Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K232208

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-16

    (265 days)

    Product Code
    Regulation Number
    864.3700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use

    Sectra Digital Pathology Module (3.3) is a software device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis.

    Sectra Digital Pathology Module (3.3) is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using Sectra Digital Pathology Module (3.3).

    Sectra Digital Pathology Module (3.3) is intended for use with Leica's Aperio GT 450 DX scanner and Dell U3223QE display, for viewing and management of the ScanScope Virtual Slide (SVS) and Digital Imaging and Communications in Medicine (DICOM) image formats.

    Device Description

    The Sectra Digital Pathology Module (3.3) [henceforth referred to DPAT (3.3)] is a digital slide viewing system. The DPAT (3.3) is intended for use together with FDA-cleared whole-slide image scanner GT 450 DX and Dell U3223QE display.

    The DPAT (3.3) can only be used as an add-on module to Sectra PACS. Sectra PACS consists of Sectra Workstation IDS7 (K081469) and Sectra Core (identified as a Class I exempt by the FDA in 2000). Sectra PACS is not part of the subject device. Sectra Workstation is the viewing workstation in which the Pathology Image Window is run. Pathology Image Window is the client component of the subject device.

    The system capabilities include:

    • retrieving and displaying digital slides,
    • support for remote intranet access over computer networks,
    • tools for annotating digital slides and entering and editing metadata associated with digital slides, and
    • displaying the scanned slide images for primary diagnosis by pathologists.

    The subject device is designed to accurately display colors. The monitor is not part of the subject device.

    Digital pathology images originating from WSI scanners other than those listed in the Indications for Use will be marked with the disclaimer "For Non-clinical Use Only" in the Pathology Image Window.

    Image acquisition will be managed by the scanner which is not part of the subject device:

    • The scanner delivers images with a tag in the file header that identifies the originating scanner.
    • The scanner includes applications for controlling the scanning process and performing related quality control (e.g., ensuring that images are sharp and cover all tissue on the slide).

    The DPAT (3.3) supports reading digital slides on a Dell U32230E display monitor, enabling pathologists to make clinically relevant decisions analogous to those they make using a conventional microscope. Specifically, the system supports the pathologist in performing a primary diagnosis based on viewing the digital slide on a computer monitor. These capabilities are provided by the Pathology Image Window.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Endpoint)Reported Device PerformanceMet?
    Primary Endpoint: The upper bound of the 2-sided 95% CI of the difference between the overall major discrepancy rates of WSIR diagnosis and MSR diagnosis when compared to the reference diagnosis shall be ≤4%.The upper bound of the 95% CI of the estimated difference in major discrepancy rates was 1.69%.Yes
    Secondary Endpoint: The upper bound of the 2-sided 95% CI of the major discrepancy rate between WSIR diagnosis and the reference diagnosis shall be <7%.The upper bound of the 95% CI for the overall estimated major discrepancy rate for WSIR diagnosis was 5.25%.Yes
    Pixel-Wise Comparison: All configurations were identical, i.e., <3ΔΕ00 to reference configuration SVS/UniView/Chrome.Based on analysis of the testing data, the 4 specified configurations (DICOM/IDS7, DICOM/UniView/Chrome, SVS/IDS7, SVS/UniView/Edge) were identical, i.e., <3∆Ε00 to reference configuration SVS/UniView/Chrome, and specifically ΔE=0 for the subject device displaying the same scanned image in both DICOM and SVS formats, in IDS7, Edge or Chrome.Yes
    Turnaround Time:
    - When selecting a slide image, it should not take longer than 3 seconds until the image is fully loaded.Reported to be "adequate for the intended use". Specific values not provided but "similar to or better than those of the predicate device."Yes
    - When panning the image (one quarter of the monitor) it should not take longer than 0.5 seconds until the image is fully loaded.Reported to be "adequate for the intended use". Specific values not provided but "similar to or better than those of the predicate device."Yes
    Measurements: Measurement accuracy has been verified using a test image containing objects with known sizes.Reported that measurement accuracy "has been verified" and show "almost identical results" to the predicate.Yes

    Note: While specific numerical results for turnaround time and measurement accuracy are not provided, the document states they were found to be "adequate" and "accurate" respectively, meeting the implicit acceptance criteria for these performance aspects.

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 258 cases.
    • Data Provenance: The document does not explicitly state the country of origin. The study was conducted at a "single site." It was a retrospective study, as the MSR diagnoses were "original sign-out diagnoses." The WSIR diagnoses were prospectively obtained using the device for the study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts for Ground Truth: Not directly stated as "experts for ground truth," but 3 reading pathologists determined the MSR diagnoses (which formed a basis for comparison with the reference diagnosis) and 3 reading pathologists created the WSIR diagnoses. A minimum of two adjudicators independently assessed concordance for the WSIR diagnosis against the reference diagnosis.
    • Qualifications of Experts: All were "pathologists." Further specific qualifications (e.g., years of experience, board certification) are not detailed in the provided text.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Minimum of two adjudicators independently assessed concordance (concordant, minor discrepancy, major discrepancy) of the WSIR diagnosis against the reference diagnosis using predefined rules. Their concordance scores for the same case were compared to determine a consensus score for major discrepancy status. This represents a form of 2-reader adjudication with consensus. The document does not explicitly mention "adjudication of ground truth" but rather adjudication of the concordance between the device's output and the reference diagnosis.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes, effectively. The study design involved 3 reading pathologists (multi-reader) reviewing 258 cases (multi-case) using the device (WSIR diagnosis) and comparing it to their traditional light microscopy review (MSR diagnosis), both against a reference diagnosis. While not explicitly termed an "MRMC study" in the classic sense of comparing AI-assisted vs. unassisted human performance, it acts as a comparative effectiveness study demonstrating non-inferiority of the digital pathology system to traditional microscopy.
    • Effect Size of Human Readers Improvement with AI vs. without AI Assistance: The study's primary endpoint was non-inferiority of the digital system (DPAT (3.3)-UniView) compared to traditional light microscopy. The estimated difference in major discrepancy rates between the two modalities (digital vs. microscope) when compared to the reference diagnosis was -0.01% (95% CI: -1.71% to 1.69%). This indicates that the digital system performed comparably to, or negligibly better than, light microscopy in terms of major discrepancy rates against a reference. It doesn't quantify improvement with AI assistance per se, but rather the non-inferiority of the digital viewing system (which is the device being cleared) to the traditional method.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    • Was a standalone performance study done? No, not in the sense of an AI algorithm making a diagnosis without human intervention. The device is a "software device intended for viewing and management of digital images... as an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis." The study specifically evaluates "interpreting digital images... for the purposes of primary diagnosis" by a human pathologist using the device, not an automated diagnostic output.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The study used "original sign-out diagnoses" made using light microscopy as the "reference diagnoses." The document clarifies that major discrepancy was defined as a "difference in diagnosis that resulted in a clinically important difference in patient management." While not explicitly stated as "pathology ground truth" established post-hoc, it strongly implies a consensus or definitive diagnosis used as the gold standard derived from clinical practice.

    8. Sample Size for the Training Set

    • Training Set Sample Size: The document does not mention the training set size for the device. The study described focuses on the performance evaluation of the final device (Sectra Digital Pathology Module 3.3) for clinical validation, not the development or training of any underlying AI or image processing models within the device.

    9. How Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: As no training set information is provided, the method for establishing its ground truth is also not mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193054

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-31

    (151 days)

    Product Code
    Regulation Number
    864.3700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use

    Sectra Digital Pathology Module device is a software intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis.

    Sectra Digital Pathology Module is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

    It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using Sectra Digital Pathology Module.

    Sectra Digital Pathology Module is intended for use with Leica's Aperio AT2 DX scanner and Dell MR2416 monitor.

    Device Description

    Sectra Digital Pathology Module is a software-only device running under the Microsoft Windows operating system for displaying and manipulating digital pathology images (scanned slides) obtained from the Aperio AT2 DX scanner.

    Sectra Digital Pathology Module may only be used in combination with Sectra PACS which consists of Sectra Workstation (K081469) and Sectra Core (identified as a Class I exempt by the FDA in 2000).

    The Sectra Pathology Import Server (SPIS) is used for importing digital pathology images (scanned slides) from the scanner. These images are viewed and manipulated by end users in the Pathology Image Window which is displayed on the Sectra Workstation IDS7 (using the Dell MR2416 monitor).

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study conducted for the Sectra Digital Pathology Module, extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Color ReproducibilityPixel-wise comparison towards Aperio ImageScopeDX (predicate) across multiple tiles, including zooming and panning operations. Sectra Digital Pathology Module was found to reproduce colors adequately and was non-inferior to ImageScopeDX.
    Turnaround Time (Case Load)When selecting a case, it should not take longer than 7 seconds until the image is fully loaded, provided system requirements are fulfilled.
    Turnaround Time (Panning)When panning the image (one quarter of the monitor), it should not take longer than 0.5 seconds until the image is fully loaded, provided system requirements are fulfilled.
    MeasurementsMeasurement accuracy has been verified using a test image containing objects with known sizes. The device was found to perform accurate measurements.
    Human FactorsTask-based usability tests showed the Sectra Digital Pathology Module user interface to be intuitive, safe, and effective for the range of intended users. The device was found to be safe and effective for the intended users, uses, and use environments.

    2. Sample Size for Test Set and Data Provenance:

    The document does not explicitly state the sample size for any specific test set regarding color reproducibility, turnaround time, or measurements. It mentions a "test image containing objects with known sizes" for measurements, and "multiple tiles" for color reproducibility.

    The data provenance is not specified (e.g., country of origin). The testing seems to be internal development and verification testing rather than involving external patient data.

    3. Number of Experts for Ground Truth and Qualifications:

    The document does not mention the use of experts to establish ground truth for the technical performance tests described (color reproducibility, turnaround time, measurements, human factors). These appear to be objective, quantifiable tests directly comparing the device's output to a known standard or the predicate device's output.

    For "Human factors testing," it states "Task-based usability tests showed the Sectra Digital Pathology Module user interface to be intuitive, safe, and effective for the range of intended users." While suggesting involvement of intended users (pathologists likely), the number and specific qualifications of these users are not detailed.

    4. Adjudication Method for Test Set:

    No adjudication method is mentioned. The tests described are primarily objective comparisons or measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done. The document explicitly states: "Substantial equivalence determination is not based upon clinical study results."

    6. Standalone (Algorithm Only) Performance Study:

    Yes, standalone (algorithm only, meaning the software itself without human interaction for diagnostic purposes) performance was done. The studies listed under "Non-clinical test results" (Color Reproducibility, Turnaround times, Measurements) evaluate the technical performance of the software. The device itself is described as a "software-only device."

    7. Type of Ground Truth Used:

    • Color Reproducibility: Ground truth was implicitly the visual output and pixel data of the predicate device (Aperio ImageScopeDX).
    • Turnaround Time: Ground truth was the pre-defined maximum acceptable time limits (7 seconds for case load, 0.5 seconds for panning).
    • Measurements: Ground truth was a "test image containing objects with known sizes."
    • Human Factors: Ground truth was established by assessing usability (intuitiveness, safety, effectiveness) through "task-based usability tests" likely with intended users.

    8. Sample Size for the Training Set:

    The document does not mention a training set. The Sectra Digital Pathology Module is described as a "software intended for viewing and management of digital images," rather than an AI/ML algorithm that requires a training set.

    9. How Ground Truth for Training Set Was Established:

    Since no training set is mentioned for an AI/ML algorithm, this question is not applicable based on the provided text. The device is a viewer and management tool, not an AI diagnostic algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K034059

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-09

    (69 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sectra Angiography and Cardiology Package provides quantitative and reproducible information regarding the calculated dimensions of arterial segments, and quantitative information regarding the calculated dimensions and calculated performance characteristics of the left vertical of the heart, imaged during angiographic x-ray procedures typically performed in cath labs.

    This information is suitable for use in the following applications:

    1. Scientific and research studies, assessing the angiographic condition of patients and the result of device and drug therapeutics,
    2. Review and analysis of patient angiographic imaging records, providing additional information to physicians and administrators.
    Device Description

    Sectra Angiography and Cardiology Package is intended to support radiologists and cardiologists in the diagnostic process to quantify findings. The device has functionality for quantitative analysis of the arteriograms (QCA and QVA) and quantitative analysis of the left ventricular angiographic images (LVA). Measurements can be applied on angiographic images.

    AI/ML Overview

    The provided text does not contain acceptance criteria for device performance or a study demonstrating that the device meets such criteria.

    The document is a 510(k) summary for the Sectra Angiography and Cardiology Package, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance studies or specific acceptance criteria.

    The "Performance Data" section merely states: "The subject device is developed according to ISO 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0." This indicates compliance with quality management systems and communication standards, not a clinical performance study with acceptance criteria.

    Therefore, I cannot populate the requested table or answer most of the questions, as the information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033088

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-10-29

    (30 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sectra IDS device is intended for the manipulation and displaying of x-ray images. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.

    Device options make possible telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing.

    Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

    Device Description

    The IDS5 10.1 Radiology Workstation is mainly a software product. It is used for visualization and processing of digital radiology images. The system runs under the Window 2000 and Windows XP operating system. The requirements on hardware are quite ordinary for a system used for displaying images. Most notably up to four monitors can be used.

    AI/ML Overview

    This document describes a 510(k) submission for the Sectra IDS5 Radiology Workstation - Version 10.1, a Picture Archiving and Communications System (PACS).

    Here's an analysis of the acceptance criteria and study information provided in the input, specifically focusing on what is present and absent from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    ISO 9001:2000 ComplianceDeveloped according to ISO 9001:2000
    ACR/NEMA Digital Imaging Communications in Medicine (DICOM) version 3.0 ComplianceComplies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0
    Substantial Equivalence to Predicate Device (Sectra IDS5 Radiology Workstation - Version 7, K002936)Concluded to be safe, effective, and substantially equivalent to the predicate device based on shared certification/conformance to standards and functional similarity as an Image Processing System (LLZ).
    Ability to manipulate and display x-ray images from different modalities, interface with image storage and printing using DICOM/similar standards.The device is intended for this purpose and its functions support it.
    Supports telecommunications, fast demonstration, prosthesis CAD, 3-D, angiography, and teleconferencing through device options.The device options make these functions possible.
    Operates on Windows 2000 and Windows XP operating systems.The system runs under these operating systems.
    Protection against unauthorized use (passwords).Passwords are required for operation.
    Image/data recovery mechanisms in case of transmission failure.Device failures may be recovered from storage or retransmission.

    2. Sample size used for the test set and the data provenance

    • Absent. The document does not describe a clinical study or a test set of images with a defined sample size. The performance data section refers to compliance with standards rather than a specific study involving image data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Absent. Since there is no described clinical study or test set for evaluation of image interpretation performance, there is no mention of experts establishing ground truth for such a set. The document states that images and information are "interpreted by a physician or trained medical personnel," implying human oversight during clinical use, but not in the context of device validation.

    4. Adjudication method for the test set

    • Absent. No test set or corresponding adjudication method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Absent. The device described is a PACS workstation, an image processing and display system. It does not appear to incorporate AI or CAD (Computer-Aided Detection/Diagnosis) directly for interpretation assistance. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable and not mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Absent. As noted above, this device is a workstation for displaying and manipulating images, not a standalone AI algorithm for diagnostic interpretation. Its performance is related to its ability to correctly process and present images according to standards, leaving interpretation to human users.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Absent. This information is not relevant in the context of the provided performance data, which focuses on compliance with technical standards (ISO, DICOM) and substantial equivalence to a predicate device for its intended function as an image display and processing system. The core function is presenting images, not independently deriving diagnostic conclusions that would require a ground truth comparison.

    8. The sample size for the training set

    • Absent. This device is a software product for viewing and manipulating images; it does not explicitly use a "training set" in the context of machine learning for diagnostic tasks. Its development followed ISO 9001:2000, which is a quality management system standard.

    9. How the ground truth for the training set was established

    • Absent. As there is no described training set for a machine learning model, the establishment of ground truth for such a set is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K001140

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-05-03

    (23 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1