Found 2 results

510(k) Data Aggregation

    K Number
    K191547

    Validate with FDA (Live)

    Device Name
    ReliefBand
    Date Cleared
    2019-08-30

    (79 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ReliefBand(R) is indicated for use in the treatment of nausea, vomiting associated with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness from pregnancy. ReliefBand is also indicated as an adjunct to antiemetics in reducing postoperative nausea. For over the counter use.

    Device Description

    ReliefBand® models 1.5 and 2.0 are battery-powered non-invasive digital therapeutic devices that are worn on the wrist like a watch. They are indicated for the counter use in the relief of mild to moderate nausea and retching associated with physician-diagnosed migraine, hangover, anxiety, motion sickness, chemotherapy and morning sickness associated with pregnancy as an adjunct to antiemetics in reducing mild to moderate postoperative nausea.

    ReliefBand 1.5 and 2.0 include two electrical contacts on the back of the device (ReliefBand 1.5) or on the clasp (ReliefBand 2.0), which must be placed firmly against the underside of the wrist at the P6 location (approximately two fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles). The electrodes provide relief of nausea and vomiting through electrical stimulation of the median nerve on the ventral side of the patient's wrist.

    The ReliefBand device has a user display that includes a five-level (ReliefBand 1.5) or 10-level (ReliefBand 2.0) LED that indicates the signal intensity. The patient can easily select the pulse amplitude of the electrical impulse and control the intensity of the stimulation via a pushbutton located on the face of the devices. A low battery indicator LED is also included on the face of the devices. ReliefBand 1.5 uses disposable batteries while ReliefBand 2.0 has a lithium-ion rechargeable battery. A fully charged battery will last for up to 18 hours of continuous use at the mid-power level or below.

    AI/ML Overview

    This FDA 510(k) summary for ReliefBand 1.5 and ReliefBand 2.0 indicates that the devices are substantially equivalent to their predicate devices (ReliefBand 1.5 K173233 and ReliefBand 2.0 K182960). The submission does not include a study proving that the device meets specific acceptance criteria based on performance because the claim for substantial equivalence is based on the lack of changes to the device's technical characteristics and performance specifications.

    Therefore, the requested information regarding acceptance criteria and the study proving the device meets them, including sample sizes, ground truth establishment, expert adjudication, or MRMC studies, is not available in the provided text.

    The document explicitly states:

    • "The subject and the predicate ReliefBand devices are identical. Only the wording in the indications for use has changed."
    • "The are no changes in technical characteristic between the predicate and the subject devices."
    • "The performance specifications for the applicant ReliefBand devices are identical, and thus substantially equivalent, to the predicate devices."
    • "Substantial equivalence was determined on the basis that there have been no changes in the physical device, technology, mechanism of action or intended use. Device modifications include only the clarified intended use, which includes an additional anticipated source of nausea (physician-diagnosed migraine) that Reliefband is intended to treat. These changes do not impact the safety of efficacy of the ReliefBand 1.5 or ReliefBand 2.0 devices."

    This means ReliefBand Technologies LLC is asserting that because the device itself (hardware, technology, mechanism of action, performance specifications) is unchanged from previously cleared devices, no new performance study is required to demonstrate its effectiveness against specific acceptance criteria for this 510(k) submission. The FDA concurred with this assessment for substantial equivalence.

    In summary, based on the provided text, the device's substantial equivalence is established by its identity to predicate devices, not by a new clinical or performance study with defined acceptance criteria.

    If such a study existed, it would likely have been part of the original 510(k) submissions for the predicate devices (K173233 and K182960).

    Ask a Question

    Ask a specific question about this device

    K Number
    K182960

    Validate with FDA (Live)

    Device Name
    ReliefBand
    Date Cleared
    2019-06-21

    (239 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ReliefBand 2.0 is indicated for use in the treatment of nausea, retching and vomiting due to motion sickness, chemotherapy and morning sickness associated with pregnancy. ReliefBand 2.0 is also indicated as an adjunct to antiemetics in reducing postoperative nausea.

    Device Description

    ReliefBand® 2.0 is a rechargeable, non-invasive digital therapeutic device that is worn on the wrist like a watch. It is a non-invasive device that is indicated for the over the counter use in the relief of mild to moderate nausea and vomiting associated with chemotherapy, motion sickness, mild to moderate nausea and vomiting associated with pregnancy, and as an adjunct to antiemetics in reducing mild to moderate postoperative nausea.

    ReliefBand 2.0® includes a hypoallergenic band with two electrical contacts on the clasp end of the band. The two contacts must be placed firmly against the underside of the P6 location, approximately two fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles. The electrodes in the wristband provide relief of nausea and vomiting through electrical stimulation of the median nerve on the ventral side of the patient's wrist.

    The ReliefBand® Device has a user display that incorporates a ten-segment linear LED that is used to display the intensity level. The patient can easily select the pulse amplitude of the electrical impulse and control the intensity of the stimulation via a pushbutton located on the edge of the device. Battery charge level is indicated by a separate four-segment circular LED. A fully charged battery will last for approximately 18 hours of continuous use at the mid-power level or below.

    AI/ML Overview

    The provided document describes the FDA 510(k) clearance for the ReliefBand 2.0 device, a transcutaneous electrical nerve stimulator for pain relief, specifically for nausea, retching, and vomiting due to various conditions.

    However, the document is a 510(k) summary (K182960), which primarily focuses on demonstrating substantial equivalence to a predicate device (ReliefBand 1.5) rather than providing detailed results from a clinical study proving its performance against pre-defined acceptance criteria.

    Therefore, many of the requested details about acceptance criteria and a study proving device performance (especially those related to diagnostic performance, such as sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, and training set details) are not present in this document.

    The document states:

    • "Substantial equivalence was determined on the basis that there have been no changes in the technology, mechanism of action or intended use."
    • "Device modifications include relocation of the electrodes from the back of the device face to the clasp end of the wrist band, changes in the size, shape and aesthetic appearance of the casing, a simpler and more streamlined user interface and a power source change from disposable to rechargeable batteries."
    • "These changes do not impact the safety of efficacy of the ReliefBand® 2.0 device."
    • "ReliefBand 2.0 was tested to the following standard: ISO 14971: Medical Devices Application of Risk Management to Medical Devices."

    This indicates that the clearance was based on comparative analysis with the predicate device and engineering/safety testing (like risk management and adherence to ISO standards), not on a clinical trial with specific performance acceptance criteria for diagnostic accuracy or efficacy. The device is a direct treatment device, not an AI/diagnostic device, so many of the questions related to AI performance are not applicable.

    Below is an attempt to address the questions based on the available information, noting where information is explicitly NOT provided.


    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative clinical acceptance criteria or a reported performance based on those criteria for the ReliefBand 2.0's therapeutic effect in a clinical study. Instead, its "performance" is demonstrated by its substantial equivalence to the predicate device (ReliefBand 1.5) and compliance with relevant standards.

    Acceptance Criteria (or Basis for Clearance)Reported Device Performance (or Basis for Substantial Equivalence Determination)
    Technological Characteristics: No changes that raise new questions about safety or effectiveness compared to predicate.ReliefBand 2.0 is substantially equivalent to ReliefBand 1.5. Device modifications (electrode relocation, size/shape/aesthetic changes, simpler UI, rechargeable battery) "do not impact the safety of efficacy of the ReliefBand® 2.0 device."
    Mechanism of Action: Same as predicate.Identical mechanism of action (electrical stimulation of the median nerve on the ventral side of the patient's wrist).
    Intended Use: Same as predicate.Identical indications for use (treatment of nausea, retching, and vomiting due to motion sickness, chemotherapy, morning sickness associated with pregnancy, and as an adjunct to antiemetics in reducing postoperative nausea).
    Safety: Device modifications maintain safety.Electrode Material: Hypoallergenic surgical grade (316L) stainless steel contacts, previously cleared material, "Does not adversely impact safety and/or effectiveness."
    Safety: Device modifications maintain safety.Wrist Band Material: Latex-free, "Conforms to ISO 10993 standard." (Predicate contained latex).
    Power/Battery: Meets standards and safe.Rechargeable 40 mA Li-Ion Polymer: "Conforms to IEC 62133 standard."
    Risk Management: Mitigate risks.ISO 14971: Medical Devices Application of Risk Management to Medical Devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable/Not provided. The document does not describe a clinical study with a "test set" of patients for measuring efficacy of the ReliefBand 2.0. The device's approval is based on substantial equivalence to a predicate, not new clinical performance data from a specific test set.
    • Data Provenance: Not applicable/Not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not provided. This device is a therapeutic device, not a diagnostic one that requires expert ground truth for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device, nor is it a multi-reader, multi-case study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a direct therapeutic device, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. As a therapeutic device cleared via substantial equivalence, new clinical efficacy data with defined ground truth is not the primary basis for this 510(k) clearance.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device that would have a "training set."

    9. How the ground truth for the training set was established

    • Not applicable.

    Summary of findings from the document:

    The FDA 510(k) clearance for ReliefBand 2.0 (K182960) is based on demonstrating substantial equivalence to a previously cleared predicate device (ReliefBand 1.5). The document highlights that there have been no changes in technology, mechanism of action, or intended use that would raise new questions about safety or effectiveness. The modifications are primarily cosmetic, user interface improvements, and a change from disposable to rechargeable batteries. Compliance with ISO 14971 (Risk Management) and IEC 62133 (Battery safety), as well as the use of ISO 10993 compliant materials for the wristband, are mentioned as part of the safety assessment. There is no indication of a new clinical study involving a test set of patients or experts that would generate the type of performance data typically associated with diagnostic AI devices.