Search Results
Found 6 results
510(k) Data Aggregation
(178 days)
This product is intended to be used in combination with compatible colonoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.
This product needs to be cut and connected to the endoscope tip of 9mm to 14mm, and medical tape needs to be prepared in advance to fix it into a cylindrical shape.
This product is intended to be attached to an endoscope.
The Disposable Distal End Tape Hood is installed at the tip of the endoscope to assist the endoscope in a clearer observation. It can push away the mucosa and keep a certain distance between the lens and the mucosa, thus maintaining the field of view.
N/A
Ask a specific question about this device
(112 days)
The Flexible Video-Choledochoscope is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.
Flexible Video-Choledochoscope is a kind of medical instrument, also known as optical camera, which can enter into the human biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends. This device must be used with a duodenoscope. The product is equipped with tiny size digital imaging parts -- photoelective sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on its monitor output for doctor observation and diagnosis.
I am sorry, but the provided text does not contain information about the acceptance criteria and the study that proves the device meets those criteria. The document is an FDA 510(k) clearance letter for a Flexible Video-Choledochoscope, which asserts substantial equivalence to a predicate device based on technical and performance characteristics.
Specifically, the document mentions:
- No Clinical Study is included in this submission. (Section 5.9 Clinical Test Data)
- Non-clinical Performance Data: "The following performance tests were conducted to substantial equivalence: mechanical performance: duodenosoone compatibility testing and flexible surgical instrument compatibility testing and comparison performance testing." (Section 5.8 Non-clinical Performance Data)
Therefore, I cannot provide the detailed information requested in your prompt regarding acceptance criteria, device performance, sample sizes, ground truth establishment, or any comparative effectiveness studies.
Ask a specific question about this device
(145 days)
The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.
Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface. Flexible Video-Choledo-Cysto-Ureteroscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human bladder, ureter, biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends. The product is equipped with tiny size digital imaging parts -- photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on it's monitor output for doctor observation and diagnosis.
This document is a 510(k) premarket notification for a new medical device, the Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100). This type of submission relies on demonstrating "substantial equivalence" to a predicate device, meaning it's as safe and effective as a device already legally marketed. Therefore, the "study" referred to is primarily a non-clinical performance evaluation comparing the proposed device to a predicate device, rather than a clinical trial or AI-specific validation study typically associated with AI/ML devices.
Here's an analysis of the provided information concerning acceptance criteria and the "study":
1. Table of Acceptance Criteria and Reported Device Performance:
The document provides a comparative table between the proposed device (PL-2100) and its primary predicate device (Flexible Video-Choledo-Cysto-Ureteroscope System, K211686, Model: PL-1000). The "acceptance criteria" are implied by the predicate device's characteristics, and the "reported device performance" are the proposed device's characteristics. The goal is to show they are "Same" or "Similar" in ways that don't raise new questions of safety or effectiveness.
| Item | Acceptance Criteria (Predicate Device) | Reported Device Performance (Proposed Device) | Comment |
|---|---|---|---|
| Imaging Processor System | Model: PL-1000 | Model: PL-2100 | |
| Indication for Use | Signals from the endoscope are converted to the image and displayed on an inspection monitor. | Signals from the endoscope are converted to the image and displayed on an inspection monitor. | Same |
| Target patients | Patients who are considered suitable for the application of this product by the physician | Patients who are considered suitable for the application of this product by the physician | Same |
| User qualifications | Physicians (experts approved by endoscopic medical safety administrator at each medical facility, or as defined by government/academic society). Specific training is not required. | Physicians (experts approved by endoscopic medical safety administrator at each medical facility, or as defined by government/academic society). Specific training is not required. | Same |
| Place of use | Medical facility | Medical facility | Same |
| Power input | 100-240V~50/60Hz45VA | 100-240V~50/60Hz48VA | Similar |
| Imaging output | HDMI | CVBS&DVI&S-VIDEO | Different |
| Light Source | LED 2W | LED 2W | Same |
| Lamp Life | 30,000 hrs | 30,000 hrs | Same |
| Screen | N/A (implied external display) | 10.1 inches | Different |
| USB Flash Memory | USB 2.0 (Max:128G; FAT32) | USB 2.0 (Max:128G; FAT32) | Same |
| illumination | ≥2000LUX | ≥2000LUX | Same |
| Resolution | d = 10mm, MTF50 value of SFR at center not lower than 200lw/pH, MTF50 value at 70% FOV not lower than 80% of center position. | d = 10mm, MTF50 value of SFR at center not lower than 200lw/pH, MTF50 value at 70% FOV not lower than 80% of center position. | Same |
| Field of View | 110°±10% | 110°±10% | Same |
| Observe the depth of field | 5-50mm, MTF50 ≥ 200 (LW/PH) | 5-50mm, MTF50 ≥ 200 (LW/PH) | Same |
| Direction of View | 0°±2° | 0°±2° | Same |
| TV distortion | < 15% | < 15% | Same |
| SNR and Dynamic Range | SNR≥10, DR≥90 | SNR≥10, DR≥90 | Same |
| Color reproducibility test | ΔE≤23, ΔC≤23 | ΔE≤23, ΔC≤23 | Same |
| Shading | Lightness non-uniformity DL≥70, Luminance non-uniformity Dγ≥95 | Lightness non-uniformity DL≥70, Luminance non-uniformity Dγ≥95 | Same |
2. Sample Size Used for the Test Set and Data Provenance:
The document explicitly states: "No Clinical Study is included in this submission." This means there isn't a "test set" in the sense of patient data used for algorithm validation. The evaluation is primarily based on non-clinical engineering and performance testing.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
Not applicable, as no clinical study or expert-adjudicated ground truth dataset was used for device efficacy or performance evaluation in this 510(k) submission.
4. Adjudication Method for the Test Set:
Not applicable, as no clinical study or expert adjudication process for a test set was performed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
Not applicable. This device is an endoscope system, not an AI/ML-assisted diagnostic device that would typically undergo an MRMC study to evaluate human reader improvement with AI assistance.
6. Standalone Algorithm Performance:
Not applicable. The device is a hardware system (endoscope and imaging processor), not a standalone algorithm.
7. Type of Ground Truth Used:
For the non-clinical performance and safety evaluations, the "ground truth" is established by international and national standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 10993) and the specified performance characteristics of the predicate device. The proposed device's performance is measured against these established engineering and safety benchmarks.
8. Sample Size for the Training Set:
Not applicable. This is not an AI/ML device, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for an AI/ML algorithm.
Ask a specific question about this device
(461 days)
The values are used in conventional gastrointestinal endoscope to control the functions of water/gas injection and suction during the endoscopic operation.
Disposable Endoscope Valves Kit consists of Water/gas button, Attraction button, Biopsy cap and Water injection adapter. Disposable Endoscope Valves Kit connect the endoscope operating handle matched with the model, and install the water gas button into the water gas port of the operating handle; Align the positioning boss of the suction button with the card slot of the operating handle and install it into the suction port; Install the biopsy cap into the clamp opening of the operating handle.
The provided text describes the acceptance criteria and a study demonstrating the substantial equivalence of the "Disposable Endoscope Valves Kit" to a predicate device. However, it does not include information typically found in studies proving an AI/ML device meets acceptance criteria, such as specific performance metrics like sensitivity, specificity, or AUC, detailed sample sizes for test sets, expert qualifications, or adjudication methods for ground truth.
This document is a 510(k) summary for a medical device that does not appear to be an AI/ML device. It seems to be a physical medical accessory. Therefore, the questions related to AI/ML specific criteria (such as effect size of human readers with AI vs without AI, standalone algorithm performance, training set details) are not applicable to this submission.
The study described is a non-clinical performance data study, not a clinical study involving AI/ML.
Here's an attempt to answer the relevant questions based on the provided text, while noting the limitations due to the nature of the device:
1. A table of acceptance criteria and the reported device performance
The document frames "acceptance criteria" as meeting regulatory standards and demonstrating substantial equivalence to a predicate device through non-clinical testing. It doesn't list quantitative performance metrics for the proposed device in the way one would for an AI/ML diagnostic tool (e.g., sensitivity, specificity).
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Biological Evaluation | Meets ISO 10993 requirements |
| Sterilization | Meets ISO 11135-1 and ISO 10993-7 requirements (EO Sterilization) |
| Appearance | Passing |
| Physical Properties | Passing |
| Substantial Equivalence | Established through comparison to predicate device K220210 |
2. Sample size used for the test set and the data provenance
The document mentions "bench tests" were performed. For these non-clinical performance tests (Appearance, Physical properties), the specific sample size is not explicitly stated, nor is the data provenance. These would typically involve testing a certain number of manufactured units.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable as the device is a physical accessory and the evaluation is based on engineering and material standards, not diagnostic interpretation requiring expert ground truth in the AI/ML context.
4. Adjudication method for the test set
This question is not applicable for the same reasons as above. The evaluation of appearance and physical properties would follow established testing protocols and standards, not a clinical adjudication process.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No such study was conducted or reported. This type of study is relevant for AI/ML diagnostic devices, which this device is not. The document explicitly states: "No Clinical Study is included in this submission."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a physical endoscope accessory, not an algorithm.
7. The type of ground truth used
For the non-clinical tests (biological evaluation, sterilization, appearance, physical properties), the "ground truth" would be compliance with established international standards (ISO series) and predefined specifications for the physical characteristics of the device. This is not "expert consensus, pathology, or outcomes data" in the typical sense of AI/ML diagnostic validation.
8. The sample size for the training set
This question is not applicable as this is not an AI/ML device and therefore does not have a "training set."
9. How the ground truth for the training set was established
This question is not applicable for the same reason as above.
Ask a specific question about this device
(266 days)
The Flexible Video-Choledochoscope System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.
The Flexible Video-Choledochoscope System comprises two components: the Flexible Video- Choledochoscope and Imaging Processor System (Including Light Source).
The Flexible Video-Choledochoscope is indicated for use in diagnostic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.
The imaging processor system (including the light source) provides illumination for Flexible Video-Choledochoscope and is also used to receive the signal from the endoscope, then convert it into an image and display it on the examination monitor.
Flexible Video-Choledochoscope System is composed of sterile Flexible Video-Choledochoscope and a non-sterile Imaging Processor System (Including Light Source). The Flexible Video-Choledochoscope is composed of Operation section, Light-guide Section, Insertion a Distal tip and bending section. The Operation Section is pulled by the wire rope to control the bending direction of the Distal tip.
The Light-guide section transmits the illumination light from the image processor to the Distal tip. The Insertion Tube is used to guide it into other instruments or cavity of the body. The Distal tip contains a camera system and a lighting system for illumination and observation.
Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface.
Flexible Video-Choledochoscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends.
The product is equipped with tiny size digital imaging parts --photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on its monitor output for doctor observation and diagnosis.
The provided FDA 510(k) summary does not contain information about acceptance criteria or a study proving the device meets those criteria. Instead, it details the device's technical specifications and compares them to a predicate device to establish substantial equivalence.
Therefore, I cannot provide the requested information.
Ask a specific question about this device
(265 days)
The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.
Flexible Video-Choledo-Cysto-Ureteroscope System is composed of sterile Flexible Video-Choledo-Cysto-Ureteroscope and an non-sterile Imaging Processor System (Including Light Source).The Flexible Video-Choledo-Cysto-Ureteroscope is composed of Control Section, Light-guiding Section, Connector Section, Insertion Tube and Distal End. The Control Section is pulled by the wire rope to control the bending direction of the Distal End.
The Light-guide section transmits the illumination light from the image processor to the Distal End. The Insertion Tube is used to guide it into other instruments or cavity of the body. The Distal End contains a camera system and a lighting system for illumination and observation.
Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface.
Flexible Video-Choledo-Cysto-Ureteroscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human bladder, ureter, biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends.
The product is equipped with tiny size digital imaging parts --photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on it's monitor output for doctor observation and diagnosis.
This document is a 510(k) Premarket Notification from the FDA for a medical device called the "Flexible Video-Choledo-Cysto-Ureteroscope System." As such, it does not contain the information requested regarding acceptance criteria, study details (sample sizes, expert involvement, ground truth methods), or the performance of an AI/algorithm.
The document primarily focuses on establishing substantial equivalence to a predicate device (Flexible Video-Uretero-Choledochoscope System, K142556) based on:
- Indications for Use: The current device has similar indications for endoscopic examination in the urinary tract and bile ducts.
- Technological Characteristics: Comparison of features like imaging technology (CMOS), illumination source (LED), distal end size, deflection, field of view, and imaging processor system capabilities.
- Non-clinical Performance Data: Mention of testing against ISO standards for function, performance, electrical safety (IEC 60601-1, IEC 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993). Sterilization process and EO residual validations are also mentioned.
Crucially, Section 5.8 "Clinical Test Data" explicitly states: "No Clinical Study is included in this submission." This means there is no data presented here about human performance (with or without AI assistance), standalone algorithm performance, or a comparative effectiveness study.
Therefore, I cannot provide a response filling in the requested details about acceptance criteria or a study proving the device meets those criteria, as that information is not present in the provided text. The device approval here is based on substantial equivalence to a predicate, not on a clinical trial demonstrating performance against specific quantitative acceptance metrics.
Ask a specific question about this device
Page 1 of 1