Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K252846

    Validate with FDA (Live)

    Date Cleared
    2025-11-20

    (73 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indications for use for TaperSet™ Hip System are:

    The TaperSet™ Hip System is designed for total hip arthroplasty and is intended to be used with compatible components of the Consensus Hip System including the CS2™ Acetabular Cup System.

    The indications for use are:

    A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
    B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
    C. Proximal femoral fractures.
    D. Avascular necrosis of the femoral head.
    E. Non-union of proximal femoral neck fractures.
    F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.

    The TaperSet™ hip stem is indicated for cementless use.

    The Consensus acetabular cup components are intended for cemented or cementless use. The CS2 Ti plasma coated acetabular shell is intended for cementless use. The Consensus all-poly acetabular cup is intended for cemented use.

    The intended use or indications for use of the TaperSet™ Hip System are similar to its original clearance K102399 except the removal of 'partial hip arthroplasty" as the DAA surgical technique is only applicable to the total hip system.

    Device Description

    Shalby Advanced Technologies, Inc. has created the DAA surgical technique for its cleared TaperSet™ Hip System including hip implants – TaperSet Femoral Stems™ and CS2™ Acetabular Components as well as other compatible components.

    There have been no changes made to the TaperSet™ Femoral Stems and CS2™ Acetabular Components or any other compatible components from their previous clearances in developing the DAA surgical technique.

    The TaperSet™ Femoral Stems are available in sizes ranging from 5 to 24mm in standard and 7mm lateralized neck options. The TaperSet™ Femoral Stems with Reduced Distal Profile (RDP) are available in sizes ranging from 10.5 to 24mm in standard and 7mm lateralized neck options. The femoral stems are manufactured from wrought or forged titanium alloy (Ti-6Al-4V), except the 5 and 6mm femoral stems which are manufactured from forged titanium alloy (Ti-6Al-4V) only, and the proximal portion is plasma sprayed with commercially pure titanium.

    The CS2™ Acetabular Components include acetabular shells and mating liners. The acetabular shells and liners have matching circumferential scallops that rotationally secure the insert in the shell and allow for dialing the insert in a desired orientation. The shells with screw holes have anatomically placed holes, which accommodate optional screws for additional fixation. The shells are manufactured from titanium alloy (Ti-6Al-4V) and are porous coated with either commercially pure titanium sintered or diffusion bonded irregular beads or commercially pure titanium plasma spray. The acetabular shells coated with titanium irregular beads are available in six configurations (1) Hemispherical with three screw holes, (2) Hemispherical with eight to ten screw holes, (3) Hemispherical without screw holes, (4) Flared rim with three screw holes, (5) Flared rim with eight to ten screw holes, and (6) Flared rim without screw holes. The acetabular shells with titanium plasma spray are available in one configuration: hemispherical with three screw holes. The acetabular shells are available in sizes ranging from 42 to 68mm.

    The CS2™ and CS2™ PLUS acetabular liners are manufactured from highly cross-linked UHMWPE or highly cross-linked UHMWPE blended with Vitamin-E (VitalitE). The CS2 acetabular liners made from highly cross-linked UHMWPE are available in two configurations: (1) Rimmed neutral and (2) Rimmed hooded. The CS2™ acetabular inserts made from VitalitE are available in four configurations: (1) Rimmed neutral, (2) Rimmed hooded, (3) Rimless neutral, and (4) Rimless hooded. The CS2™ PLUS acetabular inserts are designed with a +5mm lateralized offset and are available in two configurations: (1) Rimmed neutral and (2) Rimmed hooded.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1