Search Results
Found 4 results
510(k) Data Aggregation
(271 days)
The N10, N12, N15, N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 12-lead selectable), Arrhythmia Analysis, ST Segment Analysis, QT Analysis, Heart Rate (HR) and Heart-Rate-Variability(HRV)), interpretations of resting 12-lead ECG, Respiration rate(Resp), Temperature(Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Carbon Dioxide (CO2). The N10MPro, N12MPro, NM15Pro Multi-parameter Patient Monitors are also intended for monitoring, displaying, reviewing, storing, alarming and transferring of physiological parameters including Masimo Rainbow SpO2, Anesthesia gas (AG), oxygen (O2) respiratory gas monitoring, Bispectral Index (BIS), Respiration Mechanics (RM) and Neuromuscular Transmission Monitoring (NMT). All the parameters can be monitored on single adult, pediatric, and neonatal patient except for the following:
- Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
- NIBP measurement continual mode is not applicable to neonates.
- When using COMEN SpO2, the monitor is intended to be used on adult patient only.
- PAWP is intended for adult and pediatric patients only.
- C.O. measurement is intended for adult patients only.
- BIS monitoring is intended for adult patients only.
- RM is intended for adult and pediatric patients only.
- NMT monitoring is intended for adult and pediatric patients only.
The monitors are to be used in healthcare facilities by healthcare professionals or under their guidance.
The Multi-parameter Patient monitors are not intended for emergency and transport use, aircraft environment or home use.
The monitors are not intend for use as apnea monitors.
The monitors are not intended for use in MRI or CT environments.
The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.
There are six (6) models under evaluation, namely N10, N12, N15, N10MPro, N12MPro, N15MPro. All models share the same intended condition of use, the same intended patient population and operator profile, biological safety characteristic and principle of operation. All these models are the same on electric and electrical circuit and components, mechanical construction, software and alarm system. The only difference lies on the screen and configuration of with/without plug-in module slot and the number of battery packs.
N/A
Ask a specific question about this device
(319 days)
The SCD600 Sequential Compression System (hereby referenced as "SCD600") is designed to increase the flow rate of the patient's venous blood in order to prevent deep vein thrombosis and pulmonary embolism and treat limb edema via intermittent pneumatic compression. The SCD600 is intended for use only in professional healthcare facilities by medical staff as part of a prescribed plan of care.
The subject device Sequential Compression System includes 1 model: SCD600.
SCD600 Sequential Compression System is designed to apply intermittent pneumatic compression to increase the flow rate of the patient's venous blood, thereby helping prevent the occurrence of deep vein thrombosis, pulmonary embolism and treat limb edema in professional healthcare facility.
The system consists of the main unit, connecting tubes, battery, AC power cord and accessory compression cuffs. The compression cuffs can increase the venous blood flow by compressing the limbs. When one compression is completed, the system will start the next compression after waiting the predetermined venous fill time.
N/A
Ask a specific question about this device
(232 days)
The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
· The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
· NIBP measurement continual mode is not applicable to neonates;
· Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
The monitors are to be used in general healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The monitors are not intended for home use.
The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
• The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
• NIBP measurement continual mode is not applicable to neonates;
• Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
Those monitors provide patient monitoring capabilities by using corresponding accessories.
The multi-parameter Patient monitor, model: C50 and C80 have same design principle and technical characteristics:
But the software in the host and modules and components are different.
The differences between C50 and C80 are ICG, BIS and size of monitor.
The provided document describes the acceptance criteria and a study proving the C50 and C80 Multi-parameter Patient Monitor meets these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established through adherence to recognized international standards and clinical performance metrics. The document details a comparison with a predicate device (K170876, Passport Series Patient Monitors). While explicit "acceptance criteria" for each parameter are not always separately listed with a single performance value from the new device, the "Comparison" column in the table below infers if the new device meets or exceeds the predicate's performance or if sufficient justification is provided for differences.
| Feature | Predicate Device (K170876) Performance | Subject Device (C50 & C80) Performance | Comparison to Acceptance Criteria (Predicate) & Justification |
|---|---|---|---|
| Monitor Size | Passport 17m: 400x370x193mm; Passport 12m: 297x336x187mm | C80: Approx. 344x291x165mm; C50: Approx. 291.7x250x187mm | The monitor size of the subject device is smaller than the predicate device, but it "won't affect the safety and effectiveness of the subject device." Implied acceptance: smaller size is acceptable if safety/effectiveness are not compromised. |
| Integrated display & touch screen | Color TFT LCD (17-inch, 1280x1024 pixels for 17m; 12-inch, 800x600 pixels for 12m) | Color TFT LCD (C80: 12.1-inch, 800x600 pixels; C50: 10.4-inch, 800x600 pixels) | Screen size is smaller, but resolution for C80/C50 is the same as Passport 12m. "Same. The screen of Subject device is smaller than predicate device. Complying with IEC 60601-1 and IEC 62366-1 also indicates the clinical use is safe and effective." Implied acceptance: smaller screen size is acceptable due to compliance with relevant IEC standards for safety and usability. |
| Power supply | Two rechargeable Lithium-ion batteries or AC for Passport 17m; One rechargeable Lithium-ion battery or AC for Passport 12m | C50 and C80: Powered either by built-in battery or external AC. | "Same, both powered by battery and AC." Implied acceptance: functional equivalence. |
| Battery | 11.1V, 4500mAh | 11.1V, 2200mAh/4400mAh | "The battery capacity is different. The batteries of C50/C80 have complied with IEC 62133. C50 and C80 have conformed to IEC 60601-1." Implied acceptance: different capacity is acceptable as long as safety (IEC 62133, IEC 60601-1) is maintained. |
| ECG | 3-lead, 5-lead and 12-lead selectable, heart rate (HR) | 3-lead, 5-lead and 12-lead selectable, heart rate (HR) | "Same." Implied acceptance: functional equivalence. |
| ECG (Arrhythmia Analysis) | Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFib (24 types) | Asystole, ventricular fibrillation, R ON T, VT >2, Couplet, PVC, Bigeminy, Trigeminy, Brady (Bradycardia), PNC (Pacer Not Capture), PNP (Pacer Not Pace), Missed Beats, IHB (Irregular Heart Beat), VTAC (Ventricular Tachycardia), Tachy (Tachycardia), PVC Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm, Heart Pause (20 types) | "C50/C80 has 20 types of arrhythmias in total, 18 of which is same as the predicate device. The other two (underline) is supported by compliance with EC57." Implied acceptance: fewer types are acceptable as long as key arrhythmia detections are present and supported by EC57 compliance. |
| Respiration | Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified. | Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified. | "Same." Implied acceptance: functional and performance equivalence. |
| Pulse oxygen saturation (SpO2) | Method: red and infrared light method. Masimo SpO2: Range:1 | Method: same. Masimo SpO2: same range & accuracy. Nellcor SpO2: same range & accuracy. Comen SpO2: Range: 0%-100%. Accuracy: 70-100% ±2%(adult/pediatric, non-motion), ±3%(neonate, non-motion). | "Same." (for Masimo and Nellcor modules). For their own Comen SpO2 module, it shows equivalent accuracy. Implied acceptance: performance equivalence with predicate's different OEM modules and new Comen module meets similar accuracy. |
| Pulse rate (PR) | From Mindray SpO2 (Range 20-254 bpm, Acc ±3 bpm), Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%). | From Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), Comen SpO2 (Range 20-254 bpm, Acc ±2 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%), NIBP sensor (Range 40-240 bpm, Acc ±3 bpm or ±3%). | "The PR from Comen SpO2 is more accuracy than the PR form Mindray SpO2. The PR form Masimo SpO2, Nellcor SpO2 and IBP sensor of C50 and C80 are the same with the predicate device. In addition, C50 and C80 have one more PR data source (NIBP) than predicate device. C50 and C80 have complied with ISO 80601-2-61." Acceptance criteria met or exceeded; added NIBP source (improvement). Adherence to ISO 80601-2-61. |
| Non-invasive blood pressure (NIBP) | Method: Oscillometry. Range: Adult (systolic 25-290mmHg, diastolic 10-250mmHg), pediatric (systolic 25-240mmHg, diastolic 10-200mmHg), neonate (systolic 25-140mmHg, diastolic 10-115mmHg). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg. | Method: same. Range: Adult (systolic 40-270mmHg, diastolic 10-215mmHg), pediatric (systolic 40-200mmHg, diastolic 10-150mmHg), neonate (systolic 40-135mmHg, diastolic 10-100mmHg). Accuracy: 0-300mmHg: ±3mmHg. | "The NIBP measure range for adult/pediatric/neonate in C50 and C80 is smaller than that in predicate device. The C50 and C80 is more accuracy than predicate device. C50 and C80 have conformed to IEC 80601-2-30." Acceptance criteria met or exceeded in accuracy, and justification for smaller range implicitly accepted by meeting IEC 80601-2-30. |
| Temperature (Temp) | Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.1°C. | Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.2°C. | "The predicate device is more accuracy than C50 and C80. C50 and C80 have complied with ISO 80601-2-56." Implied acceptance: slightly lower accuracy is acceptable as long as it complies with ISO 80601-2-56. |
| Carbon dioxide (CO2) | Method: Infrared absorption. Masimo CO2: Sidestream 0-99mmHg, AwRR 0-120rpm. Microstream 0-99mmHg, AwRR 0-150rpm. Accuracy: Sidestream 0-40mmHg ±2mmHg, 41-76mmHg ±5%, 77-99mmHg ±10%, AwRR ±2rpm. Microstream 0-38mmHg ±2mmHg, 39-99mmHg ±5% +0.08%, AwRR 0-70rpm ±1rpm, 71-120rpm ±2rpm, 121-150rpm ±3rpm. | Method: same. Masimo CO2: Sidestream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Mainstream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Respironics CO2: Sidestream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Mainstream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Accuracy (Masimo): ±(2.25mmHg +reading x 4%), AwRR ±1rpm. Accuracy (Respironics): 0-40mmHg ±2mmHg, etc., AwRR ±1rpm. | "The C50 and C80 are more accuracy and measured wider than predicate device. For C50 and C80, Masimo CO2 module (K123043 and K103604) and Respironics CO2 module (K040183 and K053174) have been cleared. Both modules have complied with ISO 80601-2-55." Acceptance criteria met or exceeded; wider range and higher accuracy; existing modules previously cleared and comply with ISO 80601-2-55. |
| Invasive blood pressure (IBP) | Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor). | "Same." Implied acceptance: functional and performance equivalence. |
| Cardiac output (C.O.) | Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater. | Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater. | "Same." Implied acceptance: functional and performance equivalence. |
| Anesthetic gas (AG) | Method: Infrared absorption. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 ±0.3% ABS, N2O ±(8%REL+2%ABS), Other AG 8%REL). | Method: same. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 0-15%: ±(0.2kPa+readingx2%), N2O ±(2 kPa+readingx2%), Hal/Enf/Iso 0-8%: ±(0.15%+readingx5%), Sev 0-10%: ±(0.15%+readingx5%), Des 0-22%: ±(0.15%+readingx5%), O2 ±(1%+readingx2%), Masimo AG AwRR ±1rpm). | "For CO2/Enf/Hal/Iso/Sev/Des, The measurement range of C50 and C80 is smaller than the predicate device. For awRR, the measurement range of C80 is better than the predicate device. The C50 and C80 are more accuracy than predicate device. The measurement range of C50 and C80 is enough for most environments; it won't affect the safety and effectiveness. The C50 and C80 supports two AG modules: MASIMO ISA AX+ Sidestream module and Masimo IRMA AX+ Mainstream module, both of them have been cleared in K103604." Acceptance criteria: different ranges are acceptable with justification of sufficient range for most environments and high accuracy. Utilizes previously cleared and compliant modules. |
| BIS (C80 only) | Range: BIS, BIS L, BIS R: 0-100; SQI, SQI L, SQI R: 0-100%; EMG, EMG L, EMG R: 0-100 Db; SR, SR L, SR R: 0-100%; SEF, SEF L, SEF R: 0.5-30.0 Hz; TP, TP L, TP R: 40-100 Db; BC, BC L, BC R: 0-30; sBIS L, sBIS R: 0-10.0; Semg L, Semg R: 0-10.0; ASYM: 0-100%. Accuracy: 1% for BIS, SQI, EMG, ESR (implied). | Range and Accuracy: BIS: same; accuracy: 1%. SQI: same; accuracy: 1%. EMG: same; accuracy: 1%. ESR: 0-100%; accuracy: 1%. | "The underline parameters are what C50 and C80 doesn't have. The BIS module and sensor have been cleared by FDA and its 510k numbers are K040183 and K002734." Acceptance criteria for essential parameters are met by using previously cleared FDA modules. |
| ICG (C80 only) | Method: Indirect impedance cardiograph measurement. Range: SV: 5-250 ml; HR: 44-2m; C.O. 1.4-15 L/min. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified. | Method: Indirect impedance cardiograph measurement. Range: HR: 40-250bpm; SV: 0-250mL; C.O.:0-30L/min; TFC: 5-150 /KΩ; SVR:0-3500 dyn・s ・ cm-5. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified. | "The underline parameter is what C50 and C80 doesn't have. The ICG electrode cable and sensors are all cleared by FDA, 510(k) number of which is K110645." Acceptance criteria for essential parameters are met by using previously cleared FDA components. |
Study that Proves the Device Meets Acceptance Criteria
The essential studies to prove the device meets acceptance criteria are divided into Non-Clinical Performance Data and Clinical Studies.
Non-Clinical Performance Data:
- Software Verification and Validation Testing: Conducted in accordance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Risk analysis was performed, and the software was deemed a "major" level of concern.
- Electrical safety and Electromagnetic Compatibility (EMC): Testing conducted according to:
- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 4: 2014-02
- Bench Testing: Functional and system-level tests were performed, showing the devices meet specifications and perform equivalently to the predicate.
- Biocompatibility Testing: Performed for patient-contacting components (ECG cable, SpO2 probes, temperature probes) that were not previously cleared. Cytotoxicity, sensitization, and irritation testing were conducted based on ISO 10993-1 and FDA's 2016 biocompatibility guidance.
- Conformance with Applicable Standards: The device demonstrated compliance with numerous recognized consensus standards, in addition to those listed above for specific modules (e.g., ISO 80601-2-55 for Respiratory Gas Monitors, ISO 80601-2-56 for Thermometers, ISO 80601-2-61 for Pulse Oximeter Equipment, ANSI AAMI EC57:2012 for Cardiac Rhythm and ST-Segment measurement algorithms).
Clinical Studies:
1. NIBP Clinical Study - for adults and children
- Device Parameter: Non-invasive Blood Pressure (NIBP)
- Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
- Sample Size: 25 patients (19 adults, 6 children).
- Demographics: 10 men, 15 women. 6 patients aged 3-12, 2 aged 12-40, 10 aged 40-60, 7 above 60.
- Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The Second Affiliated Hospital of Guangzhou Medical University and The First Affiliated Hospital of Guangzhou Traditional Chinese Medical University" from 2012 to 2013 (China).
- Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
- Adjudication Method: Not specified, but standard practice for NIBP clinical validation involves multiple observers for reference measurements.
- Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."
2. NIBP Clinical Study - for neonate and infants
- Device Parameter: Non-invasive Blood Pressure (NIBP)
- Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
- Sample Size: 20 subjects.
- Demographics: 10 males, 10 females. 13 subjects <29 days, 4 subjects 29 days to 1 year, 3 subjects 1 to 3 years old.
- Data Points: 10 groups of data collected for each subject, totaling 200 groups.
- Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The First Affiliated Hospital of Guangxi Medical University and The First Affiliated Hospital of Xinxiang Medical University" from 2012 to 2013 (China).
- Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
- Adjudication Method: Not specified.
- Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."
3. SpO2 Clinical Study
- Device Parameter: Oxygen saturation (SpO2)
- Acceptance Criteria/Standard: Complies with FDA Certification Guidance Document of Photoelectric Oximeter and ISO 80601-2-61 "Medical Electrical Equipment - Part 2-61: Particular Requirements For Basic Safety And Essential Performance Of Pulse Oximeter Equipment."
- Sample Size: 30 subjects (24 adult subjects, 6 neonates).
- Demographics (Adults): 24 subjects (7 males, 17 females) aged 24-44 years old (20 yellows, 4 blacks).
- Demographics (Neonates): 6 neonates (5 males, 1 female) aged 1-24 days.
- Data Provenance: Implies prospective data collection from "the Affiliated Hospital of Guilin Medical University" from January 20, 2016 to June 10, 2016 (China).
- Ground Truth: Implicitly a controlled desaturation study where arterial blood gas samples are taken and analyzed for SaO2, which is compared to the pulse oximeter readings.
- Adjudication Method: Not specified.
- Results: "Per the clinical test analysis result, there is no adverse events, the oxygen saturation module and probe are safe and effective for the intended use, as well as demonstrating accurate pulse measurement."
Additional Requested Information:
-
2. Sample size for the test set and the data provenance:
- NIBP (Adult/Children): 25 patients from China (2012-2013).
- NIBP (Neonate/Infants): 20 subjects (200 data groups) from China (2012-2013).
- SpO2: 30 subjects (24 adults, 6 neonates) from China (Jan-June 2016).
- All clinical data appears to be prospective as it describes clinical trials "conducted in" specific hospitals during specific periods.
-
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number or qualifications of experts (e.g., radiologists) for establishing ground truth in clinical studies. For NIBP validation against ISO 81060-2, ground truth typically involves at least two trained observers performing auscultatory measurements. For SpO2, ground truth is established by laboratory CO-oximetry measurements of arterial blood, usually managed by laboratory technicians in a controlled setting, not interpretation by "experts" in the sense of image analysis.
-
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not specify the adjudication method for any of the clinical or non-clinical tests. For NIBP, ISO 81060-2 dictates specific statistical analysis methods (e.g., Bland-Altman) for comparing device readings to reference measurements, often requiring two or more observers.
-
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This device is a multi-parameter patient monitor measuring physiological parameters (ECG, SpO2, NIBP, Temp, etc.). It does not describe an AI-powered diagnostic imaging device or an AI-assistance feature for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
-
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The device itself is a standalone patient monitor that provides automated measurements and alarms. The performance of its modules (e.g., NIBP, SpO2 algorithms) was evaluated independently against reference standards (clinical validation) and through bench testing. So, yes, standalone performance of the underlying algorithms and hardware was demonstrated through the clinical and non-clinical tests.
-
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- NIBP: Likely comparative measurements against a reference standard method (e.g., auscultatory measurements by trained observers) as specified by ISO 81060-2.
- SpO2: Involves laboratory arterial blood gas analysis (CO-oximetry) during controlled desaturation studies, comparing the device's SpO2 readings to the directly measured SaO2 values.
- Other Parameters (ECG, Temp, IBP, CO2, AG, BIS, ICG): Performance is primarily verified through bench testing against calibrated simulators and reference equipment and adherence to respective ISO/IEC standards.
-
8. The sample size for the training set:
- The document describes pre-market notification (510k) for a patient monitor, not an AI/machine learning device that typically involves a training set. The clinical studies performed are for validation of the device's performance against human subjects, not for training internal algorithms. Therefore, there is no explicit "training set" mentioned in the context of machine learning.
-
9. How the ground truth for the training set was established:
- As there is no explicit mention of an AI/machine learning training set, this question is not applicable. The ground truth for validation was established by reference standard methods in clinical trials and bench testing as described in point 7.
Ask a specific question about this device
(283 days)
C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. It's used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. All parameters can be applied to adult, pediatric or neonate. The Patient Monitor is not intended to be used on the patients with pacemaker.
The monitor is used for monitoring the clinical patients, so only the doctors and nurses who are qualified through training can use these monitors.
C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is intended to be used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Resp), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. Other parameters can be applied to adult, pediatric or neonate.
The provided text is a 510(k) Summary for the Shenzhen Comen Medical Instruments Co., Ltd. Multi-parameter Patient Monitor, Model C30. It describes the device's characteristics, intended use, and comparisons to legally marketed predicate devices.
However, the document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and the results of a specific clinical study for the device's performance in isolation. The testing mentioned is primarily related to electrical safety, electromagnetic compatibility, and adherence to performance standards, along with some clinical verification for NIBP and respiration rate accuracy.
Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of an AI algorithm's performance (such as sensitivity, specificity, or reader improvement) is not directly available or applicable to this type of medical device submission. This document describes a traditional multi-parameter patient monitor, not an AI-powered diagnostic device.
Based on the provided text, here's what can be extracted and what cannot be provided:
1. A table of acceptance criteria and the reported device performance:
The document lists various performance characteristics for each parameter (ECG, Respiration, Temperature, SpO2, Pulse Rate, NIBP, CO2) as part of the technological comparison to predicate devices, including measurement ranges and accuracies. These can be considered the de facto performance criteria. While the document states the device was "tested and found to be in compliance with recognized safety, performance and electromagnetic compatibility standards," it does not provide a specific table of acceptance criteria with corresponding measured performance for each parameter from a single comprehensive study report. Instead, it infers compliance through comparison to predicate device specifications and adherence to standards.
| Parameter | Acceptance Criteria (from predicate comparison/standards) | Reported Device Performance (Implied from substantial equivalence and compliance statements) |
|---|---|---|
| ECG Heart Rate (HR) | Neonate/Pediatric: 15 to 350 bpm; Adult: 15 to 300 bpm; Accuracy: ±1% or ±1 bpm (whichever is larger) | The document states the device has "similar" characteristics and "complies" with performance standards, implying it meets these ranges and accuracies, but no direct measured performance values are provided. |
| Respiration Rate (Resp) | Measurement range: Adult: 0 to 120 rpm; Neonate: 0 to 150 rpm; Accuracy: 7 to 150 rpm: ±1 rpm (for subject device, compared to predicate's ±2 rpm or ±2%); 0 to 6 rpm: Not specified. | Clinical study indicated to verify RR accuracy by comparing subject device RR to capnography. Implies accuracy within the stated/expected range, but no specific results are quantified in the summary. |
| Temperature (Temp) | Measurement range: 0 to 50°C (32 to 122°F); Accuracy: ±0.1°C or ±0.2°F (without probe) | No specific measured performance values provided, but implied to be compliant with these specifications. |
| SpO2 | Masimo SpO2: Measurement range: 1 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric, no motion), ±3% (neonate, no motion, and with motion). Nellcor SpO2: Measurement range: 0 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric), ±3% (neonate). 0 to 69%: Not specified. | No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics. |
| Pulse Rate (PR) | From Masimo SpO2: Range 25 to 240 bpm; Accuracy: ±3 bpm (no motion), ±5 bpm (with motion). From Nellcor SpO2: Range 20 to 300 bpm (20-250 bpm: ±3 bpm, 251-300 bpm: not specified). | No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics. |
| Non-Invasive Blood Pressure (NIBP) | Measurement range (Adult/Pediatric/Neonate systolic/mean/diastolic ranges vary as per table). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg. (These are performance requirements based on ISO 81060-2:2013). | "According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This statement indicates testing was done according to the standard, implying it met the accuracy requirements specified by that standard, but explicit performance values are not provided. |
| Carbon Dioxide (CO2) | Respironics: Range 0-150 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-70 mmHg: ±5%, 71-100 mmHg: ±8%, 101-150 mmHg: ±10%. awRR: 0-150 bpm, Accuracy: ±1 rpm. Masimo: Range 0-190 mmHg; Accuracy: 0-114 mmHg: ±1.52 mmHg + 2%. awRR: 0-150 bpm, Accuracy: ±1 rpm. | No specific measured performance values provided, but implied to be compliant with these specifications. |
2. Sample size used for the test set and the data provenance:
- NIBP Clinical Study: The document states "According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This standard specifies requirements for Non-Invasive Sphygmomanometers, including clinical investigation procedures. However, the exact sample size for this specific study and the data provenance (country, retrospective/prospective) are not given in this summary.
- Respiration Rate Clinical Study: "To verify respiration rate (RR) accuracy, Clinical study was conducted comparing the RR from the subject device with that from capnography." The exact sample size and data provenance are not given.
- For other parameters, bench testing and compliance with standards are mentioned, but no specific clinical test sets or sample sizes are detailed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided as this is not an AI diagnostic device and does not inherently involve expert image interpretation or similar processes for ground truth establishment. Clinical studies mentioned (NIBP, RR) would likely use reference measurements from other validated medical devices as a "ground truth" or comparator, but details on expert involvement are absent.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable/provided. Adjudication methods are typically used in studies involving subjective assessments, especially common in AI diagnostic imaging where multiple readers reconcile disagreements. This device is a physiological monitor measuring objective parameters.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable/provided. This is a physiological monitoring device, not an AI-powered diagnostic tool intended to assist human readers in interpreting complex medical data or images. Therefore, an MRMC study is beyond the scope of this type of device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This information is not applicable/provided. This device is a multi-parameter patient monitor that measures physiological signals directly rather than an AI algorithm performing a standalone diagnostic function. The device's performance is its direct measurement output, which is then used by healthcare professionals.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For the NIBP and Respiration Rate clinical studies, the "ground truth" would likely be established through reference measurements from other validated medical devices or established clinical methods. For example, NIBP accuracy usually compares the device's readings to those obtained through an invasive arterial line or another highly accurate, validated NIBP device. For respiration rate, capnography (a precise measurement of CO2 in exhaled breath) was explicitly mentioned as the comparator.
8. The sample size for the training set:
The device described is a traditional patient monitor developed through engineering and design principles, not a machine learning or AI model that requires a "training set." Therefore, this concept is not applicable to this product.
9. How the ground truth for the training set was established:
As above, the concept of a "training set" and "ground truth for the training set" is not applicable to this traditional medical device.
Ask a specific question about this device
Page 1 of 1