510(k) Data Aggregation

    K Number
    K233477

    Validate with FDA (Live)

    Date Cleared
    2024-07-01

    (249 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vitrification Kit is indicated for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

    The Warming Kit is indicated for use in the thawing of vitrified oocvtes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

    The Cryo-straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embrvos.

    Device Description

    The Vitrification Kit and Warming Kit are assisted reproduction technology (ART) media products for freezing and thawing oocytes (MII), pronuclear (PN) zygotes through Day 3 cleavage stage embryos, and blastocyst stage embryos.

    There are two models for each Vitrification Kit and Warming Kit. The difference is the volume of the solutions in the kits. The solutions in the "A" model for each kit are in 1 ml vials, and the solutions in the "B" model are in 2 ml vials. Each model contains two sets of kit media.

    The Vitrification Kit includes three sequential media components, Washing solution (WS), Equilibrium solution (ES), and Vitrification solution (VS), containing the cryoprotectants ethylene glycol, dimethyl sulfoxide, and sucrose. Using this methodology, the permeating cryoprotectants can replace water in the oocyte and PN through blastocyst stage embryos prior to vitrification and storage in liquid nitrogen.

    The Warming Kit is composed of three media used sequentially for thawing and removing cryoprotectants from vitrified oocytes and PN through blastocyst stage embryos. The Warming Kit is composed of Thaw solution (TS), Diluent solution (DS), and Washing solution (WS).

    All media in the Vitrification Kit and Warming Kit undergo aseptic filtration and are single-use only.

    The Cryo-Straw is a single-use, sterile, cryopreservation storage device for holding and maintaining vitrified oocytes (MII) and pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. There are two models for the Cryo-straws: Type I and Type III, and each has 5 colors: brown, orange, blue, red, green. The Cryo-straws are composed of a rod and sheath with a combined length of 132.5 mm. The Type I model has a flat slide and the Type III model has a curved slide for sample loading. The Cryo-straws are sterilized using gamma radiation.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: Cryo-straw, Warming Kit, and Vitrification Kit. It details the device's indications for use, comparison to a predicate device, and non-clinical performance data.

    However, this document does not describe a study involving an AI/Machine Learning (ML) algorithm or human readers aided by AI. The acceptance criteria and performance data presented are for the physical cryopreservation device and its associated media, focusing on biological and physical properties relevant to in vitro fertilization (IVF) and cryopreservation, such as pH, osmolality, sterility, shelf-life, and mouse embryo assay (MEA) results.

    Therefore, I cannot provide the requested information about:

    1. A table of acceptance criteria and reported device performance related to AI/ML. The acceptance criteria listed are for the physical and chemical properties of the cryopreservation media and device.
    2. Sample size used for the test set and data provenance: No test set involving AI/ML is described.
    3. Number of experts used to establish ground truth & qualifications: Not applicable to this device submission.
    4. Adjudication method for the test set: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
    6. Standalone (algorithm-only) performance: Not applicable.
    7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for this device is based on laboratory assays (e.g., MEA, sterility, endotoxin) and physical testing.
    8. Sample size for the training set: Not applicable, as this is not an AI/ML device.
    9. How the ground truth for the training set was established: Not applicable.

    Summary of Device Acceptance Criteria and Performance (from document):

    The document focuses on demonstrating the substantial equivalence of the "Cryo-straw," "Warming Kit," and "Vitrification Kit" to a legally marketed predicate device (Kitazato Corporation's Vitrification Kit and Thawing Kit, K171748). This is achieved through non-clinical performance data addressing the physical, chemical, and biological properties critical for their intended use in cryopreservation of oocytes and embryos.

    Here's the relevant information extracted from the document regarding the device's acceptance criteria and reported performance:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test/ParameterAcceptance Specification (Criteria)Reported Device Performance (Typically stated as "met the requirements" or similar)
    Vitrification/Warming Kit
    SterilityAseptic processing validation, Sterility per USP<71>Met ISO 13408-1:2008/Amd1:2013 & ISO 13408-2:2018 requirements; No microbial growthMet requirements
    Shelf-lifeSupported shelf-life6 months (supported by time 0 and real-time aging data)Supported a 6-month shelf-life
    pHper USP <791>All solutions: 7.2-7.4Met specifications
    Osmolalityper USP <785>ES: 1391-1590 (1:1 dilution)VS: 1430-1710 (1:3 dilution)TS: 1,650-2,170DS: 850-925WS: 265-295Met specifications
    Endotoxinper USP <85><0.25 EU/mLMet specifications
    Biological PerformanceMouse Embryo Assay (MEA) - One-cell system≥80% embryos developed to expanded blastocyst at 96 hoursMet specifications (≥80% embryos developed to expanded blastocyst at 96 hours)
    TransportationTransportation testing per ASTM D4169-22Met requirementsMet requirements
    Seal IntegrityCap/seal leak testing/USP <1207.2> equivalentMet requirementsMet requirements
    Cryo-straw
    SterilizationProcess & Validation (ISO 11737-1/2, ISO 11137-3)Sterilization Assurance Level 10-6Met ISO standards, SAL 10-6
    Shelf-lifeSupported shelf-life2 years (supported by time 0, real-time aging, or accelerated aging per ASTM F1980-16)Supported a two-year shelf-life
    Cooling Rate(measured)Type I: 1,539°C/min; Type III: 1,543°C/minAchieved specified rates
    Warming Rate(measured)Type I: 21,430°C/min; Type III: 21,385°C/minAchieved specified rates
    DimensionalDimensional testingMet specificationsMet specifications
    DurabilityLiquid nitrogen immersionNo damage after 30 second immersion in liquid nitrogenNo damage reported
    Liquid N2 PenetrationImmersion testNo ingress after 24-hour immersion in liquid nitrogenNo ingress reported
    Endotoxinper USP<85><0.5 EU/deviceMet specifications
    Package IntegritySeal strength (F88/F88M-23), dye penetration (ASTM F1929-23)Met requirements following aging and transportation conditioningMet requirements
    Tensile StrengthRod and sheathMet specificationsMet specifications
    Biological PerformanceMouse Embryo Assay (MEA) - One-cell system≥80% embryos developed to expanded blastocyst at 96 hoursMet specifications (≥80% embryos developed to expanded blastocyst at 96 hours)

    2. Sample size used for the test set and the data provenance:
    The document does not explicitly state specific sample sizes for each test. It refers to general testing procedures (e.g., "Shelf-life testing was conducted," "Transportation testing per ASTM D4169-22"). The data provenance is implied to be from internal testing by the manufacturer, Shandong Wego Ruisheng Medical Device Co., Ltd. in China, as part of their 510(k) submission. These are non-clinical (laboratory/bench) tests, not human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. Ground truth for these tests is established by standardized laboratory methods (USP, ISO, ASTM standards) and measurements, not expert human interpretation in a clinical sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. These are objective laboratory measurements, not subjective human assessments requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No MRMC study was performed as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    No standalone algorithm performance was done as this is not an AI/ML device.

    7. The type of ground truth used:
    The ground truth used for these performance tests is based on objective scientific and engineering measurements and standardized biological assays (e.g., MEA, pH, osmolality, endotoxin levels, dimensional measurements, and physical durability tests) performed in a laboratory setting.

    8. The sample size for the training set:
    Not applicable, as this is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established:
    Not applicable, as there is no training set for an AI/ML algorithm.