Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K132439

    Validate with FDA (Live)

    Date Cleared
    2014-01-06

    (154 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sakura Steam Sterilizer ASSR-AO12, ASSR-AO12W, ASSR-AO12P and ASSR-A012PW / Skytron Integrity 270 and 270 VP Steam Sterilizer are designed and optimized for use in health care facilities. Sakura Steam Sterilizer ASSR-A012. ASSR-AO12W, ASSR-AO12P and ASSR-AO12PW / Skytron Integrity 270 and 270 VP Steam Sterilizer incorporate high-pressure steam to sterilize non-porous and porous, heat and moisture-stable items and materials used in the health care facilities.

    Device Description

    The Sakura Steam Sterilizer ASSR-A012, Sakura Steam Sterilizer ASSR-A012W, Sakura Steam Sterilizer ASSR-A012P, Sakura Steam Sterilizer ASSR-AO12PW, Skytron Integrity 270 Steam Sterilizer, and Skytron Integrity 270 VP Steam Sterilizer, are Class II, Product Code FLE Medical Devices as defined by CFR§880.6880 and defined for use in healthcare facilities. The Sakura Steam Sterilizer ASSR-A012. Sakura Steam Sterilizer ASSR-A012W. Sakura Steam Sterilizer ASSR-A012P. Sakura Steam Sterilizer ASSR-AO12PW, Skytron Integrity 270 Steam Sterilizer, and Skytron Integrity 270 VP Steam Sterilizer provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The sterilizers also dry items that have been sterilized with wet steam.

    AI/ML Overview

    The Sakura Steam Sterilizer models ASSR-A012, ASSR-A012W, ASSR-A012P, ASSR-A012PW, Skytron Integrity 270 Steam Sterilizer, and Skytron Integrity 270 VP Steam Sterilizer are assessed for safety and effectiveness through non-clinical testing and comparison to a predicate device.

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criterion for sterilization efficacy is a Sterility Assurance Level (SAL) of <10^-6, meaning less than one in one million probability of a viable microorganism after sterilization. The devices were validated according to the ANSI/AAMI ST8: 2008 standard.

    Acceptance CriteriaReported Device Performance
    Sterilization Efficacy (SAL): <10^-6 SAL based on ANSI/AAMI ST8: 2008 Section 5.5.2.5-a)Met: <10^-6 SAL
    Factory-Programmed CyclesMet: All factory-programmed cycles were validated according to ANSI/AAMI ST8: 2008.
    Safety Standards ComplianceMet: Complies with IEC 60601-1-2:2007, IEC61010-1, IEC61010-2-040: 2005, CISPR11: 2009, IEC61000-3-2:2005, IEC61000-3-3:2008, IEC61000-4-2: 2008, IEC61000-4-3: 2006, IEC61000-4-4: 2004, IEC61000-4-5: 2005, IEC61000-4-6: 2003, IEC61000-4-8: 2009, IEC61000-4-11: 2004, and ASME Boiler and Pressure Vessel Code Section VIII Division 1 2010 Edition.
    Software ValidationMet: Software validation for cycle operation performed according to FDA's moderate level of concern recommendations ("Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (May 2005)").

    2. Sample size used for the test set and the data provenance:

    The document describes "complete kill of biological indicators" but does not specify the sample size for the biological indicator tests. The data provenance is primarily from non-clinical verification, validation, and testing activities conducted by Sakura Seiki Co., Ltd. The context suggests these are prospective tests performed on the manufactured devices. The manufacturer is based in Japan.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the summary. The ground truth for sterilization efficacy is established by verifying the complete kill of biological indicators and achieving an SAL of <10^-6, which typically relies on microbiological testing rather than expert consensus on individual cases.

    4. Adjudication method for the test set:

    This information is not applicable for this type of device and study. The validation process involves laboratory testing with biological indicators against a predefined standard (ANSI/AAMI ST8: 2008), not human adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a steam sterilizer, a medical device for sterilization, not an AI-assisted diagnostic tool. Therefore, MRMC studies and assessment of human reader improvement with AI assistance are irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable in the context of AI algorithms. However, in the context of device performance, the sterilizer's function is standalone, performing sterilization cycles autonomously. Its efficacy is assessed independently of human intervention during the sterilization process itself, once programmed and initiated.

    7. The type of ground truth used:

    The ground truth for the effectiveness of the sterilization process is established by:

    • Complete kill of biological indicators: This is a direct microbiological measure of sterilization efficacy.
    • Sterility Assurance Level (SAL) of <10^-6: This is a universally accepted standard for sterilization, verified through biological indicator testing.
    • Compliance with ANSI/AAMI ST8: 2008: This industry standard dictates the methodologies and criteria for validating steam sterilization cycles.

    8. The sample size for the training set:

    This is not applicable. The device is a steam sterilizer, not a machine learning model. There is no concept of a "training set" for such a device.

    9. How the ground truth for the training set was established:

    This is not applicable for the same reason as above. The device does not utilize a training set in the machine learning sense. The ground truth for its performance is established through rigorous validation and testing against pre-defined industry standards and performance requirements using biological indicators.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122774

    Validate with FDA (Live)

    Date Cleared
    2013-05-09

    (241 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers are designed and optimized for use in health care facilities. Sakura Steam Sterilizer ASSV-AH06 / Skytron Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E / Skytron Integrity 175 SG Steam Sterilizers incorporate high-pressure steam to sterilize non-porous and porous, heat and moisture-stable items and materials used in the health care facilities.

    Device Description

    Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizers are Class II, Product Code FLE Medical Devices as defined by CFR§880.6880 and defined for use in healthcare facilities. Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E and SKYTRON Integrity 175 SG Steam Sterilizers provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The sterilizers also dry items that have been sterilized with wet steam.

    AI/ML Overview

    The Sakura Steam Sterilizer ASSV-AH06, SKYTRON Integrity 175 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AH06E, and SKYTRON Integrity 175 SG Steam Sterilizers were validated to meet the requirements of ANSI/AAMI ST8:2008.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the device are the performance specifications for each sterilization cycle, as outlined in the "Maximum Validated Loads," "Sterilize Temp," "Sterilize Time," and "Dry Time" columns. The reported device performance is that the "device passed all of the tests based on predetermined Pass/Fail criteria."

    CycleAcceptance Criteria - Maximum Validated LoadsAcceptance Criteria - Sterilize TempAcceptance Criteria - Sterilize TimeAcceptance Criteria - Dry TimeReported Device Performance
    Flash 1 (IU 1)One unwrapped, nonporous instrument no heavier than 0.22 lbs. (100 gram)270 F (132 C)3 minutes1 minutePassed
    Flash 2 (IU 2)Maximum of two (2) unwrapped instrument trays, maximum weight 25 pound (11.3 kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load.270 F (132 C)4 minutes1 minutePassed
    Prevacuum 270Maximum of two (2) double-wrapped instrument trays, maximum weight 25 pound (11.3kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load.270 F (132 C)4 minutes30 minutesPassed
    Prevacuum 275Maximum of two (2) double-wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load.275 F (135 C)3 minutes30 minutesPassed
    Express Prevacuum (IU 3)Maximum of two (2) single wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. The maximum of three (3) 2 mm inside diameter (minimum) and 400 mm long (maximum) lumen instruments per cycle load. This cycle is intended for immediate use and the wrap used for this cycle is not intended to be a sterile barrier.270 F (132 C)4 minutes3 minutesPassed
    Gravity 1Maximum of four (4) - fabric packs.250 F (121 C)30 minutes15 minutesPassed
    Gravity 2Maximum of two (2) double-wrapped instrument trays, maximum weight 25 pound (11.3 kg) each.270 F (132 C)15 minutes30 minutesPassed
    LiquidMaximum of three (3) 1,000 milliliter containers with vented closures. This cycle is intended for non-patient contacting devices only.250 F (121 C)45 minutes0 minutesPassed
    Leak TestEmpty Chamber, test time of 15 minutes.N/AN/AN/APassed
    Warm UpEmpty Chamber270 F (132C)4 minutes3 minutesPassed
    Bowie-Dick TestBowie-Dick Test Pack273 F (134 C)3.5 minutes2 minutesPassed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample size used for the test set for each validation activity. It broadly states that "Verification, validation and testing activities were conducted to establish the performance, functionality and reliability characteristics" and that "the device passed all of the tests." The validation was performed according to ANSI/AAMI ST8:2008, a standard for steam sterilization. The data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin of the data mentioned, though the manufacturer is based in Japan.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The validation relied on adherence to recommended practices, standards, and guidelines from organizations like AAMI, rather than human expert consensus on specific cases. The "ground truth" for sterilization devices is typically based on the complete kill of biological indicators and verification of a specified sterility assurance level (SAL).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and therefore not provided, as the validation method described does not involve human adjudication of results in the traditional sense of medical image analysis or similar diagnostic tasks. The efficacy is based on objective measurements related to the destruction of biological indicators.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable to the device described (a steam sterilizer). MRMC studies are typically performed for diagnostic devices where human readers interpret output, often with AI assistance. This device sterilizes medical instruments, and its performance is measured by its ability to sterilize, not by human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the testing described appears to be a standalone performance evaluation of the sterilizer itself, without a human-in-the-loop component for its primary function (sterilization). The device's efficacy is demonstrated by its ability to achieve a complete kill of biological indicators and a specified sterility assurance level.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for validating the sterilizer's effectiveness is based on the complete kill of biological indicators and verifying an appropriate sterility assurance level (SAL) of at least 10^-6 reduction. This is a microbiological standard, not expert consensus, pathology, or outcomes data related to patient diagnosis.

    8. The sample size for the training set

    This information is not applicable. The device is a steam sterilizer, not an AI/ML algorithm that requires a training set in the conventional sense. The "software validation" mentioned was performed according to FDA guidance for "Software Contained in Medical Devices," which focuses on verifying software functionality and safety, not on training an algorithm with data.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no "training set" for this type of medical device. The validation process ensures the device meets predefined sterilization standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120149

    Validate with FDA (Live)

    Date Cleared
    2012-05-08

    (111 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sakura Steam Sterilizer ASSV-AB-09 / Skytron Integrity 215 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AB09E / Skytron Integrity 215 SG Steam Sterilizer are designed and optimized for use in health care facilities. Sakura Steam Sterilizer ASSV-AB-09 / Skytron Integrity 215 Steam Sterilizer, Sakura Steam Sterilizer ASSV-AB09E / Skytron Integrity 215 SG Steam Sterilizer incorporate high-pressure steam to sterilize non-porous and porous, heat and moisture-stable items and materials used in the health care facilities.

    Device Description

    The SKYTRON Integrity 215 Steam Sterilizer. SKYTRON Integrity 215 SG Steam Sterilizer, Sakura Steam Stenlizer ASSV-AB09, Sakura Steam Sterilizer ASSV-AB09E Steam Sterilizers are Class II, Product Code FLE Medical Devices as defined by CFR§880.6880 and defined for use in healthcare facilities. The SKYTRON Integrity 215 Steam Sterilizer, SKYTRON Integrity 215 SG Steam Sterilizer, Sakura Steam Sterilizer ASSV-AB09, Sakura Steam Sterilizer ASSV-AB09E Steam Sterilizers provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The sterilizers also dry items that have been sterilized with wet steam.

    AI/ML Overview

    This document is a 510(k) Summary for the SKYTRON Integrity 215 Steam Sterilizer and its variations. It focuses on the device's capabilities to sterilize instruments and materials in healthcare facilities. It does not describe an AI/ML device, therefore, the following requested information points are not applicable as they pertain to AI/ML device studies.

    Therefore, I will provide the available information from the document regarding the device's acceptance criteria and the study conducted.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the steam sterilizers are defined by the factory-programmed sterilization cycle values, which indicate the maximum validated loads, sterilize temperature, sterilize time, and dry time for various cycles. The reported device performance is that these cycles were "validated according to ANSI/AAMI ST8: 2008" and that the device "passed all of the tests based on pre-determined Pass/Fail criteria."

    CycleAcceptance Criteria: Maximum Validated LoadsAcceptance Criteria: Sterilize TempAcceptance Criteria: Sterilize TimeAcceptance Criteria: Dry TimeReported Device Performance
    Flash 1 (IU 1)One unwrapped, nonporous instrument no heavier than 0.22 lbs. (100 gram)270 F (132 C)3 minutes1 minuteValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Flash 2 (IU 2)Maximum of six (6) unwrapped instrument trays, maximum weight 25 pound (11.3 kg) each. Max three (3) 2 mm ID (min) and 400 mm long (max) lumen instruments per cycle load.270 F (132 C)4 minutes1 minuteValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Prevacuum 270Maximum of six (6) double-wrapped instrument trays, maximum weight 25 pound (11.3kg) each. Max three (3) 2 mm ID (min) and 400 mm long (max) lumen instruments per cycle load. (Also 12 fabric packs with 15 min dry time)270 F (132 C)4 minutes30 minutes (or 15 min for fabric)Validated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Prevacuum 275Maximum of six (6) double-wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. Max three (3) 2 mm ID (min) and 400 mm long (max) lumen instruments per cycle load.275 F (135 C)3 minutes30 minutesValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Express Prevacuum (IU 3)Maximum of six (6) single wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. Max three (3) 2 mm ID (min) and 400 mm long (max) lumen instruments per cycle load. (Immediate-use, not sterile barrier)270 F (132 C)4 minutes3 minutesValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Gravity 1Maximum of twelve (12) - fabric packs.250 F (121 C)30 minutes15 minutesValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Gravity 2Maximum of six (6) double-wrapped instrument trays, maximum weight 25 pound (11.3 kg) each. (Also 12 fabric packs with adjusted times)270 F (132 C)15 minutes (or 25 min for fabric)30 minutes (or 15 min for fabric)Validated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    LiquidMaximum of three (3) 1,000 mill-liter containers with vented closures. (Non-patient contact only)250 F (121 C)45 minutes0 MinuteValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Leak TestEmpty Chamber, test time of 15 minutes.N/AN/AN/AValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Warm UpEmpty Chamber270 F (132C)4 minutes3 minutesValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.
    Bowie-Dick TestBowie-Dick Test Pack273 F (134 C)3.5 minutes2 MinuteValidated according to ANSI/AAMI ST8: 2008; Passed all pre-determined Pass/Fail criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the "sample size" in terms of number of sterilization cycles or tests performed during validation. However, it mentions that "Efficacy of sterilizer function and exposure time recommendations are ultimately shown by complete kill of biological indicators and verifying an appropriate safety factor or sterility assurance level (SAL) of at least 10^-6 reduction."

    The data provenance is not specified beyond "Sakura Seiki Co., Ltd. validates its sterilization cycles by recommended practices, standards and guidelines developed by various independent organizations such as the Association for Advancement of Medical Instrumentation (AAMI)." The document also states the testing was "Non-clinical: Verification, validation and testing activities were conducted to establish the performance, functionality and reliability characteristics" of the sterilizers. This suggests the data was collected specifically for validation purposes of this device, likely in a controlled, prospective manner, but the geographic origin of the testing is not detailed. The manufacturer is based in Japan.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided. The efficacy is demonstrated by the "complete kill of biological indicators" to achieve a sterility assurance level (SAL) of at least 10^-6, which is a quantified scientific outcome, not dependent on expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    This is not applicable as the validation focuses on objective performance measures (sterility assurance) rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size.

    This is not applicable, as this is not an AI/ML diagnostic device with human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable, as this is not an AI/ML device. The device's "standalone" performance in a sense refers to its ability to sterilize independently, which was validated.

    7. The Type of Ground Truth Used

    The ground truth used for establishing the effectiveness of the sterilizer is based on biological indicators (BIs) and achieving a sterility assurance level (SAL) of at least 10^-6 reduction. This signifies that the probability of a viable microorganism remaining after sterilization is 1 in 1,000,000. This is an objective, scientifically established outcome. The validation also adheres to ANSI/AAMI ST8: 2008.

    8. The Sample Size for the Training Set

    This is not applicable, as this is not an AI/ML device that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1