510(k) Data Aggregation

    K Number
    K051631

    Validate with FDA (Live)

    Date Cleared
    2005-07-27

    (37 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Excalibur IV Light Therapy System is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and is clinically indicated for: -- Adjunctive use in providing temporary relief of minor chronic neck and shoulder pair of musculoskeletal origin.

    Device Description

    The Excalibur IV Light Therapy System is a hand-held, non-invasive, pain therapy system which utilizes non-heating lamps consisting of four laser diodes in one applicator head. It combines the clinically accepted therapeutic treatment of numerous predicate light therapy systems currently in commercial distribution and 510(k) approved. The system consists of a basic hand-held, battery operated control unit and a permanently attached applicator head. The laser operates at a measured wavelength of 635nm (±1%) and complies with all performance, labeling, and manufacturing standards set forth in 21CFR Part 1040. Stargate International Inc. is a registered laser manufacturer with the Food and Drug Administration. The laser applicator head produces an output power of ≤4.5mw2 measured, per non-convergent beam and is classified as a Class IIIa laser.

    AI/ML Overview

    The provided document describes the Excalibur IV Light Therapy System, a Class II medical device, and its submission for 510(k) clearance by Stargate International, Inc. The core of the submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel clinical efficacy through a direct comparative study with acceptance criteria.

    Therefore, the requested information regarding acceptance criteria, a study proving the device meets them, sample sizes (test, training), number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types cannot be fully provided from the given text because the 510(k) summary for this device does not include such a clinical performance study. The device's clearance is based on its similarity to existing approved devices.

    Here's an attempt to answer the questions based on the available information, noting the limitations:


    Acceptance Criteria and Device Performance:

    The document does not detail specific quantitative acceptance criteria or a dedicated study designed to prove the device meets such criteria in terms of clinical efficacy. Instead, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence to a predicate device in terms of intended use and technological characteristics.

    Acceptance Criteria Category (Implicit from 510(k) Process)Reported Device Performance (as per submission)
    Intended Use EquivalenceThe Excalibur IV Light Therapy System has the "same intended use" as the predicate device (Quantum System, K032816). - Excalibur IV Indication: Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin. - Predicate Device Indication: Assumed to be the same, allowing for pain therapy treatment for the specified clinical presentation.
    Technological Characteristics EquivalenceThe Excalibur IV Light Therapy System has "similar technological characteristics" to the predicate device. - Wavelength: Measured at 635nm (±1%). - Power Output: ≤4.5mW per non-convergent beam. - Technology: Uses Gallium Aluminum Arsenide (GaAlAs) diodes with non-convergent beam output. - Treatment Variables: Conforms to performance specifications of predicate devices in frequency, wavelength, time, and power output. - Classification: Class IIIa laser, in compliance with 21 CFR Part 1040. - Standards Compliance: FCC Standard 47CFR Part 15B, UL® approved electrical components, Class IIIa Laser - 21CFR Part 1040.
    Safety & Performance Standards ComplianceManufactured, performs, is labeled, and tested to comply with FCC Standard 47CFR Part 15B, UL® approved electrical components, and Class IIIa Laser - 21CFR Part 1040. Includes functional performance, electrical safety, and component specification testing during manufacturing.
    No New Regulatory/Safety/Efficacy IssuesThe submission concludes that the device "does not raise any new regulatory, safety, and/or clinical efficacy issues."

    Detailed Study Information (Not applicable for this 510(k) submission):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • N/A. The submission does not describe a clinical performance study with a distinct test set of patient data to assess the device's efficacy. The substantial equivalence argument relies on the inherent characteristics and intended use being similar to an already approved device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • N/A. Since there's no clinical performance study with a test set requiring ground truth establishment for patient outcomes, this information is not relevant to this submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • N/A. No clinical study involving expert adjudication of a test set is described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. This device is a light therapy system, not an AI-assisted diagnostic tool. No MRMC study was performed or is relevant.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • N/A. The device is a physical light therapy system, not an algorithm. The submission explicitly states, "There is no software incorporated into the operation of the Excalibur IV Light Therapy System."
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • N/A. For this 510(k) summary, the "ground truth" for clearance is essentially the established safety and efficacy of the predicate device (Quantum System, K032816) for its specified indications, which the Excalibur IV aims to mirror. No new clinical ground truth from patient outcomes was established or required for this submission.
    7. The sample size for the training set:

      • N/A. This device does not use machine learning or AI that would require a "training set" in the computational sense.
    8. How the ground truth for the training set was established:

      • N/A. As no training set for an algorithm is involved, this question does not apply.

    Summary of Device Substantiation:

    The Excalibur IV Light Therapy System obtained 510(k) clearance by demonstrating "substantial equivalence" to a predicate device (Quantum System, K032816) based on same intended use (adjunctive relief of minor chronic neck and shoulder pain of musculoskeletal origin) and similar technological characteristics (wavelength, power output, diode technology, conformance to clinical parameters). The submission focused on compliance with relevant electrical, laser, and manufacturing standards, as well as functional performance and component specifications, rather than new clinical efficacy trials with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041530

    Validate with FDA (Live)

    Date Cleared
    2004-10-14

    (128 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Excalibur Light Therapy System is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and is clinically indicated for: Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain i of musculoskeletal origin.

    Device Description

    The Excalibur Light Therapy System is a band-held, non-invasive, pain therapy system which utilizes non-heating lamps consisting of two laser diodes in one applicator head. It combines the clinically accepted therapeutic treatment of numerous predicate light therapy systems currently in commercial distribution and 510(k) cleared. The system consists of a basic hand-held, battery operated control unit and a permanently attached applicator head. The laser operates at a measured wavelength of 635mm (±1%) and complies with all performance, labeling, and manufacturing standards set forth in 21CFR Part 1040. Stargate International, Inc. is a registered laser manufacturer with the Food and Drug Administration. The laser applicator head produces an output power of ≤4.5mw2 measured, per non-convergent beam and is classified as a Class IIIa laser.

    AI/ML Overview

    The provided text describes the Excalibur Light Therapy System and its predicate devices, focusing on demonstrating substantial equivalence. However, it does not contain information about formal acceptance criteria or a study designed to specifically prove the device meets such criteria through clinical performance metrics (like sensitivity, specificity, accuracy, or effect sizes).

    Instead, the submission relies on demonstrating substantial equivalence to existing predicate devices (Tuco Erchonia PL3000 and Quantum System) based on:

    • Identical Intended Use: "Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin."
    • Similar Technological Characteristics: Wavelength (635nm), power output (≤4.5mW), and the use of Gallium Aluminum Arsenide (GaAlAs) diodes with non-convergent beams.
    • Compliance with Standards: FCC, UL, and Class IIIa Laser (21CFR Part 1040).
    • Functional Performance, Electrical Safety, and Component Specification Testing: This refers to engineering and manufacturing tests, not clinical performance studies.

    Therefore, many of the requested fields cannot be filled as they pertain to clinical validation studies, which were not performed or reported in this 510(k) summary.

    Here's a breakdown of the available information based on your request:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in terms of clinical performance metrics (e.g., "device must achieve X% pain reduction in Y% of patients"). Instead, the "acceptance criteria" for this 510(k) appear to be demonstrating substantial equivalence to predicate devices based on:

    CharacteristicAcceptance Criterion (Implied by Substantial Equivalence)Reported Device Performance (Excalibur)
    Intended UseSame as predicate devices: adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.The Excalibur Light Therapy System has the identical intended use as the predicate devices: "Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin."
    Technological CharacteristicsSimilar to predicate devices in wavelength, power output, and mechanism (LLLT). Will provide same treatment benefits/regimens as cleared predicate devices.Uses two laser diodes in one applicator head. Operates at a measured wavelength of 635nm (±1%). Produces an output power of ≤4.5mW per non-convergent beam. Uses Gallium Aluminum Arsenide (GaAlAs) diodes with non-convergent beam output to emit visible photonic energy. Variables (frequency, wavelength, time, power output) conform to predicate device parameters. Classified as a Class IIIa laser.
    Safety and StandardsCompliance with relevant electrical, laser, and manufacturing standards.Manufactured, performs, is labeled, and is tested to comply with: FCC Standard - 47CFR Part 15B, All Electrical Components Utilized Are UL® Approved, Class IIIa Laser 21CFR Part 1040. System testing includes functional performance, electrical safety, and component specification, including four-staged manufacturing testing and verification. No software incorporated.
    Clinical EfficacyNo new regulatory, safety, and/or clinical efficacy issues are raised compared to predicate devices."The Excalibur System will provide the same treatment benefits and regimens for clinical presentations already cleared by the Food and Drug Administration for the predicate devices." "The performance parameters and intended use of the Excalibur Light Therapy System are identical to all predicate devices." "The Excalibur Light Therapy System is manufactured and performs as intended and does not raise any new regulatory, safety, and/or clinical efficacy issues." These claims are based on equivalence to predicate substantial devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable. No clinical test set data is provided or referenced for this device (Excalibur). The submission relies on the clinical data supporting the predicate devices.
    • Data Provenance: Not applicable for this device's clinical performance. The 510(k) refers to "clinically accepted therapeutic uses" and "clinical parameters used by the predicate devices," implying the provenance is from studies previously submitted and reviewed for the predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. No ground truth for a clinical test set for this device is established or reported.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a light therapy system, not an AI-assisted diagnostic or therapeutic tool with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical light therapy system. There is no algorithm or AI component. The document explicitly states: "There is no software incorporated into the operation of the Excalibur Light Therapy System."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable for this device. The efficacy of the underlying technology (Low Level Laser Therapy for pain relief) is implicitly accepted based on the prior clearance of the predicate devices. The "ground truth" for the predicate devices' efficacy would have been established through clinical studies, likely using patient-reported outcomes (e.g., pain scales, functional assessments). However, such data is not provided for Excalibur.

    8. The sample size for the training set

    • Not applicable. There is no AI or machine learning component requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable. There is no AI or machine learning component requiring a training set with ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K032816

    Validate with FDA (Live)

    Date Cleared
    2004-03-26

    (199 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantum Light Therapy System is a non-heating lamp, infrared as described under the provisions of 21 CFR §890.5500 and is indicated for: Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin using the QS-2; QS-4; and/or the QS-L applicator heads.

    Device Description

    The Quantum Light Therapy System is a hand-held, non-invasive, pain therapy system which utilizes non-heating lamps consisting of both laser and light emitting diodes(LED) in three optional applicator heads. It combines the clinically accepted therapentic treatment of numerous predicate light therapy callor itsacs: It commercial distribution and 510(k) approved into one complete, compact, and costeffective system. The system consists of a basic hand-held, battery operated control unit and three optional (one-LED; two-laser) and interchangeable applicator heads which can only be used with the Quantum Light Therapy System control unit. The three distinct applicator heads produce wavelengths that range between 609-645mm which falls within the spectrum as defined by 21 CFR §890.5500. The two laser applicator heads produce an output power of <Smw prom-convergent beam and I he two laser apprication neads produced wavelength of 635mm 11% and ontinued of themse are classified as Class IIIa lastis. Bour operate and included on tortly in 21CFR Part 1040. Stargeton tional, Inc. has filed as a registered laser manufacturer with the Food and Drug Administration. The distinct LED applicator head produces a non coherent diffuse light source. This applicator head operates at a measured wavelength of 628mm ±3%.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Quantum Light Therapy System, focusing on acceptance criteria and the study (or lack thereof) proving its performance.

    It's important to note that this document is a 510(k) summary for a medical device. 510(k) clearances typically establish substantial equivalence to a predicate device, rather than requiring extensive clinical trials to prove efficacy against specific acceptance criteria. Therefore, the information provided focuses more on demonstrating equivalence and compliance with standards than on a detailed clinical study demonstrating specific performance metrics.


    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in the typical sense of numerical thresholds for clinical performance (e.g., minimum sensitivity, specificity, accuracy, or reduction in a pain score). Instead, the "acceptance criteria" for a 510(k) device like this are primarily related to substantial equivalence to predicate devices, safety, and compliance with relevant technical standards.

    The "reported device performance" is framed in terms of its intended use, similarity to predicate devices, and adherence to various electromagnetic and safety standards.

    Here's a table summarizing what can be inferred as "acceptance criteria" from the document and the corresponding "performance" as stated:

    Acceptance Criterion (Inferred from 510(k) Process)Reported Device Performance (from text)
    Intended Use Equivalence: Same intended use as predicate devices."The Quantum Light Therapy System has the same intended use... It combines the clinically accepted therapeutic uses of previously FDA 510(k) approved light therapy systems... into one complete, compact, and cost-effective system.""Adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin..." (same as predicate indications).
    Technological Characteristics Equivalence: Similar technological characteristics (wavelengths, power output) to predicate devices."The technological equivalence to the predicate devices is substantiated by the wavelengths and power outputs generated by the individual Quantum System applicator heads."The two laser applicator heads produce an output power of "<5mW" and wavelength of "635nm +/- 1%".The LED applicator head operates at "628nm ±3%".These fall within the spectrum as defined by "21 CFR §890.5500."
    Safety and Functional Performance: Device is safe and performs functionally as expected."The testing of the Quantum Light Therapy System includes functional performance, software, firmware, electrical safety, and component specification.""The Quantum Light Therapy System does not raise any new regulatory, safety, and/or clinical efficacy issues."
    Compliance with Standards: Adherence to relevant national and international standards."The Quantum Light Therapy System is manufactured, performs, is labeled, and is tested to comply with the following standards: FCC Standard 47CFR Part 15B, All Electrical Components Utilized Are UL® Approved, Class IIIa Laser 21CFR Part 1040, IEC60825-1, Amendment 2, IEC 60601-2-22, EMC: IEC 60601-1-2."
    No New Safety/Efficacy Issues: Introduction of the device does not raise new questions about safety or efficacy."The Quantum Light Therapy System ... does not raise any new regulatory, safety, and/or clinical efficacy issues."

    Study Information:

    Based on the provided document, the "study" conducted for this 510(k) submission is primarily focused on bench testing and comparison to predicate device specifications, rather than a clinical trial measuring patient outcomes against specific performance criteria.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable in the context of a clinical test set with patient data for performance evaluation. The "test set" here refers to the device itself and its components undergoing engineering and safety testing.
      • Data Provenance: The data provenance is from internal testing by Stargate International, Inc. and testing by an outside independent laboratory to confirm compliance with various standards. The document does not mention human subject data or a clinical study. It's retrospective in the sense that it's comparing against existing predicate devices and established standards. The country of origin of the data would likely be the USA, where the company is based and the testing was presumably conducted, though not explicitly stated for the independent lab.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This is not a study involving expert assessment of clinical images or patient conditions to establish a ground truth. Ground truth, in this context, would be the established specifications of the device (e.g., wavelength, power output) and the requirements of the compliance standards.
    3. Adjudication method for the test set:

      • Not applicable. There's no clinical data or expert review that would require an adjudication method like 2+1 or 3+1. Adjudication was likely the process of confirming the device properties met the specified engineering and safety requirements.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a MRMC comparative effectiveness study was not done. This device is a therapeutic light system, not an AI-powered diagnostic tool, so such a study would be irrelevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not a standalone algorithm. Its performance is inherent to its physical properties and operation.
    6. The type of ground truth used:

      • The "ground truth" for this 510(k) submission is based on:
        • Technical Specifications: Measured wavelengths and power outputs of the device compared against design specifications and regulatory requirements.
        • Predicate Device Characteristics: Comparison of the Quantum Light Therapy System's features, intended use, technological characteristics (e.g., non-heating lamp, specific wavelengths and power output range) to those of the legally marketed predicate devices.
        • Compliance Standards: Meeting the requirements of various FCC, UL, IEC, and CFR standards, as verified by testing.
    7. The sample size for the training set:

      • Not applicable. There is no mention of a training set as this is not an AI/machine learning device.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set involved.

    In summary: The document describes a 510(k) submission focusing on demonstrating substantial equivalence to predicate devices and compliance with safety and performance standards. It does not present a clinical study with patient data, acceptance criteria for clinical outcomes, or expert-based ground truth establishment in the typical sense of a diagnostic or AI device study. The "study" here is primarily engineering and regulatory compliance testing.