Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K243601

    Validate with FDA (Live)

    Date Cleared
    2025-05-19

    (179 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical HiPoint Reperfusion System and Aspiration Tubing Set, when used with a compatible vacuum pump, is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease within the internal carotid artery and middle cerebral artery M1 segment within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    Device Description

    The Aspiration Catheter, Delivery Catheter, and Base Camp Sheath comprise the Route 92 Medical HiPoint Reperfusion System. The Route 92 Medical HiPoint Reperfusion System is used with the Route 92 Medical Aspiration Tubing Set and the Route 92 Medical Base Camp Sheath System.

    The Aspiration Catheter has a single-lumen, coil-reinforced polymer distal shaft section connected to a proximal control wire. The Delivery Catheter is a single-lumen variable stiffness catheter. Both catheters have hydrophilic coating. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    The Route 92 Medical Aspiration Tubing Set is made from high-pressure tubing. An on/off switch is provided to control fluid flow. A suction connector allows connection to an aspiration pump canister and a Luer fitting allows connection to the aspiration catheter.

    For the aspiration source, the Aspiration Catheter is used in conjunction with an aspiration pump with prespecified performance parameters that is connected using the Route 92 Medical Aspiration Tubing Set.

    The Route 92 Medical Base Camp Sheath System is comprised of a Sheath, a Dilator, a Navigating Catheter, and an RHV (rotating hemostasis valve). The Sheath is a single-lumen, variable stiffness catheter with a radiopaque marker on the distal end. The inner lumen of the catheter is compatible with 8 French (F) or smaller catheters. The Dilator may be placed within the Sheath to facilitate percutaneous introduction of the Sheath into a femoral artery. The Dilator has a radiopaque marker at the distal tip. The Navigating Catheter is a single-lumen, variable stiffness catheter with a radiopaque marker at the distal tip. The Navigating Catheter is compatible with the Sheath and has a shaped distal end to facilitate placement. All of the catheters are coated with hydrophilic coating.

    AI/ML Overview

    The provided text describes a 510(k) clearance for the Route 92 Medical HiPoint Reperfusion System and Aspiration Tubing Set. This is a medical device for treating acute ischemic stroke, not an AI/software device. Therefore, the requested information regarding AI acceptance criteria, data provenance, expert ground truth, MRMC studies, standalone performance, and training set details for an AI model cannot be extracted from this document, as it does not pertain to an AI device.

    The document focuses on non-clinical and clinical studies (SUMMIT MAX trial) to demonstrate the safety and effectiveness and substantial equivalence of the mechanical device to a predicate device.

    However, I can provide the acceptance criteria and performance data for the mechanical device as presented in the document:


    Device: Route 92 Medical HiPoint Reperfusion System and Aspiration Tubing Set

    Study: SUMMIT MAX, a prospective, multi-center, randomized, controlled, interventional, open-label clinical trial.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Primary Endpoint)Target/Threshold for Non-InferiorityReported Device Performance (Route 92 Medical)Reported Predicate Performance (AXS Vecta)Result (Difference, 95% CI)P-value
    Primary Effectiveness Endpoint: Successful arterial revascularization (mTICI ≥ 2b) using only the assigned study device (adjunctive therapy considered a failure).Non-inferiority margin of 12.5% vs. predicate77.5% (69/89)50.6% (39/77)26.9% (12.7%, 41.0%)< 0.0001 (one-sided)
    Primary Safety Endpoint: Symptomatic intracranial hemorrhage (sICH) within 24 hours post-procedure (Heidelberg bleeding classification).Non-inferiority margin of 12.5% vs. predicate3.6% (3/84)2.7% (2/75)0.9% (-4.5%, 6.3%)< 0.0001 (one-sided)

    Note: The efficacy endpoint was met with superiority (implied by p-value < 0.0001 in superiority context and the large positive difference outside the non-inferiority margin), and the safety endpoint met non-inferiority.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Effectiveness Endpoint (primary):
        • Route 92 arm: N=89
        • Vecta arm: N=77
      • Safety Endpoint (primary):
        • Route 92 arm: N=84 (denominator for sICH, based on sICH calculation out of 89 total patients in Route 92 arm and 77 in Vecta arm)
        • Vecta arm: N=75 (denominator for sICH, based on sICH calculation out of 89 total patients in Route 92 arm and 77 in Vecta arm)
      • The "mITT" (Modified Intent to Treat) cohort comprised 166 participants (N=89 Route 92 arm, N=77 Vecta arm), which was used for primary endpoint analysis.
    • Data Provenance: Prospective, multi-center clinical trial. Country of origin not explicitly stated but implied to be multinational given "applicable country regulations" and "IRB / Ethics Committee" implies regulatory approval beyond a single nation. The medical history table includes "Maori" as a race category, suggesting participation from New Zealand.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • For mTICI (Effectiveness Endpoint): Adjudicated by an "independent blinded core lab." Specific number and qualifications of experts not detailed.
    • For sICH (Safety Endpoint): Identified by the "core lab" and adjudicated by the "clinical events committee." Specific number and qualifications of experts not detailed beyond being "independent" and "blinded."

    4. Adjudication Method for the Test Set

    • Effectiveness (mTICI): "Adjudicated by an independent blinded core lab."
    • Safety (sICH): "Identified by the core lab and adjudicated by the clinical events committee."
    • The specific structure of the adjudication (e.g., 2+1, 3+1) is not provided. It simply states "core lab" and "clinical events committee" for adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This was a clinical trial comparing two mechanical devices (reperfusion systems), not an AI device. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a mechanical medical device, not an algorithm or AI.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Effectiveness: Angiographic outcome (modified Thrombolysis in Cerebrovascular Infarction (mTICI) score ≥ 2b), adjudicated by an independent blinded core lab. This can be considered a form of expert consensus/adjudicated imaging outcome.
    • Safety: Clinical outcome (symptomatic intracranial hemorrhage), identified by an independent blinded core lab and adjudicated by an independent blinded clinical events committee. This is based on clinical events and imaging findings interpreted by experts.
    • Other Outcomes: Modified Rankin Scale (mRS) at 90 days (patient-reported outcome/clinical assessment), time from groin puncture to final angiogram, etc., all based on clinical trial data collection.

    8. The Sample Size for the Training Set

    • Not applicable. This is a mechanical device, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. This is a mechanical device, not an AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240529

    Validate with FDA (Live)

    Date Cleared
    2024-10-02

    (222 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 8F Modified Sheath System is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The 8F Modified Sheath System is comprised of a Sheath, a Navigating Catheter, and an RHV (rotating hemostasis valve). The Sheath is a single-lumen, variable stiffness catheter with a radiopaque marker on the distal end. The inner lumen of the catheter is compatible with 8F or smaller catheters. The Dilator may be placed within the Sheath to facilitate percutaneous introduction of the Sheath into a femoral artery. The Dilator has a radiopaque marker at the distal tip. The Navigating Catheter is a singlelumen, variable stiffness catheter with a radiopaque marker at the distal tip. The Navigating Catheter is compatible with the Sheath and has a shaped distal end to facilitate placement. All of the catheters are coated with hydrophilic coating.

    AI/ML Overview

    The provided text describes the 8F Modified Sheath System and its substantial equivalence to a predicate device (Base Camp® Sheath System, K191717) for FDA clearance. The document focuses on non-clinical testing to demonstrate performance.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various performance tests and their outcomes, uniformly indicating "PASS" and that "The pre-determined acceptance criteria were met." However, it does not explicitly state the specific quantitative acceptance criteria for each test. It only reports that the device met them.

    TestTest MethodReported Device Performance
    Dimensional VerificationDevice dimensions were measured to confirm conformance to the specifications.PASS
    Luer IntegrityTested per ISO 80369-7:2016.PASS
    Tensile StrengthThe tensile strength of the catheter sections and bonds was tested.PASS
    Kink ResistanceTest specimen segments were formed into a defined bend diameter to evaluate kink resistance.PASS
    Torsion ResistanceThe test specimens were rotated to evaluate integrity after rotation.PASS
    Tip FlexibilityTest specimens were tested for tip flexibility.PASS
    Air LeakageTested per ISO 10555-1:2013 Annex D.PASS
    Liquid LeakageTested per ISO 10555-1:2013 Annex C.PASS
    Hydrophilic Coating IntegrityThe integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles in a vascular model.PASS
    Simulated Use TestingDeliverability and compatibility with accessory devices were evaluated in a vascular model.PASS
    Static Burst Pressure TestingTested per ISO 10555-1:2013 Annex F.PASS
    Particulate TestingThe size and quantity of particulate matter generated during simulated use testing were determined.PASS
    Delivery/Retrieval Force TestingThe forces to deliver and retrieve the subject device in an anatomical model were measured.PASS

    Biocompatibility Testing Acceptance

    For biocompatibility, the acceptance criteria are generally qualitative for these types of tests and are implicitly met when the "Conclusion" states no adverse reactions or that requirements were met.

    TestConclusion
    SensitizationTest article did not elicit a sensitization response.
    IrritationRequirements of the ISO intracutaneous reactivity test have been met for the test article.
    Acute Systemic ToxicityRequirements of the ISO acute systemic injection test have been met for the test article.
    Material Mediated PyrogenicityThe test article met the requirements for the absence of pyrogens.
    CytotoxicityThe test article is non-cytotoxic.
    Hemocompatibility – ASTM HemolysisThe test article is considered non-hemolytic.
    Hemocompatibility - Thrombogenicity: ASTM Partial Thromboplastin Time AssayThe test article met the requirements of the test.
    Hemocompatibility - Complement Activation AssayThe test article is a non-activator of the complement system.
    Hemocompatibility - Thrombogenicity: ASTM Platelet/Leukocyte CountThe test article met the requirements of the test.
    Hemocompatibility - Thrombogenicity: Comparative Surface AssessmentThe test article does not have any features that would increase thrombogenicity risk.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes used for each of the performance or biocompatibility tests. It refers to "test specimens" or animals (e.g., "guinea pig," "rabbits," "mice") for biocompatibility, but without specific numbers.

    • Data Provenance: The data provenance is from non-clinical bench testing and laboratory animal testing. There is no information provided about the country of origin, but given it's an FDA submission, the tests were conducted to international standards (e.g., ISO, ASTM) which are globally recognized. All testing appears to be prospective (conducted specifically for this submission).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This is a submission for a medical device (catheter system) where "ground truth" in the clinical sense (e.g., diagnosis, outcome) is not relevant for the non-clinical and biocompatibility studies presented. The "truth" or acceptance is determined by predefined engineering specifications and regulatory standards (ISO, ASTM). The document does not mention the use of human experts to establish ground truth for this type of testing.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human interpretation of data (e.g., medical imaging) to resolve discrepancies. The studies described are objective performance and biocompatibility tests where a result either meets or does not meet a predefined standard.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (catheter system), not an AI/software device. No MRMC study was conducted or is relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    As mentioned in point 3, the concept of "ground truth" in the clinical diagnostic sense is not applicable here. For the performance and biocompatibility studies, the "ground truth" is defined by:

    • Pre-determined engineering specifications: For dimensional verification, tensile strength, kink resistance, etc.
    • International standards and regulatory guidelines: Such as ISO 80369-7:2016 for Luer Integrity, ISO 10555-1:2013 for leakage and burst pressure, and ISO 10993-1 for biocompatibility.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of this traditional medical device submission. Training sets are relevant for machine learning or AI models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233329

    Validate with FDA (Live)

    Date Cleared
    2024-04-23

    (207 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set is indicated for use in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.

    Device Description

    The 054 Aspiration Catheter and Tenzing 5 Delivery Catheter comprise the Route 92 Medical Full Length 054 Reperfusion System. The Route 92 Medical Full Length 054 Reperfusion System is used with the Route 92 Medical Aspiration Tubing Set.

    The 054 Aspiration Catheter is a single-lumen, coil-reinforced variable stiffness catheter. The Tenzing 5 Delivery Catheter is a single-lumen variable stiffness catheters are hydrophilically coated. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    The Route 92 Medical Aspiration Tubing Set is made from high-pressure tubing. An on/off switch is provided to control fluid flow. A suction connector allows connection to an aspiration pump canister and a Luer fitting allows connection to the aspiration catheter.

    For the aspiration source, the Aspiration Catheter is used in conjunction with an aspiration pump with prespecified performance parameters that is connected using the Route 92 Medical Aspiration Tubing Set, along with a legally marketed canister and accessories kit.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA regarding the Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving that the device meets specific performance acceptance criteria based on clinical outcomes or expert-adjudicated ground truth.

    Therefore, many of the requested elements for describing an acceptance criteria study, especially those related to AI algorithm performance (e.g., sample size for test sets, data provenance, number of experts for ground truth, MRMC studies, standalone performance), are not applicable to this type of regulatory submission. This document describes non-clinical testing to ensure the device's functional performance, safety, and compatibility.

    Here's an analysis of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria values for each test, but rather generally states "All samples met the pre-determined acceptance criteria." The performance is reported as "PASS" for all tests.

    TestTest MethodAcceptance Criteria (Implied)Reported Device Performance
    Shelf Life and SterilityEthylene oxide sterilization cycle verified to a sterility assurance level of 1 x 10⁻⁶ in accordance with ISO 11135. Aging studies established functional integrity for labeled expiration date.Sterility assurance level of 1 x 10⁻⁶; functional for 6 months.PASS
    Biocompatibility TestingNot performed again as materials were unchanged from reference device K223530.None (re-use of prior assessment)Not applicable
    Simulated Use TestingDeliverability, clot retrieval, device integrity, kink and aspiration resistance, lumen patency, durability and compatibility with accessory devices evaluated in a neurovascular model.Meet pre-determined functional criteria in a neurovascular model.PASS
    Lumen Patency TestingTest specimens tested for lumen patency under vacuum.Meet pre-determined lumen patency criteria under vacuum.PASS
    Dimensional VerificationDevice dimensions measured to confirm conformance to specifications.Conformance to specifications.PASS
    Luer IntegrityTested per ISO 80369-7:2021.Conformance to ISO 80369-7:2021.PASS
    Tensile StrengthTensile strength of catheter sections and bonds tested.Meet pre-determined tensile strength criteria.PASS
    Kink ResistanceTest specimen segments formed into a defined bend diameter to evaluate kink resistance.Meet pre-determined kink resistance criteria.PASS
    Torsion ResistanceTest specimens rotated to evaluate integrity after rotation.Maintain integrity after rotation.PASS
    Tip FlexibilityTest specimens tested for tip flexibility.Meet pre-determined tip flexibility criteria.PASS
    Air LeakageTested per ISO 10555-1:2013 Annex D.Conformance to ISO 10555-1:2013 Annex D.PASS
    Liquid LeakageTested per ISO 10555-1:2013 Annex C.Conformance to ISO 10555-1:2013 Annex C.PASS
    Static BurstTested per ISO 10555-1:2013 Annex F.Conformance to ISO 10555-1:2013 Annex F.PASS
    Dynamic BurstMechanical integrity assessed under pressure.Maintain mechanical integrity under pressure.PASS
    Hydrophilic Coating IntegrityIntegrity of the hydrophilic coating evaluated after multiple insertion and withdrawal cycles.Maintain coating integrity after multiple cycles.PASS
    Aspiration PressureVacuum pressure at the tip of the catheter assessed during aspiration.Meet pre-determined aspiration pressure criteria.PASS
    ParticulateSize and quantity of particulates generated during simulated use evaluated.Meet pre-determined particulate limits.PASS

    2. Sample size used for the test set and the data provenance

    The document states "All samples met the pre-determined acceptance criteria" for each test but does not specify the numerical sample sizes (N) for each test set. The data provenance is described as "Non-Clinical Testing" which implies in-vitro or laboratory testing, not human-derived data like images or patient records. Therefore, information like country of origin or retrospective/prospective is not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The tests are non-clinical, mechanical, and performance-based (e.g., tensile strength, lumen patency). The "ground truth" is based on engineering specifications and adherence to international standards (like ISO) rather than expert interpretation of medical images or clinical outcomes.

    4. Adjudication method for the test set

    Not applicable. No expert adjudication was involved as the tests are objective, non-clinical measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for revascularization, not an AI-powered diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the non-clinical tests is based on:

    • Engineering specifications: Device dimensions, tensile strength, etc.
    • International standards: ISO 11135 for sterility, ISO 80369-7:2021 for Luer integrity, ISO 10555-1:2013 for air/liquid leakage and burst strength.
    • Functional performance criteria: In a simulated neurovascular model for deliverability, clot retrieval, etc.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model. There is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231583

    Validate with FDA (Live)

    Date Cleared
    2023-08-15

    (76 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Full Length 054 Access System is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical Full Length 054 Access System is comprised of a Support Catheter and a Delivery Catheter. The Support Catheter is a single-lumen, variable stiffness catheter. The Delivery Catheter is a hubbed, single-lumen, variable stiffness catheters are hydrophilically coated. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the "Route 92 Medical Full Length 054 Access System." This document focuses primarily on the comparison of the new device to a predicate device and the results of non-clinical performance testing, rather than a study proving the device meets acceptance criteria through clinical or user-based performance.

    Therefore, many of the requested details, particularly those related to AI-based device performance, human reader studies, and ground truth establishment for patient data, are not present in this document. The document focuses on physical properties and mechanical performance of the device itself.

    Based on the provided text, here's the information that can be extracted about the acceptance criteria and the "study" (non-clinical performance testing):

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each test were "pre-determined" and the performance was reported as "PASS" for all samples, indicating they met these criteria. The specific numerical values or ranges for the acceptance criteria are not provided in this summary.

    TestAcceptance Criteria (Not Explicitly Stated, Implied "Pre-determined")Reported Device Performance
    Dimensional VerificationConformance to specificationsPASS
    Luer IntegrityPer ISO 80369-7:2016PASS
    Tensile StrengthMaintain integrity of catheter sections and bondsPASS
    Kink ResistanceNo kinking in defined bend diameterPASS
    Torsion ResistanceMaintain integrity after rotationPASS
    Tip FlexibilityMet flexibility requirementsPASS
    Air LeakagePer ISO 10555-1:2013 Annex DPASS
    Liquid LeakagePer ISO 10555-1:2013 Annex CPASS
    Static BurstPer ISO 10555-1:2013 Annex FPASS
    Dynamic BurstMechanical integrity maintained up to specified pressuresPASS
    Hydrophilic Coating IntegrityCoating integrity maintained after multiple cyclesPASS
    Simulated Use TestingDeliverability and compatibility in neurovascular modelPASS

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "All samples met the pre-determined acceptance criteria" for each test. However, the specific number of samples used for each non-clinical test is not provided. The data provenance is laboratory testing of the device itself, not human or patient data. It is a non-clinical, bench-top study.

    3. Number of Experts Used to Establish Ground Truth for Test Set and Qualifications

    Not applicable for this type of non-clinical device testing. The "ground truth" here is adherence to engineering specifications and ISO standards, evaluated through physical measurements and mechanical tests, not expert interpretation of patient data.

    4. Adjudication Method for the Test Set

    Not applicable. This was not a study involving human interpretation or adjudication. The results were based on physical measurements and performance against established engineering and ISO standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not conducted. This is a non-clinical device clearance, not an AI or diagnostic imaging study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, this is purely a mechanical/physical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests was based on pre-determined acceptance criteria derived from engineering specifications and international standards (e.g., ISO 80369-7:2016, ISO 10555-1:2013).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no corresponding ground truth establishment process.

    Summary of what the document does describe related to acceptance and proof:

    The document demonstrates that the "Route 92 Medical Full Length 054 Access System" meets its non-clinical performance and safety requirements by successfully passing a series of mechanical and physical tests. These tests are conducted to ensure the device performs as intended and is equivalent to the predicate device in terms of function and safety, given its similar design and materials but different dimensions. The "acceptance criteria" are implied to be the specific thresholds or ranges for each test (e.g., tensile strength, leak rate) that the device must meet to be considered safe and effective, and all samples passed these "pre-determined" criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223530

    Validate with FDA (Live)

    Date Cleared
    2023-04-11

    (139 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set is indicated for use in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.

    Device Description

    The Aspiration Catheter and Tenzing 7 Delivery Catheter comprise the Route 92 Medical Full Length 070 Reperfusion System. The Route 92 Medical Full Length 070 Reperfusion System is used with the Route 92 Medical Aspiration Tubing Set.

    The Aspiration Catheter is a single-lumen, coil-reinforced variable stiffness catheter. The Tenzing 7 Delivery Catheter is a single-lumen variable stiffness catheters are hydrophilically coated. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    The Route 92 Medical Aspiration Tubing Set is made from high-pressure tubing. An on/off switch is provided to control fluid flow. A suction connector allows connection to an aspiration pump canister and a Luer fitting allows connection to the aspiration catheter.

    For the aspiration source, the Aspiration Catheter is used in conjunction with an aspiration pump with prespecified performance parameters that is connected using the Route 92 Medical Aspiration Tubing Set, along with a legally marketed canister and accessories kit.

    AI/ML Overview

    The provided document, a 510(k) K223530 Premarket Notification for the Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set, does not contain information about an AI-powered device or a study involving AI assistance for human readers. It describes a medical device (a catheter system for stroke treatment) and a series of non-clinical, benchtop, and animal studies to demonstrate its safety and effectiveness.

    Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving an AI device's performance, as the document does not pertain to an AI device.

    However, I can provide a summary of the acceptance criteria and the studies conducted for the medical device described in the document, as it meets the general structure of the request for device performance and testing.


    Acceptance Criteria and Study for the Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set

    This document describes the testing and acceptance criteria for a physical medical device, not an AI-powered one. The studies focus on the physical and functional performance of the catheter system.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various tests and their corresponding results, with the acceptance criterion implicitly being "PASS" for all samples meeting pre-determined criteria.

    Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
    Shelf Life & SterilitySterilization VerificationSterility Assurance Level of 1 x 10-6Verified
    Aging Studies (Packaging & Functionality)Remain functional for labeled expiration dateMet acceptance criteria
    BiocompatibilityCytotoxicity - ISO MEM ElutionTest article non-cytotoxic (grade less than 2)Non-cytotoxic
    Sensitization - ISO Guinea Pig MaximizationNo delayed dermal contact sensitizationDid not elicit sensitization
    Irritation - ISO Intracutaneous ReactivityDifferences in mean scores <= 1.0Requirements met
    Performance TestingSimulated Use TestingDeliverability, clot retrieval, integrity etc.PASS
    (Reperfusion System)Lumen Patency TestingLumen patency under vacuumPASS
    Dimensional VerificationConformance to specificationsPASS
    Luer IntegrityMeet ISO 80369-7:2016PASS
    Tensile StrengthAcceptable tensile strengthPASS
    Kink ResistanceNo kinking at defined bend diameterPASS
    Torsion ResistanceIntegrity after rotationPASS
    Tip FlexibilityAcceptable tip flexibilityPASS
    Air LeakageMeet ISO 10555-1:2013 Annex DPASS
    Liquid LeakageMeet ISO 10555-1:2013 Annex CPASS
    Static BurstMeet ISO 10555-1:2013 Annex FPASS
    Dynamic BurstMechanical integrity at specified pressuresPASS
    Hydrophilic Coating IntegrityCoating integrity after cyclesPASS
    Performance TestingDimensional VerificationConformance to specificationsPASS
    (Aspiration Tubing Set)Tensile StrengthAcceptable tensile strengthPASS
    Lumen OvalizationLumen patency under vacuumPASS
    Joint LeaksNo leakage under vacuumPASS
    Lumen PatencyLumen patency under vacuumPASS
    Simulated Use TestingUse for clot retrieval in modelPASS
    Packaging IntegrityMeet ISO 11607-1 Part 1PASS
    Canister CompatibilityConnection under vacuumPASS
    Flow SwitchAbility to stop fluid flow under vacuumPASS
    Animal StudiesSafety & Performance (Subacute & Chronic)No unacceptable adverse events; effectiveSuccessful demonstration

    2. Sample Size and Data Provenance:

    • Sample Size: The document repeatedly states "All samples" met predetermined acceptance criteria for the non-clinical (benchtop) tests. Specific numerical sample sizes are not provided for these tests. For the animal studies, it mentions "Two chronic animal studies."
    • Data Provenance: The data is derived from controlled laboratory and animal studies, not patient data. The country of origin is not explicitly stated, but the company is based in San Mateo, California, implying the studies were likely conducted in the US or under US regulatory standards. The studies are by nature prospective, as they are designed to test the device's performance according to predefined protocols.

    3. Number of Experts and Qualifications for Ground Truth for the Test Set:

    • This device clearance is based on demonstrable physical and functional properties, guided by engineering standards and biological testing, rather than human expert interpretation of data for ground truth (like in an AI imaging study).
    • Therefore, the concept of "experts establishing ground truth for the test set" in the context of human interpretation of medical images (which is relevant for AI) does not apply directly to this type of device testing. Ground truth here is established through objective measurements, chemical analyses, and direct observation of physical performance against established engineering and biological standards (e.g., ISO standards, GLP for animal studies).

    4. Adjudication Method for the Test Set:

    • Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where human readers provide subjective assessments that need to be aggregated or resolved.
    • For the non-clinical and animal studies described, no such adjudication method is mentioned or relevant. The results are quantitative measurements or observations against predefined pass/fail criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This type of study is specifically designed to assess the impact of an AI system on human reader performance, typically in diagnostic imaging.
    • The document describes testing for a physical medical device (catheter and tubing), not an AI system. Its "effectiveness" is demonstrated through its physical performance (e.g., ability to retrieve clots in a simulated environment, biocompatibility, durability) and successful outcomes in animal models.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • No, a standalone study in the context of an "algorithm only" was not done. As stated, this is a physical medical device. The "performance testing" sections describe the standalone performance of the physical device components (e.g., tensile strength, lumen patency).

    7. Type of Ground Truth Used:

    • Objective Test Results: For non-clinical (benchtop) tests, the "ground truth" is established by direct measurement and observation against pre-defined engineering specifications and international standards (e.g., ISO standards).
    • Biological Outcomes: For biocompatibility, ground truth is established through standardized biological assays (e.g., cytotoxicity, sensitization, irritation tests) with clear pass/fail criteria.
    • Animal Study Observations: For animal studies, ground truth involves in-life observations, and post-mortem assessment of relevant tissues, as well as the demonstration of safety and predetermined performance parameters (e.g., clot retrieval). The studies were conducted according to Good Laboratory Practices (GLP), which ensures scientific rigor and documented results as "ground truth."

    8. Sample Size for the Training Set:

    • Not applicable. This document describes the clearance of a medical device based on its physical and biological performance, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. There is no training set mentioned or implied in this document as it does not relate to an AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222743

    Validate with FDA (Live)

    Date Cleared
    2022-10-04

    (25 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Full Length 070 Access System is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical Full Length 070 Access System is comprised of a Support Catheter and a Delivery Catheter. The Support Catheter is a single-lumen, variable stiffness catheter. The Delivery Catheter is a hubbed, single-lumen variable stiffness catheter. Both catheters are hydrophilically coated. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    The Route 92 Medical Full Length 070 Access System is a medical device that facilitates the insertion and guidance of microcatheters in the neurovascular system. The provided document focuses on demonstrating the substantial equivalence of the device to a predicate device (Route 92 Medical Full Length 088 Access System, K210635) through non-clinical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes various performance tests conducted, with the general acceptance criterion being that "All samples met the pre-determined acceptance criteria." While specific numerical acceptance criteria are not detailed in this summary, the results consistently indicate "PASS" for all tests.

    TestAcceptance Criteria (Implied)Reported Device Performance
    Dimensional VerificationConformance to specificationsPASS
    Luer IntegrityPer ISO 80369-7:2016PASS
    Tensile StrengthSufficient strength of catheter sections/bondsPASS
    Kink ResistanceResistance to kinking at defined bend diameterPASS
    Torsion ResistanceIntegrity after rotationPASS
    Tip FlexibilityAcceptable tip flexibilityPASS
    Air LeakagePer ISO 10555-1:2013 Annex DPASS
    Liquid LeakagePer ISO 10555-1:2013 Annex CPASS
    Static BurstPer ISO 10555-1:2013 Annex FPASS
    Dynamic BurstMechanical integrity up to specified pressuresPASS
    Hydrophilic Coating IntegrityIntegrity after multiple insertion/withdrawal cyclesPASS
    Simulated Use TestingDeliverability and compatibility in neurovascular modelPASS

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the numerical sample size used for each test. Instead, it generally refers to "All samples." The provenance of the data is not specified regarding country of origin or whether it was retrospective or prospective. Given that it's non-clinical performance testing for a 510(k) submission, it's highly probable these were controlled laboratory tests conducted prospectively by the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This section is not applicable to this document. The study described is non-clinical performance testing of a physical medical device, not a study involving human interpretation of medical data where expert ground truth would be required. The "ground truth" for these tests are the established engineering specifications and international standards (e.g., ISO standards).

    4. Adjudication Method for the Test Set:

    This section is not applicable for the same reasons as point 3. Testing involves objective measurements against pre-defined engineering and performance criteria, not expert adjudication of subjective findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    No, an MRMC comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence through non-clinical performance testing of the device itself, not on evaluating the performance of human readers with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This section is not applicable. The device is a physical catheter system, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The "ground truth" for the non-clinical performance tests conducted for the Route 92 Medical Full Length 070 Access System is based on pre-determined acceptance criteria derived from established engineering specifications and relevant international standards (e.g., ISO standards). For example, Luer integrity was tested per ISO 80369-7:2016, and air/liquid leakage and static burst per ISO 10555-1:2013. Simulated use testing evaluated deliverability and compatibility against expected performance in a neurovascular model.

    8. The Sample Size for the Training Set:

    This section is not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established:

    This section is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203043

    Validate with FDA (Live)

    Date Cleared
    2020-11-11

    (36 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical 070 Access System is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical 070 Access System is comprised of a Support Catheter and a Delivery Catheter. The distal portion of the Support Catheter is a single-lumen, variable stiffness catheter. Like the predicate device, the proximal portion is a stainless-steel control wire. The Delivery Catheter is a hubbed, singlelumen variable stiffness catheter. Both catheters are hydrophilically coated. The Route 92 Medical 070 Access System is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    This document describes the Route 92 Medical 070 Access System, a percutaneous catheter intended to facilitate the insertion and guidance of microcatheters in the neurovascular system. The provided text, K203043, is a 510(k) summary from the FDA, detailing the device's substantial equivalence to a predicate device based on non-clinical testing.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally "All samples met the pre-determined acceptance criteria" for each test. The actual numerical thresholds for these criteria are not explicitly stated in this document but are implied to be established standards for such medical devices.

    TestTest MethodReported Device Performance
    Biocompatibility
    CytotoxicityISO MEM ElutionThe test article is non-cytotoxic.
    SensitizationISO Guinea Pig Maximization Sensitization Test (Normal Saline and Sesame Oil)The test article did not elicit a sensitization response.
    IrritationISO Intracutaneous Reactivity (Normal Saline and Sesame Oil)Requirements of the ISO intracutaneous reactivity test have been met for the test article.
    Acute Systemic ToxicityISO Acute Systemic Injection (Normal Saline and Sesame Oil)Requirements of the ISO acute systemic injection test have been met for the test article.
    PyrogenMaterial Mediated Pyrogen (Normal Saline)The test article is non-pyrogenic.
    Hemocompatibility – Complement ActivationComplement Activation (SC5b-9)The test article would not be expected to result in adverse effects in vivo.
    Hemocompatibility – Partial Thromboplastin TimePartial Thromboplastin TimeThe test article is considered to be a non-activator of the intrinsic coagulation pathway.
    Hemocompatibility – ASTM HemolysisASTM HemolysisThe test article is considered non-hemolytic.
    Hemocompatibility – ThromboresistanceThromboresistanceThe test articles have similar thromboresistance characteristics as the control devices.
    Performance Testing
    Dimensional VerificationDevice dimensions were measured to confirm conformance to the specifications.PASS (All samples met the pre-determined acceptance criteria)
    Tensile StrengthThe tensile strength of the catheter sections and bonds was tested.PASS (All samples met the pre-determined acceptance criteria)
    Kink ResistanceTest specimen segments were formed into a defined bend diameter to evaluate kink resistance.PASS (All samples met the pre-determined acceptance criteria)
    Torsion ResistanceThe test specimens were rotated to evaluate integrity after rotation.PASS (All samples met the pre-determined acceptance criteria)
    Tip FlexibilityTest specimens were tested for tip flexibility.PASS (All samples met the pre-determined acceptance criteria)
    Air LeakageTested per ISO 10555-1:2013 Annex D.PASS (All samples met the pre-determined acceptance criteria)
    Liquid Leakage / Static BurstTested per ISO 10555-1:2013 Annex C.PASS (All samples met the pre-determined acceptance criteria)
    Dynamic BurstMechanical integrity was maintained up to the specified pressures.PASS (All samples met the pre-determined acceptance criteria)
    Hydrophilic Coating IntegrityThe integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles.PASS (All samples met the pre-determined acceptance criteria)
    Particulate RecoveryAfter multiple insertion and withdrawal cycles, the effluent water rinsed and flushed from the devices and model was tested per USP <788>. Particulates were characterized for size ranges ≥10 um, ≥25 um, ≥50 µm, ≥100 µm, ≥200 µm, ≥500 µm and ≥1000 µm.PASS (All samples met the pre-determined acceptance criteria)
    Simulated Use TestingDeliverability and compatibility with accessory devices was evaluated in a neurovascular model.PASS (All samples met the pre-determined acceptance criteria)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document repeatedly states "All samples met the pre-determined acceptance criteria" but does not specify the exact number of samples used for each test set.
    • Data Provenance: The tests are non-clinical (laboratory/bench testing) and were conducted by the manufacturer (Route 92 Medical, Inc.) to demonstrate substantial equivalence. The country of origin of the data is not explicitly stated but is implicitly the United States, given the FDA submission. The data is prospective for the purpose of this 510(k) submission, as it was generated specifically to support the clearance of this device.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Ground Truth Establishment: For non-clinical performance and biocompatibility testing of a medical device like this catheter, "ground truth" is typically established by recognized international standards (e.g., ISO, ASTM, USP) and engineering specifications. It's not based on expert consensus in the same way as, for example, image interpretation.
    • Experts: The document does not mention the use of human experts to establish ground truth for these non-clinical tests. The tests themselves provide the "ground truth" against the established quantitative criteria from the standards.

    4. Adjudication Method

    • Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies or multi-reader reviews involving subjective assessments, especially for diagnostic devices. This is a non-clinical, bench-top testing summary. Therefore, an adjudication method as described is not applicable and was not used. The results are compared directly to predefined quantitative pass/fail criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This device is a physical medical device (catheter system), not an AI-powered diagnostic or assistive tool for human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

    6. Standalone Performance

    • Yes, a standalone performance study was done. All the listed biocompatibility and performance tests are standalone evaluations of the device's physical and biological characteristics without human intervention in the test setup (beyond conducting the test itself). The device itself is not an AI algorithm, so the term "algorithm only" is not directly applicable, but the tests were "device only" performance assessments.

    7. Type of Ground Truth Used

    • The ground truth used for the non-clinical tests is based on established industry standards and documented engineering specifications/acceptance criteria. These standards include:
      • ISO 10993-1: Biological Evaluation of Medical Devices Part 1: Evaluation and Testing for biocompatibility.
      • ISO 10555-1:2013: Intravascular Catheters - Sterile Single-Use Catheters - Part 1: General Requirements for Air and Liquid Leakage / Static Burst.
      • USP <788>: Particulate Matter in Injections for particulate recovery.
      • Generic acceptance criteria for mechanical properties (Tensile Strength, Kink Resistance, Torsion Resistance, Tip Flexibility, Dynamic Burst, Hydrophilic Coating Integrity, Dimensional Verification) which would be defined by the manufacturer based on expected performance and comparison to predicate devices, and internal design requirements.

    8. Sample Size for the Training Set

    • Not applicable. This device is a physical medical instrument, not an AI model requiring a training set.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable. As stated above, this is a physical medical device, not an AI model.

    In summary, the K203043 submission demonstrates the substantial equivalence of the Route 92 Medical 070 Access System through comprehensive non-clinical testing. The device met all pre-determined acceptance criteria based on established industry standards and internal specifications, confirming its safety and performance for its intended use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180201

    Validate with FDA (Live)

    Date Cleared
    2018-05-23

    (119 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical Access System is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical Access System is comprised of an Access Catheter and a Delivery Catheter. The distal portion of the Access Catheter is a single-lumen, variable stiffness catheter. Like the predicate device, the proximal portion is a stainless-steel control wire. The Delivery Catheter is a hubbed, singlelumen variable stiffness catheters are hydrophilically coated. The Route 92 Medical Access System is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    The provided text describes the acceptance criteria and the studies conducted for the Route 92 Medical Access System.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for all tests was "PASS". The reported device performance for all tests was also "PASS". This is explicitly stated for each test category.

    Test CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityDemonstrates non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic, minimal complement activation, non-activator of intrinsic coagulation, non-hemolytic, similar thromboresistance, non-mutagenic, and non-clastogenic.All tests passed, concluding the device meets biocompatibility requirements.
    Dimensional VerificationConformance to specificationsPASS
    Luer IntegrityPer ISO 80369-7:2016PASS
    RHV SealingSealing around catheter shaftsPASS
    Tensile StrengthStrength of catheter sections and bondsPASS
    Kink ResistanceResistance to kinking in a defined bend diameterPASS
    Torsion ResistanceIntegrity after rotationPASS
    Tip FlexibilityFlexibility of the tipPASS
    Air LeakagePer ISO 10555-1:2013 Annex DPASS
    Liquid Leakage / Static BurstPer ISO 10555-1:2013 Annex CPASS
    Dynamic BurstMechanical integrity up to specified pressuresPASS
    FlowContrast media flow rates equivalent to predicate devicesPASS
    Hydrophilic Coating IntegrityIntegrity after multiple insertion/withdrawal cyclesPASS
    Particulate RecoveryEffluent water tested per USP <788> after multiple cyclesPASS
    Corrosion ResistancePer ISO 10555-1:2013 Annex APASS
    Simulated Use TestingDeliverability and compatibility in a neurovascular modelPASS
    Packaging IntegrityPer ISO 11607-1 Part 1 & ISO 11607-2 Part 2PASS
    RadiopacityRadiopacity in an animal model under fluoroscopyPASS

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the specific "sample size" for each individual test within the performance and biocompatibility sections. Instead, it generally refers to "All samples" or "test animals" for biocompatibility.

    • For Biocompatibility: The tests used "animals" (Guinea Pigs, Rabbits, Mice, potentially others for in-vitro tests). The number of animals per test is mentioned for some (e.g., "None of the animals", "rabbits administered").
    • For Performance Testing: The results consistently state "All samples met the pre-determined acceptance criteria." without specifying the number of samples.
    • Data Provenance: The studies are non-clinical, meaning they are laboratory or animal-based tests. The document does not specify the country of origin of the data, nor whether it is retrospective or prospective, though non-clinical tests are generally prospective laboratory studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The studies described are non-clinical in nature, focusing on the physical and biological characteristics of the device. There is no mention of "experts" establishing a "ground truth" for a test set in the context of human evaluation or diagnostic accuracy, as this device is a medical instrument, not a diagnostic algorithm.

    4. Adjudication Method for the Test Set

    This information is not provided and is not applicable given the nature of the non-clinical testing performed. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies evaluating human interpretation (e.g., image reading) where disagreement among experts needs resolution.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic AI systems or those involving human interpretation of data, which is not the function of the Route 92 Medical Access System.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied to diagnostic AI is not directly applicable here. For the non-clinical tests:

    • Biocompatibility: Ground truth is established by standardized biological evaluation methods and safety endpoints (e.g., no signs of toxicity, irritation, sensitization).
    • Performance Testing: Ground truth is established by engineering specifications, recognized industry standards (e.g., ISO standards), and functional requirements (e.g., flow rates, mechanical integrity).

    8. The Sample Size for the Training Set

    This information is not applicable. This document describes the testing of a physical medical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1