510(k) Data Aggregation

    K Number
    K213159

    Validate with FDA (Live)

    Device Name
    Encore System
    Date Cleared
    2021-12-22

    (85 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and/or snoring.

    Device Description

    The Encore System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone and hyoid bone suspension to the mandible bone using a bone screw and suspension lines. The Encore System consists of 1) an integrated suture passer pre-loaded with size #2-0 braided polyester suture, 2) three (3) bone screws and two (2) bone screw inserters, 3) a suspension line lock tool, 4) a threading tool, and 5) a drill bit. In addition, the following suspension lines are provided depending on the model number: 1) #2 braided polyester suspension line and 2) #2 braided polyethylene suspension line.

    AI/ML Overview

    This document describes the Encore System, an intraoral device for treating obstructive sleep apnea (OSA) and snoring, and its substantial equivalence to a predicate device. The primary focus of this submission (K213159) is on updated MRI safety information and a minor drill bit length modification, rather than a clinical study evaluating the device's therapeutic effectiveness for OSA or snoring. Therefore, comprehensive acceptance criteria and a study proving therapeutic efficacy are not detailed in this 510(k) summary.

    However, based on the provided text, we can infer the acceptance criteria for the MRI safety update and the corresponding testing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (for MRI Safety)Reported Device Performance (from summary)
    Safe for MRI scanning under specified conditions.Encore Bone Screw may be safely scanned under specified conditions (1.5T to 3T, specific spatial field gradient, RF excitation, RF coil type, SAR, and scan duration).
    Image artifact within acceptable limits.The presence of this implant may produce an image artifact of 12mm.
    No significant risk to patient in MRI environment.Performance testing confirmed that the implantable metallic components do not pose a significant risk to a patient in a clinical MRI environment under the conditions presented in the labeling.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical testing regarding MRI safety. Therefore, there is no "test set" of patient data in the context of clinical outcomes. The "sample size" would refer to the components tested in the MRI safety experiments. The document does not specify the exact number of components or systems tested.

    • Data provenance: Non-clinical testing for MRI compatibility. The country of origin for this testing is not specified, but it would typically be conducted by a certified testing lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. As this is non-clinical MRI safety testing, there are no human experts establishing "ground truth" in the clinical sense. MRI safety evaluations typically rely on established testing standards and protocols interpreted by engineers and physicists specializing in MR compatibility.

    4. Adjudication method for the test set

    N/A. There is no clinical test set requiring adjudication in this context. The MRI safety testing follows standardized protocols, and results are typically evaluated against pre-defined thresholds.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This 510(k) submission is for an intraoral device, not an AI or imaging diagnostic device. Therefore, an MRMC study is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    N/A. This 510(k) submission is for an intraoral device, not an algorithm.

    7. The type of ground truth used

    For the MRI safety testing, the "ground truth" is defined by established international standards (e.g., ASTM F2052, ASTM F2182) and the physical properties of the materials and device. The acceptance criteria are derived from these standards and generally accepted safety limits for MRI.

    8. The sample size for the training set

    N/A. This is a non-clinical MRI safety evaluation, not a machine learning or AI algorithm development.

    9. How the ground truth for the training set was established

    N/A. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201238

    Validate with FDA (Live)

    Device Name
    Encore System
    Date Cleared
    2020-10-02

    (148 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and / or snoring.

    Device Description

    The Encore System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone and hyoid bone suspension to the mandible bone using a bone screw and suspension lines. The Encore System consists of 1) an integrated suture passer pre-loaded with size #2-0 braided polyester suture, 2) three (3) bone screws and two (2) bone screw inserters, 3) a suspension line lock tool, 4) a threading tool, and 5) a drill bit. In addition, a #2 braided polyethelene suspension line is provided. All components are provided sterile.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification from the FDA regarding the Siesta Medical Encore System. It's a regulatory clearance document and does not contain the detailed study results, acceptance criteria, or performance metrics typically found in a clinical study report.

    Therefore, I cannot provide the requested information, which includes a table of acceptance criteria and reported device performance, sample sizes for test and training sets, data provenance, number and qualifications of experts, adjudication methods, details of MRMC studies (effect size), standalone performance, or ground truth establishment.

    The document states:

    • The device is the Encore™ System.
    • Intended Use: Anterior advancement of the tongue base and hyoid suspension for the treatment of obstructive sleep apnea (OSA) and/or snoring.
    • Change: The current submission (K201238) introduces a #2 braided polyethylene suspension line (vs. #2 braided polyester in the predicate K183310).
    • Non-Clinical Testing: A "Suture Tensile Strength Test" was conducted.
    • Conclusion of Testing: The results "confirmed that the strength of the #2 braided polyethylene suspension is similar to that of the existing #2 braided polyester suspension line."

    This type of submission typically relies on demonstrating substantial equivalence to a predicate device, focusing on a specific change (in this case, the material of the suspension line) and ensuring that the new material does not negatively impact the device's fundamental function or safety. The "Suture Tensile Strength Test" is the evidence provided to support this claim, indicating that the new suspension line has comparable strength to the old one.

    Without further documentation, it is impossible to fulfill the request for detailed study information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133680

    Validate with FDA (Live)

    Device Name
    ENCORE SYSTEM
    Date Cleared
    2014-03-26

    (114 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENCORE System is intended to be used for anterior advancement of the tongue base. It is also suitable for performing hyoid suspension. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The ENCORE System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone and hyoid bone suspension to the mandible bone using a bone screw and suspension lines. The ENCORE System consists of 1) an integrated suture passer pre-loaded with size #2-0 braided polyester suture, 2) two (2) bone screws and bone screw inserters, 3) a suspension line lock tool, and 4) a threading tool. In addition, the following suspension lines are provided depending on the model number: 1) a #2 monofilament polypropylene suspension line with a radiopaque marker, 2) a size #2 braided polyester suspension line, and 3) a size #2 braided polyester suspension line with a radiopaque marker.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study details for the Siesta Medical, Inc. ENCORE System, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state quantitative "acceptance criteria" in the format one might expect for a clinical trial's primary endpoints (e.g., AHI reduction by a certain percentage with a specific p-value threshold). Instead, it relies on demonstrating substantial equivalence to predicate devices and shows that the device performs at least as well as and is as safe and effective as the predicates.

    For the non-clinical testing, the criteria were implicitly that the device met performance specifications for:

    • Suspension Line Endurance: Demonstrated to meet specifications (specific values not provided in this summary).
    • Bone Screw-to-Suspension Line Fixation Strength: Demonstrated to meet specifications (specific values not provided in this summary).

    For the clinical performance, the reported outcomes are compared against the understanding of predicate device performance and general clinical efficacy for hyoid suspension.

    Acceptance Criteria (Implicit from Submission)Reported Device Performance
    Non-Clinical:
    Meet performance specifications for Suspension Line Endurance.Demonstrated to meet performance specifications.
    Meet performance specifications for Bone Screw-to-Suspension Line Fixation Strength.Demonstrated to meet performance specifications.
    Clinical:
    Be as safe and effective as predicate devices for anterior advancement of the tongue base and hyoid suspension.Literature Review: 11 clinical studies (461 patients) involving hyoid suspension reported few complications, with six studies reporting no significant intraoperative or post-operative complications. Patient Series: 1 patient series (number of patients not explicitly stated for this series alone, but included from context of a retrospective analysis) showed similar efficacy and safety.
    Reduction in AHI/RDI from 52.8 to 11.8 (78%, p<0.01) in the patient series, with mean follow-up of 75 days.
    No new safety or performance issues compared to predicate devices.No new safety or performance issues raised during testing.
    Substantial equivalence to predicate devices (indications for use and technological characteristics).Determined to be substantially equivalent based on non-clinical and clinical data.

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-Clinical (Bench Testing): The document does not specify the exact sample sizes (e.g., number of devices or components) used for the Suspension Line Endurance Test and Bone Screw-to-Suspension Line Fixation Strength Test.
    • Clinical (Patient Series): "A series of patients treated for OSA using hyoid suspension with the Encore System." The exact number of patients in this specific patient series is not explicitly stated in the summary, but it implies a single cohort. It mentions a mean age of 58.3 years, mean BMI of 27.9 kg/m², and baseline AHI/RDI of 45.0. It's a retrospective analysis. The country of origin is not specified but implicitly assumed to be the US, given the FDA submission.
    • Clinical (Literature Review): 11 clinical studies involving 461 patients. This is a review of retrospective and/or prospective studies previously published in peer-review literature. The countries of origin for these studies are not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to a 510(k) submission for a device like the ENCORE System. The "ground truth" for non-clinical testing is based on engineering specifications and direct measurement. For clinical studies, the "ground truth" (e.g., diagnosis of OSA, AHI/RDI measurement, complication reporting) is established by the clinical investigators and standard medical diagnostic procedures, not typically by a panel of independent experts for the purpose of a 510(k) submission comparison in this context. The peer-reviewed literature would have been subject to its own expert review processes, but this is distinct from experts establishing 'ground truth' for the 510(k) submission itself.

    4. Adjudication Method for the Test Set

    Not applicable. This device's evaluation does not involve subjective "reads" that would require adjudication. Clinical outcomes (e.g., AHI reduction, complications) are typically objectively measured or reported by treating physicians.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The ENCORE System is a physical medical device (bone screw system for tongue/hyoid suspension), not an AI algorithm or imaging device requiring human reader interpretation. No MRMC study was conducted or relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The ENCORE System is a physical medical device. No algorithm is involved in its direct function for standalone performance evaluation in this context.

    7. The Type of Ground Truth Used

    • Non-Clinical: Engineering specifications and direct physical measurements for endurance and fixation strength.
    • Clinical (Patient Series & Literature Review):
      • Diagnosis of OSA: Likely based on polysomnography (PSG) and clinical evaluation according to established medical guidelines.
      • Severity of OSA: Measured by AHI (Apnea-Hypopnea Index) and/or RDI (Respiratory Disturbance Index) from PSG.
      • Clinical Outcomes: Reduction in AHI/RDI, resolution of snoring, reported complications (e.g., wound seroma).
      • Medical opinion/diagnosis by the treating physicians and authors of the peer-reviewed studies.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI model requiring a training set in the machine learning sense. The "training" in a broad sense would be the cumulative clinical experience with similar devices and procedures and the engineering design and testing cycles, not a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for a physical device in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121440

    Validate with FDA (Live)

    Date Cleared
    2012-12-06

    (205 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siesta Medical, Inc. ENCORE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The ENCORE Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The ENCORE Tongue Suspension System consists of an integrated suture passer pre-loaded with size #2-0 braided polyester suture, a titanium bone screw that is premounted on an inserter, a suspension line lock tool, and a Threading Tool. In addition, the following suspension lines are provided depending on the model number: 1) a size #1 monofilament polypropylene suspension line, 2) a size #1 monofilament polypropylene suspension line with a radiopaque marker, 3) a size #2 monofilament polypropylene suspension line with a radiopaque marker, 4) a size #2 braided polyester suspension line, and 5) a size #2 braided polyester suspension line with a radiopaque marker.

    AI/ML Overview

    This document describes the 510(k) submission for the Siesta Medical, Inc. ENCORE™ Tongue Suspension System, emphasizing its substantial equivalence to a predicate device. The information provided focuses on non-clinical testing and explicitly states that no clinical testing was performed for this submission. Therefore, many of the requested details related to clinical studies, human reader performance, and ground truth establishment are not available in this text.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this submission pertains to a modification of an existing device and relies on showing substantial equivalence through non-clinical testing, specific "acceptance criteria" for clinical performance metrics (like sensitivity, specificity, or improvement in human reader performance) are not detailed. Instead, the focus is on performance specifications met through bench testing.

    Test NamePerformance Demonstrated / Reported ResultAcceptance Criteria (Implicit)
    Suture Endurance TestMet performance specificationsSuture should withstand expected forces and duration without failure; details of exact criteria are not provided.
    Bone Screw Fixation Strength TestMet performance specificationsBone screw should maintain adequate fixation in bone; details of exact criteria are not provided.
    Radiographic Visibility of Suspension Line MarkerMet performance specificationsRadiopaque marker should be sufficiently visible under radiography; details of exact criteria are not provided.
    ENCORE System Removability in Chronic Porcine ModelMet performance specificationsThe system should be removable in a chronic setting; details of exact criteria are not provided.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of clinical performance evaluation, as no human subject clinical testing was performed. The non-clinical tests would have used specific sample sizes for their materials/components, but these are not disclosed in this summary.
    • Data Provenance: The non-clinical testing appears to be conducted by Siesta Medical, Inc. (manufacturer). No country of origin is specified for the data, but it's implied to be from the manufacturer's testing. All data is retrospective in the sense that it's from completed tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. No ground truth based on expert consensus was established as no clinical study involving human assessment was conducted.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This device is not an AI-assisted diagnostic tool, and no MRMC study was conducted. The device is a surgical implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a surgical implant, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests would be the measurement of physical properties (e.g., tensile strength, fixation force, radiographic density) against predefined engineering specifications. This is distinct from clinical ground truth like pathology, expert consensus, or outcomes data, which are not relevant here.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set in the context of this device's evaluation, as it is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121814

    Validate with FDA (Live)

    Date Cleared
    2012-11-07

    (140 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siesta Medical, Inc. ENCORE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The ENCORE Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible using a bone screw and suture. The ENCORE Tongue Suspension System consists of an integrated suture passer pre-loaded with size #2-0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter, a bone screw lock tool, and a Threading Tool. In addition, the following suspension lines are provided depending on the model number: 1) a size #1 monofilament polypropylene suspension line, 2) a size #1 monofilament polypropylene suspension line with a radiopaque marker, 3) a size #2 monofilament polypropylene suspension line with a radiopaque marker, and 4) a size #2 braided polyester suspension line.

    Similar to the predicate ENCORE System, the modified ENCORE Tongue Suspension System 1) uses a bone screw, 2) a bone screw inserter to place a bone screw into the bone, and 3) a needle to pass a suspension suture submucosally through the base of the tongue. The ENCORE uses a Suture Passer and working suture loops to pass a suture from the posterior to anterior position in the tongue. The ENCORE advances the tongue in the anterior direction with a suspension line. A Threading Tool is provided to ease passage of the suspension line through the bone screw.

    AI/ML Overview

    The provided text describes a medical device submission (K121814) for the Siesta Medical, Inc. ENCORE Tongue Suspension System. This submission is for a modification to an existing device (predicate device K111179).

    The document is a 510(k) summary, which is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. This type of submission does not typically include detailed studies proving specific performance criteria and clinical efficacy in the same way a PMA (Premarket Approval) would. Instead, it focuses on demonstrating equivalence to a device that has already been cleared.

    Therefore, much of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment is not available in this 510(k) summary. The summary explicitly states: "This submission supports the position that the Siesta Medical, Inc. modified ENCORE Tongue Suspension System is substantially equivalent to the predicate ENCORE Tongue Suspension System (K111179). The 510(k) includes a test report to support extension of the shelf-life to 2 years."

    Here's what can be inferred or explicitly stated from the provided text, while also noting what is absent:


    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated as quantifiable performance metrics for the modified device. The primary "acceptance criteria" for a 510(k) is demonstrating substantial equivalence to a predicate device in terms of intended use, operating principle, technology, and manufacturing process.
    • Reported Device Performance: The document only mentions a "test report to support extension of the shelf-life to 2 years." This is a specific performance claim related to durability/stability, but no other performance metrics (e.g., success rate of tongue advancement, reduction in AHI, or adverse event rates) are provided for the modified device in this summary.
    Acceptance Criteria (Implied for 510(k)Reported Device Performance
    Substantial Equivalence to PredicateStated to be substantially equivalent in intended use, operating principle, technology, and manufacturing process.
    Shelf-lifeExtended to 2 years (supported by a test report).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified. The document only refers to "a test report to support extension of the shelf-life." This likely refers to bench testing, not a clinical test set on patients.
    • Data Provenance: Not specified. Given it's a shelf-life test, it would typically be laboratory-based rather than patient-based.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not specified. This information is typically relevant for studies involving subjective human assessment (e.g., image interpretation), which is not the type of data presented in this 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not specified. Adjudication methods are typically for clinical trials or diagnostic accuracy studies where multiple observers or subjective assessments are involved.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a surgical implant for tongue suspension, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical medical device, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the shelf-life extension, the "ground truth" would likely be objective measurements of material degradation or functional performance over time, according to established engineering and material science standards. No clinical ground truth is discussed for the modified device itself.

    8. The sample size for the training set

    • Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not a machine learning or AI device.

    Summary of what the K121814 submission does provide:

    • It states the device's intended use: "for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring."
    • It claims substantial equivalence to the predicate device (K111179) in terms of intended use, operating principle, technology, and manufacturing process.
    • It specifically mentions "a test report to support extension of the shelf-life to 2 years" as a new piece of supporting data for the modification.
    • It provides a detailed description of the device components and how they function, highlighting similarities to the predicate.

    This 510(k) notification is primarily a regulatory filing to demonstrate that a modified device is as safe and effective as a previously cleared device, rather than a clinical study establishing new performance metrics for an entirely novel device. Therefore, the detailed study information sought is typically not part of these submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111179

    Validate with FDA (Live)

    Date Cleared
    2011-07-01

    (65 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siesta Medical, Inc. ENCORE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The ENCORE Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The ENCORE Tongue Suspension System consists of an integrated suture passer pre-loaded with size #2-0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter, a bone screw lock tool. and a Threading Tool. In addition, one of the following suspension sutures are provided depending on the model number: 1) a size #1 monofilament polypropylene suspension suture, 2) a size #1 monofilament polypropylene suspension suture with a radiopaque marker, 3) a size #2 monofilament polypropylene suspension suture with a radiopaque marker, or 4) a size #2 braided polyester suspension suture.

    Similar to the PRELUDE III System, the ENCORE Tongue Suspension System 1) uses a bone screw, 2) uses a bone screw inserter to place a bone screw into the bone, and 3) uses a needle to pass a suspension suture submucosally through the base of the tongue. The ENCORE uses a Suture Passer and working suture loops to pass a suture from the posterior position in the tongue. The ENCORE advances the tongue in the anterior direction with a suspension suture. A Threading Tool is provided to ease passage of the suspension suture through the bone screw.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Siesta Medical, Inc. ENCORE Tongue Suspension System. The submission primarily focuses on demonstrating substantial equivalence to a predicate device (PRELUDE III Tongue Suspension System) through performance testing, rather than a clinical study evaluating diagnostic or treatment effectiveness in humans.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Suture Endurance"the suture was able to endure 10.5 million cycles"
    Bone Screw to Suture Attachment Strengths"the bone screw to suture attachment strengths were no worse than the written nominal ratings" and "the performance characteristics of the ENCORE Tongue Suspension System compare favorably to the predicate device." (Specific quantitative acceptance criteria for "nominal ratings" are not provided in the excerpt, but the favorable comparison to the predicate device implies meeting equivalence standards).

    2. Sample size used for the test set and the data provenance

    The document describes device performance testing (Suture Endurance Test and Bone Screw Fixation Strength Test), not a clinical study on human subjects or a diagnostic algorithm. Therefore, "test set" in the context of clinical data or AI evaluation is not directly applicable. The "samples" would refer to the physical devices or components tested. The exact number of samples for each engineering test is not specified in the provided text, nor is their "country of origin" as they are manufactured devices. These were likely prospective engineering tests performed in a lab setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The study described is not a clinical study involving human patients or an AI device requiring expert-established ground truth. The "ground truth" here is determined by the physical properties and performance benchmarks of the medical device components themselves.

    4. Adjudication method for the test set

    This question is not applicable as it refers to a clinical or AI study with human interpretation, not a bench test for device performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not performed, nor is it applicable. This is a 510(k) submission for a physical medical device, not an AI or diagnostic tool that assists human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, this question is not applicable. The ENCORE Tongue Suspension System is a physical medical device, not a standalone algorithm.

    7. The type of ground truth used

    The "ground truth" for the device performance tests was based on engineering specifications and established performance benchmarks derived from the predicate device and potentially industry standards for medical device materials and components. For instance:

    • Suture Endurance: The acceptance criterion was enduring 10.5 million cycles, which represents a physical performance threshold.
    • Bone Screw to Suture Attachment Strengths: The ground truth was described as "written nominal ratings" and "favorable comparison to the predicate device," implying a quantitative or comparative engineering standard.

    8. The sample size for the training set

    This question is not applicable. As previously stated, this is a 510(k) submission for a physical medical device, not an AI or machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable. There was no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110127

    Validate with FDA (Live)

    Date Cleared
    2011-04-08

    (80 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siesta Medical, Inc. PRELUDE III Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The PRELUDE III Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The PRELUDE III Tongue Suspension System consists of an integrated suture passer pre-loaded with size #2-0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter, a bone screw lock tool, and size 1 monofilament polypropelyne suspension suture.

    Similar to the PRELUDE II System, the PRELUDE III Tongue Suspension System 1) uses a bone screw, 2) uses a bone screw inserter to place a bone screw into the bone, and 3) uses a needle to pass a suspension suture submucosally through the base of the tongue. The PRELUDE II uses a rigid curved needle to make an intramuscular suture pass from the oral cavity while the PRELUDE III uses a flexible needle to make an intramuscular suture pass without entering the oral cavity. The PRELUDE III uses a Suture Passer and suture loops to pass a polypropylene suture from the posterior to anterior position in the tongue. The PRELUDE III advances the tongue in the anterior direction with #1 polypropylene suture, which is passed through the base of the tongue and attached to the bone screw.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for a medical device (PRELUDE III™ Tongue Suspension System), which is a regulatory submission to demonstrate that a device is substantially equivalent to a predicate device. This type of submission focuses on comparing the new device's characteristics and performance to an already legally marketed device, rather than proving efficacy or creating a standalone AI algorithm. Therefore, many of the questions regarding AI-specific criteria (like training set size, ground truth for training, MRMC studies, or standalone algorithm performance) are not applicable to this document.

    However, based on the provided text, we can describe the acceptance criteria and the study that proves the device meets the acceptance criteria within the context of a 510(k) submission for a physical medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission, the "acceptance criteria" are typically defined by demonstrating substantial equivalence to a predicate device, meaning the new device is as safe and effective as the predicate. This is often shown through comparisons of technological characteristics and performance testing. In this case, the performance criteria were related to the mechanical properties of the device components.

    Acceptance Criteria (Performance)Reported Device Performance (PRELUDE III™)
    Suture Pass ReliabilityDemonstrated, with at least 95% confidence and 90% reliability, that the suture is able to be passed from one Suture Passer to the other.
    Suture Loop Pull-Out Force (vs. Predicate)Showed no significant difference between the average pull-out force for the suspension suture loops made with the PRELUDE II (predicate) and PRELUDE III devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the numerical sample sizes for the "Suture Pass Reliability Test" or "Suture Loop Pull-Out Force" studies. It mentions statistical confidence and reliability for the suture pass test but not the number of units tested.
    • Data Provenance: The studies are in vitro (laboratory) studies conducted by Siesta Medical, Inc. The data provenance is internal to the manufacturer. The document doesn't specify country of origin for the data, but the company is based in Los Gatos, CA, USA. The studies are prospective in the sense that they were designed and executed to evaluate the new device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable. For a physical device demonstrating mechanical performance, "ground truth" is established by direct measurement and engineering specifications, not expert consensus as would be the case for an AI diagnostic algorithm.

    4. Adjudication Method for the Test Set

    This question is not applicable as there's no human interpretation or "adjudication" in the context of these in vitro mechanical tests. The results are quantitative measurements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    This question is not applicable. MRMC studies are specific to evaluating diagnostic accuracy and reader performance, typically for imaging devices or AI algorithms. This submission is for a surgical implantable device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The PRELUDE III is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    For the in vitro studies, the "ground truth" was based on:

    • Engineering specifications and direct measurement: For Suture Pass Reliability, the "truth" was whether the suture successfully passed as designed.
    • Comparative mechanical strength: For Suture Loop Pull-Out Force, the "truth" was the measured force required for pull-out, compared directly against the predicate device's performance.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as there is no "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K103179

    Validate with FDA (Live)

    Date Cleared
    2010-12-20

    (53 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siesta Medical, Inc. PRELUDE II Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The PRELUDE II Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The PRELUDE II Tongue Suspension System consists of a pair of suture passers pre-loaded with size 2.0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter, a bone screw lock tool, and size 1 monofilament polypropelyne suspension suture.

    AI/ML Overview

    The provided document is a 510(k) summary for the Siesta Medical, Inc. PRELUDE II Tongue Suspension System. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, this document does not contain information about acceptance criteria or specific study results that would prove the device meets such criteria in terms of clinical performance or effectiveness.

    The "studies" mentioned are in vitro tests conducted to evaluate the performance characteristics of the PRELUDE II Tongue Suspension System compared to a predicate device, focusing on mechanical properties rather than clinical outcomes. These are:

    • Suture Endurance Test
    • Bone Screw Torque Strength Test
    • Bone Screw Fixation Strength Test

    Therefore, I cannot provide a table of acceptance criteria for clinical performance or details of a study proving the device meets those criteria, as this information is not present in the provided text.

    Based on the provided text, I can only state the information that is present:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions three in vitro tests and states: "The data presented demonstrate that the performance characteristics of the PRELUDE II Tongue Suspension System compare favorably to the predicate device." No specific quantitative acceptance criteria or detailed performance data from these tests are provided in this summary.

    Test TypeAcceptance Criteria (Not explicitly stated in document)Reported Device Performance (Summary)
    Suture Endurance Test(Not specified in document)Performance characteristics compare favorably to predicate.
    Bone Screw Torque Strength Test(Not specified in document)Performance characteristics compare favorably to predicate.
    Bone Screw Fixation Strength Test(Not specified in document)Performance characteristics compare favorably to predicate.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for the in vitro tests.
    • Data Provenance: The in vitro studies were conducted as part of the 510(k) submission process, likely by the manufacturer (Siesta Medical, Inc.). No country of origin is specified. The studies are described as in vitro, meaning they were laboratory tests, not human or animal studies, and thus are not retrospective or prospective in a clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the studies described are in vitro mechanical tests, not clinical studies requiring expert ground truth for diagnosis or outcome assessment.

    4. Adjudication method for the test set:

    Not applicable, as the described studies are in vitro mechanical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical implant (a tongue suspension system), not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is a medical implant, not an algorithm.

    7. The type of ground truth used:

    For the in vitro tests, "ground truth" would refer to established engineering and material science standards for evaluating suture endurance, screw torque strength, and fixation strength. The document only states that the performance "compare favorably to the predicate device," implying a comparison against an already approved device's characteristics, which serves as a benchmark rather than a "ground truth" in the clinical sense.

    8. The sample size for the training set:

    Not applicable. This document describes in vitro mechanical tests, not machine learning or AI algorithm development which would have training sets.

    9. How the ground truth for the training set was established:

    Not applicable, for the same reasons as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101060

    Validate with FDA (Live)

    Date Cleared
    2010-07-08

    (84 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siesta Medical, Inc. PRELUDE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

    Device Description

    The PRELUDE Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The PRELUDE Tongue Suspension System consists of a pair of suture passers pre-loaded with size 2.0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter and size 1 monofilament polypropelyne suspension suture.

    AI/ML Overview

    The Siesta Medical, Inc. PRELUDE Tongue Suspension System is a medical device designed for anterior tongue base suspension to treat mild or moderate obstructive sleep apnea (OSA) and/or snoring. The 510(k) submission (K101060) describes the device and the studies conducted to demonstrate its substantial equivalence to predicate devices, focusing on its performance characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document summarizes the performance tests conducted but does not explicitly define "acceptance criteria" as quantifiable thresholds. Instead, it states that the performance compares favorably to predicate devices. The reported device performance is qualitative, indicating successful completion of the tests.

    Acceptance Criteria (Implicit)Reported Device Performance
    Suture tensile strength meets or exceeds predicate devices.Post-Sterilization Suture Tensile Strength Test: Data presented demonstrates favorable comparison to predicate devices.
    Suture endurance meets or exceeds predicate devices.Suture Endurance Test: Data presented demonstrates favorable comparison to predicate devices.
    Bone screw insertion torque is appropriate for clinical use and comparable to predicate devices.Bone Screw Insertion Torque Test: Data presented demonstrates favorable comparison to predicate devices.
    Bone screw torque strength is sufficient for intended use and comparable to predicate devices.Bone Screw Torque Strength Test: Data presented demonstrates favorable comparison to predicate devices.
    Bone screw fixation strength is sufficient for intended use and comparable to predicate devices.Bone Screw Fixation Strength Test: Data presented demonstrates favorable comparison to predicate devices.
    Device components are supplied sterile.The single patient use components of the PRELUDE Tongue Suspension System are provided sterile.

    2. Sample Size Used for the Test Set and Data Provenance:

    The provided 510(k) summary does not specify the sample sizes used for each of the in vitro mechanical tests. It only states that the tests were "conducted to evaluate the performance characteristics." The data provenance for these in vitro tests would typically be from laboratory testing performed by or for the manufacturer, Siesta Medical, Inc. The document does not specify country of origin for the data, but since the company is based in Los Gatos, CA, it's likely the testing was conducted in the USA or by labs commissioned by them. The study is an in-vitro performance evaluation, not a clinical study, so the terms "retrospective" or "prospective" do not directly apply in the same way they would to human subject research.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the studies described are in vitro mechanical performance tests. They do not involve human interpretation or subjective assessment that would require expert ground truth establishment in the way an AI diagnostic or image analysis system would. The "ground truth" for these tests is based on objective physical measurements and engineering principles.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as point 3. Adjudication methods like "2+1" or "3+1" are used in clinical studies or expert review processes, typically for establishing consensus ground truth for subjective data (e.g., image interpretation). The mechanical tests performed for this device rely on objective measurements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The PRELUDE Tongue Suspension System is a physical medical device (an implantable system), not an AI or imaging diagnostic tool. Therefore, an MRMC study related to human reader improvement with AI assistance is entirely irrelevant to this device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is an implantable physical system, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the in vitro mechanical tests, the "ground truth" is established through objective physical measurements using calibrated testing equipment and adherence to standard engineering test methodologies (e.g., tensile strength, torque, fixation strength). It is based on the inherent physical properties and performance limits of the materials and design, not on expert consensus, pathology, or clinical outcomes data in the traditional sense.

    8. The sample size for the training set:

    This information is not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. The design and manufacturing processes are informed by engineering principles, material science, and preclinical testing, rather than data-driven machine learning models.

    9. How the ground truth for the training set was established:

    This information is not applicable for the same reasons as point 8.