510(k) Data Aggregation

    K Number
    K201741

    Validate with FDA (Live)

    Date Cleared
    2020-11-05

    (133 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gabriel 3-way EnFit valve is a three-way closed system valve with three ports for control of fluid flow (nutrition, medication and water) between devices with EnFit connectors and catheter tip syringe. It can be used in pediatric, adult and elderly patients, for up to 7 days.

    Device Description

    The Gabriel 3-way EnFit Valve is a three-way stopcock valve that is intended to control fluid flow between enteral devices (feeding tubes and feeding lines) with EnFit compatible connectors in a closed system fashion that reduces the need for frequent connection of the EnFit ports. It has one male EnFit port, one female EnFit port and one catheter tip port. The catheter tip port can be used for flushing and administration of medication using a catheter tip syringe. The male EnFit port is provided with a tethered cap and the catheter tip port is provided with a tethered cap. The ability to use a catheter tip syringe for flushing and administration of medicine in health care facilities that do not carry EnFit syringes, may facilitate adoption of feeding tubes and feeding lines with EnFit connectors.

    AI/ML Overview

    The provided text describes a medical device, the "Gabriel 3-Way EnFit Valve," and its 510(k) submission to the FDA. However, this document does not contain information about an AI/ML-enabled device or its acceptance criteria and study results for such a device.

    The document primarily focuses on demonstrating substantial equivalence of a physical medical device (a 3-way valve) to a predicate device based on design, materials, intended use, and bench testing for mechanical performance and biocompatibility.

    Therefore, I cannot answer the questions related to AI/ML acceptance criteria, performance studies (like MRMC or standalone performance), ground truth establishment, sample sizes for training/test sets, or expert qualifications/adjudication. These concepts are not discussed in the provided text.

    The closest information provided is the summary of bench testing for the physical valve. Below is a table of the acceptance criteria (stated as "PASS" for each test) and the reported device performance from the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied by "PASS")Reported Device Performance
    Male EnFit Measurement Verification TestMeets specified measurements for male EnFit connectorsPASS
    Female EnFit Measurements Verification TestMeets specified measurements for female EnFit connectorsPASS
    Catheter Tip Port Measurement Verification TestMeets specified measurements for catheter tip portsPASS
    Tensile TestWithstands specified tensile forcesPASS
    Cytotoxicity testNon-cytotoxicPASS
    Sensitization TestNon-sensitizingPASS
    Irritation TestNon-irritantPASS
    Leakage by Pressure Decay TestNo leakage under pressure decay conditionsPASS
    Positive Pressure Liquid Leakage TestNo leakage under positive pressure liquid conditionsPASS
    Stress Cracking TestResistant to stress crackingPASS
    Resistance to Separation from Axial Load TestResists separation from axial loadPASS
    Resistance to Separation from Unscrewing TestResists separation from unscrewingPASS
    Thread Overriding Resistance TestResistant to thread overridingPASS
    Disconnection by Unscrewing TestResists disconnection by unscrewingPASS
    Flexure modulusAbove 700 MpaPASS
    Feeding Formula Flow Rate TestAchieves specified feeding formula flow ratePASS

    The following information is NOT available in the provided text, as it pertains to AI/ML device studies, which this document does not describe:

    1. Sample sizes used for the test set and the data provenance: Not applicable, no AI/ML test set described.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, no ground truth for image/data interpretation described.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI versus without AI assistance: Not applicable.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191784

    Validate with FDA (Live)

    Date Cleared
    2019-11-15

    (135 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gabriel® Feeding Tube with Balloon functions as a conduit to facilitate enteral feeding, and may be used in adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with functioning gut who require short to moderate term feeding support, such as post-trauma patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exist, or may result, secondary to an underlying disease or condition.

    Device Description

    The Gabriel Feeding Tube with Balloon serves as conduit through which enteral feeding solutions are directly infused into patients small bowel. The tube has two distal end openings, one distal end balloon, one proximal end pilot balloon that indicates the status of the inflation of the distal end balloon and a proximal end EnFit connector that replaced our Y shaped catheter tip connector of the predicate device. The tube shaft is lined by a monofilament spiral wire to prevent occlusion from kinking. A braided 7 strands stainless steel stiffening central stylet prevents coiling in the mouth during insertion and provides column strength to facilitate placement into the stomach.The tube patient contacting material is made from DEHP-free PVC. Size 12 Fr tube is 130 cm long.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for a medical device called the "Gabriel® Feeding Tube with Balloon, and EnFit Connector." This document focuses on demonstrating substantial equivalence to a predicate device (K160787) rather than providing detailed acceptance criteria and study results for a novel AI/software-as-a-medical-device (SaMD) product.

    Therefore, the information required to answer the specific questions about acceptance criteria, study design, sample sizes, expert involvement, and ground truth for an AI device is not present in the provided text. The document describes a traditional medical device (a feeding tube) and its physical and functional characteristics, along with non-clinical tests for its performance and safety compared to its predicate.

    Here's what can be extracted from the document regarding the device and its testing, but it does not fulfill the prompt's request for AI device study details:


    Device Description and Purpose:
    The Gabriel® Feeding Tube with Balloon, and EnFit Connector is a nasoenteral feeding tube designed to facilitate enteral feeding in adult or elderly patients who cannot orally consume an adequate diet. It's intended for bedside insertion and use in patients requiring short to moderate-term feeding support. Key features include a distal end balloon for peristaltic advancement and airway misplacement detection, a flexible shaft with a spiral wire to prevent kinking, and a central stainless steel stylet for insertion. The primary changes from the predicate device are the replacement of a Y-shaped catheter tip connector with a rigid male EnFit connector to mitigate misconnection risk, a change from MR unsafe to MR conditional, and the addition of certain items to a convenience kit.

    Acceptance Criteria and Reported Device Performance (Based on Non-Clinical Testing for Substantial Equivalence):

    As this is a physical medical device and not an AI/SaMD, the "acceptance criteria" are implied by successful completion of mechanical, functional, and safety tests to demonstrate equivalence to a predicate device. The reported performance is that the device "passed" these tests.

    Acceptance Criteria Category/TestReported Device Performance
    Functional Equivalence
    Feeding formula flow rate test (model with EnFit connector vs. predicate with catheter tip connector)Pass
    Mechanical Integrity/Safety
    Tube shaft to male EnFit connector tensile testPass
    Stylet to stylet hub tensile testPass
    Leakage by pressure decay test (EnFit connector)Pass
    Positive pressure liquid leakage test (EnFit connector)Pass
    Stress cracking test (EnFit connector)Pass
    Resistance to separation from Axial load test (EnFit connector)Pass
    Resistance to separation from unscrewing test (EnFit connector)Pass
    Thread overriding resistance test (EnFit connector)Pass
    Dimensional Conformance
    Dimensional verification of EnFit connectorPass
    Material Properties
    EnFit connector material elasticity modulus above 700 MPaPass
    Shelf Life
    48 Months shelf life test (on device with catheter tip connector, applied to new device)Pass (shelf life increased to 48 months)
    Biocompatibility
    Cytotoxicity, sensitization, acute and sub-acute toxicity tests of materialsConfirmed biocompatible
    MR Safety
    MR safety testMR Conditional

    Information NOT available in the provided text (as it pertains to AI/SaMD studies):

    The prompt specifically asks for details related to AI/SaMD acceptance criteria and studies which are not relevant to this 510(k) submission for a non-AI medical device. Therefore, the following points cannot be addressed from the given document:

    1. Sample size used for the test set and the data provenance: Not applicable to this type of device submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as this is not an AI device relying on interpreted data.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    Clinical Testing:
    The document explicitly states: "No clinical testing was performed in association with this submission." This is common for 510(k) submissions where substantial equivalence can be demonstrated through non-clinical testing and comparison to an existing predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160787

    Validate with FDA (Live)

    Date Cleared
    2016-08-09

    (140 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gabriel Feeding Tube with balloon functions as a conduit to facilitate enteral feeding, and may be used in adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with functioning gut who require short to moderate term feeding support, such as post-trauma patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exist, or may result, secondary to an underlying disease or condition.

    Device Description

    The Gabriel Feeding Tube with Balloon serves as a conduit through which enteral feeding solutions are directly infused into the patients small bowel. During placement of the tube, a lubricant and or numbing gel is applied to the nostril.

    For the version of the tube with stylet with magnetic tips, an external magnet is used to assist the physician in placing the tube into the small bowel. Like the predicate device, the modified device has a stylet with a reed switch positioned near its distal tip. The reed switch is connected by wires to an external LED/battery pack that lights in response to the presence of the external steering magnet. The reed switch is encased in a lead-free glass tube and metal shield and is attached to the distal end of the stylet. The wires used to connect the distal reed switch to the LED are polyurethane insulated copper and are wrapped around the core of the stylet and contained inside the outer PTFE layer, thus keeping it out of the fluid path. The distal tip of the stylet contains magnets which are attracted to the steering magnet. The inflated feeding tube balloon allows peristalsis to advance the feeding tube distally.

    For the version of the tube with the non-magnetic stylet, the tube is manually inserted by the physician. The stylet, is non patient contacting, made out of seven braided filaments 305 stainless steel wire and is 3 cm shorter than the feeding tube.

    The stylet is removed and tube taped at the nose and placement verified by pH paper and abdominal x-ray. Like the predicate device, the modified Gabriel Feeding Tube with Balloon has a stylet. The inflated feeding tube balloon allows peristalsis to advance the feeding tube distally.

    The external tube is extruded over reinforcing monofilament stainless steel wire that prevents occlusion by kinking. The outer patient contacting layer is made from DEHP-free PVC.

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA regarding a medical device, the "Gabriel Feeding Tube with Balloon." It is not a study proving a device meets acceptance criteria using AI or advanced analytical methods. The document primarily focuses on establishing substantial equivalence to predicate (already approved) devices through non-clinical testing and biocompatibility assessments. Therefore, I cannot extract the information required to populate a table of acceptance criteria and reported "device performance" in the context of an AI/ML or a complex diagnostic device, nor can I describe a study that proves the device meets these criteria in that framework.

    Here's why I cannot provide the requested information from this document:

    • Device Type: The Gabriel Feeding Tube with Balloon is a physical medical device (a tube for enteral feeding), not a diagnostic or AI/ML-based device.
    • Approval Pathway: The 510(k) pathway establishes "substantial equivalence" to existing devices, meaning it's as safe and effective as something already on the market. It doesn't typically involve new, large-scale clinical trials or the type of "acceptance criteria" and "performance metrics" associated with complex algorithms (like sensitivity, specificity, AUC).
    • Testing: The listed "Non-Clinical Testing" focuses on physical properties (pull strength, flexibility, flow rate, leakage, tensile strength), biocompatibility, and comparison to predicate devices, not on diagnostic accuracy, human reader improvement, or algorithm-only performance.
    • No AI/ML Component: There is no mention of any artificial intelligence, machine learning, or automated diagnostic component in this device or its testing.
    • No Clinical Study Performed for this Submission: The document explicitly states: "No clinical testing was performed in association with this submission." This means there's no data from human subjects on efficacy or performance metrics in a real-world clinical setting presented here.

    In summary, this document does not contain the kind of information (AI performance metrics, human reader study data, detailed ground truth establishment, training set details) that your prompt is asking for.

    If you have a document describing an AI/ML-based device or a diagnostic device, I would be happy to help you extract that information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110005

    Validate with FDA (Live)

    Date Cleared
    2011-08-04

    (213 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Syncro-Blue Tube™ functions as a conduit to facilitate enteral feeding, and may be used in the pediatric, adult or elderly patient who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with functioning gut who require short- to moderate-term feeding support, such as posttrauma patients, post-surgical patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exists, or may result, secondary to an underlying disease or condition.

    Device Description

    The Syncro-Blue Tube serves as a conduit through which enteral feeding solutions are directly infused into the patient's small bowel. The modified device will be marketed in one length (23.6" [60cm]) and one French size (8 Fr). During placement of the tube, an external magnet is used to assist the physician in placing the tube into the small bowel. The Syncro-Blue Tube has a stylet with a reed switch positioned near its distal tip. The reed switch is connected by wires to an external LED/battery pack that lights in response to the presence of the external steering magnet. The reed switch is located at the distal end of the stylet. The distal tip of the stylet contain magnets which are attracted to the steering magnet. When the tube is positioned, the stylet is removed from the device, making the Syncro-Blue Tube MRI Safe.

    AI/ML Overview

    The Syncro-Blue Tube, The Magnetically Guided Enteral Feed Tube, as described in the provided document, underwent several performance tests to demonstrate substantial equivalence to a legally marketed predicate device (Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube cleared under K072787). The document, however, does not provide specific acceptance criteria values or detailed performance results against those criteria. It also does not involve an AI algorithm, so many of the requested criteria related to AI/ML device studies are not applicable.

    Here's a breakdown of the information that can be extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document lists the types of performance tests conducted but does not specify the numerical acceptance criteria for each test or the precise quantitative results achieved by the Syncro-Blue Tube. It generally states that "The results of the laboratory tests demonstrate that the device is as safe and effective as the legally marketed predicate device."

    Test CategoryAcceptance Criteria (Not explicitly stated in the document)Reported Device Performance (Summary from document)
    TensileNot specifiedDemonstrated substantial equivalence.
    Aspiration, FlowNot specifiedDemonstrated substantial equivalence.
    Magnet Capture / Magnetic GuidanceNot specifiedDemonstrated substantial equivalence.
    Flexibility / Column Strength / PushabilityNot specifiedDemonstrated substantial equivalence.
    BiocompatibilityNot specifiedDemonstrated substantial equivalence.
    Simulated UseNot specifiedDemonstrated substantial equivalence.
    RadiopacityNot specifiedDemonstrated substantial equivalence.
    LED Turn-on DistanceNot specifiedDemonstrated substantial equivalence.
    Connector Compatibility TestingNot specifiedDemonstrated substantial equivalence.

    Study Information:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the sample sizes used for each of the performance tests.
      • The data provenance is not explicitly mentioned, but the tests were "laboratory tests," implying they were conducted in a controlled environment, likely in the US or a similar regulatory setting. It is a retrospective summary of bench testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as the device is a physical medical device (feeding tube) and the performance tests are engineering and biological evaluations, not diagnostic interpretations requiring expert ground truth in the context of image or data analysis.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This question is not applicable for the type of device and tests described. Adjudication is typically for resolving discrepancies in expert interpretations, which is not relevant here.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This question is not applicable as the device is a physical medical device and does not involve AI or human readers for interpretation.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable as the device does not involve an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for these performance tests would be defined by established engineering and biological standards and specifications for tensile strength, flow rates, magnetic responsiveness, flexibility, biocompatibility, radiopacity, etc. The document implies compliance with these standards by stating substantial equivalence to a predicate device.
    7. The sample size for the training set:

      • This question is not applicable as the device does not involve a training set for an AI/ML algorithm.
    8. How the ground truth for the training set was established:

      • This question is not applicable as the device does not involve a training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072787

    Validate with FDA (Live)

    Date Cleared
    2007-10-31

    (30 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube is intended for direct placement into the small bowel. The tube functions as a conduit to facilitate enteral feeding, and may be used in pediatric, adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with a functioning gut who require short- to moderate-term feeding support, such as post-trauma patients, post-surgical patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom mainutrition exists, or may result, secondary to an underlying disease or condition.

    Device Description

    The Gabriel Blue Tube serves as a conduit through which enteral feeding solutions are directly infused into the patient's small bowel. The device will be marketed in one length (50" [127cm]) and one French size (11 Fr). The feeding tube contains a stylet which has magnets attached to its tip. An external magnet is used to capture the tip and guide the device to the duodenum. A reed switch and light indicate when the magnet has been captured and can be guided to its final location.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube based on the provided document:

    This 510(k) submission is a "Special 510(k)", which is used for modifications to a legally marketed device where the modifications do not affect the intended use or fundamental scientific technology of the device. Consequently, the provided information focuses on performance tests confirming the substantial equivalence of the modified device to its predicate, rather than a de novo clinical study with extensive details about ground truth establishment, reader studies, or training sets typical of entirely new AI/ML devices.

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (Tests Conducted)
    Mechanical IntegrityTensile
    Flow DynamicsFlow
    Physical CharacteristicsFlexibility
    BiocompatibilityBiocompatibility

    Conclusion: The results of the laboratory tests demonstrated that the device is as safe and effective as the legally marketed predicate devices. This implies the new device met performance benchmarks inherent to the predicate device's established safety and effectiveness.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: The document does not explicitly state the sample sizes for each performance test (Tensile, Flow, Flexibility, Biocompatibility). It refers to "Tests were performed," indicating internal laboratory testing.
      • Data Provenance: The tests were "laboratory tests," implying internal testing conducted by the manufacturer, Syncro Medical Innovations, Inc. No information is provided regarding the country of origin for the data or whether it was retrospective or prospective in terms of patient data, as these tests are primarily materials and functional performance evaluations rather than patient-based clinical trials.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not applicable in the context of this 510(k) submission. "Ground truth" in this scenario would relate to the established engineering specifications and benchmarks for the physical and biological properties of the feeding tube. These are typically defined by industry standards and regulatory requirements, not by expert consensus on clinical images or diagnoses.
    3. Adjudication Method for the Test Set:

      • Not applicable. The "adjudication method" is typically relevant for studies involving human interpretation or clinical endpoints, where discrepancies between evaluators need to be resolved. For material and functional performance tests, results are typically quantitative and compared against predefined specifications.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This type of study is not relevant for a device like a magnetically guided enteral feeding tube, which is a physical medical device and not an AI/ML diagnostic tool or software.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a physical medical device, not an algorithm or AI system.
    6. The type of ground truth used:

      • For this device, the "ground truth" for the performance tests would be established engineering specifications, material standards, and regulatory requirements for tensile strength, flow rate, flexibility, and biocompatibility. These are objective measurements compared against predefined limits or against the performance of the legally marketed predicate device.
    7. The sample size for the training set:

      • Not applicable. There is no training set mentioned or implied as this is a physical medical device being evaluated for substantial equivalence based on its physical properties and performance, not an AI/ML model.
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set exists for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K021991

    Validate with FDA (Live)

    Date Cleared
    2002-07-18

    (30 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MagnaFlow® Magnetically Guided Enteral Feeding Tube functions as a conduit to facilitate enteral feeding, and may be used in pediatric, adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with a functioning gut who require short- to moderate-term feeding support, such as post-trauma patients, post-surgical patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exists, or may result, secondary to an underlying disease or condition.

    Device Description

    The MagnaFlow® Magnetically Guided Enteral Feeding Tube system is intended for direct placement in the small bowel. The catheter functions as a conduit for enteral feeding for patients who cannot consume an adequate diet orally. The catheter has a magnet embedded in the distal tip, which through the use of an external magnet aids in catheter placement. When the external magnet is in close proximity to the catheter tip, a reed switch (located proximal to tube tip magnet) closes, causing an LED light (in manifold) to illuminate. That light indicates when the external magnet has captured the tube distal portion. The external magnet can steer the catheter through the stomach pyloric sphincter into the duodenum.

    AI/ML Overview

    The provided information pertains to a 510(k) premarket notification for a medical device called the MagnaFlow® Magnetically Guided Enteral Feeding Tube. This notification primarily focuses on establishing substantial equivalence to a predicate device already on the market (Flexiflo Magnetically Guided Enteral Feeding Tube, K972437).

    Based on the provided text, there is no specific study described that proves the device meets acceptance criteria in the way typically expected for a diagnostic or AI-driven device. The document is a regulatory submission for a physical medical device (an enteral feeding tube) and focuses on describing its design, intended use, and substantial equivalence to a previously approved device. It does not detail clinical performance metrics, accuracy, sensitivity, or specificity as would be relevant for a diagnostic or AI device.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" are not present in this document.

    Here's an analysis based on the information available:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The submission focuses on describing the device and its equivalence to a predicate, not on specific performance metrics against defined acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided. As no such performance study is detailed, there's no mention of a test set, sample size, or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided. This criterion is relevant for studies evaluating diagnostic accuracy, which is not the focus of this 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not provided. This type of study is relevant for AI-assisted diagnostic devices, which the MagnaFlow® is not.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not provided. The MagnaFlow® is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    This information is not provided.

    8. The sample size for the training set:

    This information is not provided.

    9. How the ground truth for the training set was established:

    This information is not provided.

    Summary of what is provided regarding the device's characteristics and equivalence:

    • Device Trade Name: MagnaFlow® Magnetically Guided Enteral Feeding Tube
    • Device Common Name: Enteral Feeding Tube
    • Device Classification Name: Tube, Feeding at CFR 21 876.5980
    • Legally Marketed Predicate Device: K972437: Flexiflo Magnetically Guided Enteral Feeding Tube
    • Description of Device: The MagnaFlow® system is an enteral feeding tube for direct placement in the small bowel. It features a magnet embedded in the distal tip and a reed switch that closes to illuminate an LED light when an external magnet is in close proximity. This external magnet aids in steering the catheter through the stomach pyloric sphincter into the duodenum.
    • Intended Use: Inserted orally or trans-nasally into the small bowel to provide nutrition, fluids, and medications. For pediatric, adult, or elderly patients who cannot consume an adequate diet orally. Indications include post-trauma, post-surgical, burn patients, general trauma, high-risk patients prone to tube misplacement, and malnourished patients.
    • Technological Characteristics: The submission states that the proposed device has the "same fundamental technological characteristics of the predicate devices and similar design, packaging, sterilization and labeling." This is the primary "proof" for substantial equivalence in a 510(k) for this type of device – demonstrating that it is essentially the same as a device already cleared for market, rather than proving performance against novel acceptance criteria.

    In conclusion, the provided text is a 510(k) summary for a physical medical device. It establishes substantial equivalence to a predicate device, which is the regulatory pathway for market clearance in this context. It does not contain the detailed performance study information typically associated with AI or diagnostic device submissions that would address the acceptance criteria as requested in the prompt.