510(k) Data Aggregation

    K Number
    K150002

    Validate with FDA (Live)

    Date Cleared
    2015-06-23

    (172 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iNSIGHT Color DopplerUltrasound System is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets.

    Device Description

    The iNSIGHT Color Doppler Ultrasound Systems are integrated preprogrammed color ultrasound imaging system, capable of producing a resolution intended for clinical diagnostic imaging applications. Their basic function is to acquire ultrasound signal and display the images in the following imaging operations: B, M, PW, CW, CDI, PDI, DPDI, TDI, MC and THI imaging mode, Spatial Compounding Imaging, Panoramic B and Color mode, Freehand 3D, 4D Imaging, Strain imaging, ARFI Elasticity Imaging.

    AI/ML Overview

    The provided text is a 510(k) summary for the iNSIGHT Color Doppler Ultrasound System. It describes the device, its intended use, and its substantial equivalence to predicate devices, but it does not contain acceptance criteria for device performance or a study demonstrating that the device meets such criteria.

    The document primarily focuses on regulatory compliance, outlining the device's technical characteristics, regulatory classifications, and a comparison with previously cleared predicate devices (K090059 and K130881).

    Here's a breakdown of what is and isn't present in the given text regarding your request:

    What is (partially) present:

    • Non-Clinical Test Conclusion: The document states that "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device." It also mentions "bench testing for the features providing strain and elasticity quantitative has been conducted to show that the accuracy of the strain and elasticity imaging of the proposed device; the result has been analyzed that they are accepted for clinical use."
    • Type of Ground Truth (for non-clinical testing): For the "strain and elasticity quantitative measurements," the ground truth seems to be implicitly established through the "accuracy" of the measurements, likely against a known standard or phantom in a bench test setting, rather than expert consensus, pathology, or outcomes data from human subjects.

    What is NOT present in the provided text:

    1. A table of acceptance criteria and the reported device performance: The document lists several IEC and ISO standards that the device complies with (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2-2004, ISO 10993-5, ISO 10993-10). These are general safety and performance standards for medical electrical equipment and diagnostic ultrasound, not specific performance metrics or acceptance criteria for image quality or diagnostic accuracy. While compliance with these standards could be considered a form of acceptance criteria, the document does not report specific device performance against these. The only performance mention is for "strain and elasticity quantitative measurements" stating they were "accepted for clinical use," but without quantitative details.
    2. Sample size used for the test set and the data provenance: No details are provided regarding sample sizes for any performance testing, nor the origin (country, retrospective/prospective) of any data.
    3. Number of experts used to establish the ground truth for the test set and their qualifications: Not applicable as there are no details of a test set requiring expert ground truth in the provided document beyond the general statement of "bench testing."
    4. Adjudication method: Not applicable.
    5. Multi Reader Multi Case (MRMC) comparative effectiveness study: The document explicitly states: "No clinical study is included in this submission." Therefore, no MRMC study was performed or reported.
    6. Standalone (algorithm only without human-in-the-loop performance) study: While "bench testing" was mentioned for strain and elasticity, specific details of a standalone performance study in the context of diagnostic accuracy are not provided. The device is an ultrasound system, not an AI algorithm that would typically undergo standalone algorithm-only performance testing.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): As noted, for the limited performance mention, it implies comparison to known values in a bench test, not clinical ground truth.
    8. The sample size for the training set: Not applicable, as this is an ultrasound imaging system, not an AI/machine learning model where a training set of data would be typically discussed in this context.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided FDA 510(k) summary focuses on validating the substantial equivalence of the iNSIGHT Color Doppler Ultrasound System to existing predicate devices based on its intended use, technology characteristics, and adherence to established medical device safety and performance standards. It does not include a detailed study with specific acceptance criteria for diagnostic performance outcomes (like sensitivity, specificity, accuracy) or the data supporting such performance, nor does it describe any clinical studies. The mention of "bench testing" relates to the accuracy of "strain and elasticity quantitative measurements" but lacks specific details regarding the methodology or quantitative results in relation to defined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090059

    Validate with FDA (Live)

    Date Cleared
    2009-04-30

    (111 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iMago c21 medical ultrasound system is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets. The detailed application is listed in the User Guide-and-all these applications must be operated by qualified and trained Physician or "Sonographer".

    Device Description

    The iMago c21 Diagnostic Ultrasound System is highly mobile, software-controlled, totally digital processing and PC-based ultrasound imaging system capable of the following operating modes: B, M, Pulsed Doppler, Color Flow(including amplitude Doppler). The system can generate real-time compound images and harmonic images, as well as 3D images and Wide-Field-of-View images. The system is designed for use in convex and linear scanning modes, and supports convex and linear array probes.

    AI/ML Overview

    This 510(k) summary (K090059) describes the iMago c21 Diagnostic Ultrasound System. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a dedicated study. As such, information regarding acceptance criteria and performance metrics (sensitivity, specificity, AUC) from a typical clinical study is not present. The document focuses on technical specifications, intended use, and compliance with general safety standards.

    Here's a breakdown of the available information based on your questions:

    1. Table of Acceptance Criteria and Reported Device Performance:

    No specific acceptance criteria related to diagnostic performance (e.g., sensitivity, specificity, accuracy) or a table comparing them to reported device performance metrics are provided in this 510(k) summary. The document focuses on demonstrating substantial equivalence to predicate devices by comparing intended use, principles of operation, and technological characteristics.

    Instead, the device's "performance" is implicitly evaluated through its compliance with recognized safety and performance standards for ultrasound equipment.

    Aspect of Performance/Acceptance CriteriaReported Device Performance/Compliance
    Acoustic Output Limits (MI and TI)Complies with NEMA UD2-2004; MI and TI do not exceed 1.0 in any operation modes.
    General Electrical SafetyComplies with IEC 60601-1 (2001) and IEC 60601-2-37.
    Electromagnetic Compatibility (EMC)Complies with EN60601-1-2:2001+A1:2006, EN61000-3-2:2006, and EN 61000-3-3: 1995+A1:2001+A2:2005.
    Image Quality/Diagnostic PerformanceNot explicitly reported with quantitative metrics in this document. Substantial equivalence to predicate devices (Sonosite Hand-Carried Ultrasound system (K014116), ES500 system (K020630)) implies comparable diagnostic performance for the stated indications.

    2. Sample size used for the test set and the data provenance:

    Not applicable. This 510(k) submission does not describe a clinical study with a test set of patient data to evaluate algorithmic performance. The submission relies on technical specifications and comparison to predicate devices, not on a new clinical performance study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. No ground truth establishment using experts is mentioned, as there was no clinical performance study presented in this document.

    4. Adjudication method for the test set:

    Not applicable, as no clinical performance study with a test set is outlined.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a submission for a diagnostic ultrasound system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used:

    Not applicable. No clinical ground truth was established for a performance study. The "ground truth" for the device's functionality is its adherence to industry standards and its demonstrated substantial equivalence to existing, legally marketed ultrasound systems.

    8. The sample size for the training set:

    Not applicable. This document describes a medical device (ultrasound system), not a machine learning algorithm. Therefore, there is no "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established:

    Not applicable, as no training set for an AI/ML algorithm is relevant to this submission.