510(k) Data Aggregation

    K Number
    K112469

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-03-23

    (210 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonowand Invite System is intended for use as a tool to aid intraoperative ultrasound imaging and image guided surgery during neurosurgery. It is also intended for use as a standard neuronavigation system and as a stand-alone ultrasound scanner.

    The intended use of a localizer is to enable navigation by showing the position of surgical instruments or pointers relative to MR, CT or Ultrasound images

    The General purpose localizers with adapters are intended to be used to attach surgical tools manufactured by other vendors. SonoWand Invite System will display position and orientation of the tools in the medical images on the screen.

    Device Description

    The 510(k) cleared SonoWand Invite System is a neuronavigation system with intraoperative imaging capabilities. An ultrasound scanner is physically integrated with a neuronavigation system in a single rack of equipment.

    SonoWand Invite can be used as a conventional neuronavigation system based on preoperative MR or CT-images, or as a stand-alone ultrasound scanner for real-time 2D imaging or as a combined system where 3D ultrasound data can be transferred to the navigation system for direct navigation.

    SonoWand Invite has basically two functionalities. It supports the surgeon by showing the position of tools or pointers relative to MR, CT or Ultrasound images. It also makes it possible for the surgeon to acquire ultrasound images during operation, which can be compared with other images (such as MR or CT) and can be used for supporting the surgeon by showing the position of tools and pointers. The control is primarily by a graphical touch screen. The secondary control is by a footswitch.

    The new accessories, General purpose localizers with adapters, shall be used to attach surgical tools manufactured by other vendors. SonoWand Invite will then display position and orientation of the tools in the medical images on the screen in the same way as the previous 510(k) cleared Navigator tool.

    The localizers with adapters are designed to fit surgical tools normally used in the operating room.

    There are no changes to the Intended use, function or technological aspects including software of the SonoWand Invite. The additional accessories are being added to increase usability of the SonoWand Invite system.

    AI/ML Overview

    The provided text describes a 510(k) submission for new accessories (general purpose localizers and adapters) for the SonoWand Invite System. This submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing extensive de novo clinical study data.

    Therefore, the information you've requested regarding acceptance criteria, specifics of a study proving those criteria, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, training sets, and ground truth establishment is largely not present in this 510(k) summary.

    The summary states that the new accessories have been tested to comply with standards and requirements, and that "Results of the tests performed show the General purpose localizers and adapters are as safe and effective as the accessories of the predicate device." However, it does not detail the specific performance metrics or the study design with the granularity you've requested.

    Here's what can be extracted and inferred from the provided text, while also explicitly stating what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    No specific numerical acceptance criteria (e.g., accuracy thresholds, precision ranges) are reported in a quantifiable manner, nor are specific performance values provided for the new device or the predicate. The "performance" described is qualitative, stating equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with cleaning and sterilization standardsComplies with all applicable cleaning and sterilization standards
    Software module function in cleared SonoWand Invite SystemSoftware module in cleared SonoWand Invite System supports function and complies
    Requirements of SonoWand Invite System Test (for Localizers)Localizers comply with requirements of SonoWand Invite System Test
    Fit, calibration, and accuracy (for Adapters)Adapters were tested for fit, calibration and accuracy
    Safety and effectiveness equivalence to predicate devices"as safe and effective as the accessories of the predicate device"
    Does not affect intended use of connected deviceConnecting a localizer to the device does not affect the intended use of the connected device

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The document states "The Localizers have been tested..." and "The Adapters were tested...", but does not provide the number of localizers, adapters, or test scenarios.
    • Data Provenance: Not specified. The submitter is from Norway, but the location where the tests were conducted or the origin of any data (if human subjects were involved, which is not indicated) is not mentioned. The testing is internal to the manufacturer's validation activities for demonstrating substantial equivalence.
    • Retrospective/Prospective: Not applicable, as this is a device accessory validation (likely bench testing/engineering validation), not a clinical study involving patients being followed over time.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not provided. The testing described appears to be technical/engineering validation, not a study where expert interpretations are compared to establish ground truth for image analysis or diagnostic performance.

    4. Adjudication Method for the Test Set

    • Not applicable/Not specified. This type of information is typically relevant for clinical studies involving multiple readers and often conflicting interpretations, which is not the case described here.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence of new accessories through technical validation, not on comparing reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • This is not applicable in the context of these accessories. The SonoWand Invite System itself has software, and the document states: "The devices subject to this submission do not contain software module in the cleared SonoWand Invite System that supports the function of the subject devices has been tested to and complies with requirements." This implies the existing system's software was evaluated, but not these passive localizers/adapters as standalone algorithms.

    7. The Type of Ground Truth Used

    • For the tests mentioned (cleaning/sterilization, SonoWand Invite System test, fit/calibration/accuracy), the ground truth would be established by engineering specifications, regulatory standards, and measurement references. For example, calibration accuracy would be measured against known precise standards, and fit would be verified against design specifications. It is not expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    • Not applicable. These accessories are passive mechanical components (localizers and adapters) that enable the tracking of surgical tools. There is no AI model or algorithm being "trained" with data for these specific components. The SonoWand Invite System itself is a neuronavigation system, but the submission is for its accessories.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for these accessories, there is no ground truth establishment for such a set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083597

    Validate with FDA (Live)

    Device Name
    SONOWAND INVITE
    Manufacturer
    Date Cleared
    2009-09-25

    (294 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonowand is intended for use as a tool to aid intraoperative ultrasound imaging and image guided surgery during neurosurgery. The Sonowand is intended for use as a standard neuronavigation system. The Sonowand is intended for use as a stand-alone ultrasound scanner.

    Device Description

    Invite is an integrated neuronavigation system with intraoperative imaging capabilities. The main components of the system are a navigation computer running the Invite software application, an ultrasound scanner and a tracking system in a single rack of equipment. Invite can be used as a conventional neuronavigation system based on preoperative MR or CTimages, or as a stand-alone ultrasound scanner for real-time 2D imaging. It can also be used as a combined system where high-quality 3D ultrasound data can be transferred to the navigation system for direct navigation. Thus, the surgeon can navigate not only on a preoperative 3D map (MR or CT), but also on an intraoperative 3D ultrasound map which is only a few seconds old. The acquisition time is less than one minute, which means that the surgeon can typically update the navigation map 5-10 times during the surgery. The problem with brain shift is therefore practically eliminated. The system is contained on a trolley with lockable wheels. This trolley contains power supply, power restrictor, on/off main switch, network and USB ports, an industrial standard PC (navigation computer), the ultrasound unit (GE Vivid-i), probe interface, wiring and cabling. Attached to the trolley is a steel tube column which supports two pendant arms. One of these arms supports an infrared camera (NDI Polaris Spectra). The other arm supports a touch screen. which is the primary user interface. Invite has two functions. It supports the surgeon by showing the position of tools or pointers relative to MR, CT or Ultrasound images (which will normally show features such as brain tumors). It also makes it possible for the surgeon to acquire ultrasound images during operation which can be compared with other images (such as MR or CT) and can be used for supporting the surgeon by showing the position of tools and pointers. The control is primarily by a graphical touch screen (which can be draped). The secondary control is by a footswitch. Data sets (preoperative MR/CT and intraoperative 3D ultrasound) can be used for planning the surgery as well as to navigate in the brain. As the surgery progresses, the accuracy and value of the preoperative images will normally decrease, while ultrasound images will maintain a more precise representation of the true anatomy if these images are updated regularly. Digital snapshots of the screen can easily be stored during surgery for documentation purposes. It is also possible to review the data sets and to store images after surgery.

    AI/ML Overview

    The provided 510(k) summary for the SonoWand Invite does not contain the detailed study information typically requested for acceptance criteria and device performance.

    This document describes the device, its intended use, and compares it to predicate devices to establish substantial equivalence. It confirms that the system has been "verified and validated according to specification requirements" and "tested to and complies with all applicable EMC/EMI standards," but does not provide any specific quantitative performance metrics, clinical study design, or results related to efficacy (e.g., accuracy, sensitivity, specificity, or improvement in human reader performance).

    Therefore, I cannot fulfill your request for the following information based on the provided text:

    1. A table of acceptance criteria and the reported device performance: No specific acceptance criteria or performance metrics (e.g., accuracy, precision) are mentioned.
    2. Sample size used for the test set and the data provenance: No test set or clinical data is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no test set with ground truth is described.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not mentioned.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not mentioned, and the device is described as "a tool to aid" rather than a standalone diagnostic.
    7. The type of ground truth used: Not applicable.
    8. The sample size for the training set: Not mentioned.
    9. How the ground truth for the training set was established: Not mentioned.

    What the document does state regarding testing:

    • Summary of Testing: "The SonoWand Invite system has been verified and validated according to specification requirements. The system has been also been tested to and complies with all applicable EMC/EMI standards. Applicable testing of the ultrasound components has been performed. Results of the tests performed show the SonoWand Invite is as safe and effective as the predicate devices."

    This broadly states that internal validation and verification processes were completed, and that the device meets safety standards, but it does not provide the specific performance data or study details you are asking for. This is common for 510(k) submissions focusing on substantial equivalence to predicate devices, where extensive new clinical performance data might not be required if the device's technology and intended use are similar to already cleared devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1