Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K220106

    Validate with FDA (Live)

    Device Name
    QardioArm 2
    Manufacturer
    Date Cleared
    2022-06-15

    (154 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QardioArm 2 is a fully automatic, wireless blood pressure monitor intended to measure the systolic and diastolic blood pressures and pulse rate of adult individuals. It utilizes an inflatable cuff that is wrapped around the upper arm. The arm circumference is limited to 8.7 - 18.5 inches (22 - 47 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. It is not intended to be used as a diagnostic device and users are instructed to contact their physician if hypertensive values are indicated.

    Device Description

    The QardioArm 2 is an over-the-counter (OTC) non-invasive blood pressure monitor intended for spot-checking in adults. It uses a detachable inflatable cuff that is wrapped around the upper arm to determine blood pressure based on the oscillometric method. Two cuff sizes are available: standard size for an arm circumference of 8.7 to 14.6 inches or large size for an arm circumference of 14.6 to 18.5 inches.

    AI/ML Overview

    The QardioArm 2 is a non-invasive blood pressure measurement system. The acceptance criteria and the study that proves the device meets them are detailed below.

    1. A table of acceptance criteria and the reported device performance

    Unfortunately, the provided text does not contain a specific table of acceptance criteria or reported device performance in the format of a direct comparison table. However, it states that the device "met all the requirements of the standard" and "passed all tests" for various criteria.

    Based on the general information provided, we can infer some key performance criteria and the reported performance:

    Acceptance Criteria CategorySpecific Criteria (Inferred from standards)Reported Device Performance
    Accuracy (Blood Pressure)As per ISO 81060-2:2018 (Non-invasive sphygmomanometers — Part 2: Clinical investigation of automated measurement type), which is a recognized standard for blood pressure monitor accuracy. This standard typically requires a mean difference between the device and reference standard of ≤ ±5 mmHg, and a standard deviation of differences ≤ 8 mmHg. (Implicit)"The device met all the requirements of the standard [ISO 81060-2:2018]." This implies the accuracy criteria of ISO 81060-2 were met. The technological comparison section explicitly states an accuracy of "+/-3mmHg or 2% reading for blood pressure."
    Accuracy (Pulse)As per ISO 81060-2:2018 or similar standards for pulse measurement. (Implicit)The technological comparison section explicitly states an accuracy of "+/-5% for pulse."
    Electrical SafetyCompliance with IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical electrical equipment: Part 1: General requirements for basic safety and essential performance."The device passed all tests, including the US deviations."
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests."The QardioArm 2 passed all tests."
    Battery SafetyCompliance with IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems."found to comply with all relevant sections."
    Wireless CoexistenceCompliance with AAMI TIR69:2017 and ANSI C63.27:2017."All tests passed."
    Software Verification & ValidationAdherence to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," for a "Moderate" level of concern."Software verification and validation testing were conducted, and documentation was provided as recommended... The software for this device was considered as a 'Moderate' level of concern." While not explicitly stating "passed," meeting the guidance for a moderate level of concern implies successful validation.
    Home Healthcare EnvironmentCompliance with IEC 60601-1-11:2015, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment."The device passed all tests."
    Automated Non-Invasive Sphygmomanometer PerformanceCompliance with IEC 80601-2-30:2018, Medical electrical equipment – Part 2: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers."The device passed all tests."
    Human FactorsAdherence to FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices," demonstrating the device is safe and effective for its intended users, uses, and use environments."The QardioArm 2 was found to be safe and effective for its intended users, uses, and use environments."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Clinical Data (Test Set): 86 subjects.
    • Data Provenance: The study was a "prospective, single arm paired comparison study." The document does not specify the country of origin of the data, but it was conducted to demonstrate performance in accordance with ISO 81060-2:2018 and FDA's requirements for 510(k) submissions, suggesting a standard clinical study environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The study description mentions a "same arm sequential method with dual auscultators." This implies that the ground truth for blood pressure measurements was established by two human observers (auscultators) using a reference sphygmomanometer. The document does not specify the number of experts beyond "dual auscultators" or their specific qualifications (e.g., years of experience, medical role). However, for such studies, these individuals are typically trained and validated healthcare professionals.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The method used was a "same arm sequential method with dual auscultators." While not explicitly stated as "2+1" or "3+1" (which are typically used for imaging assessments), the use of dual auscultators suggests a comparison or consensus method between the two observers for establishing reference measurements. Differences between the dual auscultators would typically be adjudicated or the reading discarded according to the ISO standard's protocol, though the specific adjudication method is not described in detail.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a non-invasive blood pressure monitor, not an AI-assisted diagnostic imaging device that involves human interpretation of images. The study focused on the device's accuracy compared to a reference standard, not on human reader performance with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in essence, a standalone performance assessment was done. The QardioArm 2 measures blood pressure automatically (oscillometric method) and provides a reading. The clinical study directly compared these device-generated readings against reference measurements obtained by human auscultators. The device operates autonomously to produce the blood pressure and pulse rate values.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for blood pressure measurements was established using a reference sphygmomanometer via a same arm sequential method with dual auscultators. This constitutes an expert consensus or expert-derived reference standard for physiological measurements, rather than pathology or outcomes data.

    8. The sample size for the training set

    The document does not provide information about a separate "training set" or its sample size. This type of device's core functionality (oscillometric measurement) is based on established principles and algorithms, often developed and refined internally. For a 510(k) submission of a blood pressure monitor, the primary focus is on the performance of the final device as compared to a reference standard in a clinical test set, rather than detailing the training data for an AI algorithm.

    9. How the ground truth for the training set was established

    As no specific training set data is mentioned for an AI algorithm, the method for establishing ground truth for a training set is not applicable or detailed in this document. The device's operation is based on physical measurement principles rather than a deep learning model requiring a large labeled training dataset in the same way an image analysis AI would.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201644

    Validate with FDA (Live)

    Device Name
    QardioCore
    Manufacturer
    Date Cleared
    2021-02-28

    (256 days)

    Product Code
    Regulation Number
    870.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QardioCore ECG ambulatory monitoring device is intended to capture, store, transmit, and display ECG information for recording periods of up to 24-hours in a single session. It is indicated for use on adult patients who may be asymptomatic or who meet clinical indications to perform an ECG-Holter monitor exam.

    The QardioCore ECG monitor is a prescription-only device, and the reported information is provided for review by a physician who will render a diagnosis based on clinical judgment and experience.

    Device Description

    The QardioCore ECG ambulatory monitoring device is intended to capture, store, transmit, and display ECG information for recording periods of up to 24-hours in a single session. It is indicated for use on adult patients who may be asymptomatic or who meet clinical indications to perform an ECG-Holter monitor exam.

    The QardioCore ECG monitor is a prescription-only device, and the reported information is provided for review by a physician who will render a diagnosis based on clinical judgment and experience.

    The QardioCore ECG Monitor is composed of six main components: i) the QardioCore sensor with Bluetooth technology, ii) a chest strap that allows fitting of OardioCore sensor, iii) a USB charging cable, iv) the Qardio App (can be downloaded and installed from the respective App store) and runs on any iOS device with iOS version 10.0 or later, v) the cloud based server where the Qardio App stores and retrieves data, and vi) the ECG Viewer which provides a web interface to the doctor to view the data sent by the iPhone application.

    The QardioCore device is a wearable device that captures information through a single-channel ECG. The data is then encrypted and transmitted via Bluetooth Low Energy to the Qardio App, installed on a compatible mobile platform. The OardioCore is supplied with chest straps accommodating chest sizes ranging from 27-5 to 43 inches. An optional XL chest strap is available from 41.7 to 59.8 inches. The device is provided with a USB Type-A cable to charge the device.

    The Qardio App (which can be installed from the user's respective app store), can transmit the data, via Wi-Fi or standard data mobile telephony, to Qardio cloud based server for storage processing and transmission to an expert medical professional.

    The ECG Viewer application provides a web interface to the doctor to view the ECG data collected from the iPhone Application. All data that a patient accumulates using the QardioCore device is stored in the central server. The ECG Viewer provides the doctor with ECG data and Heart Rate (BPM), which a doctor can use as additional information for forming a medical diagnosis. The doctor is able to see both the ECG data and Heart Rate (BPM) as soon as the data becomes available in the central server provided that the patient accepts the doctor's request for access. The device does not include automated analysis except for heart rate calculation. QardioCore is not suitable for physicians who need to perform ECG diagnoses such as myocardial ischemia, left ventricular hypertrophy or specific bundle branch blocks that require multiple and precise electrode placement and consistent wave amplitude.

    AI/ML Overview

    The provided text describes the QardioCore ECG ambulatory monitoring device and its substantial equivalence determination by the FDA. Here's a breakdown of the requested information, specifically focusing on the clinical testing to establish acceptance criteria and prove its performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document explicitly states that the purpose of the clinical study was to demonstrate the non-inferiority of the QardioCore ECG Monitor in terms of signal quality compared to a conventional, FDA-cleared, 3-channel ECG-Holter monitor device over a 24-hour recording time with continuous recording. While specific numerical acceptance criteria for "non-inferiority" are not listed, the general acceptance can be inferred from the conclusion that the device "demonstrated the substantial equivalence with the predicate device."

    Acceptance Criteria (Implied)Reported Device Performance
    Non-inferiority in ECG signal quality compared to a standard 3-channel Holter monitor over 24 hours."The data provided demonstrated the substantial equivalence with the predicate device." The study measured various qualitative and quantitative metrics, including signal quality, relevant ECG waveform amplitude and intervals, artifact burden.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in the provided text. The document mentions "a population consistent with the device indications for use," but does not give a numerical count of participants in the clinical study.
    • Data Provenance: The study was a prospective clinical study conducted to demonstrate the accuracy of the QardioCore. The country of origin of the data is not specified beyond the device being from a US-based company (San Francisco, California).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: Not explicitly stated.
    • Qualifications of Experts: Not explicitly stated. The text mentions that ECGs from both devices "were collected and analyzed in a blinded fashion," implying expert review, but does not detail their numbers or qualifications.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly stated. The text mentions that ECGs "were collected and analyzed in a blinded fashion," which is a common practice in clinical studies, but gives no specifics on how disagreements among experts (if multiple were used) were resolved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This document describes a device for capturing, storing, transmitting, and displaying ECG information. It specifically states: "The device does not include automated analysis except for heart rate calculation." Therefore, a comparative effectiveness study involving human readers improving with AI assistance would not be applicable, as the device's primary function is data acquisition and display for physician review, not AI-driven interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm-only performance study was not done. The device's current functionality, as described, is for presenting data to a physician for review, with only heart rate calculation being automated. The "reported information is provided for review by a physician who will render a diagnosis based on clinical judgment and experience."

    7. The Type of Ground Truth Used

    • The ground truth was established by comparison with a reference device, specifically the ELA Medical, Inc., Spiderview Holter ECG recorder (K032466), which is an FDA-cleared 3-channel ECG-Holter monitor. The study aimed to demonstrate non-inferiority in signal quality against this established medical device.

    8. The Sample Size for the Training Set

    • Not applicable. The document describes a medical device for data acquisition and display, not a machine learning model that requires a training set in the conventional sense. The "training" in this context would refer to the development and testing of the device's hardware and software according to recognized standards.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no mention of a machine learning training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1