Search Filters

Search Results

Found 17 results

510(k) Data Aggregation

    K Number
    K251035

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-03

    (61 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serim™ DISINTEK™ PA Test Strip is a chemical indicator for use in determining whether the concentration of peracetic acid, the active ingredient in the high-level disinfectant RAPICIDE™ PA HLD solution, is above the 850 ppm PAA minimum recommended concentration (MRC) established for RAPICIDE™ PA HLD at both 30°C and 20°C.

    Device Description

    The device is a qualitative, single use, reagent chemical indicator made up of a 0.2-inch x 0.2-inch indicator pad that has been chemically treated to detect peracetic acid above or below the Minimum Recommended Concentration established for RAPICIDE™ PA High-Level Disinfectant. The indicator pad is affixed to one end of a 3.25-inch by 0.2-inch white plastic handle.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided FDA 510(k) Clearance Letter for DISINTEK™ PA Test Strips:

    Acceptance Criteria and Device Performance for DISINTEK™ PA Test Strips

    1. Table of Acceptance Criteria and Reported Device Performance

    Study ParameterAcceptance Criteria (Implicit from "Met Acceptance Criteria")Reported Device Performance
    Dynamic RangeNegative response at concentrations at or below MRC (850 ppm PAA), positive responses at higher concentrations.100% FAIL results at and below the 850 ppm PAA (MRC) and PASS results above the MRC at both 30°C and 20°C.
    Instructions for Use ValidationNegative response at concentrations at or below MRC (850 ppm PAA), positive responses at higher concentrations, when used according to specified dip time (1-2 seconds) and sideblot time (25-30 seconds).100% FAIL results at and below the MRC and PASS results above the MRC with a 1-2 second dip time and a sideblot time at 25 seconds + 5 seconds for 30°C solutions and 30 seconds + 5 seconds for 20°C solutions.
    Closed Bottle Shelf-Life StabilityAcceptable performance (100% FAIL at/below MRC, 100% PASS above MRC) for minimum claimed shelf-life (18 months) at specified storage conditions (≥ 10°C and up to 32°C).Acceptable performance out to a minimum of 18 months shelf-life from ≥ 10°C storage up to 32°C, with 100% FAIL results at and below the MRC and PASS results above the MRC. (Data collection ongoing for extended shelf-life).
    Open Bottle Use-Life StabilityAcceptable performance (100% FAIL at/below MRC, 100% PASS above MRC) for minimum claimed open bottle use-life (18 months) at specified storage conditions (constant high humidity, repeated openings).Acceptable performance out to a minimum of 18 months open bottle use-life with 100% FAIL results at and below the MRC and PASS results above the MRC. (Data collection ongoing for extended open bottle use-life).
    Comparative Sensitivity and Specificity100% comparative sensitivity and 100% comparative specificity against the predicate device.Comparative sensitivity of 100% and comparative specificity of 100% was found for the DISINTEK PA test strips at both 30°C and 20°C.
    Analytic Specificity – BioburdenAcceptable performance (100% FAIL at/below MRC, 100% PASS above MRC) in the presence of bioburden (fetal bovine serum and hard water).100% FAIL results at and below the MRC and 100% PASS results above the MRC in the presence of fetal bovine serum and hard water.
    Analytic Specificity – Other DisinfectantsTest strip yields PASS results when exposed to other peracetic acid disinfectants at their use concentrations.Test strip yielded PASS results using these disinfectants.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample sizes (e.g., number of test strips, unique disinfectant solutions tested, or replicates) used for each test set within the "Dynamic Range," "Instructions for Use Validation," "Closed Bottle Shelf-Life Stability," "Open Bottle Use-Life Stability," "Analytic Specificity – Bioburden," or "Analytic Specificity – Other Disinfectants" studies.

    For Comparative Sensitivity and Specificity, the document also does not specify the numerical sample size.

    Data Provenance: The studies were conducted by Serim Research Corporation themselves, implying the data is generated internally from their own testing facilities. The data appears to be prospective, as it describes the testing performed on the new DISINTEK™ PA Test Strips to demonstrate their performance against defined criteria. No information regarding country of origin of the data beyond "Serim Research Corporation" is provided, but as an FDA submission, the testing would likely adhere to US regulatory standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. For chemical indicators like this, "ground truth" typically refers to the analytically determined concentration of PAA in the test solutions. This would be established by validated chemical testing methods (e.g., titration or spectroscopy) rather than expert consensus on visual interpretation. The document implies that the "ground truth" for FAIL/PASS results is based on a known PAA concentration (below/at 850 ppm MRC as FAIL, above as PASS), which would be determined analytically.

    4. Adjudication Method for the Test Set

    The document describes chemical indicator functionality (color change) and the interpretation of those changes (PASS/FAIL based on visual comparison to color blocks or a "solid black color" for the predicate). There is no indication of an adjudication method involving multiple human readers for interpreting the results of the DISINTEK™ PA Test Strips. The expectation is that the visual interpretation by a single user, following the provided instructions and color chart, is sufficient.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is usually conducted for imaging diagnostics or devices where human interpretation directly impacts effectiveness and can be improved by AI assistance. This device is a chemical indicator with a direct visual interpretation, not a diagnostic imaging AI system.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, the studies described are essentially standalone in terms of the device's chemical performance. The studies evaluate the chemical indicator's ability to accurately reflect the PAA concentration, independent of human variability in reading.

    However, the "Instructions for Use Validation" section does implicitly involve human interaction, as it validates the device when used as directed by a human with a specific dip and read time. The "performance" being measured (100% FAIL below MRC, 100% PASS above MRC) is the device's inherent chemical reaction, and the human's role is to correctly perform the test and interpret the color chart. So, while it's not an algorithm only in the AI sense, the core data is about the chemical functionality.

    7. Type of Ground Truth Used

    The ground truth used is based on analytically determined peracetic acid (PAA) concentrations in the RAPICIDE™ PA HLD solutions. Solutions were prepared with known concentrations below, at, and above the 850 ppm PAA Minimum Recommended Concentration (MRC). The "FAIL" result would be correlated with PAA concentrations ≤ 850 ppm, and the "PASS" result with PAA concentrations > 850 ppm.

    8. Sample Size for the Training Set

    This information is not applicable/not provided. The DISINTEK™ PA Test Strips are a chemical indicator, not an AI/machine learning device. Therefore, there is no "training set" in the context of an algorithm that learns from data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202565

    Validate with FDA (Live)

    Date Cleared
    2020-12-02

    (89 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When properly worn, the surgical face masks are intended to patient and healthcare workers from transfer of microorganisms, body fluids and airborne particles. This device is non-sterile and for single use only.

    Device Description

    The surgical face mask is a non-sterile, single use, three-layer, flat, pleated style with ear loops and a nose piece. The outer and inner layers are spunbond polypropylene, and the middle layer consists of melt-blown polypropylene filter. Each mask is secured to the face with ear loops with a nosepiece to conform the mask to the shape of the nose.

    AI/ML Overview

    This document describes the validation of a surgical face mask. As such, the concept of "AI assistance for human readers" or "ground truth established by experts with many years of experience" is not applicable in the context of device performance testing for such simple medical devices. The evaluations are based on standardized physical and biological tests against established criteria.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Fluid Resistance (ASTM F1862)29 out of 32 pass at 120 mm Hg31 out of 32 pass at 80 mm Hg30 out of 32 pass at 120 mm Hg
    Particulate Filtration Efficiency (ASTM F2299)≥ 98%98.8%
    Bacterial Filtration Efficiency (ASTM F2101)≥ 98%> 99.9%
    Differential Pressure (MIL- M-36954C)< 5.0 mm H20/cm²4.4 mmH20/cm²
    Flammability (16 CFR 1610)Class 1Class 1
    BiocompatibilityNon-Cytotoxic, Non-Sensitizing, Non-IrritatingNon-Cytotoxic, Non-Sensitizing, Non-Irritating

    2. Sample Size Used for the Test Set and Data Provenance

    • Fluid Resistance (ASTM F1862): 32 samples were tested for fluid resistance.
    • Other Tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability, Biocompatibility): The document does not explicitly state the sample size for these specific tests, but performance data is provided.
    • Data Provenance: The tests were conducted on the "proposed device," which implies new testing of the specific product being submitted for clearance. The document does not specify the country of origin for the testing itself, but the manufacturer is Ningbo Green Textile Co., Ltd, which is likely in China. The data would be considered prospective for the purposes of this submission as it was generated specifically for the device's clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this type of device and testing. The "ground truth" is established by adherence to standardized testing protocols (e.g., ASTM, MIL, ISO, CFR standards) with measurable outcomes, not by expert consensus or interpretation of medical images.

    4. Adjudication Method for the Test Set

    Not applicable. Testing is against objective, measurable criteria defined by international and national standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a surgical face mask, not an AI-powered diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a surgical face mask, not an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth is based on standardized physical and biological testing methods and their predefined acceptance criteria. These standards include:

    • ASTM F2100: Standard Specification for Performance of Materials Used in Medical Face Masks
    • ASTM F1862/F1862M-17: Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood
    • ASTM F2299/F2299M-03: Determining the Initial Efficiency of Materials Used in Medical Face Masks to the Penetration of Particulates Using Latex Spheres
    • ASTM F2101: Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials
    • MIL - M-36954C: Medical face masks - Requirements and test methods, Annex C
    • 16 CFR 1610: Flammability Test Method
    • ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for In Vitro Cytotoxicity
    • ISO 10993-10: Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical device (surgical face mask), not a machine learning model; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192445

    Validate with FDA (Live)

    Date Cleared
    2020-04-02

    (209 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QwikCheck™ Chemical Indicator is a chemical indicator for use in determining whether the concentration of peracetic acid, the active ingredient in CS Medical's TD-12 Solution, is above or below the manufacturer's established Minimum Recommended Concentration (MRC) of 1750 ppm.

    Device Description

    The device is a qualitative, single use, reagent chemical indicator made up of a 0.20-inch square indicator pad that has been chemically treated to detect peracetic acid above or below the Minimum Recommended Concentration. The indicator pad is affixed to one end of a 3.25 inch by 0.20-inch white opaque plastic handle.

    AI/ML Overview

    The QwikCheck™ Chemical Indicator is a qualitative chemical indicator designed to determine if the concentration of peracetic acid (PAA) in CS Medical's TD-12 Solution is above or below the Minimum Recommended Concentration (MRC) of 1750 ppm.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    StudyAcceptance CriteriaReported Device Performance
    Dynamic RangeNegative response at concentrations at or below MRC, positive responses at higher concentrations.100% FAIL results at and below the MRC and PASS results above the MRC.
    Instructions for Use ValidationNegative response at concentrations at or below MRC, positive responses at higher concentrations.Acceptable performance with an immediate dip/removal from solution time, a 5 ± 1 second side blot time, a time to blot of ≤ 1 second, and a 30 ± 5 second read time. 100% FAIL results at and below the MRC and PASS results above the MRC.
    Closed Bottle Shelf-Life StabilityMet specifications after storage of labeled shelf life of 12 months (unopened).Acceptable performance out to a minimum 15 months shelf-life from ambient room temperature up to 32°C, with data collection ongoing. 100% FAIL results at and below the MRC and PASS results above the MRC. A conservative 12-month shelf-life is claimed.
    Open Bottle Use-Life StabilityMet specification for 3 months (opened).Acceptable performance out to a minimum 5 months open bottle use-life, with data collection ongoing. 100% FAIL results at and below the MRC and PASS results above the MRC. A conservative 3-month open bottle use-life is claimed.
    Comparative Sensitivity and SpecificityComparative sensitivity and specificity of 1 (implicitly 100% agreement with predicate).Comparative sensitivity of 100% and comparative specificity of 100% was found for both devices (QwikCheck and predicate). There was 100% overall agreement with values for prepared samples in both cases.
    Analytic Specificity - BioburdenNegative response at concentrations at or below MRC, positive responses at higher concentrations (when challenged with bioburden).100% FAIL results at and below the MRC and PASS results above the MRC even with the presence of fetal bovine serum (bioburden).
    Analytic Specificity – Other Disinfectants100% FAIL results for all conditions tested when exposed to other disinfectants (i.e., not reacting to other disinfectants).100% FAIL results for all conditions tested when exposed to hydrogen peroxide and sodium hypochlorite, indicating no false positives due to other disinfectants.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test set within the "Summary of Non-Clinical Performance Data." It indicates that the performance was evaluated "in a range of concentrations below, at, and above the MRC" and that "data collection [is] ongoing" for stability studies. The provenance of the data is not specified (e.g., country of origin). The studies are non-clinical, meaning they were conducted in a laboratory setting. The data appears to be prospective in nature, as tests were performed to evaluate the device's performance under various conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of human experts to establish ground truth for the test set. Given that this is a chemical indicator designed for objective colorimetric assessment of peracetic acid concentration, the ground truth would likely be established through precise chemical analytical methods (e.g., spectrophotometry, titration) to determine the exact PAA concentration in the prepared solutions.

    4. Adjudication Method for the Test Set

    No adjudication method for the test set is mentioned, as human interpretation and multi-reader review are not applicable to the evaluation of a chemical indicator. The device's output (color change indicating PASS/FAIL) is compared directly against the known chemical concentration.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging or other human-interpreted medical devices where the variability of human perception is a factor. The QwikCheck Chemical Indicator is a passive device with a direct chemical reaction indicating a result.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies conducted are standalone evaluations of the chemical indicator's performance. The "device performance" refers to the indicator's ability to accurately change color (PASS/FAIL) in response to specific PAA concentrations, without human intervention in determining the PAA concentration itself (which is the ground truth from chemical analysis). The "read time" and "blot time" mentioned in the Instructions for Use Validation refer to user interaction, but the core performance being validated is the chemical reaction of the indicator.

    7. The Type of Ground Truth Used

    The ground truth used for performance evaluation is based on known chemical concentrations of peracetic acid in TD-12 Solution. These concentrations are precisely prepared "below, at, and above the MRC" (1750 ppm). For example, a solution known to be below 1750 ppm should result in a "FAIL" indication, and a solution known to be above 1750 ppm should result in a "PASS" indication.

    8. The Sample Size for the Training Set

    The document does not specify a "training set" or "sample size for the training set." Chemical indicators are typically developed through chemical formulation and optimization, rather than machine learning algorithms that require distinct training sets. The studies described are performance validation studies, not training exercises.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a training set for an algorithm, the concept of establishing ground truth for a training set is not applicable here. The ground truth for validating the device's accuracy would have been established through analytical chemistry methods to create solutions with verified concentrations of peracetic acid.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120652

    Validate with FDA (Live)

    Date Cleared
    2012-10-12

    (224 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RAPICIDE OPA-28 Test Strip is a chemical indicator for use in determining whether the concentration of ortho-phthalaldehyde, the active ingredient in RAPICIDE OPA-28 Solution, is above or below the established Minimum Recommended Concentration (MRC) of 0.35%.

    Device Description

    The device is a qualitative, single use, reagent test strip made up of a 0.40 inch square indicator pad that has been chemically treated to detect OPA. The pad is affixed to one end of a 3.25 inch by 0.40 inch white opaque polystyrene strip.

    AI/ML Overview

    The SERIM Research Corporation RAPICIDE OPA-28 Test Strip is a chemical indicator designed to determine if the concentration of ortho-phthalaldehyde (OPA), the active ingredient in RAPICIDE OPA-28 Solution, is above or below its established Minimum Recommended Concentration (MRC) of 0.35%.

    Here's an analysis of its acceptance criteria and supporting studies:

    1. Acceptance Criteria and Reported Device Performance

    The provided document indicates the device must reliably indicate if the OPA concentration is above or below the MRC of 0.35%.
    The clinical study demonstrated that the device met these criteria through comparative performance with a predicate device.

    Acceptance CriteriaReported Device Performance
    Reliable indication of OPA concentration (above or below MRC of 0.35% in RAPICIDE OPA-28).The RAPICIDE OPA-28 Test Strips were evaluated in OPA solutions of Rapicide OPA-28 at the MRC and above in blind studies. The performance was found to be substantially equivalent to the predicate device, Serim® DISINTEK™ OPA Test Strips (which measure potency in Cidex OPA solutions at the MEC and above). The study concludes that the test strip is "safe and effective in monitoring the OPA concentration in Rapicide OPA-28 Disinfectant Solution."

    2. Sample Size and Data Provenance

    The document does not explicitly state the exact sample size (number of test strips or OPA solution samples) used for the blind studies.
    The data provenance is not specified regarding country of origin or whether it was retrospective or prospective. Given the nature of a 510(k) submission for a diagnostic device, it is highly likely that the studies were prospective, conducted at the manufacturer's facilities.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not specify the number of experts used or their qualifications for establishing ground truth. For chemical indicators like this, the "ground truth" would likely be established through analytical chemistry methods (e.g., spectrophotometry or titration) to precisely determine the OPA concentration in the test solutions, rather than human interpretation of the strips themselves as the primary ground truth.

    4. Adjudication Method

    The document does not describe an adjudication method. For a qualitative chemical indicator where the "ground truth" is a defined chemical concentration, judgment calls by multiple experts with an adjudication method are typically not required. The comparison is against an analytical standard.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned. This type of study is typically relevant for interpretative devices where human variability in reading results is a significant factor. For a chemical indicator, the reading is expected to be relatively straightforward (color change), reducing the need for such a complex reader study. The study focused on the performance of the device itself rather than human improvement with AI assistance.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The "Performance" section explicitly states that "The performance of the RAPICIDE OPA-28 Test Strips was evaluated in OPA solutions of Rapicide OPA-28 at the MRC and above in blind studies." This evaluation determines the device's ability to accurately indicate the OPA concentration without human-in-the-loop assistance in the interpretation of the color change.

    7. Type of Ground Truth Used

    The ground truth used would be analytical determination of OPA concentration. The study involved evaluating the test strips in "OPA solutions of Rapicide OPA-28 at the MRC and above." This implies that the OPA concentrations in these solutions were precisely known through validated analytical chemistry methods, serving as the objective ground truth against which the test strip's performance was measured.

    8. Sample Size for the Training Set

    The document does not provide information about a "training set" or its sample size. This is common for chemical indicators of this type, which are typically developed based on chemical principles rather than machine learning algorithms requiring explicit training data. The development process would involve formulation and optimization of the reagents to achieve the desired color change at the target concentration.

    9. How Ground Truth for the Training Set Was Established

    As there is no mention of a "training set" in the context of machine learning, there is no corresponding ground truth establishment method described. The development of the test strip would rely on chemical formulation and empirical testing against analytically prepared OPA solutions with known concentrations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092346

    Validate with FDA (Live)

    Date Cleared
    2010-01-07

    (156 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serim® D-CIDE GTA 1.5% Test Strip is a chemical indicator for use in determining whether the concentration of glutaraldehyde, the active ingredient in Rapicide HLD and Sterilant (Medivators Reprocessing Systems), is above or at/below the minimum effective concentration (MEC) of 1.5% GTA established for this solution.

    Device Description

    The device is a qualitative, single use, reagent test strip made up of a 0.20 inch square indicator pad that has been chemically treated to detect GTA. The pad is affixed to one end of a 3.25 inch by 0.20 inch white opaque polystyrene strip.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Serim® D-CIDE GTA 1.5% Test Strip, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    The Serim® D-CIDE GTA 1.5% Test Strip is a chemical indicator for use in determining whether the concentration of glutaraldehyde, the active ingredient in Rapicide HLD and Sterilant (Medivators Reprocessing Systems), is above or at/below the minimum effective concentration (MEC) of 1.5% GTA established for this solution.The performance of the Serim D-CIDE GTA 1.5% Test Strips was evaluated in split samples blind studies and compared to test results obtained with 3M Comply Cold Sterilog 1.5% GTA Test Strips. The performance of the Serim D-CIDE GTA 1.5% Test Strips is substantially equivalent to the predicate device, 3M Comply Cold Sterilog 1.5% GTA Test Strips, for determining if glutaraldehyde concentration is above or at/below the MEC of 1.5%. Specifically, when glutaraldehyde concentration is above 1.5% (MEC), enough base is produced to result in an irreversible and distinct purple color change on the indicator pad. When glutaraldehyde concentration is at or approaching the MEC level, the indicator pad will display distinct white areas.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated in the provided document. The document mentions "split samples blind studies" but does not give a numeric count of samples.
    • Data Provenance: Not explicitly stated. There's no mention of country of origin or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document. The method of establishing ground truth relies on the chemical reaction and visual color change of the test strip, and comparison to a predicate device, rather than expert interpretation of complex data.

    4. Adjudication Method for the Test Set:

    • This information is not provided as the ground truth appears to be based on the chemical reaction and visual color change, not on subjective assessment that would require adjudication. The study involved "blind studies" and comparison to a predicate device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a chemical indicator, not an AI-assisted diagnostic tool. Its performance is based on a chemical reaction producing a visual color change, not on human interpretation that could be augmented by AI.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:

    • Yes, in essence, a standalone performance was done. The device itself, a chemical test strip, operates independently to produce a visual result based on the glutaraldehyde concentration. The "performance" described is the inherent function of the test strip (a chemical reaction) and its visual output, without a human "in the loop" for interpretation beyond simply observing the color change.

    7. The Type of Ground Truth Used:

    • The ground truth is established by the chemical concentration of glutaraldehyde in the solutions being tested. The device is designed to visually indicate whether this concentration is above or at/below the MEC of 1.5%, based on a specific chemical reaction. The validation relies on comparing this visual indication against the known concentration of glutaraldehyde and against the performance of a predicate device.

    8. The Sample Size for the Training Set:

    • This information is not provided in the document. The device is a chemical indicator, not a machine learning algorithm that requires a training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established:

    • This information is not applicable/provided. As mentioned, the device is a chemical indicator. Its development and validation likely involved chemical formulations and testing to ensure the desired color change reaction at specific glutaraldehyde concentrations, rather than a "training set" with established ground truth as understood in AI/machine learning contexts.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083295

    Validate with FDA (Live)

    Date Cleared
    2009-03-16

    (126 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serim® Disintek GTA 2.1% Test Strip is a chemical indicator for use in determining whether the concentration of glutaraldehyde, the active ingredient in CIDEXPLUS 28 Day Solution, is above or below the minimum effective concentration (MEC) established for CIDEXPLUS 28 Day Solution.

    Device Description

    The device is a qualitative, single use, reagent test strip made up of a 0.40-inch square reagent pad that has been chemically treated to detect glutaraldehyde in CIDEXPLUS 28 Day solutions. The pad is affixed to one end of a 3.25 inch by 0.40-inch white opaque polystyrene strip.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for the Serim® DISINTEK™ GTA 2.1%. Instead, it refers to performance in comparison to a predicate device. The core functionality is to indicate whether the glutaraldehyde concentration is above or below the Minimum Effective Concentration (MEC) of 2.1%. The description focuses on the qualitative visual response.

    Acceptance Criterion (Implicit)Reported Device Performance
    Reliable indication of glutaraldehyde concentration above MEC (2.1%)When glutaraldehyde concentration is above 2.1% (MEC), enough base is produced to result in an irreversible and distinct purple color change on the indicator pad. Testing of freshly activated Cidex® Plus solutions always results in a solid purple color.
    Reliable indication of glutaraldehyde concentration at or approaching MEC (2.1%)When the glutaraldehyde concentration is at or approaching the MEC level, the base is neutralized by sodium bisulfite, and the indicator pad will display a distinctly blotched yellow and purplish appearance.
    Reliable indication of glutaraldehyde concentration below MEC (approaching 1.5%)As the glutaraldehyde concentration approaches 1.5%, the indicator pad will appear mostly yellow. The device will reliably indicate if the glutaraldehyde concentration is above or below the MEC level of 2.1%.
    Substantial equivalence to predicate deviceThe performance of the Serim Disintek GTA 2.1% Test Strips was evaluated in split samples blind studies and compared to test results obtained with CIDEXPLUS Solution Test Strips. The performance is substantially equivalent to the predicate device.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document states: "The performance of the Serim Disintek GTA 2.1% Test Strips was evaluated in split samples blind studies..."

    • Sample Size: The exact sample size for the test set is not specified in the provided text.
    • Data Provenance: The text does not specify the country of origin of the data. The study was a "blind study," which implies it was conducted to objectively assess the device's performance. It is a prospective study, as it's an evaluation of the device's performance under controlled conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the "split samples blind studies". The ground truth appears to be based on the actual glutaraldehyde concentration (e.g., above 2.1%, at/approaching 2.1%, approaching 1.5%) rather than expert interpretation of the test strips.

    4. Adjudication Method for the Test Set

    The document does not explicitly mention an adjudication method (like 2+1, 3+1, none) for the test set. Given the nature of the device (a chemical indicator with a visual color change), the "ground truth" seems to be determined by the objective chemical concentration rather than requiring expert adjudication of reader interpretations, although visual interpretation is involved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

    • Was an MRMC study done? No, not in the sense of comparing human readers' performance with and without AI assistance. This device is a manual chemical indicator, not an AI-assisted diagnostic tool for Human-in-the-Loop scenarios. The study compared the new device's performance against a predicate device.
    • Effect Size: Not applicable as it's not an AI-assisted MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not directly applicable to the device as presented. The device is a "qualitative, single use, reagent test strip" that relies on a visual interpretation by a human observer. There is no "algorithm only" performance because the device itself is a passive chemical indicator that requires human observation for its intended use. The performance evaluation would inherently include the human interpretation component, or at least a comparison against a known chemical concentration.

    7. The Type of Ground Truth Used

    The ground truth used for the study appears to be the actual glutaraldehyde concentration in the CIDEXPLUS 28 Day solutions. The device is designed to indicate whether this concentration is "above or below the minimum effective concentration (MEC) established for CIDEXPLUS 28 Day Solution." This implies that the solutions used in the "split samples blind studies" had known glutaraldehyde concentrations.

    8. The Sample Size for the Training Set

    The document does not mention a training set or any machine learning/AI components. Therefore, a sample size for a training set is not applicable. The device relies on chemical reactions, not on data-driven models that require training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for this chemical indicator device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082352

    Validate with FDA (Live)

    Date Cleared
    2008-09-25

    (40 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serim® DISINTEK™ OPA Test Strip is a chemical indicator for use in determining whether the concentration of orthophthalaldehyde, the active ingredient in Metricidc OPA Plus Solution, is above or below the minimum effective concentration (MEC) established for Metricide OPA Plus Solution.

    Device Description

    The device is a qualitative, single use, reagent test strip made up of a 0.40 inch square test pad that has been chemically treated to detect OPA in Metricide OPA Plus Solution. The pad is affixed to one end of a 3.25 inch by 0.40 inch white opaque polystyrene strip.

    AI/ML Overview

    The Serim® DISINTEK™ OPA Test Strip is a chemical indicator designed to determine if the concentration of ortho-phthalaldehyde (OPA) in Metricide OPA Plus Solution is above or below its minimum effective concentration (MEC) of 0.3%.

    Here's an analysis of the provided information regarding its acceptance criteria and the study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly tied to demonstrating substantial equivalence to a predicate device, specifically the Metricide OPA Plus Solution Test Strip, by accurately indicating OPA concentration relative to the 0.3% MEC.

    Acceptance Criteria (Implicit)Reported Device Performance
    Reliable indication of OPA concentration above 0.3% MEC (green color formation).The device reliably indicates if the OPA concentration is above or below the MEC level of 0.3% OPA. OPA levels above 0.3% react with a second chemical to form a colored compound, resulting in a green color.
    Reliable indication of OPA concentration below 0.3% MEC (colorless product formation).The device reliably indicates if the OPA concentration is above or below the MEC level of 0.3% OPA. A concentration equivalent to 0.3% OPA (MEC) reacts with the OPA to form a colorless product (implying no green color if below MEC).
    Performance substantially equivalent to the predicate device (Metricide OPA Plus Solution Test Strips).The performance of the Serim DISINTEK OPA Test Strips was evaluated in split samples blind studies and compared to test results obtained with Metricide OPA Plus Solution Test Strips. The performance of the Serim DISINTEK OPA Test Strips is substantially equivalent to the predicate device, Metricide OPA Plus Solution Test Strips.
    Does not raise any new safety or effectiveness issues compared to the predicate device.The Serim DISINTEK OPA Test Strips do not raise any new safety or effectiveness issues.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the numerical sample size used for the test set. It mentions "split samples blind studies" but provides no specific numbers for the samples tested.
    • Data Provenance: Not explicitly stated. Given the context of a 510(k) submission to the US FDA, it is highly likely that the data was generated in the United States, but this is not explicitly confirmed. The studies appear to be prospective in nature, as they involve actively evaluating the performance of the new device in comparison to a predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The method for establishing ground truth for the test set is implied by the comparison to a predicate device and chemical reactions, rather than expert interpretation.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. The nature of a chemical indicator test, where the result is a color change based on a chemical reaction, typically does not involve human adjudication in the same way an imaging diagnostic might. The output is a direct chemical reading.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was it done? No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) where the AI assists in the reading process. The Serim® DISINTEK™ OPA Test Strip is a chemical indicator that produces a direct, objective result (color change), not a complex interpretation task for human readers.
    • Effect size of how much human readers improve with AI vs. without AI assistance: Not applicable, as no MRMC study was conducted.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    • Was it done? Yes, in essence, the performance evaluation described is a standalone study of the device. The "split samples blind studies" evaluated the performance of the Serim DISINTEK OPA Test Strips on their own merits and then compared the results directly to the predicate device. The device's mechanism is a chemical reaction leading to a direct color change, which is an inherent "algorithm" of the test strip itself. The outcome is not subject to human interpretation in the same way as an imaging algorithm might be reviewed by a human.

    7. Type of Ground Truth Used

    The ground truth used is based on chemical concentration levels of OPA relative to the 0.3% MEC. This is established by:

    • The known chemical properties and reactions designed into the test strip.
    • Comparison to the performance of an already legally marketed and accepted predicate device (Metricide OPA Plus Solution Test Strips) which also relies on chemical determination of OPA concentration.
    • The chemical principle: "A chemical at a concentration equivalent to 0.3% OPA (MEC) reacts with the OPA to form a colorless product. A second reaction then occurs in which OPA at levels above 0.3% reacts with a second chemical to form a colored compound, resulting in green color." This defines the objective ground truth against which the strip's response is measured.

    8. Sample Size for the Training Set

    This information is not provided. As this is a chemical indicator and not a machine learning or AI-driven device in the modern sense (which typically requires large training sets), the concept of a "training set" for an algorithm isn't directly applicable in the same way. The device's "training" would be through its chemical formulation and manufacturing tolerances, which are established through lab development and quality control, not typically a large data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable in the context of a "training set" for a traditional algorithm. The "ground truth" for the device's design and chemical formulation would have been established through:

    • Understanding the specific chemical reactions of OPA.
    • Titration or analytical chemistry methods to precisely determine OPA concentrations at and around the 0.3% MEC.
    • Empirical testing during product development to refine the chemical composition of the test strip to achieve the desired sensitivity and specificity for the color change at the MEC threshold.
    Ask a Question

    Ask a specific question about this device

    K Number
    K081370

    Validate with FDA (Live)

    Date Cleared
    2008-07-11

    (57 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serim® DISINTEK™ OPA Test Strip is a chemical indicator for use in determining whether the concentration of ortho-phthalaldehyde, the active ingredient in Cidex® OPA Solution, is above or below the minimum effective concentration (MEC) established for Cidex OPA Solution.

    Device Description

    The device is a qualitative, single use, reagent test strip made up of a 0.40 inch square test pad that has been chemically treated to detect OPA in Cidex OPA Solution. The pad is affixed to one end of a 3.25 inch by 0.40 inch white opaque polystyrene strip.

    AI/ML Overview

    The provided text describes the Serim DISINTEK OPA Test Strips and the study conducted to demonstrate its substantial equivalence to a predicate device. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" for this device are implicitly defined by the performance of the predicate device, the Cidex OPA Solution Test Strip, and the ability to accurately indicate the OPA concentration above or below the Minimum Effective Concentration (MEC). The device's performance is reported in comparison to this predicate.

    Acceptance Criteria & Reported Device Performance for Serim DISINTEK OPA Test Strips

    Acceptance Criteria (Implied)Reported Device Performance and Comparison
    Accuracy: Reliably indicate if OPA concentration is above or below the MEC of 0.3%.The device "will reliably indicate if the OPA concentration is above or below the MEC level of 0.3% OPA."
    Comparison to Predicate: Performance should be substantially equivalent to Cidex® OPA Solution Test Strip.Performance was "evaluated in split samples blind studies and compared to test results obtained with Cidex OPA Solution Test Strips."
    "The performance of the Serim DISINTEK OPA Test Strips is substantially equivalent to the predicate device, Cidex OPA Solution Test Strips."
    Intended Use Equivalence: Have the same intended use as the predicate device (measuring OPA potency in Cidex OPA Solution, above/below 0.3% MEC)."The Serim DISINTEK OPA Test Strips have the same intended use as the predicate device."
    Safety and Effectiveness: Not raise any new safety or effectiveness issues compared to the predicate."The Serim DISINTEK OPA Test Strips do not raise any new safety or effectiveness issues."
    Mechanism of Action: React with OPA to show color change based on concentration relative to MEC.The device contains chemicals that react with OPA; a "first step involves a reaction in which a chemical at a concentration equivalent to 0.3% OPA (MEC) reacts with the OPA to form a colorless product. A second reaction then occurs in which OPA at levels above 0.3% reacts with a second chemical to form a colored compound, resulting in green color."
    Physical Characteristics: Test pad size allows for easy interpretation of color change."The test pad size of 0.4" x 0.4" allows for easy interpretation of the change in color."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document states "split samples blind studies" were used, but it does not specify the exact sample size used for the test set.
    • Data Provenance: Not explicitly stated, but typical for device submissions in the US, the studies would likely be conducted in the country of origin of the submitter (Serim Research Corporation, Elkhart IN, USA). The study is described as performance evaluation, implying it was prospective in nature, as new data was generated using the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of "experts" in establishing ground truth in the context of human readers or medical professionals for this type of chemical indicator. The ground truth for chemical concentration is typically established by analytical chemistry methods rather than expert consensus on visual interpretation.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method. For a chemical indicator, "adjudication" in the sense of reconciling differing expert opinions would not typically apply. The "ground truth" for OPA concentration would be determined by quantitative chemical analysis.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This device is a chemical indicator, not an AI-assisted diagnostic tool for human readers. Therefore, the concept of human readers improving with or without AI assistance does not apply.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    The device itself is a passive chemical indicator. Its performance is inherent to its chemical reaction properties. While human interpretation of the color change is necessary, the "test strip" itself acts as a "standalone" indicator in the sense that its chemical reaction and color formation occur independently of human judgment. The study evaluated the device's performance directly by comparing its indications to those of the predicate device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for this chemical indicator would be the actual chemical concentration of ortho-phthalaldehyde (OPA) in the Cidex OPA Solution, determined by a validated analytical method (e.g., spectrophotometry, titration, or other quantitative chemical analysis). The device's color change is then compared against this known OPA concentration to determine its accuracy in indicating above or below the MEC of 0.3%.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" as this device is a chemical indicator and not a machine learning or AI model. The concept of training data does not apply here.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for this chemical indicator, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080712

    Validate with FDA (Live)

    Date Cleared
    2008-05-06

    (54 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SERIM® HiSENSE ULTRA 0.1 TM Test Strips provide a convenient means for indicating the concentration of residual chlorine (chlorine bleach) detected in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment. HiSENSE ULTRA 0.1 also provides a quick and convenient means for indicating low levels of total chlorine (chloramines/free chlorine) in water used to prepare dialysate.

    Device Description

    The device is a semi-quantitative or qualitative, single use, reagent test strip made up of a 0.20 inch square reagent pad that has been chemically treated to detect total chlorine (free chlorine/chloramines) in water. The pad is affixed to one end of a 3.25 inch by 0.20 inch white opaque polystyrene strip.

    AI/ML Overview

    This document is a 510(k) summary for the SERIM® HiSENSE ULTRA 0.1™ Test for Total Chlorine. It describes the device, its intended use, technological characteristics, and performance evaluation.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Detect total chlorine concentrationsReliably detect concentrations of 0.1 ppm total chlorine. Device detects concentrations between 0 and 3 ppm.
    Substantial equivalence to predicate device (SteriChek Sensitive Total Chlorine Reagent Strips)Performance is substantially equivalent to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific numerical sample size for the test set. It mentions: "The performance of the Serim HiSENSE ULTRA 0.1 Test Strips were evaluated using water samples in which either sodium hypochlorite or chloramines were added to give a range of total chlorine (free chlorine/chloramines) levels."

    • Sample Size: Not explicitly stated as a number of samples. The evaluation used "water samples."
    • Data Provenance: The water samples were prepared by adding sodium hypochlorite or chloramines to achieve varying total chlorine levels. While the country of origin isn't specified, it's implied to be an in-house laboratory evaluation based on the nature of the test. It is a prospective evaluation, as samples were prepared for the purpose of testing the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The ground truth for the prepared water samples would typically be established by precise chemical measurements (e.g., using a reference method or analytical chemistry techniques) rather than human experts interpreting results.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given the nature of a semi-quantitative chemical test strip, the adjudication method would likely involve comparing the color change on the strip to a color chart provided with the device, and subsequently comparing that reading to the known concentration of the prepared water sample.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a manual chemical test strip, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device requires human interpretation of the color change on the test strip against a color chart. It is not an automated algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the test set was established by prepared water samples with known concentrations of total chlorine. These known concentrations would have been determined through precise chemical methods, making it an analytical truth.

    8. The Sample Size for the Training Set

    This information is not provided in the document. As a chemical test strip, it's unlikely to have a "training set" in the sense of machine learning algorithms. The design and calibration of the test strip's chemical formulation and color chart would be based on chemical principles and experimental validation, but the term "training set" is not applicable here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8. The "ground truth" for developing such a device would reside in the chemical reactions and the correlation of color intensity to chlorine concentration, established through laboratory experiments and chemical analysis during the product's R&D phase.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052388

    Validate with FDA (Live)

    Date Cleared
    2005-12-01

    (92 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Serim ® Peracetic Acid Chemical Indicator test strips are designed for use in a STERIS SYSTEM 1® Processor with STERIS 20™ Sterilant to detect peracetic acid concentration in the use solution.

    Device Description

    The device is a qualitative, single use, reagent test strip made up of a 0.20 inch square purple reagent pad that has been chemically treated to detect the active ingredient in the Steris 20™ Sterilant® when used in the STERIS SYSTEM 1® Processor™. The pad is affixed to one end of a 3.25 inch by 0.20 inch white opaque polystyrene strip.

    The Serim* Peracetic Acid Chemical Indicator is configured as a paper pad on a plastic strip support. A dve in the chemical indicator pad is oxidized (bleached) by the use solution. The extent of bleaching indicates whether a sufficient concentration active agent (peracetic acid) was present. At the end of the processing evele the user compares the reacted indicator to a color chart composed of two color blocks. The "START" color block is similar to the unreacted indicator strip. The =COMPLETE color block is similar to an indicator strip that has been exposed to peracetic acid at or above a concentration of 1500 ppm.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the Serim® Peracetic Acid Chemical Indicator. This document focuses on establishing substantial equivalence to a previously cleared device, not on presenting a standalone study with detailed acceptance criteria and performance data as one would expect for a new, non-equivalent device or an AI/software as a medical device (SaMD).

    Therefore, many of the requested points cannot be directly addressed from the provided text because the device, a chemical indicator, is not an AI/Software device, nor is the document an MRMC or standalone study report. It's a regulatory submission for a physical device.

    However, I can extract information relevant to the intent of the questions where possible, and clearly state when information is not available due to the nature of the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission declares substantial equivalence to the STERIS PROCESS™ Chemical Monitoring Strips (K921559). The "acceptance criteria" for this submission appear to be the demonstrated identical nature and functionality to the predicate device.

    Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance
    Intended Use: Detects peracetic acid concentration in the use solution of a STERIS SYSTEM 1® Processor with STERIS 20™ Sterilant.Intended Use: The Serim® Peracetic Acid Chemical Indicator test strips are designed for use in a STERIS SYSTEM 1® Processor with STERIS 20™ Sterilant to detect peracetic acid concentration in the use solution. (Matches)
    Technological Characteristics: Identical to the STERIS PROCESS™ Chemical Monitoring strip (K921559) in composition and mechanism (dye oxidized/bleached by peracetic acid, color change compared to a chart).Technological Characteristics: "The Serim® Peracetic Acid Chemical Indicator is identical to the STERIS PROCESS™ Chemical Monitoring strip (manufactured by Serim Research Corp. for Steris Corporation) that has been previously cleared by the FDA (K921559)." The device uses a color change mechanism: dye in a pad is oxidized (bleached) by peracetic acid; user compares reacted indicator to a color chart with "START" and "COMPLETE" color blocks, where "COMPLETE" indicates exposure to ≥ 1500 ppm peracetic acid. (Matches predicate)
    Performance: Provides an indication of sufficient peracetic acid concentration (at or above 1500 ppm).Performance: The "COMPLETE" color block of the color chart corresponds to an indicator strip exposed to peracetic acid "at or above a concentration of 1500 ppm." (Implies equivalent performance to predicate)

    2. Sample size used for the test set and the data provenance

    Not applicable. This is a 510(k) for a physical chemical indicator, not a study involving a test set of data (like medical images or clinical records). The claim is about manufacturing and functional identity to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No "ground truth" as typically understood in a software or AI device study is established for a test set in this 510(k) submission. The performance relies on the chemical reaction and color change.

    4. Adjudication method for the test set

    Not applicable. No test set requiring adjudication in this context.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a chemical indicator, not an AI or software device, and no MRMC study is mentioned or relevant to this type of submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device is a standalone chemical indicator that produces a visual result. It requires human interpretation to compare the reacted strip to a color chart. There is no "algorithm only" component. The performance discussed is the chemical reaction itself, not a digital algorithm.

    7. The type of ground truth used

    The "ground truth" for this chemical indicator's functionality is the known chemical concentration of peracetic acid. The device is designed to visually indicate if this concentration is at or above a specific threshold (1500 ppm).

    8. The sample size for the training set

    Not applicable. This device is a chemical indicator, not an AI or machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set is used for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2