Found 2 results

510(k) Data Aggregation

    K Number
    K252983

    Validate with FDA (Live)

    Date Cleared
    2025-12-17

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrinkle Retreat Pro Light Therapy Face Mask is an Over-the-Counter (OTC) device intended for use in treating full-face wrinkles.

    Device Description

    The Wrinkle Retreat Pro Light Therapy Face Mask (Model: PH-m03) adopts light emitting diodes (LED) in amber(605nm±5nm), red (630nm±5nm), red (660nm±5nm) and near-infrared (830nm±5nm) spectrum to irradiate on the face to realize its therapeutic effect. The Wrinkle Retreat Pro Light Therapy Face Mask adopts the form of a mask that contains LEDs on the inner surface of the main unit. A controller is connected to the main unit to control the device, such as to turn on/off the device. To use the device, user should place the mask over the face and use the controller to operate. The device will automatically turn off after each treatment. To prevent irradiation of LED lights to eyes during the treatment, this Mask has incorporated protective eye-shield which blocks light energy from LEDs. The Mask is powered by a built-in rechargeable lithium battery which is located in the controller.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K222316

    Validate with FDA (Live)

    Date Cleared
    2022-09-28

    (57 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair.

    Device Description

    The IPL Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The IPL Hair Removal Device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the treated skin, the device will not emit the light pulse. Also, the subject device is equipped with ice compress function to cool down the skin temperature and give a user a better experience, but this function has no other therapeutic use. The IPL Hair Removal Device includes ZHF-IPL-II and ZHF-IPL-III two models. Their intended use, performance, structure design and operation are basically identical, with the differences mainly in product appearance and its spot size of the treatment window, output energy, and display form, but these differences do not affect or change the intended use of the device. The model difference is embodied in #1) the spot size of the ZHF-IPL-II is 3.6cm2 and the corresponding output energy is 2.64.4J/cm²; the spot size of the ZHF-IPL-III is 4.1cm² and the corresponding output energy is 2.43.7J/cm²; #2) the display form of ZHF-IPL-II is LCD display screen and indicator light, and the display form of ZHF-IPL-III is indicator bar and light.

    AI/ML Overview

    This FDA 510(k) summary for the IPL Hair Removal Device (K222316) focuses on establishing substantial equivalence to a predicate device (K221001) rather than detailing a study with specific acceptance criteria related to efficacy (hair removal performance). The performance data provided primarily addresses safety.

    Therefore, for aspects related to hair removal efficacy, the document does not contain the level of detail typically found in a clinical study report proving the device meets specific performance acceptance criteria. It relies on the substantial equivalence to the predicate device, which presumably demonstrated efficacy.

    However, I can extract the acceptance criteria and performance related to device safety and general design.

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (from recognized standards)Reported Device Performance
    Biocompatibility SafetyISO 10993-5: Tests for In Vitro CytotoxicityPassed
    ISO 10993-10: Tests for Irritation and Skin SensitizationPassed
    Electrical SafetyIEC 60601-1-2: Electromagnetic compatibilityPassed
    IEC 60601-1: Basic safety and essential performancePassed
    IEC 60601-1-11: Home Healthcare EnvironmentPassed
    IEC 60601-2-83: Home Light Therapy EquipmentPassed
    Eye SafetyIEC 62471: Photobiological safety of lamps and lamp systemsPassed
    Software Verification & ValidationAll software requirement specifications met and hazards mitigated to acceptable risk levels (for moderate level of concern software)All software requirement specifications met and all software hazards have been mitigated to acceptable risk levels.

    Missing Information Regarding Efficacy Studies:

    The document does not provide the following information directly related to demonstrating hair removal efficacy:

    1. Sample size used for the test set and the data provenance: Not applicable as no efficacy study detailed.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable for an efficacy study outlined in this document.
    7. The sample size for the training set: Not applicable as no efficacy study detailed.
    8. How the ground truth for the training set was established: Not applicable.

    Explanation for Missing Efficacy Data:

    This document is a 510(k) summary for a premarket notification. The primary goal of a 510(k) is to demonstrate "substantial equivalence" to a legally marketed predicate device, not necessarily to re-prove efficacy through new, extensive clinical trials if the predicate device's efficacy is well-established for the same intended use.

    The statement under "VI. Comparison of Technological Characteristics With the Predicate Device" clarifies this: "Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use."

    The "Performance Data" section (VII) explicitly states: "The following performance data were provided in support of the substantial equivalence determination." This section then details the safety and software validation tests listed in the table above, but not a separate clinical trial for hair removal efficacy. The assumption is that because the device is substantially equivalent to a cleared predicate device with demonstrated efficacy, new efficacy studies are not required for this 510(k) clearance.