Search Results
Found 2 results
510(k) Data Aggregation
(89 days)
Coating Material is a light-cured, transparent resin sealant for acrylic denture bases that produces a smooth, glossy surface finish on the denture.
Coating Material is a light-cured denture base coating that is composed of two liquids that are mixed prior to use, and then photo-polymerized by a UV light. The coated surface provides smoothness and glossiness to the surface of an acrylic denture base. The coated layer is transparent, so it does not affect the original color of the denture.
This document describes a 510(k) premarket notification for a device called "Coating Material," which is a light-cured, transparent resin sealant for acrylic denture bases. The submission aims to demonstrate substantial equivalence to a predicate device, Heraeus Kulzer, Inc., Palaseal® (K892452).
Here's an analysis of the provided text in relation to acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (from ISO 7405/10993) | Reported Device Performance (Coating Material) |
|---|---|
| Biocompatibility: | |
| Cytotoxicity (ISO 10993-5) | Non-cytotoxic; all test method acceptance criteria met. |
| Skin Sensitization (ISO 10993-10) | No sensitization under experimental conditions. |
| Intracutaneous Reactivity (ISO 10993-10) | No signs of irritation; met requirements; classified as not irritant. |
| Subchronic Toxicity (ISO 10993-11) | No significant systemic toxicity under experimental conditions. |
| Genotoxicity (In Vitro Mammalian Chromosome Aberration Test) (ISO 10993-3) | Non-clastogenic under experimental conditions. |
| Genotoxicity (Bacterial Reverse Mutation (AMES) Test) (ISO 10993-3) | Non-mutagenic (non-genotoxic and non-clastogenic) under experimental conditions. |
| Performance Bench Testing: | |
| Coating applicability (implied from predicate comparison) | Confirmed to conform to required specifications and suitable for intended use. |
| Properties of the coated surface (implied from predicate comparison) | Smooth, high gloss surface with no evidence of cracking or peeling after being light-cured. |
| Risk Management (ISO 14971): | Outcomes considered acceptable; all potential risks mitigated. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not explicitly state the sample sizes for the individual biocompatibility tests or performance bench tests. It refers to "extending testing" and "experimental conditions."
- Data Provenance: The studies are described as "Non-Clinical Performance Data." The sponsor is Sun Medical Co., Ltd. in Japan, suggesting the studies were likely conducted or overseen in Japan, but this is not explicitly stated for each test. The studies appear to be prospective as they were conducted specifically to support this 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document. For non-clinical lab tests like biocompatibility and bench testing, "ground truth" is typically established by adhering to standardized test protocols (e.g., ISO standards) and comparing results against predefined acceptance criteria outlined in those standards. The expertise would lie in the accredited laboratories and personnel performing these specific tests rather than a panel of clinical experts establishing a ground truth in the way it would be for diagnostic imaging.
4. Adjudication Method for the Test Set
This information is not applicable in the context of these non-clinical, objective laboratory tests. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human interpretation of data where consensus among multiple reviewers is needed to establish a "ground truth" for a diagnosis or specific outcome.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device." Therefore, there is no information on how much human readers improve with AI vs. without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This question is not applicable. The "Coating Material" is a physical medical device (a resin sealant for dentures), not a software algorithm or AI-based device. Performance testing was conducted directly on the material itself.
7. The Type of Ground Truth Used
For the biocompatibility tests and performance bench tests, the "ground truth" used was adherence to established international standards and their predefined acceptance criteria. For example, for cytotoxicity, the ground truth for "non-cytotoxic" is defined by the criteria within ISO 10993-5. For physical properties like gloss and cracking, the "ground truth" is based on visual assessment against predefined quality metrics.
- Biocompatibility: Ground truth established by compliance with ISO 7405, ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-10, ISO 10993-11 test methodologies and their respective acceptance limits.
- Performance Bench Testing: Ground truth established by adherence to "required specifications" (which would be internal to the manufacturer, but validated against intended use) and visual/tactile assessment of "smooth, high gloss surface with no evidence of cracking or peeling."
8. The Sample Size for the Training Set
This question is not applicable. As stated previously, this is a physical medical device, not an AI or software device that uses training sets.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable as there is no training set for this device.
Ask a specific question about this device
(152 days)
Acrylic Block is a cured base acrylic resin that is indicated for use to fabricate a denture.
The Acrylic Block is a solid block of an acrylic polymer block that includes small quantities of color pigments. This Acrylic Block is a high polymer material made from quality poly methyl methacrylate. Poly (methyl methacrylate) is commonly used for thermopolymerizable acrylic resin denture bases because of its properties, including color and durability. In vitro tests have also shown its biocompatibility and enhancement of flexural behavior properties.
To fabricate the denture base, the Acrylic Block is milling machine utilizing the CAM data. After a 3D denture image is structured using the 3D CAD software, the artificial teeth are factored out, and a 3D denture base image is obtained. The Acrylic Block is then milled to the shape of the denture base. The artificial teeth are then bonded to the milled denture base using a bonding material and polished.
The provided document is a 510(k) Pre-market Notification for the "Acrylic Block" device, a cured base acrylic resin used to fabricate dentures. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study purely on the device's performance against specific acceptance criteria in a clinical setting in the way an AI/ML device would.
Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set information is largely not applicable in the context of this traditional medical device submission. The submission relies on non-clinical (bench) testing and established standards for materials.
However, I can extract and present the relevant information that is available, framed within the context of a traditional device submission.
Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria for Acrylic Block (K143265)
The "Acrylic Block" device is a cured base acrylic resin intended for fabricating dentures. Its regulatory submission (510(k)) primarily demonstrates substantial equivalence to a legally marketed predicate device (Yamahachi Denture Base Resins - K131036) by showing comparable material properties and biocompatibility, rather than assessing diagnostic performance or clinical effectiveness through human reader studies.
The acceptance criteria for this device are established through compliance with recognized international and national standards for dental materials, specifically focusing on biocompatibility and mechanical properties. The "study" proving the device meets these criteria consists of a series of non-clinical, bench-top tests.
1. Table of Acceptance Criteria and Reported Device Performance
| Category | Acceptance Criteria (Standard Reference) | Reported Device Performance (Acrylic Block) |
|---|---|---|
| Biocompatibility | EN ISO 7405:1997 / ISO 7405:2008: Dentistry; Preclinical evaluation of biocompatibility of medical devices used in dentistry, test methods for dental materials. ISO 10993-1:2009: Biological evaluation of medical devices - Part 1: Evaluation and testing. ISO 10993-5 (Cytotoxicity): Non-cytotoxic. ISO 10993-10 (Skin Sensitization): No sensitization. ISO 10993-10 (Intracutaneous Reactivity): Not irritant. ISO 10993-11 (Subchronic Toxicity): No significant systemic toxicity. ISO 10993-3 (Genotoxicity - Chromosome Aberration): Non-clastogenic. ISO 10993-3 (Genotoxicity - AMES Test): Non-mutagenic (non-genotoxic and non-clastogenic). | - Non-cytotoxic: Met acceptance criteria per ISO 10993-5, JIS T 6001, Yakushokukihatsu No. 0301-20. - No Sensitization: Considered to cause no sensitization per ISO 10993-10, IIST T 6001:2012 Yakushokukihatsu No. 0301-1, and Yakushokukihatsu No. 0301-20. - Not Irritant: Caused no signs of irritation and met requirements per ISO 10993-10, JIS T 6001, and Yakushokukihatsu No. 0301-20. - No Significant Systemic Toxicity: Considered to not cause significant systemic toxicity per ISO 10993-11, JIS T 6001, Yakushokukihatsu No. 0301-1. And Yakushokukihatsu No. 0301-20. - Non-clastogenic: Considered non-clastogenic per ISO 10993-3, JIST T 6001:2012 Yakushokukihatsu No. 0301-1, and Yakushokukihatsu No. 0301-20. - Non-mutagenic: Considered non-mutagenic per ISO 10993-3, JJST T 6001:2012 Yakushokukihatsu No. 0301-1, and Yakushokukihatsu No. 0301-20. |
| Mechanical Properties | EN ISO 1567:1995: Denture Base Polymers. ISO 20795-1: Dentistry -- Base polymers -- Part 1: Denture base polymers. ANSI/ADA Specification No. 12:2002 (Reaffirmed 2008): Denture Base Polymers (including surface characteristics, color, translucency, freedom from porosity, flexural strength, flexural modulus, residual methyl methacrylate monomer, sorption and solubility). | - Met all acceptance criteria for denture base polymers as outlined in ISO 20795-1 and ANSI/ADA Specification No. 12:2002. - Demonstrated substantially similar performance to predicate denture base resins during non-clinical bench testing. |
| Risk Management | ISO 14971: Medical devices - Application of risk management to medical devices. | - Risk analysis conducted. Outcomes considered acceptable, and all potential risks mitigated to the lowest form. |
| Quality Systems | ISO 9001:2007 ISO 13485:2002 | - Confirmed compliance (as listed for both subject and predicate device). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a numerical "sample size" in terms of patients or independent data points originating from specific cases, as would be common for AI/ML or clinical studies. Instead, samples refer to material specimens created and tested according to the requirements of the cited ISO and ANSI/ADA standards. These standards typically define the number of test specimens required for each specific mechanical or biological test (e.g., a certain number of flexural bars, cytotoxic extracts, etc.). The exact number for each test is not detailed in this summary.
- Data Provenance: Not applicable in the traditional sense of patient data. The data originates from laboratory testing (bench testing) performed on material specimens of the Acrylic Block. The location of these testing facilities is not explicitly stated but would typically be in certified labs in Japan (Sun Medical Co. Ltd.'s country of origin) or elsewhere. The testing is prospective in the sense that it was conducted specifically for this submission against established standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts & Qualifications: Not applicable. For material science and biocompatibility testing, "ground truth" is established by adherence to validated standardized test methods (e.g., ISO, ANSI/ADA) and objective measurements by trained laboratory personnel, rather than expert consensus on interpretive tasks. The results are quantitative measurements or assessments of biological reactions.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. Material testing results are typically objective measurements or observations (e.g., cell viability; flexural strength in MPa, absence/presence of irritation) and do not involve human interpretation or adjudication in the way clinical image analysis might.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data, and the goal is to assess if the device improves reader performance. The Acrylic Block is a material used for fabrication, not a diagnostic tool.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: Not applicable. The Acrylic Block is a physical material, not an algorithm or software device. Its "performance" is assessed through its inherent physical, chemical, and biological properties, not a standalone algorithmic output.
7. The Type of Ground Truth Used
- Type of Ground Truth: The "ground truth" for this device's evaluation is derived from objective, quantitative measurements and standardized biological assays as defined by the referenced international and national standards (ISO, ANSI/ADA, JIS). For example:
- Biocompatibility: Absence of cytotoxicity, sensitization, irritation, systemic toxicity, and genotoxicity as determined by specific, validated lab tests (e.g., MEM Elution, Guinea Pig Maximization, Intracutaneous Reactivity, Chromosome Aberration Test, AMES Test).
- Mechanical Properties: Quantitative values for flexural strength, flexural modulus, sorption, solubility, etc., measured against specified thresholds in the standards like ISO 20795-1 and ANSI/ADA 12.
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable. The Acrylic Block is not an AI/ML device that requires a "training set" to learn from data.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth Establishment for Training Set: Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1