    Intended Use

    The RS 3000 Shielded Cabinet X-ray Radiation Source is intended for the irradiation of blood or blood products packaged in transfusion bags in accordance with "Recommendations Regarding License Amendments and Procedures for Gamma Irradiation of Blood Products" (22 July 1993 memorandum from Acting Director, Office of Blood Research and Review, Center for Biologics Evaluation and Research, FDA to all registered blood establishments) when irradiation to reduce the risk of Graft Versus Host Disease is indicated.

    Device Description

    The RS 3000 Shielded Cabinet X-ray Radiation Source consists of A shielded enclosure containing 2 vertically opposed x-ray tubes with provision for a sample holder (canister) between them, a power supply and a controller.

    AI/ML Overview

    This document is a 510(k) premarket notification for the RS 3000 Shielded Cabinet X-ray Radiation Source, a blood irradiator. It focuses on demonstrating substantial equivalence to a predicate device (IBL 437C Blood Irradiator), rather than presenting a study with acceptance criteria and device performance in the typical sense of a diagnostic or therapeutic device that would involve a clinical study with human subjects.

    Therefore, many of the requested categories for a clinical study are not applicable to this type of submission. However, I can extract information related to the device's technical specifications and how they compare to the predicate, which serves as the "performance" data in this context for demonstrating substantial equivalence.

    Here's the breakdown of the information provided in the input, tailored to what's available for this specific type of device submission:

    1. Table of "Acceptance Criteria" (Predicate Device Specifications) and Reported Device Performance (RS 3000)

    In the context of a 510(k) for a blood irradiator, "acceptance criteria" are effectively the specifications and performance characteristics of the predicate device, which the new device (RS 3000) aims to meet or be substantially equivalent to.

    Criteria (Predicate: IBL 437C)RS 3000 Performance
    Source:160 kVdc x-rays, .38 mm Cu filter, hvl app. 4 cm H₂0
    Cs-137, 662 keV
    Dose rate:3 Gy min⁻¹
    > 4 Gy min⁻¹
    Max/min dose ratio:< 1.3
    < 1.67
    Sample holder:fixed, presents maximum width, minimum depth
    rotates in beam
    Radiation safety:Pb shielding, interlocks
    Pb shielding, interlocks
    Federal Regulatory Environment:Requires 510(k). Must comply with 21 CFR 1020.40
    Requires 510(k), user must have NRC license and radiation safety officer, operators must be film-badged.

    2. Sample size used for the test set and the data provenance: Not applicable. This is a technical comparison for a blood irradiator, not a diagnostic or therapeutic device evaluated on patient data. The "test set" here refers to the device's physical and functional characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications: Not applicable. "Ground truth" in this context is established by the known specifications and performance of both the RS 3000 and the predicate IBL 437C, which are typically derived from engineering specifications, calibration, and testing.

    4. Adjudication method for the test set: Not applicable. There's no "adjudication" in the sense of expert review of clinical cases. The comparison is based on objective, measurable technical specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a blood irradiator, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This phrase usually applies to AI algorithms. The RS 3000 is a physical device that performs irradiation; its "performance" is its ability to deliver the specified radiation dose safely and effectively, which is likely validated through physical measurements and calibration.

    7. The type of ground truth used:
    The "ground truth" for this submission are the established technical specifications and performance data of both the new device (RS 3000) and the predicate device (IBL 437C), verified through physical testing, engineering design, and adherence to relevant standards (e.g., radiation safety regulations). The primary "ground truth" for demonstrating substantial equivalence rests on objective, measurable characteristics rather than clinical outcomes on human subjects for this type of device. The FDA's "Recommendations Regarding License Amendments and Procedures for Gamma Irradiation of Blood Products" (22 July 1993 memorandum) also serves as a regulatory "ground truth" for the intended use and irradiation parameters.

    8. The sample size for the training set: Not applicable. This device is not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established: Not applicable. As above, this phrase refers to AI algorithms, which this device is not.