Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K202293

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-07

    (55 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

    Device Description

    This is a new generation of Ceiling Suspension Radiographic System. This system is characterized by its simple and functional design. Thanks to its vertical and horizontal displacements, the suspension can cover almost all the room positions in which it is installed allowing all radiographic procedures. The system is modular and supports different configurations, such as radiographic system without radiographic table or without Wall Stand. The X-ray image receptors used in this system are digital detectors. X-ray film and Computed Radiography (CR) image receptors can be used but they rarely are these days. The device software used is the CANON CXDI which is supplied unmodified by CANON (Clearance numbers above). It has a moderate level of concern. The Radiographic System ChallengeX AP is provided with Auto-positioning, Auto-centering and Auto-tracking functions and it is composed of: Ceiling Suspension (OTC), Radiographic Table, Wall Stand, High Voltage X-ray Generator and acquisition image software. Auto-tracking allows the X-ray Tube to follow the Receptor when it changes position or the other way around while the SID remains constant. The "Auto" features were present and validated in the predicate system.

    AI/ML Overview

    The document provided is a 510(k) premarket notification for a medical device called "Radiographic System Challenge X." This notification aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific acceptance criteria in a clinical study with an AI component.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for AI-based analysis is not present in the provided text.

    The document discusses non-clinical testing for compliance with international standards and FDA guidance for traditional medical device aspects such as safety, electrical compatibility, radiation protection, and software validation. It explicitly states that clinical testing was not required to establish substantial equivalence because the digital x-ray receptor panels already had previous FDA clearance.

    Key takeaway from the document regarding studies:

    • No AI component or AI-specific acceptance criteria are documented.
    • The device is a traditional X-ray system, not an AI-powered diagnostic tool.
    • Clinical testing was not performed for this submission. The substantial equivalence was established through non-clinical bench testing and compliance with existing standards and previously cleared components (like the detectors and software).
    Ask a Question

    Ask a specific question about this device

    K Number
    K192011

    Validate with FDA (Live)

    Device Name
    PhoeniX
    Manufacturer
    Date Cleared
    2019-09-25

    (58 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    0 - 0.08
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technician on both adult and pediatic subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography

    Device Description

    These are redesigned versions of our predicate mobile digital diagnostic xray systems. They feature motorized movement and full battery operation. Various Canon digital Xray panels are supplied with the system. (See list above.) The Mobile X-Ray Unit PhoeniX has a Basic configuration or Advanced configuration. Advanced configurations include:

    • Second screen on head assembly.
    • Smooth movements of head assembly.
    • Telescopic arm in four steps instead three steps for basic configuration.

    The Mobile X-Ray unit is provided with touch screen to operate as a control console. The Digital Imaging System is composed by image receptors and application for image acquisition (control console & image processing controller). The Image acquisition software "CANON CXDI Control Software NE" runs on Mobile X-ray unit and it is displayed on touch screen. It is the user interface and compatible digital detectors are listed above. All have FDA Clearance. The Advanced configuration has a second Touch Screen Monitor located on head-assembly.

    The PhoeniX Mobile X-Ray Unit is provided with separate battery packs for X-Ray generation and motorized movements of the unit can operate connected to mains or in stand-alone mode, that is, operating without mains being present or unplugged from mains. The unit is connected to mains to charge the batteries; New rating: the input voltage range goes from 100 V~ to 240 V~, 1 phase, 1.1 kVA. New X-ray generator (model SHFM) is mounted on Battery Mobile X-Ray unit PhoeniX. This new X-Ray generator for PhoeniX Mobile X-Ray Unit has a radiogenic unit mounted on headassembly and comprising an X-ray tube with rotating anode and its circuit for high voltage. The electronic and associated software to control the X-ray generation are placed on mobile cart. The Battery Mobile X-Ray Unit PhoeniX is provided with different output powers: 20 kW, 32 kW, 40 kW and 50 kW. There are available two X-ray tube inserts with rotating anode manufactured by CANON ELECTRON TUBES & DEVICES:

    • New: XRR-3331 insert.
    • New: E7886 insert.
      New: Manual Beam Limiting Device from Ralco, model: R108 F. External interface (controls) and covers are provided by Sedecal. There are two versions of collimator assembly, Basic and Advanced.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a mobile X-ray system named PhoeniX. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study for novel acceptance criteria.

    Therefore, many of the requested items (acceptance criteria, specific study details with sample sizes, expert involvement, and ground truth information) are not directly addressed in the provided FDA submission as they would be for a de novo marketing authorization or a more extensive clinical study. The submission relies heavily on non-clinical testing and the prior clearance of its components and predicate devices.

    However, I can extract information related to the device's technical specifications and the basis for its substantial equivalence, which implicitly acts as its "acceptance criteria" for regulatory clearance.

    Here's an analysis based on the provided text, addressing the points where information is available:

    1. Table of Acceptance Criteria (Implicit) and Reported Device Performance

    The device is considered acceptable if its performance and specifications are substantially equivalent to the predicate device and meet relevant international standards. The "reported device performance" in this context refers to its technical specifications and compliance with safety standards, rather than clinical performance metrics like sensitivity or specificity.

    Acceptance Criteria (Implicitly based on Predicate & Standards)Reported Device Performance (PhoeniX)
    Indications for Use:Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography. (SAME as Predicate)
    Configuration:Mobile System with digital x-ray panel and image acquisition computer. (SAME as Predicate)
    kW Rating:20 kW, 32 kW, 40 kW and 50 kW. (Predicate: 40 kW only)
    kV Range:From 40 kV to 150 kV in 1 kV steps. (SAME as Predicate)
    mA Range:From 10 mA to 630 mA / 640 mA / 650 mA. (Predicate: From 10 mA to 500 mA)
    Collimator:Ralco R108F (Two versions: Basic and Advanced). (Predicate: Ralco R221 DHHS)
    Digital X-ray Panels Supplied:CANON CXDI-401C Wireless, CANON CXDI-701C Wireless, CANON CXDI-801C Wireless. (SAME as Predicate). Plus: CANON CXDI-710C Wireless, CANON CXDI-810C Wireless, CANON CXDI-410C Wireless. (New panels, but all previously FDA cleared).
    Software:Canon control software CXDI-NE. (SAME as Predicate, updated in K190368)
    Panel Interface:Ethernet or Wi-Fi wireless. (SAME as Predicate)
    Meets US Performance Standard:YES 21 CFR 1020.30. (SAME as Predicate)
    Power Source:Universal power supply, from 100 V~ to 240 V~. 1 phase, 1.1 kVA. (Predicate: Input transformer with 7 input voltage taps; AC 20 amp and Batteries)
    Safety and Effectiveness:"The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination."
    Compliance with International Standards:IEC 60601-1:2005+A1:2012 (Edition 3.1), IEC 60601-1-2:2014 (Edition 4.0), IEC 60601-1-3:2008+A1:2013 (Edition 2.1), IEC 60601-2-54:2009+A1:2015 (Edition 1.1), IEC 60601-2-28:2010 (Edition 2.0), IEC 60601-1-6:2010 + A1:2013 (Edition 3.1), IEC 62304:2006 + A1:2016 (Edition 1.1). (Compliance affirmed through non-clinical testing)

    Regarding the Study Proving Acceptance Criteria:

    The "study" in this submission is primarily a non-clinical bench testing and comparative analysis to establish substantial equivalence to a legally marketed predicate device (K161345 RadPRO® Mobile 40kW; RadPRO® Mobile 40kW FLEXPLUS).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified in terms of number of images or patients. The non-clinical testing involved "Systems covering all generator/panel combinations were assembled and tested." This suggests testing of device configurations rather than a "test set" of medical images or patient data in the typical sense of an AI/CAD device.
    • Data Provenance: Not applicable in the context of clinical data for a "test set." The testing是在实验室环境中,使用设备进行物理和功能验证。

    3. Number of Experts and Qualifications to Establish Ground Truth:

    • Not applicable as this was a non-clinical submission for substantial equivalence based on technical specifications and functionality. Clinical trials requiring expert reads for ground truth were not performed.

    4. Adjudication Method:

    • Not applicable as there was no clinical reading study or consensus required for ground truth determination.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. The document states: "Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance." This device is a mobile X-ray system, not an AI or CAD software that assists human readers.

    6. Standalone (Algorithm Only Without Human-in-the-loop) Performance:

    • Not applicable. This device is an X-ray imaging system, not an algorithm that performs a diagnostic task independently. Its performance is related to image quality, radiation output, mechanical safety, and electrical safety.

    7. Type of Ground Truth Used:

    • Technical Specifications and Compliance with Standards: The "ground truth" for this submission are the established performance characteristics of the predicate device and the specified requirements of relevant international safety and performance standards (e.g., IEC 60601 series, 21 CFR 1020.30). The "truth" is that the device meets these engineering and regulatory benchmarks.
    • The document mentions that the modified combination of components "produced diagnostic quality images as good as our predicate generator/panel combination," implying an internal assessment of image quality, but no explicit detailed ground truth generation method is described for clinical image interpretation.

    8. Sample Size for the Training Set:

    • Not applicable. This is not an AI/ML device that requires a training set of data. The device's components (X-ray generator, panels, software) are either updates to existing technology or previously cleared devices.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1