Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K003087
    Manufacturer
    Date Cleared
    2000-11-07

    (41 days)

    Product Code
    Regulation Number
    880.6885
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K983194
    Manufacturer
    Date Cleared
    1999-03-05

    (175 days)

    Product Code
    Regulation Number
    880.6885
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sporicidin® Sterilizing and Disinfecting Solution (SSDS) is intended for sterilization or high level disinfection of medical and surgical instruments that require submersion. SSDS is intended for reprocessing only heat-sensitive medical devices.

    Sporicidin® Sterilizing and Disinfecting Solution (SSDS) should be used under the following contact conditions:

    TimeTemperature
    Sterilization12 hours25 degrees C
    High LevelDisinfection20 minutes25 degrees C

    SSDS can be reused for a maximum of 7 days or until Sporicidin Test Strips indicate that either the glutaraldehyde or phenol level falls below its Minimum Recommended Concentration (MRC). The glutaraldehyde MRC is 0.6%; the phenol MRC is 1.3%.

    Sporicidin Indicator Strips are intended for verifying the MRC of glutaraldehyde and phenol in SSDS during use. Sporicidin Indicator strips should be used before each use of SSDS.

    Device Description

    Sporicidin® Sterilizing and Disinfecting Solution (SSDS) is a liquid chemical sterilant and high level disinfectant. The product contains two solutions: a buffer containing 1.7% phenol/phenate, and an activator containing 25% glutaraldehyde. When the two solutions are mixed, or activated, SSDS contains 0.95 % glutaraldehyde and 1.64% phenol/phenate. The solutions are combined at the user level, resulting in the activated germicide.

    SSDS is a liquid chemical sterilant and a high level disinfectant when used or reused according to the DIRECTIONS FOR USE included in the package insert. SSDS can be reused for a maximum of 7 days or until Sporicidin Test Strips indicate that either the glutaraldehyde or phenol level falls below its Minimum Recommended Concentration (MRC). The MRC for glutaraldehyde is 0.6%, and 1.3% for phenol/phenate.

    Sporicidin® Sterilizing and Disinfecting Solution (SSDS) must be used with Sporicidin Indicator Strips, used to verify that the MRC of active ingredients is present. It is recommended that SSDS be tested with Sporicidin Indicator Strips prior to each use.

    AI/ML Overview

    The provided documentation describes the efficacy testing and acceptance criteria for Sporicidin® Sterilizing and Disinfecting Solution (SSDS).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for SSDS are based on its ability to act as a liquid chemical sterilant and high-level disinfectant. The performance is assessed against various microorganisms under specified conditions. The Minimum Recommended Concentration (MRC) levels for glutaraldehyde (0.6%) and phenol (1.3%) are also key acceptance criteria, along with specific contact times and temperatures.

    Acceptance Criteria (Intended Use Conditions)Target Microorganism TypeReported Device Performance
    Sterilization: 12 hours immersion at 25°C with MRC levels (0.6% glutaraldehyde, 1.3% phenol)Spores (Bacillus subtilis, Clostridium sporogenes)Effective at 12 hours at 25°C (freshly activated and reused solution)
    High-Level Disinfection: 20 minutes immersion at 25°C with MRC levels (0.6% glutaraldehyde, 1.3% phenol)High-Level Disinfection (Mycobacterium bovis, Mycobacterium terrae)Effective within 20 minutes at 25°C (freshly activated and reused solutions)
    Vegetative Organisms (Pseudomonas aeruginosa, Salmonella cholerasuis, Staphylococcus aureus)Effective within 10 minutes at 20°C (freshly activated and reused solutions)
    Fungi (Trichophyton mentagrophytes)Effective within 10 minutes at 20°C (freshly activated and reused solutions)
    Non-Lipid Small Virus (Poliovirus Type II, Coxsackie, Rotavirus)Effective within 3 minutes at 23°C (Poliovirus Type II)Effective within 2 minutes at 23°C (Coxsackie)Effective within 2 minutes at 20°C (Rotavirus) (reused solutions)
    Lipid Medium Virus (Herpes simplex Type I, Herpes simplex Type II, HIV-1, Influenza Type A2)Effective within 3 minutes at 25°C (Herpes simplex Type I & II)Effective within 1 minute at 23°C (HIV-1 (HTLV-IIIRF) strain)Effective within 10 minutes at 20-25°C (Influenza Type A2) (reused solutions)
    Lipid Large Virus (Vaccinia, Cytomegalovirus)Effective within 10 minutes at 20-25°C (Vaccinia)Effective within 2 minutes at 20°C (Cytomegalovirus) (reused solutions)
    Simulated In-Use Testing: Effective in reducing Mycobacterium terrae by 10^6 on endoscopes within 15 minutes at 25°C with active ingredient levels near or below MRC.Mycobacterium terrae (on endoscopes)Solutions with active ingredient levels near or below MRC were effective in reducing Mycobacterium terrae by 10^6 on endoscopes in 15 minutes at 25°C.
    Residue Testing: No detectable glutaraldehyde, phenol levels < 24 ppm.Glutaraldehyde and Phenol ResiduesNo detectable glutaraldehyde, phenol < 24 ppm (0.0024%) after 12-hour soak.
    Material Compatibility: Compatible with a wide range of materials and endoscopes.Various Materials (Plastics, Synthetic Rubber, Glass, Alumite, Stainless Steel, Adhesive, Sealing material)Compatible with the materials and devices listed in the document.
    Shelf Life/Stability: 2 yearsRetention of EfficacySupported by real-time stability studies and accelerated aging.
    Test Strips: Verify MRC of glutaraldehyde (0.6%) and phenol (1.3%)Glutaraldehyde and Phenol MRCSporicidin Indicator Strips should be used prior to each use to verify MRC.

    2. Sample size used for the test set and the data provenance

    • Microbiological Testing: The document does not specify a precise sample size in terms of number of organisms tested per run, but it states that testing was conducted under "worst case conditions with solution batches that were at the end of, or beyond, their two year expiration dates."
    • Simulated In-Use Testing (Endoscopes):
      • Sample Size: Mentions "Flexible fiber endoscopes made of a variety of material types were used." No specific number provided.
      • Data Provenance: Retrospective (implied, as the tests were conducted for market submission). The country of origin is not specified, but the submission is to the FDA in the USA.
    • Clinical (In-Use) Testing (Endoscopes):
      • Sample Size: Refers to "Lower GI endoscopes used in patient care." No specific number provided.
      • Data Provenance: Prospective (implied, "under actual use conditions" of patient care). The country of origin is not specified.
    • Residue Testing:
      • Sample Size: "devices were soaked". No specific number provided.
      • Data Provenance: Retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document describes laboratory efficacy testing rather than studies requiring expert adjudication for ground truth (e.g., image interpretation). Therefore, no information is provided on the number or qualifications of experts for establishing ground truth as it would be understood in a diagnostic accuracy study. The "ground truth" for the microbiological efficacy studies would be the established methods for determining microbial kill and reduction.

    4. Adjudication method for the test set

    Not applicable, as the studies are primarily laboratory efficacy tests and chemical residue analysis, not requiring expert adjudication in the manner of diagnostic accuracy studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for a chemical sterilant/disinfectant, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a chemical product, not an algorithm. The product itself, when used according to instructions, is "standalone."

    7. The type of ground truth used

    The ground truth used for efficacy testing is based on microbiological assay results (e.g., reduction in microbial count, complete kill) following standardized testing procedures. For residue testing, the ground truth is established by chemical analysis (e.g., HPLC, spectroscopy) to detect and quantify residues.

    8. The sample size for the training set

    Not applicable. This is a chemical sterilant/disinfectant, not a machine learning model. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1