Search Filters

Search Results

Found 49 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EVO Q30 Diagnostic Ultrasound System; EVO Q20 Diagnostic Ultrasound System; EVO Q10 Diagnostic Ultrasound System; EVO XQ30 Diagnostic Ultrasound System; EVO XQ20 Diagnostic Ultrasound System; EVO XQ10 Diagnostic Ultrasound System; EVO QH30 Diagnostic Ultrasound System; EVO QH20 Diagnostic Ultrasound System; EVO QH10 Diagnostic Ultrasound System and transducers are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Dermatology, Trans-esophageal (Cardiac) and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals (including emergency rooms), private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode.

    Device Description

    EVO Q30, EVO Q20, EVO Q10, EVO XQ30, EVO XQ20, EVO XQ10, EVO QH30, EVO QH20, EVO QH10 Diagnostic Ultrasound System are general purpose, portable (without cart)/mobile (with cart), software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode.

    EVO Q30, EVO Q20, EVO Q10, EVO XQ30, EVO XQ20, EVO XQ10, EVO QH30, EVO QH20, EVO QH10 Diagnostic Ultrasound System also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals.

    EVO Q30, EVO Q20, EVO Q10, EVO XQ30, EVO XQ20, EVO XQ10, EVO QH30, EVO QH20, EVO QH10 Diagnostic Ultrasound System have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HERA Z20 Diagnostic Ultrasound System; HERA Z20e Diagnostic Ultrasound System; HERA Z20s Diagnostic Ultrasound System; R20 Diagnostic Ultrasound System; HERA Z30 Diagnostic Ultrasound System; R30 Diagnostic Ultrasound System and probes are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac), Peripheral vessel and Ophthalmic.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, M mode, Color Doppler mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, Power Doppler (PD) mode, ElastoScan Mode, MV-Flow Mode, Multi Image mode(Dual, Quad), Combined modes, 3D/4D mode.

    Device Description

    HERA Z20 Diagnostic Ultrasound System; HERA Z20e Diagnostic Ultrasound System; HERA Z20s Diagnostic Ultrasound System; R20 Diagnostic Ultrasound System; HERA Z30 Diagnostic Ultrasound System; R30 Diagnostic Ultrasound System are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode.

    HERA Z20 Diagnostic Ultrasound System; HERA Z20e Diagnostic Ultrasound System; HERA Z20s Diagnostic Ultrasound System; R20 Diagnostic Ultrasound System; HERA Z30 Diagnostic Ultrasound System; R30 Diagnostic Ultrasound System also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. HERA Z20 Diagnostic Ultrasound System; HERA Z20e Diagnostic Ultrasound System; HERA Z20s Diagnostic Ultrasound System; R20 Diagnostic Ultrasound System; HERA Z30 Diagnostic Ultrasound System; R30 Diagnostic Ultrasound System have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for each AI-ML based software feature described in the FDA 510(k) summary:

    Overview of Acceptance Criteria and Device Performance (AI-ML Features)

    The provided document details the testing and performance for several new and updated AI-ML based software features. The information given for each feature constitutes the acceptance criteria and the device's reported performance against those criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    AI/ML FeatureAcceptance CriterionReported Device Performance
    AbdomenAssist - Kidney Length MeasurementSuccess Rate: Within a pre-specified clinical error margin compared to reference standard.100% [94.17%, 100.00%] (Success Rate)
    Bland-Altman mean difference (Bias) / 95% Limits of Agreement (LoA).1.5% (Bias); [-4.53%, 7.64%] (95% LoA)
    AbdomenAssist - Spleen Length MeasurementSuccess Rate: Within a pre-specified clinical error margin compared to reference standard.96.77% [88.98%, 99.11%] (Success Rate)
    Bland-Altman mean difference (Bias) / 95% Limits of Agreement (LoA).2.8% (Bias); [-7.02%, 12.61%] (95% LoA)
    BladderAssist - Bladder Width Measurement (First Instance)Success Rate: Within a pre-specified clinical error margin compared to reference standard.95.24% [86.91%, 98.37%] (Success Rate)
    Bland-Altman mean difference (Bias) / 95% Limits of Agreement (LoA).-4.81% (Bias); [-12.94%, 3.31%] (95% LoA)
    BladderAssist - Bladder Width Measurement (Second Instance)Success Rate: Within a pre-specified clinical error margin compared to reference standard.96.83% [89.14%, 99.13%] (Success Rate)
    Bland-Altman mean difference (Bias) / 95% Limits of Agreement (LoA).-3.2% (Bias); [-11.86%, 5.47%] (95% LoA)
    QualityCheck - View Classification (Manual Images)Sensitivity: > 0.80 (implied, by meeting 0.85-1.00 range)Ranged from 0.85 to 1.00
    Specificity: > 0.80 (implied, by meeting 0.85-1.00 range)Ranged from 0.85 to 1.00
    Positive Predictive Value (PPV): > 0.80 (implied, by meeting 0.80-1.00 range)Ranged from 0.80 to 1.00
    Negative Predictive Value (NPV): > 0.90 (implied, by meeting 0.90-1.00 range)Ranged from 0.90 to 1.00
    QualityCheck - Structure Detection (Manual Images)Sensitivity: > 0.80 (implied, by meeting 0.81-1.00 range)Ranged from 0.81 to 1.00
    Specificity: > 0.99 (implied, by meeting 0.99-1.00 range)Ranged from 0.99 to 1.00
    Positive Predictive Value (PPV): > 0.89 (implied, by meeting 0.89-1.00 range)Ranged from 0.89 to 1.00
    Negative Predictive Value (NPV): > 0.99 (implied, by meeting 0.99-1.00 range)Ranged from 0.99 to 1.00
    QualityCheck - View Classification (LVA Images)Sensitivity: > 0.80 (implied, by meeting 0.86-1.00 range)Ranged from 0.86 to 1.00
    Specificity: > 0.80 (implied, by meeting 0.85-1.00 range)Ranged from 0.85 to 1.00
    Positive Predictive Value (PPV): > 0.80 (implied, by meeting 0.80-1.00 range)Ranged from 0.80 to 1.00
    Negative Predictive Value (NPV): > 0.91 (implied, by meeting 0.91-1.00 range)Ranged from 0.91 to 1.00
    QualityCheck - Structure Detection (LVA Images)Sensitivity: > 0.80 (implied, by meeting 0.82-1.00 range)Ranged from 0.82 to 1.00
    Specificity: = 1.00 (implied, by meeting 1.00-1.00 range)Ranged from 1.00 to 1.00
    Positive Predictive Value (PPV): > 0.89 (implied, by meeting 0.89-1.00 range)Ranged from 0.89 to 1.00
    Negative Predictive Value (NPV): > 0.99 (implied, by meeting 0.99-1.00 range)Ranged from 0.99 to 1.00
    PelvicAssist - Volume AlignmentAcceptance rate96.67%
    PelvicAssist - LH Measurement (ICC)Intraclass Correlation Coefficient (ICC) for six measurements.LH Area (0.9802), LH Circ. (0.9837), LH AP (0.9910), LH Lat. (0.9536), Right LUG (0.9423), Left LUG (0.9596)
    PelvicAssist - LH Measurement (Bland-Altman)Mean difference near zero; majority of data points within 95% LoA.Mean difference near zero (up to 1.1cm² and 0.53cm); 95.8% of data points fell within 95% LoA.
    EzVolume - Measurement TestBias (Mean Difference) for each label: Does not exceed ±2%.Did not exceed ±2%
    95% confidence interval for mean error includes zero.Included zero
    95% Limits of Agreement (LoA) for all labels: fell within ±15%.Fell within ±15%
    UterineAssist - Segmentation (Sagittal)Average Dice-score of uterus.96.7%
    UterineAssist - Segmentation (Transverse)Average Dice-score of uterus.95.8%
    UterineAssist - Segmentation (Endometrium)Average Dice-score of endometrium.86.8%
    UterineAssist - Feature Points Extraction (Uterus)Average error range of uterus feature points.1.5 – 2.6 mm
    UterineAssist - Feature Points Extraction (Endometrium)Average error range of endometrium feature points.0.9 - 1.7 mm
    UterineAssist - Size Measurement (Uterus)Average error range of Measurements.0.87 – 1.79 mm
    Widest 95% LoA range for uterus measurements; Largest Mean difference.[-2.96, 4.04] (widest 95% LoA); 1.23 mm (largest Mean difference)
    UterineAssist - Size Measurement (Endometrium)95% LoA range for endometrium measurements; Mean difference.[-1.59, 2.09] (95% LoA); 0.25 mm (mean difference)
    NerveTrack - DetectionLocalization accuracy success rate (95% CI); Processing speed.92.19% (95% CI: [90.03%, 94.34%]) (Success Rate); ~3.98 FPS

    2. Sample Sizes and Data Provenance

    AI/ML FeatureSample Size (Test Set)Data Provenance
    AbdomenAssist62 individual patients; 124 ultrasound images (62 kidney, 62 spleen)United States and Germany; Mix of retrospective and prospective
    BladderAssist63 individual patients; 63 ultrasound bladder transverse imagesUnited States and Germany; Mix of retrospective and prospective
    QualityCheck283 individual patients; 43,737 static 2D B-mode images (25,786 manual, 17,951 Live ViewAssist)United States; Mix of retrospective and prospective
    PelvicAssist40 individual patients; 120 volumes (40 rest, 40 contraction, 40 Valsalva)United States and Italy; Mix of retrospective and prospective
    EzVolume200 individual patients/3D volumes (100 1st trimester, 100 2nd/3rd trimesters)South Korea and United States; Mix of retrospective and prospective
    UterineAssist60 individual patients; 120 static images (60 sagittal, 60 transverse)South Korea and United States; Mix of retrospective and prospective
    NerveTrack46 individual patients; At least two nerve views per patient, with 2D sequences of at least 10 images. At least 24 and up to 42 ultrasound images for each of the 10 nerves.South Korea and United States; Mix of retrospective and prospective

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    AI/ML FeatureNumber of ExpertsQualifications of Experts
    AbdomenAssist3Two sonographers (one with >20 years exp., one with >10 years exp.); One senior expert radiologist (>20 years exp.)
    BladderAssist3Two sonographers (one with >20 years exp., one with >10 years exp.); One senior expert radiologist (>20 years exp.)
    QualityCheck3Two sonographers (each with >20 years exp.); One Obstetrician-Gynecologist (>10 years exp.)
    PelvicAssist3Three clinical experts (each with >20 years exp.)
    EzVolume3Three clinical experts (each with >10 years exp.)
    UterineAssist(At least) 2One sonographer (>10 years exp.) for view classification; Two sonographers (>10 years exp.) for manual drawing of anatomy areas/ground truth for validation images.
    NerveTrack3Two clinical experts (extensive experience in musculoskeletal ultrasound); One senior clinical expert (extensive experience in the field)

    4. Adjudication Method for the Test Set

    AI/ML FeatureAdjudication Method
    AbdomenAssist2+1 (2 independent measurements, 1 senior expert adjudication). The third senior expert reviewed and adjudicated the two measurements to determine the final value.
    BladderAssist2+1 (2 independent measurements, 1 senior expert adjudication). The third senior expert reviewed and adjudicated the two measurements to determine the final value.
    QualityCheckThe expert panel for the validation ground truth consisted of two sonographers, who performed the annotation, and an Obstetrician-Gynecologist, who provided the review and final confirmation. (Implies a 2+1 model, where the third expert reviews and confirms).
    PelvicAssistGTs were permuted and sent to the experts for peer review. Rejected data were re-labeled by the initial assigned expert, and the process is repeated. (This suggests an iterative consensus approach rather than a strict 2+1 or 3+1 structure initially, but aims for consensus among the 3 experts.)
    EzVolumeConsensus process of 3 experts. Initial annotation by one expert, then reviewed independently and blindly by the other two. If both accept, it's final. If any propose modifications, all three convene for unanimous agreement.
    UterineAssistFor images, two sonographers manually drew anatomy areas. (Implies agreement or an internal process, but not explicitly stated as 2+1 or 3+1. Ground truth was "made by sonographer" and then "manually drawn for each of the image by two sonographers" - implies dual annotation to establish GT.)
    NerveTrack2+1 (2 independent manual segmentations, 1 senior clinical expert adjudication). The senior expert resolved any discrepancies to establish the definitive ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with AI vs. without AI assistance for any of the features described. The studies focused on the standalone performance of the AI algorithms.

    6. Standalone (Algorithm Only) Performance

    Yes, standalone (algorithm only without human-in-the-loop performance) was done for all the AI/ML features described.

    • AbdomenAssist: Evaluated success rate and Bland-Altman agreement of the algorithm's measurements.
    • BladderAssist: Evaluated success rate and Bland-Altman agreement of the algorithm's measurements.
    • QualityCheck: Evaluated sensitivity, specificity, PPV, and NPV of the algorithm's view classification and structure detection.
    • PelvicAssist: Evaluated acceptance rate for volume alignment and ICC/Bland-Altman for LH measurement of the algorithm.
    • EzVolume: Evaluated error rate, bias, and LoA for the algorithm's measurements based on its segmentation results.
    • UterineAssist: Evaluated Dice-score for segmentation, average error range for feature point extraction, and error range/Bland-Altman for size measurements of the algorithm.
    • NerveTrack: Evaluated localization accuracy success rate and processing speed of the algorithm's detection.

    7. Type of Ground Truth Used

    AI/ML FeatureType of Ground Truth
    AbdomenAssistExpert consensus / manual measurement by clinical experts
    BladderAssistExpert consensus / manual measurement by clinical experts
    QualityCheckExpert consensus / classifications and annotations by clinical experts
    PelvicAssistExpert consensus / annotations by clinical experts
    EzVolumeExpert consensus / 3D segmentation annotation by clinical experts
    UterineAssistExpert consensus / manual segmentation and feature point annotation by sonographers
    NerveTrackExpert consensus / manual segmentation (ROI drawing) by clinical experts

    8. Sample Size for the Training Set

    The document explicitly states for each feature that "Data used for test and training/tuning purpose are completely separated from the ones during training process and there is no overlap between the two." or "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three."

    However, the specific sample sizes for the training sets are not provided in this FDA 510(k) summary document.

    9. How the Ground Truth for the Training Set was Established

    Similar to the training set size, the document does not explicitly describe how the ground truth for the training set was established. It only details the process for establishing the ground truth for the test/validation sets. The inference is that a similar expert-driven annotation process would have been used for training data, but the specifics are absent from this document.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    V8 Diagnostic Ultrasound System; cV8 Diagnostic Ultrasound System; V7 Diagnostic Ultrasound System; cV7 Diagnostic Ultrasound System; V6 Diagnostic Ultrasound System; cV6 Diagnostic Ultrasound System:

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.

    V5 Diagnostic Ultrasound System; cV5 Diagnostic Ultrasound System; V4 Diagnostic Ultrasound System; cV4 Diagnostic Ultrasound System:

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac), Peripheral vessel and Dermatology. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode, MV-Flow Mode.

    Device Description

    ■ V8/cV8, V7/cV7, V6/cV6
    The V8/cV8, V7/cV7, V6/cV6 Diagnostic Ultrasound Systems are general purpose, mobile, software controlled, diagnostic ultrasound systems. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V8/cV8, V7/cV7, V6/cV6 also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V8/cV8, V7/cV7, V6/cV6 have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    ■ V5/cV5, V4/cV4
    The V5/cV5, V4/cV4 Diagnostic Ultrasound Systems are general purpose, mobile, software controlled, diagnostic ultrasound systems. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S-Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes. The V5/cV5, V4/cV4 diagnostic ultrasound systems also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V5/cV5, V4/cV4 diagnostic ultrasound systems have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    The provided FDA 510(k) clearance letter describes the Samsung Medison Ultrasound Systems (V8/cV8, V7/cV7, V6/cV6, V5/cV5, V4/cV4) and their AI-powered features: BiometryAssist, ViewAssist, HeartAssist (fetus), and EzNerveMeasure of NerveTrack.

    Below is a breakdown of the acceptance criteria and study details for each AI feature, based on the provided text.


    BiometryAssist

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature AreaAcceptance Criteria (Threshold)Reported Device Performance
    SegmentationAverage Dice-score ≥ 0.8Average Dice-score = 0.869
    Size Measurement (Circumference)Error Rate ≤ 8%Error Rate = 8% or less
    Size Measurement (Distance)Error Rate ≤ 4%Error Rate = 4% or less
    Size Measurement (NT)Error Rate ≤ 1mmError Rate = 1mm or less

    2. Sample Size for Test Set and Data Provenance

    • Individual Patients: 52 (17 from South Korea, 35 from the US)
    • Images: 320 static images
    • Data Provenance: Mix of retrospective and prospective data collection from hospitals in South Korea and the United States.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: 3 (one obstetrician, two sonographers) + 1 senior obstetrician for supervision.
    • Qualifications:
      • One obstetrician with over 20 years of experience in fetal cardiology.
      • Two sonographers, each with more than 10 years of experience.
      • One senior obstetrician with over 25 years of clinical experience (supervised and corrected annotations).

    4. Adjudication Method for Test Set

    • Expert reviewers manually performed view classification and measurements following the same protocol as used during actual device operation. The entire process was supervised by a senior obstetrician who reviewed and corrected all annotations to ensure consistency and accuracy. This implies an expert consensus-based adjudication, likely 3+1 or similar, though not explicitly stated as a formal (X+Y) method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not explicitly mentioned. The study focuses on standalone algorithm performance against expert ground truth.

    6. Standalone Performance (Algorithm Only)

    • Yes, performance metrics (Dice-score, error rates) are reported for the algorithm's output compared to established ground truth.

    7. Type of Ground Truth Used

    • Expert Consensus: Ground truth was established by expert reviewers manually classifying views and performing measurements, supervised and corrected by a senior obstetrician. This closely reflects the clinical use environment.

    8. Sample Size for Training Set

    • Not specified. The document only states that data used for training, tuning, and validation were completely separated.

    9. How Ground Truth for Training Set Was Established

    • All images used for training and tuning were first classified into the correct ultrasound views by three participating experts. Following view classification, the corresponding anatomical structures were manually annotated for each image.

    ViewAssist

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature AreaAcceptance Criteria (Threshold)Reported Device Performance
    View Recognition (Sensitivity)Not explicitly stated (thresholds 75.9%, 88.2% defined)Achieved sensitivity = 93.97%
    View Recognition (Specificity)Not explicitly stated (thresholds 75.9%, 88.2% defined)Achieved specificity = 99.62%
    SegmentationAverage Dice-score ≥ 0.8Average Dice-score = 0.863

    2. Sample Size for Test Set and Data Provenance

    • Individual Patients: 102 (42 from South Korea, 60 from the US)
    • Images: 680 static images
    • Data Provenance: Mix of retrospective and prospective data collection from hospitals in South Korea and the United States.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: 3 (one obstetrician, two sonographers) + 1 senior obstetrician for supervision.
    • Qualifications:
      • One obstetrician with over 20 years of experience in fetal cardiology.
      • Two sonographers, each with more than 10 years of experience.
      • One senior obstetrician with over 25 years of clinical experience (supervised and corrected annotations).

    4. Adjudication Method for Test Set

    • Expert reviewers manually performed view classification and measurements following the same protocol as used during actual device operation. The entire process was supervised by a senior obstetrician who reviewed and corrected all annotations to ensure consistency and accuracy. This implies an expert consensus-based adjudication, likely 3+1 or similar.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not explicitly mentioned. The study focuses on standalone algorithm performance against expert ground truth.

    6. Standalone Performance (Algorithm Only)

    • Yes, performance metrics (sensitivity, specificity, Dice-score) are reported for the algorithm's output compared to established ground truth.

    7. Type of Ground Truth Used

    • Expert Consensus: Ground truth was established by expert reviewers manually classifying views and performing measurements, supervised and corrected by a senior obstetrician. This closely reflects the clinical use environment.

    8. Sample Size for Training Set

    • Not specified. The document only states that data used for training, tuning, and validation were completely separated.

    9. How Ground Truth for Training Set Was Established

    • All images used for training and tuning were first classified into the correct ultrasound views by three participating experts. Following view classification, the corresponding anatomical structures were manually annotated for each image.

    HeartAssist (fetus)

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature AreaAcceptance Criteria (Threshold)Reported Device Performance
    View Recognition (Sensitivity)Not explicitly stated (thresholds 75.9%, 88.2% defined)Achieved sensitivity = 94.29%
    View Recognition (Specificity)Not explicitly stated (thresholds 75.9%, 88.2% defined)Achieved specificity = 99.62%
    SegmentationAverage Dice-score ≥ 0.8Average Dice-score = 0.865
    Size Measurement (Area)Error Rate ≤ 8%Error Rate = 8% or less
    Size Measurement (Angle)Error Rate ≤ 4%Error Rate = 4% or less
    Size Measurement (Circumference)Error Rate ≤ 11%Error Rate = 11% or less
    Size Measurement (Diameter)Error Rate ≤ 11%Error Rate = 11% or less

    2. Sample Size for Test Set and Data Provenance

    • Individual Patients: 70 (26 from South Korea, 44 from the US)
    • Images: 280 static images
    • Data Provenance: Mix of retrospective and prospective data collection from hospitals in South Korea and the United States.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: 3 (one obstetrician, two sonographers) + 1 senior obstetrician for supervision.
    • Qualifications:
      • One obstetrician with over 20 years of experience in fetal cardiology.
      • Two sonographers, each with more than 10 years of experience.
      • One senior obstetrician with over 25 years of clinical experience (supervised and corrected annotations).

    4. Adjudication Method for Test Set

    • Expert reviewers manually performed view classification and measurements following the same protocol as used during actual device operation. The entire process was supervised by a senior obstetrician who reviewed and corrected all annotations to ensure consistency and accuracy. This implies an expert consensus-based adjudication, likely 3+1 or similar.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not explicitly mentioned. The study focuses on standalone algorithm performance against expert ground truth.

    6. Standalone Performance (Algorithm Only)

    • Yes, performance metrics (sensitivity, specificity, Dice-score, error rates) are reported for the algorithm's output compared to established ground truth.

    7. Type of Ground Truth Used

    • Expert Consensus: Ground truth was established by expert reviewers manually classifying views and performing measurements, supervised and corrected by a senior obstetrician. This closely reflects the clinical use environment.

    8. Sample Size for Training Set

    • Not specified. The document only states that data used for training, tuning, and validation were completely separated.

    9. How Ground Truth for Training Set Was Established

    • All images used for training and tuning were first classified into the correct ultrasound views by three participating experts. Following view classification, the corresponding anatomical structures were manually annotated for each image.

    EzNerveMeasure of NerveTrack

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature AreaAcceptance Criteria (Threshold)Reported Device Performance
    Flattening Ratio (FR)Not explicitly statedAverage error rate = 8.05% (95% CI: [7.64, 8.46]), Std Dev = 0.87
    Cross-Sectional Area (CSA)Not explicitly statedAverage error rate = 13.11% (95% CI: [11.83, 14.38]), Std Dev = 2.73

    2. Sample Size for Test Set and Data Provenance

    • Individual Patients: 20 (10 from South Korea, 10 from the US)
    • Images: 200 static images (10 static images per patient from a 2D sequence)
    • Data Provenance: Prospective data collected in clinical practice using Samsung ultrasound systems (RS80A and V8) from South Korea and the United States.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: 3 anesthesiologists + 1 senior anesthesiologist for resolution.
    • Qualifications: All three anesthesiologists had over 10 years of experience. The senior anesthesiologist had over 10 years of extensive clinical experience in regional anesthesia and ultrasound-guided procedures.

    4. Adjudication Method for Test Set

    • The ground truth (GT) for median nerve locations was initially drawn by one anesthesiologist performing the ultrasound scans, then verified by the other two anesthesiologists. Disagreements were resolved by a senior anesthesiologist. This is a consensus-based adjudication process, effectively a 3+1 method for resolving disagreements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not explicitly mentioned. The study focuses on standalone algorithm performance against expert ground truth.

    6. Standalone Performance (Algorithm Only)

    • Yes, performance metrics (average error rate, confidence interval, standard deviation) are reported for the algorithm's output compared to established ground truth.

    7. Type of Ground Truth Used

    • Expert Consensus: Ground truth (median nerve locations, FR, CSA measurements) was established by three anesthesiologists, with disagreements resolved by a senior anesthesiologist. A clinical evaluation was also conducted by experienced doctors to assess clinical significance.

    8. Sample Size for Training Set

    • Not specified. The document only states that data used for training, tuning, and validation were completely separated.

    9. How Ground Truth for Training Set Was Established

    • Not explicitly detailed for the training set itself, but it implies a similar expert annotation process as described for validation, with clear separation between data sets.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative. Pediatric, Small Organ, Neonatal Cephalic, Trans-rectal, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode. Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode. Tissue Doppler Imaging (TDI) mode. Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.

    Device Description

    The V8/cV8, V7/cV7, V6/cV6 are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V8/cV8, V7/cV7, V6/cV6 also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V8/cV8, V7/cV7, V6/cV6 have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    The provided text describes the acceptance criteria and study proving the device meets those criteria, specifically for the 'EzNerveMeasure' functionality of the V8/cV8, V7/cV7, V6/cV6 Diagnostic Ultrasound System.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    MetricAcceptance Criteria (Implicit)Reported Device Performance
    Flattening Ratio (FR) Error RateNot explicitly stated an acceptance criterion, but implicitly that the performance is acceptable for clinical use.Average: 8.31% (95% CI: [7.29, 9.34]) Standard Deviation: 5.22
    Cross-Sectional Area (CSA) Error RateNot explicitly stated an acceptance criterion, but implicitly that the performance is acceptable for clinical use.Average: 13.12% (95% CI: [10.90, 15.34]) Standard Deviation: 11.33

    Note: The document states, "We tested on the flattening ratio (FR) and cross-sectional area (CSA) of NerveTrack EzNerveMeasure." and then presents the average error rates. While explicit acceptance criteria values are not given (e.g., "FR error < 10%"), the presentation of these results implies that they are considered acceptable for the device's intended use and substantial equivalence claim.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: 100 static images.
      • These 100 images were derived from 10 individual patients, with 10 static images randomly selected from each case.
    • Data Provenance:
      • Country of Origin: Seoul National University (implying South Korea), with patients identified as "Koreans."
      • Retrospective or Prospective: Mix of data from retrospective and prospective data collection in clinical practice.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Three participating experts.
    • Qualifications of Experts:
      • Initially, an anesthesiologist with more than 10 years of experience in pain management manually drew the nerve areas.
      • For verification of the ground truth (GT), "other doctors with more than 10 years of experience" checked every frame of each scanned sequence.

    4. Adjudication method for the test set

    • Adjudication Method: Consensus.
      • The document states, "If they did not agree on median nerve locations, necessary corrections were made to make the final GT." This indicates a process where disagreements among experts were reconciled to reach a final, agreed-upon ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • An MRMC comparative effectiveness study was not performed and is not mentioned in the provided text. The evaluation focuses on the algorithm's performance against expert-defined ground truth, rather than human reader improvement with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone evaluation was done. The reported error rates for FR and CSA (8.31% and 13.12% respectively) represent the performance of the deep learning-based segmentation algorithm independently against the established ground truth.

    7. The type of ground truth used

    • Type of Ground Truth: Expert Consensus / Manual Annotation.
      • The ground truth (GT) data was "manually drawn by an anesthesiologist with more than 10 years of experience in pain management."
      • This was then "verified" by "other doctors with more than 10 years of experience," who performed "necessary corrections... to make the final GT" in cases of disagreement. This indicates an expert consensus approach to establishing the ground truth.

    8. The sample size for the training set

    • The exact sample size for the training set is not specified. The document only mentions that images were collected for "training, tuning and validation." It does not provide the specific number of images or cases used for training.

    9. How the ground truth for the training set was established

    • The method for establishing ground truth for the training set is described as: "Nerve areas in all acquired images for training, tuning and validation were manually drawn by an anesthesiologist with more than 10 years of experience in pain management."
    • Similar to the validation set, "For verification of GT, other doctors with more than 10 years of experience checked every frame of each scanned sequences. If they did not agree on median nerve locations, necessary corrections were made to make the final GT."
    • This suggests an expert manual annotation and consensus verification process was used for the entire dataset, including data destined for training.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Peripheral vessel and Dermatology.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode, MV-Flow Mode.

    Device Description

    V5 Diagnostic Ultrasound System; H5 Diagnostic Ultrasound System; XV5 Diagnostic Ultrasound System: XH5 Diagnostic Ultrasound System; V4 Diagnostic Ultrasound System; H4 Diagnostic Ultrasound System: XV4 Diagnostic Ultrasound System; XH4 Diagnostic Ultrasound System are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode. The V5/H5/XV5/XH5, V4/H4/XV4/XH4 diagnostic ultrasound system also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V5/H5/XV5/XH5, V4/H4/XV4/XH4 diagnostic ultrasound system have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and supporting study details for each AI-powered feature of the V5/H5/XV5/XH5 and V4/H4/XV4/XH4 Diagnostic Ultrasound Systems, as provided in the document:


    HeartAssist (Fetal and Adult)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureModalityAcceptance CriteriaReported Performance
    View RecognitionFetalAverage recognition accuracy ≥ 89%93.21%
    AdultAverage recognition accuracy ≥ 84%98.31%
    SegmentationFetalAverage Dice-score ≥ 0.80.88
    AdultAverage Dice-score ≥ 0.90.93
    Size MeasurementFetalError rate of area measured value ≤ 8%≤ 8%
    (Error Rate)FetalError rate of angle measured value ≤ 4%≤ 4%
    FetalError rate of circumference measured value ≤ 11%≤ 11%
    FetalError rate of diameter measured value ≤ 11%≤ 11%
    Size MeasurementAdultPCC value (B-mode) ≥ 0.8Passed (PCC ≥ 0.8)
    (Pearson Correlation)AdultPCC value (M-mode) ≥ 0.8Passed (PCC ≥ 0.8)
    AdultPCC value (Doppler-mode) ≥ 0.8Passed (PCC ≥ 0.8)

    2. Sample Sizes and Data Provenance:

    • Test Set (Fetal):

      • Number of individuals: 80
      • Number of static images: 280
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from five hospitals.
      • Country of Origin: Not explicitly stated, but clinical experts are from multiple countries including US and Korea, suggesting data could be from these regions.
    • Test Set (Adult):

      • Number of individuals: 30
      • Number of static images: 540
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from five hospitals.
      • Country of Origin: Not explicitly stated, but clinical experts are from multiple countries including US and Korea, suggesting data could be from these regions.

    3. Number of Experts and Qualifications for Ground Truth:

    • Fetal:

      • 3 participating experts:
        • 1 obstetrician with >20 years of experience (in fetal cardiology)
        • 2 sonographers with >10 years of experience (in fetal cardiology)
      • Supervised by: 1 obstetrician with >25 years of experience.
    • Adult:

      • 4 experts:
        • 2 cardiologists with at least 10 years of experience
        • 2 sonographers with at least 10 years of experience.

    4. Adjudication Method:

    • Fetal: All acquired images were first classified into correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn. The entire process was supervised by another obstetrician.
    • Adult: Experts manually traced the contours of the heart and the signal outline on the images. (Implicitly, this suggests an expert consensus or expert-defined ground truth, but a specific adjudication method like "2+1" is not detailed.)

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned for evaluating human readers with and without AI assistance for HeartAssist. The study focused on standalone performance metrics against ground truth.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (recognition accuracy, Dice-score, error rates, PCC values) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-defined outlines. For fetal images, experts classified views and manually drew anatomy areas. For adult images, cardiologists and sonographers manually traced contours. The PCC for adult size measurement was calculated against "the cardiologist's measurements."

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • Not explicitly stated, but it can be inferred that a similar expert-based process was used for generating ground truth for training data, as noted for the validation data: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn..."

    BiometryAssist (Fetal)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    SegmentationAverage Dice-score ≥ 0.80.919
    Size Measurement (Circumference)Error rate ≤ 8%≤ 8%
    Size Measurement (Distance)Error rate ≤ 4%≤ 4%
    Size Measurement (NT)Error rate ≤ 1mm≤ 1mm

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 77
      • Number of static images: 320
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from two hospitals.
      • Country of Origin: Americans and Koreans, suggesting data could be from these regions.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 obstetrician with >20 years of experience (in fetal cardiology)
      • 2 sonographers with >10 years of experience (in fetal cardiology)
    • Supervised by: 1 obstetrician with >25 years of experience.

    4. Adjudication Method:

    • All acquired images were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each image. The entire process was supervised by another obstetrician.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (Dice-score, error rates) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-defined outlines. Experts classified views and manually drew anatomy areas.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn..."

    ViewAssist (Fetal)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    View RecognitionAverage accuracy ≥ 89%94.26%
    Anatomy Annotation (Segmentation)Average Dice-score ≥ 0.80.885

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 77
      • Number of static images: 680
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from two hospitals.
      • Country of Origin: Americans and Koreans, suggesting data could be from these regions.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 obstetrician with >20 years of experience (in fetal cardiology)
      • 2 sonographers with >10 years of experience (in fetal cardiology)
    • Supervised by: 1 obstetrician with >25 years of experience.

    4. Adjudication Method:

    • All acquired images were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each image. The entire process was supervised by another obstetrician.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (recognition accuracy, Dice-score) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-defined outlines. Experts classified views and manually drew anatomy areas.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn..."

    UterineAssist

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    Segmentation (Uterus)Average Dice-score not explicitly stated, but 96% is reported.96%
    Segmentation (Endometrium)Average Dice-score not explicitly stated, but 92% is reported.92%
    Feature Points Extraction (Uterus)Errors ≤ 5.8 mm≤ 5.8 mm
    Feature Points Extraction (Endometrium)Errors ≤ 4.3 mm≤ 4.3 mm
    Size MeasurementErrors ≤ 2.0 mm≤ 2.0 mm

    2. Sample Sizes and Data Provenance:

    • Test Set (Segmentation):

      • Number of static images: 450 sagittal uterus images and 150 transverse uterus images (total 600 images).
      • Number of individuals: 60 contributed to the validation dataset (for segmentation AND feature points/size measurement)
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from three hospitals.
      • Country of Origin: All Koreans.
    • Test Set (Feature Points Extraction & Size Measurement):

      • Number of static images: 48 sagittal and 44 transverse plane images of the uterus (total 92 images).
      • Number of individuals: 60 individuals contributed to the validation dataset (same as segmentation).
      • Provenance: Mix of retrospective and prospective data collection in clinical practice from three hospitals.
      • Country of Origin: All Koreans.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating OB/GYN experts with >10 years' experience.

    4. Adjudication Method:

    • Segmentation of the ground truth was generated by three participating OB/GYN experts. The image set was divided into three subsets, and each expert drew ground truths for one subset. Ground truths drawn by one expert were cross-checked by the other two experts. Any images not meeting inclusion/exclusion criteria were excluded.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported performance metrics (Dice-score, error rates) are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-drawn contours and measurements.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for test/training validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "Segmentation of the ground truth was generated by three participating OB/GYN experts with more than 10 years' experience." This process for ground truth establishment is also applicable to training data generation.

    NerveTrack (Nerve Detection)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    DetectionAverage accuracy from 10 image sequence not explicitly stated as an acceptance criterion, but 89.91% is reported. A detection is considered correct if Dice coefficient is ≥ 0.5.Average accuracy from 10 image sequence: 89.91% (95% CI: 86.51, 93.35)
    SpeedNot explicitly stated as an acceptance criterion, but 3.98 fps is reported.Average speed (fps): 3.98 (95% CI: 3.98, 3.99)

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 22
      • Number of images: 3,999 (extracted from 2D sequences, with each individual contributing at least ten images per sequence).
      • Provenance: Prospective data collected in clinical practice from eight hospitals.
      • Country of Origin: Koreans.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 anesthesiologist with >10 years of experience in pain management (for drawing GT).
      • "Other doctors" with >10 years of experience (for GT verification).
      • Doctors who scanned the ultrasound were directly involved in GT construction.

    4. Adjudication Method:

    • Manual drawing of nerve areas by an anesthesiologist. For verification of GT, "other doctors" checked every frame. If they disagreed, corrections were made to finalize the GT.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported accuracy and speed are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-annotated rectangular regions for nerve locations.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "The GT data were built by three participating experts. Nerve areas in all acquired images for training, tuning and validation were manually drawn by an anesthesiologist... For verification of GT, other doctors with more than 10 years of experience checked every frame... corrections were made to make the final GT."

    NerveTrack (Nerve Segmentation)

    1. Table of Acceptance Criteria and Reported Device Performance:

    FeatureAcceptance CriteriaReported Performance
    SegmentationAverage accuracy from nine image sequences not explicitly stated as an acceptance criterion, but 98.30% is reported. A segmentation is considered correct if Dice coefficient is ≥ 0.5.Average accuracy from nine image sequences: 98.30% (95% CI: 95.43, 100)
    SpeedNot explicitly stated as an acceptance criterion, but 3.98 fps is reported.Average speed (fps): 3.98 (95% CI: 3.98, 3.98)

    2. Sample Sizes and Data Provenance:

    • Test Set:
      • Number of individuals: 20
      • Number of images: 1,675 (extracted from 2D sequences, with each individual contributing at least ten images per sequence).
      • Provenance: Prospective data collected in clinical practice from ten hospitals.
      • Country of Origin: Koreans.

    3. Number of Experts and Qualifications for Ground Truth:

    • 3 participating experts:
      • 1 anesthesiologist with >10 years of experience in pain management (for drawing GT).
      • "Other doctors" with >10 years of experience (for GT verification).
      • Doctors who scanned the ultrasound were directly involved in GT construction.

    4. Adjudication Method:

    • Manual drawing of nerve areas (contours) by an anesthesiologist. For verification of GT, "other doctors" checked every frame. If they disagreed on nerve and other organ contours, corrections were made to finalize the GT.

    5. MRMC Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly mentioned.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The reported accuracy and speed are direct measurements of the algorithm's performance against expert-defined ground truth.

    7. Type of Ground Truth Used:

    • Expert consensus / Expert-annotated contours for nerve regions.

    8. Sample Size for Training Set:

    • Not explicitly stated ("Data used for training, tuning and validation purpose are completely separated").

    9. How Ground Truth for Training Set Was Established:

    • It's inferred that a similar expert-based process was used as for validation data: "The GT data were built by three participating experts. Nerve areas in all acquired images for training, tuning and validation were manually drawn by an anesthesiologist... For verification of GT, other doctors with more than 10 years of experience checked every frame... corrections were made to make the final GT."
    Ask a Question

    Ask a specific question about this device

    K Number
    K242444

    Validate with FDA (Live)

    Date Cleared
    2024-11-27

    (103 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan+™ Mode, Combined modes, Multi-Image modes (Dual, Quad), 3D/4D modes.

    Device Description

    The HERA W9/ HERA W10 are general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as Bmode, M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S- Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes.

    The HERA W9/HERA W10 also give the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W9/HERA W10 have real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    The provided FDA 510(k) summary (K242444) describes the acceptance criteria and study proving the device meets these criteria for three AI-powered features: HeartAssist, BiometryAssist, and ViewAssist, as well as a non-AI feature, SonoSync.

    Here's a breakdown of the requested information for each AI feature:

    HeartAssist

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    View recognition accuracy89%96.07%
    Segmentation dice-score0.80.88
    Size measurement (area) error rate8% or less8% or less
    Size measurement (angle) error rate4% or less4% or less
    Size measurement (circumference) error rate11% or less11% or less
    Size measurement (diameter) error rate11% or less11% or less

    2. Sample size used for the test set and data provenance

    • Individuals: 69
    • Static Images: 315 (at least 1 static image per view location per individual)
    • Provenance: Data collected at two hospitals in the United States and South Korea. Mixed retrospective and prospective data collection.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    • Number of Experts: Three active participating experts for initial classification and manual drawing, supervised by one additional expert.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (primary classification/drawing).
      • Two sonographers with more than 10 years of experience in fetal cardiology (primary classification/drawing).
      • One obstetrician with more than 25 years of experience (supervising the entire process).

    4. Adjudication method for the test set

    Not explicitly stated. The process mentions "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the images." It doesn't specify if there was a consensus mechanism or independent review and adjudication if the experts disagreed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the provided data describes the standalone performance of the HeartAssist algorithm against established ground truth.

    7. The type of ground truth used

    Expert consensus based on manual classification of views and manual drawing of corresponding anatomy areas.

    8. The sample size for the training set

    Not explicitly stated for HeartAssist, but it is mentioned that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three."

    9. How the ground truth for the training set was established

    Not explicitly detailed for the training set, but it is implied to be the same method as for the validation set: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the images."


    BiometryAssist

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Segmentation dice-score0.80.91
    Size measurement (circumference) error rate8% or less8% or less
    Size measurement (distance) error rate4% or less4% or less
    Size measurement (NT, NB, IT) error rate1mm or less1mm or less

    2. Sample size used for the test set and data provenance

    • Individuals: 33
    • Static Images: 360 (at least 1 static image per view location per individual)
    • Provenance: Data collected at two hospitals in South Korea and the United States. Mixed retrospective and prospective data collection.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    • Number of Experts: Three active participating experts for initial classification and manual drawing, supervised by one additional expert.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (primary classification/drawing).
      • Two sonographers with more than 10 years of experience in fetal cardiology (primary classification/drawing).
      • One obstetrician with more than 25 years of experience (supervising the entire process).

    4. Adjudication method for the test set

    Not explicitly stated. The process mentions "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image." It doesn't specify if there was a consensus mechanism or independent review and adjudication if the experts disagreed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the provided data describes the standalone performance of the BiometryAssist algorithm against established ground truth.

    7. The type of ground truth used

    Expert consensus based on manual classification of views and manual drawing of corresponding anatomy areas.

    8. The sample size for the training set

    Not explicitly stated for BiometryAssist, but it is mentioned that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three."

    9. How the ground truth for the training set was established

    Not explicitly detailed for the training set, but it is implied to be the same method as for the validation set: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image."


    ViewAssist

    1. Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    View recognition accuracy89%94.92%
    Anatomy annotation (segmentation) dice-score0.80.89

    2. Sample size used for the test set and data provenance

    • Individuals: 98
    • Static Images: 1,485 (at least 1 static image per view location per individual)
    • Provenance: Data collected at two hospitals in South Korea and the United States. Mixed retrospective and prospective data collection.

    3. Number of experts used to establish the ground truth for the test set and their qualifications

    • Number of Experts: Three active participating experts for initial classification and manual drawing, supervised by one additional expert.
    • Qualifications:
      • One obstetrician with more than 20 years of experience (primary classification/drawing).
      • Two sonographers with more than 10 years of experience in fetal cardiology (primary classification/drawing).
      • One obstetrician with more than 25 years of experience (supervising the entire process).

    4. Adjudication method for the test set

    Not explicitly stated. The process mentions "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image." It doesn't specify if there was a consensus mechanism or independent review and adjudication if the experts disagreed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the provided data describes the standalone performance of the ViewAssist algorithm against established ground truth.

    7. The type of ground truth used

    Expert consensus based on manual classification of views and manual drawing of corresponding anatomy areas.

    8. The sample size for the training set

    Not explicitly stated for ViewAssist, but it is mentioned that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap between the three."

    9. How the ground truth for the training set was established

    Not explicitly detailed for the training set, but it is implied to be the same method as for the validation set: "All acquired images for training, tuning and validation were first classified into the correct views by three participating experts. Afterwards, corresponding anatomy areas were manually drawn for each of the image."


    General Notes applicable to all AI features:

    • No Multi-Reader Multi-Case (MRMC) Study: The document explicitly states that "The subject of this premarket submission, HERA W9/ HERA W10, did not require clinical studies to demonstrate the substantial equivalence." The studies described are technical performance evaluations of the AI algorithms, not comparative effectiveness studies with human readers.
    • Ground Truth Consistency: For all three AI features, the ground truth establishment process is described identically, relying on a small panel of experienced experts.
    • Independence of Data: For all three AI features, it is stated that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three."
    • Demographics: For all three AI features, the demographic distribution of the validation dataset indicates Female Gender, Reproductive age (specific age not collected), and Ethnicity/Country as Not Available / United States and South Korea. ICUOG and AIUM guidelines were used to divide fetal ultrasound images into views. BMI and Gestational Age distributions are also provided.
    • Equipment: Data was acquired with SAMSUNG MEDISON's ultrasound systems (HERA W9/HERA W10) to secure diversity.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241971

    Validate with FDA (Live)

    Date Cleared
    2024-10-11

    (98 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode. Color Doppler mode. Pulsed Wave (PW) Doppler mode. Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, Power Doppler (PD) mode, ElastoScan™ Mode, MV-Flow Mode, Multi Image mode (Dual Quad) Combined modes 3D/AD model

    Device Description

    The HERA Z20, R20, HERA Z30, R30 diagnostic ultrasound system are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode.

    The HERA Z20, R20, HERA Z30, R30 diagnostic ultrasound system also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA Z20, R20, HERA Z30, R30 diagnostic ultrasound system have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the studies conducted for the AI-powered features of the HERA Z20, R20, HERA Z30, R30 Diagnostic Ultrasound System, based on the provided text.

    This document describes several AI-powered features: Live ViewAssist, EzVolume, UterineContour, ViewAssist, HeartAssist, and BiometryAssist. Each feature has its own acceptance criteria and study findings.


    1. A table of acceptance criteria and the reported device performance

    Note: Some performance metrics were not explicitly stated as "acceptance criteria" but rather as "summary test statistics or other test results," indicating the device's measured performance against implicit or internal targets.

    AI FeatureMetricAcceptance CriteriaReported Device Performance
    Live ViewAssistQuality assessment (Cohen's kappa)Threshold 0.7Average Cohen's kappa coefficient: 0.818
    Time/duration (Frames per Second - FPS)Threshold 20 FPSAverage speed: 30.06 FPS
    EzVolumeAcceptance Rate (segmentation)Higher than 70% for each label1st Trimester: Fluid 98%, Fetus 96%, Umbilical-cord 80%, Placenta 86%, Uterus 89% 2nd/3rd Trimester: Fluid 92%, Head 94%, Body 84%, Limbs 83%, Umbilical-cord 82%, Placenta 85%, Uterus 87%
    Mean DSC (segmentation)(No explicit numerical criterion provided, but correlation with acceptance rate indicates adequacy)1st Tri (Accepted): Fluid 0.96, Fetus 0.91, Umbilical-cord 0.68, Placenta 0.74, Uterus 0.93 1st Tri (Rejected): Fluid 0.17, Fetus 0.55, Umbilical-cord 0.37, Placenta 0.33, Uterus 0.32 2nd/3rd Tri (Accepted): Fluid 0.78, Head 0.94, Body 0.68, Umbilical-cord 0.67, Limbs 0.66, Placenta 0.75, Uterus 0.80 2nd/3rd Tri (Rejected): Fluid 0.25, Head 0.46, Body 0.29, Umbilical-cord 0.38, Limbs 0.39, Placenta 0.32, Uterus 0.30
    UterineContourSegmentation (uterus Dice-score)Not explicitly stated as acceptance criteriaAverage dice-score of uterus: 96%
    Segmentation (endometrium Dice-score)Not explicitly stated as acceptance criteriaAverage dice-score of endometrium: 92%
    3D coronal view adaptationProportion of appropriateness evaluated as clinically diagnosable, with over 90% of all casesOver 90% of all cases
    ViewAssistView recognition accuracyThreshold 89%Average recognition accuracy: 94.50%
    Anatomy annotation (Dice-score)Threshold 0.8Average Dice-score: 0.892
    HeartAssistView recognition accuracyThreshold 89%Average recognition accuracy: 95.00%
    Segmentation (Dice-score)Threshold 0.8Average Dice-score: 0.876
    Size measurement (Area error rate)Not explicitly stated as acceptance criteria8% or less
    Size measurement (Angle error rate)Not explicitly stated as acceptance criteria4% or less
    Size measurement (Circumference error rate)Not explicitly stated as acceptance criteria11% or less
    Size measurement (Diameter error rate)Not explicitly stated as acceptance criteria11% or less
    BiometryAssistSegmentation (Dice-score)Threshold 0.8Average Dice-score: 0.928
    Size measurement (Circumference error rate)Not explicitly stated as acceptance criteria8% or less
    Size measurement (Distance error rate)Not explicitly stated as acceptance criteria4% or less
    Size measurement (NT, NB, IT error rate)Not explicitly stated as acceptance criteria1mm or less

    2. Sample size used for the test set and the data provenance

    AI FeatureTest Set Sample SizeData Provenance
    Live ViewAssist3,900 fetal ultrasound imagesMix of retrospective and prospective data collection in clinical practice from Americans and Koreans (gender: female, reproductive age; BMI 17-45.4)
    EzVolume200 test volumes (100 in 1st trimester, 100 in 2nd/3rd trimester)Mix of retrospective and prospective data collection in clinical practice from Koreans, Americans, Italians, and British (gender: female, reproductive age)
    UterineContour450 sagittal uterus images (for segmentation) and 30 sagittal images (for 3D coronal view)Mix of retrospective and prospective data collection in clinical practice from three hospitals in Korea (gender: female, reproductive age)
    ViewAssist1,600 fetal ultrasound and fetal biometry imagesMix of retrospective and prospective data collection in clinical practice from two hospitals in America and Korea (gender: female, reproductive age; BMI 17-45.4)
    HeartAssist440 fetal heart imagesMix of retrospective and prospective data collection in clinical practice from America and Korea (gender: female, reproductive age; BMI 17-45.4)
    BiometryAssist360 fetal biometry imagesMix of retrospective and prospective data collection in clinical practice from two hospitals in America and Korea (gender: female, reproductive age; BMI 17-45.4)

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    AI FeatureNumber of ExpertsQualifications of Experts
    Live ViewAssist3 primary experts, 1 supervising expertAn obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor).
    EzVolume4 primary experts, 1 supervising expertAn obstetrician with more than 20 years of experience (primary). Three examiners (clinical experts) with more than 10 years of experience, all in fetal diagnosis (primary). Another obstetrician with more than 25 years of experience (supervisor).
    UterineContour3 OB/GYN expertsThree participating OB/GYN experts with more than 10 years' experience.
    ViewAssist3 primary experts, 1 supervising expertAn obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor).
    HeartAssist3 primary experts, 1 supervising expertAn obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor).
    BiometryAssist3 primary experts, 1 supervising expertAn obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor).

    4. Adjudication method for the test set

    AI FeatureAdjudication Method
    Live ViewAssistGround truth established by consensus of 3 experts, supervised by 1. Exact method (e.g., 2+1, 3+1) not explicitly detailed, but implied by "manual drawing" and "classified into acceptable and not-acceptable views by three participating experts."
    EzVolumeGround truths were drawn manually by four participating clinical experts, supervised by one.
    UterineContourEach of the 3 experts delineated structures. Conflicts were resolved by a consensus of the three experts ("fixed the wrong part with consensus").
    ViewAssistGround truth established by consensus of 3 experts, supervised by 1.
    HeartAssistGround truth established by consensus of 3 experts, supervised by 1.
    BiometryAssistGround truth established by consensus of 3 experts, supervised by 1.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly stated as performed to compare human readers with and without AI assistance. The studies described focus on the standalone performance of the AI algorithms.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, standalone performance studies of the AI algorithms were done for all the described features: Live ViewAssist, EzVolume, UterineContour, ViewAssist, HeartAssist, and BiometryAssist. The reported metrics like Cohen's kappa, FPS, acceptance rates, Dice scores, and error rates are all measures of the algorithm's performance without human intervention during the assessment phase (though human experts were used to establish ground truth).


    7. The type of ground truth used

    For all features, the ground truth was established by expert consensus based on manual classification, delineation, or drawing by qualified clinical experts (obstetricians, sonographers, and examiners).


    8. The sample size for the training set

    The document explicitly states that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three." However, the exact sample size for the training set itself is not provided for any of the features. The sample sizes listed in Section 2 are for the test/validation sets.


    9. How the ground truth for the training set was established

    For all features, the ground truth for the training set (and validation/evaluation sets) was established through manual classification, delineation, or drawing by the same groups of qualified clinical experts mentioned in section 3, following similar expert consensus processes as described for the test sets. For UterineContour, initial delineations by 3 experts were then reviewed and fixed with consensus for unmatched results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241302

    Validate with FDA (Live)

    Device Name
    SonoSync
    Date Cleared
    2024-08-23

    (106 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SonoSync is a software solution for live streaming images acquired through ultrasound systems to healthcare professionals for diagnostic image viewing, remote access, review, consultation, quidance, support and education in real time.

    It is required to follow the technical and operator requirements in the User Manual.

    As users of this remote solution, the healthcare professionals take responsibility for ensuring that image quality, display contrast and ambient light conditions meet the generally accepted practices of the ultrasound imaging.

    Device Description

    SonoSync is a software solution that allows medical professionals to stream ultrasound images live, providing real-time access to diagnostic imaging, remote viewing, reviewing, consulting, guiding, supporting, and educating.

    SonoSync supports four main features: remote diagnostic viewing and guiding, remote clinical training and education, collaboration, and remote service support.

    With permission from the person operating the ultrasound equipment, a remote user can control the ultrasound equipment using a virtual remote touch panel and remote control panel.

    Furthermore, under the technical, operational, and environmental conditions described in the manual, healthcare professionals can provide clinical diagnostics at a remote location as if directly using the ultrasound equipment.

    AI/ML Overview

    The provided text does not contain detailed information about a study that proves the device meets specific acceptance criteria in the format requested. The document is an FDA 510(k) submission summary for SonoSync, a software solution for live streaming ultrasound images.

    While it mentions "Pre-set criteria were utilized in validation tests to assess whether remote viewing and reviewing with SonoSync matched the performance of local ultrasound systems," it does not provide the specific acceptance criteria, the reported device performance against those criteria, the sample size or provenance of data, the number or qualifications of experts, adjudication methods, or other detailed study information.

    It states:

    • "10. Summary of Non-Clinical Testing: Pre-set criteria were utilized in validation tests to assess whether remote viewing and reviewing with SonoSync matched the performance of local ultrasound systems. Specifications for remote displays and required network bandwidth to ensure diagnostic image quality were identified. Labeling materials have been prepared to inform users about the necessary specifications for safely and effectively conducting remote diagnostic reviews and viewing."

    This indicates that some testing was performed, but the results, methodology, and specific criteria are not elaborated in this summary. Therefore, I cannot generate the requested table or answer the specific questions about the study design.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240631

    Validate with FDA (Live)

    Date Cleared
    2024-06-21

    (107 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and transducers are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode

    Device Description

    The V8/XV8/XH8, V7/XV7/XH7, V6/XV6/XH6 diagnostic ultrasound system are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. The V8/XV8/XH8, V7/XV7/XH7, V6/XV6/XH6 diagnostic ultrasound system also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The V8/XV8/XH8, V7/XV7/XH7, V6/XV6/XH6 diagnostic ultrasound system have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the AI-based NerveTrack feature, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance (NerveTrack - Detection)

    Validation TypeDefinitionAcceptance CriteriaReported Average PerformanceStandard Deviation95% Confidence Interval
    Accuracy (%)$\frac{\text{Number of correctly detected frames}}{\text{Total number of frames with nerve}} \times 100$≥ 80%90.34.8887.28 to 93.33
    Speed (FPS)$\frac{1000}{\text{Average latency time of each frame (msec)}}$≥ 2 FPS3.660.253.51 to 3.82

    1. Table of Acceptance Criteria and Reported Device Performance (NerveTrack - Segmentation)

    Validation TypeDefinitionAcceptance CriteriaReported Average PerformanceStandard Deviation95% Confidence Interval
    Accuracy (%)$\frac{\text{Number of correctly segmented frames}}{\text{Total number of frames with nerve}} \times 100$≥ 80%98.423.9997.13 to 99.71
    Speed (FPS)$\frac{1000}{\text{Average latency time of each frame (msec)}}$≥ 2 FPS3.640.403.48 to 3.80

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the numerical sample size for the test set (number of frames or cases). It only mentions that the "test images" were used and that evaluation included "BMI subgroups" (underweight, healthy weight, overweight, and obesity) to assess generalizability.

    • Sample Size: Not explicitly quantified.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, though Samsung Medison Co., Ltd. is based in the Republic of Korea. The document mentions "All the nerve data were acquired on Samsung ultrasound devices including V8."
      • Retrospective or Prospective: Not explicitly stated, however, the description of "collecting scan data" and the involvement of experts in "establishing the ground truth" suggests these were real-world scans that were then annotated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Number of Experts: 12 anesthesiologists and 5 sonographers (total of 17 experts).
    • Qualifications: Over ten years' experience.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: For each dataset, the doctors who performed the ultrasound scans directly drew the ground truth (GT) of the nerve locations. These were then reviewed and verified as accurate by at least two additional anesthesiologists and sonographers. Any errors or discrepancies were corrected until satisfactory results were achieved. This effectively describes a form of expert consensus and review process.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not reported. The study focused on the standalone performance of the AI algorithm.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    • Yes, a standalone performance study was done. The reported accuracy and speed metrics are for the NerveTrack algorithm itself, without human intervention or assistance during the test. The document states, "The standalone performance of NerveTrack was evaluated..."

    7. Type of Ground Truth Used:

    • Type of Ground Truth: Expert consensus. Specifically, "The ground truth made by anesthesiologists" and established by the manual drawing/delineation of nerve locations by doctors who performed the scans, followed by review and verification by at least two additional experts.

    8. Sample Size for the Training Set:

    • Sample Size: Not explicitly stated. The document only mentions "The training data used for the training of the NerveTrack algorithm are independent of the data used to test the NerveTrack algorithm."

    9. How the Ground Truth for the Training Set Was Established:

    • Ground Truth Establishment: The document does not explicitly describe how the ground truth for the training set was established. It only states that the training data and test data are independent. However, given the robust process described for the test set's ground truth (expert drawing and multi-expert review/verification), it is highly probable that a similar, if not identical, expert-driven ground truth establishment method was used for the training data as well.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240516

    Validate with FDA (Live)

    Date Cleared
    2024-06-12

    (110 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.

    The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans vaginal, Muscular Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans esophageal(Cardiac) and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, Power Doppler (PD) mode, ElastoScan™ Mode, Multi Image mode(Dual, Quad), Combined modes. 3D/4D mode.

    Device Description

    The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Dopler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode, MV-Flow Mode or as a combination of these modes. The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals. the RS85 has real time acoustic output display with two basic indices, a mechanical index, which are both automatically displayed.

    AI/ML Overview

    This document describes the validation of AI-based features for the Samsung Medison RS85 Diagnostic Ultrasound System, specifically focusing on NerveTrack's detection, segmentation, and EzNerveMeasure functionalities, as well as SonoSync.

    Here's a breakdown of the requested information:

    Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Device Performance
    NerveTrack (Detection)Not explicitly stated as a numerical acceptance criterion, but the validation study aimed to demonstrate a high accuracy and reasonable speed for nerve detection. A detection is considered correct if the Dice coefficient is 0.5 or more.Average accuracy from 10 image sequences: 89.6% (95% Confidence Interval: 86.41, 92.79) Average speed (fps): 14.49 (95% CI: 13.86, 15.11)
    NerveTrack (Segmentation)Not explicitly stated as a numerical acceptance criterion, but the validation study aimed to demonstrate high accuracy, Dice similarity coefficient, and low Hausdorff distance for nerve segmentation. A segmentation is considered correct if the Dice coefficient is 0.5 or more.Average accuracy from nine image sequences: 99.78% (95% Confidence Interval: 99.34, 100) Average speed (fps): 10.44 (95% CI: 10.03, 10.85) Average Dice similarity coefficient: 90.44% (95% Confidence Interval: 86.19, 94.69) 95% Hausdorff distance (excluding bone): 18.69 pixels (95% Confidence Interval: 9.21, 28.17) Hausdorff distance (for bone): 93.51 pixels (95% Confidence Interval: 27.31, 159.72)
    NerveTrack (EzNerveMeasure - FR)Not explicitly stated as a numerical acceptance criterion, but the validation study aimed to demonstrate a low error rate for Flattening Ratio (FR) measurements.Average error rate of FR: 6.11% (95% Confidence Interval: 5.10, 7.12)
    NerveTrack (EzNerveMeasure - CSA)Not explicitly stated as a numerical acceptance criterion, but the validation study aimed to demonstrate a low error rate for Cross-sectional Area (CSA) measurements.Average error rate of CSA: 9.75% (95% Confidence Interval: 8.54, 10.96)
    SonoSyncPre-determined criteria were utilized to assess whether remote viewing and reviewing matched the performance of local ultrasound systems. (Specific numerical criteria not provided in the document).Labeling materials are provided to inform users about the necessary specifications for safely and effectively conducting remote diagnostic reviews and viewing. The document states that validation tests assessed if performance matched local systems.

    Study Details

    1. Sample Sizes and Data Provenance

    FeatureTest Set Sample SizeData ProvenanceRetrospective/Prospective
    NerveTrack (Detection)3,999 nerve imagesEight hospitals in Korea (Ethnicity: Koreans)Prospective
    NerveTrack (Segmentation)1,753 nerve imagesTen hospitals in Korea (Ethnicity: Koreans)Prospective
    NerveTrack (EzNerveMeasure)50 median nerve imagesA hospital in Korea (Ethnicity: Koreans)Prospective

    2. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Three participating experts.
    • Qualifications of Experts:
      • One anesthesiologist with more than 10 years of experience in pain management performed the initial manual drawing of nerve areas.
      • Other doctors with more than 10 years of experience performed verification of the ground truth.

    3. Adjudication Method for the Test Set

    The adjudication method appears to be a 2+1 process:

    1. Initial ground truth (GT) data was manually drawn by one anesthesiologist.
    2. Other doctors (presumably at least two distinct individuals to form a consensus if needed, though the document only states "other doctors") with more than 10 years of experience checked every frame.
    3. If discrepancies ("did not agree on nerve locations/contours") arose, necessary corrections were made to finalize the GT. This indicates a consensus-based approach for disagreement resolution, but the specific number for resolving disagreements is not explicitly 2+1; it just states "other doctors."

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed according to the provided text. The studies focus on the standalone performance of the AI algorithms.

    5. Standalone Performance Study

    Yes, standalone (algorithm only without human-in-the-loop) performance studies were done for NerveTrack's detection, segmentation, and EzNerveMeasure functionalities. The reported results (accuracy, speed, Dice coefficient, Hausdorff distance, and error rates) are direct measurements of the algorithm's performance against established ground truth.

    6. Type of Ground Truth Used

    The ground truth used was expert consensus. It was established by manual delineation (drawing of nerve areas/contours) by an anesthesiologist with over 10 years of experience, followed by verification and correction (if needed) by other doctors also with over 10 years of experience.

    7. Sample Size for the Training Set

    The document explicitly states: "Data used for training, tuning and validation purpose are completely separated from the ones during training process, and there is no overlap among the three." However, the exact sample size for the training set is not provided in this document. It only gives the test set sizes.

    8. How Ground Truth for the Training Set Was Established

    While the exact size of the training set is not given, the method for establishing its ground truth is described similarly to the validation set:

    • The GT data for training, tuning, and validation were built by three participating experts.
    • Nerve areas were manually drawn by an anesthesiologist with more than 10 years of experience in pain management.
    • The doctors who scanned the ultrasound were directly involved in the construction of GT data.
    • For verification, other doctors with more than 10 years of experience checked every frame.
    • If they did not agree on nerve locations/contours, necessary corrections were made to create the final GT.

    This suggests that the ground truth for the training set was also established through expert consensus and manual delineation, following the same "Truthing" process as the test set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 5