Search Results
Found 2 results
510(k) Data Aggregation
(132 days)
Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular, transvaginal and trans-rectal applications in B, M, PWD and Color Doppler modes.
The Color Doppler Ultrasound Systems are integrated preprogrammed color ultrasound imaging system, capable of producing a resolution intended for clinical diagnostic imaging applications. Their basic function is to acquire ultrasound signal and display the images in the imaging operation modes of B, M, PWD and Color Doppler.
This document describes the Color Doppler Ultrasound System Models mEye1 and mEye2 (K182165) and its substantial equivalence determination to a predicate device. There is no study described within this document that performs a comparative effectiveness study with human readers or standalone algorithm performance, or establishes acceptance criteria based on diagnostic performance.
The acceptance criteria and study described are for non-clinical performance and safety standards, not for the diagnostic accuracy of the ultrasound system's output. The document explicitly states: "No clinical study is included in this submission." This means that no study was conducted to demonstrate the device's diagnostic performance for the specified indications for use, nor were acceptance criteria related to diagnostic performance established or met through a clinical study.
Therefore, many of the requested items cannot be answered given the provided text.
Here's a breakdown of what can be extracted:
1. A table of acceptance criteria and the reported device performance
The "acceptance criteria" here refer to conformance with established medical device safety and performance standards, rather than diagnostic accuracy metrics.
| Acceptance Criteria (Standard/Requirement) | Reported Device Performance |
|---|---|
| IEC 60601-1-2005+A1:2012 (Electrical Safety) | Complied |
| IEC 60601-1-2:2014 (EMC) | Complied |
| IEC 60601-2-37:2007 (Performance - Ultrasonic Medical Diagnostic) | Complied |
| ISO 10993-5:2009 (Biocompatibility - Cytotoxicity) | Complied |
| ISO 10993-10:2010 (Biocompatibility - Irritation & Sensitization) | Complied |
| NEMA UD 2-2004 (R2009) (Acoustic Output) | Complied |
| Software Level | Moderate level of concern |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable. No clinical test set or data provenance related to diagnostic performance is mentioned. The tests were non-clinical, likely performed in manufacturing facilities.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. No clinical test set or establishment of ground truth by experts for diagnostic performance is mentioned.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. No clinical test set requiring adjudication for diagnostic performance is mentioned.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader, multi-case comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." This device is a diagnostic ultrasound system, not an AI-powered CAD (Computer-Aided Detection) or CADx (Computer-Aided Diagnosis) system, so the concept of "human readers improve with AI vs without AI assistance" does not apply here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No, a standalone (algorithm only) performance study was not done. This is a medical imaging device, not an AI algorithm. Its performance is evaluated through compliance with technical and safety standards, and by demonstrating substantial equivalence to a predicate device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For the non-clinical tests conducted, the "ground truth" was established by the technical specifications and requirements of the referenced standards (e.g., whether the device successfully met electromagnetic compatibility limits, or whether biocompatibility tests showed no cytotoxicity). There was no clinical ground truth established for diagnostic accuracy.
8. The sample size for the training set
Not applicable. This document describes a medical imaging device, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. See point 8.
Ask a specific question about this device
(48 days)
WED-3100 Digital Ultrasound Scanner is intended for diagnostic ultrasound imaging analysis for abdomen, gynecology, obstetric, urology, small-parts.
The system is intended to use for the following type of studies: fetal, abdominal, pediatric, small organs, peripheral vascular, and musculo-skeletal (both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional. This device is not intended for home use.
This equipment is high resolution linear/convex ultrasound scanning diagnostic equipment. It adopts micro-computer control and digital scan converter (DSC), digital beam-forming (DBF), real time dynamic aperture (RDA), real time dynamic receiving apodization, real time dynamic receiving focusing (DRF), Digital frequency scan (DFS), frame correlation technologies to endue its image with clarity, stability and high resolution.
Here's an analysis of the provided text regarding the acceptance criteria and study for the KLOOO79 device (WED-3100 Digital Ultrasound Scanner).
It's important to note that this document is a 510(k) summary, which often provides a high-level overview. Detailed study protocols and results are typically found in the full 510(k) submission, which is not provided here. Therefore, some information requested might be absent or generalized.
Acceptance Criteria and Device Performance
The acceptance criteria for the device are generally implied by the technical specifications and the statement that "Laboratory testing was conducted to validate and verify that the proposed devices met all design specifications, including electrical safety, EMC, specifications." For specific performance criteria, we look at the 'Accuracy' and 'Resolution' values provided in the technical specifications.
Table of Acceptance Criteria and Reported Device Performance (Implied from Technical Specifications)
| Feature/Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Accuracy (Measurements) | Error ≤ ±10% | Error ≤ ±10% (Stated as met) |
| Resolution (C1-11 Probe) | Lateral ≤1 mm (Depth≤60mm), Axial ≤1 mm (Depth≤80mm) | Lateral ≤1 mm (Depth≤60mm), Axial ≤1 mm (Depth≤80mm) (Stated as met) |
| Resolution (L1-5 Probe) | Lateral ≤3 mm (depth≤80mm), ≤5 mm (80mm<depth≤130mm); Axial ≤1 mm (depth≤80mm) | Lateral ≤3 mm (depth≤80mm), ≤5 mm (80mm<depth≤130mm); Axial ≤1 mm (depth≤80mm) (Stated as met) |
| Dead Zone (C1-11 Probe) | ≤3 mm | ≤3 mm (Stated as met) |
| Dead Zone (L1-5 Probe) | ≤6 mm | ≤6 mm (Stated as met) |
| Detect Depth (C1-11 Probe) | ≥80 mm | ≥80 mm (Stated as met) |
| Detect Depth (L1-5 Probe) | ≥140 mm | ≥140 mm (Stated as met) |
| Geometric Position Precision (C1-11 Probe) | Horizontal ≤5 mm, Vertical ≤5 mm | Horizontal ≤5 mm, Vertical ≤5 mm (Stated as met) |
| Geometric Position Precision (L1-5 Probe) | Horizontal ≤7.5 mm, Vertical ≤5 mm | Horizontal ≤7.5 mm, Vertical ≤5 mm (Stated as met) |
| Electrical Safety | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37 | Met (Stated as tested and compliant) |
| EMC | Compliance with IEC 60601-1-2 | Met (Stated as tested and compliant) |
| Acoustic Output | Compliance with NEMA UD 2: 2004 | Met (Stated as tested and compliant) |
| Biocompatibility | Compliance with ISO 10993-5, ISO 10993-10, ISO 10993-12 | Met (Stated as tested and compliant) |
Study Details
The primary study mentioned to demonstrate effectiveness is a "Clinical Measurement Effectiveness Test." However, the details provided for this test are very limited.
-
Sample size used for the test set and the data provenance:
- Sample Size: Not specified.
- Data Provenance: Not specified. The manufacturer is based in China, but it's not explicitly stated where the clinical measurement effectiveness test was conducted or where data originated. The submission date is September 28, 2009, with the K number assigned on March 1, 2010; this doesn't indicate if the data was retrospective or prospective.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not specified. The document only states "Clinical Measurement Effectiveness Test was conducted." There is no mention of experts or their qualifications for establishing ground truth.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not specified.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC or human reader study is mentioned. This device is a digital ultrasound scanner, not an AI-assisted diagnostic tool.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The "Clinical Measurement Effectiveness Test" is likely a standalone performance assessment of the device's measurement capabilities. However, it's not an AI algorithm. It assesses the accuracy of measurements performed by the ultrasound system itself.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the "Clinical Measurement Effectiveness Test," the ground truth is implied to be highly accurate reference measurements against which the device's measurements (Distance, circumference, area, volume, heart rate, GA, FW, EDD) are compared. The common practice for such tests involves phantoms with known dimensions or established clinical methods for comparison, but the specific type is not detailed beyond "Effectiveness Test."
-
The sample size for the training set:
- Not applicable. This device is a diagnostic ultrasound scanner, not a machine learning or AI product that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable. (See #7).
Additional Safety and Effectiveness Studies Mentioned:
Beyond the "Clinical Measurement Effectiveness Test," the document highlights several other studies/tests crucial for demonstrating safety and equivalence:
- Electrical Safety Standards: Testing against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37.
- Acoustic Output: Testing against NEMA UD 2: 2004.
- Biocompatibility Standards: Testing against ISO 10993-5, ISO 10993-10, ISO 10993-12.
These studies demonstrate compliance with recognized industry standards related to the physical and biological safety of the device, rather than diagnostic effectiveness in a clinical human-in-the-loop context. The "Clinical Measurement Effectiveness Test" is the sole mention of a study to prove the device's diagnostic capability.
Ask a specific question about this device
Page 1 of 1