510(k) Data Aggregation

    K Number
    K240910

    Validate with FDA (Live)

    Date Cleared
    2024-06-12

    (71 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EO Chex Indicator Tape for Ethylene Oxide gas sterilization is designed to secure wrapped packs by a health care provider. EO Chex indicator tape demonstrates that the packs have been exposed to an EO gas sterlization process, and to distinguish between processed and unprocessed packs. The EO process indicator turns from pink to orange-brown when exposed to EO gas sterilization conditions, thus providing an indication of processed items.

    The EO Chex indicator tape can be used in the following EO sterilization cycles:

    1. 38°C, 736 mg/L EO, 40-80% RH, 4.5 hours exposure
    2. 38°C, 759 mg/ L EO, 40-80% RH, 4.5 hours exposure
    3. 54℃, 600 mg/ L EO, 40-60% RH, 45 min or longer
    4. 55°C, 736 mg/ L EO, 40-80% RH, 1 hour exposure
    5. 55°C, 759 mg/ L EO, 40-80% RH, 1 hour exposure
    Device Description

    The EO Chex "1" indicator tape is a single-use, beige color pressure-sensitive adhesive tape with pink stripes that can be used to secure packages for Ethylene Oxide Gas (EO) sterilization.

    The pink stripes are process chemical indicators for monitoring of exposure to Ethylene Oxide sterilization. After sterilization exposure the pink stripes turn to orange-brown.

    The EO Chex indicator tape is made in rolls ¾ inch and 1 inch wide and 55-60 yards long. Each roll has a core with 3-inch internal diameter that allows for use with any standard tape dispenser.

    The indicator responds to the critical parameters of an EO sterilization cycle: exposure time, temperature, relative humidity, and concentration of Ethylene Oxide gas. During EO sterilization process indicator ink chemicals react forming an orange-brown compound. The degree of the reaction depends on the sterilization exposure. When the parameters achieve the required level, the indicator ink chemistry changes color from pink to orange-brown. If the parameters do not achieve the required level, the indicator color remains as any shade of pink.

    AI/ML Overview

    This FDA 510(k) summary describes the acceptance criteria and the study proving the device (EO Chex™ Indicator Tape) meets these criteria. Since this is for a physical/chemical sterilization process indicator, the "study" is a series of non-clinical performance tests rather than a clinical trial with human subjects. Thus, many of the typical questions for AI medical devices (e.g., sample size for test set, number of experts, MRMC studies) are not applicable.

    Here's the breakdown based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for the EO Chex™ Indicator Tape is its conformance to the ANSI/AAMI/ISO 11140-1:2014 standard for Type 1 process indicators, along with successful performance in typical EO sterilization cycles, biocompatibility, endpoint stability, shelf-life, and adhesive tests.

    Table of Acceptance Criteria and Reported Device Performance:

    TestPurposeAcceptance CriteriaReported Device Performance (Result)
    Performance testing in BIER vessel (ANSI/AAMI/ISO 11140-1:2014 for Type 1 Indicator)To demonstrate conformance to the requirements specified in ISO 11140-1:2014 for process indicators.Specific Pass Conditions: - 37°C, RH 60%, EO 600mg/L, 25 min: orange-brown - 54°C, RH 60%, EO 600mg/L, 20 min: orange-brown Specific Fail Conditions (no color change or markedly different): - 37°C, RH 60%, EO 600mg/L, 3 min - 54°C, RH 60%, EO 600mg/L, 2 min - 60°C, RH ≥85%, EO - none, 90 minPassed
    Testing in EO cycles typical for healthcare useTo demonstrate that EO Chex indicator tape achieves specified end color in typical cycles.Color change from pink to orange-brown (in specified cycles: 38°C, 736 mg/L EO, 40-80% RH, 4.5 hrs; 38°C, 759 mg/L EO, 40-80% RH, 4.5 hrs; 54°C, 600 mg/L EO, 40-60% RH, 45 min+; 55°C, 736 mg/L EO, 40-80% RH, 1 hr; 55°C, 759 mg/L EO, 40-80% RH, 1 hr).Passed
    Biocompatibility study and ink transfer testTo demonstrate that the indicator does not create biocompatibility issues for healthcare professionals and patients; and to confirm no ink transfer.Evaluation of individual components for biocompatibility and indicators with similar formulation with long history on the market. Ink Transfer Requirement (ISO 11140-1:2014, 6.2.2): No ink transfer should be observed on unprocessed and EO processed samples.Passed
    End point stability and shelf-life studyTo confirm acceptable stability of color after processing (passed/failed conditions) and to demonstrate performance parameters are met throughout the declared shelf-life using real-time exposure.Endpoint Stability: EO Chex indicator tape processed in Pass and Fail cycles at various time points after production and at the end of shelf-life should demonstrate stable color for 180 days. Shelf-Life: Met specifications after 24 months of real-time shelf-life exposure.Passed
    Adhesive test healthcare applicationsTo evaluate if the adhesive is suitable for its function and does not deteriorate during the sterilization process.The test is considered a pass if at least 95% of the indicators at different stages of shelf life remain on surfaces of sterilization packaging materials before and after EO processing.Passed

    Study Details (Non-Clinical Performance Testing):

    This section addresses the specific points requested, noting their applicability to a physical/chemical indicator.

    1. Sample size used for the test set and the data provenance:

      • The document does not explicitly state the numerical sample size for each test (e.g., number of indicator strips tested). However, the testing was conducted according to specified industrial standards (ANSI/AAMI/ISO 11140-1:2014), which implicitly define the required sample sizes and methodologies for such performance evaluations.
      • Data Provenance: Not explicitly stated, but typical for 510(k) submissions, the testing would have been conducted by the manufacturer (Propper Manufacturing Co., Inc.) or a qualified testing lab, likely within the US, as the submitter is based in New York. The data is retrospective in the sense that it was collected prior to the 510(k) submission for the purpose of demonstrating substantial equivalence.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable in the context of this device. Ground truth for a chemical indicator is established by physical/chemical reactions under controlled conditions (e.g., color change based on precise exposure to sterilant). It does not involve expert human interpretation of complex data like medical images. The "ground truth" is defined by the physical/chemical properties and the specified color change according to the ISO standard.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable as the evaluation relies on objective color change criteria and physical performance metrics rather than subjective human interpretation requiring adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable as the device is a chemical indicator, not an AI-assisted diagnostic tool or system that involves human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is not applicable. The device is a physical indicator. Its "performance" is its ability to change color as designed, which is a standalone function in itself, but not an "algorithm."
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth is established by objective physical and chemical criteria, specifically:
        • Defined color changes: Pink to orange-brown under specific EO sterilization conditions (Pass).
        • Absence of color change or markedly different color: Under defined challenging or insufficient sterilization conditions (Fail).
        • Adherence/physical integrity standards: For the adhesive and ink transfer.
        • Biocompatibility standards: Related to material safety.
      • These are based on the requirements of ANSI/AAMI/ISO 11140-1:2014 and the manufacturer's internal specifications derived from these standards.
    7. The sample size for the training set:

      • This is not applicable. The device is a chemical indicator that operates based on material properties and chemical reactions. There is no "training set" in the machine learning sense. The development of the indicator chemistry and manufacturing process would involve internal R&D and optimization, but this is not analogous to training an AI model.
    8. How the ground truth for the training set was established:

      • This is not applicable for the reasons stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222137

    Validate with FDA (Live)

    Date Cleared
    2022-12-05

    (139 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OK Plus indicator is a chemical indicator designed for monitoring the efficacy of steam sterlization process. The indicator responds to all critical sterilization parameters. The OK Plus indicator changes color from off-white to black to indicate that the conditions of the cycle have been met. The indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

    *Gravity: 121 C/250 F - 30 minutes

    • Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes

    • Vacuum assisted (prevacuum): 132 C/270 F - 10 minutes

    • Vacuum assisted (prevacuum): 134 C/273 F - 3.5 minutes

    • Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes

    • *Vacuum assisted (prevacuum): 135 C/275 F 3 minutes
      The OK Plus indicator has the following Stated Values determined in Resistometer:
    • 121 C/250 F - 15 minutes

    • 132 C/270 F - 4 minutes

    • 134 C/273 F - 3 minutes

    • 135 C/275 F - 2.9 minutes

    Device Description

    The OK Plus indicator is a single use chemical indicator designed for steam sterilization monitoring. Each indicator consists of reactive steam indicator ink printed on a 4" x 9/16" substrate paper strip. It can be also printed on other substrate sizes, for example 8" x 9/16" paper. OK Plus indicators are sold in boxes of 250 strips.

    The indicator responds to all critical parameters of a steam sterilization cycle: exposure time, temperature, and presence of saturated steam sterilization process indicator ink chemicals react forming a black compound. The degree of the reaction depends on the sterilization exposure. When the parameters achieve required level, the indicator ink chemistry changes color from off-white to black. If the parameters do not achieve the required level, the indicator color will be light brown or brown.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of a study for the "OK Plus Indicator" device, a chemical indicator for steam sterilization monitoring.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestPurposeAcceptance CriteriaResult
    Performance testing in Steam BIER vesselDemonstrate conformance with half-cycle requirements. Indicator should show end-point color when bacteria/spores are deactivated.Color changes - end-point colors: 121°C - 15 min 00 s - black; 132°C - 4 min 00 s - black; 134°C - 3 min 00 s - black; 135°C - 2 min 54 s - blackPassed
    Testing in Dry Heat ovenConfirm that indicator does not change color in absence of steam.When tested in cycles at 140°C+/-2°C and time 30+/1 min as per ANSI/AAMI/ISO 11140-1:2014, the indicator should not achieve end color. Also, should not demonstrate color change to end point in 180°C - 60 min dry heat typical cycle.Passed
    Testing against biological indicatorEstablish correlation between performance of OK Plus indicator and Steam Biological indicator.The OK Plus indicator should not achieve end-point color before the biological indicator is inactivated.Passed
    Single parameter variation testingConfirm that OK Plus indicator is sensitive to critical sterilization parameters.Variation of one parameter while other ones are maintained steady. OK Plus indicator should not reach specified end-point black color.Passed
    Testing in cycles typical for healthcareDemonstrate OK Plus indicator achieves specified end color in typical hospital-size and table-top sterilizers.Color change from off-white to black.Passed
    Biocompatibility study and ink transferDemonstrate that the indicator does not create biocompatibility issues and no ink transfer occurs.Evaluation of individual components for biocompatibility and review of biocompatibility of indicators with similar formulation. Testing according to ISO 11140-1:2014 (Requirement: 6.2.2. No ink transfer should be observed on unprocessed and Steam processed samples).Passed
    End-point stability and shelf-life studyConfirm acceptable stability after processing ("Pass" and "Fail" conditions) and maintenance of performance over shelf life.OK Plus indicators processed in Pass and Fail cycles at various time points after production and at the end of shelf life should demonstrate stable color for at least 6 months. Meet specifications after real-time 48 months shelf-life exposure.Passed

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes used for each individual test in the "Summary of non-clinical testing" section. It broadly discusses "testing" and "cycles" without providing numerical quantities for the indicators tested in each scenario. No information on data provenance (e.g., country of origin, retrospective/prospective) is provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The device's performance is assessed based on objective color changes and correlation with established standards (like biological indicators), rather than expert interpretation of a ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable and therefore not provided. The assessment of the indicator's performance is based on a clear, objective color change (off-white to black) as per pre-defined criteria, not subjective human adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study is mentioned. This type of study is typically for diagnostic imaging devices where human interpretation is a key component. The OK Plus indicator is a chemical indicator with an objective color change.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    The studies described are inherently "standalone" in the context of the device's function. The OK Plus indicator's performance is evaluated based on its chemical reaction and color change under specific conditions, without requiring human intervention for its primary function of indicating sterilization efficacy. The "result" column indicates whether the indicator "Passed" or "Failed" specific criteria, demonstrating its inherent performance.

    7. Type of Ground Truth Used:

    The ground truth used for these tests is based on:

    • Physical/Chemical Standards: Defined temperature, time, and steam conditions (e.g., 121°C - 15 min 00 s for black color).
    • ANSI/AAMI/ISO 11140-1:2014: International standards for chemical indicators.
    • FDA Guidance: "Pre-market Notification [510(k)] Submissions for Chemical Indicators, 2003."
    • Correlation with Biological Indicators: The "inactivation" of biological indicators serves as a biological ground truth for sterilization efficacy.

    8. Sample Size for the Training Set:

    This information is not provided. Chemical indicators are typically not "trained" in the same way machine learning algorithms are. Their performance is inherent to their chemical formulation, and verification is done through rigorous testing, not a training/validation split.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no mention of a "training set" in the context of this chemical indicator, the establishment of ground truth for a training set is not applicable. The device's efficacy is proven by meeting the predetermined performance criteria against established chemical and biological standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212592

    Validate with FDA (Live)

    Date Cleared
    2022-01-20

    (157 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gas-Chex Indicator is an integrating indicator designed to respond to all critical parameters over a specified range of Ethylene Oxide sterilization cycles. The Gas-Chex Indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

    1. 37°C, 736 mg/l EO, ≥35% RH, 3 hours exposure
    2. 37°C, 759 mg/l EO, ≥35% RH, 3 hours exposure
    3. 38°C, 736 mg/l EO, 40-80% RH, 4.5 hours exposure
    4. 38°C, 759 mg/l EO, 40-80% RH, 4.5 hours exposure
    5. 55°C, 736 mg/l EO, ≥35% RH, 1 hour exposure
    6. 55°C, 759 mg/l EO, ≥35% RH, 1 hour exposure
    7. 55°C. 600 mg/l EO. 60% RH. 4 hours exposure
      The Gas-Chex Indicator for EO sterilization has the following minimum Stated Values determined in a Resistometer:
      54℃, 42 min, EO 600 mg/l, RH 60±10%
      37℃,75 min. EO 600 mg/l. RH 60±10%
      55°C, 36 min, EO 736 mg/l, RH 50±10%
      37°C, 75 min, EO 736 mg/l, RH 50±10%
      55°C, 35 min, EO 759 mg/l, RH 50±10%
      37°C, 75 min, EO 759 mg/l, RH 50±10%
    Device Description

    The Gas-Chex® Indicator is a single use chemical integrating indicator designed for Ethylene Oxide (EO) sterilization monitoring. Each indicator consists of reactive EO indicator ink printed on a 4" x 9/16" substrate paper strip. It can be also printed on other substrate sizes, for example 8" x 9/16" paper. Gas-Chex® Indicators are sold in boxes of 250 strips.
    The indicator responds to all critical parameters of an EO sterilization cycle: exposure time, temperature, relative humidity and amount of Ethylene Oxide gas. During sterilization process the indicator ink chemicals react to EO gas forming a green compound. The degree of the reaction depends on the sterilization exposure. When the parameters achieve required level, the indicator ink chemistry changes color from brown-red to green. If the parameters do not achieve the required level, the indicator color will be brown-red, or brown-yellow.

    AI/ML Overview

    The acceptance criteria and the study proving the device meets these criteria are detailed in the "Summary of Non-Clinical Testing" section of the provided document.


    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Performance testing in EO BIER vesselColor changes:- Pass Condition (Green): - 54°C, 42 min, EO 600 mg/l, RH 60% - 37°C, 75 min, EO 600 mg/l, RH 60% - 55°C, 36 min, EO 736 mg/l, RH 50% - 37°C, 73 min, EO 736 mg/l, RH 50% - 55°C, 35 min, EO 759 mg/l, RH 50% - 37°C, 73 min, EO 759 mg/l, RH 50%- Fail Condition (Brown, Brown-red, Brown-yellow): - 49°C, 33.6min, EO 510 mg/l, RH 60% - 32°C, 60 min, EO 510 mg/l, RH 60% - 50°C, 28.8min, EO 625mg/l, RH 50% - 32°C, 58.4min, EO 625mg/l, RH 50% - 50°C, 28 min, EO 645 mg/l, RH 50% - 32°C, 58.4min, EO 645mg/l, RH 50%Passed
    Testing in "No EO Gas" cyclesWhen tested in cycle with temperature 60°C±2°C, RH≥85%, time 90 min±1min, the indicator should not achieve end-point color.Passed
    Testing against biological indicatorThe integrator does not achieve end-point color before the biological indicator is inactivated and demonstrates parallel performance to the biological indicator.Passed
    Single parameter variation testingVariation of one parameter while other ones are maintained steady. Gas-Chex Indicator should not reach specified end-point green color.Passed
    Testing in cycles with parameters typical for healthcareColor change from brown-red to green.Passed
    Biocompatibility study and ink transfer testEvaluation of individual components for biocompatibility and review of biocompatibility of indicators with similar formulation with history on the market.Testing according to ISO 11140-1:2014. Requirement: 6.2.2. No ink transfer should be observed on unprocessed and EO processed samples.Passed
    End-point stability and shelf-life studyGas-Chex indicators processed in Pass and Fail cycles at various time points after production and at the end of shelf life should demonstrate stable color for at least 3 months.Meet specifications after real-time 36 months shelf-life exposure.Passed

    Detailed Information on the Study:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the exact number of individual Gas-Chex Indicators or test runs used for each test. Instead, it states that various conditions (e.g., specific temperatures, EO concentrations, RH, and exposure times) were tested.
      • Data Provenance: The tests were performed to demonstrate conformance with FDA guidance for industry and FDA Staff: Pre-market notification [510(k)] submissions for chemical indicators, 2003. These are laboratory-based, non-clinical tests conducted in controlled environments (Ethylene Oxide BIER vessel). The specific country of origin is not explicitly stated for the testing location, but the applicant company is located in Long Island City, New York, USA. The study design is prospective in nature as it involves planned testing under specific conditions.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This type of device (chemical indicator for sterilization) does not typically involve human expert interpretation of complex data for ground truth establishment.
      • The "ground truth" is established by the known, objective physical and chemical parameters of the sterilization cycles (e.g., temperature, time, EO concentration, RH) and the expected color change response of the indicator based on its design and the established standards (e.g., ISO 11140-1:2014 and FDA guidance). The color change itself (brown-red to green, or remaining brown-red/brown-yellow) is a direct, observable, and objective outcome.
      • Therefore, no human experts are described as being used in an "adjudication" sense to establish fundamental ground truth for the device's performance in these non-clinical tests.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • No adjudication method from human experts is described since the assessment of the chemical indicator's color change is a direct observation against defined criteria (e.g., "green" for pass, "brown, brown-red, brown-yellow" for fail). This is a purely objective assessment against pre-defined color states.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This type of study is not applicable to a chemical sterilization indicator, which is a passive device with a direct visual output, rather than an AI-driven diagnostic or interpretative tool that assists human readers.
    5. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

      • Not applicable. The device is a chemical indicator that changes color; there is no algorithm involved. Its performance is inherent to its chemical formulation and physical design.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth is based on the known, controlled physical and chemical parameters of the Ethylene Oxide sterilization cycles (temperature, Ethylene Oxide concentration, relative humidity, and exposure time) generated in a Resistometer (BIER vessel). The expected outcome (color change to green for successful sterilization parameters, or no change/partial change for insufficient parameters) is derived from established sterilization science and regulatory standards for chemical indicators.
    7. The sample size for the training set:

      • Not applicable. This device is a passive chemical indicator, not an AI/machine learning model, so there is no concept of a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210866

    Validate with FDA (Live)

    Date Cleared
    2021-06-21

    (90 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steri-Dot Process Indicator for EO gas sterilization is designed for use by a health care provider to demonstrate that the unit or load has been exposed to an EO gas sterilization process, and to distinguish between processed and unprocessed units or loads. The indicator dots turn from purple-red to green when exposed to EO gas sterilization conditions, thus providing an indication of processed items.

    The Steri-Dot process indicator can be used in the following EO sterilization cycles:

    1. 37°C, 736 mg/l EO, ≥35% RH, 3 hours exposure
    2. 37°C, 759 mg/l EO, ≥35% RH, 3 hours exposure
    3. 54°C, 600 mg/l EO, 40-60% RH, 45 min or longer exposure
    4. 54°C, 736 mg/l EO, ≥35% RH, 1 hour exposure
    5. 55°C, 759 mg/l EO, ≥35% RH, l hour exposure
    6. 55°C, 600 mg/l EO, 40-60% RH, 4 hours exposure
    Device Description

    The Steri-Dot Process Indicator is a single use chemical indicator designed for Ethylene Oxide (EO) sterilization monitoring. Each indicator consists of reactive EO indicator ink printed on a substrate paper circle, 3/4" or 1" in diameter, with adhesive backing. Individual indicators are printed with reactive ink only.

    The indicator responds to the critical parameters of an EO sterilization cycle: exposure time, temperature, relative humidity and amount of Ethylene Oxide gas. During EO sterilization process indicator ink chemicals react forming a green compound. The degree of the reaction depends on the sterilization exposure. When the parameters achieve required level, the indicator ink chemistry changes color from purple-red to green. If the parameters do not achieve the required level, the indicator color will be brown or red.

    AI/ML Overview

    The provided document describes the Steri-Dot® Process Indicator, a chemical indicator for Ethylene Oxide (EO) gas sterilization. The information primarily focuses on non-clinical testing to demonstrate its performance and substantial equivalence to a predicate device.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a "Summary of Non-clinical Testing" table.

    TestPurposeAcceptance CriteriaResult
    ANSI/AAMI/ISO 11140-1:2014 testing for Type 1 indicator.To demonstrate conformance of Steri-Dot indicator to the requirements specified in ISO 11140-1:2014 for process indicators.37°C, RH 60%, EO 600mg/l, 3 min: no color change or color markedly different compared to green 37°C, RH 60%, EO 600mg/l, 25 min: green color 54°C, RH 60%, EO 600mg/l, 2 min: no color change or color markedly different compared to green 54°C, RH 60%, EO 600mg/l, 20 min: green color 60°C, RH ≥85%, EO - none, 90 min: no changePassed
    Testing in EO cycles with parameters used in healthcare sterilization.To demonstrate that Steri-Dot Process indicator achieves specified end color in typical cycles.Color change from purple-red to greenPassed
    Biocompatibility study and ink transfer testTo demonstrate that the indicator does not create biocompatibility issues to health care professionals and patients. The exposure to health care professional is minimal and well below any identified toxic thresholds for the compounds.Evaluation of individual components for biocompatibility and indicators with similar formulation with long history of on the market. Testing according to ISO 11140-1:2014. Requirement: 6.2.2. No ink transfer should be observed on unprocessed and EO processed samples.Passed
    End point stability and shelf-life studyTo confirm that Steri-Dot process indicator has acceptable stability after processing when achieved and not achieved end point color (parameters when tested using real-time shelf-life exposure method).Criteria not explicitly stated in the provided text for this row.Implied Passed, as overall conclusion is positive.
    Adhesive test – healthcare applicationsThe purpose of the test is to evaluate if the adhesive is suitable for its function and does not deteriorate during sterilization process.The test is considered a pass if at least 95% of the indicators at different stages of shelf life remain on surfaces of sterilization packaging materials before and after Steri processing.Passed

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical laboratory testing of a device rather than a study involving human data or a "test set" in the context of AI/machine learning. Therefore:

    • Sample size: Not explicitly stated in terms of number of indicators tested for each condition. The testing is described as fulfilling the requirements of ISO standards and demonstrating performance.
    • Data provenance: Not applicable in the sense of patient data. The tests are laboratory-based, performed on the Steri-Dot Process Indicator itself.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable to this document. The device is a chemical indicator that changes color. The "ground truth" for its performance is determined by the physical chemical reaction and visual observation based on established standards (e.g., ISO 11140-1:2014). There is no mention of expert readers or human interpretation of complex images/data to establish ground truth as would be the case for an AI medical device.

    4. Adjudication Method for the Test Set

    This section is not applicable. As it's a chemical indicator with a clear color change endpoint, there is no ambiguity requiring an adjudication method by multiple human readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

    This section is not applicable. An MRMC study is relevant for AI devices where human readers interpret medical images or data. The Steri-Dot Process Indicator is a chemical indicator, not an AI-powered diagnostic tool. There are no human readers involved in its direct "performance" evaluation beyond visual confirmation of the chemical reaction.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    This section is not applicable. The Steri-Dot Process Indicator is a physical chemical indicator, not an algorithm or software. Its performance is inherent to its material composition and chemical reaction, not a computational model.

    7. The Type of Ground Truth Used

    The ground truth for this device's performance is based on:

    • Standardized Chemical Reaction: The chemical indicator's color change in response to specific sterilization parameters (temperature, humidity, EO concentration, exposure time).
    • Compliance with International Standards: Specifically, ANSI/AAMI/ISO 11140-1:2014 for Type 1 process indicators. This standard defines the expected color changes under various exposure and non-exposure conditions to ethylene oxide.
    • Visual Observation: The "result" column in the table simply states "Passed," implying that the visual color change (or lack thereof) met the specified acceptance criteria.

    8. The Sample Size for the Training Set

    This section is not applicable. This is a physical device, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable for the same reason as point 8.


    In summary: The provided document is a 510(k) summary for a physical chemical indicator. It details the product's function, mechanism, and non-clinical performance testing against established ANSI/AAMI/ISO standards. Many of the questions you posed (e.g., sample size for AI test/training sets, expert consensus, MRMC studies) are primarily relevant for AI/software as a medical device (SaMD) rather than a simple chemical indicator. The "acceptance criteria" and "study proving device meets acceptance criteria" here refer to laboratory testing demonstrating compliance with performance standards for physical chemical indicators.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210553

    Validate with FDA (Live)

    Date Cleared
    2021-05-19

    (83 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steam-Dot Process Indicator for steam sterilization is designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed units or loads. The indicator dots turn from white to dark brown/black when sterilization conditions, thus providing an indication of processed items.

    The Steam-Dot process indicator can be used in the following steam sterilization cycles:

    *Gravity: 121 C/250 F - 30 minutes

      • Vacuum assisted (pre-vacuum): 132 C/270 F 3 minutes
      • Vacuum assisted (pre-vacuum): 132 C/270 F 4 minutes
      • Vacuum assisted (pre-vacuum): 134 C/273 F 3 minutes
      • Vacuum assisted (pre-vacuum): 134 C/273 F 4 minutes
      • Vacuum assisted (pre-vacuum): 135 C/275 F 3 minutes
    Device Description

    The Steam-Dot Process Indicator is a single use chemical indicator designed for steam sterilization monitoring. Each indicator consists of reactive steam indicator ink printed on a substrate paper circle, 3/4" in diameter, with adhesive backing. Individual indicators are printed with reactive ink only.

    The indicator responds to the critical parameters of a steam sterilization cycle: exposure time, temperature, and presence of saturated steam sterilization process indicator ink chemicals react forming a black compound. The degree of the reaction depends on the sterilization exposure. When the parameters achieve required level, the indicator ink chemistry changes color from white to black/dark brown. If the parameters do not achieve the required level, the indicator color will be light brown or crème.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called the "Steam-Dot Process Indicator." The document outlines the device's indications for use, performance criteria, and the results of non-clinical testing to demonstrate its substantial equivalence to a predicate device.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    ANSI/AAMI/ISO 11140-1:2014 testing for Type 1 indicator121°C-10 min: dark brown or black color121°C-2 min: no color change or color markedly different compared to dark brown or black134°C-2 min: dark brown or black color134°C-0.3 min: no color change or color markedly different compared to dark brown or black140°C-30 min Dry heat: no color changePassed
    FDA Guidance for industry for chemical indicators. Steam process indicator performance test.121°C-10 min: dark brown or black color121°C-2 min: no color change or color markedly different compared to dark brown or black132-135°C-2 min: dark brown or black color132-135°C-20sec: no color change or color markedly different compared to dark brown or black140°C-30 min Dry heat: no color changePassed
    Testing in hospital type sterilizers in gravity and pre-vacuum 510k cleared cycles.Color change from white to dark brown or blackPassed
    Biocompatibility study and ink transfer testIndividual components should not create biocompatibility issues.Testing according to ISO 11140-1:2014. Requirement: 6.2.2. No ink transfer should be observed on unprocessed and steam processed indicators.Passed
    End point stability and shelf-life studySteam-Dot indicators processed in Pass and Fail cycles at various time points after production and at the end of shelf life should demonstrate stable color for 110 days.Meet specifications after real-time 48 months shelf-life exposure.Passed

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (e.g., number of indicators or test runs) used for each test. It broadly states that the "performance of the Steam-Dot process indicator meets the requirements of ANSI/AAMI/ISO 11140-1:2014 for Type 1 process indicators and the requirements of FDA guidance for industry and FDA Staff: Pre-market notification [510(k)] submissions for chemical indicators, 2003."

    The provenance of the data is not explicitly stated in terms of country of origin or whether it's retrospective or prospective. However, given that it's a 510(k) submission, the testing would be prospective and conducted under controlled laboratory conditions, likely in the US or in a facility adhering to US regulatory standards for medical device testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided in the document. The "ground truth" for a chemical indicator like the Steam-Dot is based on the physical and chemical reaction of the indicator ink to specific temperature, time, and steam exposure parameters, as defined by international standards (ANSI/AAMI/ISO 11140-1:2014) and FDA guidance. There are no human "experts" establishing a subjective ground truth in this context; rather, instrumental measurements of physical parameters and observable color changes define the outcome.

    4. Adjudication method for the test set

    Not applicable. As described above, the determination of "pass" or "fail" for the chemical indicator is objective, based on measurable physical parameters and a defined color change endpoints. There is no subjective interpretation requiring an adjudication process by multiple reviewers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a chemical indicator that changes color to confirm exposure to sterilization conditions. It is not an AI-powered diagnostic device or an imaging system requiring human interpretation or assistance from AI. Therefore, an MRMC study is irrelevant for this product.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm-only device. Its function is based on a chemical reaction, not a computational algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device is based on defined physical and chemical parameters of steam sterilization as established by international standards (ANSI/AAMI/ISO 11140-1:2014) and regulatory guidance (FDA Guidance for industry for chemical indicators). This involves exposing the indicators to precise temperature, time, and steam conditions in controlled laboratory settings and observing the specified color change. It is an objective, laboratory-defined ground truth, not reliant on expert consensus, pathology, or clinical outcomes data.

    8. The sample size for the training set

    Not applicable. This device is a chemical indicator, not an AI/machine learning model that requires a training set. Its function is based on a chemical formulation and its reaction to sterilization conditions.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.