Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243063

    Validate with FDA (Live)

    Date Cleared
    2025-02-10

    (136 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for aesthetic use, including facial and neck stimulation or body skin stimulation.

    Device Description

    The subject device, the MyoLight Microcurrent Handpieces are intended to be used together with a main control unit that is equipped with a color touch screen that significantly facilitates the use of the device. There are two MyoLight Microcurrent Handpieces that connect to the main unit. The small handpiece (diameter: 3.9 cm, Length: 14 cm) is generally used on the face and the large handpiece (diameter: 6.2 cm, Length: 15 cm) is generally used on the body. The handpieces are different in dimensions but not function. The handpieces contain the electrodes that deliver microcurrent for aesthetic use. The subject device is intended to be used together with a main control unit that is equipped with a color touch screen that significantly facilitates the use of the handpieces. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about therapy type, remaining therapy time and main therapy parameters on the screen.

    AI/ML Overview

    The provided context does not offer information about the acceptance criteria or a study proving that the device meets specific acceptance criteria in the typical format of a clinical or performance study with metrics like sensitivity, specificity, or AUC.

    The document is a 510(k) summary for a medical device (MyoLight Microcurrent Handpiece) and focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed performance study with acceptance criteria.

    However, based on the non-clinical testing section, we can infer some "acceptance criteria" through compliance with recognized standards.

    Here's the breakdown of the information that can be extracted, and where the requested information is not available:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) for substantial equivalence, the "acceptance criteria" are implicitly met by demonstrating compliance with recognized standards and similarity to predicate devices. The "reported device performance" is essentially the device's conformance to these standards and its specifications aligning with the predicate devices.

    Acceptance Criterion (Inferred from Standards & Equivalence)Reported Device Performance (MyoLight Microcurrent Handpiece)
    Electrical Safety and Essential Performance (IEC 60601-1:2005/A1:2012)Compliant with IEC 60601-1 leakage current requirements and isolation requirements.
    Electromagnetic Compatibility (EMC) (IEC 60601-1-2:2014/EN 60601-1-2:2015)Compliant with IEC 60601-1-2 (implied by listing the standard as tested).
    Particular Requirements for Nerve and Muscle Stimulators (IEC 60601-2-10 Edition 2.1 2016-04)Compliant with IEC 60601-2-10 (implied by listing the standard as tested).
    Biocompatibility (ISO 10993 and FDA Guidance)Biocompatibility tests performed (exact results not detailed, but implied to be acceptable).
    Software Validation & VerificationSoftware V&V tests performed (exact results not detailed, but implied to be acceptable).
    Bench Testing for PerformanceBench testing performed to verify performance (exact details not provided, but used to support substantial equivalence).
    Specifications within Acceptable Range of Predicate Devices (e.g., Max output voltage, current, power density, net charge, phase charge, pulse duration, frequency)Max output voltage (30V @ 500 Ω, 33V @ 1,000 Ω, 35V @ 10,000 Ω), Max output current (60 mA @ 500 Ω, 33 mA @ 1000 Ω, 3.5 mA @ 10,000 Ω), Max current density (82 µA/mm² @ 500 Ω), Max average power density (0.012W/cm² @ 500 Ω), Net Charge (0 µC), Max phase charge (23 µC), Duration of primary (128, 256, 384 µs), Pulse Duration (128, 256, 384 µs), Frequency (1 ~ 63 Hz). These are compared to predicate devices and deemed "substantially equivalent."
    Safety - Automatic Shut OffYES
    Safety - Patient Contact MaterialStainless Steel 304, PMMA (biocompatibility validated).

    2. Sample size used for the test set and the data provenance
    The document does not describe a clinical test set in terms of patient samples. The testing described is non-clinical (bench testing, software V&V, and compliance with electrical safety and EMC standards). Therefore, there is no information on sample size or data provenance in the context of a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable, as no clinical test set using expert-established ground truth is described.

    4. Adjudication method for the test set
    Not applicable, as no clinical test set with adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. The device is a microcurrent handpiece, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable, as the device is not an algorithm for diagnostic or prognostic purposes. It is a physical device delivering electrical stimulation, with software controls for operation.

    7. The type of ground truth used
    For non-clinical testing:

    • Electrical Safety, EMC, and Nerve/Muscle Stimulator Standards: Compliance is based on objective measurements against defined limits in the respective IEC standards.
    • Biocompatibility: Likely based on laboratory assays and tests according to ISO 10993 standards.
    • Software Validation & Verification: Based on predefined software requirements specification and test protocols.
    • Bench Testing: Based on objective measurements of physical and electrical parameters against design specifications.

    8. The sample size for the training set
    Not applicable, as this is not an AI/machine learning device. The "training set" concept doesn't apply.

    9. How the ground truth for the training set was established
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202175

    Validate with FDA (Live)

    Device Name
    OxyLight 2.0
    Date Cleared
    2021-12-02

    (485 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oxylight 2.0 is intended for dermatological use by physicians and healthcare professionals for the following: LED Technology is intended for:

    -Blue LED 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

    -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.

    • Yellow LED 590nm - treatment of periorbital wrinkles and rhytides.

    Microdermabrasion is intended for exfoliation of the skin.

    Oxygen spray is intended to refresh the skin.

    Device Description

    The subject device Oxylight 2.0 is a skin therapy system that was modified from the predicate device OxyLight cleared previously under K200104 for the LED Light Panel Therapy (Class 2), Microdermabrasion (Class 1, 510k exempt) and Oxygen Spray (General Wellness). The OxyLight 2.0 system includes two new LED handpieces called MyoLight. The MyoLight handpieces connect to the main unit. The small handpiece (diameter: 3.9 cm, Length: 14 cm) is generally used on the face and the large handpiece (diameter: 6.2 cm, Length: 15 cm) is generally used on the body.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called OxyLight 2.0. It is primarily focused on demonstrating substantial equivalence to a previously cleared predicate device (OxyLight, K200104) and thus does not contain the typical structure of a study proving a device meets specific clinical acceptance criteria for an AI/algorithm-based device.

    The "acceptance criteria" and "device performance" in this context refer to the technical specifications and safety standards that the OxyLight 2.0 must meet to be considered substantially equivalent to its predicate. There is no AI or algorithm involved, and therefore, no performance metrics like accuracy, sensitivity, or specificity relative to a diagnostic task are presented.

    Instead, the "study" proving the device meets acceptance criteria primarily involves non-clinical bench testing, electrical safety, electromagnetic compatibility, and biocompatibility.

    Here's an analysis based on the provided document, addressing your points where applicable, and noting where the information is not relevant to this type of device submission:


    1. A table of Acceptance Criteria and the Reported Device Performance

    For this device (OxyLight 2.0), the "acceptance criteria" are not clinical performance metrics for an AI, but rather technical specifications and safety standards aligned with the predicate device. The "reported device performance" is the confirmation that the new device meets these technical specifications and passes the required non-clinical tests.

    Criteria Type (Not "Acceptance Criteria" in AI Sense)Specific Criterion (Based on Predicate)Reported Device Performance (OxyLight 2.0)Comment (from document)
    Device NameOxyLightOxyLight 2.0N/A
    510(k) NumberK200104K202175N/A
    ApplicantRAJA Trading Company, Inc.RAJA Trading Company, Inc.N/A
    Intended use/Product CodePowered laser surgical instrument,Power dermabrasionClass II/GEX/878.4810Class I/GFE/878.4820Powered laser surgical instrument,Power dermabrasionClass II/GEX/878.4810Class I/GFE/878.4820Identical
    Indications for Use (LED)(Specific wavelengths & conditions)(Specific wavelengths & conditions)Identical (for listed LED indications)
    Design and Mode of ActionPanel with LED arrayPanel with LED array + 2 handheld handpiecesSimilar (addition of handpieces)
    Number of LEDs (Panel)840840Similar power density and energy flux
    Number of LEDs (Handpieces)N/A (Predicate had no handpieces)Face Handpiece - 8, Body Handpiece - 18Implicitly, these must meet safety standards and produce specified light output.
    Treatment Time20 minutes20 minutesIdentical
    Light SourceLEDLEDIdentical
    Operation InterfaceDisplay ScreenDisplay ScreenIdentical
    Wavelength (Blue)465 nm +/- 5nm465 nm +/- 5nmIdentical
    Energy Output (Blue)54 J/cm²54 J/cm²Identical
    Power Output (Blue)45 mW/cm²45 mW/cm²Identical
    Wavelength (Red)625 nm +/- 5nm625 nm +/- 5nmIdentical
    Energy Output (Red)120 J/cm²120 J/cm²Identical
    Power Output (Red)100 mW/cm²100 mW/cm²Identical
    Wavelength (Yellow)590 nm +/- 5nm590 nm +/- 5nmIdentical
    Energy Output (Yellow)42 J/cm²42 J/cm²Identical
    Power Output (Yellow)35 mW/cm²35 mW/cm²Identical
    Non-Clinical TestingCompliance with standards (e.g., electrical safety, EMC, biocompatibility)Compliance was verified via testing.IEC 60601-1, IEC 60601-1-2, ISO 10993, Software V&V, Bench Testing.

    Note: This table reflects the comparison presented in the 510(k) summary, which is a key part of demonstrating substantial equivalence for physical devices.


    Regarding AI/Algorithm-Specific Questions (Not applicable to this 510(k) submission):

    This 510(k) submission is for a physical medical device (LED phototherapy and microdermabrasion system), not an AI/algorithm-based diagnostic or assistive software. Therefore, the following points are not applicable to this document:

    1. Sample sizes used for the test set and data provenance: No test set of patient data (e.g., images for classification) was used for performance evaluation of an algorithm. The testing was physical device validation.
    2. Number of experts used to establish the ground truth for the test set and their qualifications: No ground truth established by experts for an algorithm's performance, as there is no algorithm.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable. There is no human-in-the-loop AI assistance being evaluated.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. "Ground truth" in the AI sense is not relevant here.
    7. The sample size for the training set: Not applicable, as there is no AI algorithm being developed or trained.
    8. How the ground truth for the training set was established: Not applicable, as there is no AI algorithm being developed or trained.

    Summary of the Study Proving Device Meets Criteria (Based on the document):

    The "study" conducted for the OxyLight 2.0 to prove it meets acceptance criteria (i.e., is substantially equivalent and safe/effective for its intended use) is documented under "8. Non-Clinical Testing".

    • Type of Study: Non-clinical validation and verification.
    • Tests Conducted:
      • IEC 60601-1:2005/A1:2012: General Requirements for Basic Safety and Essential Performance (electrical safety).
      • IEC 60601-1-2:2014/EN 60601-1-2:2015: Electromagnetic Compatibility (EMC).
      • Biocompatibility Tests per ISO 10993 and FDA Guidance: To ensure materials in contact with skin are safe.
      • Software Validation & Verification Test: For the device's operational software.
      • Bench Testing: To verify the performance of the device's components and outputs (e.g., light wavelengths, energy/power outputs, and the function of the new handpieces).
    • Assertions from the document: The applicant states these tests were "conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device."
    • Data Provenance (Implicit): The tests were performed in a controlled laboratory setting, adhering to international standards. The data is prospective as it was generated specifically for this submission. The origin is likely the manufacturer or accredited testing labs.
    • Clinical Testing: The document explicitly states, "No clinical study is included in this submission." This indicates that substantial equivalence was demonstrated without new clinical trials, relying on the established safety and efficacy of the predicate device and the non-clinical testing of the modifications.

    In conclusion, for this specific 510(k) submission, the "acceptance criteria" are the established technical and safety specifications for similar devices, primarily the predicate. The "study" involves comprehensive non-clinical engineering and safety testing to demonstrate that the new device adheres to these specifications and introduces no new safety or efficacy concerns that would preclude a substantial equivalence determination.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200104

    Validate with FDA (Live)

    Device Name
    Oxylight
    Date Cleared
    2020-06-04

    (139 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oxylight is intended for dermatological use by physicians and healthcare professionals for the following:

    LED Technology is intended for:

    -Blue LED = 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

    -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.

    • Yellow LED 590nm - treatment of periorbital wrinkles and rhytides.

    Microdermabrasion is intended for exfoliation of the skin.

    Oxygen spray is intended to refresh the skin.

    Device Description

    The Oxylight system is a skin therapy system that uses three major modalities delivered by different attachments powered from a single energy source and operated from a universal control unit. The treatments include, LED Light Panel Therapy (Class 2), Microdermabrasion (Class 1, 510k exempt) and Oxygen Spray. The Oxygen Spray is delivered through the same panel used for LED light therapy and separate handpieces.

    AI/ML Overview

    The provided document is a 510(k) Pre-market Notification, which focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than providing extensive clinical study data to prove novel efficacy or meet specific acceptance criteria for a new clinical claim. Therefore, much of the requested information regarding acceptance criteria, specific device performance against those criteria, and detailed clinical study methodologies is not available in this document.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present a table of specific clinical acceptance criteria or human-centered performance metrics. Instead, it compares the technical characteristics of the Oxylight device (the "subject device") with predicate devices to argue for substantial equivalence in safety and efficacy.

    The relevant "performance data" section focuses on safety testing against recognized electrical and electromagnetic compatibility standards.

    StandardTest TitleTest ArticleAcceptance Criteria (Implicit)Reported Performance
    IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012Medical electrical equipment Part 1: General requirements for basic safety and essential performanceOxyLightCompliance with standard IEC 60601-1Pass
    EN 60601-1-2:2015, EN 61000-3-2:2014, EN 61000-2-3:2013Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and testsSapphire 3 OxyLightCompliance with standard EN 60601-1-2 et al.Pass
    ISO 14971:2012Medical devices – Application of risk management to medical devicesNot directly tested (risk management applied)Application of risk management principlesApplied

    2. Sample Size Used for the Test Set and the Data Provenance

    The document explicitly states: "Clinical testing was not performed with this device."

    Therefore, there is no test set in the sense of a patient cohort, nor data provenance from such a set. The "test article" for safety testing was the device itself (OxyLight / Sapphire 3 OxyLight).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since no clinical testing was performed, no ground truth was established by experts for a test set in a clinical context.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done, as indicated by the statement "Clinical testing was not performed with this device."
    The document does not mention any human reader performance or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The Oxylight is a physical device (LED panel, microdermabrasion, oxygen spray) for dermatological use, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    For the purpose of substantial equivalence, the "ground truth" implicitly relies on the established safety and efficacy of the predicate devices with similar technical characteristics and indications for use. For the safety tests conducted, the ground truth was compliance with the specified international safety and EMC standards.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/algorithm-based device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1