510(k) Data Aggregation

    K Number
    K182542

    Validate with FDA (Live)

    Date Cleared
    2019-02-22

    (158 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EPAD 2 system is intended for use in monitoring neurological status by recording somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open or percutaneous, lumbar, thoracic, and cervical surgical procedures.

    Device Description

    The EPAD 2 System consists of the EPAD 2 Headbox and power supply, a tablet computer, electrodes, stimulating probe or clip and cables. The Headbox is responsible for the stimulation and acquisition of signals for the intraoperative neurophysiologic monitoring (IONM) functions. The Headbox contains onboard firmware and is mounted on the operating room (OR) bed by clipping it to the bed side bar. The EPAD application provides the primary graphical user interface and controls for the EPAD II System. The application runs on a touchscreen tablet mobile device which connects wirelessly or via a wired USB cable to the Headbox. The application serves as the interface to the EPAD Headbox, enabling both user input (e.g., patient and procedure information, adjustment of stimulus and acquisition parameters) and display of output (e.g., display of acquired waveforms, data, messages and alerts to the clinician).

    AI/ML Overview

    The provided text primarily focuses on the 510(k) clearance of the EPAD 2 System, demonstrating its substantial equivalence to predicate devices rather than presenting data from a specific study designed to meet acceptance criteria for novel performance claims. Therefore, much of the requested information regarding acceptance criteria, reported performance, and study details is not explicitly available in the provided document.

    However, I can extract information related to the device's characteristics and the types of testing performed to support its substantial equivalence.

    Here's an analysis based on the given document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding reported device performance metrics in the way one would expect for a study proving new performance claims. Instead, it relies on demonstrating substantial equivalence to predicate devices. The performance data mentioned are related to meeting functional, system, and software requirements, and compliance with safety standards.

    CharacteristicAcceptance Criteria (Implied by substantial equivalence)Reported Device Performance (Summary from text)
    Intended UseSimilar to predicate devicesEPAD 2 system is intended for monitoring neurological status by recording SSEP, EMG, or assessing NMJ during various surgical procedures. (Matches expanded indications of predicates)
    TechnologySimilar to predicate devices (waveform, pulse width, frequency, current range)Monophasic, Rectangular waveform; 50-300 µsec pulse width; 0.1-50 Hz frequency; 0-100 mA current range. (Within ranges of predicate devices and their combinations)
    SafetyCompliance with recognized electrical safety standards (e.g., 60601-1)EMC and Electrical Safety Testing performed to ensure all functions are electrically safe and comply with recognized electrical safety standards.
    UsabilityNo adverse effect within the intended environmentUsability testing performed to demonstrate no adverse effect.
    Functional, System, Software RequirementsMeeting defined requirementsNonclinical performance testing demonstrates that the EPAD 2 System meets these requirements.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document states: "Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data."
    This indicates that:

    • There was no clinical test set in the traditional sense for evaluating novel performance claims against acceptance criteria.
    • The data provenance, sample size, or retrospective/prospective nature of a clinical study are not specified because such a study was not the basis for clearance.
    • The performance data cited are from nonclinical performance testing, EMC, electrical safety, and usability testing, which typically involve device-level assessments rather than patient-based test sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Since there was no clinical test set for evaluating novel performance claims, there is no information provided regarding experts establishing ground truth for such a test set. The substantial equivalence determination relies on comparisons to already cleared predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set requiring ground truth adjudication for novel performance claims was used as the basis for this 510(k) clearance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The EPAD 2 System is an Evoked Response Electrical Stimulator used for neuromonitoring, not an AI-assisted diagnostic imaging device that would typically involve a multi-reader multi-case study or "human readers improve with AI" metrics.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable to the EPAD 2 System. The device is an instrument used by qualified medical personnel for neuromonitoring and does not operate as a standalone "algorithm only" device for making diagnoses or performing interventions without human-in-the-loop interaction. Its function is to acquire and display physiological signals for interpretation by a clinician.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the nonclinical performance, EMC, electrical safety, and usability testing, the "ground truth" would be established by:

    • Engineering specifications and regulatory standards: For functional, system, and software requirements, and for EMC/electrical safety, the device's performance is compared against pre-defined engineering specifications and national/international standards (e.g., IEC 60601-1).
    • User feedback/observation: For usability testing, the "ground truth" is typically the observation of user interaction and compliance with usability engineering principles, identifying potential use errors or safety concerns.

    There is no mention of pathology or outcomes data as ground truth for this 510(k) submission, as it explicitly states that clearance is "not based on an assessment of clinical performance data."

    8. The sample size for the training set

    Not applicable. This device is cleared through the 510(k) pathway which demonstrates substantial equivalence to predicate devices. It is not an AI/ML device that typically requires a large "training set" in the machine learning sense for its core functionality. While there might be firmware/software development which internally used data, the submission does not detail such a training set for regulatory review.

    9. How the ground truth for the training set was established

    Not applicable, as no training set in the AI/ML context is described or relevant for this 510(k) cleared device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132616

    Validate with FDA (Live)

    Device Name
    EPAD
    Date Cleared
    2014-01-24

    (156 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EPAD is intended for use in monitoring neurological status by recording somatosensory evoked potentials (SSEP) or assessing the neuromuscular junction (NMJ).

    Device Description

    The EPAD system consists of the following components/accessories:

    EPAD Headbox EPAD Headbox Power Supply EPAD Tablet Computer (includes power supply and USB cable) Stimulator Left Blue Cable Assembly (110 inches) Stimulator Right Yellow Cable Assembly (110 inches) Acquisition Left Red Cable Assembly (91 inches) Acquisition Right White Cable Assembly (92 inches) Stimulator Left Blue Short Cable Assembly (67 inches) Stimulator Right Yellow Short Cable. Assembly (67 inches) Acquisition Left Red Short Cable Assembly (44 inches) Acquisition Right White Short Cable Assembly (72 inches) Adapter Cable for EPAD Headbox (for leakage current testing)

    Upper Limb Electrodes package

    Lower Limb Electrodes package

    The EPAD Headbox contains a complete data acquisition system that has built-in amplifiers, analog to digital converters, and digital signal processors. User interface is via tablet touchscreen computer provided with the EPAD System and running the Android operating system. The EPAD software application is preloaded onto the tablet. Data can be transferred to an external computer for archiving purposes. Communication between the EPAD Headbox and tablet is via Bluetooth wireless or USB connection.

    Electrode cables are provided for left and right stimulation and left and right acquisition, color coded for correct connection to the EPAD Headbox: Custom cutaneous electrodes for stimulus and acquisition are provided by SafeOp for use with the EPAD. A total of 11 electrodes are applied for full patient monitoring (upper and lower limbs). The SafeOp electrodes are wet qel, single patient use, disposable, and biocompatible for short term (<24 hours) use on intact skin.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the EPAD™ device, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary for the EPAD™ primarily establishes substantial equivalence through a comparison with predicate devices and nonclinical testing that confirms safety and performance against established standards. Direct "acceptance criteria" for diagnostic performance (like sensitivity/specificity) are not explicitly stated for a clinical study comparing the AI's diagnostic ability to a gold standard. Instead, the acceptance is based on meeting technical specifications and demonstrating safety and effectiveness.

    However, we can infer performance criteria from the non-clinical testing performed and the comparison to predicate devices. The "reported device performance" refers to the results of these tests.

    CategoryAcceptance Criteria (Inferred/Stated from Predicates/Standards)Reported Device PerformanceStudy proving compliance
    Alert Algorithm PerformanceAbility to detect waveform changes of reduced amplitude (>50%) and increased latency (>10%) under minimal, moderate, and extreme noise conditions.The EPAD was able to detect waveform changes of reduced amplitude (>50%) and increased latency (>10%) under all noise conditions (minimal, moderate, and extreme).Alert Algorithm Testing
    Software EfficacyAll software requirements are fulfilled and all software hazards mitigated with no unresolved anomalies.All software requirements have been fulfilled and all software hazards have been mitigated. There are no unresolved anomalies in the EPAD software.Software verification and validation testing
    BiocompatibilitySafe for short-term (<24 hours) contact with intact skin (meeting ISO 10993-1 requirements for cytotoxicity, sensitization, and irritation/intracutaneous reactivity).Electrodes are safe for short-term (<24 hours) contact with intact skin. Specific testing included cytotoxicity, sensitization, and irritation/intracutaneous reactivity, meeting ISO 10993-1.Biocompatibility testing on patient-contacting materials
    Shelf Life (Electrodes)15-month shelf life with all functional tests passed following accelerated aging (including Impedance monitoring, visual inspection, ANSI/AAMI EC12:2000 Section 5.2.2 for Disposable ECG Electrodes tests, Electrode Impedance values on forearm, and Electrode evoked potential responses).All tests conducted following accelerated aging of the electrodes in their final packaging were passed, establishing a 15-month shelf life. This included impedance monitoring, visual inspection, ANSI/AAMI EC12:2000 tests, electrode impedance values on forearm, and electrode evoked potential responses.Accelerated aging testing
    Electrical SafetyCompliance with UL60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-40 standards.All tests passed with no need for device modifications for UL60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-40.Electrical Safety and Electromagnetic Compatibility Testing
    EMCCompliance with IEC 60601-1-2 standard.All tests passed with no need for device modifications for IEC 60601-1-2.Electrical Safety and Electromagnetic Compatibility Testing
    FCC CertificationCompliance with FCC regulations under 47 CFR Part 15 for Bluetooth wireless technology.All tests passed related to FCC Certification for the Bluetooth wireless technology.FCC Certification Testing (conducted by tablet manufacturer)
    Signal QualityAcceptable noise level peak-to-peak, CMRR adjustment, LF/HF filter performance, DC offset, cross-talk, gain, and impedance circuit qualification on each channel.Tests validate the quality of waveform signals from Headbox to Tablet, including noise level peak-to-peak, CMRR adjustment, filter performance (LF, HF), DC offset, cross-talk, gain, and impedance circuit qualification. All test results met the test protocol acceptance criteria. (Specific quantitative results not provided in summary).EPAD Signal Quality Tests (Functional Performance Testing)
    Wireless CoexistenceNo observed decline in data transmission between the EPAD Headbox and Tablet in the presence of multiple simultaneously transmitting wireless devices.There was no observed decline in data transmission between the EPAD Headbox and Tablet under the test conditions (multiple wireless devices transmitting simultaneously).Wireless Coexistence Testing (Functional Performance Testing)
    Electrode FunctionalMeet requirements per AAMI/ANSI EC12:2000: Disposable ECG Electrodes (Section 5.2.2: 10-Hz AC impedance, individual pair; Combined offset instability and internal noise; DC voltage offset, Bias test).EPAD electrodes meet functional test requirements per AAMI/ANSI EC12:2000: Disposable ECG Electrodes (specifically Section 5.2.2 tests: 10-Hz AC impedance, combined offset instability and internal noise, and DC voltage offset/bias test).Accelerated aging testing (as part of shelf-life) and likely separate functional performance testing specifically for electrodes.
    Max VoltageWithin acceptable range compared to predicates (similar expected max current). Protektor 32 has 400VDC with 4kΩ load.380 VDC with 3.8 kΩ load impedance. Deemed substantially equivalent as "Slightly lower maximum driving voltage but same expected maximum current."(Comparison to Predicate)
    Pulse DurationWithin range offered by predicates and appropriate for EPAD modes. Predicates range from 0.05 msec to 1.0 msec.Choose from 0.1, 0.2 or 0.3 msec settings. Deemed appropriate for EPAD modes.(Comparison to Predicate)
    Repetition RateWithin range offered by predicates and appropriate for EPAD modes. Predicates range from 0.1 Hz to 1000 Hz.0.1 to 50 Hz. Deemed appropriate for EPAD modes.(Comparison to Predicate)
    TimebaseWithin range offered by predicates and appropriate for EPAD modes. Protektor 32 has 0.5 ms/division to 500 ms/division.2 ms/division to 10 ms/division. Deemed appropriate for EPAD modes.(Comparison to Predicate)
    SensitivityWithin range offered by predicates and appropriate for EPAD modes. Protektor 32 has 0.1 µV/division to 5 mV/division.0.5 µV/division to 20 µV/division. Deemed appropriate for EPAD modes.(Comparison to Predicate)
    High Frequency FilterAppropriate for EPAD modes (SSEP and NMJ tests requiring HFF of 3 kHz), even if different from general-purpose predicate devices. Predicate Protektor 32 has 30 Hz to 15 kHz.30 Hz to 3 kHz. Justified as appropriate for SSEP and NMJ tests.(Comparison to Predicate)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set: Non-clinical testing mainly involved hardware, software, and component-level assessments rather than a clinical "test set" of patient data for diagnostic performance.
      • For the Alert Algorithm Testing, the summary states it was tested "under minimal, moderate, and extreme noise conditions," implying simulated or laboratory conditions rather than patient data. No specific number of cases or data provenance is mentioned.
      • For Electrode Functional Testing (AAMI/ANSI EC12:2000), these are standard bench tests on electrode samples. The sample size would be number of electrodes tested, but this is not specified.
      • For Biocompatibility Testing, the sample size would be biological assays and animal tests. Not specified in summary.
      • Data Provenance: Not applicable for most non-clinical tests mentioned. For any implicitly simulated data (like for the alert algorithm), no provenance (country, retrospective/prospective) is provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • This information is not provided in the summary. Since the testing described is primarily non-clinical (engineering, software, biocompatibility), there was no "ground truth" derived from human experts for a diagnostic task in the traditional sense. The alert algorithm's performance criteria (amplitude and latency changes) are predefined thresholds not requiring expert adjudication in the context of this summary.

    4. Adjudication Method for the Test Set

    • Not applicable / None described. The non-clinical testing does not appear to involve human observers or an adjudication process for ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done or described in this 510(k) summary. The EPAD incorporates an alert algorithm, but its performance is evaluated in a standalone manner against predefined thresholds, not as an AI-assisted diagnostic tool for human readers in a clinical study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a form of standalone performance was assessed for the "Alert Algorithm Testing." The algorithm's ability to detect waveform changes (reduced amplitude and increased latency) was tested independently under various noise conditions.
      • Performance: "The EPAD was able to detect waveform changes of reduced amplitude (>50%) and increased latency (>10%) under all noise conditions."

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • For the Alert Algorithm: The "ground truth" was defined by predefined quantitative thresholds for waveform changes: >50% amplitude reduction and >10% latency increase. This is a technical, rather than a clinical, ground truth.
    • For other non-clinical tests (e.g., electrical safety, biocompatibility, signal quality), the ground truth is compliance with established engineering standards and specifications.

    8. The Sample Size for the Training Set

    • The summary does not provide information on a training set sample size. As a 510(k) submission for a device primarily focused on evoked potential recording and basic alert logic, it doesn't describe the development of a complex machine learning model requiring a distinct training set. If the alert algorithm involved any learned parameters, this information is not detailed.

    9. How the Ground Truth for the Training Set was Established

    • Not detailed / Not applicable as a distinct training set and its ground truth establishment are not discussed in this 510(k) summary. If the alert algorithm's thresholds were derived from clinical data, the method for establishing that ground truth is not provided.