Found 4 results

510(k) Data Aggregation

    K Number
    K251129

    Validate with FDA (Live)

    Date Cleared
    2025-11-07

    (210 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pro Zygoma dental implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. The Pro Zygoma dental implants may be used with single-stage or two-stage procedures and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to seek initial clearance for S.I.N. Tapered Pro Conical Zygoma Implant System which includes various endosseous dental implants and corresponding prosthetic components. S.I.N. Tapered Pro Conical Zygoma Implant System adds to the S.I.N. Dental Implant System, which includes several components previously cleared in K231127 and K240609, as well as other features of the Tapered Pro Conical Implant System previously cleared in K240187. This submission includes Pro Zygoma dental implants for placement in the maxillary arch, with corresponding Pro Conical Multi-unit Abutments with up to 60° angulation.

    The subject Pro Zygoma dental implants have an internal conical abutment connection, with a 15° cone taper. The Pro Zygoma dental implants are provided with body/platform diameters of 3.8 and 4.2 mm, and each body/platform size is provided in overall lengths of 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 55, 57.5 and 60 mm. For all implants, the threads start at the apex and extend 16.8mm coronally.

    The external machined surface of the subject Pro Zygoma implants, as well as the external surface of the Pro Conical Angled multi-unit abutments are colored yellow for aesthetic and identification purposes by a standard anodization process in which the devices are submerged in an electrolytic solution and exposed to an electric current to increase the thickness of the natural oxide layer on the surface and impart a distinctive color. No dyes are used in this process. The multi-unit abutment screw (part number SCMUAS) is anodized blue by the same process. The anodization process for the subject Pro Zygoma implants is identical to that used on the reference implant devices cleared in K240187. The anodization process for the subject Pro Conical Angled multi-unit abutments and multi-unit abutment screw is identical to that used on the primary predicate abutment devices cleared in K231127.

    Resorbable Blast Texturing (RBT) is applied to the threaded surface of all subject device implants creating a random, roughened texture which increases the implant surface area and helps achieve hard tissue (bone) attachment with the implant. RBT is the application, under pressure, of biocompatible hydroxylapatite (HA) particles (conforming to ASTM F1185 Standard Specification for Composition of Hydroxylapatite for Surgical Implants) to the exterior of the machined implant threads. Not to be confused with HA coating, RBT processing uses HA particles to blast the implant surface without depositing HA onto the surface. The HA used for RBT processing does not remain on the surface post-processing. HA media grain size is between 180 – 300 µm for the surface trademarked as RBT (surface finish at 60 Ra minimum).

    The subject Pro Conical Angled abutments are multi-unit, indexed abutments for use only with the subject Pro Zygoma implants. The 45°, 52° and 60° Pro Conical Angled Multi-unit Abutments have a prosthetic platform diameter of 4.8 mm, and a gingival height ranging from 1.5 mm to 3.0 mm and are designed only for use with the Pro Zygoma implants. The internal conical implant-abutment connection of the subject devices is identical to that of the BioHorizons reference devices included in K240187, thus the subject Pro Zygoma dental implants are fully compatible with the BioHorizons Conical Multi-unit Angled Abutments (17° and 30°).

    All subject dental implants are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. All subject implants have a resorbable blast textured (RBT) surface treatment, identical to that cleared in K240187. All subject multi-unit abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136. All subject implants and abutments are provided sterile to the end user.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252266

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (88 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System may be immediately loaded when good primary stability is achieved and the functional load is appropriate.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components previously cleared K200992, K193096, K170398 and K170392.

    This submission adds Interface Abutments and Healing Abutments to the S.I.N. Dental Implant System. The subject device Interface abutments are provided with MT 16º and CM11.5º connection compatible with the S.I.N. Dental implants.

    Subject Device Interface Abutments

    Subject device Interface Abutments are two-piece abutments for which the second part (or top half) is the ceramic superstructure. The second part is designed and manufactured using CAD-CAM techniques. The titanium component of the Interface Abutments has a cementable platform diameter of 3.5mm, 4.0mm, 4.5mm, or 5.5mm, cementable post-height of 4.0mm or 6.0mm, and a gingival height built into the titanium base component from 0.5mm to 4.0 mm. The two pieces of the Ti-Base are bonded using Panavia Universal Dual-Cure Cement (K150704), as referenced in K190936.

    All subject devices are provided only in straight designs, with no pre-angled conventional abutments or pre-angled CAD-CAM abutments. No subject device abutment is to be customized to create an angled abutment or to correct for angulation, and no superstructure for use with any subject device Interface abutment (CAD-CAM) is to be fabricated to create a final abutment with angulation or to correct for angulation. Therefore, this model has a slight design difference to improve screw insertion and removal. It was purposely designed to optimize fixation and removal of the screw connection. This new presentation allows screw fixation even in cases of anterior aesthetic rehabilitation, avoiding the need to create screw access holes in the sublingual phase. However, it is confirmed that all implant-abutment interfaces are intended exclusively for straight configurations and are not indicated for any angled model. These minor differences in the exact dimensions of the subject device interface abutment as compared to the predicate device and reference devices do not impact safety, effectiveness, or substantial equivalence.

    The subject device interface abutments with the CM11.5º interface are compatible with implants cleared in K200992, K221453, K211921 and K170392 in terms of abutment-implant connection. The subject device with the MT16º interface is compatible with implants cleared in K222231, K193096 and K170392 in terms of abutment-implant connection.

    The design parameters for the CAD-CAM fabrication of patient-specific abutments from CADCAM Abutments are the same for the 11.5° CM implant connection and the MT 16° implant connection. The design limit parameters are:

    • Minimum wall thickness – 0.5 mm
    • Minimum abutment post height for single-unit restoration – 4.0 mm
    • Maximum angle – 0°. Straight only
    • Maximum angle for screw insertion/removal – 20º
    • Maximum gingival height – 5.0mm
    • Minimum Gingival Height – 0.5 mm
    • Maximum allowable abutment Post Height – 6 mm
    • Total Abutment Height – 10mm

    By definition, the abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.

    All subject device Interface Abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and the same zirconia that is specified for superstructures in K190396, and are to be used with subject abutment screws. All subject devices are provided sterile. All Interface abutments are provided sterile by gamma irradiation; the same sterilization method used in predicate device K190396 and in the additional devices K200992 and K170392.

    All Interface abutments manufactured form titanium alloy are anodized using a standard anodization process is identical to the anodization process used on abutments cleared in K190396.

    The subject device abutments screws are to be used with Interface Abutments present in this submission and they are substantially equivalent to material (titanium alloy) and designs to abutments screws cleared in K170392.

    Subject Device Healing Caps

    The subject device Healing Cap is substantially equivalent to the corresponding design Healing Cap cleared in K170392. The subject device healing cap is designed for the healing period between implant placement and final abutment placement and is not intended to support a temporary prosthesis.

    All Healing Caps are provided sterile by gamma irradiation; the same sterilization method used in predicate device K170392 and additional predicate device K200992.

    All subject device components are provided sterile by gamma irradiation, the same sterilization method used in K190936, K170392 and K200992.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251262

    Validate with FDA (Live)

    Date Cleared
    2025-09-26

    (156 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.- validated milling center for manufacture.

    The Abutment Mini Angled Morse 450 is intended for use with the S.I.N. Dental Implant System Zygomatic implants. S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K222231, K231127, and K170392.

    This submission includes dental implants Epikut S with a Morse taper (MT) abutment interface and an acid-etched endosseous surface, and Epikut S Plus implants with an endosseous surface produced by acid- etching followed by application of a hydroxyapatite coating (HAnano) and the same described connection. The implant design and endosseous surfaces are nearly identical to the Epikut S and Epikut S Plus implants cleared in K222231, with the exception of the additional body/platform diameter (6.0 /5.2 mm) with lengths of 8.5, 10, 11.5, 13.

    The subject device dental implants are summarized in the following table.

    Implant LinesBody Ø, mmPlatform Ø, mmLengths, mm
    Epikut S6.05.28.5, 10, 11.5, 13
    Epikut S Plus6.05.28.5, 10, 11.5, 13

    This submission includes Abutment Mini Angled Morse 45° in sizes from 1.5, 2.0, 2.5 with a 16° Morse taper connection mating abutments for screw retained, multi-unit prostheses and a prosthetic platform diameter of 4.8 mm, that is for exclusively use with cleared Zygomatic Plus implant. The corresponding abutment screw is PFMAAM 16A.

    This submission includes Interface abutment MT 16° prosthetic components for fabrication of patient-specific abutments that are compatible with implants from S.I.N. Dental Implant System. Subject device Interface Abutments are two-piece abutments for which the second part (or top half) is the ceramic superstructure. The second part is designed and manufactured using CAD-CAM techniques. The titanium component of the Interface Abutments Morse Taper has a cementable platform diameter of 3.5mm, 4.5mm, or 5.5mm, cementable post-height of 4.0mm or 6.0mm, and a gingival height built into the titanium base component from 0.5mm to 4.0mm. The two pieces of the Ti-Base are bonded using Panavia Universal Dual-Cure Cement (K150704), as referenced in K190936. These abutments are used with a fixation screw PFMT 02A.

    After scanning the intraoral area, the S.I.N.-validated milling center will receive the STL file and design the coping/superstructure using computer software (CAD) according to the design parameters below and the space available.

    Design parameters for all Interface CAD-CAM Abutments superstructure are:

    • Minimum wall thickness – 0.5 mm
    • Minimum abutment post height for single-unit restoration – 4.0 mm
    • Maximum angle – 0°. Straight only
    • Maximum gingival height – 5.0 mm
    • Minimum Gingival height – 0 mm
    • Maximum allowable abutment Post Height – 6 mm
    • Total abutment height – 10mm
    • Abutment diameter – depends on available space (for the patient)

    By definition, the abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251046

    Validate with FDA (Live)

    Date Cleared
    2025-08-13

    (132 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. dental implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    S.I.N. Dental Implant System implants with lengths of 18, 20, 22 or 24mm may be tilted up to 30º. When used in the mandible or maxilla with implants with lengths of 18, 20, 22 or 24 mm at an angulation of 30º, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22 or 24 mm at angulations between 0º and less than 30º, the S.I.N. Dental Implant System implants are only indicated for multiple unit restoration in splinted applications that utilize at least two implants.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes various endosseous implants with 16° Morse tapers interface connection. This submission adds to the S.I.N. Dental Implants System, the Versalis S implants line.

    The Versalis S implant line has the Versalis S and Versalis S Plus implants, and both have a Morse taper connection with an internal 16° cone taper. The Versalis S implants have implant body lengths from 8.5 to 24mm and body diameter Ø from 3.5 to 7.0mm., they are manufactured from unalloyed titanium conforming to ASTM F67, with a double acid-etched surface treatment. The Versalis S Plus implants have implant body lengths from 8.5 to 24mm and body diameter Ø from 3.5 to 7.0mm., they are manufactured from unalloyed titanium conforming to ASTM F67, with a double acid-etched surface treatment plus hydroxyapatite surface coating.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700), Grade 4. The material used to manufacture the dental implants in this submission (conforming to ASTM F67) is identical to the material used to manufacture the device implants cleared in primary predicate K170392 and in additional predicate K222231.

    AI/ML Overview

    The provided text is a 510(k) clearance letter for a dental implant system. It does not contain any information about acceptance criteria or a study proving that an AI/ML device meets acceptance criteria. The "Performance Data" section discusses non-clinical testing for the dental implant itself, such as sterilization validation, shelf-life testing, and biocompatibility, as well as fatigue testing and MRI environment compatibility. This is for a physical medical device, not a software device that relies on AI/ML.

    Therefore, I cannot provide the requested information as it is not present in the given document.

    To reiterate, the document is about a physical dental implant system and the performance data pertains to its physical properties, sterility, and biocompatibility, not to the performance of an AI/ML algorithm.