Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K130237

    Validate with FDA (Live)

    Device Name
    PONTIC PAINT
    Date Cleared
    2013-04-05

    (64 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pontic Paint is intended for use as a light cure coloring agent of thermoplastic aligners.

    Device Description

    The Pontic Paint is a light cure coloring agent for thermoplastic aligners. The Pontic Paint is formulated in a viscosity that is easily mixed and bonds chemically to a typically difficult to adhere to surface, a thermoplastic aligner now common in the dental industry. This paint, upon polymerization, is in Vita Shades A1. A2, and Bl allowing the appearance of a realistic tooth under the thermoplastic aligner thereby making a missing tooth situation appear aesthetically pleasing to the patient undergoing orthodontic treatment using clear aligner systems. The light cure property of the paint allows the user to determine the polymerization time required by simply exposing the adhesive to an LED light source until set. The Pontic Paint will be marketed in a jar for ease of mixing and application with a brush.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called "Pontic Paint." However, the document does not contain any information about acceptance criteria, device performance results, or a study to prove the device meets acceptance criteria.

    The document primarily focuses on:

    • Identifying the applicant (Reliance Orthodontic Products, Inc.)
    • Describing the device ("Pontic Paint" - a light-cure coloring agent for thermoplastic aligners)
    • Stating its intended use and population
    • Identifying the predicate devices (Flowtain™ and Plastic Conditioner)
    • The FDA's decision of substantial equivalence.

    Therefore, I cannot provide the requested information in the tables or bullet points because it is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123348

    Validate with FDA (Live)

    Date Cleared
    2013-02-28

    (120 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLEAR ALIGNER ADHESIVE is intended for use as an orthodontic bonding adhesive for brackets and appliances to thermoplastic aligner surfaces.

    Device Description

    The Clear Aligner Adhesive is a light cure, orthodontic adhesive for metal, composite and porcelain brackets and appliances to bond to a thermoplastic aligner. The Clear Aligner Adhesive is formulated in a viscosity that flows easily from the syringe and bonds chemically to a typically difficult to adhere to surface. a thermoplastic aligner now common in the dental industry. This adhesive, upon polymerization, is clear and virtually indistinguishable from the thermoplastic aligner thereby making it aesthetically pleasing to the patient. The light cure property of the adhesive allows the user to determine the polymerization time required by simply exposing the adhesive to an LED light source until set. The Clear Aligner Adhesive will be marketed in a push syringe for dispensing.

    AI/ML Overview

    The provided text describes a medical device called "Clear Aligner Adhesive" and its 510(k) submission for premarket notification. The information pertains to the adhesive's composition, intended use, and a comparison to a predicate device. However, the document does not contain information about acceptance criteria or specific studies that would typically be described for a diagnostic AI/ML device approval (e.g., sensitivity, specificity, clinical endpoints, reader studies).

    Therefore, I cannot provide a detailed answer to your request in the format specified. The information about sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set sizes, and ground truth establishment for a diagnostic AI/ML device is not present in the provided text.

    The document primarily focuses on establishing substantial equivalence for a physical medical adhesive based on its technical characteristics and an assessment of shear bond strength and cytotoxicity.

    Here's what can be extracted from the provided text about the device:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly list quantitative "acceptance criteria" for regulatory approval in the typical sense for an AI/ML device. Instead, it compares the new device's properties to a predicate device and reports on specific tests.

    PropertyAcceptance Criteria (Not explicitly stated as such for regulatory approval, but implied for comparison)Reported Device Performance
    Intended UseOrthodontic Bonding Adhesive for Thermoplastic Aligners (similar to predicate)Orthodontic Bonding Adhesive for brackets and appliances to thermoplastic aligner surfaces.
    Method of CureLight Cure (similar to predicate)Light Cure
    Aesthetic FeaturesClear when polymerized (improvement over predicate's A1 shade)Clear when polymerized
    Physical Properties / BondChemical bond to thermoplastic aligner and metal, composite or porcelain (similar to predicate)Chemical bond to thermoplastic aligner and metal, composite or porcelain.
    Shear Bond Strength(Implicitly: demonstrate adequate or improved bond strength compared to predicate)Testing resulted in highly increased bond strength values for the Clear Aligner Adhesive when compared to the use of Flowtain with a Plastic conditioner (predicate). (Specific quantitative values are not provided in this summary).
    Biocompatibility / Cytotoxicity (ISO 10993-5)No evidence of causing cell lysis or toxicityTested via the ISO 10993-5 elution test method and showed no evidence of causing cell lysis or toxicity.

    Regarding the study proving the device meets criteria, the document states:

    "Clear Aligner Adhesive was tested and compared to Reliance Orthodontic Products, Inc. Flowtain in conjunction with Plastic Conditioner for Shear Bond Strength. Testing resulted in highly increased bond strength values for the Clear Aligner adhesive when compared to the use of Flowtain with a Plastic conditioner. Clear Aligner Adhesive has been tested for cytotoxicity via the ISO 10993-5 elution test method and showed no evidence of causing cell lysis or toxicity."

    The following information cannot be provided from the given text:

    • 2. Sample size used for the test set and the data provenance: Not mentioned.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/mentioned for this type of device and submission.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/mentioned.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For Shear Bond Strength, the "ground truth" would be the measured force. For cytotoxicity, it would be the cellular response observed in the lab. These are not clinical "ground truths" in the diagnostic sense.
    • 8. The sample size for the training set: Not applicable for this type of device.
    • 9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111814

    Validate with FDA (Live)

    Device Name
    PADLOCK
    Date Cleared
    2011-09-14

    (79 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Light Bond is intended for use as an orthodontic bracket adhesive.

    Pad Lock® is intended for use as a fluorescing, light cure Orthodontic Bracket Adhesive.

    Device Description

    Light Bond and Pad Lock are light cure, orthodontic bracket adhesives available in a variety of viscosities for bonding to metal, ceramic and composite surfaces. Both Light Bond and Pad Lock are available in fluoride and non-fluoride formulas and come in syringe style or tip dispensing.

    In addition, Pad Lock fluoresces to ease clean-up of flash for the user.

    AI/ML Overview

    The provided document is a 510(k) summary for orthodontic bracket adhesives (Light Bond™ and Pad Lock®). It is not a study report for an AI-powered device. Therefore, I cannot extract information about acceptance criteria or a study proving that an AI device meets acceptance criteria, as an AI device is not described in this document.

    The document discusses the substantial equivalence of the Light Bond™ and Pad Lock® products to a predicate device (QUICK CURE, K001048), focusing on their intended use and mechanical/physical properties as orthodontic bracket adhesives.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101951

    Validate with FDA (Live)

    Date Cleared
    2010-10-26

    (106 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    L.C.R." is a flowable, light-cure orthodontic adhesive intended to be used within an orthodontic, dental or pediatric dental office for the bonding of lingual retainers, creation of occlusal buildups and for the retention of thermo-plastic aligners.

    Device Description

    L.C.R."" is a light-cure orthodontic adhesive that is flowable and a highly filled resin in order to provide durability. These properties make it ideal for bonding lingual retainers, creating occlusal buildups and for the retention of a thermoplastic aligner.

    L.C.R is available in push syringe, luer-lok syringe or tips for preferences in dispensing.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called L.C.R.™, a light-cure orthodontic adhesive. The primary purpose of this submission is to demonstrate substantial equivalence to a legally marketed predicate device, Flowtain™.

    Here's an analysis of the acceptance criteria and the study information provided, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    PropertyAcceptance Criteria (from Predicate Device)L.C.R.™ Reported Performance
    Intended UseLight Cure adhesive for bonding lingual retainers, and retention of thermoplastic alignersLight Cure adhesive for bonding lingual retainers, creation of occlusal buildups and retention of thermoplastic aligners
    Mechanical / Physical PropertiesFlowable Composite Light CureFlowable Composite Light Cure
    StorageRoom TemperatureRoom Temperature
    Shelf Life2 years2 years
    DeliverySyringeSyringe and Tips
    Flexural StrengthPerformance consistent with ISO 4049:2009 requirementsPerformance consistent with ISO 4049:2009 requirements (tested via ISO 4049:2009(E))
    Depth of CurePerformance consistent with ISO 4049:2009 requirementsPerformance consistent with ISO 4049:2009 requirements (tested via ISO 4049:2009(E))
    Bonding of Lingual RetainersSuccessful bond of wire to tooth withstanding multidirectional forceSuccessful bond of wire to tooth withstanding multidirectional force
    Bonding of Thermoplastic AlignersAdhesive created retentive surface for thermoplastic aligner.Adhesive created retentive surface for thermoplastic aligner.
    Compressive Strength (for occlusal buildups)(Implicitly, "effective" performance for the intended use)"Acceptable results were obtained"
    Oral Toxicity(Implicitly, no significant toxicity)"no significant evidence of toxicity"

    2. Sample Size Used for the Test Set and Data Provenance

    • Flexural Strength & Depth of Cure: The document states "L.C.R.™ was tested and compared to Flowtain™ for Flexural Strength and Depth of Cure via ISO 4049:2009(E) test method." While the standard ISO 4049:2009(E) specifies methodology, the exact sample size used for this specific study is not explicitly mentioned in the provided text.
    • Compressive Strength: "five replicates" were conducted for the Compressive Strength Test.
    • Oral Toxicity: "a 10 Mouse, 7 Day Method (Solid)" was used.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of material property testing, these would typically be laboratory-based prospective studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of experts being used to establish a "ground truth" in the traditional sense of clinical outcome assessment for the performance tests (Flexural Strength, Depth of Cure, Compressive Strength, Oral Toxicity). These are objective, quantitative laboratory measurements.

    For the "Bonding of Lingual Retainers" and "Bonding of Thermoplastic Aligners" properties, the descriptions ("Successful bond," "Adhesive created retentive surface") suggest qualitative assessments, but no information about experts or their qualifications is provided.

    4. Adjudication Method for the Test Set

    Not applicable. The reported tests are primarily objective laboratory measurements rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an orthodontic adhesive, not an AI software or imaging interpretation tool that would involve human readers or MRMC studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a physical medical device (an adhesive), not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance characteristics (Flexural Strength, Depth of Cure, Compressive Strength) is based on objective, quantitative measurements obtained through standardized laboratory test methods (e.g., ISO 4049:2009(E)). For oral toxicity, the ground truth is established by the observed biological response in the mouse model. For bonding efficacy, the "successful bond" or "retentive surface" acts as a qualitative ground truth.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical product, not a machine learning model, so there is no training set in the AI sense. The development of the adhesive likely involved internal R&D tests, but these are not described as a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083051

    Validate with FDA (Live)

    Date Cleared
    2009-02-20

    (129 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Flowtain™ is a flowable, light-cure orthodontic adhesive intended to be used within an orthodontic, dental or pediatric dental office for the bonding of lingual retainers and splinting materials, creation of archwire stops and indirect bonding of orthodontic brackets. Flowtain™ can also be used for the retention of thermoplastic aligners.

    Flowtain™ LV is a lower viscosity, flowable light cure orthodontic adhesive intended to be used for indirect bonding of orthodontic brackets.

    Device Description

    Flowtain™ is a light-cure orthodontic adhesive with a low modulus of elasticity that is flowable. This low modulus of elasticity makes it ideal for bonding lingual retainers, splinting materials and for indirect bonding of brackets. Thixotropic design properties of Flowtain™ also make it highly polishable which allows it to be used for comfortable archwire stops and as an aid for retention of a thermoplastic aligner.

    Flowtain™ LV is a low viscosity version of the Flowtain™ product used primarily for indirect bonding of brackets.

    Both are available in push syringe or tips for preferences in dispensing.

    AI/ML Overview

    The provided text describes the 510(k) summary for "Flowtain™ and Flowtain™ LV," orthodontic adhesives. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than defining and proving acceptance criteria for a novel AI or software device. As such, many of the requested categories for AI/software device evaluation are not directly applicable or explicitly stated in the document.

    However, I can extract and infer information relevant to the device's performance evaluation as presented:


    Acceptance Criteria and Study for Flowtain™ and Flowtain™ LV

    The acceptance criteria for Flowtain™ and Flowtain™ LV relate to demonstrating comparable performance, specifically in shear bond strength, to the legally marketed predicate device, 3M Unitek Transbond™ Supreme LV. The study aimed to show similar performance between the new devices and the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Shear Bond StrengthSimilar performance to 3M Unitek Transbond™ Supreme LV.Testing "resulted in similar performance between the two adhesives."
    ToxicityNon-toxic."Flowtain™ has been tested and proven to be non-toxic."

    2. Sample Size and Data Provenance

    • Sample Size for Test Set: Replicates of 5 for Flowtain™, Flowtain™ LV, and Transbond™ Supreme LV. (Total of 15 samples for shear bond strength testing, assuming 5 replicates for each of the three products).
    • Data Provenance: Not explicitly stated, but the testing method ("in-vivo performance test method") suggests biological or simulated biological conditions. The manufacturer is based in Itasca, Illinois, USA, implying the study was likely conducted in the USA. The study design (comparison against a predicate device) is typical for a 510(k) submission. It's a retrospective comparison against a known predicate device's performance data or a prospective comparative study where both were tested simultaneously under the same conditions. Given the phrasing "tested against," it implies a comparative study.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable for this type of device. The assessment of shear bond strength is a quantitative mechanical test, not dependent on expert interpretation or ground truth established by human experts in the way an image analysis or diagnostic AI would require.

    4. Adjudication Method

    • Not applicable for this type of device. Adjudication methods like 2+1 or 3+1 are used for expert consensus on qualitative assessments (e.g., medical image interpretation). Shear bond strength is a physical measurement.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This type of study is relevant for AI systems that assist human readers in tasks like medical image interpretation. Flowtain™ and Flowtain™ LV are orthodontic adhesive materials, not AI systems.

    6. Standalone Performance Study (Algorithm Only)

    • Not applicable. This relates to AI algorithms. The performance of the adhesive materials was evaluated directly through mechanical testing.

    7. Type of Ground Truth Used

    • Mechanical Test Results / Physical Property Measurement: The "ground truth" for shear bond strength is the objective measurement obtained from a standardized mechanical test. For toxicity, it's based on toxicology testing.

    8. Sample Size for Training Set

    • Not applicable. This device is a physical product (orthodontic adhesive), not a machine learning model, so there is no "training set."

    9. How Ground Truth for Training Set was Established

    • Not applicable. As there is no training set for an AI model, this question is irrelevant to the described device.

    Summary of the Study:

    The study detailed in the 510(k) summary for Flowtain™ and Flowtain™ LV involved in-vivo performance testing to assess shear bond strength. This testing was conducted using an indirect bonding procedure for replicates of 5 for Flowtain™, Flowtain™ LV, and the predicate device, 3M Unitek Transbond™ Supreme LV. The results indicated that the new adhesives achieved "similar performance" to the predicate device, thereby meeting the implied acceptance criterion for substantial equivalence in this critical mechanical property. Additionally, Flowtain™ was explicitly tested and proven to be non-toxic.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073243

    Validate with FDA (Live)

    Device Name
    LED PRO SEAL
    Date Cleared
    2008-02-08

    (81 days)

    Product Code
    Regulation Number
    872.3310
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LED PRO SEAL® is intended to seal enamel surfaces around orthodontic brackets.

    Device Description

    LED PRO SEAL® is a fluoride containing light cure filled surface sealant provided in a light impervious bottle dispenser. Upon delivery of one drop of LED PRO SEAL to a mixing pad, a thin coat of LED PRO SEAL is applied to a prepared enamel surface using a brush applicator and light cured for a minimum of 20 seconds with any halogen, plasma arc or cordless light capable of a 440-490 NM wavelength.

    AI/ML Overview

    The provided document describes the 510(k) summary for the LED PRO SEAL® dental sealant. This submission seeks to demonstrate substantial equivalence to a predicate device, not through an AI/ML study, but through traditional laboratory testing of material properties. Therefore, many of the requested fields related to AI/ML specific studies will be marked as "Not Applicable".

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    PropertyAcceptance Criteria (Predicate Device Performance - K021503)Reported Device Performance (LED PRO SEAL®)
    Mechanical/Physical:
    Shear bond strengthEquivalent to original PRO SEAL®Equivalent to original PRO SEAL®
    Compressive strengthEquivalent to original PRO SEAL®Equivalent to original PRO SEAL®
    Flexural strengthEquivalent to original PRO SEAL®Equivalent to original PRO SEAL®
    Cumulative fluoride releaseEquivalent to original PRO SEAL®Equivalent to original PRO SEAL®
    Depth of cureEquivalent to original PRO SEAL®Equivalent to original PRO SEAL®
    Biocompatibility:
    Cytotoxicity (ISO 10993-5)Grade 2 material (mild reactivity)Grade 2 material (mild cell lysis or toxicity)

    Study Proving Device Meets Acceptance Criteria:

    A series of comparative tests were conducted between LED PRO SEAL® and the predicate device, original PRO SEAL® (K021503). These tests included:

    • Shear bond strength
    • Compressive strength
    • Flexural strength
    • Cumulative fluoride release
    • Depth of cure

    The document states that these comparisons "demonstrate that the applicant device is equivalent to the legally marketed device in terms of effectiveness."

    Additionally, for biocompatibility, a Cytotoxicity Study using the ISO 10993-5 Agarose Overlay Method (Solid) was performed. The LED PRO SEAL® was found to cause "mild cell lysis or toxicity (generating grade 2 for mild reactivity)" and "met the requirements of ISO10993-5 for a grade 2 material showing evidence that the product is safe to use."

    2. Sample size used for the test set and the data provenance

    • Sample size: Not explicitly stated in the provided text for the material property tests.
    • Data provenance: The tests were conducted internally by Reliance Orthodontic Products, Inc., comparing their new device to their own predicate device. The country of origin of the data is implied to be USA given the company's location. The study is prospective in nature as it tests the newly developed device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable: This type of medical device submission (dental sealant) relies on objective laboratory measurements of material properties and biological tests, not expert interpretation of images or clinical outcomes that would require a ground truth established by human experts in the context of an AI/ML study.

    4. Adjudication method for the test set

    • Not Applicable: See explanation for item 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is not an AI/ML device requiring human reader assessment. It's a dental material.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is not an AI/ML device.

    7. The type of ground truth used

    • Not Applicable: For material properties, the "ground truth" is established by the objective measurements obtained from standardized testing methods (e.g., measuring force for bond strength, chemical analysis for fluoride release, cell viability for cytotoxicity). For the purpose of regulatory submission, the performance of the predicate device acts as the benchmark ("ground truth" for comparison).

    8. The sample size for the training set

    • Not Applicable: This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable: This is not an AI/ML device, so there is no training set or associated ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071688

    Validate with FDA (Live)

    Device Name
    RELIANCE S.E.P.
    Date Cleared
    2007-09-07

    (79 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reliance S.E.P. is a one step self etching primer intended to be used for the preparation of a tooth surface prior to bonding an orthodontic appliance(s) with a light cure adhesive.

    Device Description

    The Reliance S.E.P. is a one step self etching primer contained in a handheld dispenser. Upon manual compression into a light impervious mixing well, enough solution is dispensed to prepare the enamel surface of up to 20 teeth.

    AI/ML Overview

    The provided document describes the Reliance S.E.P. (Self Etching Primer) and its substantial equivalence to a predicate device, 3M Unitek Transbond Plus Self Etching Primer. The primary study conducted to demonstrate this equivalence is a "Shear bond strength testing comparison."

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Equivalence in Effectiveness (Shear Bond Strength)The Reliance S.E.P. demonstrates equivalence to the legally marketed predicate device (3M Unitek Transbond™ Plus Self Etching Primer) in terms of effectiveness, as shown by shear bond strength testing comparisons.
    Non-toxicityThe device was found to be non-toxic based on an Oral Toxicity Test method.

    Note: The document states that the acceptance criterion for effectiveness is equivalence to the predicate device's effective bond strength, but it does not provide specific numerical thresholds or performance targets for either device's bond strength. It simply states that the Reliance S.E.P. "demonstrate[s] that the applicant device is equivalent to the legally marketed device in terms of effectiveness."


    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the sample size used for the shear bond strength testing.
      • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not applicable to the type of study conducted. The ground truth for this performance study (shear bond strength testing) would be objective measurements of bond strength, not expert consensus.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Adjudication methods like 2+1 or 3+1 are typically used for studies involving subjective human assessment (e.g., image interpretation). This is a physical/mechanical performance study, so such adjudication is not applicable. The measurements would be objective.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a self-etching primer, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This question is not applicable. The device is a physical product (self-etching primer), not an algorithm. The "standalone" performance here refers to the device's inherent mechanical properties.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for effectiveness was based on objective mechanical/physical measurements of shear bond strength. For safety, it was an Oral Toxicity Test method.
    7. The sample size for the training set:

      • This information is not applicable. This is a study of a physical medical device, not a machine learning algorithm that requires a "training set."
    8. How the ground truth for the training set was established:

      • This information is not applicable, as there is no "training set" in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071590

    Validate with FDA (Live)

    Device Name
    THERMA-CURE
    Date Cleared
    2007-07-31

    (50 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Therma-Cure is a heat cure orthodontic adhesive intended to be used within an orthodontic, dental or pediatric dental office for the indirect bonding of orthodontic brackets and appliances.

    Device Description

    Therma-Cure™ is a heat curable orthodontic adhesive used for the indirect bonding of orthodontic brackets and appliances.

    AI/ML Overview

    The provided text describes the 510(k) summary for the device "Therma-Cure™", a heat-curable orthodontic adhesive, and its comparison to a predicate device, "Light Bond™ Orthodontic Adhesive". The study focused on demonstrating substantial equivalence, primarily through performance testing.

    Here's an analysis of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    PropertyAcceptance Criteria (Implied by Predicate Performance)Reported Device Performance (Therma-Cure)
    Intended UseLight Cure Adhesive for indirect bonding (as per predicate device, Light Bond)Heat Cure Adhesive for indirect bonding
    Mechanical / Physical PropertiesSingle paste in a hand-held delivery (as per predicate device, Light Bond)Single paste in a hand-held delivery
    Chemical CompositionBisphenol A Diglycidyl Methacrylate / Urethane Dimethacrylate Resin Monomer and Silica Filler Triethyleneglycol Dimethacrylate Diluent 2-3-Di-tert-butyl-4-methylphenol Inhibitor (as per predicate device, Light Bond)Bisphenol A Diglycidyl Methacrylate resin Monomer and Silica Filler Triethyleneglycol Dimethacrylate Diluent 2-3-Di-tert-butyl-4-methylphenol Inhibitor
    Shear Bond StrengthNot explicitly stated as a numerical acceptance criterion, but the implied criterion is similar performance to the predicate device, Light Bond, when tested using an in-vivo performance test method. This means the shear bond strength should be within an acceptable range of the predicate, indicating comparable adhesive strength for orthodontic bracket retention.The study reported "similar performance" to the predicate device for shear bond strength using an in-vivo performance test method. No specific numerical values were provided.
    Non-toxicityThe device must be non-toxic, conforming to applicable biocompatibility standards. The implied acceptance criterion would be passing the requirements of ISO 10993-5 (tests for in vitro cytotoxicity).Therma-Cure has been proven to be non-toxic when tested according to ISO 10993-5.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: "replicates of 5 for both Therma-Cure and Light Bond" were used for the shear bond strength testing. This means 5 samples for the Therma-Cure device and 5 samples for the Light Bond device, for a total of 10 samples in the test.
      • Data Provenance: The study states "in-vivo performance test method". While it doesn't explicitly state the country of origin, the FDA submission context implies the data would be relevant to U.S. regulatory standards. It is a prospective study as it involves conducting new tests for regulatory submission.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • This information is not provided in the given text. The study describes "in-vivo performance test method" for shear bond strength, which is a laboratory or clinical measurement rather than an expert interpretation task. For the non-toxicity test, reference to ISO 10993-5 implies standard toxicology protocols were followed, not expert consensus on ground truth.
    3. Adjudication Method for the Test Set:

      • Not applicable. The tests described are objective performance measurements (shear bond strength, non-toxicity) rather than subjective assessments requiring adjudication among experts.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This was not an MRMC study. The device is an orthodontic adhesive, not an AI-powered diagnostic tool. Therefore, human reader improvement with or without AI assistance is not relevant to this submission.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device is a physical product (adhesive), not an algorithm or AI system.
    6. The Type of Ground Truth Used:

      • For Shear Bond Strength: The ground truth was established through a direct, objective measurement using an "in-vivo performance test method." The predicate device's performance served as the comparative "ground truth" for demonstrating substantial equivalence.
      • For Non-Toxicity: The ground truth was established by adherence to the objective test standards outlined in ISO 10993-5.
    7. The Sample Size for the Training Set:

      • Not applicable. The device is a physical product, not a machine learning model, so there is no concept of a "training set" in this context.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable. As there is no training set for a physical adhesive, this question does not apply.
    Ask a Question

    Ask a specific question about this device

    K Number
    K033551

    Validate with FDA (Live)

    Device Name
    EVOLVE
    Date Cleared
    2003-12-16

    (36 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fluoride-releasing, light-cured bracket bonding system in the oral cavity of dental patients.

    Device Description

    EVOLVE is a fluoride-releasing, light-cured, color changing adhesive for bonding any type of orthodontic bracket to any enamel, porcelain, composite or metal surfaces when properly conditioned. It is a highly filled, smooth, tacky paste that helps prevent bracket flotation and delivers maximum bond strength on curing. Its contrasting blue color facilitates the ease of placement and clean-up once cured. The material is supplied in push syringes or syringe tips.

    AI/ML Overview

    The provided text describes a 510(k) submission for a dental orthodontic adhesive called EVOLVE and does not contain information about acceptance criteria or a study proving that the device meets such criteria.

    The document is a premarket notification for a Class II dental device, asserting its substantial equivalence to a predicate device ("Light Bond with Fluoride"). The submission focuses on comparing the technological characteristics and intended use of EVOLVE with the predicate device to demonstrate similar safety and effectiveness.

    Therefore, I cannot provide the requested information. The text does not include:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes or data provenance for test sets.
    • The number or qualifications of experts for ground truth establishment.
    • Adjudication methods.
    • Information on a multi-reader multi-case (MRMC) comparative effectiveness study or related effect sizes.
    • Results from a standalone algorithm performance study.
    • The type of ground truth used (beyond general statements about biocompatibility and physical properties).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    The study presented is essentially a comparison of the new device (EVOLVE) with a previously cleared predicate device (Light Bond with Fluoride) based on:
    * Intended Use: Both are "Dental Bracket Adhesive."
    * Chemical Composition: Both are "Light cured, fluoride-releasing, glass filled, Dimethacrylate-based composite."
    * Mechanical/Physical Properties: Both are "Highly viscous, smooth, tacky paste. High bond strength."
    * Biocompatibility: EVOLVE was tested and found to be "non-toxic."

    The conclusion drawn by the applicant is that EVOLVE is as safe and effective as the predicate device because the proposed changes do not affect safety and efficacy. This is the basis for the FDA's "substantial equivalence" determination, not a study against specific acceptance criteria for performance metrics as would be seen for diagnostic AI/ML devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K001048

    Validate with FDA (Live)

    Date Cleared
    2000-04-27

    (35 days)

    Product Code
    Regulation Number
    872.3750
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1