    K Number
    K100995

    Validate with FDA (Live)

    Date Cleared
    2010-06-25

    (77 days)

    Product Code
    Regulation Number
    868.5800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To allow airflow over the vocal cords for speaking function. To be used with a standard 15 mm connection on tracheostomy tubes.

    Device Description

    The Shikani speaking valve consists of a plastic housing with a captured ball that acts as a check valve. Once attached to the tracheostomy tube connector the speaking valve allows the patient to inhale through the tracheostomy tube and exhale across the vocal cords which facilitates speech or phonation.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Shikani Speaking Valve

    This report details the acceptance criteria and the study (or, in this case, comparison to a predicate device) that demonstrates the Shikani Speaking Valve meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission for a device that is essentially identical to a previously cleared predicate and is considered a line extension, the "acceptance criteria" are primarily established by the performance of the predicate device. The study performed is a demonstration of substantial equivalence based on identical design and performance parameters.

    Acceptance Criteria (based on Predicate)Reported Device Performance (Shikani Speaking Valve)Comment
    Intended Use: Allow airflow over vocal cords for speaking function.Allows airflow over vocal cords for speaking function.Identical intended use.
    Indications for Use: Allow airflow... with 15mm connection.Allows airflow... with a standard 15 mm connection.Identical indications for use, reflecting the predicate device's later line extension with 15mm connection.
    Environment of Use: Hospital, Sub-acute, Home.Hospital, Sub-acute Institutions, Home.Identical.
    Patient Population: Tracheostomy patients.Tracheostomy patients.Identical.
    Contraindications: Not for use in-line with a ventilator.Not for use with inflated tracheostomy tube cuff; Not for use with unconscious patients; Not for use with patients unable to exhale around the tube/cuff.The proposed device includes additional contraindications considered appropriate for speaking valves in general, not due to new risk analysis of the device itself.
    Standard 22/15mm Connections (per ISO 5356-1)Yes - standard fitting to connect to breathing circuit and 15mm tracheostomy tube connector.Identical. The device is confirmed to conform to ISO 5356-1. The manufacturer states "Conformance was not tested as the device is unchanged, but this standard is referenced." This implies previous testing or inherent design conformity.
    Weight: 32 gm32 gmIdentical.
    Valve type: BallBallIdentical.
    Attachment to tube: Attachment clipAttachment clipIdentical.
    Pressure drop: 1.3 cm H2O @ 30 lpm; 2.0 cm H₂O @ 40 lpm1.3 cm H2O @ 30 lpm; 2.0 cm H₂O @ 40 lpmIdentical.
    Materials: Housing - Celcon C251-A Acetyl Copolymer; Ball - Nylon 6/6 Grade 2; Pin - 316 Stainless steelHousing - Celcon C251-A Acetyl Copolymer; Ball - Nylon 6/6 Grade 2; Pin - 316 Stainless steelIdentical.
    Biocompatibility: Passed ISO 10993 (for materials)Materials passed ISO 10993 (as per predicate testing).The proposed device uses identical materials, thus existing biocompatibility testing for the predicate applies.

    2. Sample Size Used for the Test Set and Data Provenance

    This submission is for a device explicitly stated to be "the same device" as the legally marketed predicate (Shikani-French 15, which was a line extension of K982128). Therefore, there isn't a new "test set" in the traditional sense for evaluating AI/algorithm performance. The "test" is an assertion of identicality to the predicate device, which itself had performance testing.

    • Sample Size for Test Set: Not applicable as a distinct new test set for the "proposed device" was not used. The performance claims rely on the predicate device's testing.
    • Data Provenance: The original performance data for the predicate device (K982128) would have been generated during its initial clearance, likely from laboratory testing. Specific details on the country of origin or whether it was retrospective/prospective are not provided for the original K982128 submission in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. This device is a mechanical speaking valve, not an AI/imaging device requiring expert interpretation for ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. This device is a mechanical speaking valve, not an AI/imaging device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a mechanical device, not an AI algorithm assisting human readers.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This is a mechanical device, not an algorithm.

    7. Type of Ground Truth Used

    For the original predicate device (K982128), and by extension for this identical device, the "ground truth" for performance was based on physical measurements such as:

    • Resistance to flow (pressure drop)
    • Weight
    • Material composition
    • Biocompatibility test results (laboratory assays)

    These are objective, quantifiable physical properties rather than expert consensus, pathology, or outcomes data typically associated with diagnostic or AI devices.

    8. Sample Size for the Training Set

    Not applicable. This is a mechanical device, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is a mechanical device, not an AI algorithm.