Found 13 results

510(k) Data Aggregation

    K Number
    K250844

    Validate with FDA (Live)

    Date Cleared
    2025-06-02

    (74 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PULLUP®, BT LOOP®, PULLUP® CLIP and PULLUP® TEX CLIP cortical fixation devices are indicated in adults for reconstruction of the anterior cruciate ligament.

    • PULLUP®, PULLUP® CLIP, PULLUP® TEX CLIP must be used with a Soft Tissue Transplant.
    • BT LOOP® must be used with a Bone Tendon Bone Transplant.

    The RIGIDLOOP™ T, RIGIDLOOP™ BTB, RIGIDLOOP™ Clip and RIGIDLOOP™ Suture Loop cortical fixation devices are indicated in adults for reconstruction of the anterior cruciate ligament.

    • RIGIDLOOP™ T, RIGIDLOOP™ Clip, RIGIDLOOP™ Suture Loop must be used with a Soft Tissue Transplant.
    • RIGIDLOOP™ BTB must be used with a Bone Tendon Bone Transplant.
    Device Description
    • PULLUP®: comprises a titanium button-plate, and an adjustable nonabsorbable braided loop. The system is preloaded with traction and flip threads, and is designed to be used with a soft tissue graft. Standard and XL sizes.

    • BT LOOP®: comprises a titanium button-plate, and an adjustable nonabsorbable braided loop. The system is preloaded with traction and flip threads, as well as a temporary split suture, and is designed to be used with a bone-tendon-bone graft. It is assembled on a holder that can be mounted on the GraftTech® preparation station. Standard and XL sizes.

    • PULLUP® CLIP: comprises a titanium button-plate only, is designed to be used with a soft tissue graft, and can either be connected to another PULLUP® CLIP button-plate by using the PULLUP® TEX CLIP nonabsorbable braided loop, or can be connected to another PULLUP® device. Standard and XL sizes.

    • PULLUP® TEX CLIP: Comprises a nonabsorbable braided loop. One size only.

    Standard models are used for cortical tunnels with a diameter of 4.5 mm; XL models are used for cortical tunnels with a diameter between 5 and 10 mm.

    The implants are supplied sterile, individually packaged, ready to use.

    The PULLUP®, PULLUP® CLIP, BT LOOP® and PULLUP® TEX CLIP devices are also sold under the trade names RIGIDLOOP™ T, RIGIDLOOP™ Clip, RIGIDLOOP™ BTB and RIGIDLOOP™ Suture Loop respectively.

    AI/ML Overview

    I apologize, but based on the provided FDA 510(k) Clearance Letter (K250844), the device in question is a bone fixation fastener, specifically for anterior cruciate ligament (ACL) reconstruction.

    The document describes non-clinical tests performed to demonstrate substantial equivalence, such as:

    • Ultimate failure load (post-fatigue tensile strength) and fatigue displacement
    • Biocompatibility testing (ISO 10993-1:2018)
    • Bacterial endotoxin testing (ISO 11737-3:2023)
    • Shelf-life, storage, and stability testing (ISO 11607-1:2019 for packaging)
    • Sterile adoption (ISO 11137-1:2006)
    • MRI safety evaluation (ASTM F2503 - 20, IEC 62570:2014, FDA guidance)

    The provided text explicitly states "Not Applicable" for clinical tests and indicates that "Changes in design are minor, and do not raise new or different questions of safety and effectiveness compared with the predicate device." This means that no human clinical study was conducted or reported in this 510(k) submission to assess the device's performance in terms of diagnostic accuracy or human-AI interaction.

    Therefore, I cannot provide the information requested regarding:

    1. A table of acceptance criteria and reported device performance directly related to AI or diagnostic accuracy.
    2. Sample size for a test set (as no clinical test set for AI performance was mentioned).
    3. Number of experts or their qualifications for establishing ground truth.
    4. Adjudication method.
    5. MRMC comparative effectiveness study.
    6. Standalone AI performance.
    7. Type of ground truth used (in an AI context).
    8. Sample size for a training set (AI model training).
    9. How ground truth for a training set was established (AI model training).

    The 510(k) summary focuses on the mechanical, material, and safety aspects of the physical implant device, demonstrating its equivalence to existing predicate devices through bench testing and material science principles, not through AI performance or diagnostic accuracy studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231787

    Validate with FDA (Live)

    Device Name
    ecoFIX®
    Date Cleared
    2023-07-26

    (36 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ecoFIX® threaded anchor system is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow, the following procedures:

    Shoulder: Rotator Cuff Repair; SLAP Lesion Repair; Biceps Tenodesis; Acromio-clavicular Separation Repair; Deltoid Repair; Capsule Shift or Capsulolabral Reconstruction;

    Ankle/Foot: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy

    Knee: Anterior Cruciate Ligament Repair; Medial Collateral Ligament Repair; Lateral Collateral Ligament Repair; Patellar Tendon Repair; Posterior Oblique Ligament Repair; Iliotibial Band Tenodesis;

    Wrist/Hand: Scapholunate Ligament Reconstruction; Ulnar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.

    Elbow: Biceps Tendon Reattachment; Ulnar or Radial Collateral Ligament Reconstruction; Tennis Elbow Repair and Lateral Epicondylitis Repair.

    Device Description

    ecoFIX® is an extension the FIXIT® / ComposiTCP™ threaded anchor system previously cleared by the FDA (K170868 & K180960). ecoFIX® is a threaded anchor system which comprises an anchor made of DUOSORB® 30, a composite bioabsorbable material composed of 30 % Beta-tricalcium phosphate (β-ΤCP) and 70 % Poly L,DL-lactic acid (PLDLLA), and sutures made of ultra-high molecular weight polyethylene (UHMWPE). Different variants of the product are available depending on the number and type of suture combination (2 round sutures, 2 round needled sutures, 3 round sutures, or knotless anchor).

    The implant is supplied sterile, ready to use.

    The implant is available in the following configurations, supplied with a reusable screwdriver:

    · ecoFIX®: hexagonal footprint anchor (3 sizes Ø4.5, Ø5.5, Ø6.5) suppplied in an implant holder with 2 round sutures ;

    • · ecoFIX® Needle: hexagonal footprint anchor (3 sizes Ø4.5, Ø6.5) supplied in an implant holder with 2 round needled sutures ;
      · ecoFIX® +: hexagonal footprint anchor (1 size Ø5.5) supplied in implant holder with 3 round sutures ;

    · ecoFIX® Band: hexagonal footprint anchor (3 sizes Ø4.5, Ø5.5, Ø6.5) supplied in an implant holder with 2 flat sutures ;

    · ecoFIX® Knotless: hexagonal footprint anchor (3 sizes Ø4.5, Ø5.5) supplied in an implant holder.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study proving a digital health device meets those criteria.

    Instead, the document is a 510(k) premarket notification for a medical device called "ecoFIX®", which is a threaded anchor system for fixing soft tissue to bone. The document focuses on demonstrating substantial equivalence to a predicate device (FIXIT® / ComposiTCP™ Threaded Anchor System).

    Here's an analysis of the information related to performance testing from the text, as it pertains to the ecoFIX® device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly list specific acceptance criteria in a quantitative table format. Instead, it states that non-clinical performance testing was conducted to determine substantial equivalence.

    Acceptance Criteria (Inferred)Reported Device Performance
    Functional within intended useDevices were functional within their intended use.
    Equivalent to predicate devicesDevices were equivalent to the predicate devices.
    Meet endotoxin limitsBacterial endotoxin testing completed and results demonstrated proposed devices meet endotoxin limits.
    Ease of usePerformance tests carried out to evaluate ease of use. Results demonstrated substantial equivalence to predicate device.
    Driver holdPerformance tests carried out to evaluate driver hold. Results demonstrated substantial equivalence to predicate device.
    Needle holdPerformance tests carried out to evaluate needle hold. Results demonstrated substantial equivalence to predicate device.
    Insertion performancePerformance tests carried out to evaluate insertion performance. Results demonstrated substantial equivalence to predicate device.
    Pullout/slippage performancePerformance tests carried out to evaluate pullout/slippage performance. Results demonstrated substantial equivalence to predicate device.
    Mechanical propertiesThe ecoFIX® Threaded Anchor System is substantially equivalent to its predicate device in terms of mechanical properties.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not specified. The document mentions "non-clinical performance testing" and "performance tests" but does not give specific sample sizes for these tests.
    • Data Provenance: Not specified, but given these are non-clinical (laboratory/bench) tests for a physical implant, the data would originate from the testing facility. The document does not specify country of origin or whether it's retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. The document describes non-clinical performance testing of a physical medical device (an anchor system), not a digital health device requiring expert human input for ground truth establishment.

    4. Adjudication Method for the Test Set:

    Not applicable. This concept (e.g., 2+1, 3+1) is relevant for studies involving human interpretation or review, typically for medical imaging or similar data, which is not the case for this physical medical device's non-clinical performance tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No. An MRMC study is a clinical study design often used for diagnostic imaging devices to compare the performance of human readers with and without AI assistance. This document describes non-clinical testing of a physical implant, not a diagnostic AI device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Not applicable. This device is a physical implant, not an algorithm.

    7. The Type of Ground Truth Used:

    For the non-clinical tests mentioned, the "ground truth" would be established by the physical testing methodologies and measurements themselves, according to established engineering and medical device testing standards (e.g., measuring pullout strength, insertion force, etc.). It would not involve expert consensus, pathology, or outcomes data in the way a diagnostic AI device would.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device. It does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140226

    Validate with FDA (Live)

    Date Cleared
    2014-06-03

    (125 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NEOTIS Plate is intended to be used in conjunction with bone screws to provide fixation following Proximal Tibial Opening wedge osteotomies.

    Device Description

    The NEOTIS plate is designed for medial approach High Tibial Osteotomy stabilization. Anatomically shaped, thin and short, the NEOTIS plate enables minimally invasive surgery. The changes brought by the NEOTIS plate are as follows (vs. OTIS-C Plus plate):

    • The proximal temporary screw is replaced with a pin that offers a simpler and faster placement of the plate. The area of the cortical bone drilled through the process is also reduced.
    • The distal temporary screw is replaced with an AO screw in order to maintain compression. The compression will be maintained by the 4 permanent 6.5 mm screws (OTIS screws).
    AI/ML Overview

    The provided documentation describes a traditional 510(k) submission for the NEOTIS Plate and Screws, a medical device for bone fixation. Here's a breakdown of the acceptance criteria and the study information as extracted from the text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria TypeAcceptance CriteriaReported Device Performance
    Mechanical PropertiesNot explicitly stated as acceptance criteria, but implied to be equivalent to the predicate device."The mechanical properties are the same as for the previous OTIS-C Plus"
    Placement Process (Temporary Screw/Pin)Improved ease and speed of placement, reduced drilled cortical bone area."The proximal temporary screw is replaced with a pin that offers a simpler and faster placement of the plate. The area of the cortical bone drilled through the process is also reduced."
    Compression Maintenance (Distal Screw)Compression to be maintained."The distal temporary screw is replaced with an AO screw in order to maintain compression. The compression will be maintained by the 4 permanent 6.5 mm screws (OTIS screws)."
    Substantial EquivalenceDemonstrate that the modified NEOTIS plate is as safe, as effective, and performs as safely and effectively as its predicate device."Verification and validation tests demonstrate that modified the NEOTIS plate is as safe, as effective, and performs as safely and effectively as its predicate device."
    Confirmation of Compression (Cadaver Study)Not explicitly stated as an acceptance criterion, but the cadaver study's purpose was to confirm this."Non-clinical performance testing includes a cadaver study that confirms the level of compression induced during the placement of the plate."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions a "cadaver study" but does not provide details on the number of cadavers or specimens used.
    • Data Provenance: The cadaver study is implied to be laboratory-based (non-clinical performance testing). The document does not specify the country of origin for the cadaver data, nor does it explicitly state if it was retrospective or prospective, though "non-clinical performance testing" typically implies prospective experimental testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • The documentation does not provide information about experts establishing ground truth for the cadaver study. The study appears to be a mechanical performance test rather than one requiring clinical interpretation by experts to establish ground truth.

    4. Adjudication Method for the Test Set

    • Not applicable/Not mentioned. The study described is a non-clinical performance test, not a study involving human interpretation with adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The device is hardware (plate and screws), not an AI algorithm or diagnostic tool that would typically undergo an MRMC study.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, this is not applicable. The device is a physical medical implant, not an algorithm.

    7. The Type of Ground Truth Used

    • For the cadaver study, the "ground truth" was the level of compression induced by the device during placement. This is a direct physical measurement/observation in a controlled experimental setting, not expert consensus, pathology, or outcomes data in the clinical sense.

    8. The Sample Size for the Training Set

    • Not applicable. The device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. The device is not an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120818

    Validate with FDA (Live)

    Device Name
    OTIS-C PLUS
    Date Cleared
    2012-04-19

    (31 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OTIS-C Plus is intended to be used in conjunction with bone screws to provide fixation following Proximal Tibial Opening wedge osteotomies

    Device Description

    The OTIS-C Plus plate is designed for the stabilization of High Tibial Osteotomy with a medial approach. Anatomically shaped, thin and short, the OTIS-C Plus plate enables mini-invasive surgery. Its locking system provides immediate compression of the graft as well as stable fixation, thus allowing early weight-bearing. The design of the self-tapping OTIS-C Plus screws (unmodified) allows easy and reliable one step locking, without counter-nut, in a simple and concise approach. With its range of twelve screws lengths, fixation can be either mono or bi-cortical, upon the choice of the surgeon.

    AI/ML Overview

    This 510(k) submission describes a medical device, not a diagnostic AI/ML device. Therefore, the questions related to AI/ML specific performance criteria, ground truth, and expert evaluation are not applicable.

    Here's an analysis of the provided text in the context of device acceptance criteria and study information:

    Device Acceptance Criteria and Performance

    The document does not explicitly present a table of acceptance criteria with numerical targets. Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device through non-clinical performance testing. The criteria are implicitly derived from the established safety and effectiveness of the predicate.

    Acceptance CriteriaReported Device Performance
    Fretting Corrosion (comparable to predicate)Demonstrated that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.
    Torsional Yield Strength (comparable to predicate)Demonstrated that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.
    Ultimate Torque (comparable to predicate)Demonstrated that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.
    Insertion Torque (comparable to predicate)Demonstrated that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.
    Removal Torque (comparable to predicate)Demonstrated that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.
    Pull-Out Strength (comparable to predicate)Demonstrated that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.
    Mechanical Behavior (unchanged by modification)Finite Element Method demonstrates that modification to OTIS-C Plus plate do not alter its mechanical behavior.
    Fretting Corrosion (unchanged by modification)Finite Element Method demonstrates that modification to OTIS-C Plus plate do not alter its fretting corrosion.

    Study Description:

    1. Sample size used for the test set and the data provenance: Not explicitly stated as this is a non-clinical, mechanical testing study. The "test set" would refer to the physical samples of the OTIS-C Plus device and its components used for the performance testing. The provenance of the data is from SCIENCE FOR BIOMATERIALS, a French company. The study is prospective in the sense that the testing was performed on the newly designed modified device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is a mechanical engineering study, not an imaging or diagnostic study requiring expert interpretation for ground truth. The "ground truth" is established by the physical laws and engineering principles governing the materials and their interactions.
    3. Adjudication method for the test set: Not applicable. This refers to consensus methods for expert opinions, which is not relevant for mechanical testing.
    4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device or a diagnostic device involving human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
    6. The type of ground truth used: For the mechanical performance tests (fretting corrosion, torsional yield strength, etc.), the "ground truth" is the established scientific and engineering principles for material properties and device performance under specified conditions, as measured by standard laboratory tests. For the Finite Element Method (FEM) analysis, the "ground truth" or reference for comparison is the mechanical behavior and fretting corrosion of the predicate device (K100604).
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090994

    Validate with FDA (Live)

    Date Cleared
    2009-05-06

    (29 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LIGAFIX® interference screws are designed for the interference fixation of grafts in anterior cruciate ligament reconstruction.

    Device Description

    LIGAFIX® / ComposiTCP® range of products consists of resorbable cannulated screws available in several models designed for the interference fixation of grafts in anterior cruciate ligament reconstruction.

    LIGAFIX® / ComposiTCP® interference bone screw is made of a ceramic (beta-TCP) / polymer (Poly Lactic Acid -PLA) composite.

    LIGAFIX® / ComposiTCP® interference screw is available in several sizes and in two beta-TCP/ PLA ratios 60/40 (LIGAFIX® 60/ ComposiTCP® 60) and 30/70 (LIGAFIX® 30/ ComposiTCP® 30)

    LIGAFIX® / ComposiTCP® interference bone screws are supplied sterile and individually packaged in double heat sealed pouches.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the LIGAFIX® / ComposiTCP® Resorbable Interference Screw. This document is a regulatory submission for a medical device seeking substantial equivalence to existing devices, not a study report proving the device meets an acceptance criterion with detailed findings. Therefore, much of the requested information regarding specific acceptance criteria and detailed study parameters for performance is not present in the provided text.

    However, based on the information available, here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "Biological, mechanical and biocompatibility tests confirmed that LIGAFIX® / ComposiTCP® screws are highly biocompatible and presents the requisite strength to provide sustained fixation of the graft. LIGAFIX®/ ComposiTCP® screws strength retention profiles are compatible with the healing process."

    This is a high-level summary of the reported performance, but specific quantitative acceptance criteria (e.g., "tensile strength must be >X MPa") and corresponding reported numerical performance values are not detailed in the provided text.

    Acceptance Criterion (Inferred)Reported Device Performance
    Biocompatibility"highly biocompatible"
    Requisite strength for sustained graft fixation"presents the requisite strength to provide sustained fixation of the graft"
    Strength retention profile compatibility with healing process"strength retention profiles are compatible with the healing process"

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in the document. The text mentions "Biological, mechanical and biocompatibility tests" but does not detail the methodology, sample sizes, or data provenance (e.g., country of origin, retrospective/prospective nature) of these tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable as the document describes a regulatory submission for a physical medical device (interference screw) and not an AI/diagnostic software. "Ground truth" in the context of expert review for diagnostic accuracy is not relevant here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable for the reasons stated above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as the device is an interference screw, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable for the reasons stated above.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    This information is not applicable as the document pertains to a physical medical device. The "truth" for the performance claims would be established through laboratory testing (mechanical, biological) rather than expert review of images or pathology.

    8. The sample size for the training set:

    This information is not applicable as the document is for a physical medical device and does not involve AI or machine learning models that require training sets.

    9. How the ground truth for the training set was established:

    This information is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071155

    Validate with FDA (Live)

    Date Cleared
    2007-07-03

    (69 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K070507

    Validate with FDA (Live)

    Date Cleared
    2007-03-29

    (36 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LigaFix® interference screws are designed for the interference fixation of grafts in anterior cruciate ligament reconstruction.

    Device Description

    LIGAFIX® are resorbable cannulated screws made of ceramic (β-ΤCP) / polymer (Poly Lactic Acid) composite and designed for the interference fixation of grafts in anterlor cruciate ligament reconstruction.

    To better fit the surgeon's needs and surgical preferences and patient's anatomy the full range of LIGAFIX Interference screws comprise:

    • 2 material specifications: LIGAFIX® 30 (30% TCP/70% PLA) and LIGAFIX® 60 (60% o TCP/40% PLA)
    • 3 head design: standard reduced -- rounded ಂ
    • Diameters from 7 mm to 11 mm 0
    • Lengths from 20 mm to 35 mm ಂ

    LIGAFIX® interference screws are supplied sterile and individually packaged In double heat sealed pouches.

    AI/ML Overview

    This 510(k) Special submission is for an extension of the range of products for the LIGAFIX® Resorbable Interference Screw. It focuses on demonstrating substantial equivalence to previously cleared devices (K050407 and K061262) rather than presenting a de novo study with new acceptance criteria and performance data for a novel device.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment, which are typical for studies validating the performance of a new device, are largely not applicable in the context of this specific 510(k) submission.

    This submission asserts that the expanded product range (new material specifications, head designs, diameters, and lengths) of the LIGAFIX® Interference Screws shares the "same fundamental scientific technology, operating principle and intended use" as the predicate devices. The review process for such a submission typically involves evaluating whether the new product variations introduce any new questions of safety or effectiveness that would necessitate new performance studies. Since this document is a 510(k) summary, it doesn't contain detailed study reports.

    Here's a breakdown of why most of the requested points cannot be answered from the provided text, and what can be inferred:


    1. A table of acceptance criteria and the reported device performance

    • Not Applicable in this context. The document does not provide a table of acceptance criteria or reported device performance for a new study. The submission relies on the substantial equivalence to previously cleared devices which would have undergone performance testing during their initial clearance. The "performance" being demonstrated here is the similarity to the predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No new test set or associated data provenance is mentioned as this is a Special 510(k) for an expansion of an existing device line, not a new device requiring a new clinical or performance study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No test set or ground truth establishment by experts is described in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a resorbable interference screw, not an AI-powered diagnostic or assistive technology. Therefore, MRMC studies or AI-related effectiveness are entirely irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. As above, this is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable. No new ground truth is established for this submission. The "ground truth" for the clearance of the predicate devices would have been based on established engineering principles, biocompatibility testing, mechanical testing, and potentially animal or clinical studies to demonstrate safety and effectiveness for their intended use.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device; thus, there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. No training set is involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K061022

    Validate with FDA (Live)

    Date Cleared
    2006-05-24

    (41 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOSORB ® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure.

    BIOSORB ® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure stabilization of the defect in all plans.

    The BIO 1-KIT ® BIOSORB® filled cartridge is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This cartridge provides the surgeon with a convenient way to mix autologous blood or bone marrow with the BIOSORB ® Resorbable Void Filler and deliver the material to the orthopedic surgical site.

    Device Description

    BIOSORB ® bone void filler (BIO1 range of products) (K021963) is an osseo-conductive macroporous implant made of synthetic beta tri Calcium Phosphate (βTCP) indicated for Bone Void Filling.

    BIOSORB ® bone void filler presents a multidirectional interconnected porosity structure (45% and 70%), similar to that of the human cancellous bone. BIOSORB ® bone void filler implant slowly resorbs during the remodeling and bone defect repair process and is progressively replaced with bone and soft tissues. The progressive resorption within 6-12 months prevents premature resorption and failed filling.

    The impregnation of bone substitute with marrow is largely practiced by the surgeons. Usually this operation is done by immersion of the bone substitute in a container containing the osseous marrow. In order to facilitate this operation and to limit the contact of the mixture with the ambient air at the origin of the degradation of the cells and proteins contained in osseous marrow, S.B.M has developed a system allowing the and procession of autologous osseous marrow, the impregnation of the calcium phosphate pellets and the impaction of the mixture in the cavity to be filled. This system allows the impregnation of porous pellets with osseous marrow under aseptic conditions.

    BIO 1-Kit® comprises :

    • a pre-filled syringe of BIOSORB® (β-ΤCP porous ceramic pellets)
    • a connector with an adapter and a needle
    • a bone marrow aspiration needle

    BIO 1-Kit® is delivered sterile and intended for a single use.

    BIO 1-Kit® is available either with BIOSORB® granules (1.5 and 3.0 mm in diameter) or BIOSORB® blocks (4x4x4 mm) and in quantities ranging from 5 cc to 45 cc.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the BIO 1-KIT ® BIOSORB® filled cartridge:

    Acceptance Criteria and Device Performance for BIO 1-KIT ® BIOSORB® filled cartridge

    The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device rather than explicitly defining and meeting specific quantitative acceptance criteria for the BIO 1-KIT® BIOSORB® filled cartridge itself (the new device being reviewed). Instead, it relies heavily on the established safety and effectiveness of its main component, BIOSORB® bone void filler (which had its own prior 510(k) clearance, K021963), and the substantial equivalence of the new syringe system to a similar predicate.

    Therefore, the "acceptance criteria" presented below are inferred from the types of tests performed and the claims made about the device's functionality and safety, especially regarding its new features (the cartridge system).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from tests/claims)Reported Device Performance
    For BIOSORB® (bone void filler component):
    BiocompatibilityExcellent biocompatibility; non-toxic, non-allergenic, no inflammation, no adverse systemic effects.
    Material CharacteristicsConforms to ASTM F 1088-2004 (Standard Specification for Beta-Tricalcium Phosphate for Surgical Implantation).
    Osseo-integration & Bone ReplacementClinical and radiological data demonstrate osseo-integration and replacement by osseous tissues.
    Resorption ProfileProgressive resorption within 6-12 months.
    For BIO 1-KIT® (cartridge system component):
    Material Compatibility (syringe & contents)Assessed by infra-red spectrometry. Results imply compatibility and absence of particle release.
    Aspiration and Impregnation FunctionalityTesting performed; implicit successful performance for aspiration of autogenous blood/marrow and impregnation of pellets.
    SterilityDelivered sterile.
    Single UseIntended for single use.
    Substantial EquivalenceSubstantially equivalent to VITOSS-Filled cartridge K032130 in terms of intended use, material, design, and function.

    2. Sample Size Used for the Test Set and Data Provenance

    • For BIOSORB® (prior clearance K021963):
      • Sample Size: A "human clinical trial" was performed. The specific number of subjects/cases is not provided in this 510(k) summary.
      • Data Provenance: Not explicitly stated (e.g., country of origin). Since the submitter is French, it's possible this was an international or European trial, but no details are given. The trial was prospective, as it was conducted to investigate safety and effectiveness.
    • For BIO 1-KIT® cartridge (new clearance K061022):
      • Sample Size: Not specified for the "compatibility," "aspiration," and "impregnation" tests. These appear to be bench or lab-based tests, not clinical trials.
      • Data Provenance: Not specified beyond being performed by S.B.M. (the manufacturer). Retrospective or prospective is not applicable for these bench tests.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • For BIOSORB® (prior clearance K021963):
      • No information is provided on the number and qualifications of experts involved in establishing ground truth for the clinical trial data discussed. The text refers to "clinical and radiological data" but doesn't detail the expert review process.
    • For BIO 1-KIT® cartridge (new clearance K061022):
      • Not applicable as no clinical ground truth was established for the specific cartridge functionality tests described. These were likely internal engineering or quality control assessments.

    4. Adjudication Method for the Test Set

    • For BIOSORB® (prior clearance K021963): Not described.
    • For BIO 1-KIT® cartridge (new clearance K061022): Not applicable, as there were no clinical test sets requiring adjudication described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The submission focuses on substantial equivalence of a medical device (bone void filler cartridge) for its physical and functional properties, not on AI-assisted diagnostic or therapeutic interpretation by human readers. Therefore, the concept of "effect size of how much human readers improve with AI vs without AI assistance" is not relevant to this device submission.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, a standalone algorithm-only performance study was not done. This device is a physical medical product (a cartridge system for bone void filler), not a software algorithm or AI-driven tool.

    7. The Type of Ground Truth Used

    • For BIOSORB® (prior clearance K021963):
      • Clinical and Radiological Data: The summary mentions "Clinical and radiological data demonstrate the biocompatibility, the osseo integration and the resorption of BIOSORB® bone void filler." This implies a combination of direct clinical observations (e.g., patient outcomes, inflammatory response) and interpretation of medical imaging (e.g., X-rays to assess bone consolidation and resorption). Pathology results or specific outcome measures (beyond "stable osseous fusion") are not explicitly detailed as the ground truth, but rather general clinical/radiological findings.
    • For BIO 1-KIT® cartridge (new clearance K061022):
      • Bench Test Results/Specifications: The ground truth for tests like "compatibility" (infra-red spectrometry) and "aspiration/impregnation" would be based on predefined physical or chemical specifications and successful functional operation, not clinical ground truth.

    8. The Sample Size for the Training Set

    • Not applicable. This submission describes a physical medical device. There is no mention of an algorithm or machine learning component requiring a "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as no training set for an AI algorithm is involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K051304

    Validate with FDA (Live)

    Date Cleared
    2005-06-03

    (15 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Ophthalmic. The OIS provides real-time, high-resolution images of the intraocular anatomy and is, therefore, useful in evaluation of pathologies of the anterior segment such as trauma, tumors, cysts, and glaucoma. It can also be used during ophthalmic surgery to monitor the changes in anatomical structures such as the anterior angle, Schlemm's canal and related outflow structures of the eye, for glaucoma management.

    Device Description

    The Ophthalmic Imaging System (OIS) is an ultrasound system designed for noninvasive imaging of the anterior segment of the eye through the use of standard ultrasound imaging pulse-echo techniques. The system consists of a handpiece containing a high-frequency transducer, connected to interfacing electronics in the Transducer Interface Module (TIM) which is in turn connected to a personal computer with video display. Software developed by iScience Surgical runs on the computer controlling the system and generating the image. The OIS is used with a sterile disposable tissue interface (Imaging Gel Cap) which is attached to the handpiece for direct contact with the eye during imaging. The OIS may also be used with a sterile Sheath Adapter Assembly for sterile field applications.

    AI/ML Overview

    The provided text is a 510(k) summary for the iScience Surgical Ophthalmic Imaging System (OIS). It does not contain information about acceptance criteria or a study that proves the device meets such criteria. Instead, it focuses on the device description, its intended use, and its substantial equivalence to predicate devices, which is the basis for 510(k) clearance.

    Therefore, I cannot provide the requested information from the text provided as it does not contain details about:

    1. Acceptance criteria and reported device performance: This document discusses the intended use and technological characteristics but no specific performance metrics or acceptance thresholds.
    2. Sample size, data provenance, number of experts, adjudication method for a test set: These details are typically found in a clinical study report, which is not present here.
    3. MRMC comparative effectiveness study: No mention of such a study.
    4. Standalone performance study: No mention of such a study.
    5. Type of ground truth: Not applicable as no performance study is described.
    6. Sample size for the training set: Not applicable as no AI/algorithm training is described.
    7. How ground truth for the training set was established: Not applicable.

    The document is a regulatory submission demonstrating substantial equivalence to existing devices, not a performance study report with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020881

    Validate with FDA (Live)

    Date Cleared
    2002-03-27

    (9 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Science Incorporated Personal Infusor Local Pain Management Procedural Kit is intended to provide continuous infusion of physician-prescribed local anesthetic medications directly into the intraoperative site of a patient for management of postoperative pain. Medication is delivered percutaneously, through the infusor's administration line, which is attached to a catheter.

    The Personal Infusor is portable and is suitable for use by ambulatory patients.

    The Personal Infusor is intended for single use only.

    The Personal Infusor is not intended for rapid infusion, intravenous or intra-arterial drug infusion, for the delivery of blood, blood products, lipids, or fat emulsions, or for use by patients with a history of allergic reactions to anesthetics.

    Device Description

    The Science Incorporated Personal Infusor Local Pain Management Procedural Kit consists of a Personal Infusor provided in a kit with medical device components including a catheter set, catheter introducer, 60 mL syringe, dressings and accessories. Each of the additional medical device components is purchased in finished form and is packaged with the Personal Infusor in the Personal Infusor Local Pain Management Procedural Kit.

    The Personal Infusors provided in the Personal Infusor Local Pain Management Procedural Kit will be offered in three volume / flow rate configurations.

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria and the study that proves the device meets them, as requested in your prompt. The document is primarily a 510(k) summary and the FDA's substantial equivalence letter for a medical device (Personal Infusor Local Pain Management Procedural Kit). It describes the device, its intended use, and its predicate device, but it does not include details on performance studies, acceptance criteria, or ground truth establishment.

    Therefore, I cannot populate the table or answer the specific questions about sample size, expert qualifications, adjudication methods, or MRMC studies.