Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K223859

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-02-13

    (52 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rhino Non-Sterile Powder-Free Nitrile Exam Gloves – Periwinkle Color Tested for Use with Chemotherapy Drugs and Fentanyl is non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The tested chemotherapy drugs are as follows:

    Carmustine (BCNU) (3.3mg/mL 3,300 ppm) 14.2
    Cisplatin (1.0 mg/mL 1,000 ppm) > 240
    Cyclophosphamide (Cytoxan) (20 mg/mL 20,000 ppm) > 240
    Dacarbazine (10 mg/mL 10,000 ppm) > 240
    Doxorubicin Hydrochloride (2.0 mg/mL 2,000 ppm) > 240
    Etoposide (Toposar) (20.0 mg/mL 20,000 ppm) > 240
    Flurouracil (50.0 mg/mL 50,000 ppm) > 240
    Paclitaxel (Taxol) (6.0 mg/mL 6,000 ppm) > 240
    Thiotepa (10.0 mg/mL 10,000 ppm) 43.2
    Fentanyl Tested as Follows:
    Fentanyl Citrate Injection (100 mcg/2mL) > 240

    Note: Carmustine and Thiotepa have extremely low permeation times of 14.2 and 43.2 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa

    Device Description

    Rhino Non-sterile Powder-Free Nitrile Exam Glove - Periwinkle Color Tested for Use with Chemotherapy Drugs and Fentanyl are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are offered in three sizes, extra small, small, and extra large. The color for all sizes is periwinkle. The model number is RHI-UNB5 with specific part sizes based on size as follows:

    Glove SizePart Numbers
    XSRHI-UNB5XS100BX1
    SRHI-UNB5S100BX1
    XLRHI-UNB5XL100BX1
    AI/ML Overview

    The provided document describes the acceptance criteria and study results for the "Rhino Non-sterile Powder-Free Nitrile Exam Glove - Periwinkle Color Tested for Use with Chemotherapy Drugs and Fentanyl." This is a medical device, and the criteria relate to its physical properties, biocompatibility, and resistance to chemotherapy drugs and Fentanyl.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance (Results)
    ASTM D6319
    Physical Dimensions TestEnsure gloves meet size specificationsExtra Small: Length: $\geq 220$ mm, Width: $70 \pm 10$ mm Small: Length: $\geq 220$ mm, Width: $80 \pm 10$ mm Extra Large: Length: $\geq 230$ mm, Width: $120 \pm 10$ mmPass
    ThicknessEnsure adequate material thicknessFinger: $\geq 0.05$ mm Palm: $\geq 0.05$ mmPass
    Physical Properties (Tensile Strength)Measure strength before and after agingBefore Aging: $\geq 14$ MPa After Aging: $\geq 14$ MPaPass
    Physical Properties (Ultimate Elongation)Measure elasticity before and after agingBefore Aging: $\geq 500$ % After Aging: $\geq 400$ %Pass
    ASTM D5151Freedom from holesMeet the requirements for AQL 2.5Pass
    ASTM D6124Powder Residue< 2.0 mgPass
    ISO 10993-10IrritationNon-irritating; Primary irritation Index $\leq 1.0$Under the conditions of the study, not an irritant/Pass
    ISO 10993-10SensitizationNon-sensitizing; Grade < 1Under the conditions of the study, not a sensitizer/Pass
    ISO 10993-11Systemic ToxicityNon-acute systemic toxicity, no animals treated with test extracts exhibit greater reaction than control animals.Under the conditions of the study, did not show acute systemic toxicity in vivo/Pass
    ASTM D6978-05Permeation rate of specific chemotherapy drugs and FentanylMinimum breakthrough detection time of >240 min on all three (3) specimens for tested drugs.All drugs, except Carmustine and Thiotepa, had permeation time of > 20 minutes. Carmustine permeation time was 14.2 minutes and Thiotepa permeation time was 43.2 minutes. Note: Manufacturer issues a warning: Do Not Use with Carmustine, Thiotepa.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document outlines that "Non-clinical tests were conducted to verify that the proposed device met all design specifications." However, specific sample sizes for each test are generally not explicitly stated with numerical values in the provided summary beyond "3 specimens" for the permeation test and "50," "100," or "150" gloves for "Freedom from holes" for the predicate device. For example, for "Freedom from holes" for the subject device: "XS: No leakers in 50, S: 3 leakers in 100, XL: No leakers in 50."

    The data provenance is from non-clinical laboratory studies conducted to evaluate the device's adherence to various ASTM and ISO standards for medical gloves. The studies are prospective in nature, as they involve testing the manufactured device against established standards. The country of origin of the data is not specified in this document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not applicable to this type of device and study. The "ground truth" for the performance of these examination gloves is established by adherence to recognized industry standards (ASTM, ISO). There are no human "experts" establishing a subjective ground truth in the way there would be for a diagnostic AI device. The tests are objective measurements and assessments against predefined physical, chemical, and biological criteria.

    4. Adjudication Method for the Test Set:

    This is not applicable. As mentioned above, the "ground truth" is determined by objective measurements compared to established standard criteria, not by human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This document describes the testing of a physical medical device (nitrile exam gloves), not an artificial intelligence (AI) or software-as-a-medical-device (SaMD) that would involve human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This document describes the testing of a physical medical device (nitrile exam gloves), not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The ground truth is established by adherence to internationally recognized standards and specifications for medical gloves. These include:

    • ASTM D6319 (Standard Specification for Nitrile Examination Gloves for Medical Application)
    • ASTM D5151 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124 (Standard Test Method for Residual Powder on Medical Gloves)
    • ASTM D6978-05 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs)
    • ISO 10993-10 (Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization)
    • ISO 10993-11 (Biological evaluation of medical devices - Part 11: Tests for Systemic Toxicity)

    For the chemotherapy drug permeation test, a "minimum breakthrough detection time of >240 min on all three (3) specimens" is described as the expectation, although the nature of the test is to determine the permeation, not necessarily to meet a strict pass/fail for all drugs (as evidenced by the warning for Carmustine and Thiotepa). Therefore, the ground truth is based on the objective measurements derived from these standardized test methodologies.

    8. The Sample Size for the Training Set:

    This is not applicable. This filing is about a physical medical device, not a machine learning model. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221082

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-11-30

    (231 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rhino Non-Sterile Powder-Free Nitrile Exam Gloves – Periwinkle Color Tested for Use with Chemotherapy Drugs and Fentanyl is non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The tested chemotherapy drugs are as follows:

    Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU)(3.3mg/mL 3,300 ppm)14.2
    Cisplatin(1.0 mg/mL 1,000 ppm)> 240
    Cyclophosphamide (Cytoxan)(20 mg/mL 20,000 ppm)> 240
    Dacarbazine(10 mg/mL 10,000 ppm)> 240
    Doxorubicin Hydrochloride(2.0 mg/mL 2,000 ppm)> 240
    Etoposide (Toposar)(20.0 mg/mL 20,000 ppm)> 240
    Flurouracil(50.0 mg/mL 50,000 ppm)> 240
    Paclitaxel (Taxol)(6.0 mg/mL 6,000 ppm)> 240
    Thiotepa(10.0 mg/mL 10,000 ppm)43.2
    Fentanyl Tested as Follows:Fentanyl Citrate Injection (100 mcg/2mL)> 240

    Note: Carmustine and Thiotepa have extremely low permeation times of 14.2 and 43.2 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa

    Device Description

    Rhino Non-sterile Powder-Free Nitrile Exam Glove - Periwinkle Color Tested for Use with Chemotherapy Drugs and Fentanyl are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are offered in two sizes, medium and large. The color for all sizes is periwinkle. The model number is RHI-UNB5 with specific part sizes based on size as follows:

    Glove SizePart Numbers
    MRHI-UNB5M100BX1
    LRHI-UNB5L100BX1
    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device: "Rhino Non-sterile Powder-Free Nitrile Exam Glove – Periwinkle Color Tested for Use with Chemotherapy Drugs and Fentanyl." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing the extensive clinical study data one would find for novel, higher-risk devices. As such, the information you're requesting regarding AI-specific criteria, human expert adjudication, and large-scale training/test data for an AI algorithm is largely not applicable to this document because it describes a physical examination glove, not an AI/ML medical device.

    However, I will extract and present the available information related to acceptance criteria and the "study" (non-clinical testing) that proves the device meets them, while also clearly indicating what information is not present or applicable.

    Acceptance Criteria and Device Performance (Non-Clinical)

    The acceptance criteria for this medical glove are based on established ASTM and ISO standards for physical properties, barrier integrity, and biocompatibility, as well as an assessment for chemical permeation.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and reported performance are derived from Section VI of the 510(k) Summary: "Comparison of Technological Characteristics with Predicate Device" and Section VII: "Non-Clinical Studies."

    Characteristic / Test MethodAcceptance CriteriaReported Device Performance (Subject Device K221082)Basis for Acceptance (Standard)
    Material/Use
    Sterile vs. Non-SterileNon-SterileNon-SterileSimilar to Predicate
    Prescription or OTCOTC (Over-The-Counter)OTCSimilar to Predicate
    Single-use DisposableYesYesSimilar to Predicate
    Physical Dimensions
    Overall LengthMinimum: 230 mm (ASTM D6319)Large: 235 mm; Medium: 230 mmASTM D6319
    Palm WidthXS: 70 ± 10; S: 80 ± 10; M: 95 ± 10; L: 110 ± 10; XL: 120 ± 10 (ASTM D6319)Medium: 90-100 mm; Large: 103-113 mmASTM D6319
    Palm ThicknessMinimum: 0.05 mm (ASTM D6319)Large: 0.07 mm; Medium: 0.07 mmASTM D6319
    Finger ThicknessMinimum: 0.05 mm (ASTM D6319)Large: 0.10 mm; Medium: 0.10 mmASTM D6319
    Physical Properties
    Tensile Strength (Before Aging)Minimum: 14 MPa (ASTM D6319)Large: 33.4 MPa; Medium: 40.9 MPaASTM D6319
    Tensile Strength (After Aging)Minimum: 14 MPa (ASTM D6319)Large: 39.0 MPa; Medium: 35.5 MPaASTM D6319
    Ultimate Elongation (Before Aging)Minimum: 500% (ASTM D6319)Large: 520%; Medium: 550%ASTM D6319
    Ultimate Elongation (After Aging)Minimum: 400% (ASTM D6319)Large: 520%; Medium: 510%ASTM D6319
    Freedom from HolesG1, AQL 2.5 (ASTM D6319)Medium: 2 leakers in 50, 3 leakers in 100; Large: 4 leakers in 50, 4 leakers in 100, 4 leakers in 150. All "Pass".ASTM D6319, ASTM D5151
    Powder-Free (Powder Residue)< 2.0 mg/glove (ASTM D6319)Medium: 0.9 mg; Large: 0.6 mg. All "Pass".ASTM D6319, ASTM D6124
    Biocompatibility
    Acute Systemic ToxicityNon-acute systemic toxicity; no animals treated with test extracts exhibit greater reaction than control animals. (ISO 10993-11)Under conditions of study, did not show acute systemic toxicity in vivo/Pass.ISO 10993-11
    Primary Skin IrritationNon-irritating; Primary irritation Index $\le$ 1.0 (ISO 10993-10)Under conditions of study, not an irritant/Pass.ISO 10993-10
    Guinea Pig SensitizationNon-sensitizing; Grade < 1 (ISO 10993-10)Under conditions of study, not a sensitizer/Pass.ISO 10993-10
    Chemotherapy Drug Permeation
    Fentanyl Citrate Injection (100 mcg/2mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Cisplatin (1.0 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Cyclophosphamide (20 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Dacarbazine (10 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Doxorubicin HCl (2.0 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Etoposide (20.0 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Flurouracil (50.0 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Paclitaxel (6.0 mg/mL)Minimum Breakthrough Detection Time > 240 min (ASTM D6978-05)> 240 minutesASTM D6978-05
    Carmustine (3.3mg/mL)No explicit "acceptance criteria" beyond general permeation assessment, but breakthrough time given.14.2 minutes (Noted as "extremely low permeation time" and "Do Not Use with Carmustine")ASTM D6978-05; Warning/Labeling
    Thiotepa (10.0 mg/mL)No explicit "acceptance criteria" beyond general permeation assessment, but breakthrough time given.43.2 minutes (Noted as "extremely low permeation time" and "Do Not Use with Thiotepa")ASTM D6978-05; Warning/Labeling

    Note on Chemotherapy Permeation: For most chemotherapy drugs, the acceptance criteria are implicitly met by achieving a breakthrough detection time of >240 minutes. For Carmustine and Thiotepa, while permeation times are reported, the device is explicitly warned against use with these drugs, meaning the acceptance criterion for safe use with these specific drugs is effectively "not intended."

    Information Not Applicable or Not Provided

    The request is tailored for AI/ML device studies. This submission is for a physical medical glove. Therefore, most of the requested points are not applicable or the information is not provided in the 510(k) summary.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Specific sample sizes for each physical test (e.g., how many gloves for tensile strength, or how many for hole detection) are not explicitly detailed beyond the general "Pass" results against AQLs (Acceptable Quality Levels) or ASTM minimums. For permeation tests, breakthrough times are given for specific (presumably triplicate or similar standardized) specimens as per ASTM D6978-05.
    • Data Provenance: Not specified, but standard testing for medical devices typically occurs in specialized labs following good laboratory practices (GLP). It would be considered prospective testing for the purpose of this submission.
    • Retrospective or Prospective: The testing reported is prospective, specifically performed to demonstrate conformance to standards for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is a hardware device, not an AI algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" (i.e., whether the glove meets specifications) is established through standardized laboratory testing protocols, not human expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. See point 3. Testing is based on objective measurements against engineering and material standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-driven device. No human-in-the-loop studies or MRMC studies were conducted as part of this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Physical/Chemical Standards Compliance: The "ground truth" is compliance with established international and national standards (ASTM D6319, ASTM D5151, ASTM D6124, ASTM D6978-05, ISO 10993-10, ISO 10993-11) verified through laboratory testing.

    8. The sample size for the training set:

    • Not Applicable. There is no AI model or training set for this product.

    9. How the ground truth for the training set was established:

    • Not Applicable. There is no AI model or training set for this product.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1