    Intended Use

    Guardian 150 electrode is intended for use with TENS and NMES, including for Muscle reeducation by application of external stimulation to the muscles necessary for pharyngeal contraction.

    Device Description

    Guardian 150 electrode is a single-patient, single use, self-adhering electrode for use with TENS and NMES, and available in cloth (models 150. 151. 152 and 3637), pre-wired (model 3637), snap (models 150, 151 and 152) and pin style (model 3637) configurations. It is latexfree.

    AI/ML Overview

    The provided text does NOT describe acceptance criteria for a medical device or a study that proves the device meets specific acceptance criteria.

    Instead, the documents are:

    • A 510(k) summary for the "Guardian 150 electrode" from SelectiveMed™ Components, Inc.
    • An FDA letter confirming the substantial equivalence of the Guardian 150 electrode to predicate devices.
    • The "Indications for Use" for the Guardian 150 electrode.

    These documents establish the device's classification, intended use, and substantial equivalence to existing devices. However, they do not contain any information regarding specific performance acceptance criteria or a study that outlines how these criteria were met.

    Therefore, I cannot provide the requested table or answer the specific questions about sample size, ground truth, expert qualifications, or comparative effectiveness studies, as that information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031053

    Validate with FDA (Live)

    Date Cleared
    2003-05-20

    (48 days)

    Product Code
    Regulation Number
    890.5525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Selective Med Components, Inc. Buffered Iontophoresis Drug Delivery Electrodes are intended to be used to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.
    Selective Med Components, Inc., Buffered Iontophoresis Drug Delivery System Electrodes are indicated to introduce ions of soluble salts or other drugs into the body.

    Device Description

    The Selective Med Components, Inc. Buffered Iontophoresis Drug Delivery Electrode System consists of an active drug delivery electrode and a passive return electrode. These electrodes are designed for one use by a single patient for the local administration of ionic drug solutions into the body for medical purposes. There are multiple sizes and shapes of drug delivery electrodes to accommodate placement various sites on the body. The size of the return electrode is the same for all drug delivery electrode sizes. The SMC Buffered Iontophoresis Drug Delivery Electrodes have technological characteristics equivalent to those of the predicate devices, including comparable performance specifications, comparable materials including the same buffering agent (Ag/AgCl) and fiberous polyester reservoir materials on the active drug delivery electrodes and the same buffering, self-adhering polymer on the return electrode, multiple shapes and sizes of active the drug delivery electrodes, and equivalent packaging and labeling.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria, device performance, a study to prove acceptance criteria, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth details.

    The document is a 510(k) summary for a medical device (Selective Med Components, Inc. Buffered Iontophoresis Drug Delivery Electrodes), which primarily focuses on demonstrating substantial equivalence to already legally marketed predicate devices. It describes the device, its intended use, and lists predicate devices, but it does not include performance data or studies against specific acceptance criteria.

    Therefore, I cannot fulfill your request for that information based on the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020476

    Validate with FDA (Live)

    Device Name
    MEDION 6000
    Date Cleared
    2002-05-03

    (80 days)

    Product Code
    Regulation Number
    890.5525
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended to use a direct current to introduce ions of soluble salts or other drugs into the body for medical purposes, as an alternative to hypodermic injection when it is advisable to avoid the pain that may accompany needle insertion and drug injection, and when it is advisable to minimize the infiltration of carrier fluids or to avoid the damage that may be caused by needle insertion when tissue is traumatized.

    Device Description

    The iontophoresis device is intended to use a direct current to introduce ions of soluble salts or other drugs into the body for medical purposes. The iontophoretic technology operates on the principle that an electic potential will cause ions in solution to migrate according to their electrical charges. Further, the quantity and distribution of a drug delivered into or across the skin by iontophoresis is dependent upon the charge and size (molecular weight) of the ion, strength of the electircal current being applied, electrode composition, duration of current flow and several other factors. The Medion™ is a battery-powered, solid state, microprocessor controlled device which controls current strength and duration, calculates total charge delivered and monitors current flow and electrode/tissue impedance.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Medion™ 6000 Series Iontophoresis Device. This submission focuses on demonstrating substantial equivalence to predicate devices based on bench testing. As such, it does not involve the types of studies typically associated with AI/ML-driven devices, which would require the specific data points you've requested (such as sample sizes for test/training sets, expert qualifications, MRMC studies, etc.).

    Therefore, I cannot provide all the requested information because the provided document does not contain information related to AI/ML device performance or clinical studies with human readers, training sets, or ground truth establishment in the way you've outlined.

    However, I can extract the available information regarding acceptance criteria and the study performed:

    Acceptance Criteria and Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Electrical OutputMeets specifications fully
    Device OperationOperates as described
    Substantial EquivalenceSubstantially equivalent to Iomed Phoresor II Model PM 900 (K982668), Iomed Phoresor II Model PM 800 (K933445), and R. A. Fischer MD-2 Iontophoresis Unit (K895365). Most closely similar to the Phoresor Model PM 800/900 units.

    Study Details (Based on available information in the document)

    • Sample size used for the test set and the data provenance: Not applicable. The "test set" in this context refers to the device itself and its performance against specifications, not a dataset for an algorithm. The study was non-clinical and involved bench testing.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of AI/ML, refers to human-labeled data. Here, the "ground truth" equivalent would be the engineering specifications and established performance of predicate devices.
    • Adjudication method for the test set: Not applicable. This concept is relevant for reconciling expert disagreements on ground truth annotations, which isn't present in this type of study.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No. This document describes a traditional medical device (iontophoresis device) and not an AI/ML-driven device. Therefore, no MRMC study, human readers, or AI assistance is mentioned.
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No. This section is not applicable as the device is not an algorithm.
    • The type of ground truth used: For the purposes of demonstrating substantial equivalence, the "ground truth" was the engineering specifications for the device's electrical output and operational characteristics, and the established regulatory status and performance of the predicate devices.
    • The sample size for the training set: Not applicable. There is no AI/ML algorithm involved that would require a training set.
    • How the ground truth for the training set was established: Not applicable, as there is no training set for an AI/ML algorithm.

    In summary, the provided document describes a non-clinical bench test to confirm that the Medion™ 6000 Series Iontophoresis Device meets its electrical output specifications and operates as described, demonstrating substantial equivalence to its predicate devices. This is a common pathway for traditional medical device clearances and does not involve AI/ML-specific study designs.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993081

    Validate with FDA (Live)

    Device Name
    IO-DRIVE
    Date Cleared
    2000-03-03

    (170 days)

    Product Code
    Regulation Number
    890.5525
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Io-Drive electrode is indicated to introduce ions of soluble salts and other drugs into body.

    Device Description

    The IO-Drive Iontophoresis electrode system consists of an active drug delivery electrode and a return electrode. These electrodes are designed for single patient use. There are multiple sizes of drug delivery electrodes to accommodate placement at different body sites. The size of the return electrode is the same size for all drug delivery electrodes.

    AI/ML Overview

    The provided document, K993081, describes the IO-DRIVE Iontophoresis electrode system. This submission is a 510(k) premarket notification, indicating that the device aims to demonstrate substantial equivalence to existing legally marketed predicate devices, rather than proving novel safety and effectiveness through extensive clinical trials.

    Based on the provided information, the following can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Equivalence to Predicate Devices: The IO-Drive electrode system should function similarly to the Iomed TransQ (K91-4621) and Empi Dupel (K91-2014) in terms of delivering ionic drug solutions into the body via iontophoresis."A performance evaluation was performed and based on the evaluation the IO-Drive electrode is similar to the predicate electrodes."
    Intended Use Equivalence: The device should have the same intended use as predicate devices."The electrode is intended to be used in the clinic. Iontophoresis drug delivery systems are indicated for the local administration of ionic drug solutions into the body for medical purposes and can be used as an alternative to injections." This matches the general indications for iontophoresis devices.
    Safety and Material Compatibility: While not explicitly detailed as acceptance criteria, substantial equivalence implies the materials and design are safe for single patient use as an external electrode.The device consists of an active drug delivery electrode and a return electrode, designed for single patient use, with multiple sizes to accommodate different body sites. This description is consistent with typical iontophoresis electrode designs.
    Manufacturing Quality: Compliance with Current Good Manufacturing Practice (GMP) requirements.The FDA letter states: "A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size for a test set or data provenance (e.g., country of origin, retrospective/prospective study details). The "performance evaluation" mentioned is very high-level and does not provide these specifics. For a 510(k) submission focused on substantial equivalence, such detailed clinical trial data with specific test sets and provenance is often not required if the device is sufficiently similar to predicates and performance can be established through bench testing or limited clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Given the nature of a 510(k) submission relying on "performance evaluation" for substantial equivalence, it's unlikely that an extensive expert-adjudicated test set in the traditional sense was used to establish ground truth for a novel "device performance" metric. Instead, the "ground truth" for comparison would be the established performance characteristics of the predicate devices.

    4. Adjudication Method for the Test Set

    This information is not provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not performed or reported in this document. The submission is for a medical device (electrode system), not an imaging or diagnostic AI algorithm, so an MRMC study would generally not be applicable in this context.

    6. Standalone (Algorithm Only) Performance Study

    A standalone performance study was not performed or reported. This device is an electrode system for drug delivery, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this 510(k) submission is primarily performance characteristics and established safety/effectiveness profiles of the predicate devices. The IO-Drive electrode system's "performance evaluation" aimed to show similarity to these predicates. The FDA's substantial equivalence determination implies that the new device is as safe and effective as the predicate devices based on available information, which might include design comparisons, material testing, and limited functional testing, rather than an independent "ground truth" based on clinical outcomes.

    8. Sample Size for the Training Set

    This information is not applicable/not provided. The IO-Drive system is an electrode, not an AI algorithm, so there is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the same reasons as above.