Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K151069

    Validate with FDA (Live)

    Date Cleared
    2016-02-05

    (290 days)

    Product Code
    Regulation Number
    868.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle is intended for use in injecting a single dose of local anesthetic or analgesic into a patient for regional anesthesia or pain therapy with the echogenic reflective pattern providing visualization of the needle tip using Ultrasound Imaging. The needles are intended to mate with the male connector of a Luer Lock or Luer Slip Syringe or an extension set.

    Device Description

    The SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle consists of a stainless steel needle and a clear or colored (translucent) standard female Luer lock square or round Hub locking connector for rapid (flashback) visualization. The stainless steel needles are available with and without an echogenic feature (i.e. mechanical dimpling treatment to the needle point surface), and with or without lubrication. The needles are available in a range of wall thicknesses, gauges and lengths to match the end-user need (16 to 30 gauges).

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (SPECTRA Disposable Single Lumen Echogenic Injection Anesthesia Conduction Needle). It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria and device performance in the context of AI/ML or diagnostic accuracy.

    Therefore, many of the requested categories for AI/ML study details, such as sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or specific performance metrics (like sensitivity, specificity, AUC), are not applicable or available in this document. This document describes a traditional medical device, not an AI/ML system.

    However, I can extract information related to the performance data presented to support the substantial equivalence claim.

    Here's the information derived from the document:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" with numerical thresholds for performance, but rather lists the mechanical and biocompatibility tests performed and concludes that the device "exhibited comparable mechanical and functional characteristics to the predicate device in addition to being biocompatible acceptable."

    Test CategorySpecific Tests PerformedReported Device Performance
    Mechanical Testing- Bond strength of hub and cannula- ISO 594-1 (Luer Fittings)- ISO 7864:11 (Sterile hypodermic needles)- Luer compatibility- Hub and Needle Strength- Penetration Force"Exhibited comparable mechanical and functional characteristics to the predicate device." The device's performance in these tests was deemed acceptable and comparable to the predicate device, supporting substantial equivalence.
    Biocompatibility Testing- ISO 10993-5: Cytotoxicity Study Using the ISO Elution Method- ISO 10993-4: In Vitro Hemolysis Study (Modified ASTM-Extraction Method)- ISO 10993-11: USP and ISO Systemic Toxicity Study - Extracted- ISO 10993-10: ISO Intracutaneous Study - Extract- ISO 10993-10: ISO Maximization Sensitization Study - Extract"Biocompatible acceptable." The product contact materials were "well characterized chemically and physically," and test results confirmed "compliance with ISO 10993 requirements."
    Sterility"Supplied Packaged Sterile"The device is supplied as sterile. (Sterilization method: EO (Ethylene Oxide))

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated for each test. The document refers to "the following performance data was provided" without specifying the number of units tested for each mechanical or biocompatibility assessment.
    • Data Provenance: Not specified. This is a submission to the FDA for a new device, so the testing would have been conducted by the manufacturer or their contracted labs. No information about country of origin of data (e.g., patient data) is relevant here as this is a physical device. The studies are prospective in the sense that they were conducted for this specific submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This is not applicable for this type of medical device submission. Ground truth in the context of diagnostic accuracy or AI/ML is not relevant here. The "truth" for mechanical and biocompatibility tests is based on the results of validated laboratory methods and adherence to international standards (ISO, ASTM).

    4. Adjudication method for the test set

    • This is not applicable. No human interpretation or adjudication common in AI/ML validation studies is mentioned. The assessment is based on objective laboratory measurements against established standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable. The device is a physical anesthesia needle, not an AI system. No MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This is not applicable. The device is a physical anesthesia needle, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the performance data in this context refers to established international standards and validated laboratory test methods (e.g., ISO 594-1, ISO 7864:11, ISO 10993 series). The device's characteristics were measured against these standards to demonstrate compliance and comparability to the predicate.

    8. The sample size for the training set

    • This is not applicable. There is no "training set" as this is a physical medical device, not an AI/ML model.

    9. How the ground truth for the training set was established

    • This is not applicable. There is no "training set" or corresponding ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103740

    Validate with FDA (Live)

    Date Cleared
    2011-07-29

    (219 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated only for use in flushing compatible intravenous tubing systems and in dwelling intravascular access devices. Not to be used for any other purposes.

    Device Description

    Spectra Medical's Device, Sodium Chloride Injection, 0.9%, USP 5 mL, 10 mL Ampule.

    AI/ML Overview

    This document is a 510(k) summary for a medical device: Spectra Medical's Sodium Chloride Injection, 0.9%, USP 5 mL, 10 mL Ampule.

    Based on the provided text, there is no information about acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of an AI/ML device.

    This document describes a pharmaceutical product (sterile saline solution for flushing), not an AI/ML powered medical device. The 510(k) summary focuses on demonstrating "substantial equivalence" to a predicate device, which is a regulatory pathway for low-risk medical devices in the US. This pathway typically involves comparing the new device's indications, technological characteristics, and performance data to a legally marketed predicate, rather than conducting new performance studies with specific acceptance criteria that would be relevant for an AI/ML system.

    Therefore, I cannot populate the requested table and answer the study-related questions, as they are not applicable to the content of this 510(k) summary. The document explicitly states:

    • "There has been no change to the performance characteristics of the device system." (Section 6)
    • "There has been no change to the fundamental scientific technology." (Section 7)

    These statements indicate that no new performance studies (of the kind expected for an AI/ML device) were conducted or are being presented in this submission. The device is claimed to be "identical to the predicate device" with only a difference in ampule size.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082689

    Validate with FDA (Live)

    Date Cleared
    2009-03-20

    (186 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated only for use in flushing compatible intravenous tubing systems and in dwelling intravascular access devices. Not to be used for any other purposes.

    Device Description

    Spectra Medical's, Sodium Chloride Injection, 0.9%, USP, 10 mL Ampule

    AI/ML Overview

    This document is a 510(k) summary for a medical device called "Sodium Chloride Injection, 0.9%, USP, 10 mL Ampule" by Spectra Medical. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting new performance data from a specific study. Therefore, most of the requested information about acceptance criteria and a study proving their fulfillment is not present in this document.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    This document does not specify any acceptance criteria or report new performance data for Spectra Medical's device. The core argument for clearance is "substantial equivalence" to a predicate device (K023740 - Syrex Pre-filled Syringe with 0.9% Sodium Chloride).

    The document explicitly states:

    • "There has been no change to the performance characteristics of the device system."
    • "There has been no change to the fundamental scientific technology."
    • "Spectra Medical's Chloride Injection, 0.9%, USP, 10 mL Ampule is identical to the predicate device. The difference is that the Spectra medical product is in a 10 mL ampule whereas the predicate device is in a syringe."

    Therefore, no acceptance criteria or new reported performance are provided for this specific submission as it relies on the predicate device's established performance.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    No test set or new data is presented in this 510(k) summary. The submission relies on the existing data and performance characteristics of the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable. No new test set or ground truth establishment is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. No new test set or adjudication process is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a Sodium Chloride Injection, not an AI-assisted diagnostic or therapeutic tool. Therefore, an MRMC study related to human readers and AI is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a Sodium Chloride Injection, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. No new ground truth was established for this submission. The "truth" for this submission is that its formulation and intended use are identical to a legally marketed predicate device, with only a change in packaging (ampule vs. syringe).

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is not an AI/machine learning device.

    In summary:

    This 510(k) submission is for a device that is considered "substantially equivalent" to a previously cleared predicate device. The basis for this equivalence is that the product (0.9% Sodium Chloride Injection) is identical to the predicate device in terms of its composition, performance characteristics, and fundamental scientific technology. The only difference highlighted is the packaging (10 mL ampule vs. syringe). Therefore, no new clinical studies, performance data, or detailed acceptance criteria are presented in this document for the Spectra Medical device itself, as it is deemed to meet the established safety and effectiveness of its predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082580

    Validate with FDA (Live)

    Date Cleared
    2008-12-08

    (94 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spectra Guidewire Introducer Needle is intended to be used by medical professionals as introducers/cannula for percutaneous introduction and placement of guidewires in vascular procedures.

    The Spectra Guidewire Introducer Needle is indicated for percutaneous introduction and placement of guidewires in vascular procedures.

    Device Description

    The Spectra Guidewire Introducer Needle consists of a stainless steel needle and a colored translucent standard female Luer lock hub locking connector for rapid (flashback) visualization. The Spectra Guidewire Introducer Needle is available in a single-wall style with or without a Seldinger shield. The stainless steel needles are available with and without an echogenic feature. The needles are available in a range of wall thicknesses, gauges and lengths to match the end-user need.

    Spectra Guidewire Introducer Needles will be marketed as sterile, non-pyrogenic, and single use devices. Additionally, the device will be marketed as a guidewire introducer needle for use as an accessory in procedure kits. In the case of being used in procedural kits, the product will be shipped bulk non-sterile to the kitting manufacturer. The Spectra Guidewwire Introducer Needle will be incorporated into the procedure kit, packaged and sterilized.

    AI/ML Overview

    The provided text describes the submission of a 510(k) premarket notification for the Spectra Guidewire Introducer Needle, arguing its substantial equivalence to predicate devices rather than claiming novel effectiveness. Therefore, the document does not contain a study that proves the device meets specific performance acceptance criteria in the way a clinical trial or algorithm validation study would. Instead, it focuses on demonstrating that the device is safe and effective for its intended use by meeting established standards and exhibiting similar technological characteristics to legally marketed predicate devices.

    Here's a breakdown based on your request, highlighting what information is available and what is not:


    1. Table of Acceptance Criteria and Reported Device Performance

    The submission mentions that "The Spectra Guidewire Introducer Needle will meet all established acceptance criteria for performance testing" and that "Performance data gathered demonstrated that the Spectra Guidewire Introducer Needle is substantially equivalent to the noted predicate devices." However, the specific quantitative acceptance criteria (e.g., tensile strength, flow rates, sharp-ness specifications) and the precise reported performance values are not detailed in the provided text.

    The acceptance criteria are generally framed around adherence to recognized standards for medical devices and comparison to predicate devices, rather than novel performance benchmarks.

    Acceptance Criteria CategoryReported Device Performance
    Sterilization RequirementsMeets ISO 11135:2007 (Ethylene Oxide Sterilization).
    Biocompatibility RequirementsMeets ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. (Implicitly, the device passed these tests and is considered biocompatible).
    Technological CharacteristicsSubstantially equivalent design, materials, and method of use to predicate devices (Teleflex Medical Introducer Needle (K851140) and Cook Percutaneous Entry Needle (Pre-amendment device, Reference Medical Device Listing No. 187040)). This implies performance in aspects like needle stiffness, tip geometry, hub connection, flashback visualization, etc., are comparable to the predicates.
    Safety and EffectivenessDemonstrated safe and effective for its intended use, based on FDA's decision tree for substantial equivalence. This is a conclusion based on the combined assessment of materials, design, manufacturing, and comparison to predicates.
    Intended Use & IndicationsShares the same intended use and indications for use as predicate devices: "intended to be used by medical professionals as introducers/cannula for percutaneous introduction and placement of guidewires in vascular procedures."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards and protocols." However, the sample size used for the test set (i.e., how many devices were tested for performance, how many units of materials, etc.) is not specified.

    Regarding data provenance, the testing would have been conducted by Spectra Medical Devices, Inc. or contractors they employed, likely in the US (given the company's address in Wilmington, Massachusetts). The data is retrospective in the sense that it was generated for the purpose of this 510(k) submission, not from a prospective clinical trial on human subjects. The type of data is primarily in vitro and bench testing data to demonstrate compliance with standards and equivalence.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable in this context. The "ground truth" for a medical device like an introducer needle in a 510(k) submission is established by engineering specifications, material standards, and comparisons to legally marketed predicate devices, not by expert consensus on clinical findings. Performance is measured against physical and chemical parameters, not against interpretations of medical images or patient outcomes by experts.


    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human expert assessment of data (e.g., radiology reads). In the context of a 510(k) for an introducer needle, "adjudication" is not a standard part of the device's technical performance testing. Device performance is determined by meeting pre-defined physical and material specifications.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical data, often with AI assistance. The Spectra Guidewire Introducer Needle is a physical surgical instrument, not a diagnostic imaging device or an AI algorithm intended to assist human interpretation.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, this is not applicable. The Spectra Guidewire Introducer Needle is a physical medical device, not an algorithm or AI software. Therefore, there is no "standalone algorithm" performance to report.


    7. The Type of Ground Truth Used

    For this device, the "ground truth" for performance assessment is primarily based on:

    • Established engineering specifications and material standards: These define what constitutes acceptable strength, biocompatibility, sterility, dimensions, etc. (e.g., ISO 11135, ISO 10993).
    • Performance characteristics of predicate devices: The "ground truth" for equivalency is that the new device performs similarly to legally marketed devices that have already been deemed safe and effective. This is assessed via various physical and chemical bench tests (e.g., material compatibility, physical dimensions, functional tests implicitly covered by "design, materials, and method of use").

    There is no "pathology" or "outcomes data" ground truth in the context of this 510(k) submission to demonstrate substantial equivalence of a guidewire introducer needle. Clinical outcomes would typically be part of a PMA (Premarket Approval) submission for a novel, high-risk device.


    8. The Sample Size for the Training Set

    This is not applicable. A "training set" refers to data used to train an algorithm or AI model. As established, this device is a physical medical instrument, not a software or AI product.


    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082674

    Validate with FDA (Live)

    Date Cleared
    2008-11-28

    (74 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spectra, "GLASS LOR SYRINGE" [Loss of Resistance Syringe] is intended for use, in conjunction with an epidural needle, to verify the needle tip placement in the cpidural space by the Loss of Resistance technique as explained in standard medical textbooks. These syringes are not intended for injection or aspiration.

    Device Description

    Spectra, GLASS LOR SYRINGE

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called "Spectra, GLASS LOR SYRINGE". This type of letter generally does not include detailed study results, acceptance criteria, or performance data beyond stating that the device is substantially equivalent to a predicate. The letter confirms the device can be marketed based on its equivalence.

    Therefore, many of the requested items (e.g., acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) will not be available in the provided text.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided. This document is an FDA clearance letter based on substantial equivalence, not a detailed study report. It does not contain specific acceptance criteria or performance metrics beyond the general statement of substantial equivalence to a predicate device.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. The 510(k) letter does not include details about specific test set sample sizes or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Cannot be provided. Information on expert involvement for ground truth establishment for a test set is not present in this regulatory clearance letter.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Cannot be provided. Details on adjudication methods are not included in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Cannot be provided. This device is a "GLASS LOR SYRINGE" (Loss of Resistance Syringe), a physical medical device. It is not an AI algorithm, and therefore, an MRMC comparative effectiveness study involving AI assistance would not be applicable or performed for this product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Cannot be provided. As a physical syringe, the concept of "standalone algorithm performance" is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Cannot be provided. This information is not part of an FDA clearance letter for a physical device. For a device like a syringe, performance is typically assessed through physical and mechanical testing against recognized standards or direct comparison to a predicate, rather than through complex ground truth establishment methods relevant to diagnostic algorithms.

    8. The sample size for the training set

    • Cannot be provided. As this is a physical device and not an AI algorithm, there would be no "training set" in the computational sense.

    9. How the ground truth for the training set was established

    • Cannot be provided. This is not applicable to a physical device.

    Summary based on the provided text:

    The provided document is an FDA 510(k) clearance letter for the "Spectra, GLASS LOR SYRINGE." This letter indicates that the device has been deemed substantially equivalent to legally marketed predicate devices. The basis for this determination is typically a comparison of the device's technological characteristics and performance to those of the predicate.

    While the letter confirms the device meets the regulatory requirements for market entry, it does not contain the detailed study data, acceptance criteria, or performance reports that would include information on sample sizes, expert involvement, ground truth establishment, or specific metrics like sensitivity, specificity, or effect sizes. Such detailed information would typically be found in the manufacturer's 510(k) submission itself, which is not provided here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081524

    Validate with FDA (Live)

    Date Cleared
    2008-07-29

    (57 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SPECTRA-LOR, Loss of Resistance Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.

    Device Description

    The construction of the SPECTRA-LOR, Loss of Resistance Syringe is similar to that of a standard piston syringe; it has a barrel, a plunger and a bung. The barrel has graduations which correspond to a capacity of 10 ml and is available in a Luer Lock or Luer Slip tip.

    AI/ML Overview

    The provided text is a 510(k) summary for the SPECTRA-LOR, Loss of Resistance Syringe. It describes the device, its intended use, and its equivalence to a predicate device. However, it does not contain any information about acceptance criteria, device performance metrics, or any studies conducted to prove the device meets specific acceptance criteria.

    The document explicitly states: "The SPECTRA-LOR, Loss of Resistance Syringe is equivalent to the predicate device, AVID-NIT Loss of Resistance Syringe in materials of construction, intended use, capacity and bench testing." This indicates that the regulatory clearance was based on substantial equivalence to a predicate device, rather than new studies proving pre-defined acceptance criteria for the SPECTRA-LOR syringe's performance.

    Therefore, I cannot provide the requested information from the text.

    If this were a typical AI device submission, the response would look like this (but based on the actual information provided, such details are absent):

    1. Table of Acceptance Criteria and Reported Device Performance
    (No specific acceptance criteria or performance metrics were provided in the document for the SPECTRA-LOR syringe. Its approval is based on substantial equivalence to a predicate device.)

    2. Sample Size Used for the Test Set and Data Provenance
    (Not applicable, as no dedicated study with a test set was described to prove specific performance metrics for the SPECTRA-LOR syringe.)

    3. Number of Experts Used to Establish Ground Truth and Qualifications
    (Not applicable, as no specific study involving expert-established ground truth was described.)

    4. Adjudication Method for the Test Set
    (Not applicable.)

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
    (Not applicable. This device is a physical syringe, not an AI-assisted diagnostic tool.)

    6. Standalone (Algorithm Only) Performance Study
    (Not applicable. This device is a physical syringe, not an algorithm.)

    7. Type of Ground Truth Used
    (Not applicable, as no ground truth was established for performance metrics in a study for this device.)

    8. Sample Size for the Training Set
    (Not applicable. This device is a physical syringe, not a machine learning model requiring a training set.)

    9. How the Ground Truth for the Training Set Was Established
    (Not applicable.)

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1