Found 11 results

510(k) Data Aggregation

    K Number
    K201468

    Validate with FDA (Live)

    Device Name
    Pulse Oximeter
    Date Cleared
    2021-01-25

    (237 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Pulse Oximeter is intended for measuring the functional oxygen saturation (SpO2) and pulse rate (PR). It is intended for spot check and continuous recording of SpO2, PR of adult or pediativ patients in hospitals, clinics, or home. This device is not intended for continuous monitoring.

    Device Description

    This AP-10 Pulse Oximeter is wrist-worn. It measures the SPO2 physiological parameter, and the measurement results are displayed and stored. The data can be transferred to a phone or PC via a USB connection or via Bluetooth. The AP-10 Pulse Oximeter measures SPO2 and Pulse Rate simultaneously. The device can store up to 500 hours of SPO2 and Pulse rate data.

    AI/ML Overview

    The Shenzhen Creative Industry Co., Ltd. Pulse Oximeter, AP-10, underwent a clinical study to establish its accuracy for measuring oxygen saturation (SpO2). The study was conducted in accordance with Annex EE of ISO 80601-2-61:2018 (or 2011, as listed in the provided text).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (ISO 80601-2-61)Reported Device Performance (AP-10)
    SpO2 Accuracy±3% (during 70%-100% SpO2)±3% (during 70%-100% SpO2)
    Pulse Rate Accuracy±2bpm or ±2% (whichever is greater)±2bpm or ±2% (whichever is greater)

    Note: The document explicitly states that the AP-10's reported performance for SpO2 and Pulse Rate accuracy meets the ISO 80601-2-61 standard's requirements, which are the acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that clinical testing was conducted. However, it does not explicitly mention the sample size for the test set or the country of origin. It refers to the "Guideline for evaluating and documenting SpO2 accuracy in human subjects of ISO 80601-2-61:2011." This guideline typically involves healthy volunteers studied under controlled hypoxia conditions. Therefore, the data would be prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The provided text does not contain information regarding the number of experts or their qualifications used to establish ground truth for the test set. For pulse oximetry, the "ground truth" for SpO2 is usually established via a co-oximeter measuring arterial blood gas (PaO2), overseen by medical professionals, but this detail is not present in the document.

    4. Adjudication Method

    The document does not specify an adjudication method. For pulse oximetry studies, adjudication typically refers to how reference SpO2 values are determined from blood samples, which doesn't usually involve multiple human readers in the same way imaging studies might.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or described. This type of study is typically used for diagnostic imaging devices where human readers interpret results, and the AI's impact on their performance is being evaluated. The AP-10 is a direct measurement device (pulse oximeter), so this type of study is not applicable.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance evaluation was conducted. The "Reported Device Performance" in the table above reflects the algorithm's accuracy in measuring SpO2 and pulse rate as a standalone device. The clinical study mentioned in "Clinical: Clinical testing was conducted per Annex EE Guideline for evaluating and documenting SpO2 accuracy in human subjects of ISO 80601-2-61:2011" specifically evaluates the device's intrinsic accuracy.

    7. Type of Ground Truth Used

    The type of ground truth used, though not explicitly stated as "co-oximetry," is implied by the reference to Annex EE of ISO 80601-2-61. This standard's Annex EE describes a method where arterial oxygen saturation (SaO2) measured by a co-oximeter from arterial blood samples is considered the ground truth against which the pulse oximeter's SpO2 readings are compared.

    8. Sample Size for the Training Set

    The document does not provide any information about the sample size for the training set. It refers to non-clinical and clinical testing for the current device's performance validation, but not to machine learning model training data. Given that pulse oximeters primarily rely on established biophysical principles (light absorption at different wavelengths by oxygenated and deoxygenated hemoglobin) rather than complex machine learning models directly trained on vast datasets in the same way an AI diagnostic tool might, the concept of a "training set" for the core physiological measurement might not be directly applicable or detailed in this type of submission.

    9. How the Ground Truth for the Training Set Was Established

    As with the previous point, the document does not provide information on how ground truth for a training set (if one was used for specific algorithm development within the device) was established. The focus of the submission is on the device's adherence to established performance standards for pulse oximetry.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192343

    Validate with FDA (Live)

    Device Name
    Patient Monitor
    Date Cleared
    2019-12-31

    (125 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Patient Monitor is a multi-functional instrument designed for monitoring the vital physiological signs of adult and pediatric (but not neonatal) patients. With the functions of real-time recording and displaying parameters, such as ECG, heart rate(HR), non-invasive blood pressure (NIBP), functional oxygen saturation (SpO2), respiration rate (RESP), body temperature (TEMP), end-tidal CO2 concentration (EtCO2), it allows comprehensive analysis of patient's physiological conditions.

    This instrument is applicable for use in hospitals and clinical institutions. The operation should be performed by qualified professionals only.

    Device Description

    This Patient Monitor is a multi-functional instrument designed for monitoring the vital physiological signs of adult and pediatric (not neonatal) patients. With the functions of real-time recording and displaying parameters, such as ECG, pulse rate (PR), noninvasive blood pressure (NIBP), functional oxygen saturation (SpOz), respiration rate (RESP), body temperature (TEMP), end-tidal CO2 concentration (EtCO2), it allows comprehensive analysis of patient's physiological conditions.

    This instrument is applicable for use in hospitals and clinical institutions. The operation should be performed by qualified professionals only.

    There are three versions of the K serial Patient Monitor, K10, K12 and K15. The primary difference is physical dimension and display TFT size; all other specifications remain unchanged. All versions have the same indications for use.

    AI/ML Overview

    This document, a 510(k) summary for the Shenzhen Creative Industry Co., Ltd. Patient Monitor, Models K10, K12, and K15, primarily focuses on demonstrating substantial equivalence to a predicate device (Shenzhen Creative Industry UP-7000 Patient Monitor, K123711) rather than detailing the specific acceptance criteria and study proving performance for a novel AI/software device.

    Therefore, much of the requested information regarding AI device performance (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, standalone performance, ground truth establishment for training) is not applicable or present in this document. This document describes a traditional medical device (patient monitor) and its components, and the "study" referred to is non-clinical and clinical testing to ensure compliance with relevant performance standards for vital sign monitoring.

    However, I can extract the information that is applicable based on the provided text, particularly focusing on the performance criteria for the integrated vital sign modules.

    Here's an analysis based on the provided text:

    Device: Patient Monitor, Models K10, K12, K15

    Indications for Use: Monitoring the vital physiological signs of adult and pediatric (but not neonatal) patients, including ECG, heart rate (HR), non-invasive blood pressure (NIBP), functional oxygen saturation (SpO2), respiration rate (RESP), body temperature (TEMP), and end-tidal CO2 concentration (EtCO2). Applicable for use in hospitals and clinical institutions, operation by qualified professionals only.

    Study Type: This is a 510(k) submission seeking substantial equivalence to a predicate device. The "studies" involve non-clinical (safety and performance) and clinical (NIBP validation) testing against recognized standards rather than a comparative effectiveness study of a novel AI algorithm's diagnostic performance.


    1. Table of acceptance criteria and the reported device performance:

    The document lists performance specifications for each physiological parameter module, often directly comparing them to the predicate device. The acceptance criteria are implicitly that the devices meet or are substantially equivalent to the established performance requirements of the predicate device and relevant industry standards.

    CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (Subject Device)Difference Discussion (Relevance)
    SpO2 AccuracyAdult & Pediatric: ±3% (70%-100%), Undefined (0-70%)Adult & Pediatric: ±3% (70%-100%), Undefined (0-70%)No change.
    Pulse Rate Accuracy (SpO2)±2 bpm or ±2% (whichever is greater)±2 bpm or ±2% (whichever is greater)No change.
    CO2 Accuracy (0-40 mmHg)±2 mmHg±2 mmHgNo change.
    CO2 Accuracy (41-70 mmHg)±5% of reading±5% of readingNo change.
    CO2 Accuracy (71-100 mmHg)±8% of reading±8% of readingNo change.
    CO2 Accuracy (101-150 mmHg)±10% of reading±10% of readingNo change.
    Respiration Rate Accuracy (CO2)±2 rpm±2 rpmNo change.
    Heart Rate Precision (ECG)±1% or ±2 bpm, whichever is greater±1% or ±2 bpm, whichever is greaterNo change.
    RESP Rate Accuracy±2% or ±2 rpm, whichever is greater±2% or ±2 rpm, whichever is greaterNo change.
    NIBP BP Accuracy (Mean Deviation)±5 mmHgMax. mean deviation values: ±5 mmHgIdentical.
    NIBP BP Accuracy (Standard Deviation)<= 8 mmHgMax. standard deviation: 8 mmHgIdentical.
    Temperature Measurement Accuracy±0.1C (35.0C-39.0C); ±0.2C (rest)±0.2C (25.0C-45.0C); ±0.4C (rest)Difference noted, but stated to "not raise any new questions of safety and effectiveness and still complies with the ISO80601-2-56 requirement."
    NIBP Overpressure Protector (Adult Cuff Pressure)Cuff pressure exceeds 300mmHgCuff pressure exceeds 300mmHgIdentical.
    NIBP Overpressure Protector (Pediatric Cuff Pressure)Cuff pressure exceeds 300mmHgCuff pressure exceeds 240mmHgSubject device is more stringent (lower max pressure).

    Note: The document explicitly states for the NIBP and TEMP modules that "The new [module] used in subject devices complies with the same ISO [standard] and design requirement as Predicate device." This implies the acceptance criterion for these components is compliance with the cited ISO standards (e.g., ISO 80601-2-30 for NIBP, ISO 80601-2-56 for TEMP).


    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size:
      • For NIBP (Non-invasive sphygmomanometers) clinical validation (ISO 81060-2), this standard typically requires a minimum of 85 subjects (adults and children combined) for a full validation. While the exact number isn't stated in the provided text, the reference to this specific standard implies adherence to its sample size requirements.
      • For other parameters (SpO2, CO2, ECG, RESP, TEMP), the "performance data" section refers to IEC/ISO standards (e.g., 60601-2-49, 80601-2-61, 80601-2-55, 60601-2-27, 80601-2-56). These standards define particular requirements for the performance of the devices and often involve bench testing, simulations, or phantom studies rather than large-scale human clinical trial "test sets" in the AI sense.
    • Data Provenance: Not specified in the provided text (e.g., country of origin). The testing is referred to as "Non-Clinical Study" and "Clinical Study" without detail on whether retrospective or prospective data was used, except for the NIBP clinical validation which would be prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable in the AI/Software Sense: This document describes a traditional patient monitor. The "ground truth" for a patient monitor's vital sign measurements is established against recognized reference measurement methods or calibrated instruments (e.g., a known SpO2 reading from a CO-oximeter, a calibrated pressure sensor for NIBP, or a physiological simulator for ECG).
    • For NIBP (ISO 81060-2): This standard for clinical validation of automated sphygmomanometers typically requires measurements to be taken by trained observers (not necessarily "experts" in the diagnostic sense, but highly skilled in manual auscultatory measurements) and compared against a reference standard. The number of observers is also specified in the standard (e.g., at least 3 observers taking sequential measurements). Their "qualification" would be their training and demonstrated proficiency in the reference method.

    4. Adjudication method for the test set:

    • Not Applicable in the AI/Software Sense: There is no "adjudication" of expert interpretations as seen in AI studies, as the device measures physiological parameters directly.
    • For NIBP (ISO 81060-2): The standard defines specific statistical methods for comparing the device's measurements to the reference measurements, often involving Bland-Altman analysis and calculating mean difference and standard deviation. There's no "adjudication" in the sense of resolving conflicting human interpretations; rather, it's a statistical comparison to a pre-defined reference.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable: This device is a patient monitor, not an AI diagnostic imaging tool or a system that assists human "readers" with interpretation. Therefore, no MRMC study was performed or is relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable in the AI/Software Sense: The described device functions as a standalone vital sign monitor. Its "performance" is inherently "algorithm only" in that its internal processing generates the vital sign values. However, this is not an "AI algorithm" in the context of complex pattern recognition or decision support. The performance is assessed by comparing its output against reference standards, which is a form of standalone evaluation for a monitor.

    7. The type of ground truth used:

    The ground truth for the performance evaluations of this patient monitor would be derived from:

    • Reference Measurements/Calibrated Standards:

      • Physiological Simulators: For ECG, RESP, and potentially SpO2 and NIBP, to generate known, controlled physiological signals.
      • Calibrated Instruments: For SpO2 (e.g., CO-oximeter), CO2 (calibrated gas mixtures), NIBP (auscultatory measurements by trained observers or calibrated pressure transducers), and Temperature (calibrated thermometers).
      • In-vitro/Bench Testing: As implied by adherence to IEC/ISO performance standards, which often dictate specific test setups and reference instruments.
    • Clinical Data (for NIBP): As per ISO 81060-2, the ground truth for NIBP accuracy relies on trained human observers taking simultaneous or sequential auscultatory blood pressure measurements (the "reference method").


    8. The sample size for the training set:

    • Not Applicable: This device is a traditional patient monitor, not an AI/machine learning device that requires a "training set" in the context of deep learning models. Its algorithms are based on established physiological signal processing principles.

    9. How the ground truth for the training set was established:

    • Not Applicable: As stated above, there is no "training set" for this type of device. The ground truth for its performance verification (as opposed to AI training) is established using validated, calibrated reference standards and methods as described in point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172792

    Validate with FDA (Live)

    Date Cleared
    2018-11-13

    (424 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Handheld Pulse Oximeter is intended for measuring the pulse rate, functional oxygen saturation (SpO2) and temperature (optional). It is intended for spot checking and recording of SpO2, pulse rate and temperature of adult and pediatric patients in clinical institutions and homes. The Handheld Pulse Oximeter is not intended for active continuous monitoring.

    Device Description

    This SP-20 Handheld Pulse Oximeter is an integrated product with a modular design. It simultaneously measures physiological parameters by integrated modules and corresponding accessories. The measurement results are displayed, stored and transferred by the central processing unit. The SP-20 Handheld Pulse links with the compatible 510(k) cleared IR Ear thermometer Device (option). Since the SP-20 receives data from the 510(k) cleared IR Ear thermometer and no data is transmitted to the thermometer, it does not affect the operations or accuracy of the linked IR Ear thermometer Device.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Shenzhen Creative Industry Co., Ltd. Handheld Pulse Oximeter, SP-20 based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text focuses on demonstrating substantial equivalence to predicate devices and adherence to various safety and performance standards rather than explicitly stating numerical acceptance criteria for SpO2 and Pulse Rate accuracy with corresponding numerical performance results. However, it does state compliance with the general requirements of the standards.

    Based on the information, the primary performance standard for the pulse oximeter function is:

    Acceptance Criteria (Standard & Description)Reported Device Performance
    IEC 60601-1:1990+A1+A2+A13: Medical devices Part1: General requirements for safetyComplied with standard's requirements (Electrical Safety)
    IEC 60601-1-2:2014: Medical electrical equipment - part 1-2: General requirements for basic safety and essential performance - Collateral standards: Electromagnetic compatibility -Requirements and testsComplied with standard's requirements (EMC)
    IEC 60601-1-11:Ed. 2.0, 2015: Medical electrical equipment-part 1-11: General requirements for basic safety and essential performance -collateral standard: Requirement for medical electrical equipment and medical electrical systems used in home healthcare environmentComplied with standard's requirements (Home Healthcare Electrical Safety)
    ISO80601-2-61_SpO2: Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipmentComplied with standard's requirements (General requirements: Alarm system and SpO2/PR performance)
    IEC/TR60878:2003: Graphical symbols for electrical equipment in medical practiceComplied with standard's requirements (Graphical symbols)
    ISO 10993-1:2009: Biological evaluation of medical devices--Part 1: Evaluation and testingComplied with standard's requirements (Biological evaluation)
    ISO 10993-5:2009: Biological Evaluation of Medical Devices - Part 5 Tests for In Vitro CytotoxicityComplied with standard's requirements (Biological Estimation)
    ISO 10993-10:2010: Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Delayed-Type HypersensitivityComplied with standard's requirements (Biological Estimation)
    Cleaning and Disinfection ValidationComplied with standards (details not given on specific criteria)
    Performance of SpO2 and TEMP through 510(k) cleared modules (normal operating states, appearance, structure, basic performance, system performance)Test results demonstrate compliance with the standards.

    Important Note for SpO2 and Pulse Rate: While ISO 80601-2-61 covers the particular requirements for pulse oximeter equipment, the document does not explicitly state the specific accuracy (e.g., Arms value) or bias results for SpO2 or Pulse Rate as acceptance criteria or reported performance results. It only states general compliance with the standard. Typically, pulse oximeter standards define specific accuracy requirements (e.g., Arms ≤ 3% for SpO2 between 70-100%). It's implied that the device meets these within the standard, but the specific numbers are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Twelve (12) subjects.
    • Data Provenance: The subjects consisted of volunteers (3 women and 9 men) with no known illnesses and no anemia. They represented a mixture of ethnic backgrounds with skin colors ranging from light to dark, and ages from 22 to 38. A total of 24 samples were obtained per subject.
    • Study Type: This appears to be a prospective clinical testing for performance validation. The location of the test is not explicitly stated as a country, but it says "Data was recorded by Bickler-Ye Lab and provided for analysis," which might hint at a US-based lab given Dr. Jeffrey Bickler's affiliation with UCSF.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish ground truth for the clinical testing in the traditional sense of consensus reading for diagnostic imaging. Instead, for pulse oximetry, the "ground truth" for oxygen saturation is typically established through direct arterial blood gas analysis, a recognized clinical standard. However, this document does not explicitly state how ground truth for SpO2 was established in the clinical test. It states "Testing validated the submitted device was effective in determining SPO2."

    4. Adjudication Method for the Test Set

    Not applicable in the context of this pulse oximeter performance testing. Adjudication methods like "2+1" or "3+1" are typically used in studies where human readers are interpreting images or data and their disagreements need to be resolved. For pulse oximetry, the comparison is usually against a physiological standard (e.g., co-oximetry).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study was not done. The study described is a clinical validation of the device's accuracy in measuring SpO2 and pulse rate, not a comparative effectiveness study involving human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the described "Clinical testing" focuses on the device's performance in measuring SpO2. This is a standalone performance test of the device (algorithm/sensor system) without human interpretation in the loop. The device directly provides numerical readings (SpO2, Pulse Rate).

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The document does not explicitly state the method for establishing ground truth for SpO2 in the clinical trial. For pulse oximeters, the accepted gold standard for SpO2 is typically fractional arterial oxygen saturation (SaO2) measured by a co-oximeter from an arterial blood sample. While the document mentions "Testing validated the submitted device was effective in determining SPO2," it doesn't detail the comparison method.

    8. The Sample Size for the Training Set

    No information about a training set is provided. This device is a pulse oximeter, not an AI/ML-driven diagnostic algorithm that typically requires a separate training set. Its principle of operation is based on optoelectronics and signal processing, which doesn't usually involve a "training phase" in the same way an AI model does.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a training set for an AI/ML algorithm is not relevant to this type of device based on the provided information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170047

    Validate with FDA (Live)

    Date Cleared
    2017-09-09

    (247 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The All-in-One Health Monitor PC-303 is a device designed for spot-checking measuring of the patient's physiological parameters, such as Non-Invasive Blood Pressure (NIBP), Oxygen saturation (SpO2), Pulse Rate (PR) and Body Temperature (TEMP);

    Additionally, the device is available to communicate with the compatible Blood Glucose Monitoring System and ECG monitor to make the measurement.

    This device is applicable for Adult and Pediatic (age>3 years old) use in clinical institutions and has no conditions or factors of contraindication.

    Device Description

    The All-in-One Health Monitor, PC-303 is mobile modular designed patient monitor. It monitors the patient's blood oxygen saturation (SpO2) and pulse rate (PR) non-invasively by the photoelectric method. It can also measure non-invasive blood pressure (NIBP, the pressures of systolic, diastolic and mean) by the oscillating method and body temperature (TEMP) by the infrared radiation energy technology.

    Additionally, the device is also capable to link and communicate with the compatible Blood Glucose Meter and ECG monitor to make the measurement.

    This device system measure the patient's physical parameters with variety modules. It takes digital signals from NIBP, SpO2, Temperature module through COM ports, and receives ECG data, blood glucose data from compatible legally marketed ECG Monitor or BLOOD GLUCOSE MONITORING Device. The software to process the data, then display parameters on the screen and storage some useful information.

    This device is applicable for use in clinical institutions and has no conditions or factors of contraindication.

    AI/ML Overview

    The provided document describes the All-in-One Health Monitor, PC-303, and its substantial equivalence to a predicate device (UP-7000 Patient Monitor, K123711). The acceptance criteria are primarily based on meeting relevant international standards for medical electrical equipment and specific performance characteristics in comparison to the predicate device.

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are not explicitly stated as distinct numerical targets in a single table within the document but are inferred from the comparison with the predicate device and the adherence to recognized international standards. The "reported device performance" is the claim of compliance with these standards and the equivalent performance to the predicate.

    Parameter/CharacteristicAcceptance Criteria (Implied by Predicate & Standards)Reported Device Performance (PC-303)
    General
    SafetyCompliance with IEC 60601-1"All the requirements of IEC 60601-1 were evaluated in this test report" (with specific exclusions)
    EMCCompliance with IEC 60601-1-2"Testing was performed to ascertain Electromagnetic Compatibility to other electrical devices... Emissions tests... Immunity tests... Harmonics current... Voltage fluctuation" - all passed.
    BiocompatibilityCompliance with ISO 10993-5, ISO 10993-10"Test article Contact Lens Case extract did not show potential toxicity to L-929 cells." "Test article extracts showed no significant evidence of causing ocular irritation." "Test article extract showed no significant evidence of causing skin sensitization."
    PerformanceEquivalent to predicate device"Each function operated within the design parameters, within the performance criteria established by the pertinent IEC standards." "established the equivalency to the predicate device."
    SpO2 Measurement
    Patient TypeAdult and PediatricAdult and Pediatric patients
    Measurement Accuracy$\pm$3% (70%-100%), Undefined (0-70%)$\pm$3% (during 70%-100%), Undefined (during 0-70%)
    Pulse Rate Range30 bpm-240 bpm30 bpm-240 bpm
    Pulse Rate Accuracy$\pm$2bpm or $\pm$2% (whichever is greater)$\pm$2bpm or $\pm$2% (whichever is greater)
    AlarmsFunctional alarms (Predicate has High/Low Limit)No Alarm (for SpO2 and Pulse Rate)
    NIBP Measurement
    MethodOscillometric methodOscillometric method
    Patient TypeAdult and Pediatric PatientsAdult and Pediatric patients
    Mean Deviation$\pm$5 mmHg$\pm$5 mmHg
    Standard Deviation$\le$8 mmHg$\le$8 mmHg
    Cuff Pressure Range0 to 300mmHg0 to 300mmHg
    Over Pressure ProtectionInternal operating software with limits, circuitryCuff pressure exceeds 300mmHg (Adult & Pediatric modes) at any time. (Similar to predicate's circuitry)
    AlarmsFunctional alarms (Predicate has High/Low Limit)No Alarm (for NIBP Systolic/Diastolic)
    Temperature Measurement
    Patient TypeAdult, PediatricAdult, Pediatric
    Unit of Measure°C or °F°C or °F
    Measurement SiteEar (for PC-303)Ear
    Measurement Range32.0°C to 43.0°C (90°F to 109.5°F)32.0°C to 43.0°C (90°F to 109.5°F)
    Measurement Accuracy$\pm$0.2°C (36.0-39.0°C), $\pm$0.3°C (rest)$\pm$0.2°C (36.0°C to 39.0°C), $\pm$0.3°C (the rest); $\pm$0.4°F (96.8°F to 102.2°F), $\pm$0.5°F (the rest)
    Blood Glucose (External Linked)
    IntegrationCommunication with compatible system"receives data from Blood Glucose Meter via the USB data interfaces for data communication"
    PerformanceAs per G-777G BLOOD GLUCOSE MONITORING SYSTEM (K113077)The linked device is a legally marketed system (K113077)
    ECG (External Linked)
    IntegrationCommunication with compatible system"receives data from Easy ECG Monitor via the USB data interfaces for data communication"
    PerformanceAs per PC-80 Easy ECG Monitor (K073152)The linked device is a legally marketed system (K073152)

    2. Sample size used for the test set and the data provenance

    The document states that "Clinical testing was performed for the NIBP functions and are submitted as part of this submission." However, it does not specify the sample size used for this clinical testing or the data provenance (e.g., country of origin, retrospective/prospective).

    For biocompatibility testing, the "In Vitro Cytotoxicity Test" used L929 mouse fibroblast cells, and the "Tests for irritation and skin sensitization" used guinea pigs. These are standard biological test methods, not human clinical sample sizes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information on the number or qualifications of experts used to establish ground truth for any of the performance or clinical studies.

    4. Adjudication method for the test set

    The document does not describe any adjudication method used for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This device is a patient monitor with built-in and external measurement capabilities for physiological parameters (SpO2, NIBP, TEMP, Blood Glucose, ECG). It is not an AI-assisted diagnostic device, and therefore, an MRMC comparative effectiveness study involving human readers improving with AI assistance would not be applicable and was not reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a medical instrument that provides physiological measurements. Its performance is inherent in its measurement capabilities and adherence to performance standards. The "standalone" performance is essentially its ability to accurately measure the parameters as specified, which is verified through testing against standards and comparison to a predicate device. The document mentions "Performance testing was performed to ensure that all of the device operated within specified parameters independently and with all other functions of the device operating." This indicates standalone performance testing of its direct measurement capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the measurements would typically be established by highly accurate reference instruments or established clinical methods, as dictated by the specific IEC/ISO/ASTM standards (e.g., for NIBP, a reference sphygmomanometer; for SpO2, a co-oximeter with induced hypoxia studies; for temperature, a reference thermometer). The document implies adherence to these standards for validation rather than explicitly stating ground truth derivation methods for each measurement. For NIBP, the reference to "clinical testing" would imply comparison to a clinically accepted gold standard.

    8. The sample size for the training set

    This device is a physiological monitor, not an AI/machine learning device that typically requires a "training set" in the context of algorithm development. Therefore, the concept of a training set sample size does not apply here.

    9. How the ground truth for the training set was established

    As the device does not employ a machine learning algorithm requiring a "training set," this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150093

    Validate with FDA (Live)

    Date Cleared
    2015-06-03

    (138 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pulse Oximeters are intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, indicating the pulse intensity by a bar-graph display.

    These devices are used for spot-checking a patient's SpO2 and pulse rate in a home or clinical environment. They are used by adult and pediatric patients. These devices are recommended for use on the index fingers of 1.0 - 2.2 cm thick.

    Device Description

    A blood-oxygen saturation reading indicates the percentage of hemoglobin molecules in the arterial blood which are saturated with oxygen. The reading may be referred to as SaO2. Readings vary from 0 to 100%. Normal readings in a healthy adult, however, range from 94% to 100%. The term SpO2 means the SaO2 measurement determined by pulse oximetry. In its most common (transmissive) application mode, a sensor device is placed on a thin part of the patient's body. The device passes two wavelengths of light through the body part to a photodetector. It measures the changing absorbance at each of the wavelengths, allowing it to determine the absorbances due to the pulsing arterial blood alone, excluding venous blood, skin, bone, muscle and fat. For the purpose of measuring the SpO2 more easily and accurately, our company developed the Fingertip Oximeter. The device can measure SpO2 and body pulse simultaneously.

    The PC-66A; PC-66B; PC-66C;PC-60B1; PC-60B5; PC-60D; PC-60D2; PC-60E; PC-60N; POD-1; POD-2; POD-3; PC-68A; PC-68B; PC-68C; POD-1W; POD-60NW-1 Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, indicating the pulse intensity by a bar-graph display. This device is powered by 2 AAA batteries. This device is applicable for spotchecking of SpO2 and pulse rate in a home and clinical environment. The index finger is the recommended site. It is intended for spot-checking adult and pediatric patients on fingers between 1.0 -2.2 cm thick.

    This oximeter has a wireless data transmission function. The user can transmit the SPO2 and pulse rate data to a computer through the wireless communication module.

    AI/ML Overview

    This document, a 510(k) summary for various Pulse Oximeter models from Shenzhen Creative Industry Co., Ltd., primarily asserts substantial equivalence to predicate devices and focuses on mechanical, electrical, and biocompatibility testing. It does not describe a study to prove the device meets acceptance criteria for its core function (SpO2 and pulse rate measurement) beyond a "Bench Test of SpO2 and Pulse Rate" which references accuracy ranges using a simulator.

    Therefore, many of the requested details about a study proving the device meets acceptance criteria are not available in this document. The document explicitly states: "In that the measuring device in the electro-circuit module, cushion rubber, emitter and receiver are the same as predicate devices, we concluded that this will not affect the device performance. No new clinical validation needed." This implies reliance on the predicate device's clinical validation.

    Here's an analysis of the provided information based on your requests:

    1. A table of acceptance criteria and the reported device performance

    Based on the document, the "performance" for SpO2 and Pulse Rate is presented as a target accuracy range tested using a simulator.

    Measurement ParameterAcceptance Criteria (Target)Reported Device Performance (Tested via Simulator)
    SpO2 Measuring RangeNot explicitly stated0%-99%
    SpO2 Measuring AccuracyNot explicitly stated70%-100% ±3%; 69% no defined 0%
    PR Measuring RangeNot explicitly stated30bpm-240bpm
    PR Measuring AccuracyNot explicitly stated±2bpm or ±2% (whichever is greater)

    Note: The document only provides the tested accuracy range with the simulator, not explicit acceptance criteria from an independent study. The wording "we conducted additional verification to show it complies with the requirement" implies these ranges are the requirements, but they are presented as results from a bench test with a simulator, not from a clinical study on human subjects.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not available in the provided document for the SpO2 and pulse rate accuracy claims. The document states a "Bench Test of SpO2 and Pulse Rate" was performed using an "SpO2 simulator method (simulator model: Index 2 series)". This is a laboratory test, not a clinical study on a human test set. Therefore, there is no information on human sample size, data provenance, or retrospective/prospective nature.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not available. Since the primary claims for SpO2 and pulse rate accuracy were validated using a simulator, there were no human experts involved in establishing ground truth for a clinical test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not available. As there was no clinical test set involving human subjects to establish ground truth, an adjudication method for such a test set is not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable to the device described. This document pertains to a pulse oximeter, which is a standalone measurement device, not an AI-assisted diagnostic tool that would involve human "readers" or interpretation in the way an MRMC study evaluates.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in the context of a pulse oximeter, its performance is inherently standalone. The "Bench Test of SpO2 and Pulse Rate" using a simulator can be considered a standalone performance evaluation of the device's measurement capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the "Bench Test of SpO2 and Pulse Rate," the ground truth was based on the output of an SpO2 simulator (Index 2 series). This is a synthetic ground truth, not derived from biological sources like expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    This information is not available and likely not applicable in the traditional sense of machine learning. The device is a hardware-based measurement instrument, and while it contains embedded software/firmware, the document does not describe it as an AI/ML device that requires a "training set." The performance relies on the physics of pulse oximetry and the calibration of its components.

    9. How the ground truth for the training set was established

    This information is not available and not applicable. As explained above, the device is not described as an AI/ML device that typically uses a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123711

    Validate with FDA (Live)

    Device Name
    PATIENT MONITOR
    Date Cleared
    2013-10-01

    (301 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Patient Monitor is a multi-functional instrument designed for monitoring the vital physiological signs of adult and pediatric patients. With the functions of real-time recording and displaying parameters, such as ECG, heart rate, non-invasive blood pressure, functional oxygen saturation, end-tidal CO2 concentration, respiration rate, body temperature, and so on, it allows comprehensive analysis of patient's physiological conditions.

    This instrument is applicable for use in hospitals and clinical institutions. The operation should be performed by qualified professionals only.

    Device Description

    UP-7000 Patient Monitor is a modular designed patient monitor. It monitors the patient's Electrocardiograph (ECG), respiratory rate (RR) and bo dy temperature (TEMP) by measuring physical parameters with variety modules. It can also measure non-invasive blood pressure (NIBP, the pressures of systolic, diastolic and mean) by the oscillating method. It can detect the blood oxygen saturation (SpO₂) and pulse rate (PR) non-invasively by the photoelectric method. Finally, it can extend CO2 monitor which measures the End-tidal Carbon dioxide, Inspired CO2 and Respiratory Rate. The accessories and the sensors will transfer the physical parameters into electrical signal, which will be collected and amplified by the circuit in the device. After CPU analyzing and calculating, the parameters are displayed on the screen in a graphical representation and it can record and/or print if necessary. The alarm activates if the monitored parameters go over the specified limits, alerting the medical professional.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Patient Monitor, Model UP-7000

    This submission for the Patient Monitor, Model UP-7000, primarily focuses on demonstrating substantial equivalence to predicate devices through adherence to recognized performance and safety standards. The provided documentation does not detail specific, quantitative acceptance criteria for device performance parameters like accuracy, sensitivity, or specificity, nor does it present detailed study results proving a device meets such criteria to the level often seen for diagnostic AI/ML devices. Instead, the submission relies on compliance with established international and medical device standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Function/ParameterAcceptance Criteria (Standard Compliant)Reported Device Performance
    General Safety and PerformanceCompliance with general requirements for safety and essential performance of medical electrical equipment. (e.g., IEC 60601-1: 1990+A1+A2+A13, IEC 60601-1-2: 2007 (EMC), IEC 60601-1-4: 2000 (programmable medical electrical systems), IEC 60601-1-6:2006 (usability), IEC 60601-1-8: 2006 (alarms))"Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1. Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device. All test results were satisfactory." The device is concluded to be "safe and effective and substantially equivalent to predicate devices." This implies that the device successfully passed all tests stipulated by these general safety and performance standards.
    ECG/Heart RateCompliance with particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment. (e.g., IEC 60601-2-27 :2005)"Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device. All test results were satisfactory." This confirms compliance with IEC 60601-2-27. Specific performance metrics (e.g., heart rate accuracy range, artifact rejection) are not enumerated in this summary.
    SpO2 (Functional Oxygen Saturation)Compliance with particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use. (e.g., ISO 9919: 2005)"Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device. All test results were satisfactory." This confirms compliance with ISO 9919. Specific performance metrics (e.g., SpO2 accuracy, perfusion index range) are not enumerated.
    NIBP (Non-invasive Blood Pressure)Implied compliance with relevant standards for non-invasive blood pressure measurement. (Not explicitly listed in relation to a specific NIBP standard, but "Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device.")"Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device. All test results were satisfactory." NIBP is measured by the oscillating method. Specific performance metrics (e.g., accuracy against a reference method) are not enumerated.
    CO2 (End-tidal CO2 concentration, Respiration Rate)Compliance with particular requirements for the basic safety and essential performance of respiratory gas monitors. (e.g., ISO 21647: 2004(E))"Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device. All test results were satisfactory." This confirms compliance with ISO 21647. Specific performance metrics (e.g., EtCO2 accuracy, respiration rate accuracy) are not enumerated.
    Biological Evaluation (Biocompatibility)Compliance with biological evaluation of medical devices standards. (e.g., ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010)The device was tested for biocompatibility according to ISO 10993 series, with all test results reported as satisfactory. This means it passed tests for evaluation and testing, in vitro cytotoxicity, and irritation and delayed-type hypersensitivity.

    2. Sample Size for Test Set and Data Provenance

    The provided 510(k) summary does not specify the sample size used for any performance testing or the provenance of any data (e.g., country of origin, retrospective or prospective). The submission largely relies on the declaration of compliance with recognized performance and safety standards, rather than presenting a detailed clinical study with a specific test set.

    3. Number of Experts and Their Qualifications for Ground Truth

    The provided document does not mention the use of experts to establish ground truth for a test set. This type of submission, pertaining to a patient monitor (a hardware device measuring physiological parameters), typically relies on established calibration standards and validated reference measurement systems for performance testing, rather than expert consensus on diagnostic interpretations.

    4. Adjudication Method

    Given the nature of the device and the absence of expert involvement in establishing ground truth, there is no mention of an adjudication method in the provided document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done as described in the provided summary. This type of study would be relevant for devices that assist human readers in making diagnostic interpretations, which is not the primary function of this patient monitor.

    6. Standalone (Algorithm Only) Performance Study

    The device itself is a standalone patient monitor that directly measures physiological parameters. Therefore, the "standalone performance" is essentially the device's ability to accurately measure these parameters. The submission indicates that "Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device. All test results were satisfactory." This implies that the device's performance was evaluated independently against these standards. However, no specific detailed results of such a standalone study (e.g., accuracy metrics for each parameter) are provided in the summary.

    7. Type of Ground Truth Used

    The ground truth used for testing this patient monitor would generally be based on:

    • Reference standard measurements: Comparison against highly accurate and calibrated reference devices or methods for each physiological parameter (e.g., a calibrated ECG simulator for ECG and heart rate testing, a reference oximeter for SpO2, a known pressure source for NIBP, calibrated gas mixtures for CO2).
    • Biocompatibility testing: Laboratory assays for cytotoxicity, irritation, and hypersensitivity.
    • Electrical and mechanical safety testing: Verification against defined thresholds and criteria in the relevant IEC standards.

    The summary states that "Testing to specific ISO standards was used to validate the effectiveness and accuracy of the device." These standards implicitly define the required ground truth and evaluation methods.

    8. Sample Size for the Training Set

    This device is a hardware patient monitor that measures physiological parameters, not an AI/ML device that requires a "training set" in the conventional sense of machine learning. Therefore, the concept of a "training set" as it applies to AI/ML is not relevant to this submission and no sample size for training is provided.

    9. How Ground Truth for the Training Set Was Established

    As explained above, this device does not typically utilize a "training set" in the context of AI/ML. The "ground truth" for its development and validation would have been established through engineering design, component validation, and testing against physical standards and known physiological inputs, in accordance with the cited international electrical and medical device standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120502

    Validate with FDA (Live)

    Date Cleared
    2012-12-07

    (290 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PC-60NW Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, and indicating the pulse intensity by a bar-graph display. This device is powered by 2 AAA batteries. This device is applicable for spotchecking of SpO2 and pulse rate in home and clinic for adult and pediatric patients. This device is recommended for use on the index finger, for patients with fingers of 1.0 - 2.2 cm thick.

    Device Description

    The PC-60NW Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, and indicating the pulse intensity by a bar-graph display. This device is powered by 2 AAA batteries, it is small in size, convenient to use, and easy to carry. This device is applicable for spot-checking of SpO2 and pulse rate in home and clinic. The index finger is the recommended site. It is intended for spotchecking adult and pediatric patients on fingers between 1.0 -2.2 . cm thick. This oximeter has the function of wireless data transmission. The user can effectively transmit the data to computer through the wireless communication module.

    AI/ML Overview

    The provided 510(k) summary for the PC-60NW Fingertip Oximeter does not contain the detailed study information typically requested for acceptance criteria. It primarily focuses on demonstrating substantial equivalence to a predicate device (PC-60 Fingertip Oximeter K063641) based on meeting relevant standards and general performance rather than presenting a specific clinical or technical study with detailed performance metrics against predefined acceptance criteria.

    The document states: "The Model PC-60 Fingertip Oximeter substantially has been tested in accordance with the system V & V plan and summary included with the submission using production equivalent units prior to release to market." It also mentions "Testing performed to Electrical Safety IEC 60601-1-1:1988, EMC Requirements IEC 60601-1-2, Biocompatibility IEC 10993-1, SPO2 Basic Safety and performance ISO 9919:2005, Basic Electrical safety and performance for Pulse Oximeters, ISO 80610-2-61 and pertinent FDA quidance documents, complied with the specific requirements for safety and performance."

    Therefore, much of the requested information regarding detailed study design, specific performance metrics, sample sizes for test/training sets, ground truth establishment, expert involvement, and comparative effectiveness is not explicitly available in this 510(k) summary. The summary focuses on compliance with established standards rather than a specific performance study with a standalone acceptance criteria table.

    However, based on the information provided, here's what can be extracted and inferred:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a specific table of acceptance criteria with corresponding performance results. Instead, it states that the device "complied with the specific requirements for safety and performance" according to various standards (IEC 60601-1-1, IEC 60601-1-2, IEC 10993-1, ISO 9919:2005, ISO 80610-2-61) and FDA guidance documents. These standards typically define the acceptance criteria for pulse oximeters, such as accuracy limits for SpO2 and pulse rate under various conditions. Without access to the full submission or the specific standards, the exact numerical acceptance criteria and detailed performance against them cannot be precisely tabulated from this summary.

    Inferred Performance (based on compliance with standards):

    • Pulse Oximetry Accuracy (SpO2): Expected to meet the accuracy requirements specified in ISO 9919:2005 and ISO 80610-2-61 for pulse oximeters, which typically outline a range of accuracy (e.g., ±2% or ±3% for SpO2 values between 70-100%).
    • Pulse Rate Accuracy: Expected to meet the accuracy requirements specified in the relevant standards.
    • Electrical Safety, EMC, Biocompatibility: Complies with the specified IEC standards.

    2. Sample Size Used for the Test Set and Data Provenance:

    The 510(k) summary does not specify a sample size for a test set (e.g., a specific number of patients or recordings used for performance evaluation). It mentions "system V & V plan" and "production equivalent units" but does not elaborate on the details of human subject testing or data provenance for performance validation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not provided in the 510(k) summary. The summary refers to compliance with standards, which might imply the use of reference devices or ground truth established by methodologies recognized within those standards, but it doesn't detail expert involvement.

    4. Adjudication Method for the Test Set:

    This information is not provided in the 510(k) summary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in this 510(k) summary. This type of study is more common for diagnostic imaging AI systems where human interpretation is a key component. For a pulse oximeter, the focus is on the device's direct measurement accuracy, not on human interpretation or improvement with AI assistance.

    6. Standalone Performance Study:

    Yes, a standalone performance assessment (algorithm only without human-in-the-loop performance) was implicitly done. The statement "The Model PC-60 Fingertip Oximeter substantially has been tested in accordance with the system V & V plan and summary included with the submission using production equivalent units prior to release to market" indicates that the device's performance characteristics (SpO2 and pulse rate measurement) were evaluated independently against specified requirements. However, the details of this evaluation (e.g., number of subjects, conditions tested, specific accuracy results) are not elaborated in the summary.

    7. Type of Ground Truth Used:

    The document does not explicitly state the type of ground truth used. For pulse oximeters, the typical ground truth for SpO2 accuracy studies involves co-oximetry measurements from arterial blood samples (gold standard for oxygen saturation) compared to the oximeter readings under induced hypoxia conditions in healthy volunteers. While not stated, it is highly probable that expert consensus, pathology, or outcomes data were not directly used for establishing SpO2 ground truth; rather, a reference co-oximeter would be the expected ground truth method to which the device's measurements are compared.

    8. Sample Size for the Training Set:

    This information is not provided in the 510(k) summary. For a device like a pulse oximeter, "training set" is generally not applicable in the sense of machine learning algorithms that learn from data. The device's algorithms are typically based on well-established physiological models and empirical calibration, not a "training set" in the modern AI sense.

    9. How the Ground Truth for the Training Set Was Established:

    As "training set" is not applicable in the typical AI sense, this information is not provided. The calibration and empirical basis for the oximetry algorithm would have been established during the device's development using reference methods, likely involving human subject testing with co-oximetry as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093016

    Validate with FDA (Live)

    Date Cleared
    2010-03-30

    (182 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vital Signs Monitor is designed for monitoring the vital physiological signs of the patient. It is used for non-invasive continuous monitoring of oxygen saturation (SpO2), pulse rate, CO2 and respiration rate.

    The Vital Signs Monitor is adaptable to adult and pediatric usage in a hospital environment. It is intended to be used only under regular supervision of clinical personnel.

    Device Description

    PC-900A vital signs monitor is a small Multi-parameter Patient Monitor, which can monitor the vital physiological parameters: Carbon Dioxide (CO2), Pulse Oxygen Saturation (SpO2), respiration and pulse rate. The accessories and the sensors will transfer the physical parameters into electrical signal, which will be collected and amplified by the circuit in the device. The specific sensors have been previously cleared by the FDA 510(k) process. (For specifics, please refer to the Description Section). After CPU analyzing and calculating, the parameters can display on the screen in a graphical way, record and/or print if necessary. The alarm will work if the parameters over the limits to take medical practitioner's attention.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text for the Vital Signs Monitor, Model PC-900A:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing specific acceptance criteria for a new clinical trial. However, it does list performance specifications and accuracy claims for the device. These can be interpreted as the functional acceptance criteria the manufacturer aims for, by comparing them to predicate devices.

    ParameterAcceptance Criteria (Implied by equivalence and typical performance claims)Reported Device Performance (PC-900A)
    SpO₂ Measurement RangeSimilar to predicate devices (e.g., 1%-100%)70%~99%
    SpO₂ AccuracySimilar to predicate devices (e.g., Adult/Pediatric: ±2% (70-100%), Neonate: ±3% (70-100%))Adult and Pediatric: $\pm$ 3% (during 70%99%)Undefined (during 070%)
    SpO₂ AlarmsHigh and lower alarms, adjustable limitsHigh and lower alarms. The limits are adjustable.
    Pulse Rate Display RangeSimilar to predicate devices (e.g., 20 bpm ~ 250 bpm)30 bpm~240 bpm
    Pulse Rate AccuracySimilar to predicate devices (e.g., ±3 bpm)$\pm$ 2bpm or $\pm$ 2% (whichever is greater)
    Pulse Rate AlarmsHigh and lower alarms, adjustable limitsHigh and lower alarms. The limits are adjustable.
    CO₂ ModuleUse of previously cleared modules (e.g., Respironics LoFlo™ EtCO2, CAPNOSTAT 5 EtCO2)Respironics LoFlo™ EtCO2 (Side-stream) Module(K053174) and CAPNOSTAT 5 EtCO2 (Main-stream) Module(K042601)
    CO₂ Measurement MethodInfrared absorption methodInfrared absorption method
    CO₂ Measure ModeSidestream or MainstreamSidestream or Mainstream
    CO₂ Response TimeSidestream: <3s (includes transport and rise); Mainstream: <60ms (rise time)Sidestream: <3seconds (includes transport time and rise time).Mainstream: <60ms (rise time)
    CO₂ Accuracy (EtCO₂: 0-150mmHg)0-40 mmHg ±2mmHg, 41-70 mmHg ±5% of reading, 71-100 mmHg ±8% of reading, 101-150mmHg ±10% of reading (for Respironics LoFlo/CAPNOSTAT)040 mmHg ±2mmHg4170 mmHg ±5% of reading71100 mmHg ±8% of reading101150mmHg±10% of reading
    CO₂ Accuracy (EtCO₂: 0-99mmHg)0-38 mmHg: ± 2 mmHg; 39-99 mmHg: ± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg) (Oridion Capnostream) OR 0-40mmHg ±2mmHg; 41-76mmHg ±8% of reading; 77-99mmHg ±10% of reading (Edan M3B)038 mmHg: ± 2 mmHg; 3999 mmHg: ± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg) (from Oridion column)040mmHg ±2mmHg; 4176mmHg ±8% of reading; 77~99mmHg±10% of reading (from Edan column)
    Respiration Rate Measure Range2-150 rpm or 0-150 rpm2150rpm (Sidestream) or 0150rpm (Mainstream)
    Respiration Rate AccuracySimilar to predicate (e.g., 0-70 rpm: ±1 rpm; 71-120 rpm: ±2 rpm; 121-150 rpm: ±3 rpm for Capnostream)±2rpm (General Statement)070 rpm: ±1 rpm; 71120 rpm: ±2 rpm; 121~150 rpm: ±3 rpm (Detailed breakdown, presumably for the CO2 module)
    Flow Rate (Sidestream)50ml/min ±10 ml/min50ml/min ±10 ml/min (Sidestream)50 (42.5≤flow≤65) ml/min, flow measured by volume (From predicate description, likely applicable to the module)
    Suffocation Alarm Delay10~60s10~60s
    Alarms of EtCO₂, RRHigh and lower alarms, adjustable limitsHigh and lower alarms. The limits are adjustable.

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not provide details on the sample size for a "test set" or data provenance (e.g., country of origin, retrospective/prospective clinical data).

    The document states: "The PC 900A has undergone Third Party safety testing in accordance with IEC standards and completed performance testing in accordance with IEC standards." This typically refers to bench testing and electrical safety, not clinical performance studies with a specific test set of patients whose data is then explicitly analyzed for the stated performance metrics. The performance metrics are likely derived from the specifications of the integrated modules (e.g., SpO2 module, CO2 module) and validated through benchtop testing or limited in-vivo testing to meet the module's specified accuracy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the summary. As the submission focuses on substantial equivalence based on technical specifications and predicate device comparison, a formal process of establishing ground truth by a panel of clinical experts for a dedicated test set is not described. The accuracy claims for SpO2 and CO2 modules are derived from their own clearances and internal validation data, which typically involve comparisons to reference methods rather than expert consensus on subjective data.

    4. Adjudication Method for the Test Set

    This information is not provided. Given the lack of a detailed clinical test set and expert review, an adjudication method would not be applicable in the context described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or described in this 510(k) summary. This type of study is more common for diagnostic imaging AI systems where human interpretation is a key component. The Vital Signs Monitor is a device that provides numerical physiological parameters, not an interpretation of complex data by a human reader.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The performance data presented are standalone performance specifications of the device and its integrated modules. For example, the SpO2 accuracy and CO2 accuracy are direct measurements of the device's ability to provide accurate physiological readings, independent of a human operator's interpretation, other than their ability to correctly apply the sensors. No "algorithm only" study is explicitly mentioned as a separate entity, but the performance claims are inherently about the device's measurement capabilities.

    7. The Type of Ground Truth Used

    For the SpO2 and CO2 accuracy claims, the ground truth would typically be established by a reference method (e.g., co-oximetry for SpO2, calibrated gas analyzers for CO2) during controlled laboratory or clinical studies for the respective sensors/modules. The 510(k) refers to the use of "Creative SpO2 module with the same principle as PC-60(K063641) Creative SpO2 sensor only" and "Respironics LoFlo™ EtCO2 (Side-stream) Module(K053174) and CAPNOSTAT 5 EtCO2 (Main-stream) Module(K042601)." The accuracy statements for these components are derived from their own prior regulatory clearances and the studies that supported them, which would have used appropriate gold standard reference methods. The document itself does not detail these ground truth methodologies for this specific submission, but rather references the cleared modules.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. The Vital Signs Monitor is a hardware device integrating sensors and processing units, not an AI software system that undergoes "training" in the machine learning sense. The underlying algorithms for parameter measurements (like SpO2 and CO2) are based on established physiological principles and signal processing, not a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable due to the nature of the device (hardware monitor, not a machine learning model).

    Ask a Question

    Ask a specific question about this device

    K Number
    K082055

    Validate with FDA (Live)

    Date Cleared
    2009-02-06

    (200 days)

    Product Code
    Regulation Number
    884.2660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.

    Device Description

    This Doppler tests the fetal heart rate through non-invasive ultrasonic Doppler Effect. As is known, ultrasonic wave propagating at a given frequency will be reflected when encountering an obstacle. If it is still an obstacle, the back wave will share the same frequency with the transmitted wave. Once the obstacle moves, the frequency of the back wave will be changed. The higher rate the object moves at, the bigger frequency change will take place. This is the so-called Doppler Effect. With the apparatus, the ultrasonic probe is placed on the abdomen of the pregnant woman. The ultrasonic probe can perceive the fetal heartbeat. When the transmitted wave encounters the fetal heart, the back wave will develop offset frequency. With the offset frequency, the fetal heart rate and frequency can be worked out.

    AI/ML Overview

    The provided text from K082055 is a 510(k) summary for the PC-860 Fetal Doppler. It primarily focuses on demonstrating substantial equivalence to a predicate device and safety/performance based on standards.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for Fetal Heart Rate (FHR) detection accuracy. Instead, it broadly states that "Clinical testing was used to validate the effectiveness and accuracy of the device. Specific testing relating to Ultrasound equipment was use to verify performance to recognized standards. All test results were satisfactory."

    Given this, the table below reflects what can be inferred or is directly stated:

    Acceptance Criteria (Inferred/Stated)Reported Device Performance
    Effectiveness and Accuracy of FHR DetectionClinical testing validated effectiveness and accuracy. Test results were satisfactory.
    Safety (Electrical, Mechanical, Environmental)Met standard EN/IEC 60601-1. Test results were satisfactory.
    Performance to Recognized Ultrasound StandardsPerformance verified to recognized standards. Test results were satisfactory.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Clinical testing was used to validate the effectiveness and accuracy of the device" but does not specify the sample size for this clinical testing.

    The data provenance (e.g., country of origin, retrospective or prospective) is not mentioned.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not provide any information regarding the number of experts, their qualifications, or their role in establishing ground truth for any test set.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method used for a test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being conducted or its results. The device appears to be a standalone Fetal Doppler, not an AI-assisted interpretation tool for human readers.

    6. Standalone Performance Study

    The document states "Clinical testing was used to validate the effectiveness and accuracy of the device." This implies a standalone performance study was done to evaluate the device's ability to detect Fetal Heart Rate independently. However, specific performance metrics (e.g., sensitivity, specificity, accuracy against a reference) are not provided beyond the general statement of "satisfactory."

    7. Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for the clinical testing. Given it's a Fetal Doppler, the ground truth would typically be a reference method for FHR measurement (e.g., a cardiotocograph (CTG), a more advanced ultrasound, or expert auscultation with a stopwatch), but this is not detailed.

    8. Sample Size for the Training Set

    The document does not mention a training set or its sample size. Fetal Dopplers of this type typically do not involve machine learning models that require a distinct training set in the same way modern AI systems do. Their function is based on physical principles (Doppler effect) and signal processing.

    9. How the Ground Truth for the Training Set Was Established

    Since a training set is not mentioned as part of the device's development or validation, there is no information on how its ground truth might have been established.


    Summary of Missing Information:

    This 510(k) summary is from 2009, a period before the widespread requirement for detailed AI/ML validation studies. As such, it lacks specific quantitative acceptance criteria, detailed clinical study methodologies (sample size, expert involvement, ground truth specifics), and any information relating to AI-specific study paradigms like MRMC or training sets. The submission focuses on substantial equivalence to a predicate device and adherence to general safety and performance standards for medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073152

    Validate with FDA (Live)

    Date Cleared
    2008-01-10

    (63 days)

    Product Code
    Regulation Number
    870.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a handheld, personal electrocardiograph (ECG) unit, which can measure electrical activities of the heart easily and conveniently. It's a single LEAD ECG measuring device for home health care use, which can detect, display and store ECG signal, and if possible, provide average heart rate message after ECG measurement. The users can use it themselves to check their heart condition.

    It is suitable for the adult users (age 18 or older), who suffers from cardio-vascular diseases, or the adult people who are caring about their heart working conditions during their daily life. This device is not intended for use as a conventional diagnostic tool, but use as a healthcare tool which can provide doctor the recorded data as references. This Device is also not intended for recording and transmission of user's ECG signal simultaneously, and it's not recommended to use with implanted pacemaker.

    Device Description

    PC-80 / Prince 180 Easy ECG Monitor is a handheld electrocardiograph unit which can measure, display, and store ECG data by the user. The device can calculate the average heart rate and display the heart rhythm after analysis. Furthermore, it's small, light, and easy to use. A 30-second ECG record is measured and stored in this device. The waveform data and heart rate value can be displayed on LCD screen, and those ECG data records can be reviewed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the PC-80 /Prince 180 Easy ECG Monitor, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific acceptance criteria (e.g., sensitivity, specificity, accuracy targets) for its performance in detecting or interpreting ECG signals. Instead, the performance testing focuses on compliance with established medical device standards.

    Acceptance Criteria (Not explicitly stated as performance targets, but implied by compliance)Reported Device Performance (Summary of compliance)
    Safety based on IEC 60601-1Device has undergone and validated against IEC 60601-1.
    Electromagnetic Compatibility (EMC) based on IEC 60601-1-2Device has undergone and validated against IEC 60601-1-2.
    Performance of Ambulatory Electrocardiographs based on AAMI EC 38Device has undergone and validated against AAMI EC 38.
    Risk management and mitigation complianceRisk analysis identifying potential hazards and documenting mitigation has been developed and applied.
    Quality system compliance (21CFR820, ISO 13485)Quality system conforms to 21CFR820, MDCAS ISO 13485 certified by ORKI.
    Heart Rate Monitoring RangeMonitors from 30-240 bpm (Predicate: 2-200 bpm).
    ECG measurement capabilitiesMeasures, displays, and stores ECG data by user. Calculates average heart rate and displays heart rhythm after analysis.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide details on the sample size used for any clinical test set or the data provenance (e.g., country of origin, retrospective/prospective). The performance testing described is primarily focused on engineering and quality system compliance rather than a clinical performance study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not mention any clinical test set involving expert review to establish ground truth for ECG interpretation. The device's intended use is for home health care and as a reference tool for doctors, not as a diagnostic tool itself.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set with human review requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. The submission focuses on substantial equivalence based on technical standards and device features, not on improved human reader performance with AI assistance.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) was Done

    The testing described is focused on the device as a whole, including its ability to measure, display, and store ECG data, and to provide average heart rate and rhythm information. There's no specific mention of a standalone algorithm performance study in terms of diagnostic accuracy metrics (e.g., sensitivity, specificity) for automatically detecting specific cardiac events. The performance validation is against engineering and safety standards (IEC, AAMI).

    7. The Type of Ground Truth Used

    For the engineering and safety standards (IEC 60601-1, IEC 60601-1-2, AAMI EC 38) to which the device was validated, the "ground truth" would be the specifications and requirements outlined in those standards. This would involve calibrated test equipment and established protocols to verify electrical safety, EMC, and the accuracy of ECG signal acquisition and basic heart rate calculation as per the AAMI standard. There is no mention of clinical ground truth (e.g., expert consensus, pathology, or outcomes data) in the context of diagnostic accuracy.

    8. The Sample Size for the Training Set

    The document does not mention a training set, which would typically be associated with AI/ML algorithm development. The device's description suggests a direct measurement and display function, with basic analysis for heart rate and rhythm, rather than a sophisticated AI-driven interpretive algorithm that would require a large training dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set is mentioned.