Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K213088

    Validate with FDA (Live)

    Device Name
    ZQuiet Advance
    Manufacturer
    Date Cleared
    2022-06-15

    (264 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ZQuiet Advance is intended as an aid in the reduction of snoring for adults at least 18 years old.

    Device Description

    The ZQuiet Advance is an intraoral appliance designed to reduce snoring by advancing the lower jaw. The principal effect of mandibular advancement devices is the protrusion of the lower jaw, thereby widening the upper airway to decrease air turbulence, a causative factor in snoring. This device combines several essential functions: Separate upper and lower dental trays with thermal impression material to engage the upper and lower dentition, a mechanism to provide the forward positioning of the lower jaw relative to the upper jaw in 3 positions determined by the user's bite providing up to +6 mm of possible advancement. The design allows the maintenance of the forward positioning of the lower jaw while allowing minor sagittal and vertical movement. The ZQuiet Advance is a "Boil-and-Bite" device that uses thermal impression resin for device retention. The upper and lower trays are both constructed out of a rigid plastic outer tray to provide the device structure and attachment points for the connecting semi-rigid straps that provide the adjustability. Both the upper and lower trays are lined with material that is moldable when heated in a water bath and provides the ability to conform the device to the individual's teeth.

    AI/ML Overview

    The provided text is a 510(k) summary for the ZQuiet Advance intraoral device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing the details of a study with specific acceptance criteria and performance metrics typically found for AI/ML-based medical devices or devices undergoing rigorous clinical trials to prove efficacy.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and training set details are not present in this document. The device in question is a physical intraoral device, not an AI/ML algorithm requiring such detailed performance characteristics.

    The document explicitly states:

    • "No clinical testing was performed in association with this submission." (Page 15, Section 3.13)
    • The justification for clearance relies on "Substantial Equivalence" to legally marketed predicate devices, meaning it has similar indications for use and technological characteristics.

    Instead of a study proving the device meets acceptance criteria in the sense of an AI/ML study, the non-clinical testing performed focuses on mechanical properties and human factors:

    • Non-Clinical Testing (Page 15, Section 3.11):

      • Strap compression testing (compared to similar designs).
      • Physical properties testing (by material manufacturer).
      • Human Factors Evaluation: Concluded that identified risks associated with critical tasks were adequately mitigated.
    • Biocompatibility (Page 15, Section 3.12):

      • Testing conducted by the device manufacturer.
      • Met requirements of ISO 10993-1 for mucosal membrane contacting devices with prolonged duration (cytotoxicity, sensitization, irritation).

    Conclusion based on the provided text:

    This document is a regulatory submission for a physical medical device. It does not provide the type of performance study details requested for an AI/ML-based device. The device's acceptance is based on demonstrating substantial equivalence to a predicate device through comparison of features, materials, and non-clinical testing (mechanical and human factors), rather than a clinical study with specific performance metrics against an established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140777

    Validate with FDA (Live)

    Device Name
    ZQUIET-SA
    Manufacturer
    Date Cleared
    2014-07-25

    (119 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ZQuiet-SA is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in adults.

    Device Description

    The ZQuiet-SA oral appliance consists of an upper and lower tray constructed in one piece and joined by a flexible hinge. The trays engage with the maxillary and mandibular dentition and the device maintains an anterior positioning of the mandible which widens the pharyngeal airway to prevent occlusion. The device is presented in multiple models allowing the dentist to recommend different degrees of mandibular advancement and preference of anterior posts.

    AI/ML Overview

    The provided 510(k) summary for the ZQuiet-SA device does not contain acceptance criteria or a study proving that the device meets such criteria in the traditional sense of a performance study with defined metrics. This submission relies on demonstrating substantial equivalence to predicate devices. Therefore, the information requested in some categories cannot be extracted or is not applicable.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Technological Characteristics Identical to Predicate DevicesThe ZQuiet-SA has identical technological characteristics as the predicate devices (ZQuiet K090503, ZQuiet K093407, TOMED SomnoGuard K061688) in terms of basic design (one-piece upper and lower tray, flexible hinge), materials (thermoplastic elastomer compound), and differential mandibular advancement capabilities. The ZQuiet-SA offers more advancement options (0mm, 2mm, 4mm, 6mm) compared to the predicate ZQuiet devices (6mm) and is comparable to SomnoGuard (0-5mm).
    Intended Use Identical to Predicate DevicesThe ZQuiet-SA is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in adults, which aligns with the intended use of comparable predicate devices.
    Biocompatibility Established by Predicate DevicesMaterials and manufacturing methods are identical to predicate devices K093407 and K090503, thus biocompatibility is considered established.
    Risk Analysis ConductedA Risk Analysis was conducted to demonstrate that product risks have been identified and appropriately accounted for. (No specific performance metrics are provided for the risk analysis itself beyond its completion).

    Detailed breakdown of requested information:

    1. A table of acceptance criteria and the reported device performance: See table above. It frames "acceptance criteria" as meeting the criteria for substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • No test set (clinical or performance) was used for this submission. The submission relies on substantial equivalence to predicate devices, which means no new clinical or non-clinical performance data was generated for this specific device to demonstrate its safety and effectiveness.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. No clinical or performance test set was used that required expert ground truth establishment.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. No test set was used.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. The ZQuiet-SA is a medical device (an intraoral appliance), not an AI/imaging diagnostic tool. No MRMC study was conducted.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The ZQuiet-SA is an intraoral appliance; it is not an algorithm and does not operate in a "standalone" or "human-in-the-loop" context in the way AI algorithms do.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable. No specific ground truth determination was made for this submission, as it relied on substantial equivalence. The "ground truth" for the predicate devices' effectiveness would have been established through clinical trials (e.g., outcomes data related to snoring and AHI reduction), but these studies were not conducted for the ZQuiet-SA as part of this 510(k) submission.
    8. The sample size for the training set:

      • Not applicable. No training set was used as this is not an algorithm-based device.
    9. How the ground truth for the training set was established:

      • Not applicable. No training set was used.

    Summary of the K140777 Submission:

    This 510(k) submission for the ZQuiet-SA device is a direct demonstration of substantial equivalence. The manufacturer asserts that the ZQuiet-SA possesses "identical technological characteristics" and "identical materials and methods of manufacture" to its previously cleared ZQuiet devices (K093407 and K090503) and also compares it to another predicate, TOMED SomnoGuard (K061688).

    The "study" that proves the device meets the acceptance criteria is the comparison of its technical characteristics and intended use to those of legally marketed predicate devices. The acceptance criteria are, implicitly, that these characteristics and intended use are sufficiently similar to deem it substantially equivalent, thereby implying comparable safety and effectiveness without the need for new clinical data.

    Specifically, the document states:

    • "No additional non-clinical testing was performed for this submission."
    • "No additional biocompatibility testing was conducted" because materials and manufacturing are identical to predicate devices.
    • "No clinical testing was performed in association with this submission."

    The FDA's letter concurs with the substantial equivalence determination, allowing the device to be marketed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122475

    Validate with FDA (Live)

    Device Name
    ZQUIET PRO-PLUS
    Manufacturer
    Date Cleared
    2012-12-07

    (115 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZQuiet® Pro-Plus oral appliance is intended to reduce or alleviate nighttime snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older.

    Device Description

    The ZQuiet® Pro-Plus is an intra oral device used for treating snoring and sleep apnea. It functions to reposition the mandible to increase the patient's pharyngeal space to increase airflow, reduce the vibrations of the soft palate associated with snoring, and/or prevent the collapse of the airway associated with apneic events. The device consists of two (2) custom orthotics; one for the maxillary arch and one for the mandibular arch, made from acrylics widely used and approved for use in the dental industry. The orthotic employs a buccal cam mechanism, consisting of an apposing fin and buttress, that maintains the orthotics in an anteriorized relationship when worn during sleep.

    AI/ML Overview

    The provided text is a 510(k) Summary for the ZQuiet® Pro-Plus oral appliance, seeking substantial equivalence to a predicate device. It defines the device, its intended use, and compares it to a legally marketed predicate device. However, it does not contain information about acceptance criteria or a study proving the device meets specific performance criteria in terms of clinical outcomes.

    Instead, the submission focuses on demonstrating substantial equivalence to an existing predicate device based on similar intended use, technological characteristics, and materials. The non-clinical testing mentioned is related to material properties rather than clinical performance for treating snoring or sleep apnea.

    Therefore, the following information cannot be extracted from the provided document:

    1. A table of acceptance criteria and the reported device performance: This document does not describe performance metrics and associated acceptance criteria related to reducing snoring or OSA, nor does it present results from a study designed to measure these.
    2. Sample size used for the test set and the data provenance: No clinical test set is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no clinical test set with ground truth is described.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an oral appliance, not an AI-assisted diagnostic tool for human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used: Not applicable as no clinical study is detailed.
    8. The sample size for the training set: Not applicable as there is no mention of a training set for a machine learning algorithm.
    9. How the ground truth for the training set was established: Not applicable.

    What the document does indicate regarding "study" and "proof":

    • Non-Clinical Testing: The document mentions "Biocompatibility test results are known for this device as well as tests for tear resistance tear strength, tensile strength, elongation yield, and modulus of elasticity." These tests relate to the material properties and safety of the device, not its clinical efficacy in treating snoring or OSA.
    • Proof of Substantial Equivalence: The primary "proof" presented in this 510(k) submission is the comparison to a predicate device, the SomnoMed MAS RxA (K050592). The conclusion states: "By comparing designs to the predicate device and the materials used it's concluded that the ZQuiet® Pro-Plus is as safe, as effective, and performs as well as the predicate device." This is a regulatory pathway that relies on the predicate device's established safety and effectiveness, rather than a new clinical study for the ZQuiet® Pro-Plus.
    Ask a Question

    Ask a specific question about this device

    K Number
    K093407

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-22

    (232 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZQuiet mandibular advancement device is intended for the treatment of nighttime snoring in adults 18 years or older.

    Device Description

    The ZQuiet anti-snoring device is a single piece anti-snoring device, which moves the lower jaw forward and helps reduce the likelihood of snoring. This is achieved by covering the upper and lower teeth with a resilient non-toxic thermoplastic elastomer compound. The ZQuiet is easy to wear and simple to use. The single shot manufacturing process incorporates a resilient hinge in the molar area to provide a single piece device. The anti-snore device comprises an upper member adapted to engage the maxillary dentition of a human and a lower member adapted to engage the mandibular dentition of the human, the upper and lower members being resiliently hinged together.

    AI/ML Overview

    The ZQuiet Mouthpiece is an anti-snoring device. The study performed for its clearance was a non-clinical evaluation rather than a clinical study. The device claims substantial equivalence to its predicate device (K090503 ZQuiet Mouthpiece) based on identical intended use and design, with a material change that was tested to ensure safety.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedStandard/MethodologyAcceptance Criteria (Implied)Reported Device Performance
    Agar Diffusion Test (in vitro cytotoxicity)ISO 10993-5: 1999 "Tests for in vitro cytotoxicity"Not cytotoxicPass - Not considered cytotoxic
    Primary Dermal Irritation in RabbitsFederal Hazardous Substances Act Regulations (16 CFR 1500.41)Not a primary dermal irritantPass - Not a primary dermal irritant
    Guinea Pig Closed Patch Sensitization TestISO 10993-10: 2002 Tests for irritation and delayed type hypersensitivityNo sensitizing propertiesPass - No sensitizing properties
    Oral Mucosal Irritation StudyISO 10993 – Part 10 – Tests for Irritation and Delayed-Type HypersensitivityMinimal irritant or lessConsidered to be a minimal irritant.

    2. Sample size used for the test set and the data provenance:

    • Agar Diffusion Test: Not specified (in vitro test).
    • Primary Dermal Irritation in Rabbits: Rabbits (specific number not provided but typically a small group for such tests). The data provenance is pre-clinical animal testing.
    • Guinea Pig Closed Patch Sensitization Test: Guinea pigs (specific number not provided). The data provenance is pre-clinical animal testing.
    • Oral Mucosal Irritation Study: Not specified (likely animal model or in vitro, but typically animal models for irritation studies). The data provenance is pre-clinical testing.

    The document does not specify country of origin for the data, but these are standard international bio-compatibility tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. The tests performed were bio-compatibility and safety studies (in vitro and animal models), not human clinical studies involving expert interpretation of results. The "ground truth" was established by the scientific methodologies of the respective ISO standards and regulations.

    4. Adjudication method for the test set:

    Not applicable. These were laboratory tests with objective endpoints (e.g., cytotoxicity, irritation scores) rather than subjective assessments requiring adjudication by multiple experts.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    No. This was a non-clinical safety evaluation for a material change, not a study assessing diagnostic or treatment effectiveness involving human readers or cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. This device is a physical anti-snoring mouthpiece, not a software algorithm.

    7. The type of ground truth used:

    The ground truth used was based on established scientific methodologies and standards for biocompatibility and safety testing (e.g., ISO 10993 standards, Federal Hazardous Substances Act Regulations). The "truth" refers to the cytotoxic, irritating, or sensitizing potential of the material under test.

    8. The sample size for the training set:

    Not applicable. This device does not use machine learning, so there is no "training set."

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set as it's not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090503

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-03-10

    (13 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZQuiet Anti-Snoring device is intended for the treatment of nightime snoring in adults.
    The ZQuiet mandibular advancement device is intended for the treatment of nighttime snoring in adults.

    Device Description

    The ZQuiet ® Mouthpiece is an Anti-Snoring device consisting of :

    • Two trayed plates fitted in front and between the upper and lower teeth and gums . and integrated with each other with the same material as the upper and lower plates.
    • The device is made of Thermoplastic elastomer .
    • ◆ May be used as supplied
    AI/ML Overview

    Here's an analysis of the provided text regarding the ZQuiet® Mouthpiece, focusing specifically on acceptance criteria and supporting studies.

    Important Note: The provided document is a 510(k) summary and subsequent FDA clearance letter. It is designed to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to prove efficacy or establish novel performance criteria through extensive clinical trials. Therefore, the information regarding acceptance criteria and performance studies will be limited by the nature of this regulatory submission.


    Acceptance Criteria and Device Performance

    Based on the provided text, specific quantitative acceptance criteria for device performance (e.g., reduction in snoring decibels, AHI improvement) are not explicitly stated or provided in a table format. The submission focuses on demonstrating substantial equivalence to a predicate device (SomnoGuard Series, K061688) based on shared intended use, materials, design, and theory of operation, implying that if the new device is substantially equivalent, it meets the inherent safety and effectiveness standards already established for the predicate.

    Implicit Acceptance Criteria (based on substantial equivalence):

    Acceptance Criteria CategoryDescriptionReported Device Performance (ZQuiet® Mouthpiece)
    Intended UseTreatment of nighttime snoring in adults.Intended for the treatment of nighttime snoring in adults. (Identical to predicate)
    Device TypeMandibular Advancement Device; intraoral device.Mandibular Advancement Device; intraoral device. (Identical to predicate)
    Core FunctionalityReduces or helps alleviate snoring by advancing the lower jaw to open the upper airway.Functions by advancement of lower jaw to open upper airway. (Identical to predicate)
    Material CompositionMade of Thermoplastic elastomer.Made of Thermoplastic elastomer. (Identical to predicate)
    BiocompatibilityExpected to be biocompatible for intraoral use, not raising new safety concerns."ZQuiet® Anti-Snoring Device has been evaluated for safety through in vitro tests and animal safety studies." The conclusion states it "does not raise any new issues concerning safety and effectiveness" compared to the predicate device, implying satisfactory biocompatibility.
    Design CharacteristicsKey design features such as being ready-to-use, having upper and lower trays, a one-piece design, allowing breathing through the mouth, daily placement/removal, and daily cleaning. (Differences noted for "Boil and Bite fitting" and indication for "mild to moderate" sleep apnea are highlighted as not raising new safety/effectiveness issues for this non-Boil & Bite, snoring-only device).Ready-to-Use; Upper and lower trays; One piece design; Permits patient to breathe through the mouth; Placed in user's mouth each evening; Cleaned daily; Easily removed from the mouth. (All match predicate, with noted differences in fitting method and specific apnea indication that are argued not to raise new issues).
    Regulatory ClassificationClass II medical device, product code LRK.Classified as Class II, product code LRK.

    Study Details:

    1. Sample Size and Data Provenance for Test Set:

      • The submission does not mention a clinical study with a "test set" or a comparison against a "ground truth" to establish the device's efficacy in reducing snoring.
      • The "Non-Clinical Performance Data" section only mentions "in vitro tests and animal safety studies" for biocompatibility. No sample sizes are provided for these.
      • This is typical for a 510(k) submission where substantial equivalence to a predicate device is the primary argument, rather than a de novo efficacy claim requiring human clinical trials for performance.
    2. Number of Experts and Qualifications for Ground Truth:

      • Not applicable. There is no mention of establishing ground truth using experts for performance claims, as no performance study is described. The assessment focuses on the device's technical characteristics and safety relative to a predicate.
    3. Adjudication Method for Test Set:

      • Not applicable. No test set or related adjudication process is described for performance evaluation.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No such study was done or is reported. This type of study is common for imaging or diagnostic devices where human interpretation is involved. The ZQuiet® Mouthpiece is a therapeutic device, and the submission does not involve human interpretation of data where MRMC would be relevant.
    5. Standalone Performance Study (Algorithm Only):

      • No such study was done or is reported. This device is a physical medical device, not an algorithm or software. Its performance is directly tied to its physical interaction with the user's anatomy.
    6. Type of Ground Truth Used:

      • For the non-clinical safety evaluation, the "ground truth" for biocompatibility would be established by standard biological safety testing protocols (e.g., ISO 10993 series), rather than clinical outcomes or expert consensus.
      • For the overall regulatory clearance, the "ground truth" is the legally marketed predicate device (SomnoGuard Series K061688), against which the new device is deemed substantially equivalent in terms of intended use, technology, and safety/effectiveness.
    7. Sample Size for Training Set:

      • Not applicable. This device is not an AI/ML algorithm that requires a "training set" of data.
    8. How Ground Truth for Training Set was Established:

      • Not applicable. As above, no training set is relevant to this type of device and submission.

    In summary: The K090503 submission for the ZQuiet® Mouthpiece is a 510(k) premarket notification. Its "study" for acceptance criteria is primarily a comparison to a legally marketed predicate device to demonstrate substantial equivalence, rather than a clinical trial designed to prove independent efficacy against quantitative acceptance criteria. The acceptance criteria are implicit: that the device's characteristics (components, materials, intended use, mechanism of action, design) are sufficiently similar to the predicate that it raises no new questions of safety or effectiveness. The only explicit performance data mentioned are non-clinical (in vitro and animal safety studies) to assess biocompatibility, but no details or sample sizes are provided for these.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1