Found 2 results

510(k) Data Aggregation

    K Number
    K131643

    Validate with FDA (Live)

    Date Cleared
    2014-12-17

    (560 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Durex® RealFeel™ Pleasure Gel Personal Lubricant is intended for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease of comfort of intimate sexual activity and supplement the body's natural lubrication. This lubricant is compatible with polyurethane condoms. It is not compatible with natural rubber latex and polyisoprene condoms.

    Device Description

    Durex® RealFeel™ Pleasure Gel is a non-sterile, personal silicone base lubricant. The lubricant is comprised of cyclopentasiloxane and dimethiconol. Durex® RealFeel™ Pleasure Gel contains a blend of silicone fluid ingredients similar to ingredients found in the predicate device The specifications for Durex® RealFeel™ Gel include appearance, odor, viscosity, specific gravity, total aerobic microbial count, total yeast and mold count, and absence of pathogenic organisms (Pseudomonas aeroginosa, Staphylococcus aureus, and Candida albicans).

    AI/ML Overview

    This document describes the review of Durex® RealFeel™ Pleasure Gel Personal Lubricant for 510(k) premarket notification. The review focuses on establishing substantial equivalence to a legally marketed predicate device, ONE® Silicone Lubricant (K110690).

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance
    Intended UseSame as predicate device (moisturize and lubricate for intimate sexual activity, supplement natural lubrication). Condom compatibility is the key differentiating factor.Durex® RealFeel™ Pleasure Gel is intended for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease of comfort of intimate sexual activity and supplement the body's natural lubrication.
    Material CompositionSimilar to predicate device (silicone-based).Comprised of cyclopentasiloxane and dimethiconol; a blend of silicone fluid ingredients similar to ingredients found in the predicate device.
    Physical SpecificationsNot explicitly stated, but assumed to be acceptable (e.g., appearance, odor, viscosity, specific gravity)."The specifications... include appearance, odor, viscosity, specific gravity..." (Specific values not provided in this document but implied to meet internal standards).
    Microbiological PurityAbsence of pathogenic organisms; acceptable total aerobic microbial count, total yeast and mold count."total aerobic microbial count, total yeast and mold count, and absence of pathogenic organisms (Pseudomonas aeroginosa, Staphylococcus aureus, and Candida albicans)." (Specific limits not provided but implied to meet internal standards).
    BiocompatibilityPass relevant ISO 10993-1 standards for cytotoxicity, systemic toxicity, sensitization, and irritation."The results demonstrate that the subject device is biocompatible." (Based on four studies: cytotoxicity, systemic toxicity in mice, maximization in guinea pigs, and vaginal irritation in rabbits).
    Condom CompatibilityMust be compatible with polyurethane condoms.Compatible with polyurethane condoms. Not compatible with natural rubber latex and polyisoprene condoms (This is a differentiating factor from the predicate but is acceptable as it's clearly stated in the indications for use).
    Stability/Shelf-LifeMaintains specifications over stated shelf life."The results of accelerated and real time studies demonstrate that the subject lubricant maintains its specifications over the duration of its shelf life."

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not specify the exact sample sizes for each of the performance tests (biocompatibility, condom compatibility, stability). However, it states that:

    • Biocompatibility Studies: Four studies were conducted. The specific number of animals/samples for each endpoint (cytotoxicity, systemic toxicity in mice, maximization in guinea pigs, and vaginal irritation in rabbits) is not detailed.
    • Condom Compatibility Testing: The sample size for this testing is not provided.
    • Stability Studies: The sample size for the accelerated and real-time studies is not provided.

    The data provenance is not explicitly stated in terms of country of origin but is implied to be from studies conducted by or for Reckitt Benckiser, LLC. These studies are retrospective to the submission date of November 18, 2014.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation is for a personal lubricant, not a diagnostic device requiring expert interpretation of medical images or data for ground truth establishment. The ground truth (or reference standard) for this device's performance is based on standardized testing methods and laboratory measurements, not expert consensus on medical conditions.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated above. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies where expert consensus is needed to establish a ground truth for diagnostic accuracy, which is not the case here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    This information is not applicable. This is a submission for a personal lubricant and does not involve AI, human readers, or diagnostic interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. This is not an algorithmic or AI-based device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Standardized Laboratory Testing and Measurement:
      • Biocompatibility: Adherence to GLP requirements and applicable ISO 10993-1 standards for endpoints like cytotoxicity, systemic toxicity, sensitization, and irritation.
      • Condom Compatibility: Testing conducted in accordance with ASTM D7661-2010.
      • Physical and Microbiological Specifications: Internal quality control standards and established methods for appearance, odor, viscosity, specific gravity, microbial counts, and absence of specific pathogens.
      • Stability: Laboratory assessment of product specifications over time under accelerated and real-time conditions.

    8. The Sample Size for the Training Set

    This information is not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As this is not an AI/ML device, there is no training set or associated ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140193

    Validate with FDA (Live)

    Date Cleared
    2014-11-04

    (281 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Durex Embrace Pleasure Gels, which includes the Warming Sensation Gel and Tingling Experience Gel, are intended for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubricants are compatible with natural rubber latex, synthetic polyisoprene and polyurethane condoms.

    Device Description

    Durex Embrace Pleasure Gels are clear odorless slightly viscous liquids that are packaged together and consist of two different formulation is water-based with multisensate (Tingling Experience Gel) and the female formulation (Warming Sensation Gel) is Vanillyl Butyl Ether in Glycerine base. The female lubricant (Pink Bottle) offers a warming experience, and one for the male partner (Purple bottle) offers a tingling experience. Devices are supplied in 60 ml PETG pump bottles. Both lubricant bottles are covered with a shrink sleeve including graphics that also provides tamper evident protection. Bottles are then placed in clear trays, which are then shrink-wrapped and placed in printed cartons.

    AI/ML Overview

    The provided text describes the submission for a 510(k) premarket notification for Durex Embrace Pleasure Gels, seeking substantial equivalence to a predicate device (KY Yours + Mine). It outlines performance testing, but it is not a study that proves the device meets specific acceptance criteria in the context of an AI/ML device.

    The document is for a personal lubricant, not an AI/ML device, and therefore does not have acceptance criteria or studies related to AI/ML performance metrics (like sensitivity, specificity, F1-score, AUC, etc.).

    However, I can extract the relevant information regarding the device's acceptance criteria and the studies performed to demonstrate its safety and performance based on the provided document.

    Here's an interpretation of the request in the context of this non-AI medical device submission:

    1. A table of acceptance criteria and the reported device performance

    The document lists "Specifications" for which "acceptance specifications" have been developed, but it does not explicitly state the numerical acceptance criteria or the specific reported device performance values for each specification. It only states that "testing on samples collected up to 12 months of aging... showed that devices met their acceptance criteria" and that test results "demonstrated device materials to be biocompatible" and "non-irritating" or "did not elicit sensitization reactions."

    SpecificationAcceptance Criteria (Not explicitly stated numerically in this document)Reported Device Performance (Summary)
    Appearance(Implied: clear, odorless, slightly viscous liquids)"clear odorless slightly viscous liquids" (Description of Device section)
    Odor(Implied: odorless)"clear odorless slightly viscous liquids" (Description of Device section)
    pH @ 25 degree C(Implied: within a specified range)Not detailed, but document states Durex Embrace Pleasure Gels are "substantially equivalent to the predicate device KY Yours + Mine in terms of technological characteristics such as... pH" and "met their acceptance criteria" during stability testing.
    Total Aerobic Microbial Count(Implied: Below a specified threshold)Not detailed, but implied to have met acceptance criteria during stability testing.
    Total Yeast and Mold Count(Implied: Below a specified threshold)Not detailed, but implied to have met acceptance criteria during stability testing.
    Lack of Pathogenic Organisms (Pseudomonas aeruginosa, Staphylococcus aureus, Candida albicans)(Implied: Absence of these organisms)Not detailed, but implied to have met acceptance criteria during stability testing.
    Preservative Effectiveness(Implied: Meets specified criteria, e.g., USP <51>)Not detailed, but implied to have met acceptance criteria during stability testing.
    Viscosity-RV4, 10rpm at 25 degree C(Implied: Within a specified range)Not detailed, but implied to have met acceptance criteria during stability testing.
    Osmolality(Implied: Within a specified range, e.g., WHO guidelines for lubricants)Not detailed, but implied to have met acceptance criteria during stability testing.
    Biocompatibility (Acute Systemic Toxicity)No biological reactivity at tested time points for test group animals.All test group animals survived the test period and none exhibited any biological reactivity (for Tingling Experience Gel, Warming Sensation Gel, and 1:1 mixture) at any tested time points.
    Biocompatibility (Vaginal Irritation)Considered non-irritating.Durex Embrace Pleasure Gel, Warming Sensation Gel, is considered non-irritating.
    Biocompatibility (Penile Irritation)Considered non-irritating.Durex Embrace Pleasure Gel, Tingling Experience Gel, is considered non-irritating.
    Biocompatibility (Cytotoxicity)Non-cytotoxic, or if cytotoxic, mitigated by other in vivo biocompatibility results.Cytotoxicity results for Warming Sensation Gel (alone and in combination) may have the potential to produce a biological response. However, other in vivo biocompatibility results were non-irritant, and the overall profile does not indicate a significant risk.
    Biocompatibility (Sensitization)Did not elicit sensitization reactions.Durex Embrace Pleasure Gels did not elicit sensitization reactions in study animals (for Tingling Experience Gel, Warming Sensation Gel, and 1:1 mixture).
    Condom Compatibility (Natural Rubber Latex, Synthetic Polyisoprene, and Polyurethane Condoms)Does not affect the mechanical and physical integrity of the condoms as per ASTM D7661-10.Both formulations demonstrated they do not affect the mechanical and physical integrity of natural rubber latex, synthetic polyisoprene and polyurethane condoms.
    Stability/Shelf Life (up to 12 months)Maintain all specifications/acceptance criteria (Appearance, Odor, pH, Microbial Counts, Pathogens, Preservative Effectiveness, Viscosity, Osmolality) for the target shelf life.Testing on samples collected up to 12 months of aging (real-time and accelerated) showed that devices met their acceptance criteria.

    Regarding AI/ML specific information, this document does not contain any of the following because it is not an AI/ML device:

    • Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    • Adjudication method for the test set
    • Multi reader multi case (MRMC) comparative effectiveness study
    • Standalone (i.e. algorithm only without human-in-the-loop performance) study
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for test set
    • Sample size for the training set
    • How the ground truth for the training set was established

    Detailed information from the document (non-AI related):

    2. Sample sizes used for the test set and data provenance:

    • Acute Systemic Toxicity: "test group animals" (number not specified). Data provenance: Not specified, but likely conducted in a laboratory setting as per GLP requirements.
    • Vaginal Irritation Testing: Not specified, but involved "rabbits" and assessed the Warming Sensation Gel.
    • Penile Irritation Study: Not specified, but involved "rabbits" and assessed the Tingling Experience Gel.
    • Cytotoxicity: Not specified, but tested using "tissue culture agar diffusion test."
    • Sensitization: "study animals" (number not specified). Involved "Guinea Pig Maximization test."
    • Condom Compatibility: Not specified for number of condoms or samples. Testing was "conducted in accordance with ASTM D7661-10."
    • Stability: "samples collected up to 12 months of aging (real-time and accelerated at 40 C-75%RH)" (number of samples not specified).
    • Data Provenance: Not explicitly stated, but all studies were conducted in accordance with GLP requirements and applicable ISO 10993 standards or ASTM D7661-10, implying controlled laboratory conditions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a physical/chemical/biological performance study, not an AI/ML ground truth establishment. The "ground truth" is determined by standardized test methods and observations/measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for this type of performance study. Results are based on direct observation, measurement, or established biological outcomes according to standard protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable as this is not an AI/ML device. The performance tests are for the product itself (lubricant).

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • The "ground truth" for these studies is based on observable biological reactions (e.g., irritation, sensitization, systemic toxicity), physical/chemical properties (e.g., pH, viscosity), and mechanical integrity (condom compatibility), as defined by established international standards (ISO 10993, ASTM D7661).

    8. The sample size for the training set:

    • Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable as this is not an AI/ML device.