Found 2 results

510(k) Data Aggregation

    K Number
    K160770

    Validate with FDA (Live)

    Date Cleared
    2016-06-29

    (100 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SinuSys Vent-Os® Sinus Dilation family is intended to provide a means to access the sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.

    • The Vent-Os Maxillary Dilation System is indicated for dilation of Maxillary sinus and associated spaces.
    • · The Vent-Os Frontal Dilation System is indicated for dilation of the Frontal recess and associated spaces.
    • · The Vent-Os Sphenoid Dilation System is indicated for dilation of Sphenoid sinus and associated spaces.
    Device Description

    The Vent-Os® Sinus Dilation family comprises a family of osmotically driven devices that access and dilate the sinus ostia and spaces associated with the paranasal sinus cavities. Each member of the Vent-Os® Sinus Dilation family is a system that includes a Dilation Device preloaded on a Placement Instrument. The Placement Instrument accesses the target site through the nasal passageway and delivers the Dilation Device. The Vent-Os® Sinus Dilation Device then expands and remodels the target tissues and then is removed.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided text.

    Based on the provided text, the device in question is the "Vent-Os® Sinus Dilation Family."

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for outcomes like "patent" rates for the clinical trial. Instead, it compares the device's performance to a literature control to demonstrate "substantial equivalence." The "acceptance criteria" can be inferred as achieving comparable safety and effectiveness to the predicate device and the literature control.

    Attribute Tested/OutcomeAcceptance Criteria (Inferred: Comparable to Literature Control / Passed)Reported Device Performance (Vent-Os Study)
    BiocompatibilityNon-cytotoxic, Non-sensitizing, Non-irritatingCytotoxicity (MEM Elution): Non-cytotoxicSensitization-Maximization: Non-sensitizingIntracutaneous Reactivity: Non-irritating
    Bench TestingAll attributes PassedDevice weight: PassedPlacement System Working Length Profile: PassedPlacement System Working Length: PassedCrossing Profile: PassedDeployment Force: PassedDilation Device Diameter @ 1 hour: PassedDilation Device Working Length @ 1 hour: PassedDilation Device Integrity: PassedBond Strength: Passed
    Transportation, Sterilization, & Shelf-Life TestingAll attributes Passed (Sterility Assurance Level (SAL) of 10⁻⁶)Sterilization: Achieved SAL of 10⁻⁶Transit Testing: PassedSterile Barrier Integrity (Bubble Test): PassedPackaging & Product Visual Inspection: PassedPouch Seal Testing: Passed
    Cadaver TestingSuccessful delivery and placementEvaluators confirmed successful delivery and placement of the devices to the targeted sinus.
    Clinical Outcomes (Frontal Recess - Acute)Comparable to Bolger Study (69% patent)Vent-Os Study: 33/33 (100%) patent (95% CI: 92.4% - 100%)Bolger Study (Literature Control): 85/124 (69%) patent (95% CI: 59.9% - 76.1%)
    Clinical Outcomes (Frontal Recess - 3 Months)Comparable to Bolger Study (84% patent)Vent-Os Study: 28/30 (93%) patent (95% CI: 78.7% - 98.2%)Bolger Study (Literature Control): 82/98 (84%) patent (95% CI: 75.1% - 89.7%)
    Clinical Outcomes (Sphenoid Sinus Ostium - Acute)Comparable to Bolger Study (55% patent)Vent-Os Study: 13/15 (87%) patent (95% CI: 62.1% - 96.3%)Bolger Study (Literature Control): 41/75 (55%) patent (95% CI: 43.5 - 65.4%)
    Clinical Outcomes (Sphenoid Sinus Ostium - 3 Months)Comparable to Bolger Study (59% patent)Vent-Os Study: 10/15 (67%) patent (95% CI: 41.7% - 84.8%)Bolger Study (Literature Control): 32/54 (59%) patent (95% CI: 46.0% - 71.3%)

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Clinical Trial Test Set: 30 patients.
      • Data Provenance: The study was a "multi-center, prospective, single arm, non-randomized trial conducted at three (3) clinical sites." The country of origin is not explicitly stated, but given the FDA submission, it's typically understood to be within or recognized by the US regulatory framework.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document states "three (3) investigators" for the clinical trial. It implies these investigators were physicians involved in treating patients with chronic rhinosinusitis and evaluating the outcomes. Specific qualifications (e.g., years of experience, specific specialization beyond "physician users") are not detailed.
      • For cadaver testing, "multiple physician users" were involved in validating usability.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • The document does not describe any specific adjudication method for the clinical outcomes. It implies the "investigators" assessed the success of the treatment, but details on how agreement was reached for indeterminate cases or divergences in assessment are not provided.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is a surgical instrument, not an AI-assisted diagnostic tool for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is not applicable as the Vent-Os® Sinus Dilation Family is a surgical instrument and not an algorithm or AI system.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the clinical trial, the outcomes (patent, non-patent, indeterminate) appear to be based on clinical assessment by the physician investigators (outcomes data) at the 3-month follow-up. This type of assessment often relies on endoscopic evaluation and investigator judgment.
      • For cadaver testing, the ground truth was expert observation/confirmation of successful delivery and placement by "multiple physician users."
    7. The sample size for the training set:

      • As this is a surgical device and not an AI or machine learning model, there is no "training set" in the conventional sense. The development of the device would involve engineering, material science, and iterative testing, but not data-driven training of an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133016

    Validate with FDA (Live)

    Date Cleared
    2013-12-09

    (75 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vent-Os™ Sinus Dilation System is an instrument intended to provide a means to access the sinus space and to dilate the maxillary sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.

    Device Description

    The Vent-OsTM Sinus Dilation System is an osmotically driven device that dilates maxillary sinus ostia and associated spaces. The Vent-Os™ Sinus Dilation System is comprised of a Dilation Device preloaded on a Placement Instrument. The Placement Instrument accesses the target site through the nasal passageway and delivers the Dilation Device. The Vent-Os™ Sinus Dilation Device then expands and remodels the target tissues and then is removed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Vent-Os™ Sinus Dilation System, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary focuses on demonstrating "substantial equivalence" rather than providing explicit numeric acceptance criteria for each test. Instead, it reports "Passed" or specific outcomes that are compared to a literature control for clinical efficacy.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility
    Non-cytotoxicNon-cytotoxic
    Non-sensitizingNon-sensitizing
    Non-irritatingNon-irritating
    In Vitro Bench Evaluation
    Device weight acceptablePassed
    Placement System Working Length Profile acceptablePassed
    Placement System Working Length acceptablePassed
    Crossing Profile acceptablePassed
    Deployment Force acceptablePassed
    Dilation Device Diameter @ 1 hour acceptablePassed
    Dilation Device Working Length @ 1 hour acceptablePassed
    Dilation Device Integrity acceptablePassed
    Bond Strength acceptablePassed
    Proximal Anchoring Force acceptablePassed
    Sterilization & Packaging
    Sterilization compliantCompliant with ANSI/AAMI/ISO 11137-2
    Transportation integrityPassed (Transit Testing)
    Sterile barrier integrityPassed (Bubble Test)
    Packaging & System visual inspection acceptablePassed
    Pouch seal integrityPassed
    Clinical Efficacy (Maxillary Ostia Patency)SinuSys Study (Per Protocol)
    Acute Patency comparable to literature54/57 (95%) with 95% CI: 85.6% - 98.2%
    3 Months Patency comparable to literature51/55 (93%) with 95% CI: 82.7% - 97.1%

    Study That Proves the Device Meets Acceptance Criteria:

    The 510(k) submission details several studies to demonstrate safety and effectiveness for substantial equivalence:

    • Biocompatibility Studies: In compliance with EN ISO 10993 requirements (ISO 10993-5:2009 and ISO 10993-10:2010).
    • In Vitro Bench Evaluation: Performance testing was completed on various physical and mechanical attributes of the device.
    • Transportation, Sterilization, and Shelf-life Testing: Sterilization compliant with ANSI/AAMI/ISO 11137-2. Transportation and packaging validation included transit testing, sterile barrier integrity, packaging/system visual inspection, and pouch seal testing.
    • Animal Studies: Four (4) Dilation Devices were deployed into four (4) separate sheep maxillary sinuses to confirm safe placement and dilation.
    • Cadaver Testing: Performed by multiple users (investigators) using two (2) adult cadaveric head specimens. The device was placed four (4) times into each target site.
    • Clinical Trial: A prospective, non-randomized, single-arm, open-label study was conducted.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Trial Test Set Sample Size:
      • Patients: 34 patients
      • Maxillary Ostia Treated: 57 maxillary ostia
      • Data Provenance: The study was conducted at five (5) clinical sites. The summary does not specify the country of origin, but given the applicant is a US company (Palo Alto, CA) and the FDA submission is for the US market, it's highly probable the study was conducted in the United States.
      • Retrospective or Prospective: Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The summary states that five (5) investigators conducted the clinical trial at five (5) clinical sites.
    • Their specific qualifications are not explicitly stated beyond being "investigators," but for a study involving sinus dilation, they would typically be Ear, Nose, and Throat (ENT) specialists or Otolaryngologists. Their years of experience are not mentioned.
    • The "ground truth" for the clinical trial was likely established by these investigators through post-procedure assessments of ostia patency, potentially using endoscopy or other clinical evaluation methods.

    4. Adjudication Method for the Test Set

    • The information provided does not specify an explicit adjudication method (e.g., 2+1, 3+1). For a single-arm, open-label study, it's common for the treating physician/investigator to evaluate the outcome. There is no mention of independent reviewers or a consensus process for the clinical trial results presented in the table.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a surgical instrument, not an AI diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. This device is a manual surgical instrument, not an algorithm, so a standalone algorithmic performance study was not performed.

    7. Type of Ground Truth Used

    • Biocompatibility: Standardized laboratory test results based on established ISO guidelines.
    • In Vitro Bench: Quantitative measurements and qualitative assessments ("Passed") against pre-defined performance specifications.
    • Animal Studies: Direct observation and assessment of safe placement and dilation in sheep maxillary sinuses by qualified personnel.
    • Cadaver Testing: Endoscopic delivery and placement of the device confirmed by investigators.
    • Clinical Trial: Clinical assessment of maxillary ostia patency at acute and 3-month follow-ups, likely by the treating investigators based on clinical examination (e.g., endoscopy). This falls under expert clinical assessment/outcomes data.

    8. Sample Size for the Training Set

    • Not applicable. This is a physical medical device, not an AI/Machine Learning algorithm, so there is no "training set" in the computational sense. The "training" for the device would involve manufacturing process controls, design iterations, and perhaps early prototype testing, none of which are detailed as a formal "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for an algorithm. The "ground truth" during device development and manufacturing would be established through engineering specifications, material standards, and quality control metrics.