510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DAVID One Step Home Use Pregnancy Test Strip is an in-Vitro diagnostic test device for the qualitative detection of human chorionic gonadotropin (HCG) in the urine. The test device is intended for use as an aid in early detection of pregnancy. in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

    This product is intended for over-the-counter use.

    DAVID One Step Home Use Pregnancy Test Cassette is an in-Vitro diagnostic test device for the qualitative detection of human chorionic gonadotropin (HCG) in the urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

    This product is intended for over-the-counter use.

    DAVID One Step Home Use Pregnancy Test Midstream is an in-Vitro diagnostic test device for the qualitative detection of human chorionic gonadotropin (HCG) in the urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. This product is intended for over-the-counter use only.

    DAVID One Step Prescription Pregnancy Test Strip is an in-Vitro diagnostic test device for the qualitative detection of human chorionic gonadotropin (HCG) in the urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

    This product is intended for prescription use.

    DAVID One Step Prescription Pregnancy Test Cassette is an in-Vitro diagnostic test device for the qualitative detection of human chorionic gonadotropin (HCG) in the urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. This product is intended for prescription use.

    Device Description

    DAVID One Step Pregnancy Test is designed in three formats: strip, cassette, and midstream. Each of the devices (strip, cassette and midstream) contains a pouch with the test device and instructions for use, additional test cassette also contains a disposable plastic dropper. The test cassette and test midstream consist of a chromatographic test strip enclosed in plastic housing. The test strip contains mouse monoclonal anti-ß-hCG antibody, mouse monoclonal anti-a-hCG antibody, goat anti-mouse IgG polyclonal antibody and colloidal gold.

    The added plastic housing does not affect the performance of the product, which is determined by the test strip. The only difference between the different formats is test procedure.

    The test uses two lines to indicate results. The appearance of two color bands, one at the "Test line" and one at the "Control line" region means the test is "positive". The appearance of a color band at the "Control line" region but not at the "Test line" region means the test is "negative". If there are no colored bands in the "Control line" area and "Test line" region or if there is no color band in "Control line" region even there is a band in the "Test line" region; this means "invalid" test result.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies performed, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in a formal table with pass/fail metrics. However, based on the performance data presented, the implicit acceptance criteria for a qualitative pregnancy test device would be high accuracy, analytical sensitivity, and no significant interference. The reported performance is summarized below:

    Criteria/Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    Analytical Sensitivity (Lowest concentration for 100% positive)Achieve 100% positive results at 10 mIU/mL and 50% positive at 7.5 mIU/mL.Strips, Cassettes, Midstreams (all methods): 100% positive at 10 mIU/mL, ~48-50% positive at 7.5 mIU/mL (across 3 lots and different operators/methods).
    Specificity (Cross-reactivity)No cross-reactivity with LH (500mIU/mL), FSH (1000mIU/mL), and TSH (1000μIU/mL).No cross-reaction observed with LH at 500mIU/mL, FSH at 1000mIU/mL, and TSH at 1000μIU/mL.
    Specificity (Interference from common substances)No interference from specified endogenous and exogenous compounds at relevant concentrations.No interference observed from various compounds including Glucose, Albumin, Hemoglobin, Bilirubin, Estriol, common birth control pills, progestational drugs, prescription/OTC substances (Acetaminophen, Aspirin, Caffeine, etc.), antibiotics (Amikacin, Penicillin G, etc.) at specified concentrations.
    Specificity (pH Effects)No interference across a physiological urine pH range.No interference observed with urine pH ranging from 4 to 9.
    Specificity (Specific Gravity Effects)No interference across a physiological urine specific gravity range.No interference observed with urine specific gravity ranging from 1.000 to 1.050.
    High Dose Hook EffectNo hook effect up to a very high hCG concentration.No hook effect observed up to 2,000,000 mIU/mL hCG.
    Interference from hCG ß-core FragmentLimited or no interference from hCG ß-core fragment.Interference demonstrated at concentrations of 1,000,000 pmol/L.
    Method Comparison with Predicate DeviceHigh agreement with the predicate device.100% agreement (Positive: 52-73, Negative: 47-68) with the predicate device for all formats.
    Clinical Sensitivity in Early PregnancyHigh detection rate on or after the expected period, and acceptable detection rate in early days before the expected period.Strips: 100% from EMP to EMP+2, declining to 0% at EMP-10. Cassettes: 100% from EMP-2 to EMP+2, declining to 0% at EMP-10. Midstreams (dip): 100% from EMP-1 to EMP+2, declining to 0% at EMP-10. Midstreams (simulated stream): 100% from EMP-3 to EMP+2, declining to 1% at EMP-10.
    Lay User Study (Accuracy)High agreement between lay users and professionals.100% agreement between lay users and professionals for all formats (50 positive, 70 negative cases). For spiked samples, 100% positive at 10mIU/mL and 11.5mIU/mL, 1-2% positive at 5mIU/mL, and 43-46% positive at 8mIU/mL.
    Lay User Study (Usability)Easy to use, understandable labeling, correct interpretation of results by lay users.Questionnaire results reflect that consumers found the device easy to use and had no trouble understanding labeling or interpreting results. Flesch-Kincaid reading analysis indicated an 8th-grade reading level for instructions.
    Non-pregnant Urine Sample AnalysisLow to zero false positive rate in non-pregnant individuals.0% false positive in pre-menopausal and menopausal groups, 1 false positive out of 300 (0.33%) in the post-menopausal group (>55 years old) due to naturally higher hCG levels.

    2. Sample Size Used for the Test Set and Data Provenance

    • Analytical Performance (Precision/Reproducibility): For each hCG concentration, 50 samples were tested per lot, across 3 lots, for each of 3 formats (strips, cassettes, midstreams). This totals 450 samples per hCG concentration for each format (50 samples/lot * 3 lots * 3 sites * 1 investigator/site * 5 days = not explicitly detailed, but 50/50 for each lot implies 50 samples). This was conducted at 3 sites.
    • Analytical Performance (Detection Limit): 30 samples for each concentration, for 10 concentrations, per lot, across 3 lots, for each of 3 formats via 4 test methods (strip, cassette, midstream dip, midstream stream). (30 samples/concentration * 3 lots * 4 test methods = 360 samples per concentration level). Urine samples were from healthy non-pregnant women.
      • Provenance: Data appears to be prospective as samples were spiked and tested. The country of origin for the samples is not explicitly mentioned but the applicant is from China, so it's likely China.
    • Method Comparison with Predicate Device: 120 urine samples tested simultaneously with the predicate device and the proposed devices.
      • Provenance: Not specified if retrospective or prospective, but likely prospective for comparison.
    • Detection of hCG in Early Pregnancy Clinical Samples: 1287 urine samples collected from 99 different women (20-40 years old).
      • Provenance: Prospective collection from women followed throughout their conception cycles. Specific country of origin is not mentioned, but given the applicant, likely China.
    • Lay User Study:
      • First study (professional vs. non-professional comparison): 120 volunteers per site (50 pregnant, 70 non-pregnant) across 3 sites (one format per site). This totals 360 volunteers.
      • Second study (spiked samples with lay users): 120 volunteers per site, detecting 4 spiked samples, across 3 sites.
      • Provenance: Prospective, conducted on Chinese mainland with women volunteers (20-40 years old).
    • Non-pregnant Urine Sample Analysis: 900 samples tested across three clinical trial sites, with 100 samples from each of three age groups (pre-menopausal, menopausal, post-menopausal) per site, per format (not explicitly stated if per format, but described as "one format of device from 3 lots").
      • Provenance: Not specified if retrospective or prospective. Likely collected specifically for the study.

    3. Number of Experts and Qualifications for Ground Truth

    • Clinical Samples (Early Pregnancy): "laboratory technicians (one for strip, one for cassette format, and one for midstream for both dip and stream methods)" performed the initial testing. The ultimate ground truth for positive samples was confirmed with "ultrasound scan." No specific number or qualifications for the laboratory technicians or sonographers are provided beyond "laboratory technicians."
    • Lay User Study: "professionals" tested the simultaneously collected urine samples, and their results were used as ground truth for comparison with lay users. No specific number or qualifications of these "professionals" are detailed.
    • Non-pregnant Urine Sample Analysis: "professionals" tested the samples. For the false positive in the post-menopausal group, the concentrations of these samples "were determined to be higher than the cut-off of the device by quantitative test," implying a quantitative lab test was the ground truth for that specific follow-up.

    4. Adjudication Method for the Test Set

    • Analytical Performance (Precision/Reproducibility & Detection Limit): "1 investigator" per site for precision tests, and "12 laboratory operators" divided into 4 groups for detection limit tests. There is no explicit mention of an adjudication method among multiple readers for these analytical studies. Results are aggregated.
    • Clinical Samples (Early Pregnancy): One laboratory technician per format/method combination performed the initial testing. The ultimate confirmation for positive results was "ultrasound scan." There is no mention of explicit adjudication among multiple human readers for initial interpretations.
    • Lay User Study: Professionals served as the ground truth for comparison with lay users. There is no mention of an adjudication method among these professionals if multiple were involved in reading the "ground truth" results for each sample. For the lay user study where spiked samples were tested by lay users, the ground truth was the known concentration of hCG in the spiked samples.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No formal MRMC comparative effectiveness study was done to compare human readers with and without AI assistance. This device is a rapid diagnostic test (lateral flow immunoassay) that provides a visual qualitative result (lines present or absent) directly to the user or a healthcare professional, not an AI-powered image analysis system. Therefore, the concept of "AI assistance" in the context of improving human reader performance is not applicable here.

    6. Standalone Performance

    Yes, a standalone performance was done. The entire analytical and clinical performance sections detail the algorithm's (device's) performance without human-in-the-loop directly influencing the result generation. For instance, the analytical sensitivity, specificity, and comparison with a predicate device are all examples of the device's standalone performance. The "lay user study" and "non-pregnant urine sample analysis" also contribute to understanding the device's performance when interpreted by humans (lay users or professionals).

    7. Type of Ground Truth Used

    • Analytical Performance:
      • Precision/Reproducibility & Detection Limit: Known concentrations of spiked HCG standard in negative urine samples.
      • Cross-reactivity & Interference: Known concentrations of interfering substances spiked into negative and positive (10 mIU/mL hCG) urine samples.
      • pH & Specific Gravity: Known pH and specific gravity values of urine samples.
      • High Dose Hook Effect: Known high concentrations of hCG.
    • Clinical Samples (Early Pregnancy): "Ultrasound scan" confirmed positive pregnancy. For non-pregnant cases, it implies absence of ultrasound-confirmed pregnancy.
    • Lay User Study:
      • Professional Comparison: Results from testing by "professionals" using the same urine samples as lay users.
      • Spiked Samples: Known concentrations of hCG in the spiked urine samples.
    • Non-pregnant Urine Sample Analysis: Absence of pregnancy (based on age group and general health status) and, for the false positive case, validation by "quantitative test" for hCG concentration.

    8. Sample Size for the Training Set

    The provided document describes premarket testing and clinical studies to evaluate the device. It does not mention a training set or a machine learning (AI) algorithm that would typically require such a set. This is a traditional in-vitro diagnostic device.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set for an AI algorithm is mentioned, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172627

    Validate with FDA (Live)

    Date Cleared
    2018-03-28

    (208 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    David Home Pregnancy Test Cassette is an in-vitro diagnostic test for the qualitative determination of human chorionic gonadotropin (HCG) in the urine to aid in early detection of pregnancy, based on lateral flow immunoassay. It is intended for the Over-the-Counter Use only.

    David Professional Pregnancy Test Cassette is an in-vitro diagnostic test for the qualitative determination of human chorionic gonadotropin (HCG) in the urine to aid in early detection of pregnancy, based on lateral flow immunoassay. It is intended for prescription use only.

    Device Description

    David Pregnancy Test Cassette will be designed in one format:Cassette.The Cassette kit consists of one test device, a disposable plastic dropper, and a package insert. Each test strip contains mouse monoclonal anti-a-hCG antibody coated membrane and a pad containing mouse monoclonal anti-ß-hCG antibody colloidal gold conjugate. The control antibodies are goat anti-mouse IgG polyclonal antibody.

    AI/ML Overview

    The provided document describes the performance characteristics and studies for the "David Home Pregnancy Test Cassette" and "David Professional Pregnancy Test Cassette".

    Here's an analysis of the acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    | Acceptance Criteria Category | Specific Criteria | Reported Device Performance and Results |
    |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
    | Sensitivity (Detection Limit) | The positive ratio of samples containing concentrations higher than 25 mIU/mL HCG should be more than 95%. | 100% positive ratio for samples with 25 mIU/mL HCG and above (for both professional and home use, detected by both professional and lay users). At 22.5 mIU/mL HCG, positive ratios were 50% (professional), 56% (home by professional), and 45% (home by lay user). The minimum detection limit is stated as 25 mIU/mL HCG. |
    | Specificity (Cross-reactivity) | No cross-reaction with LH, FSH, and TSH at specified high concentrations. | No cross-reaction observed with LH at 500 mIU/mL, FSH at 1000 mIU/mL, and TSH at 1000 μIU/mL for both 0 mIU/mL HCG and 25 mIU/mL HCG spiked samples across 3 lots. |
    | Interference | No interference from various exogenous compounds at specified concentrations. | No interference observed from a list of substances (e.g., Glucose, Albumin, Bilirubin, Hemoglobin, Acetaminophen, Aspirin, etc.) at their tested concentrations, in both negative and 25 mIU/mL HCG positive samples. |
    | pH Interference | No interference across a specified pH range. | pH ranging from 3 to 10 in urine samples does not interfere with the performance of the test in negative and 25 mIU/mL HCG positive samples. |
    | Specific Gravity Interference | No interference across a specified specific gravity range. | Specific gravity ranging from 1.000 to 1.050 in urine samples does not interfere with the performance of the test in negative and positive urine. |
    | High Dose Hook Effect | No high dosage hook effect up to a very high HCG concentration. | No hook effect observed up to 2,000,000 mIU/mL hCG. |
    | hCG ß-core fragment effects | No interference from high levels of beta core fragment. | No hook effect observed at an hCG ß-core fragment concentration up to 100,000 pmol/L. |
    | Precision/Reproducibility | Good repeatability between lots and within lot. For concentrations higher than 25 mIU/mL HCG, all results should be positive. For concentrations below 20 mIU/mL HCG, 98% results should be negative. | For HCG concentrations 25 mIU/mL and above, 100% positive results were obtained across all 3 lots (150/150 for each lot). For HCG concentrations 16 mIU/mL and below, 100% negative results were obtained (150/150 for each lot). At 18.75 mIU/mL, 1/150 positive results were observed across all 3 lots. At 20 mIU/mL, 5/150, 3/150, and 2/150 positive results were observed across lots 1, 2, and 3 respectively, meaning >97% negative results. The study concluded good repeatability. |
    | Method Comparison (Professional vs. Predicate) | 100% agreement between the candidate device (Professional) and the predicate device. | Positive, Negative, and Total coincidence rates were all 100% (153 positive, 207 negative, total 360 samples). |
    | Method Comparison (Home vs. Predicate) | 100% agreement between the candidate device (Home) and the predicate device. | Positive, Negative, and Total coincidence rates were all 100% (153 positive, 207 negative, total 360 samples). |
    | Lay-user Concordance | High concordance between lay-user results and professional results for the Home Pregnancy Test Cassette. Lay users find the test easy to use and understand instructions/labeling. | Positive, Negative, and Total coincidence rates between non-professionals and professionals were all 100% (153 positive, 207 negative, total 360 samples). Lay-user study with spiked samples showed 100% positive for 27.5 mIU/mL HCG and 99% negative for 20 mIU/mL HCG. Questionnaire results indicated lay-users found the test easy to use and understood labeling. Flesch-Kincaid showed readability for 8th-grade education. |

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility Study Test Set:

      • Sample Size: For each HCG concentration level (0, 16, 18.75, 20, 25, 27.5, 100 mIU/mL), 150 tests were performed per lot, across 3 lots. Total tests per concentration = 150 * 3 = 450. Total tests in this study = 7 concentrations * 450 tests/concentration = 3150 tests.
      • Data Provenance: Samples were urine spiked with HCG. The study was conducted at three (3) sites. The country of origin is not explicitly stated for the samples themselves but the applicant is from China. It is a prospective study (samples were used in the study, implying prepared for the study).
    • Detection Limit Study Test Set:

      • Sample Size: For each HCG concentration level (0, 12.5, 16, 18.75, 20, 22.5, 25, 27.5, 50, 100 mIU/mL), 90 tests were performed at each of the 3 test sites (total 270 tests per concentration). Each 90 tests at a site comprised:
        • Professional Test Kit: 30 tests by professionals
        • Home Use Test Kit (by professionals): 30 tests by professionals
        • Home Use Test Kit (by lay users): 30 tests by lay users
      • Data Provenance: Samples were urine spiked with HCG. The study was conducted at three (3) sites. Prospective nature, similar to the precision study.
    • Analytical Specificity (Cross-reactivity) Test Set:

      • Sample Size: 20 fresh urine specimens (10 healthy non-pregnant women, 10 healthy men) were used for each hormone concentration (LH, FSH, TSH) spiked into negative (0 mIU/mL HCG) and positive (25 mIU/ml HCG) samples. This was done for 3 lots.
      • Data Provenance: Urine specimens from healthy non-pregnant women and men. Prospective.
    • Interference Study Test Set:

      • Sample Size: Not explicitly stated, but "Negative and positive hCG urine samples (25 mIU/mL) were individually spiked with the substances listed... Three devices of each format were tested."
      • Data Provenance: Urine samples spiked with HCG and interfering substances. Prospective.
    • pH Interference Study Test Set:

      • Sample Size: Not explicitly stated, but "Negative and positive hCG urine samples (25 mIU/mL) were tested across a pH range of 3-10."
      • Data Provenance: Urine samples pH adjusted and spiked with or without HCG. Prospective.
    • Specific Gravity Interference Study Test Set:

      • Sample Size: Not explicitly stated, but "The device was challenged with negative urine and positive urine with specific gravity from 1.000 to 1.050."
      • Data Provenance: Urine samples with adjusted specific gravity and spiked with or without HCG. Prospective.
    • High Dose Hook Effect Study Test Set:

      • Sample Size: Not explicitly stated, but "Negative urine specimens were spiked with high hCG concentration of ranging from 50 mIU/mL to 2000,000 mIU/mL. All samples yielded the expected positive results."
      • Data Provenance: Urine samples spiked with high HCG concentrations. Prospective.
    • Effects of hCG ß-core fragment Test Set:

      • Sample Size: Not explicitly stated how many actual urine samples, but "Negative urine hCG (containing 0 mIU/mL hCG,16mIU/mL HCG) and positive urine samples (containing 27.5 mIU/mL hCG) were spiked with hCG beta core fragment standard... Samples of each concentration were tested 10 times with the professional and lay-user respectively." (total 20 tests per concentration per lot, across 3 lots)
      • Data Provenance: Urine samples spiked with HCG and beta-core fragment. Prospective.
    • Method Comparison Study Test Set:

      • Sample Size: 360 fresh women urine samples.
      • Data Provenance: "fresh women urine samples". Country of origin not specified, but likely from China given the applicant. This seems to be a prospective collection or at least fresh samples used in a current study.
    • Lay-user Study Test Set:

      • Sample Size: 120 female volunteers (40 from each of three sites) for their own urine samples. Additionally, each volunteer tested 3 other double-blind numbered samples spiked with HCG (concentrations not specified but one example given is 27.5mIU/mL and 20mIU/mL HCG). This implies 120 * 3 = 360 spiked tests.
      • Data Provenance: "fresh women urine samples" and urine samples spiked with HCG. Prospective, involving human volunteers.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of "experts" used to establish the ground truth in the traditional sense of medical image interpretation or pathology review.

    • For analytical performance studies (Precision, Detection Limit, Specificity, Interference, etc.), the "ground truth" for HCG concentration is established by the known concentrations of spiked HCG in the urine samples, traceable to the WHO International Standard.
    • For the Method Comparison Study, the "ground truth" is the result from the predicate device (Wondfo One Step HCG Urine Pregnancy Test, K043443) when tested by professionals. The results from the candidate device (both professional and home use) were compared against this predicate result.
    • In the Precision/Reproducibility and Detection Limit studies, "9 operators" and "3 operators" (plus 30 volunteers for detection limit) were involved. Their qualifications are not specified beyond being "operators" or "professionals".
    • For the Lay-user study, the ground truth for the volunteers' own samples would be their actual pregnancy status (which would likely be confirmed by clinical methods or laboratory testing, though not specified as such in this document). For the spiked samples, the ground truth is the known spiked HCG concentration.

    Therefore, for the analytical studies, the ground truth is based on known sample characteristics. For the method comparison, it's based on a predicate device's performance handled by "professionals". "Experts" with specific qualifications (e.g., radiologists) are not applicable to this type of in-vitro diagnostic device.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used in studies where human readers independently interpret images or clinical data, and then discrepancies are resolved. This is not directly applicable to a lateral flow immunoassay pregnancy test, which provides a qualitative "positive" or "negative" result based on visual lines.

    • For the Precision/Reproducibility and Detection Limit studies: The results are reported as positive or negative counts across multiple tests. There's no mention of an adjudication process to resolve differing readings on the same test, suggesting direct reading of the test results.
    • For the Method Comparison Study: The comparison is directly between the result of the candidate device and the predicate device. If there were discrepancies, the document does not describe a formal adjudication process; instead, it reports a 100% agreement, implying no discrepancies or that any were resolved instantly during the testing by professionals.
    • For the Lay-user study: Lay users conducted the test according to instructions. The results from their tests were then compared to reference (likely professional reading of their own samples and known HCG content of spiked samples). No formal "adjudication" in the sense of a third-party expert review of conflicted readings is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    An MRMC comparative effectiveness study is not applicable to this device. This device is an in-vitro diagnostic test for HCG, not an AI-assisted diagnostic imaging device that involves interpretation by human readers. Therefore, there is no AI component or human reader improvement with AI to report effect sizes for.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    This question is not applicable. The device is a lateral flow immunoassay. Its "performance" is determined by the biochemical reaction on the strip and the visual interpretation of lines, which is inherently a "human-in-the-loop" visual interpretation (even for the "home use" version). There is no "algorithm only" component.

    7. The Type of Ground Truth Used

    • For analytical performance studies (Precision, Detection Limit, Specificity, Interference, Hook Effect, Beta-core fragment): The ground truth was based on known concentrations of HCG or other substances spiked into urine samples, traceable to the WHO International Standard (5th WHO Chorionic Gonadotrophin, NIBSC code: 07/364).
    • For the Method Comparison Study: The ground truth was established by the results from a legally marketed predicate device (Wondfo One Step HCG Urine Pregnancy Test) when tested by professionals.
    • For the Lay-user study: The ground truth for volunteer's own samples would implicitly be their actual clinical pregnancy status, but this isn't detailed. For the spiked samples used in the lay-user study, the ground truth was the known spiked HCG concentration.

    8. The Sample Size for the Training Set

    The document describes performance studies for device validation. It does not mention any "training set" in the context of an algorithm or AI model development. This is an IVD device, not a machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for an algorithm.