process to define true concentrations.

    4. Adjudication Method for the Test Set

    • Not applicable for this type of in vitro diagnostic assay. Adjudication methods like 2+1 or 3+1 are typically used in clinical trial settings where human interpretation or consensus for a diagnosis is required.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices (often imaging AI) where multiple human readers interpret cases and their performance is compared with and without AI assistance. The MAGLUMI 2000 25-OH Vitamin D is an automated immunoassay, not a device requiring human interpretation in this manner.

    6. Standalone (Algorithm Only) Performance

    • Yes, a standalone performance was done. The entire analytical performance section (Precision, Linearity, Stability, Detection Limit, Interference) directly assesses the algorithm's (the immunoassay's) performance without human intervention in the measurement process. The "MAGLUMI 2000 Fully-auto chemiluminescence immunoassay analyzer" is an automated system, meaning the results are generated directly by the device. The method comparison study is also a standalone assessment of the device's output against a predicate.

    7. Type of Ground Truth Used

    • The ground truth for the analytical validation aspects (e.g., calibrators, controls, linearity samples) is based on pre-defined concentrations, often traceable to reference methods like ID-LC-MS/MS (Isotope Dilution Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) and reference materials such as NIST RMP 2972.
    • For interference studies, the "ground truth" is inferred by the known addition of interferents and measuring the deviation from the expected value.

    8. Sample Size for the Training Set

    • The document does not mention a training set in the context of machine learning or AI. This device is an immunoassay, which functions based on established biochemical principles and reagents, not on a machine learning model that requires a separate training set. The "development" or "optimization" of the assay would involve various experiments, but not a formally defined "training set" like in AI/ML validation.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no mention of a training set for an AI/ML model for this immunoassay.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182423

    Validate with FDA (Live)

    Device Name
    MAGLUMI 2000 FT4
    Date Cleared
    2018-10-04

    (28 days)

    Product Code
    Regulation Number
    862.1695
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGLUMI 2000 FT4 assay is for in vitro diagnostic use in the quantitative determination of free thyroxine (FT4) in human serum. The measurement of FT4 is used in the diagnosis of thyroid disorders.

    Device Description

    MAGLUMI 2000 FT4 kit consists of the following reagents: Magnetic Microbeads- coated with T4 antigen, BSA, NaN3(<0.1%) Calibrator Low-Containing BSA and T4 antigen, NaN3 (<0.1%) Calibrator High- Containing BSA and T4 antigen, NaN3 (<0.1%) Buffer- Containing BSA and NaN3 (<0.1%) ABEI Label- Anti-T4 monoclonal antibody labeled with ABEI, containing BSA, NaN3(<0.1%) Control 1- Containing BSA and T4 antigen, NaN3 (<0.1%) Control 2- Containing BSA and T4 antigen, NaN3 (<0.1%)

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study findings for the MAGLUMI 2000 FT4 device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    PrecisionCV% within acceptable ranges for clinical chemistry devices at various concentrations (implied by CLSI EP5-A2). For example, Total CV for Control 1: ≤ 10% (often a general guideline for low-concentration analytes). Total CV for other controls/samples: ≤ 8-10% depending on concentration. Specific pre-defined limits are not explicitly stated, but the study aims to demonstrate acceptable precision.Control 1 (1.006 ng/dL): Total CV 9.34%Control 2 (1.9952 ng/dL): Total CV 7.81%Control 3 (3.9978 ng/dL): Total CV 5.04%Calibrator low (0.3333 ng/dL): Total CV 9.99%Calibrator high (7.1929 ng/dL): Total CV 3.20%Serum Pools/Patient Pools: Total CVs ranged from 3.76% to 9.96%
    LinearityThe assay should demonstrate linearity across its claimed measuring range (0.19 to 10.0 ng/dL) with a strong correlation (R² close to 1) and a slope close to 1, and y-intercept close to 0 (implied by CLSI EP6-A).Linear between 0.19 and 10.0 ng/dL. Observed = 0.9898 (Expected) + 0.01364, R² = 0.9991.
    Stability (Reagents)Reagents, calibrators, and controls should be stable for a specified shelf life at recommended storage conditions (e.g., 2-8°C for 12 months for accelerated stability, confirmed by real-time studies).Accelerated stability at 37°C showed controls, calibrators, and reagents are stable for 12 months at 2-8°C. Real-time stability is on-going.
    Detection Limit (Limit of Blank - LOB)95th percentile value from analyte-free samples (implied by CLSI EP17-A guidelines). A low LOB is desired.0.087 ng/dL (highest of 3 lots).
    Detection Limit (Limit of Detection - LOD)Lowest analyte concentration that can be detected (implied by CLSI EP17-A guidelines). A low LOD is desired.0.143 ng/dL (highest of 3 lots).
    Detection Limit (Limit of Quantitation - LOQ)Lowest analyte concentration that can be reproducibly measured with an intermediate precision CV of ≤ 20% (as per CLSI EP17-A guidelines).0.190 ng/dL (highest of 3 lots).
    Interference (Cross-Reactivity)% Cross-reactivity for relevant compounds should be very low (e.g., <0.02% as shown for the specified substances). No significant interference (recovery ± 10% of initial value) from common drugs and endogenous substances at specified concentrations.All tested cross reactants (Monoiodotyrosine, L-Triiodothyronine, Diiodotyrosine, 3,5-Diiodo-L-thyronine, Reverse Triiodothyronine) showed <0.02% cross-reactivity. No significant interference (recovery ± 10%) from listed common drugs (e.g., Bilirubin, Hemoglobin, Triglycerides, Cefoxitin) and endogenous interferents (HAMA, RF, Total protein) at specified concentrations.
    Method ComparisonStrong correlation and good agreement between the candidate device and the predicate device, demonstrated by a linear regression equation with an R² close to 1 (e.g., >0.95 or >0.98), a slope close to 1, and a y-intercept close to 0 (implied by CLSI EP9-A2).Y = 1.0451X - 0.05375, R² = 0.9919, for MAGLUMI FT4 (y) vs. ADVIA CENTAUR FT4 (x).

    Note on Acceptance Criteria: The document often refers to CLSI guidelines (e.g., EP5-A2, EP6-A, EP17-A, EP9-A2) for performance characteristic studies. While explicit numerical acceptance criteria are not always stated directly in the text, compliance with these guidelines implicitly means that the performance falls within industry-accepted ranges and statistical thresholds. For precision, for instance, a total CV of less than 10% (or even lower for higher concentrations) is generally considered acceptable in clinical chemistry. For linearity and method comparison, an R² value close to 1, a slope near 1, and a y-intercept near 0 are standard indicators of good performance.


    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Study:
      • Sample Size: 240 measurements per control/calibrator/pool (N=240, from 80 samples analyzed per level on each of 3 instruments).
      • Data Provenance: Not explicitly stated, but implied to be patient serum pools and native patient samples from clinical settings, likely domestic to the manufacturer's testing location or within industry standard practices for clinical evaluation. The study was conducted on "three controls, two calibrators, four spiked patient serum pools and four native patient sample pools."
    • Linearity Study:
      • Sample Size: 11 levels with FT4 concentrations from 0.19 to 10.0 ng/dL, each measured in quadruplicate on 3 lots of reagent.
      • Data Provenance: Samples prepared by spiking T4 USP Standard into T4-free human serum samples.
    • Detection Limit Study (LOB, LOD, LOQ):
      • LOB: 80 measurements of T4 free human serum samples.
      • LOD: Four levels of low samples, measured in 80 replicates over 5 days per sample.
      • LOQ: Six low serum samples, in six replicates per run, one run per day, over 5 days.
      • Data Provenance: T4 free human serum samples and low serum samples.
    • Interference Study:
      • Sample Size: Variable, involves preparing solutions, and for common drugs/interfering substances, three serum samples (low, medium, high FT4) per substance. For HAMA/RF/Protein, specific FT4 human serum samples.
      • Data Provenance: Spiked solutions, human serum pools, FT4 human serum samples.
    • Method Comparison Study:
      • Sample Size: 224 human serum samples.
      • Data Provenance: Human serum samples. Provenance (e.g., country of origin, retrospective/prospective) is not explicitly stated.
    • Expected Values/Reference Range Study:
      • Sample Size: 157 serum samples.
      • Data Provenance: Normal, apparently healthy adult individuals (22 years and older). Provenance (e.g., country of origin, retrospective/prospective) is not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. For an in vitro diagnostic device like the MAGLUMI 2000 FT4, the "ground truth" for analytical performance tests (precision, linearity, detection limits, interference) is typically established by the reference methods, spiked concentrations, or consensus values derived from the testing procedures themselves, rather than human expert opinion. For method comparison, the "ground truth" is typically the result from the established predicate device.


    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1 for clinical diagnoses based on expert review) is not applicable to the analytical performance studies described for this in vitro diagnostic device. The studies rely on quantitative measurements and statistical analysis rather than subjective expert interpretation of results for ground truth.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for medical imaging devices or other diagnostic tools where human readers interpret results, and the AI system acts as an aid to improve human performance. For an automated in vitro diagnostic assay like MAGLUMI 2000 FT4, the performance is evaluated through direct analytical comparisons.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are for the standalone performance of the MAGLUMI 2000 FT4 device. These are analytical performance studies (precision, linearity, detection limits, interference, method comparison) that evaluate the device's ability to accurately and precisely measure FT4 concentrations in human serum, independent of human interpretation or intervention in the measurement process itself. The "algorithm" here refers to the immunoassay's chemistries and the instrument's measurement principles.


    7. The Type of Ground Truth Used

    • Precision: Reference materials (controls, calibrators) with established target values, and pooled human serum samples where the mean serves as a statistical "ground truth" for variability assessment.
    • Linearity: Gravimetrically or volumetrically prepared "expected" concentrations of T4 in serum.
    • Stability: Initial measurements at time 0, with subsequent measurements compared to these baselines.
    • Detection Limits (LOB, LOD, LOQ): Analyte-free samples and low-concentration samples measured extensively to statistically determine minimum detectable/quantifiable levels.
    • Interference (Cross-Reactivity): Known amounts of potential interferents/cross-reactants.
    • Method Comparison: Results from the predicate device (Siemens ADVIA Centaur FT4) are used as the reference "ground truth" for comparison.
    • Expected Values/Reference Range: Statistical distribution of measurements from a reference population (157 apparently healthy adults) to establish a normal range.

    8. The Sample Size for the Training Set

    This information is not provided. The document describes performance characteristic studies (test set), but does not detail the development or "training" specific to an AI algorithm for this device. For in vitro diagnostics, "training set" would typically refer to samples used during the assay development and optimization phase to establish parameters like calibration curves, reagent formulations, or instrument settings. The document focuses on validation studies.


    9. How the Ground Truth for the Training Set Was Established

    This information is not provided as a "training set" isn't explicitly defined in the context of this submission. If referring to the development/optimization phase of a typical immunoassay, ground truth would be established through:

    • Reference materials: Highly purified FT4 standards.
    • Known concentrations: Samples prepared with precise, known concentrations of FT4.
    • Comparative data: Initial comparisons against existing reference methods or well-characterized assays during the research and development phase.
    • Clinical samples: Use of a large number of clinical samples to optimize the assay's dynamic range and specificity.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1