Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K123805

    Validate with FDA (Live)

    Date Cleared
    2013-03-20

    (99 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessing Products Corporation (RPC) E-Z Chek® Blood Leak Test Strips are designed to indicate the presence or absence of blood in dialysate during the hemodialysis procedure. The test strips will measure equal to 0 mg HGB/dL and 0.25 mg HGB/dL in dialysate.

    The Reprocessing Product Corporation (RPC) E-Z Chek® Blood Leak Test Strips are indicated for detecting the presence or absence of blood in dialysate during hemodialysis treatments.

    Device Description

    Device is semi-quantitative, reagent test strip comprised of a pad impregnated with chemicals which change color upon contact with blood in dialysate. The pad is attached to a plastic strip for handling.

    AI/ML Overview

    The provided document describes the Reprocessing Products Corporation (RPC) E-Z Chek® Blood Leak Test Strips. Here's an analysis of the acceptance criteria and the study outlined:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a quantitative pass/fail format typical for medical device verification. Instead, it describes the intended performance and the study's findings regarding that performance.

    Acceptance Criteria (Implied from Intended Use)Reported Device Performance
    Indicate the presence or absence of blood in dialysate.The product consistently generates color change which meets the color block for the reference solution concentration.
    Measure equal to 0 mg HGB/dL (no blood).Reference solutions were used to evaluate performance between 0 mg HGB/dL and 0.25 mg HGB/dL Bicarbonate/Acetate dialysate. The data confirm the product meets the color block for this concentration.
    Measure 0.25 mg HGB/dL (presence of blood at a specific detectable threshold).Reference solutions were used to evaluate performance between 0 mg HGB/dL and 0.25 mg HGB/dL Bicarbonate/Acetate dialysate. The data confirm the product meets the color block for this concentration.
    Appropriate performance for use in hemodialysis dialysate during treatment.These data demonstrate appropriate performance for use in hemodialysis dialysate used in treatment.
    Technological characteristics: Detection based on pseudoperoxidative activity of hemoglobin and myoglobin, catalyzing oxidation of an indicator by an organic hydroperoxide, producing a color change consistent with color blocks.Confirmed through the performance study where the device generates color change that meets the color block for reference solutions, implying the underlying technological principle is functional and produces the expected visual outcome.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "Reference solutions" were used, and "data confirm the product consistently generates color change," but it doesn't quantify the number of tests performed or the sample size of distinct reference solutions/concentrations tested.
    • Data Provenance: Not explicitly stated. The study appears to be a laboratory-based performance evaluation using reference solutions, rather than human clinical data from a specific country. It is almost certainly retrospective in the sense that the test strips were manufactured and then tested.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for this device is established by the known concentrations of hemoglobin (HGB) in the prepared reference solutions, not by expert human interpretation. The "color block" comparison itself would be a visual assessment, but the ground truth is the chemical concentration.

    4. Adjudication Method for the Test Set

    Not applicable. The performance is evaluated by observing a color change and comparing it to a visual "color block" corresponding to known HGB concentrations. There is no mention of human adjudication for the results beyond this direct comparison.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a diagnostic test strip where performance is assessed by comparison to reference color blocks, not by interpretation by multiple human readers. Therefore, an effect size of human readers improving with AI vs. without AI assistance is not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this appears to be a standalone performance evaluation. The device's performance is determined by its inherent color-changing reaction to blood in dialysate, and its comparison to established color blocks. There is no "human-in-the-loop" aspect to its fundamental operation or the evaluation method described beyond visually comparing the strip's color to a reference.

    7. The Type of Ground Truth Used

    The ground truth used is known concentrations of hemoglobin (HGB) in prepared reference solutions. These solutions serve as the objective standard against which the test strip's color-changing reaction is evaluated.

    8. The Sample Size for the Training Set

    Not applicable. This device is a chemical reagent test strip, not an AI or machine learning algorithm that requires a "training set." Its chemical formulation dictates its performance, which is then verified.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101750

    Validate with FDA (Live)

    Date Cleared
    2011-01-07

    (199 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessing Products Corporation Test Strips are intended for use in the monitoring of pH in Dialysate and Water.
    The Reprocessing Product Corporation (RPC) pH Test Strips are indicated for determining the pH of acid dialysates, bicarbonate dialysates, or water. These Test Strips are indicated for testing acid/bicarbonate dialysates and water.
    pH Test Strips are comprised of two different pH ranges:
    pH = 6.8-8.5 (K100-0117) is for acid/bicarbonate dialysate, bicarbonate concentrate and water
    pH = 0-14 (K100-0104) is for water used to make up dialysate.

    Device Description

    Device is semi-quantitative, reagent test strip comprised of a pad impregnated with chemicals which change color upon contact with Dialysate or Water respectively. The pad is attached to a plastic strip for handling.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Reprocessing Products Corporation pH Test Strips, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of specific quantitative metrics like sensitivity or accuracy percentages. Instead, the performance is described qualitatively.

    Acceptance Criteria (Implied)Reported Device Performance
    Accurate pH determination (correlation to Hydrogen ion concentration)"The test strip contains a specialized chemical formulation that reacts with the Hydrogen ion concentration in solutions of dialysate and water. The reaction results in a color change which is correlatable to the concentration of Hydrogen ions in the solution."
    Consistent and correct color changes across the specified pH ranges"The data confirm that the product consistently generates color changes which match the color blocks for the Reference solution pH. Reference solutions were used to evaluate pH performance between 6.8 and 8.5 in acid/bicarbonate dialysate, bicarbonate concentrate and water, and the pH performance between 0 and 14 in water used to prepare dialysate."
    Appropriate performance for intended use (hemodialysis dialysate and water)"These data demonstrate appropriate performance for use in hemodialysis dialysate and water."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a numerical sample size for the test set. It mentions "Reference solutions were used to evaluate pH performance." This implies a series of solutions at various pH levels within the specified ranges, but the exact number is not provided.
    • Data Provenance: The document does not explicitly state the country of origin. It can be inferred that the testing was conducted by the manufacturer, Reprocessing Products Corporation, based in Tucson, AZ, USA. The data appears to be prospective, as it describes the evaluation of the product's performance against reference solutions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts to establish ground truth for the test set. The ground truth appears to be established by the reference solutions themselves, whose pH values are presumably known through independent, calibrated means (e.g., pH meters).

    4. Adjudication Method for the Test Set

    No adjudication method is described or implied, as the ground truth is established by reference solutions and the device's color changes are matched against these.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned. This device is a test strip with a visual color change, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study was done. The "Performance" section explicitly states: "The data confirm that the product consistently generates color changes which match the color blocks for the Reference solution pH." This refers to the device's inherent ability to produce accurate results independent of human interpretation or assistance, by visually matching the color to a provided chart.

    7. Type of Ground Truth Used

    The ground truth used was reference solutions with known pH values. The device's color changes were compared to the color blocks corresponding to these reference pH values.

    8. Sample Size for the Training Set

    The concept of a "training set" is not applicable here in the context of machine learning. This device is a chemical reagent test strip, not a machine learning algorithm. Its performance is based on chemical reactions, not trained data.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there is no "training set" in the machine learning sense for this type of device. The chemical formulation of the test strip is designed to react in a specific way to different pH levels, and its performance is evaluated against chemically prepared reference solutions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090338

    Validate with FDA (Live)

    Date Cleared
    2009-05-08

    (87 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessing Product Corporation (RPC) Total Chlorine Test Strips are indicated for detection of Free Chlorine and Total Chlorine in water. These Test Strips are indicated for testing water used to prepare dialysate (Free Chlorine and Chloramines) and for testing rinse water following dialysis equipment disinfection (Free Chlorine).

    Device Description

    Device is semi-quantitative, reagent test strip comprised of a pad impregnated with chemicals which change color upon contact with Free Chlorine or Chloramines. The pad is attached to a plastic strip for handling. The color change is interpreted by the use of color blocks on two separate (ranges) color charts. Gradations for the Ultra-Low Total Chlorine (K100-0118) color chart include 0 ppm, 0.01 ppm, 0.02 ppm, 0.05ppm, 0.1ppm, 0.15ppm, 0.2 ppm. Gradations for the E-Z Chek® Sensitive Total Chlorine (K100-0106) color chart include 0 ppm, 0.1 ppm, 0.5 ppm and 3.0 ppm.

    AI/ML Overview

    The provided document describes the Reprocessing Products Corporation (RPC) Total Chlorine Test Strips. Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" in a quantitative pass/fail format with specific thresholds. Instead, it describes the intended performance and then presents data confirming that performance.

    Acceptance Criteria (Implied)Reported Device Performance
    Detection of Chlorine Concentrations:The data confirm the product consistently generates color change which correlates with the color block for the Reference solution concentration.
    For water to prepare dialysate:These data demonstrate accurate detection performance, below, equal to, and above the dialysis level of 0.1 ppm for Chloramines (measured as Total Chlorine).
    Detection of Total Chlorine (Chloramines) at 0.1 ppm
    For dialysis equipment rinse water:These data demonstrate accurate detection performance, below, equal to, and above the dialysis level of 0.5 ppm for Free Chlorine.
    Detection of Free Chlorine at 0.5 ppm
    Consistency of Color Change and Correlation with Reference SolutionsThe data confirm the product consistently generates color change which correlates with the color block for the Reference solution concentration. Reference solutions were used to evaluate performance between 0 and 10 ppm Chlorine. These data demonstrate appropriate performance for use in testing dialysis equipment rinse water and for testing water for dialysate preparation.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document states that "Reference solutions were used to evaluate performance between 0 and 10 ppm Chlorine." However, it does not specify the number of individual tests or measurements performed with these reference solutions. It does list the specific gradations for each test strip range (e.g., 0 ppm, 0.01 ppm, 0.02 ppm for Ultra-Low Total Chlorine).
    • Data Provenance: The document does not specify the country of origin of the data. It also does not explicitly state if the study was retrospective or prospective. Given the nature of a 510(k) submission for a test strip, it's highly likely that the performance evaluation was conducted prospectively in a laboratory setting using prepared reference solutions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of testing for chemical reagent strips does not typically involve human experts establishing "ground truth" in the same way as medical imaging or clinical diagnoses. The "ground truth" is established by the known concentrations of the reference solutions themselves. Therefore, information on the number and qualifications of experts for ground truth is not applicable in this context.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used when there's subjective interpretation or disagreement among human readers/experts. For these test strips, "ground truth" is determined by the precisely known concentrations of the reference solutions. The interpretation of the color change on the strip is then compared against a pre-defined color chart.

    The document indicates: "The color change is interpreted by the use of color blocks on two separate (ranges) color charts." This suggests a direct comparison method rather than an adjudication process. Any discrepancies would likely point to a failure of the strip or the reading process, not an issue with establishing ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed, nor is it applicable for this type of device. MRMC studies are used to evaluate the impact of a device (often AI) on human reader performance in interpreting medical images or clinical data. This device is a semi-quantitative chemical test strip; its performance is assessed against known chemical concentrations, not through human diagnostic accuracy with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study was performed. The entire performance section "The data confirm the product consistently generates color change which correlates with the color block for the Reference solution concentration. Reference solutions were used to evaluate performance between 0 and 10 ppm Chlorine..." describes the device's performance in isolation using reference solutions. It does not involve human readers interpreting results in a variable "human-in-the-loop" setting.

    7. Type of Ground Truth Used

    The type of ground truth used is known chemical concentrations (reference solutions). The performance of the test strips is validated by comparing the color change produced by the strip when exposed to these reference solutions against the expected color block for that known concentration.

    8. Sample Size for the Training Set

    The concept of a "training set" is primarily relevant for machine learning algorithms. Since this device is a chemical reagent test strip, it does not involve a machine learning algorithm and therefore does not have a "training set" in the conventional sense. The development of the chemical formulation and color chart would involve R&D and calibration, but not a dataset for algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" for a machine learning algorithm, this question is not applicable. The ground truth for the performance evaluation (test set) was established by using precisely prepared reference solutions with known concentrations of chlorine.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974426

    Validate with FDA (Live)

    Date Cleared
    1999-02-10

    (471 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Micro-X Potency Test Strips are intended for use as a semi-quantitative test of the presence of peracetic acid in dialyzers after reprocessing. This device may be used with all peroxyacetic/peracetic acid germicides used in the reprocessing of dialyzers as part of an established quality assurance program.

    Device Description

    Micro X Potency Test Strips are approximately 0.25" x 2.0" packaged 100strips /container. The strips are labeled for use in Reprocessing Hemodialyzers. Micro Xen Potency Test Strips are impregnated with an indicator solution which is activated by the presence of peracetic acid. The reaction results in a color change of the indicator which provides visual confirmation of the presence of peracetic acid.

    AI/ML Overview

    The provided text describes a medical device called "Micro X Potency Test Strips" and its 510(k) submission. However, the document does not contain explicit "acceptance criteria" in a tabulated format or a detailed description of a "study that proves the device meets the acceptance criteria" in the way one would expect for a quantitative performance study involving AI or complex diagnostics.

    The core of the submission is a demonstration of substantial equivalence to a predicate device, not necessarily a demonstration of meeting predefined numerical performance targets in a standalone study.

    Here's an attempt to extract and infer the requested information based on the provided text, while acknowledging the limitations of what's available:


    1. A table of acceptance criteria and the reported device performance

    Based on the document, the primary "acceptance criterion" is the demonstration of "identical performance" to the predicate device. The device's fundamental characteristic is a binary output (YES/NO) for the presence of peracetic acid above a minimum level.

    Acceptance Criteria (Inferred from function)Reported Device Performance
    Visual confirmation of peracetic acid presence upon activation above a minimum level. (YES/NO indication)Identical performance to Renalin Indicator Test Strips
    Activates only in the presence of peracetic acid.Identical performance to Renalin Indicator Test Strips
    Does not provide a measure of concentration (i.e., not quantitative).The test is not quantitative.

    2. Sample size used for the test set and the data provenance

    The document states: "Comparative testing between Micro X Potency Strips, Renalin Indicator Test Strips indicate identical performance among the test strips."

    • Sample Size for Test Set: Not specified. The phrase "among the test strips" suggests multiple strips were tested, but no specific number is provided.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). It appears to be a laboratory comparison performed by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. The "ground truth" for this type of chemical indicator would typically be established by known concentrations of peracetic acid solutions, rather than expert interpretation of a complex image or signal.

    4. Adjudication method for the test set

    Not applicable/Not specified. Given the nature of a chemical indicator changing color, visual observation is the method. There's no indication of multiple readers adjudicating results or complex decision-making.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No.
    • Effect Size: Not applicable. This device is a simple chemical indicator, not an AI-powered diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This is a manual chemical indicator, requiring human visual interpretation of a color change. It is not an algorithm or an AI device.

    7. The type of ground truth used

    The implicit ground truth is the known presence/absence and concentration of peracetic acid solutions. The device is activated by the presence of a minimum level of peracetic acid.

    8. The sample size for the training set

    Not applicable. This device does not use machine learning or AI, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974064

    Validate with FDA (Live)

    Date Cleared
    1998-12-07

    (406 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Micro-X Residual Test Strips are intended for use as a semi-quantitative test of the residual presence of peracetic acid in dialyzers after reprocessing. This device may be used with all peroxyacetic/peracetic acid germicides used in the reprocessing of dialyzers, sanitation of water treatment systems and dialysis machine disinfection as part of an established quality assurance program.

    Device Description

    Micro Xtm Residual Test Strips are approximately 0.25" x 4.75" packaged 100strips /container. The strips are labeled for use in Reprocessing Hemodialyzers. Micro X " Test Strips have a test pad, which is impregnated with peroxidase, which liberates oxygen from peracetic acid upon contact with the solution. The oxygen reacts with the chromophore, color indicator, which is also present on the test pad, resulting in a color change. The color change is compared to the color scale on the container and the result is read directly in mg/L.

    AI/ML Overview

    The provided text focuses on the 510(k) summary for Micro X™ Residual Test Strips and the FDA's clearance letter. It describes the device, its intended use, and states that comparative testing was done. However, it does not contain enough information to fully describe the acceptance criteria and the study in the detail requested.

    Here's what can be extracted and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document mentions "Comparative testing between Micro Xtm Residual Strips" and states the device is "Substantially equivalent to Renalin Residual Test Strips." This implies the acceptance criterion was likely meeting or closely approximating the performance of the predicate device (Renalin Residual Test Strips). However, the specific metrics for comparison (e.g., sensitivity, specificity, accuracy, precision, detection limits, agreement percentage) and the actual numerical results are not provided.

    Acceptance Criteria (Implied)Reported Device Performance
    Performance substantially equivalent to Renalin Residual Test StripsMet (based on 510(k) clearance)
    (Specific metrics and thresholds not provided in the document)(Specific results not provided in the document)

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document only states "Comparative testing." It does not specify the number of samples (e.g., how many tests were run, across what range of concentrations, or how many dialyzers/systems were tested).
    • Data Provenance: Not specified. It's unclear if the data was retrospective or prospective, or from what country/locations the samples originated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not mentioned. The testing seems to be a chemical/analytical comparison rather than relying on human interpretation directly for primary ground truth.

    4. Adjudication method for the test set:

    • Not mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:

    • No. This type of study is not applicable as the device is a chemical test strip, not an imaging or diagnostic device requiring human interpretation of complex data.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device itself is a standalone, non-algorithmic chemical test strip. Its performance is inherent to its chemical reaction and color change. Users read the color change against a scale, but the core "algorithm" is the strip's chemistry. Therefore, its performance is its standalone performance.

    7. The type of ground truth used:

    • Implied Ground Truth: The ground truth for this type of device would typically be:
      • Known concentrations of peracetic acid: prepared standards measured by analytical chemistry methods (e.g., titration, spectrophotometry) as the gold standard.
      • Comparison to a predicate device: The performance of the Renalin Residual Test Strips would serve as a reference in this comparative study.
    • The document doesn't explicitly state how the ground truth was established, but it would logically involve creating solutions with known peracetic acid concentrations.

    8. The sample size for the training set:

    • Not applicable/Not mentioned. Chemical test strips typically do not have "training sets" in the same way machine learning algorithms do. Their performance is based on their chemical composition, which is developed and optimized, but not "trained" with data.

    9. How the ground truth for the training set was established:

    • Not applicable/Not mentioned for the reasons above.

    In summary: The provided text is a high-level 510(k) summary and FDA clearance letter. It confirms that comparative testing was done to establish substantial equivalence with a predicate device for Residual Test Strips, but it lacks the detailed technical specifications and study results that would describe specific acceptance criteria and performance data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1