510(k) Data Aggregation

    K Number
    K222944

    Validate with FDA (Live)

    Date Cleared
    2024-10-18

    (753 days)

    Product Code
    Regulation Number
    878.5030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SafePath Suturing System - Silk Suture: The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues.
    SafePath Suturing System - Polyamide (Nylon) Suture: The SafePath Suturing System is intended for use in placement of a nylon suture in the skin and subcutaneous tissues.
    SafePath Suturing System - Polypropylene Suture: The SafePath Suturing System is intended for use in placement of a polypropylene suture in the skin and subcutaneous tissues.

    Device Description

    The silk suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of the organic protein, fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. The silk suture is silicone coated, braided and dyed (black) with logwood extract. The silk suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP size 2-0. The suture is provided as a 36 in. Jength, swaged on a single-armed 1½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.

    The nylon suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of long chain aliphatic polymers - nylon 6,6. SafePath nylon sutures are dyed with Logwood extract, FDA approval number §73.1410. The nylon suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a singlearmed ½ circle. 26 mm reverse cutting needle. The suture is pre-loaded in a handheld, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.

    The polypropylene suture provided with the SafePath Suturing System is a monofilament synthetic non-absorbable, sterile surgical suture composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin. The molecular formula is (C3H6)n. The Safepath suture is dyed with [Phthalocyaninato(2-)] copper, FDA approval number §74.3045. The polypropylene suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a single-armed ½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.

    Finished devices may be packaged individually, or in multi-unit cartons or procedure packs. The SafePath Suturing System is designed to place a suture in skin and subcutaneous tissue.

    AI/ML Overview

    This document discusses the FDA clearance for the SafePath Suturing System, which includes silk, polyamide (nylon), and polypropylene sutures. However, the provided text does not contain information about a study based on AI/ML. All the performance criteria and testing mentioned relate to the physical and biological properties of the sutures and the suturing system, not to an AI component.

    Therefore, I cannot answer your request with the specific details about acceptance criteria and studies related to AI/ML performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.

    The document highlights the following non-clinical performance criteria and testing for the sutures:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (General)Reported Device Performance
    USP Nonabsorbable Sutures - DiameterMet or exceeded USP criteria
    USP Sutures - Needle AttachmentMet or exceeded USP criteria
    USP Sutures - Tensile StrengthMet or exceeded USP criteria
    Biocompatibility - CytotoxicityMet test criteria
    Biocompatibility - SensitizationMet test criteria
    Biocompatibility - IrritationMet test criteria
    Biocompatibility - Systemic ToxicityMet test criteria
    Biocompatibility - HemocompatibilityMet test criteria
    Biocompatibility - PyrogenicityMet test criteria
    Biocompatibility - EndotoxicityMet test criteria

    2. Sample size used for the test set and the data provenance: Not applicable as this is a non-AI/ML device. The testing was physical and biological, not based on data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is a non-AI/ML device. Ground truth as typically understood in AI/ML is not relevant here.

    4. Adjudication method for the test set: Not applicable as this is a non-AI/ML device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is a non-AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as this is a non-AI/ML device.

    7. The type of ground truth used: For physical testing, ground truth would be established by the specifications of the USP standards. For biocompatibility, it's about meeting pre-defined safety thresholds as per recognized standards (e.g., ISO 10993).

    8. The sample size for the training set: Not applicable as this is a non-AI/ML device.

    9. How the ground truth for the training set was established: Not applicable as this is a non-AI/ML device.

    The study described is a non-clinical performance testing study that evaluated the physical and biocompatibility characteristics of the sutures and the suturing system. The conclusions state that the device met or exceeded USP criteria for physical properties and met the test criteria for a battery of biocompatibility evaluations. The safety evaluation for the nylon and polypropylene sutures leveraged information from a previously cleared device (K180701) and used identical raw materials, packaging, and sterilization conditions as the predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180701

    Validate with FDA (Live)

    Date Cleared
    2018-06-25

    (101 days)

    Product Code
    Regulation Number
    878.5030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues.

    Device Description

    The silk suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of the organic protein, fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. The silk suture is silicone coated, braided and dyed (black) with logwood extract. The silk suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP size 2-0. The suture is provided as a 30 in. length, swaged on a single-armed ½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.

    Finished devices may be packaged individually, or in multi-unit cartons or procedure packs. The SafePath Suturing System is designed to place a suture in skin and subcutaneous tissue.

    The System Includes:

    (2) Suturing Device Pre-loaded with one (1) 30 in. strand of USP size 2-0 black braided silk suture Swaged on a ½ circle, 26 mm reverse cutting needle

    AI/ML Overview

    The provided document describes the SafePath Suturing System, a medical device. Based on the content, the acceptance criteria and the study that proves the device meets those criteria are focused on physical properties and biocompatibility of the suture itself, rather than AI/algorithm performance. There is no mention of an AI component in this device.

    Therefore, the requested information regarding AI-specific criteria (such as multi-reader multi-case studies, human-in-the-loop performance, standalone algorithm performance, number of experts for ground truth, adjudication methods, and training set details) is not applicable to this device.

    Here's a breakdown of the acceptance criteria and study information that is applicable:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (USP Standards)Reported Device Performance
    Nonabsorbable Sutures - DiameterMet or exceeded USP criteria
    Sutures - Needle AttachmentMet or exceeded USP criteria
    Tensile StrengthMet or exceeded USP criteria
    Biocompatibility Testing (ISO-10993 Guidance)
    CytotoxicityMet test criteria
    SensitizationMet test criteria
    IrritationMet test criteria
    Systemic ToxicityMet test criteria
    HemocompatibilityMet test criteria
    PyrogenicityMet test criteria
    EndotoxicityMet test criteria

    2. Sample Size Used for the Test Set and Data Provenance:

    The document refers to "Physical Testing" and "Biocompatibility Testing," but does not specify the sample sizes used for these tests.
    The data provenance is implied to be laboratory testing of the device components, not human or patient data.

    • Data Provenance: Laboratory testing of the SafePath Suturing System components (sutures and overall system).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable in the context of device physical and biocompatibility testing. Ground truth here refers to established scientific/engineering standards (USP and ISO).

    4. Adjudication Method for the Test Set:

    Not applicable. Testing appears to be based on adherence to quantitative standards rather than expert consensus on interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, this type of study is for evaluating diagnostic or interpretive AI systems. The SafePath Suturing System is a physical medical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    No, this pertains to AI/algorithm performance, which is not relevant to this device.

    7. The Type of Ground Truth Used:

    • Physical Testing: United States Pharmacopeia (USP) standards for Nonabsorbable Surgical Suture (Diameter, Needle Attachment, Tensile Strength).
    • Biocompatibility Testing: FDA Guidance regarding the use of International Standards ISO-10993.

    8. The Sample Size for the Training Set:

    Not applicable. This device does not use machine learning, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set.

    In summary, the SafePath Suturing System is a traditional medical device, and its acceptance criteria and testing focus on its physical properties, sterility, and biocompatibility in accordance with established medical device standards and regulations. The questions related to artificial intelligence and machine learning performance are not relevant to this specific device.