Search Filters

Search Results

Found 38 results

510(k) Data Aggregation

    K Number
    K253693

    Validate with FDA (Live)

    Date Cleared
    2026-03-11

    (110 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Strut Suture: The Strut Suture is intended to be used for fixation of bone to bone, and is intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL Repair.

    No-Tie Button The No-Tie Button is intended to be used for fixation of bone to bone, intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL Repair when used with Strut Suture (or equivalent) and soft tissue fixation to bone in the knee for meniscal root repair when used with RootMend MRR (or equivalent).

    Device Description

    Strut Suture devices are a suture and button construct composed of uncoated UHMWPE (ultra-high molecular weight polyethylene) with UHMWPE tracer material and an attachable fixation anchor composed of a titanium (Ti-6Al-4V) button. Strut Suture devices will be provided sterile using ethylene oxide, for single use only, and are not to be re-sterilized.

    No-Tie Button devices are an attachable fixation anchor composed of a titanium (Ti-6Al-4V) button. No-Tie Button devices will be provided sterile using ethylene oxide, for single use only, and are not to be re-sterilized.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253695

    Validate with FDA (Live)

    Device Name
    LigaMend
    Date Cleared
    2026-02-27

    (98 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LigaMend devices are intended to be used for fixation of soft tissue to bone and are intended as fixation over areas of ligament or tendon repair, in the following procedures:
    • Shoulder: Biceps Tenodesis, Acromio-Clavicular Separation Repair
    • Foot/Ankle: Achilles Tendon Repair.
    • Knee: Meniscal Root Repair, MPFL Repair, ACL/PCL repair

    Device Description

    LigaMend is a fixation loop composed of UHMWPE (ultra-high molecular weight polyethylene) with UHMWPE tracer material. Fixation loops are available with an attachable fixation anchor composed of a titanium (Ti-6Al-4V) button, or with an attached UHMWPE fixation button. LigaMend devices will be provided sterile using ethylene oxide, for single use only, and are not to be re-sterilized.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253024

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-16

    (27 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ProZip Knotless Implant is intended for use in soft-tissue approximation.

    Device Description

    The ProZip Knotless Implant is comprised of braided ultra-high molecular weight polyethylene (UHMWPE) and Polyester sutures. It contains one working suture, also referred to as the repair strand, and shuttle strand that is used to shuttle the repair strand around the tissue. Additionally, the repair strand has a suture attached at either end to aid in passing through tissue, known as adaptor tail.

    Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252201

    Validate with FDA (Live)

    Device Name
    HS Fiber
    Manufacturer
    Date Cleared
    2025-08-13

    (30 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riverpoint Medical HS Fiber® sutures are indicated for injury and/or reconstruction of soft tissue, excluding the ones listed in contraindications, where soft tissue ligation or approximation is required. The Riverpoint Medical HS Fiber® sutures are also indicated for conditions where allograft tissues are required to be fixated, such as ACL tears.

    Device Description

    The Riverpoint Medical HS Fiber® sutures are non-absorbable, sterile, surgical sutures composed of multiple single strands of ultra-high molecular weight polyethylene (UHMWPE) braided together to form the implant. HS Fiber sutures are available in common sizes and lengths with or without pre-attached needles.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a surgical suture, the HS Fiber® Suture, and not an AI/ML powered device. Therefore, the request to describe acceptance criteria and a study proving device performance in the context of AI/ML (e.g., test set, ground truth, expert consensus, MRMC study, training set) cannot be fulfilled based on the provided document.

    The document states: "The updates being proposed are limited to the labeling, no additional performance testing or biological evaluation was required." This indicates that no new performance studies (of any kind, let alone AI/ML specific ones) were conducted for this particular 510(k) submission (K252201). The device is deemed substantially equivalent to a previously cleared predicate device (K231163) based on the technological identity and minor labeling changes.

    Therefore, I cannot provide the requested information for an AI/ML device from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K252144

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-30

    (22 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Iconix Speed Anchors are intended to be used for soft-tissue to bone fixation in the shoulder. They are indicated for use in rotator cuff repair.

    Device Description

    The Iconix Speed Anchors are soft-tissue to bone fixation devices, provided preloaded on a disposable self-punching inserter. The device is composed of a braided polyester anchor body that contains two or three working sutures. The anchor is inserted into the bone using a self-punching mechanism, and the polyester sheath bunches as the anchor is deployed to allow for fixation in bone.

    Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserters are comprised of metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. Iconix Speed Anchors are available in common sizes and lengths and will be sold sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    The classification for the Iconix Anchor is FDA Class II device with product classification 21 CFR §888.3040: Smooth or threaded metallic bone fixation fastener, Product Code MBI.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Iconix Speed Anchor does not describe a study that involves software performance, AI algorithms, or human-in-the-loop assessments. Instead, it describes mechanical and biological performance testing for a physical medical device (bone anchor).

    Therefore, I cannot provide details for most of your requested points, as they are specific to AI/software device studies.

    However, I can extract the information related to the device's performance testing and acceptance criteria as described in the document.


    1. Table of Acceptance Criteria and the Reported Device Performance

    The 510(k) summary states that "Results of performance testing for the Iconix Speed Anchor device concluded that the device performed comparably to the predicate device in insertion, cyclic and pullout testing and the validations performed demonstrated that the Iconix Speed Anchor met all requirements for its intended use."

    However, the document does not explicitly state specific quantitative acceptance criteria (e.g., "pullout strength > X N") or the exact reported performance values for the Iconix Speed Anchor in these tests. It only states that the device performed "comparably" to the predicate.

    Acceptance Criteria (Implied)Reported Device Performance
    Comparable insertion strength to predicate device.Performed comparably to the predicate device.
    Comparable cyclic strength to predicate device.Performed comparably to the predicate device.
    Comparable pullout strength to predicate device.Performed comparably to the predicate device.
    Met all requirements for its intended use.Met all requirements for its intended use.
    Sutures meet USP performance requirements for tensile strength (except diameter).UHMWPE sutures tested per USP performance requirements for tensile strength.
    Sterilization adoption valid.Sterilization adoption validated.
    Biocompatible per ISO10993-1:2018.Biocompatibility testing per ISO10993-1:2018 completed.
    Packaging stable per ISO 11607-1:2006.Stability testing on product packaging per ISO 11607-1:2006 completed.
    Usability validated in simulated use per EN62366:2015.Usability engineering validation with simulated use in cadaveric models performed per EN62366:2015.

    The following information cannot be provided as the provided document describes a physical medical device (bone anchor), not an AI/software device study.

    1. Sample size used for the test set and the data provenance: Not applicable. The "test set" in this context refers to physical units of the device subjected to mechanical and biological testing, not a dataset for an AI model. The document does not specify the number of samples for each mechanical test.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's mechanical performance is typically established through direct physical measurement under controlled conditions, not expert consensus.
    3. Adjudication method: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: For mechanical tests (insertion, cyclic, pullout), the "ground truth" is the measured physical performance of the device. For biological tests, it's compliance with ISO standards. For usability, it's observation in cadaveric models.
    7. The sample size for the training set: Not applicable. The device is a physical product, not an AI algorithm requiring a training set.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243988

    Validate with FDA (Live)

    Device Name
    RootMend MRR
    Date Cleared
    2025-03-20

    (84 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RootMend MRR devices are intended to be used for suture (soft tissue) fixation to bone in the knee for meniscal root repair.

    Device Description

    The RootMend Meniscal Root Repair (MRR) devices are composed of uncoated UHMWPE (ultra-high molecular weight polyethylene) with UHMWPE tracer material. RootMend MRR devices are available with a fixation button composed of a titanium (Ti-6Al-4V) button, or with an attached UHMWPE fixation button. RootMend MRR devices will be provided sterile using ethylene oxide, for single use only, and are not to be resterilized. The UHMWPE is available undyed (white), black, or blue, with trace filaments of black or blue. RootMend MRR devices are intended to be used for suture (soft tissue) fixation to bone in the knee for meniscal root repair.

    AI/ML Overview

    The provided text from the FDA 510(k) summary for the RootMend MRR device states that "Non-clinical mechanical testing was performed to verify the fixation strength of the RootMend MRR devices and is compared to the predicate device. Testing conducted includes cyclic and ultimate tensile strength (UTS) testing, and assessment of displacement for the subject button and predicate anchor constructs. Usability engineering testing with simulated use in cadaver lab was performed on the subject device per EN62366-1. In all instances of the testing referenced above, the acceptance criteria were met, and the proposed RootMend MRR devices performed as intended."

    However, the provided document does not contain the specific acceptance criteria or the detailed results (e.g., reported device performance values) from the mechanical testing and usability testing. It also lacks information regarding sample size, ground truth establishment, expert qualifications, or MRMC studies that would be relevant for an AI/software device evaluation.

    Therefore, I cannot fulfill the request the following information based on the provided text:

    • A table of acceptance criteria and the reported device performance: Not provided.
    • Sample sized used for the test set and the data provenance: Not provided.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable to this type of device and testing described.
    • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a hardware device, not an AI/software device assisting human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For mechanical testing, the "ground truth" is the specified performance thresholds (acceptance criteria) and the measured physical properties. For usability testing, the 'ground truth' is successful completion of tasks by users in a cadaver lab. Specifics are not detailed beyond "performed as intended."
    • The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
    • How the ground truth for the training set was established: Not applicable.

    Based on the available information, here is what can be inferred or directly stated:

    1. Acceptance Criteria and Device Performance:

    The document states: "In all instances of the testing referenced above, the acceptance criteria were met, and the proposed RootMend MRR devices performed as intended."

    • Acceptance Criteria: Not explicitly stated in quantitative terms. They likely relate to minimum cyclic and ultimate tensile strength, and maximum displacement relative to the predicate device.
    • Reported Device Performance: Not explicitly stated in quantitative terms. The performance met the unstated acceptance criteria.

    2. Sample Size and Data Provenance:

    • Sample Size: Not specified for any of the tests (cyclic, UTS, displacement, usability).
    • Data Provenance: The tests were "non-clinical mechanical testing" and "usability engineering testing with simulated use in cadaver lab." This implies laboratory/bench testing rather than human clinical data. The location of the testing or data origin (country) is not specified.

    3. Experts for Ground Truth / Adjudication / MRMC studies / Standalone Performance:

    • These concepts (experts, adjudication, MRMC, standalone performance) are typically relevant for AI/software-based medical devices that interpret medical images or data. The RootMend MRR is a physical orthopedic implant (metallic bone fixation fastener). Therefore, these criteria are not applicable to the device and studies described.

    4. Type of Ground Truth:

    • For mechanical testing: The ground truth is established by engineering specifications and comparative performance to a predicate device. This would involve objective measurements of strength, fatigue, and displacement against predefined thresholds.
    • For usability testing: The ground truth is successful completion of surgical tasks and safe operation in a simulated environment (cadaver lab) as assessed by the test protocol and observations.

    5. Training Set Information:

    • Sample Size for Training Set: Not applicable. This is not an AI/machine learning device that uses a "training set."
    • How Ground Truth for Training Set was Established: Not applicable.

    In summary, the provided FDA 510(k) summary provides a high-level statement that the device met acceptance criteria through specified non-clinical mechanical and usability testing, but it does not disclose the quantitative details of those criteria, the measured performance, or the sample sizes used. The remaining questions are not relevant given the nature of the device as a physical implant rather than an AI/software product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242494

    Validate with FDA (Live)

    Date Cleared
    2024-11-26

    (96 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct are intended for use in soft tissue to bone fixation for the following indications: Shoulder Rotator Cuff Repair

    Device Description

    The Riverpoint Medical Interfacing Anchor is comprised of a pre-deployed all-suture anchor with attached repair sutures and 302 stainless steel needles. The Interfacing Anchor is inserted into either the humerus or into passing slots available in devices intended for total shoulder arthroplasty. The Riverpoint Medical Bone Tunnel Construct is comprised of a double armed Ultra-High Molecular Weight Polyethylene (UHMWPE) suture with a preassembled polypropylene passing loop and 302 stainless steel needles. Suture supplied meet United States Pharmacopeia (USP) requirements for nonabsorbable suture except for diameter. Suture dyes are FDA authorized. The device is sterilized by ethylene oxide gas and is provided sterile for single use. Interfacing Anchor and Bone Tunnel Construct are available in common sizes and lengths with preattached 302 stainless steel needles and will be sold sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    The provided text is a 510(k) summary for the OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct. This document details the regulatory review process for a medical device and its substantial equivalence to a predicate device.

    It does NOT describe an AI/ML-driven medical device, nor does it contain information about diagnostic performance metrics (e.g., sensitivity, specificity, AUC), test set characteristics, expert adjudication processes, or MRMC studies.

    The "Performance Data" section of the document refers to non-clinical performance testing for a surgical device (an anchor and tunnel construct), not a diagnostic AI. The tests mentioned are:

    • Usability engineering validation with simulated use in cadaveric models (EN62366:2015)
    • Endotoxin/pyrogenicity testing (ANSI/AAMI ST72:2019, USP <161>, USP <151>, USP <85>)
    • Mechanical testing to verify fixation strength (cyclic and pullout testing)

    Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI/ML device's performance, as the provided text pertains to a physical surgical device.

    To directly answer your prompt based on the provided text, the following cannot be extracted:

    • A table of acceptance criteria and reported device performance related to diagnostic accuracy (as it's not a diagnostic device). The acceptance criteria mentioned are for mechanical and usability performance of a surgical implant.
    • Sample sizes used for a "test set" in the context of image analysis or diagnostic performance. The "test set" here refers to cadaveric models and mechanical testing samples.
    • Data provenance (country of origin, retrospective/prospective) for a diagnostic dataset.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for a test set.
    • MRMC comparative effectiveness study, including effect size.
    • Standalone (algorithm only) performance.
    • Type of ground truth (expert consensus, pathology, outcomes data).
    • Sample size for a training set.
    • How ground truth for the training set was established.

    However, based on the provided text, here's what can be inferred or directly stated about the "acceptance criteria" and "study" for this specific device (a surgical implant):

    Acceptance Criteria and Device Performance (for a surgical implant):

    Acceptance Criterion (Type)Reported Device Performance
    Material/Suture PropertiesSuture supplied meets USP requirements (except for diameter).
    UHMWPE sutures tested per USP for needle attachment & tensile strength.
    Biocompatibility/SterilityEndotoxin/pyrogenicity testing per various standards; results below recommended limits.
    Mechanical Performance (Fixation Strength)Performed comparably to the predicate device in cyclic and pullout testing.
    Usability/Clinical PerformanceUsability engineering validation with simulated use in cadaveric models performed as per EN62366:2015. Results support use and IFU.
    General Performance (Overall Summary)"In all instances of the testing referenced above, the acceptance criteria were met, and the proposed OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct performed as intended."
    "Validations performed demonstrated that the OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct met all requirements for its intended use."

    Study Details (for this surgical implant):

    1. Sample sizes used for the test set and data provenance:

      • The "test set" refers to samples for mechanical testing and cadaveric models for usability. Specific numerical sample sizes are not provided in this summary.
      • Data provenance is not described (e.g., where the cadavers came from, or if the mechanical tests were done by a specific lab).
      • The studies were non-clinical ("bench testing" and "simulated use").
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable in the context of diagnostic "ground truth." The "ground truth" for a surgical implant's mechanical performance is defined by engineering standards and comparative testing to predicates. Usability testing would involve typical users (surgeons, OR staff), but the number or qualifications are not specified.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as this is not a diagnostic study requiring human expert consensus/adjudication of results. Performance is measured against physical test standards and engineering specifications.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a MRMC comparative effectiveness study was not done. This is a physical surgical device, not an AI/ML diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable as this is a physical surgical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For mechanical testing: Engineering specifications and comparative performance to a legally marketed predicate device.
      • For usability testing: Performance in simulated use conditions in cadaveric models against predefined usability objectives (derived from EN62366:2015).
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device with a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243323

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-19

    (27 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Iconix Knotless Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

    Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/ Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repair

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

    Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair

    Device Description

    The Iconix Knotless Anchors are soft-tissue to bone fixation devices, provided preloaded on a disposable inserter. The device is composed of a braided polyester anchor body that contains one working suture, also referred to as the repair strand, and a shuttle strand that is used to shuttle the repair strand around tissue and through the braided anchor body.

    Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserters are comprised of metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The size of the anchor of the subject device is 1.4mm. Iconix Knotless Anchor will be sold sterile for single use with no components or accessories. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    The provided text pertains to an FDA 510(k) premarket notification for a medical device called the "Iconix Knotless Anchor." It outlines the device's description, intended use, and the basis for its substantial equivalence to a previously cleared predicate device.

    However, the document does not describe acceptance criteria or a study that proves the device meets those criteria in the context of an AI/ML medical device. The product is a physical medical device (knotless anchor for soft-tissue to bone fixation), not an AI/ML algorithm.

    Therefore, I cannot provide the requested information related to acceptance criteria, device performance, sample sizes, ground truth establishment, expert adjudication, or MRMC studies for an AI/ML device, as these concepts are not applicable to the physical medical device described in the document.

    The document discusses non-clinical mechanical testing (insertion, cyclic, and pullout testing) and other validations (shelf-life, sterilization, biocompatibility, packaging stability, usability engineering) to demonstrate that the device performs comparably to its predicate and meets requirements for its intended use. This is typical for physical medical devices but does not align with the AI/ML-focused questions in your prompt.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243203

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-29

    (28 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OC JuggerKnotless Soft Anchor and OC JuggerLoop Soft Anchor are intended for use in soft tissue to bone fixation in the following indications:
    Knee MPFL
    Knee Patellar tendon repair
    Knee MCL
    Knee Quadriceps tendon repair
    Foot and Ankle Achilles tendon repair
    Foot and Ankle Medial/lateral repair and reconstruction
    Foot and Ankle Plantar plate repair
    Foot and Ankle Mid- and forefoot repair
    Foot and Ankle Metatarsal ligament/tendon repair or reconstruction
    Shoulder Rotator Cuff
    Shoulder Shoulder Instability
    Shoulder Biceps Tenodesis
    Elbow Lateral epicondylitis repair
    Elbow Biceps tendon reattachment

    Device Description

    The OC JuggerKnotless® Soft Anchor is comprised of a suture sleeve structure and working suture. Non-absorbable braided ultra-high molecular weight polyethylene (UHMWPE) sutures are spliced through a non-absorbable braided polyester suture anchor sleeve. Up to two non- absorbable round and flat braided UHMWPE working sutures can be added inside the suture anchor sleeve. The UHMWPE sutures are available undyed (white), blue, or black Available Suture sizes are standard according to USP requirements (dependent on suture type). The hydroxyapatite and bioceramic particles are embedded in the surface of the suture anchor material.

    The OC JuggerLoop® Soft Anchor is comprised of a suture sleeve structure and shuttling and cinching suture. Non-absorbable braided ultra-high molecular weight polyethylene (UHMWPE) sutures are spliced through a non-absorbable braided polyester suture anchor sleeve. Up to two non-absorbable round and flat braided UHMWPE working sutures can be added inside the suture anchor sleeve. The UHMWPE sutures are available undyed (white), blue, or black. Available Suture sizes are standard according to USP requirements (dependent on suture type). The hydroxyapatite and bioceramic particles are embedded in the surface of the suture anchor material.

    Suture supplied meet United States Pharmacopeia (USP) requirements for nonabsorbable suture except for diameter. Suture dyes are FDA approved. The inserter is comprised of a metallic shaft with a standard or self-punching tip and an overmolded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use, OC JuggerKnotless Soft Anchor and OC JuggerLoop Soft Anchor are available in common sizes and lengths and will be sold sterile for single use with no components or accessories. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    The provided text describes a medical device, the OC JuggerKnotless Soft Anchor and OC JuggerLoop Soft Anchor, and its substantial equivalence to a predicate device. However, it does not contain information typically associated with the acceptance criteria and study design for AI/ML-based medical devices. The document is an FDA 510(k) clearance letter and summary for a physical medical device (bone fixation fastener), not a software or AI/ML device.

    Therefore, the requested information regarding acceptance criteria, reported device performance in a table, sample sizes for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment cannot be extracted from this document because these concepts are not applicable to the type of device described.

    The document discusses the following:

    • Device Type: Soft Anchors for soft tissue to bone fixation.
    • Regulatory Classification: Class II medical device (21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener).
    • Performance Data: Non-clinical mechanical testing (insertion, cyclic, and pullout testing) and usability engineering validation in cadaveric models.
    • Conclusion: The device is substantially equivalent to a predicate device based on intended use, principle of operation, materials, packaging, sterilization, and comparable performance in non-clinical testing.

    If you have a document pertaining to an AI/ML medical device, please provide that for analysis.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241282

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-19

    (104 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RibFix Titan Fixation System is indicated for the fixation, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.

    Device Description

    The Riverpoint Medical RibFix Titan Fixation System is comprised of titanium bone plates that are secured to bone by a deplovable ultra-high molecular weight polyethylene (UHMWPE) suture anchor with a titanium washer and button. Bone plates are provided in configurations for intrathoracic and extrathoracic use. Intrathoracic bone plates are provided in 60mm and 94mm lengths. Extrathoracic bone plates are provided in 60mm, 94mm, and 120mm lengths.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or results of a study proving the device meets those criteria in a tabular format as requested for all points. The document is a 510(k) summary for the RibFix Titan™ Fixation System, which describes performance testing generally but does not provide specific metrics, acceptance criteria, or detailed study results.

    However, based on the limited information available, here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "In all instances of the testing referenced above, the acceptance criteria were met, and the proposed RibFix Titan Fixation System performed as intended." However, no specific numerical acceptance criteria or performance metrics are provided in the document.

    Acceptance CriteriaReported Device Performance
    Not specifiedMet all acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    The document does not specify sample sizes for mechanical testing or usability validation, nor does it provide information on the data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable. The reported studies are mechanical and usability engineering, not clinical studies requiring expert ground truth for diagnosis/assessment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. This is not a clinical study involving human judgment on cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a medical device (fixation system), not an AI/software device involving human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Not applicable. For the mechanical and usability testing mentioned:

    • Mechanical testing: Ground truth would be defined by engineering specifications and physical measurements according to standards like ASTM F382 (for plate components) and ASTM F1717 (for assembled constructs).
    • Usability engineering validation: Ground truth would be the defined usability goals and safety requirements, assessed through expert evaluation and simulated use.

    8. The sample size for the training set:

    Not applicable. This is not a machine learning/AI study involving a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is not a machine learning/AI study.

    Summary of available information regarding studies:

    The document mentions the following non-clinical performance data studies were conducted:

    • Four-point bend testing (static and fatigue) of bone plate components per ASTM F382.
    • Cyclic and Ultimate Tensile Strength (UTS) testing of anchor constructs.
    • Cyclic fatigue testing of assembled constructs using ASTM F1717 as a guide.
    • Usability engineering validation with simulated use in cadaver lab performed per EN62366-1.
    • MR Compatibility testing per ASTM F2052, ASTM F2213, ASTM F2119, ASTM F2182.

    The document states that "In all instances of the testing referenced above, the acceptance criteria were met, and the proposed RibFix Titan Fixation System performed as intended." It also mentions that the testing compared the RibFix Titan Fixation System to the predicate device, especially regarding fixation strength. For usability, it notes that the "usability study and performance testing of the RibFix Titan support the use of the RibFix Titan Fixation System and eIFU."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4