Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K124044

    Validate with FDA (Live)

    Date Cleared
    2013-08-30

    (242 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aloe Cadabra® Lubricant and Aloe Cadabra® Flavored/Scented Lubricants are personal lubricants, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Aloe Cadabra® Lubricant and Aloe Cadabra® Flavored/Scented Lubricants are non-sterile, aloe-based vaginal lubricants designed to supplement the body's own natural lubrication fluids, and to enhance the ease and comfort of intimate sexual activity. Aloe Cadabra® Lubricant and Aloe Cadabra® Flavored/Scented Lubricants are specifically formulated with 95% organic aloe. The remaining five percent of ingredients are in all cases food grade and in many cases certified organic. This device is not a contraceptive or spermicide, nor does it contain any such component. All ingredients in Aloe Cadabra Flavored/Scented Lubricants are GRAS (Generally Recognized as Safe by FDA). The specifications for the Aloe Cadabra Lubricants include appearance, odor, pH, viscosity, osmolality, total microbial count, fungal/yeast/mold limits, and absence of pathogenic organisms (straphylococcus aureus, pseudomonas aeruginosa, escheria coli, salmonella, bile tolerant gram negative bacteria, c. albicans and clostridia). The product is bottled in an HDPE opaque white plastic bottle with a white flip top cap. The bottle is packaged into a carton for sale to consumers. The product comes in five varieties of scents and flavors including Natural Aloe, Pina Colada, Tahitian Vanilla, French Lavender, and Peppermint.

    AI/ML Overview

    The provided text describes a 510(k) summary for a personal lubricant, not an AI-powered medical device that requires a study to prove acceptance criteria. Therefore, most of the requested information (such as sample size, ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) is not applicable to this document.

    However, I can extract the acceptance criteria and performance data related to the lubricant's safety and compatibility.

    Acceptance Criteria and Reported Device Performance for Aloe Cadabra® Lubricants

    The acceptance criteria for the Aloe Cadabra® Lubricants are established through various performance tests, primarily focusing on biocompatibility and condom compatibility, in comparison to the predicate device. The device aims to meet these criteria to demonstrate substantial equivalence to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device PerformanceStudy Type/Standard Applied
    BiocompatibilityAcute Systemic ToxicityNo systemic responses or biological reactivity in test animals.All test group animals survived the test period and exhibited no biological reactivity at any tested time points.ISO 10993-11: 2006
    Vaginal Irritation TestingNon-irritating to vaginal mucosal tissue.Considered non-irritating to the vaginal mucosa in female New Zealand White Rabbits.ISO 10993-10: 2010
    CytotoxicityNo cytotoxic effect.Not considered to have a cytotoxic effect based on qualitative evaluation.ANSI/AAMI/ISO 10993-5: 2009
    Skin SensitizationNo sensitization reactions.Did not elicit sensitization reactions in the animals used.ISO 10993-10: 2010
    Product StabilityShelf LifeDemonstrate a two-year shelf life.Achieved a two-year shelf life based on real-time aging study results.Real-time aging study
    Condom CompatibilityNatural Rubber Latex CondomsCompatible.Compatible with natural rubber latex condoms.ASTM D7661-10
    Polyisoprene CondomsCompatible.Compatible with polyisoprene condoms.ASTM D7661-10
    Polyurethane CondomsCompatible (or specified as incompatible).Not compatible with polyurethane condoms.ASTM D7661-10

    2. Sample Size Used for the Test Set and Data Provenance

    • Acute Systemic Toxicity: Mice were used. Specific sample size not provided, but the standard implies a sufficient number for statistical relevance per ISO 10993-11.
    • Vaginal Irritation Testing: Female New Zealand White Rabbits were used. Specific sample size not provided, but the standard implies a sufficient number per ISO 10993-10.
    • Cytotoxicity: Not applicable to in vivo subjects; likely cell culture tests. Specific sample size (e.g., number of cell lines or replicates) not provided.
    • Skin Sensitization: Guinea pigs were used. Specific sample size not provided, but the standard implies a sufficient number per ISO 10993-10.
    • Shelf Life: Not applicable to biological samples; involves product samples over time. Specific number of product samples not provided.
    • Condom Compatibility: Involved all 5 versions of the lubricants. Specific number of condoms tested per type not provided.

    Data Provenance: The studies are described as performance tests performed by the submitter (Seven Oaks Ranch, Inc.) or contracted labs according to international ISO standards. The country of origin for the studies is not explicitly stated, but the submission is to the US FDA. The studies are prospective in nature, actively testing the device to gather performance data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable. The "ground truth" for these tests refers to established biological responses (e.g., irritation, toxicity) as measured by standard laboratory methods outlined in the ISO documents. It does not involve expert consensus on image interpretation or similar qualitative assessments performed by medical professionals for diagnostic devices.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication typically refers to the process of resolving disagreements among experts when establishing ground truth. For these objective, standardized biological and physical tests, the results are determined by laboratory observations and measurements against defined criteria, not by human expert interpretation that requires adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. These studies are typically conducted for diagnostic devices where human readers (e.g., radiologists) interpret cases with and without AI assistance. The Aloe Cadabra® Lubricants are personal care products, not diagnostic devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, this question is not applicable. Standalone performance refers to the performance of an algorithm without human intervention, usually in the context of an AI/ML-powered device. The Aloe Cadabra® Lubricants are physical products, not software algorithms.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests relies on objective, scientific measurements and observations performed according to established international standards (ISO, ASTM) for biological reactivity, irritation, toxicity, and material compatibility. For example:

    • Biocompatibility: Absence of adverse biological reactions in animal models or cell cultures, as defined by the specific ISO standards.
    • Shelf Life: The product maintaining its specifications (e.g., pH, viscosity, microbial count) over a defined period.
    • Condom Compatibility: The condoms maintaining their physical integrity and performance after exposure to the lubricant, as defined by ASTM D7661-10.

    8. The Sample Size for the Training Set

    This information is not applicable. A "training set" refers to data used to train an AI/ML algorithm. The Aloe Cadabra® Lubricants are physical products and do not involve AI/ML.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1