510(k) Data Aggregation

    K Number
    K251331

    Validate with FDA (Live)

    Date Cleared
    2025-10-27

    (180 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate by using the arm cuff. The device can be used in medical facilities or at home, and only for indoor use. It is supplied for OTC use.

    Device Description

    The blood pressure monitor is a fully automatic, non-invasive upper arm measurement device using oscillometric methodology to measure systolic pressure, diastolic pressure and pulse rate. The device features an inflatable cuff that wraps around the arm, with a built-in pressure sensor and transducer that analyze arterial pulsations to determine blood pressure values. Measurement results are clearly displayed on the LCD screen.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233823

    Validate with FDA (Live)

    Date Cleared
    2024-06-28

    (210 days)

    Product Code
    Regulation Number
    884.2660
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultrasonic Fetal Doppler (Model M520, M521, M523) is intended to detect and display fetal heart rate, and used by health care professionals in a hospital, clinic, community, and home setting for singleton pregnancies after 16 weeks gestation.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for an "Ultrasonic Fetal Doppler." This document primarily focuses on regulatory approval and does not contain specific information about the acceptance criteria, study details, or performance metrics of the device as typically found in clinical validation studies.

    Therefore, I cannot extract the requested information regarding:

    • A table of acceptance criteria and the reported device performance
    • Sample size used for the test set and data provenance
    • Number of experts used to establish ground truth and their qualifications
    • Adjudication method
    • Multi-reader multi-case (MRMC) comparative effectiveness study results or effect size
    • Standalone performance details
    • Type of ground truth used
    • Sample size for the training set
    • How the ground truth for the training set was established

    The document states the device's intended use: "to detect and display fetal heart rate, and used by health care professionals in a hospital, clinic, community, and home setting for singleton pregnancies after 16 weeks gestation." However, it does not provide the technical validation data that would be used to establish equivalence and meet performance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233146

    Validate with FDA (Live)

    Date Cleared
    2024-03-19

    (174 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the of adult by using the arm cuff (22-42cm), it can be used in medical facilities or at home. It is supplied for OTC use.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a Blood Pressure Monitor. It does NOT contain information about an AI/ML medical device, nor does it describe an acceptance criteria study with the requested details.

    Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth, or training set information.

    The provided text focuses on the regulatory clearance process for a standard medical device (blood pressure monitor), detailing regulations, marketing provisions, and administrative contact information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230409

    Validate with FDA (Live)

    Date Cleared
    2023-08-25

    (191 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult by using the wrist circumference 12.5-21.5cm, it can be used in medical facilities or at home. It is supplied for OTC use.

    Device Description

    The subject device, Wrist Type Blood Pressure Monitor, is a battery driven automatic noninvasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure and pulse rate of adults at wrist within its claimed range and accuracy via the oscillometric technique. The device has data storage function for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The subject device is intended to be used in medical facilities or at home. And it is provided non-sterile, and not to be sterilized by the user prior to use.

    The proposed blood pressure monitor includes two models, which are W05, W1101L. All models contain the same software, measurement principle and NIBP algorithm. The main differences are product appearance, specification of pump and specification of the battery. W05 is alkaline battery driven device and W1101L is lithium battery driven device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Shenzhen Jamr Technology Co., Ltd. Wrist Type Blood Pressure Monitor (W05, W1101L):

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Standard ISO 81060-2:2018)Reported Device Performance (Clinical Test Summary)
    Mean error for blood pressure measurementsFulfilled required limits
    Standard deviation for blood pressure measurementsFulfilled required limits
    Pressure Accuracy: ± 3 mmHgPressure Accuracy: ± 3 mmHg (Same)
    Pulse Rate Accuracy: ± 5 % BPMPulse Rate Accuracy: ± 5 % BPM (Same)

    2. Sample Size and Data Provenance for Test Set:

    • Sample Size: 95 patients.
    • Data Provenance: The document does not explicitly state the country of origin. The study was a clinical investigation performed to verify accuracy according to ISO 81060-2:2018, which is a standard for clinical investigation of intermittent automated measurement type for non-invasive sphygmomanometers. It describes patients with "adequately accounting for demographic variables, such as age, gender, limb size, blood pressure, population characteristics and measurement site." It is a prospective clinical test, as new measurements were taken for the study.

    3. Number of Experts and Qualifications for Ground Truth:

    • The document implies that the ground truth was established by the standard auscultation method, which is performed by trained medical professionals. However, the exact number of experts and their specific qualifications (e.g., number of years of experience) used to establish the ground truth for this particular test set are not explicitly stated in the provided text.

    4. Adjudication Method for the Test Set:

    • The document states, "Standard auscultation method was used as the reference blood pressure monitor measuring in the arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2 Third edition 2018-11." This indicates a direct comparison approach, but no specific adjudication method (like 2+1 or 3+1 for resolving discrepancies) is mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for interpretation by humans.

    6. Standalone Performance Study:

    • Yes, a standalone performance study was done. The clinical test summarized in the document directly evaluates the device's accuracy in measuring blood pressure and pulse rate against a reference standard (auscultation) without human-in-the-loop assistance. The device's measurement principle is described as "oscillometric technique," which is an automated process.

    7. Type of Ground Truth Used:

    • Expert consensus (indirectly through a standard method): The ground truth was established using the standard auscultation method. This method relies on the auditory interpretation of Korotkoff sounds by trained human observers, which can be considered an expert-derived ground truth.

    8. Sample Size for the Training Set:

    • The document does not provide information about a separate training set or its sample size. Blood pressure monitors typically rely on established algorithms for oscillometric measurements, rather than machine learning models that require a distinct training set in the same way an AI diagnostic tool would.

    9. How the Ground Truth for the Training Set Was Established:

    • As a training set is not explicitly mentioned or implied for this type of device (a traditional blood pressure monitor using an oscillometric algorithm), the document does not explain how ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200437

    Validate with FDA (Live)

    Date Cleared
    2020-06-25

    (122 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Pressure Monitor is intended to measure the systolic blood pressure as well as the pulse rate by using the oscillometric method during inflation. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults at medical facilities or at home. The intended upper arm circumference is 22-40cm (8.7" - 15.7"). Not suitable for neonatal and infants.

    Device Description

    The blood pressure monitor model C02 and F01 are sphygmomanometers with electronic manometer intended to be used for the indirect (non-invasive) measurement of diastolic, systolic blood pressure and pulse rate using a standard oscillometric method for adults during inflation. The inflatable cuff is wrapped around the upper arm of an individual. The systolic and diastolic blood pressures are transmitted via air pressure and determined by the transducer integrated on the monitor with the oscillometric method. The systolic and diastolic blood pressure values and pulse rate per minute are displayed on the LED panel. The monitors C02 and F01 have the same measurement principle, specification, structure, intended use and similar software, the main differences are appearance.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    Device: Blood Pressure Monitor (Models C02 and F01)
    Manufacturer: Shenzhen Jamr Technology Co., Ltd.
    Predicate Device: Full Automatic (NIBP) Blood Pressure Monitor HL858CL (K190507)


    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criteria for the clinical performance of the blood pressure monitor are derived from the ISO 81060-2 standard.

    Acceptance Criteria (from ISO 81060-2)Reported Device Performance (Blood Pressure Monitor C02)
    Mean error and standard deviation within required limits"The results of this clinical investigation show that the required limits for mean error and standard deviation are fulfilled by the subject device C02."
    BP Accuracy: ±3mmHg (±0.4kPa)±3mmHg (±0.4kPa)
    PR Accuracy: ±5% of reading±5% of reading
    BP Measurement Range: 60-255mmHg (SYS); 40-200mmHg (DIA)60-255mmHg (SYS); 40-200mmHg (DIA)
    PR Measurement Range: 40-170 pulses/min40-170 pulses/min

    Note: The document explicitly states that the device passes various other non-clinical standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-11, IEC 80601-2-30, ISO 81060-1, IEC 62304, ISO 14971, IEC 62366-1) which imply acceptance criteria for electrical safety, EMC, performance, software, usability, and risk management, but specific quantitative acceptance values for these are not detailed in this summary.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 90 patients (42 males and 48 females)
    • Data Provenance: Not explicitly stated, but the submission is from a Chinese company (Shenzhen Jamr Technology Co., Ltd.) and refers to a "Clinical Test report" and "clinical investigation," implying this data was collected during a clinical study to support this particular device. It is most likely prospective data collected for the purpose of regulatory submission. The location of the actual clinical trial is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document states that the "Standard auscultation method was used as the reference blood pressure monitor measuring in the arm according to the sequence in ISO 81060-2."

    • Number of Experts: Not explicitly stated. The ISO 81060-2 standard typically requires measurements by multiple trained observers (usually at least two trained observers taking simultaneous or sequential measurements) to establish the reference values using the auscultatory method. However, the specific number of experts involved in this particular study is not detailed in the provided summary.
    • Qualifications of Experts: Not explicitly stated, but implies individuals trained in the auscultatory method of blood pressure measurement, as per the ISO 81060-2 standard.

    4. Adjudication Method for the Test Set

    The adjudication method is implicitly that of the ISO 81060-2 standard for assessing blood pressure devices using the auscultatory method. This standard typically involves comparisons between the automated device and a reference measurement obtained by trained observers using a sphygmomanometer and stethoscope. While the document doesn't explicitly state an "adjudication method" in terms of resolving discrepancies between multiple expert readings, the ISO 81060-2 protocol itself serves as the framework for establishing the reference values against which the device is compared. It generally involves the average of readings from multiple observers.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This study is a standalone clinical validation of the device's accuracy against a standard reference method (auscultation), not a comparison of human reader performance with and without AI assistance.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone clinical validation (algorithm only) was done. The study evaluated the accuracy of the Blood Pressure Monitor (an automated device) against the reference auscultation method. This evaluates the device's inherent performance.


    7. The Type of Ground Truth Used

    The type of ground truth used is expert consensus/clinical reference measurement established by the standard auscultation method. This is considered the gold standard for non-invasive blood pressure measurement in such validation studies.


    8. The Sample Size for the Training Set

    The document does not provide any information regarding a training set. This is a clinical validation study for a medical device's accuracy, not a machine learning algorithm that typically requires a distinct training set. The device itself (Blood Pressure Monitor) is a hardware device with an embedded algorithm; the summary focuses on the validation of the device rather than the development of the algorithm.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned or implied for this type of device validation, this information is not applicable.