Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K202453

    Validate with FDA (Live)

    Date Cleared
    2021-03-24

    (209 days)

    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine500ng/mL
    Oxazepam300 ng/mL
    Cocaine150ng/mL
    Marijuana50 ng/mL
    Methamphetamine500ng/mL
    Morphine300ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxy-methamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL
    2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    The tests are intended for over-the-counter use.

    SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine. Oxazepam. Marijuana. Methamphetamine, Morphine. Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

    Drug(Identifier)Cut-off level
    Amphetamine500ng/mL
    Oxazepam300 ng/mL
    Cocaine150ng/mL
    Marijuana50 ng/mL
    Methamphetamine500ng/mL
    Morphine300ng/mL
    Oxycodone100 ng/mL
    Secobarbital300 ng/mL
    Buprenorphine10 ng/mL
    Methylenedioxy-methamphetamine500 ng/mL
    Phencyclidine25 ng/mL
    Methadone300 ng/mL
    Nortriptyline1000 ng/mL
    d-Propoxyphene300 ng/mL
    2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    The tests are intended for over-the-counter use.

    Device Description

    The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    AI/ML Overview

    The provided document describes the performance characteristics and studies for the SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup. It does not describe an AI/ML device but rather an in-vitro diagnostic device (IVD) for drug screening. Therefore, several of the requested categories for AI/ML device evaluation are not applicable (e.g., number of experts, adjudication method, MRMC study, standalone performance, training set).

    Here's the information extracted from the document, tailored to the nature of the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for qualitative drug tests are typically defined by precision around the cutoff concentration. The device is expected to consistently classify samples below the cutoff as negative and above the cutoff as positive. For samples near the cutoff, some variability in classification is expected.

    Test Parameter / Drug (Cut-off)Acceptance Criteria (Implicit from study design)Reported Device Performance (Precision Study - Example for Dip Card, Lot 1)
    PrecisionSamples < -25% Cut-off: 100% negative calls. Samples > +25% Cut-off: 100% positive calls. Samples within +/-25% of Cut-off: Expected variability.Amphetamine 500: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Cocaine 150: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Methamphetamine 500: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Morphine 300: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive)
    Lay-User Study (Accuracy)For samples far from cutoff (e.g., -50% and +50%), approximately 100% correct results. For samples near cutoff (e.g., -25% and +25%), high accuracy (e.g., >90%).AMP500: -50% Cutoff: 100% correct (0 Pos, 170 Neg)+50% Cutoff: 100% correct (40 Pos, 0 Neg)-25% Cutoff: 90% correct (2 Pos, 18 Neg)+25% Cutoff: 95% correct (19 Pos, 1 Neg)(Similar results reported for COC150, THC, BAR, BZO, MET500, MTD, MOP300, MDMA, OXY, BUP, PCP, TCA, PPX, EDDP)

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Study (Test Set):

      • For each drug (Amphetamine, Cocaine, Methamphetamine, Morphine), 9 concentrations were tested around the cutoff (-100%, -75%, -50%, -25%, Cut-off, +25%, +50%, +75%, +100%).
      • For each concentration, tests were performed two runs per day for 25 days, using 3 different lots of the device. This equates to 50 replicates per concentration per lot, totaling 450 tests per lot per drug.
      • Total replicates for 4 drugs (AMP, COC, MET, MOP): 9 concentrations x 50 replicates/concentration x 3 lots = 1350 tests per drug. (Some data for other drugs were referenced from prior 510(k)s: K182654, K181968, K153646, K201120).
      • Data Provenance: The document states "in-house" for comparison studies and "urine samples were prepared by spiking drug in negative samples" for precision studies. This suggests a controlled laboratory setting (likely prospective, artificial samples). The document does not specify the country of origin of the data.
    • Method Comparison Study (Clinical/Test Set):

      • 80 "unaltered clinical samples" were used for each drug. These samples were split into categories: 10 negative, 10 low negative, 20 near cutoff negative, 20 near cutoff positive, 20 high positive for each drug.
      • Total samples per device type (Dip Card or Cup) for 4 drugs mentioned: 80 samples x 4 drugs = 320 samples per device type.
      • Data Provenance: "in-house" and "unaltered clinical samples," implying real-world samples but processed within the manufacturer's lab. The document does not specify the country of origin or whether these clinical samples were retrospective or prospectively collected for the study.
    • Lay-User Study (Test Set):

      • 310 lay persons participated for each device format (Dip Card and Cup).
      • Urine samples were prepared at 7 concentrations: negative, +/-75%, +/-50%, +/-25% of the cutoff.
      • Total samples: For each drug, the number of samples varied across concentrations. For example, for AMP500, 20 samples at -100% cutoff, 20 at -75%, 170 at -50%, 20 at -25%, 20 at +25%, 40 at +50%, 20 at +75%. This totals 310 samples per drug per device format.
      • Data Provenance: "at three intended user sites." Samples were "prepared by spiking drugs into drug free-pooled urine specimens," making them artificial but intended to mimic a range of concentrations. This suggests a prospective study design, mimicking real-world use conditions but with controlled samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Precision Study: Ground truth was established by preparing urine samples with known drug concentrations confirmed by LC/MS. No human experts were involved in establishing ground truth, as it was an analytical study.
    • Method Comparison Study: The ground truth for clinical samples was established by LC/MS results. No mention of human experts for ground truth.
    • Lay-User Study: Ground truth was established by LC/MS results of the prepared spiked urine samples.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This device is a qualitative diagnostic test read directly by users, not an AI/ML imaging device requiring expert adjudication. In the method comparison study, three laboratory assistants "ran" the samples, implying they performed the test, but the ground truth was LC/MS, not their consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device, and no MRMC study comparing human readers with and without AI assistance was mentioned. The lay-user study evaluated the device's performance with lay users, not an "AI assistance" scenario.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The device is a physical, lateral flow immunochromatographic assay. Its performance inherently involves a human interpreting the result line, even if it's a simple positive/negative visual interpretation. It is not an algorithm-only device. The precision and method comparison studies evaluate the device's analytical performance, which is its inherent "standalone" capability.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • LC/MS (Liquid Chromatography-Mass Spectrometry) was used as the ground truth method to confirm drug concentrations in both spiked samples (for precision and lay-user studies) and clinical samples (for method comparison studies). This is a highly accurate and quantitative analytical method.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a "training set" in the machine learning sense. The device is based on immunoassay principles.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201120

    Validate with FDA (Live)

    Date Cleared
    2020-05-27

    (30 days)

    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations listed. The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations listed. The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    Device Description

    The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    AI/ML Overview

    The provided text describes the performance characteristics of the SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup. The document details analytical performance (precision, linearity, stability, interference, specificity, effect of specific gravity and pH) and comparison studies (method comparison and lay-user study).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each analytical performance characteristic are implicitly defined by the successful results reported in the study. The study aims to demonstrate that the device performs as expected according to its intended use and analytical specifications.

    • Precision:

      • Acceptance Criteria for -100% to -25% Cut-off: All samples should test negative.
      • Acceptance Criteria for +25% to +100% Cut-off: All samples should test positive.
      • Acceptance Criteria for Cut-off (nominal concentration): Approximately 50% positive and 50% negative results (reflecting the nature of the cut-off).
      • Reported Device Performance (for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine as an example):
        • Dip Card (Lot 1, 2, 3):
          • -100% Cut-off to -25% Cut-off: 50-/0+ (All negative)
          • +25% Cut-off to +100% Cut-off: 50+/0- (All positive)
          • Cut-off: 24-/26+, 25-/25+, 26-/24+ (Split between positive and negative, as expected)
        • Cup (Lot 1, 2, 3):
          • -100% Cut-off to -25% Cut-off: 50-/0+ (All negative)
          • +25% Cut-off to +100% Cut-off: 50+/0- (All positive)
          • Cut-off: 27-/23+, 24-/26+, 26-/24+ (Split between positive and negative, as expected)
    • Interference:

      • Acceptance Criteria: No interference from common physiological or pathological substances at specified concentrations.
      • Reported Device Performance: No interference observed for a comprehensive list of compounds at 100ug/mL (except albumin at 100mg/dL and ethanol at 1% volume).
    • Specificity (Cross-Reactivity):

      • Acceptance Criteria: Related compounds should either not cross-react or cross-react at concentrations significantly higher than the cut-off, as defined by medical relevance.
      • Reported Device Performance (for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine as an example):
        • Methadone: Positive at 300000 ng/mL (0.1% cross-reactivity)
        • EMDP: Positive at 300000 ng/mL (0.1% cross-reactivity)
        • Doxylamine, Disopyramide, LAAM HCl, Alpha Methadol: Positive at >100,000 ng/mL (<0.3% cross-reactivity)
    • Effect of Urine Specific Gravity and Urine pH:

      • Acceptance Criteria: Reliable results (positive above cut-off, negative below cut-off) across a specified range of specific gravity and pH.
      • Reported Device Performance: All samples at and above +25% Cut-Off were positive, and all samples at and below -25% Cut-Off were negative, for specific gravity from 1.000 to 1.035 and pH 4 to 9.
    • Method Comparison Study (Clinical Samples):

      • Acceptance Criteria: High agreement (minimal discordant results) between the device results and LC/MS (Liquid Chromatography/Mass Spectrometry) results, especially for samples near the cut-off.
      • Reported Device Performance (for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine as an example):
        • Dip Card: High concordance with LC/MS, with few discordant results (e.g., Viewer A: 1 false positive at 291 ng/mL; Viewer B: 1 false negative at 309 ng/mL).
        • Cup: High concordance with LC/MS, with few discordant results (e.g., Viewer A: 1 false positive at 285 ng/mL; Viewer B: 1 false negative at 372 ng/mL).
    • Lay-user Study:

      • Acceptance Criteria: High percentage of correct results for samples well below and well above the cut-off, and reasonable performance for samples near the cut-off, when interpreted by lay users. Ease of understanding instructions.
      • Reported Device Performance (for specific drugs like Amphetamine (AMP), Cocaine (COC), THC, BAR, BZO, MET, MTD, MDMA, OXY, OPI, BUP, PCP, TCA, PPX, EDDP as examples):
        • For samples at -100% to -50% Cut-off and +50% to +75% Cut-off: 100% correct results across all drugs.
        • For samples at -25% Cut-off: generally 90-95% correct negative results (some false positives).
        • For samples at +25% Cut-off: generally 85-95% correct positive results (some false negatives).
        • All lay users indicated the device instructions were easy to follow (Flesch-Kincaid Grade Level 7).

    2. Sample Sizes Used for the Test Set and the Data Provenance

    • Precision Study: For each drug and concentration, two runs per day for 25 days per device (Dip Card and Cup), implying a total of 50 tests per concentration per lot. With 3 lots, this is 150 tests per concentration per format. There were 9 concentrations tested. So 150*9 = 1350 tests per drug per format.
    • Interference Study: Not explicitly stated, but tests were performed on "three batches of each device" for various interfering substances spiked into drug-free and drug-spiked urine samples.
    • Specificity Study: Not explicitly stated how many tests per compound, but "three batches of each device" were used.
    • Effect of Urine Specific Gravity and Urine pH Study: "three lots of each device" were used for samples spiked with target drugs at +/- 25% Cut-Off levels across the range of specific gravity and pH.
    • Method Comparison Study: 80 "unaltered clinical samples" for each drug (40 negative and 40 positive relative to cut-off). This was conducted "in-house." The provenance of the clinical samples is not specified beyond being "clinical samples."
    • Lay-user Study: 310 lay persons for each device format. The samples were prepared by spiking drugs into "drug free-pooled urine specimens." The concentrations were confirmed by LC/MS. Thus, the samples were prepared and not naturally occurring clinical samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • For the Precision, Interference, Specificity, and Effect of Urine Specific Gravity and Urine pH Studies: The ground truth was established by precise laboratory preparation of samples where drug concentrations were known and confirmed by LC/MS. There were no human experts classifying these samples for ground truth.
    • For the Method Comparison Study: The ground truth was established by LC/MS results. LC/MS is a highly accurate and definitive analytical method. The text does not mention human experts establishing ground truth for these samples; it was based on the objective chemical analysis. The "three laboratory assistants" were operators performing the device tests, not establishing ground truth.
    • For the Lay-user Study: The ground truth was the known drug concentration of the prepared urine samples, confirmed by LC/MS.

    4. Adjudication Method for the Test Set

    • Precision, Interference, Specificity, and Effect of Urine Specific Gravity and Urine pH Studies: No human adjudication was mentioned as the ground truth was based on precisely prepared samples with known concentrations.
    • Method Comparison Study: The device results were compared directly to LC/MS results. There was no adjudication mentioned involving human experts to reconcile discrepancies between device and LC/MS. Discrepancies were simply reported as "discordant results."
    • Lay-user Study: Comparisons were made against the known LC/MS confirmed concentrations of the prepared samples. No human adjudication was mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is a rapid diagnostic test (lateral flow immunoassay) for drug detection, where the "reading" is a visual interpretation of lines by a user (or laboratory assistant in the method comparison study). It does not appear to involve AI or sophisticated image analysis where human readers would be "assisted" by AI. The "lay-user study" involved multiple readers (310 lay persons), but it was designed to assess the ease of use and accuracy of the device itself by untrained individuals, not to compare human performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • This device is a lateral flow immunoassay intended for visual interpretation. It does not have an "algorithm only" component in the sense of a software-based AI system. Its performance inherently involves a "human-in-the-loop" to view and interpret the test lines.

    7. The Type of Ground Truth Used

    • The primary ground truth used throughout the studies (precision, interference, specificity, method comparison, and lay-user study) was LC/MS (Liquid Chromatography/Mass Spectrometry), which is a definitive analytical method for confirming drug concentrations in urine samples. For analytical studies, ground truth was also established by careful preparation of samples with known drug concentrations.

    8. The Sample Size for the Training Set

    • This document describes performance studies for an immunochromatographic assay, which is a chemical/biological test, not a machine learning or AI model. Therefore, there is no "training set" in the context of AI model development. The assays are "trained" during their manufacturing and chemical formulation processes to react at specific cut-off levels.

    9. How the Ground Truth for the Training Set Was Established

    • As stated above, this device is not an AI/ML product and thus does not have a "training set" with ground truth in that context. The "training" of the assay refers to its inherent chemical design to react to certain drug concentrations, which relies on established analytical chemistry principles and known drug properties.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181968

    Validate with FDA (Live)

    Date Cleared
    2018-08-20

    (27 days)

    Product Code
    Regulation Number
    862.3100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital and Methadone in human urine at the cutoff concentrations of:

    Drug (Identifier)Cut-off level
    Amphetamine1000 ng/mL
    Oxazepam300 ng/mL
    Cocaine300 ng/mL
    Cannabinoids50 ng/mL
    Methamphetamine1000 ng/mL
    Morphine2000 ng/mL
    Secobarbital300 ng/mL
    Methadone300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Oxazepam and Secobarbital when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cannabinoids, Methamphetamine, Morphine, Secobarbital and Methadone in human urine at the cutoff concentrations of:

    Drug (Identifier)Cut-off level
    Amphetamine1000 ng/mL
    Oxazepam300 ng/mL
    Cocaine300 ng/mL
    Cannabinoids50 ng/mL
    Methamphetamine1000 ng/mL
    Morphine2000 ng/mL
    Secobarbital300 ng/mL
    Methadone300 ng/mL

    Configuration of SAFECARE® Multi-Drug Urine Test Cup combination of the above listed drug analytes.

    The test may yield positive results for the prescription drugs Oxazepam and Secobarbital when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    Device Description

    The SAFECARE Dip Card Tests and SAFECARE Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Secobarbital and Methadone (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    AI/ML Overview

    The document describes the SAFECARE Multi-Drug Urine Test Dip Card and SAFECARE Multi-Drug Urine Test Cup devices, which are qualitative lateral flow immunochromatographic assays for detecting various drugs in human urine.

    Here's an analysis of the acceptance criteria and study data provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of a minimum performance threshold (e.g., "sensitivity must be >90%"). Instead, it presents the results of precision, specificity, linearity, stability, interference, and comparison studies, implying that the observed performance across these studies is considered acceptable for substantial equivalence. For the qualitative tests, the key performance indicator is the ability to correctly identify samples both above and below the cut-off concentrations.

    Below is a summary of the reported device performance for selected drugs based on "Precision" and "Lay-user study" data. The "Precision" study shows how consistently the device performs near the cut-off, while the "Lay-user study" assesses the device's accuracy when used by the intended over-the-counter users.

    Acceptance Criteria (Implied for consistency and accuracy related to cut-off)

    • Precision: High agreement (ideally 100% negative/positive) for samples far from the cutoff (e.g., -100%, -75%, +75%, +100% cutoff). Some variability (mixed positive/negative results) is expected and acceptable near the cutoff (e.g., -25%, cutoff, +25%).
    • Lay-user Performance (Percentage of correct results): High percentage of correct results, especially for samples significantly above or below the cutoff. Some deviation might be acceptable near the cutoff.

    Reported Device Performance (Excerpt for Secobarbital Dip Card and for all drugs in Lay-user study)

    Precision Study - Secobarbital Dip Card (Illustrative Example from document, similar patterns for other drugs)

    Concentration % of Cut-offLot 1 (Negative/Positive)Lot 2 (Negative/Positive)Lot 3 (Negative/Positive)
    -100% Cut-off50-/0+50-/0+50-/0+
    -75% Cut-off50-/0+50-/0+50-/0+
    -50% Cut-off50-/0+50-/0+50-/0+
    -25% Cut-off50-/0+50-/0+50-/0+
    Cut-off25-/25+26-/24+26-/24+
    +25% Cut-off50+/0-50+/0-50+/0-
    +50% Cut-off50+/0-50+/0-50+/0-
    +75% Cut-off50+/0-50+/0-50+/0-
    +100% Cut-off50+/0-50+/0-50+/0-

    Interpretation for Precision (e.g., Secobarbital Dip Card):

    • For samples far below the cutoff (-100%, -75%, -50%, -25%), all 50 tests consistently showed negative results (50-/0+), indicating perfect agreement for negative samples sufficiently below the cutoff.
    • For samples far above the cutoff (+25%, +50%, +75%, +100%), all 50 tests consistently showed positive results (50+/0-), indicating perfect agreement for positive samples sufficiently above the cutoff.
    • At the exact cutoff, there was an expected mix of positive and negative results (e.g., 25-/25+, 26-/24+), demonstrating the device's performance around the critical concentration.

    Lay-user Study - Percentage of Correct Results (Summary across drugs for Dip Card)

    Drug% of CutoffPercentage of Correct Results (%)
    Amphetamine-100%100
    -75%100
    -50%100
    -25%95 (19 negative out of 20)
    +25%85 (17 positive out of 20)
    +50%100
    +75%100
    Cocaine-100%100
    -75%100
    -50%100
    -25%90 (18 negative out of 20)
    +25%90 (18 positive out of 20)
    +50%100
    +75%100
    Cannabinoids-100%100
    -75%100
    -50%100
    -25%90 (18 negative out of 20)
    +25%95 (19 positive out of 20)
    +50%100
    +75%100
    Secobarbital-100%100
    -75%100
    -50%100
    -25%95 (19 negative out of 20)
    +25%90 (18 positive out of 20)
    +50%100
    +75%100
    Oxazepam-100%100
    -75%100
    -50%100
    -25%85 (17 negative out of 20)
    +25%95 (19 positive out of 20)
    +50%100
    +75%100
    Methamphetamine-100%100
    -75%100
    -50%100
    -25%85 (17 negative out of 20)
    +25%90 (18 positive out of 20)
    +50%100
    +75%100
    Methadone-100%100
    -75%100
    -50%100
    -25%95 (19 negative out of 20)
    +25%90 (18 positive out of 20)
    +50%100
    +75%100
    Morphine-100%100
    -75%100
    -50%100
    -25%95 (19 negative out of 20)
    +25%95 (19 positive out of 20)
    +50%100
    +75%100

    Interpretation for Lay-user Study:
    The lay-user study for both Dip Card and Cup formats consistently shows 100% correct results for samples well above (+50%, +75%) and well below (-50%, -75%, -100%) the cutoff. Near the cutoff point (-25% and +25%), the percentage of correct results typically drops to between 85% and 95%, which is expected given the concentrations are close to the detection threshold. The study indicates the devices perform as intended for over-the-counter use.

    2. Sample size used for the test set and the data provenance

    • Precision Study: For each drug and each device type (Dip Card/Cup), the samples were prepared at 9 different concentrations relative to the cut-off (-100%, -75%, -50%, -25%, cut-off, +25%, +50%, +75%, +100%). For each concentration, tests were performed "two runs per day for 25 days per device in a randomized order." This implies a total of 50 tests per concentration per drug per device, for each of the three lots.
      • Example for Secobarbital Dip Card: 9 concentrations * 50 tests/concentration * 3 lots = 1350 test results.
      • Data Provenance: "These samples were prepared by spiking drug in negative samples." The document specifies these are "spiked" samples into "negative samples" (presumably drug-free urine), and concentrations were confirmed by LC/MS. This suggests laboratory-controlled samples (retrospective fabrication) rather than patient-derived clinical samples.
    • Comparison Studies (Lab Assistant/Clinical Performance): For each drug, "Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug."
      • This means 80 clinical samples per drug were used.
      • Data Provenance: "unaltered clinical samples." The document states this was "performed in-house," but does not specify the country of origin. Given the submitter is Safecare Biotech (Hangzhou) Co. Ltd. from China, the in-house clinical samples likely originated from China. The samples were compared to LC/MS results.
    • Lay-user Study: "A lay user study was performed at three intended user sites with 240 lay persons for each device format."
      • The test set comprised urine samples prepared at 7 different concentrations relative to the cut-off (negative, +/-75%, +/-50%, +/-25% of the cutoff). The document indicates for each concentration, approximately 20 to 100 samples were tested for each drug (e.g., 20 at -100% cutoff, 100 at -50% cutoff, 40 at +50% cutoff, totaling around 20 * 5 + 100 * 2 = 300 test results per drug).
      • Data Provenance: "Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens." Concentrations confirmed by LC/MS. This implies laboratory-controlled spiked samples, similar to the precision study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Precision Studies & Lay-user Study: The ground truth (drug concentration) was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is a highly accurate analytical method. The document does not specify human experts for establishing ground truth for these studies, as the LC/MS directly provides quantitative measurements.
    • Comparison Studies (Lab Assistant/Clinical Performance): The ground truth for the "unaltered clinical samples" was established by LC/MS results. The study mentions "three laboratory assistants" who "ran" the tests and "compared to LC/MS results," but these assistants are the operators of the device under test, not the ground truth experts.

    4. Adjudication method for the test set

    • Precision Studies & Lay-user Study: The ground truth was based on objective analytical measurements (LC/MS). There was no human "adjudication" in the sense of reconciling differing expert opinions. The comparison was directly between the device's qualitative result and the known quantitative drug concentration relative to the cut-off.
    • Comparison Studies (Lab Assistant/Clinical Performance): The document does not describe an explicit adjudication method. The results from the device, interpreted by "three laboratory assistants," were directly compared to the LC/MS results, which served as the gold standard. Discordant results are individually reported (e.g., Viewer A found Positive, LC/MS was 289 ng/mL), but no adjudication of discrepant viewer readings against each other or a higher authority is mentioned. Each viewer's interpretation was compared against the LC/MS.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a rapid diagnostic test (lateral flow immunochromatographic assay), not an AI-assisted diagnostic tool. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers was not performed or described. The "readers" in the "Comparison Studies" are "three laboratory assistants" interpreting the visual lines on the test.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a manual, visually interpreted lateral flow assay. It does not employ an algorithm for interpretation. The performance is inherently "standalone" in the sense that the device produces a visual result, but it requires human interpretation. The "Lay-user study" specifically assesses performance without expert human-in-the-loop, relying on typical users' interpretation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The primary ground truth used in all analytical and comparative studies (Precision, Comparison, Lay-user) was Liquid Chromatography-Mass Spectrometry (LC/MS) results. This is a gold standard analytical method for precise and accurate quantification of drug concentrations.
    • For the comparison studies, "unaltered clinical samples" were used, and their status (positive/negative relative to cutoff) was determined by LC/MS.

    8. The sample size for the training set

    • The document does not describe any "training set." These are diagnostic devices that do not involve machine learning or AI models requiring a training phase for their interpretation logic. Their "training" is in their manufacturing process and the inherent biochemical reactions.

    9. How the ground truth for the training set was established

    • Not Applicable as there is no training set described for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1