Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K202644

    Validate with FDA (Live)

    Device Name
    Acetaminophen
    Date Cleared
    2022-02-18

    (525 days)

    Product Code
    Regulation Number
    862.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acetaminophen assay is used for the quantitative determinophen in human serum or plasma on the ARCHITECT c Systems.

    The Acetaminophen assay is to be used as an aid in the diagnosis and treatment of acetaminophen overdose toxicity.

    Device Description

    The Acetaminophen assay is an enzymatic, spectrophotometric assay for the measurement of acetaminophen concentration in human serum and plasma. The assay consists two working reagents, an enzyme reagent and a color reagent.

    The enzyme reagent contains aryl acylamidase, which cleaves the amide bond of acetaminophen, forming p-aminophenol which then reacts with the 2,5-dimethylphenol (contained the color reagent) in the presence of manganese. The product of that reaction causes increased absorbance at 660 nm which is directly proportional to the acetaminophen concentration in the sample.

    Testing is performed on the ARCHITECT c8000 clinical chemistry analyzers in conjunction with a calibrator (510(k) exempt) which is provided separately.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an Acetaminophen assay (K202644). The document details various performance studies conducted to demonstrate the device's substantial equivalence to a legally marketed predicate device.

    Here's an analysis of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Stated or Implied)Reported Device Performance
    Precision (Within-Run %CV)Not explicitly stated as acceptance criteria, but typically within acceptable laboratory limits for clinical assays. For example, for Control Level 1, %CV (0.9) is excellent; for Panel A, %CV (5.6) is higher but likely acceptable for low concentrations.Control Level 1: Mean 15 µg/mL, SD 0.1, %CV 0.9Control Level 2: Mean 73 µg/mL, SD 0.5, %CV 0.6Control Level 3: Mean 227 µg/mL, SD 0.8, %CV 0.3Panel A: Mean 5 µg/mL, SD 0.3, %CV 5.6Panel B: Mean 51 µg/mL, SD 0.3, %CV 0.6Panel C: Mean 84 µg/mL, SD 0.4, %CV 0.4Panel D: Mean 278 µg/mL, SD 1.0, %CV 0.4Panel E: Mean 362 µg/mL, SD 1.6, %CV 0.4
    Precision (Within-Laboratory %CV)Not explicitly stated as acceptance criteria, but typically within acceptable laboratory limits.Control Level 1: SD 0.2, %CV 1.3Control Level 2: SD 0.6, %CV 0.8Control Level 3: SD 1.3, %CV 0.6Panel A: SD 0.5, %CV 9.1Panel B: SD 0.4, %CV 0.7Panel C: SD 0.6, %CV 0.7Panel D: SD 2.0, %CV 0.7Panel E: SD 2.6, %CV 0.7
    Reproducibility (%CV)Not explicitly stated as acceptance criteria, but typically within acceptable laboratory limits.Control 1: Mean 14 µg/mL, SD 0.5, %CV 3.2Control 2: Mean 68 µg/mL, SD 0.6, %CV 0.9Control 3: Mean 208 µg/mL, SD 1.6, %CV 0.8Panel: Mean 163 µg/mL, SD 1.2, %CV 0.7
    Limit of Blank (LoB)Not explicitly stated, but should ideally be very low.0 µg/mL (0 µmol/L)
    Limit of Detection (LoD)The sponsor chose to use 1 µg/mL (7 µmol/L) for reporting purposes.Scientific LoD: 0.2 µg/mL (1.3 µmol/L)Reporting LoD: 1 µg/mL (7 µmol/L)
    Limit of Quantitation (LoQ)The sponsor chose to use 3 µg/mL (20 µmol/L) for reporting purposes. This is the lower end of the analytical measuring interval.Scientific LoQ: 1.9 µg/mL (12.6 µmol/L)Reporting LoQ (lower end of AMI): 3 µg/mL (20 µmol/L)
    Linearity/Assay RangeDevice demonstrated linearity across the range of 0 to 386 µg/mL, which spans the analytical measuring interval of 3 to 377 µg/mL. Implicit acceptance is that the assay is linear across its claimed analytical measuring interval.Linear across 0 to 386 µg/mL. Analytical Measuring Interval: 3 to 377 µg/mL (20 to 2496 µmol/L).
    Analytical Specificity (Interference)Interference within ± 7.5% for acetaminophen samples > 20 µg/mL, ORwithin ± 1.50 µg/mL for acetaminophen samples < 20 µg/mL.Specific endogenous and exogenous substances (e.g., conjugated bilirubin ≤ 14 mg/dL, hemoglobin ≤ 570 mg/dL, N-Acetylcysteine ≤ 1663 mg/L) tested and demonstrated to meet the acceptance criteria at the specified interferent levels.
    Method Comparison (Correlation)Not explicitly stated, but the predicate device has a strong correlation (0.9993) with the candidate. Implicit acceptance would be a high correlation coefficient (e.g., >0.975 or >0.98), slope close to 1, and intercept close to 0.Correlation Coefficient: 0.9993Slope: 1.042Intercept: -0.034(Comparing 3.50 - 356.26 µg/mL (predicate) to 3.58 - 375.28 µg/mL (candidate))
    Matrix ComparisonRecovered within ± 7.5% for values ≥ 20 µg/mL, ORwithin ± 1.50 µg/mL for values < 20 µg/mL.Serum, lithium heparin, lithium heparin mechanical separator, lithium heparin (separator tube), serum (separator tube), and sodium heparin tube types were found suitable for analysis.
    Automated/Manual DilutionNot explicitly stated, but results showed % difference of 1.5% to 5.3%, implying these met specific internal criteria for analytical accuracy of diluted samples. The range of 1.5% to 5.3% is generally considered excellent for dilution recovery.Demonstrated % difference results of 1.5% to 5.3% when measuring samples above the analytical measuring interval by 1:10 autodilution vs. manual dilution methods.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision: 5 human serum panels and 3 control levels. Each assayed in a minimum of 2 replicates, twice per day on 20 days. N=120 for controls and panels.
    • Reproducibility: 1 human serum panel and 3 control levels. Each assayed in a minimum of 3 replicates at 2 separate times per day on 5 different days. N=146-150 for controls and panel.
    • Analytical Sensitivity (LoB, LoD, LoQ): Zero analyte samples and low analyte samples tested in replicates of 10.
    • Linearity/Assay Reportable Range: One unique sample set, 13 sample levels, tested in a minimum of 2 replicates.
    • Analytical Specificity (Interference): Test and reference samples tested in a minimum of 12 replicates.
    • Method Comparison: 119 patient specimens (114 unaltered native samples, 5 contrived).
    • Matrix Comparison: 78 matched specimen sets.
    • Automated/Manual Dilution: 5 samples with high acetaminophen concentrations, tested in replicates of 42.

    Data Provenance: The document does not explicitly state the country of origin of the data. The studies appear to be prospective as they were specifically designed and executed to evaluate the performance of the device for this submission (e.g., "A study was performed based on guidance...", "Testing was conducted according to CLSI EP05-A3...").

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number and qualifications of experts) is typically associated with studies involving subjective interpretation, such as imaging or pathology studies. For an in vitro diagnostic (IVD) like this Acetaminophen assay, the "ground truth" is generally established using reference methods or gravimetric preparation of standards, rather than expert consensus on patient data.

    • For Value Assignment: The Acetaminophen Calibrator is prepared gravimetrically from USP grade reference acetaminophen material to a specific concentration, and verified against a master calibrator (internal reference standard). This is the analytical "ground truth" for calibration.
    • For Method Comparison: The predicate device is considered the "reference" for comparison, not a "ground truth" established by experts.

    Therefore, the concept of "experts establishing ground truth" as you've defined it (e.g., radiologists) is not applicable to this type of IVD device performance study.

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple readers or experts provide independent assessments, and discrepancies need to be resolved. Since this is an IVD device measuring an analyte quantitatively, and the "ground truth" is established analytically (gravimetric, reference methods, or predicate device results), an adjudication method in the traditional sense is not applicable. The studies involve quantitative measurements from instruments and comparisons to established analytical standards or methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic (IVD) device, specifically an assay for measuring acetaminophen levels, not an imaging or AI-assisted diagnostic tool that involves human readers. Therefore, the concept of "human readers improving with AI vs. without AI assistance" is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device itself (the Acetaminophen assay on the ARCHITECT c Systems) is a standalone system for quantitative measurement. The performance studies detailed (precision, linearity, method comparison, etc.) represent the standalone performance of the algorithm/reagent system as it processes samples on the automated platform without human intervention beyond sample loading and system operation. There is no "human-in-the-loop" component for interpretation of results in the direct sense; the device provides a quantitative number.

    7. The Type of Ground Truth Used

    The ground truth for the performance studies is established via:

    • Gravimetric Preparation/Reference Standards: For analytical performance characteristics like calibrator value assignment, LoB, LoD, LoQ, and linearity. For example, calibrators are "prepared gravimetrically from USP grade reference acetaminophen material."
    • Predicate Device Results: For the method comparison study, the results from the legally marketed predicate device (SEKURE Acetaminophen L3K Assay on Hitachi 717) serve as the comparative "reference" or "ground truth" against which the new device's performance is measured.
    • Controlled Samples: For interference studies, samples with known concentrations of acetaminophen and specific interferents are used.

    8. The Sample Size for the Training Set

    The document does not mention a training set because this is not a machine learning or artificial intelligence (AI) device that requires extensive supervised learning. This is an enzymatic spectrophotometric assay, which relies on chemical reactions and optical measurement principles, not statistical pattern recognition trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for this type of IVD device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182702

    Validate with FDA (Live)

    Date Cleared
    2019-04-17

    (202 days)

    Product Code
    Regulation Number
    862.1215
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the in vitro quantitative measurement of creatine kinase activity in serum and plasma on the SK500 Clinical Chemistry System. Measurements of creatine kinase are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

    Device Description

    The SEKURE Creatine Kinase Assay (CK Assay) is a spectrophotometric, coupled enzyme assay for the quantitative measurement of creatine kinase (CK) activity. The assay consists of two working reagents, a buffer solution (R1) and a substrate reagent (R2). The SEKURE CK Assay employs the reverse reaction of CK, to produce adenosine triphosphate (ATP). The reaction is coupled to hexokinase and G6PDH which consumes ATP to generate NADPH. The rate of NADPH formation is monitored at 340 nm and is directly proportional to CK activity. Testing is performed on the SK500 in conjunction with calibrator and controls which are provided separately.

    The SK500 Analyzer is manufactured as Clinical Chemistry Analyzer Tokyo Boeki Medisys Inc. Biolis 50i Superior. "SK500" is the Sekisui Diagnostics labelled name for the Tokyo Boeki Medisys Inc. Biolis 50i Superior instrument.

    AI/ML Overview

    This document describes the SEKURE Creatine Kinase Assay, an in vitro diagnostic device, and its performance study to demonstrate substantial equivalence to a predicate device.

    1. Acceptance Criteria and Reported Device Performance

    The device performance is evaluated against various analytical metrics. While explicit "acceptance criteria" for each study are not individually listed as pass/fail thresholds in a table, the document reports the results of these studies, implying that the observed performance met internal or regulatory expectations for demonstrating substantial equivalence. The predicate device's characteristics serve as an implicit benchmark for similarity.

    Here's a table summarizing the reported device performance, with implied acceptance based on the submission being cleared:

    Feature/StudyAcceptance Criteria (Implied)Reported Device Performance (SEKURE Creatine Kinase Assay)Predicate Device Performance (Creatine Kinase-SL Assay)
    Intended UseQuantitative measurement of CK activity in serum.Quantitative measurement of CK activity in serum and plasma on SK500.Quantitative measurement of CK activity in serum.
    MethodologyMust be similar to predicate.Colorimetric (NADPH), Enzymatic (coupled hexokinase-G6PD)Colorimetric (NADPH), Enzymatic (coupled hexokinase-G6PD)
    Specimen TypeSuitable for reported specimen types.Serum and lithium heparin plasma.Serum
    Measuring IntervalWithin a clinically relevant and acceptable range, similar to predicate.11-1500 U/L2-1500 U/L
    PrecisionAcceptable repeatability and within-laboratory variability (%CV, SD). CLSI EP05-A3 guidelines.See Section "Precision" table (e.g., Repeatability %CV ranging from 0.6-1.3%, Within Laboratory %CV ranging from 1.2-3.6%).Not explicitly detailed but assumed to meet similar standards.
    Limit of Blank (LoB)Must be low and clinically acceptable. CLSI EP17-A2 guidance.3 U/LNot explicitly detailed but assumed to meet similar standards.
    Limit of Detection (LoD)Must be low and clinically acceptable. CLSI EP17-A2 guidance.5 U/LNot explicitly detailed but assumed to meet similar standards.
    Limit of Quantitation (LoQ)Clinically acceptable level. CLSI EP17-A2 guidance.11 U/LNot explicitly detailed but assumed to meet similar standards.
    Linearity/Reportable RangeObserved mean values within ±10% or ±3 U/L of theoretical values. CLSI EP06-A guidance.Supported measuring range up to 1500 U/L for serum and plasma.2-1500 U/L
    Analytical Specificity (Interference)Significant interference < 10% difference from control. CLSI EP07-A2 guideline.No significant interference observed at or below specified concentrations (e.g., Conjugated Bilirubin 40 mg/dL, Hemoglobin 200 mg/dL).Not explicitly detailed but assumed to meet similar standards.
    Method Comparison (vs. Predicate)High correlation (near 1), low bias, slope near 1, small y-intercept. CLSI EP09-A3 guidance.Lot 1: y-Intercept 2.4 U/L, Corr. Coeff. 0.9998, Bias -2.7%, Slope 0.964. Lot 2: y-Intercept 2.3 U/L, Corr. Coeff. 0.9997, Bias -2.2%, Slope 0.970.N/A (this is the predicate for comparison)
    Matrix Comparison (Serum vs. Plasma)High correlation (near 1), low bias, slope near 1, small y-intercept.Lot 1: Slope 1.013, y-Intercept -1.8 U/L, Corr. Coeff. 0.9991, Mean Bias 0.9%. Lot 2: Slope 1.015, y-Intercept -3.4 U/L, Corr. Coeff. 0.9991, Mean Bias 0.7%.N/A
    Calibration StabilityClinically acceptable.30 daysNot explicitly detailed.
    Onboard StabilityClinically acceptable.30 daysNot explicitly detailed.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision:

      • Sample Size: 80 replicates per sample type (Serum Control 1, Serum Control 2, Serum High Pool, Plasma Low Pool, Plasma Med Pool, Plasma High Pool) for each of two reagent lots. Total of 480 measurements per lot.
      • Data Provenance: Not explicitly stated, but typical for in vitro diagnostic device studies would be laboratory-generated or purchased control materials and pooled human specimens. The study was conducted in-house by SEKISUI DIAGNOSTICS P.E.I. INC.
      • Retrospective/Prospective: Analytical studies like precision are typically prospective, performed under controlled laboratory conditions.
    • Analytical Sensitivity (LoB, LoD):

      • Sample Size: 60 measurements for each level (low level samples, saline blanks, dilutions of pooled serum/plasma) for each reagent lot and matrix type.
      • Data Provenance: Not explicitly stated, but likely laboratory-generated materials.
      • Retrospective/Prospective: Prospective laboratory study.
    • Limit of Quantitation (LoQ):

      • Sample Size: 40 replicates (8 linear serum samples, 8 linear plasma samples assayed in 5 runs over 3 days) for each reagent lot and matrix type.
      • Data Provenance: Not explicitly stated, but likely laboratory-generated materials.
      • Retrospective/Prospective: Prospective laboratory study.
    • Linearity/Assay Reportable Range:

      • Sample Size:
        • Serum: 8 dilutions of pooled serum, tested in quadruplicate on two lots.
        • Plasma: 10 dilutions and admixtures of pooled plasma, tested in quadruplicate on two lots.
      • Data Provenance: Pooled serum and plasma likely from human donors, obtained through commercial biospecimen providers or internal collection (with ethics approval).
      • Retrospective/Prospective: Prospective laboratory study.
    • Analytical Specificity (Interference):

      • Sample Size: Minimum of seven interferent concentrations in replicates of five on each of two reagent lots.
      • Data Provenance: Laboratory spiked samples using serum.
      • Retrospective/Prospective: Prospective laboratory study.
    • Method Comparison with Predicate Device:

      • Sample Size: 112 serum specimens, tested in duplicate.
      • Data Provenance: Human serum specimens, source not further specified (e.g., country of origin).
      • Retrospective/Prospective: Likely prospective collection or use of banked retrospective samples under controlled laboratory conditions.
    • Matrix Comparison (Serum vs. Lithium Heparin Plasma):

      • Sample Size: 75 matched serum and lithium heparin plasma specimens, assayed in duplicate.
      • Data Provenance: Matched human serum and plasma specimens, source not further specified.
      • Retrospective/Prospective: Likely prospective collection or use of banked retrospective samples under controlled laboratory conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of in vitro diagnostic device (IVDD) for Creatine Kinase measurement, which measures a biochemical analyte, does not typically rely on "experts" in the same way an imaging AI device would. The "ground truth" for these studies is established through:

    • Reference Methods: For value assignment (e.g., DC-Cal Multi-Analyte Calibrator traceable to an IFCC reference method).
    • Highly Characterized Materials: Pooled serum/plasma, control materials with known target values.
    • Statistical Analysis: CLSI guidelines (e.g., EP05-A3, EP17-A2, EP06-A, EP07-A2, EP09-A3) provide the statistical framework for defining performance metrics like precision, linearity, and limits of detection/quantitation.
    • Comparison to Predicate: The predicate device's established performance serves as a comparative benchmark.

    Therefore, there were no specific "experts" (like radiologists interpreting images) establishing ground truth for individual test cases in the context of this 510(k) submission.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the ground truth for this IVDD study is based on quantitative measurements against reference methods, characterized materials, and statistical analysis, not on subjective interpretations requiring adjudication by multiple readers or experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are primarily relevant for imaging devices or AI tools where human interpretation of medical images is involved, and the AI's impact on reader performance is being evaluated. This 510(k) is for an in vitro diagnostic assay, which directly measures an analyte concentration.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are all "standalone" in the sense that they evaluate the performance of the SEKURE Creatine Kinase Assay and the SK500 analyzer system based on their analytical capabilities to directly measure CK activity. There is no "human-in-the-loop" component in the direct measurement process or the evaluation of its analytical performance. The device provides a quantitative result without human interpretive input for the measurement itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used here is primarily analytical ground truth based on:

    • Reference Methods: Specifically, the CK activity value assignment for DC-Cal is traceable to an IFCC reference method.
    • Characterized Materials: Pooled serum and plasma with known or expected CK activity levels derived from rigorous analytical testing.
    • Statistical Models: Ground truth for linearity (theoretical values) and detection limits (through statistical calculation as per CLSI guidelines).
    • Comparison to a Legally Marketed Predicate Device: The predicate device's performance (Creatine Kinase-SL Assay on a Hitachi 717 Analyzer) serves as a comparative ground truth for demonstrating substantial equivalence.

    8. The sample size for the training set

    This submission pertains to the performance validation of a diagnostic assay (reagents and analyzer), not a machine learning or AI algorithm in the traditional sense that requires distinct "training sets" and "test sets" for model development and validation. Therefore, there is no explicit separate "training set" described for an AI model.

    The "training" of such a system involves the development and optimization of the assay chemistry and instrument parameters. The data presented here are part of the verification and validation (V&V) studies conducted on the final product to demonstrate its performance characteristics and substantial equivalence, akin to a "test set" for a traditional product.

    9. How the ground truth for the training set was established

    As there isn't a "training set" in the context of an AI algorithm, the concept of establishing ground truth for it doesn't apply directly here. The development of the assay involves standard analytical chemistry principles, optimization of reagent concentrations, reaction conditions, and instrument calibration, with performance iteratively assessed against established analytical standards and clinical relevance. This is an engineering and chemistry development process rather than an AI model training process that relies on labeled datasets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163078

    Validate with FDA (Live)

    Device Name
    DC-UIBC-CAL
    Date Cleared
    2016-12-08

    (35 days)

    Product Code
    Regulation Number
    862.1150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For IN VITRO diagnostic use as a calibrator in clinical chemistry assays for UIBC (Unsaturated Iron Binding Capacity). DC-UIBC-CAL may be used to check the linearity over the reportable patient range of UIBC assays.

    Device Description

    DC-UIBC-CAL is a human based, lyophilized serum intended for use as a calibrator with Sekisui Diagnostics UIBC Assays. The concentrations and activities of the UIBC (Unsaturated Iron Binding Capacity) is optimized for the calibration of clinical chemistry systems - Beckman Coulter AU® Systems, Roche/Hitachi® Beckman Coulter SYNCHRON® Systems, and Cobas Mira®.

    The DC-UIBC-CAL kit consists of the following:

    Six vials with 10mL of lyophilized serum.

    Human serum was used in the manufacture of this product. The human serum in this product has been prepared exclusively from the blood of donors tested individually and shown by FDAapproved methods to be free from HBsAg and antibodies to HCV and HIV. However, as no test method can rule out the potential risk of infection with absolute certainty, the material should be treated just as carefully as a patient sample. In the event of exposure, the directives of the responsible Health Authorities should be followed.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device called DC-UIBC-CAL, a calibrator for clinical chemistry assays. The submission focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than a full de novo performance study. As such, many of the typical elements requested for software-based AI/ML device studies are not present.

    Here's an analysis based on the provided text, indicating where information is present and where it is not applicable or not provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative "acceptance criteria" in the format typically seen for diagnostic performance metrics (e.g., sensitivity, specificity, AUC). Instead, the acceptance is based on demonstrating "substantial equivalence" to a predicate device.

    The reported performance is primarily around stability and traceability/value assignment.

    Acceptance CriteriaReported Device Performance
    Substantial Equivalence to PredicateThe information presented in the premarket notification demonstrates that the performance of the Sekisui Diagnostics modified DC-UIBC-CAL is substantially equivalent to the cleared predicate device.
    Shelf Life StabilityAccelerated stability studies for shelf life were conducted, and acceptance criteria were met. The DC-UIBC-CAL is stable at 2-8 °C for 36 months or until the expiration date. Real-time studies are ongoing.
    Open Vial Stability (Reconstituted)Acceptance criteria were met. Reconstituted product is stable at 2-8 °C for 14 days.
    Traceability of UIBC ValuesUIBC values are traceable to NIST SRM 937. Sekisui verifies UIBC is within acceptable target range using Sekisui protocol.
    Target Range Consistency Across InstrumentsFor UIBC-SL Reagent 153-10/30/50/90, the target range across Beckman Coulter AU, Beckman Coulter SYNCHRON, Roche/Hitachi, and Cobas Mira systems is 208 – 312 µg/dL. Specific target values are provided (e.g., 238 µg/dL for Beckman Coulter AU Systems).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document relates to a calibrator, not a diagnostic test for patient samples. Therefore, the concept of a "test set" for patient data, sample size, and data provenance (country, retrospective/prospective) as typically applied to diagnostic algorithms is not applicable here.

    The stability studies would have involved multiple units of the calibrator product. The traceability and value assignment involved using two reagent lots on instruments. No specific sample sizes for these internal validation tests are provided.

    The device itself is manufactured using human serum, and mentions that "Human serum was used in the manufacture of this product."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable. The device is a calibrator, not an interpretative AI/ML diagnostic device. Ground truth for a calibrator relates to its chemical composition and assigned values, which are established through metrological traceability to reference materials (like NIST SRM 937) and internal protocols, not through expert human review of diagnostic outputs.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable for a calibrator device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML diagnostic device for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a calibrator, a physical reagent device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the calibrator's UIBC concentration is established through metrological traceability to a reference material, specifically NIST SRM 937 (National Institute of Standards and Technology Standard Reference Material 937). Internal Sekisui protocols are also used for value assignment. This is a chemical/physical ground truth, not a clinical ground truth like pathology or patient outcomes.

    8. The sample size for the training set

    Not applicable. This is a calibrator, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111915

    Validate with FDA (Live)

    Device Name
    MAGNESIUM ASSAY
    Date Cleared
    2011-12-02

    (149 days)

    Product Code
    Regulation Number
    862.1495
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sekisui Magnesium Assay is for the quantitative determination of magnesium in human serum and plasma (Lithium Heparin) on automated chemistry analyzers. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). This device is intended for professional use and IN VITRO diagnostic use only.

    Device Description

    The Sekisui Magnesium assay kit consists of the following: Magnesium Reagent: A solution containing buffer (pH 11.2 at 25°C), 0.14 mmol/L xylidy/ blue-1, 0.1 mmol/L EGTÅ, and a surfactant.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Sekisui Magnesium Assay, based on the provided text:

    Acceptance Criteria and Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Sekisui Magnesium Assay)
    PrecisionTotal CV%:
    0.8 mg/dL (low)N/A (evaluated per CLSI EP5-A2)6.1%
    2.0 mg/dL (mid)N/A (evaluated per CLSI EP5-A2)3.7%
    4.7 mg/dL (high)N/A (evaluated per CLSI EP5-A2)3.2%
    Within Run CV%:
    0.9 mg/dL (low)N/A (evaluated per CLSI EP5-A2)7.0%
    2.0 mg/dL (mid)N/A (evaluated per CLSI EP5-A2)2.2%
    4.9 mg/dL (high)N/A (evaluated per CLSI EP5-A2)1.6%
    Linearity / Reportable RangeDemonstrably linear over the specified range for a linear equation (Nonlinearity < allowable nonlinearity).Linear from 0.2 mg/dL to 8.0 mg/dL. Reportable range confirmed as 0.3 - 8.0 mg/dL.
    Limit of Detection (LoD)Reliably detect the smallest amount of analyte (LoD) and quantify at a specified CV (LoQ).LoD: 0.25 mg/L, LoQ: 0.30 mg/L
    Analytical Specificity / InterferenceNo significant interference (within +/- 10% of control result) at specified concentrations of interfering substances.No significant interference found for: - Hemoglobin up to 1000 mg/dL - Conjugated Bilirubin up to 40 mg/dL - Unconjugated Bilirubin up to 40 mg/dL - Ascorbic Acid up to 3000 µg/dL - Intralipid up to 1000 mg/dL (or 3000 mg/dL simulated triglycerides)
    Method Comparison (vs. Predicate)Substantially comparable with the predicate device.Slope: 0.981, Intercept: 0.00 mg/dL, Correlation: 0.9848
    Matrix Comparison (Serum vs. Plasma)Substantially comparable between serum and Li Heparin plasma.Slope: 1.009, Intercept: -0.12 mg/dL, Correlation: 0.9903
    Expected Values / Reference RangeLiterature reports of adult reference ranges for magnesium are supported by a study with no outliers.Reference range supported by a study of 20 healthy donors with no outliers.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Precision: Not explicitly stated as a test set size for a single study, but samples (serum pools and control materials) were tested twice per day in duplicate for 20 days.
      • Linearity: 11 concentrations of magnesium (dilutions of a high concentration serum pool).
      • Detection Limit: 60 replicates of a zero magnesium sample and 60 replicates of a low concentration serum control (total 120 replicates).
      • Analytical Specificity: Three samples of human serum at different magnesium concentrations were tested in quadruplicate for each interfering substance.
      • Method Comparison: 100 serum samples.
      • Matrix Comparison: Not explicitly stated as a number, but refers to "fresh samples from non-fasting donors."
      • Reference Range Validation: 20 serum samples from apparently healthy donors.
      • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be retrospective as they use acquired samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This is an in vitro diagnostic device for quantitative chemical measurement. The "ground truth" for analytical performance tests (precision, linearity, detection limits, specificity, method comparison) relies on established reference methods, spiked samples, or known concentrations, rather than expert consensus on diagnostic images or interpretations.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is an in vitro diagnostic device for quantitative chemical measurement. Adjudication methods are typically used for qualitative or interpretive assessments (e.g., in imaging studies) where human disagreement needs resolution.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader, multi-case comparative effectiveness study was not done. This type of study is relevant for devices involving human interpretation (e.g., radiology AI), not for a quantitative chemical assay.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance characteristics described (precision, linearity, detection limit, analytical specificity) are standalone performance of the assay itself on an automated chemistry analyzer (Roche/Hitachi 717). The method comparison and matrix comparison also evaluate the device's performance against a predicate or different matrix, not human readers.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Precision: Internal controls and serum pools with target values.
      • Linearity: Serially diluted samples from a high concentration pool, where relative concentrations are known.
      • Detection Limit: Zero magnesium samples and low concentration serum control samples.
      • Analytical Specificity: Spiked samples with known concentrations of interfering substances, compared to unspiked controls.
      • Method Comparison: Comparison against the predicate device (Beckman Coulter Magnesium Reagent, K944407)
      • Matrix Comparison: Comparison between the candidate device run on serum vs. plasma samples.
      • Reference Range: Samples from "apparently healthy donors" (implying clinically normal magnesium levels).
    7. The sample size for the training set:

      • Not applicable. This device is a chemical reagent and does not involve machine learning or AI that requires a "training set" in the conventional sense. Its performance is based on chemical reactions and assay optimization.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1