510(k) Data Aggregation

    K Number
    K131341

    Validate with FDA (Live)

    Date Cleared
    2014-02-06

    (273 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Yellow Synthetic Vinyl Patient Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Yellow Synthetic Vinyl Patient Examination Gloves that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).

    AI/ML Overview

    Please find the acceptance criteria and study details based on the provided text. Due to the nature of the device (patient examination gloves), the "study" is a series of performance tests against established standards, rather than a clinical study involving human readers or AI.

    Device: Powder Free Yellow Synthetic Vinyl Patient Examination Gloves (K131341)

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is gauged against established ASTM standards for patient examination gloves, primarily ASTM D5250-06 (Reaffirmation 2011), ASTM D5151-06 (Reaffirmation 2011), and ASTM D6124-06 (Reaffirmation 2011). The acceptance criteria are the minimum requirements specified by these standards.

    Feature & DescriptionAcceptance Criteria (from ASTM Standard)Reported Device Performance (Subject Device)Result of Comparison
    Intended UseDisposable device for medical purposes, worn to prevent contamination between patient and examiner.Disposable device for medical purposes, worn to prevent contamination between patient and examiner.Substantially equivalent
    Length≥230mm min. (ASTM D5250-06)230mm min for all sizesSubstantially equivalent
    WidthMeets ASTM D5250-06 specifications: Small 80-90 mm, Medium 90-100mm, Large 100-110mm, X large 110-120 mmSmall 80-85 mm, Medium 95-97 mm, Large 102-108mm, X large 114-118 mmSubstantially equivalent
    ThicknessFinger 0.05mm min., Palm 0.08mm min. (ASTM D5250-06)Finger 0.05mm min., Palm 0.08mm min.Substantially equivalent
    Physical Properties (Before aging/after aging)Elongation ≥300%, Tensile Strength ≥14MPa (ASTM D5250-06)Elongation ≥300%, Tensile Strength ≥ 14MPaSubstantially equivalent
    Freedom from PinholesMeets 21 CFR 800.20, ASTM D5250-06, ASTM D5151-06 (Inspection Level I, AQL 2.5)Meets ASTM D5151 (Holes Inspection Level I, AQL 2.5)Substantially equivalent
    Residual PowderMeets ASTM D6124-06 (≤2mg of residual powder, for powder-free classification)Results generated values below 2mg of residual powderSubstantially equivalent
    Materials UsedPVCPVCSubstantially equivalent
    Dusting/Donning PowderPU (as surface coating agent)PU (as Surface Coating Agent)Substantially equivalent
    Performance Data Supporting Substantial EquivalenceMeets ASTM D5151-06, ASTM D5250-06, ASTM D6124-06Meets ASTM D5151-06, ASTM D5250-06, ASTM D6124-06Substantially equivalent
    Single Patient UseSingle Patient UseSingle Patient UseSubstantially equivalent
    Biocompatibility(Expected to be non-irritant/non-sensitizer per ISO 10993-10)SKIN IRRITATION DERMAL and SENSITIZATION STUDIES Meets ISO 10993-10. The test article was a non-irritant or non-sensitizer.Substantially equivalent

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each physical and performance test (e.g., number of gloves tested for tensile strength, pinholes, etc.). However, it indicates that the tests were conducted in accordance with ASTM standards. These standards typically specify the sampling plans and test methods.

    Data Provenance: The tests are implied to be conducted by the manufacturer, Shijiazhuang Winful Plastic Co., Ltd. The data is retrospective, as it pertains to measurements taken on the manufactured gloves to demonstrate compliance with standards for the 510(k) submission. The company's address is listed as: No.6 Cangshi Road, Jinzhou City, Hebei, 052260, China. Therefore, the data originates from China.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable as the device is a medical glove, not an AI/imaging device requiring expert interpretation for ground truth establishment. The "ground truth" for the performance characteristics of the gloves is established by the specified ASTM standards and the validated methodologies for performing the tests detailed within those standards.

    4. Adjudication Method for the Test Set

    This is not applicable. The "adjudication method" concept is relevant for studies involving subjective human interpretation or AI outputs, where multiple experts might disagree on a "ground truth" label. For this device, the tests involve objective measurements against quantitative physical and chemical specifications from established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This is not applicable. An MRMC study is designed to evaluate the diagnostic performance of a system (often AI-assisted) compared to human readers. This device is a physical product (medical glove) and does not involve diagnostic interpretation or human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. This concept applies to AI algorithms. The device is a physical product.

    7. The Type of Ground Truth Used

    The ground truth used for proving the device meets acceptance criteria is compliance with established national and international consensus standards:

    • ASTM D5250-06 (Reaffirmation 2011) - Standard Specification for Poly(vinyl chloride) Patient Examination Gloves
    • ASTM D5151-06 (Reaffirmation 2011) - Standard Test Method for Detection of Holes in Medical Gloves
    • ASTM D6124-06 (Reaffirmation 2011) - Standard Test Method for Residual Powder on Medical Gloves
    • 21 CFR 800.20 - General provisions regarding medical devices. (specifically for waterleak test on pinhole AQL)
    • ISO 10993-10: 2002/Amd. 1:2006(E) - Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (for biocompatibility).

    These standards define the acceptable physical properties, performance, and safety requirements for patient examination gloves.

    8. The Sample Size for the Training Set

    This is not applicable. This device is a physical product, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    This is not applicable. As there is no training set for a physical product, there is no corresponding ground truth establishment process. The "ground truth" for the product's design and manufacturing quality is derived from its adherence to the aforementioned ASTM and ISO standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043475

    Validate with FDA (Live)

    Date Cleared
    2005-02-01

    (47 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00c4.

    AI/ML Overview

    Requested information about the acceptance criteria and study that proves the device meets the acceptance criteria:

    1. Table of acceptance criteria and the reported device performance:
    CharacteristicsAcceptance Criteria (Standard)Device Performance
    DimensionASTM standard D 5250-00e4Meets
    Physical PropertiesASTM standard D 5250-00e4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-00e4 & D6124-01<2mg/glove
    Biocompatibility:
    - Primary Skin Irritation(ISO10993-10 mentioned implicitly)Passes (Not a Primary Skin Irritation, tested in rabbits)
    - Dermal Sensitization(ISO10993-10 mentioned implicitly)Passes (Not a Dermal sensitization, tested in the guinea pig)
    1. Sample size used for the test set and the data provenance: Not applicable. This document pertains to patient examination gloves, which undergo non-clinical testing rather than classification with a "test set" in the context of AI/ML or medical imaging devices. The tests are performed on samples of the manufactured gloves. The document does not specify exact sample sizes for each test but refers to standard testing methods. The data provenance is derived from direct testing of the manufactured gloves.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For medical device manufacturing like gloves, "ground truth" is established by adherence to specified physical, chemical, and biological standards (e.g., ASTM, CFR). No human experts are used to subjectively determine a "ground truth" for the test set in the way they would for diagnostic imaging.

    3. Adjudication method for the test set: Not applicable. Since the evaluation relies on objective measurements against predefined standards, no subjective adjudication method (like 2+1 or 3+1 consensus) is required.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a patient examination glove, not an AI or diagnostic imaging device. Therefore, MRMC studies and AI assistance are irrelevant.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm or AI device.

    6. The type of ground truth used: The ground truth is defined by objective, quantitative, and qualitative criteria established in international and national standards (ASTM standard D 5250-00e4, 21 CFR 800.20, ASTM D6124-01, ISO10993-10). These standards specify acceptable ranges for dimensions, physical properties, limits for pinholes, maximum powder residual, and biological response criteria for biocompatibility tests.

    7. The sample size for the training set: Not applicable. There is no "training set" in the context of manufacturing and testing patient examination gloves.

    8. How the ground truth for the training set was established: Not applicable. There is no "training set" for this type of device. The established standards are the "ground truth" against which all manufactured gloves are evaluated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043477

    Validate with FDA (Live)

    Date Cleared
    2005-01-07

    (22 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00c4

    AI/ML Overview

    The provided text describes the regulatory submission for Powdered Vinyl Patient Examination Gloves. It details the device's characteristics and its equivalence to a predicate device, primarily through adherence to established standards.

    Here's an analysis of the acceptance criteria and the study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 5250-00e4Meets
    Physical PropertiesASTM standard D 5250-00e4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder AmountASTM standard D 5250-00e4Meets
    BiocompatibilityPrimary Skin Irritation in rabbitsPasses, Not a Primary Skin Irritation (<10mg/dm2)
    Dermal sensitization in the guinea pigPasses, Not a Dermal sensitization

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" sample size in the context of typical AI/software performance studies. The testing described pertains to physical and biological properties of the gloves, and the sample sizes for these tests are not explicitly stated. The data provenance is internal testing by the manufacturer, SHIJIAZHUANG WINFUL PLASTIC CO.,LTD, located in China. The studies are by nature retrospective as they are conducted on manufactured products to ensure compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the device is a physical medical device (examination gloves), not an AI/software device requiring expert interpretation of data to establish ground truth for a test set. The "ground truth" for the device's performance is objective measurements against established ASTM and CFR standards, as performed by laboratory testing.

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as point 3. No adjudication method is described for determining the "ground truth" of the physical and biological characteristics of the gloves.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is an examination glove, not an AI/software device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is an examination glove, and there is no algorithm or human-in-the-loop involved in its function or evaluation.

    7. The type of ground truth used

    The ground truth used for evaluating the device is based on objective measurement against established regulatory and industry standards. Specifically:

    • ASTM standard D 5250-00e4 for dimensions, physical properties, and powder amount.
    • 21 CFR 800.20 for freedom from pinholes (waterleak test).
    • ISO 10993-10 for biocompatibility (Primary Skin Irritation and Dermal sensitization tests).

    8. The sample size for the training set

    This information is not applicable. There is no "training set" in the context of this device, as it is a physical product and not an AI/machine learning model.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as point 8.