510(k) Data Aggregation

    K Number
    K242712

    Validate with FDA (Live)

    Device Name
    LubriZenzs
    Manufacturer
    Date Cleared
    2025-06-04

    (268 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LubriZenzs is a personal lubricant for vaginal and/or penile application, intended to moisturize and lubricate to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    LubriZenzs is a non-sterile, water-based, over-the-counter personal lubricant that is compatible with natural rubber latex and polyisoprene condoms. Its formulation consists of water, propanediol, cellulose gum, algin, 1,2-hexanediol, sodium hyaluronate, lactic acid, gluconolactone, Tremella fuciformis (mushroom) extract, caprylhydroxamic acid, sodium benzoate, calcium gluconate, sodium hydroxide, xylitylglucoside, anhydroxylitol, glycogen, xylitol, and glucose.

    The Lubricant comes in a 100 mL white polypropylene plastic airless pump bottle, with a clear overcap. The lubricant has a shelf-life of 2 years.

    Specifications:

    • Appearance: Clear, flowable
    • Color: Colorless
    • Odor: Odorless
    • pH: 4.5-5.0
    • Viscosity: 10,000 – 25,000 cps
    • Osmolality: 135 - 380 mOsm/kg (1:10 dilution)
    • Total Microbial Count (TAMC): < 100 cfu/g
    • Fungal/Yeast/Mold Limits (TYMC): < 10 cfu/g
    • Absence of Pathogenic Organisms (Pseudomonas aeruginosa, Staphylococcus aureus, Salmonella, Escherichia coli, Candida albicans): Absent
    • Antimicrobial Effectiveness: Meets USP <51> criteria for category 2. Bacteria: No less than 2.0 log reduction from initial count at 14 days, and no increase from the 14-day count at 28 days. Yeast and Molds: No increase from the initial calculated count at 14 days and 28 days.
    AI/ML Overview

    The FDA 510(k) clearance letter for LubriZenzs (K242712) pertains to a personal lubricant, not an AI-powered medical device. Therefore, the information requested in points 2, 3, 4, 5, 6, 7, 8, and 9 (related to AI study design, experts, ground truth, and training sets) is not applicable to this submission.

    However, I can extract the acceptance criteria and the summary of performance testing for the LubriZenzs device as described in the provided text.


    Acceptance Criteria and Device Performance for LubriZenzs

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance CriteriaReported Device Performance
    AppearanceClear, flowableMeets specification
    ColorColorlessMeets specification
    OdorOdorlessMeets specification
    pH4.5 – 5.0Meets specification (USP <791>)
    Viscosity10,000 – 25,000 cpsMeets specification (USP <912>)
    Osmolality135 - 380 mOsm/kg (1:10 dilution)Meets specification (USP <785>)
    Total Microbial Count (TAMC)< 100 cfu/gMeets specification (USP <61>)
    Fungal/Yeast/Mold Limits (TYMC)< 10 cfu/gMeets specification (USP <61>)
    Absence of Pathogenic OrganismsAbsent (Pseudomonas aeruginosa, Staphylococcus aureus, Salmonella, Escherichia coli, Candida albicans)Meets specification (USP <62>)
    Antimicrobial EffectivenessMeets USP <51> criteria for category 2 (Bacteria: No less than 2.0 log reduction from initial count at 14 days, and no increase from the 14-day count at 28 days. Yeast and Molds: No increase from the initial calculated count at 14 days and 28 days.)Meets USP <51> criteria
    BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritating, not acutely systemically toxicDemonstrated to be non-cytotoxic, non-sensitizing, non-irritating, and not acutely systemically toxic
    Condom CompatibilityCompatible with natural rubber latex and polyisoprene condoms; not compatible with polyurethane condomsCompatible with natural rubber latex condoms and synthetic polyisoprene condoms; not compatible with polyurethane condoms
    Shelf Life2 yearsDemonstrated 2-year shelf-life, meeting specifications at endpoint

    2. Sample size used for the test set and the data provenance:

    • Biocompatibility: The document does not specify the exact sample sizes for each biological test (Cytotoxicity, Sensitization, Vaginal Irritation, Acute Systemic Toxicity). The provenance of the data (e.g., country of origin, retrospective/prospective) is not mentioned.
    • Condom Compatibility: The document does not specify the sample size of condoms or lubricant used for testing. The provenance of the data is not mentioned.
    • Shelf-Life Testing: The document does not specify the number of samples or batches tested for real-time stability. The provenance of the data is not mentioned.
    • Other Device Specifications (Appearance, Color, Odor, pH, Viscosity, Osmolality, Microbial Counts): The document does not specify sample sizes for these tests or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. This is a lubricant, and its performance is assessed via standardized chemical, physical, and biological tests, not expert interpretation of diagnostic images or clinical data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. Performance is based on objective measurements against predefined specifications using established test methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is a personal lubricant, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is a personal lubricant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The "ground truth" for this device is based on compliance with predefined physical, chemical, and biological specifications established by relevant industry standards (e.g., USP, ASTM, ISO) and regulatory guidance. For example:

    • Chemical/Physical properties (pH, viscosity, osmolality): Measured against numerical ranges.
    • Microbiological purity: Measured against limits for colony-forming units and absence of specific pathogens.
    • Biocompatibility: Assessed through in-vitro and in-vivo tests designed to detect cytotoxic, sensitizing, irritating, or toxic effects.
    • Condom compatibility: Tested according to a specific ASTM standard (ASTM D7661-18) which defines compatibility criteria.

    8. The sample size for the training set:
    Not applicable. This is a physical product (lubricant), not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established:
    Not applicable. As above, no training set is involved.