    Intended Use

    The STERLINK Plus sterilizer with STERLOAD cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

    The STERLINK Plus can sterilize *:

    • · Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • · Medical devices with a single stainless steel lumen with:
    • o An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

    *The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 3 .97 lbs.

    Tyvek® Roll with CI for STERLINK Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

    The materials compatibility for use in the Tyvek® Roll with CI for STERLINK Sterilizer, when used in CHAMBER mode, are as follows: Aluminum 6061, Stainless Steel 304, Stainless Steel 316L, Titanium, Acrylonitrile Butadiene Styrene (ABS), High Density Polyethylene (HPDE), Polypropylene (PP), Polytetrafluorethylene (PTFE) and Silicone (Hardness 50).

    The maximum load weight that can be placed in the Tyvek® roll is:

    • · 3.97 pounds (1.8kg) for CHAMBER mode of FPS-15s Plus
    • 3.97 pounds (1.8kg) for CHAMBER mode of STERLINK plus sterilizer
    • 1.54 pounds (0.7kg) for CHAMBER mode of STERLINK mini sterilizer

    The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1 :2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK sterilizer.

    The Tyvek® Roll with CI for STERLINK Sterilizer is offered in the follow 1 type: · Sterilization roll. Flat

    Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 10^6 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

    Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

    Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.

    Terragene Cintape® (CT40) is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterlization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

    The self-contained biological indicator and chemical processing indicators are intended for use with.

    • · CHAMBER mode of FPS-15s Plus sterilizer
    • · CHAMBER mode of STERLink plus sterilizer
    • · CHAMBER mode of STERLINK mini sterilizer
    Device Description

    The STERLINK plus sterilizer with STERLOAD™ cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments. This sterilizer offers an effective, safe, fast, economical, easy to use, reliable, and flexible sterilization method.

    This system consists of a main device connected pump module and cassette which are called the STERLINK plus and STERLOAD™ cassette, respectively. The STERLOAD™ cassette contains 58-59.5% (weight concentration) of hydrogen peroxide (HzO2) which is utilized as the sterilant.

    Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to be used to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices for up to 1 month post sterilization.

    The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINKTM sterilizer.

    The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).

    Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FRLCD or Mini-Bio Auto-Readers Incubators.

    Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.

    Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper, and plastic.

    AI/ML Overview

    This document describes the acceptance criteria and supporting study for the STERLINK™ plus Sterilizer with STERLOAD™ Cassette and associated components (Tyvek® Roll with CI for STERLINK™ Sterilizer, Terragene Bionova® SCBI (BT96), Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye® (CD42), and Terragene Cintape® (CT40)).

    The document asserts substantial equivalence to a predicate device (STERLINK™ FPS-15s Plus Sterilizer with STERLOAD™ Cassette, K212200), and therefore, the acceptance criteria are generally met by demonstrating compliance with recognized standards and establishing equivalence in performance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission demonstrating substantial equivalence to a predicate, specific numerical "acceptance criteria" for novel performance metrics are not explicitly stated in the provided text as they would be for a de novo device. Instead, the acceptance criteria are implicitly that the device performs as well as or better than the predicate device and meets established standards for sterilization, packaging, and indicators. The reported device performance is presented as "Pass" for compliance with these standards.

    Acceptance Criteria CategorySpecific Criteria / Standard MetReported Device Performance
    STERLINK™ plus Sterilizer with STERLOAD™ Cassette
    Risk managementISO 14971Pass
    Human factors and usability engineeringGuidance / IEC 62366-1Pass
    BiocompatibilityISO 10993-5Pass
    Software validationIEC 62304 / FDA GuidancePass
    Electrical safetyIEC 60601-1 / IEC 61010-1Pass
    Electromagnetic compatibility (EMC)IEC 60101-1-2Pass
    Resistance validation for biological indicator testISO 11138-1:2017Pass
    Lumen sterilizationISO 14937:2009Pass
    Surface sterilizationISO 11737-1:2018 / ISO 11737-2:2009Pass
    Mated surface sterilizationISO 14937:2009 / ISO 11737-1:2018 / ISO 11737-2:2009Pass
    Simulated use testASTM E1837-96(2014) / ISO 11737-1:2018Pass
    In-use testASTM E1837-96(2014)Pass
    Sporicidal activity testAOAC 966.04Pass
    Bacteriostasis testISO 11737-1:2018Pass
    Material compatibility testASTM D638, ASTM E8/E8M-16ae1, ASTM D790, ASTM E290-14, ASTM D256, ASTM E23-18, ASTM E1164, ASTM D3985, ASTM F1249Pass
    Delivery validationASTM D4169-14Pass
    Hydrogen peroxide gas detection testOSHA analytical method 1019Pass
    Tyvek® Roll with CI for STERLINK™ Sterilizer
    Sterilization efficacy test (for sterilant penetration)ISO 14937:2009Pass
    CI of sterilization roll validationANSI/AAMI/ISO 11140-1:2014Pass
    Shelf-life (Internal pressurization test)ASTM F1980-16:2016 / ASTM F1140/F1140M-13:2013Pass
    Shelf-life (Visual inspection test)ASTM F1980-16:2016 / ASTM F1886/F1886M-16:2016Pass
    Shelf-life (Dye penetration test)ASTM F1980-16:2016 / ASTM F1929-15:2015Pass
    Shelf-life (Tensile strength of Tyvek®)ASTM F1980-16:2016 / ASTM D5035-11:2019Pass
    Shelf-life (Tensile strength of plastic film)ASTM F1980-16:2016 / ASTM D882Pass
    Shelf-life (Seal strength)ASTM F1980-16:2016 / ASTM F88Pass
    Shelf-life (Tear resistance)ASTM F1980-16:2016 / ASTM D1922-20Pass
    Shelf-life (Microbial Barrier Test)ASTM F1980-16:2016 / DIN 58953-6:2016Pass
    CI of sterilization roll validationANSI/AAMI/ISO 11140-1:2014Pass
    Residual sterilant on Tyvek® validationManufacturer's internal test standardPass
    Sterilization Process Indicator (BT96, CD42, CT40)
    Resistance validation for biological indicator test (BT96)ISO 11138-1:2017Pass
    BI & Test pack validation testManufacturer's internal standardPass
    Chemical indicator validation (CD42, CT40)Manufacturer's internal standardPass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific test or the country of origin/provenance of the data. It refers to various ISO and ASTM standards, which typically specify test methodologies and sample size requirements. The studies are non-clinical, likely laboratory-based evaluations to demonstrate performance against these standards. The submission originates from Plasmapp Co., Ltd. in South Korea, suggesting studies were likely conducted either internally or by contract research organizations, potentially in South Korea, in accordance with international standards. The studies would be prospective in nature, as they are tests performed on the device to prove its compliance with standards and equivalence.

    • Sample Size: Not explicitly stated for individual tests.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The applicant is Plasmapp Co., Ltd., BVC-111, 125, Gwahak-ro, Yuseong-gu, Daejeon, 34141, Rep. of Korea (South Korea).
    • Retrospective or Prospective: Implied to be prospective as the tests were conducted to demonstrate device performance against defined standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This submission is for a sterilization system and associated components, which are evaluated based on defined physical, chemical, and biological performance standards (e.g., sporicidal activity, material compatibility, chemical indicator color change) rather than interpretation of complex medical images or patient data. Therefore, the concept of "experts establishing ground truth" in the diagnostic AI sense (e.g., radiologists for an imaging device) is not directly applicable here. The "ground truth" is established by the well-defined parameters and pass/fail criteria of the referenced ISO and ASTM standards for sterility, material science, and indicator performance.
    The experts involved would be laboratory technicians and engineers performing and verifying the tests, and potentially subject matter experts in sterilization science and engineering within Plasmapp and the regulatory body. Their qualifications are not specified but would be implicit in their roles and adherence to the testing standards.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is a technical performance study against objective standards, not a diagnostic study requiring adjudication of expert interpretations.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a sterilization system and its accessories, not an AI-based diagnostic device intended for use by human readers.


    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical sterilization system, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance evaluation of sterilization devices and accessories typically refers to:

    • Biological inactivation: Complete kill of test microorganisms (e.g., Geobacillus stearothermophilus spores) for sterilizers and biological indicators, confirmed through culture methods.
    • Physical and chemical parameters: Confirmation that specified physical conditions (temperature, pressure, time) were met and chemical indicators changed appropriately, as defined by industry standards and internal specifications.
    • Material integrity: No degradation or adverse effects on device materials after sterilization, as per material compatibility tests.
    • Packaging integrity: Maintenance of sterility barrier, confirmed by physical and microbial barrier tests.

    These 'ground truths' are established through validated laboratory tests against international and national standards (ISO, ASTM, AOAC).


    8. The sample size for the training set

    Not applicable. This is a hardware device (sterilizer, packaging, indicators) and not an AI/machine learning model that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI/machine learning model, there is no training set and therefore no ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223015

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-09

    (222 days)

    Product Code
    Regulation Number
    880.6860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STERLINK™ FPS-15s Plus sterilizer is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal medical devices used in healthcare facilities. A preprogrammed sterilization lumen cycle operates at low pressure and is thus suitable for processing medical devices sensitive to heat and moisture.

    The STERLINK™ FPS-15s Plus can sterilize*:

    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices with a single stainless steel lumen with:
      • STERPACK™ plus: An inside diameter of 1.6 mm or larger and a length of 200 mm or shorter
      • STERLOAD™: An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

    *The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 1 1b and 3.97 lbs with STERPACK™ plus and STERLOAD™, respectively,

    Device Description

    The STERLINK™ FPS-15s Plus sterilizer with STERPACK™ plus cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments.

    This system consists of a main device and cassettes which are called the STERLINK™ FPS-15s Plus and STERPACK™ plus or STERLOAD™, respectively. The STERPACK™ plus and STERLOAD™ cassette contain 58-59.5% (weight concentration) of hydrogen peroxide (H2O2) which is utilized as the sterilant.

    AI/ML Overview

    Based on the provided text, the device in question is the STERLINK™ FPS-15s Plus sterilizer with STERPACK™ plus cassette, a vaporized hydrogen peroxide sterilizer. The acceptance criteria and the study proving the device meets these criteria are primarily focused on the non-clinical performance of the sterilizer, specifically its ability to sterilize medical devices. The information provided is typical of a 510(k) summary for a medical device.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally "Pass" for compliance with various standards and guidance documents. The reported device performance is that it "Passes" these tests.

    Test CategoryStandard/Guidance DocumentAcceptance CriteriaReported Device Performance
    Human factors and usability engineeringGuidance for Industry and Food and Drug Administration Staff, "Appling Human Factors and Usability Engineering to Medical Devices", issued on February 3, 2016; IEC 62366-1:2015PassPass
    Shelf-life test (Sterilant Preservation Test)Manufacturer's internal standardPass1 year
    BiocompatibilityISO 10993-5:2009PassPass
    Software validationIEC 62304:2006 + AMD1:2015; Guidance for Industry and Food and Drug Administration Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device", issued on May 11, 2005PassPass
    Electrical safetyEN 61010-1:2010; EN 61010-2-040:2015; IEC 60601-1:2005 (3rd Ed.) + CORR.1:2006 + CORR.2:2007 + A1:2012; EN 60601-1-2:2015PassPass
    RF disturbanceEN 55011:2009/A1:2010PassPass
    Electromagnetic compatibility (EMC)EN 61326-1:2013; EN 61000-3-2:2014; EN 61000-3-3:2013; EN 60601-1-2:2015PassPass
    Resistance validation for biological indicator testISO 11138-1:2017PassPass
    Lumen sterilizationISO 14937:2009PassPass
    Surface sterilizationISO 14937:2009; ISO 11737-1:2018; ISO 11737-2:2009PassPass
    Mated surface sterilizationISO 14937:2009; ISO 11737-1:2018; ISO 11737-2:2009PassPass
    Simulated use testASTM E1837-96:2014; ISO 11737-1:2018PassPass
    In-use testASTM E1837-96:2014PassPass
    Sporicidal activity testAOAC 966.04 (AOAC 6.3.05:2013)PassPass
    Bacteriostasis testISO 11737-1:2018PassPass
    Material compatibility testASTM D638; ASTM E8/E8M-16ae1; ASTM D790; ASTM E290-14; ASTM D256; ASTM E23-18; ASTM E1164; ASTM E313; ASTM D3985; ASTM F1249PassPass
    Delivery validation (Sterilizer)ASTM D4169-14PassPass
    Delivery validation (Sterilant)Manufacturer's internal standardPassPass
    Hydrogen peroxide gas detection testOSHA analytical method 1019PassPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific "sample size" in terms of number of medical devices or test runs for each individual performance test (e.g., lumen sterilization). Instead, it states for lumen sizes: "The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load." It also mentions "a validation load with a total weight of 1 lb and 3.97 lbs with STERPACK™ plus and STERLOAD™, respectively."

    • Sample Size for performance tests: For lumen sterilization, it mentions a maximum of five lumens per load. Other tests don't specify a detailed sample size breakdown within this document. The "Pass" results indicate that the testing met the minimum requirements for the specific standards.
    • Data Provenance: The manufacturer is "Plasmapp Co., Ltd." in "Rep. of Korea (South Korea)." The document implies that the testing was conducted to meet the requirements of the U.S. FDA, but the location where the actual tests were performed is not explicitly stated. The nature of these tests (e.g., sterilization efficacy) strongly suggests they were performed as prospective validation studies to demonstrate the device's performance.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    This type of information (number and qualifications of experts establishing ground truth for a test set) is generally relevant for AI/ML-driven diagnostic devices where human expert consensus forms the ground truth for image interpretation or diagnosis. For a sterilizer, the "ground truth" for its performance is typically established through objective, laboratory-based biological and physical testing using standardized methods and biological indicators (e.g., Geobacillus stearothermophilus). Therefore, the concept of "experts establishing ground truth for a test set" in the context of human interpretation of data is not applicable here. The "experts" would be the scientists and technicians conducting the validation tests according to the specified standards.

    4. Adjudication Method for the Test Set

    As the "ground truth" for a sterilizer's performance is objective (e.g., kill/no kill of microorganisms, meeting sensor specifications), adjudication methods like 2+1 or 3+1 (common in image interpretation studies) are not applicable. The device's performance is measured against established scientific and engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC study was not done. MRMC studies are typically performed for diagnostic devices (especially those involving image interpretation) to evaluate the impact of a device (e.g., AI assistance) on human reader performance. This document is for a medical device sterilizer, which does not involve human interpretation of cases or AI assistance in that context.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the performance tests described are standalone in the sense that they evaluate the sterilizer's ability to sterilize medical devices independently of human subjective assessment. The "algorithm" here is the sterilization cycle itself, and its effectiveness is measured directly through biological and chemical indicators. The results are based on the device's inherent design and operational parameters.

    7. The Type of Ground Truth Used

    The ground truth used for proving the sterilizer's efficacy is based on:

    • Biological Activity: The ability to inactivate microorganisms (e.g., Geobacillus stearothermophilus spores) as measured by standard biological indicator tests (e.g., ISO 11138-1:2017, ISO 11737-1:2018, AOAC 966.04).
    • Physical Parameters: Confirmation that the sterilizer operates within specified parameters (temperature, pressure, time) and that the sterilant is delivered effectively.
    • Material Compatibility: Ensuring that the sterilization process does not degrade the materials of the medical devices being sterilized.
    • Engineering Standards: Compliance with electrical safety, EMC, software validation, and other relevant engineering standards (e.g., IEC, EN, ASTM).

    8. The Sample Size for the Training Set

    This document does not refer to a "training set." The concept of training sets (and validation/test sets derived from them) is specific to machine learning or AI models that learn from data. This document is about a hardware device (sterilizer) whose performance is validated through defined physical and biological tests, not through training a model on a dataset.

    9. How the Ground Truth for the Training Set was Established

    As there is no "training set" in the context of an AI/ML model for this device, this question is not applicable. The sterilizer's performance is established through controlled, laboratory-based validation studies against regulatory and industry standards for sterilization efficacy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223025

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-09

    (222 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in pouch plus mode and chamber mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

    The maximum load weight that can be placed in the Tyvek® roll is 1 lb (450g) for pouch plus mode and 3.97 bs (1.8kg) for chamber mode.

    The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2015) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.

    The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the following one type:

    · Sterilization roll, Flat

    Device Description

    Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to be used to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices for up to 1 month post sterilization.

    The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINKTM sterilizer.

    The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).

    AI/ML Overview

    This prompt describes a medical device, specifically a "Tyvek® Roll with CI for STERLINK™ Sterilizer," and provides information about its regulatory submission (K223025) and comparison to a predicate device (K212198). The document focuses on regulatory approval based on demonstrating substantial equivalence, primarily through non-clinical testing.

    However, the provided text does not describe an AI/ML powered medical device and therefore does not contain information typically associated with acceptance criteria or studies used to prove performance for such devices. The questions in the prompt (sample size for test set, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, ground truth types) are all relevant to the validation of AI/ML algorithms, not to the physical properties and sterilization efficacy of a sterilization wrap.

    Therefore, I cannot directly answer your detailed questions about acceptance criteria and study design as if this were an AI/ML device. The information provided is for a traditional medical device (sterilization wrap with a chemical indicator).

    Here's how I can address the request based on the actual content of the document:

    The document provides a summary of non-clinical testing performed to demonstrate that the subject device (Tyvek® Roll with CI for STERLINK™ Sterilizer) is substantially equivalent to a predicate device, focusing on its ability to maintain sterility and the performance of its chemical indicator.

    Acceptance Criteria and Study for a Traditional Medical Device (Sterilization Wrap):

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" for each test but states "Pass" for all performed tests, indicating that the device met the requirements of the standards cited. The acceptance criteria are implicitly defined by the successful completion of the referenced standards.

    Acceptance Criteria (Implied by Standard)Reported Device Performance
    Sterility Maintenance/Integrity:
    Sterilant Penetration (efficacy of sterilization) (ISO 14937)Pass
    Internal pressurization test (ASTM F1980-16, ASTM F1140/F1140M-13)Pass
    Visual inspection test (ASTM F1980-16, ASTM F1886/F1886M-16)Pass
    Dye penetration test (ASTM F1980-16, ASTM F1929-15)Pass
    Tensile strength of Tyvek® (ASTM F1980-16, ASTM D5035-11)Pass
    Tensile strength of plastic film (ASTM F1980-16, ASTM D882)Pass
    Seal strength (ASTM F1980-16, ASTM F88)Pass
    Tear resistance (ASTM F1980-16, ASTM D1922-20)Pass
    Microbial Barrier Test (ASTM F1980-16, DIN 58953-6)Pass
    Chemical Indicator Performance:
    Chemical indicator (CI) of sterilization roll validation (ANSI/AAMI/ISO 11140-1)Pass
    Biocompatibility/Safety:
    Toxicological Properties (Biocompatibility Tests) (ISO 10993-5)Pass
    Residual sterilant on Tyvek® validation (Internal test standard)Pass

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the exact sample sizes for each of the non-clinical tests listed.
    • Data provenance is not explicitly stated in terms of country of origin or retrospective/prospective. The tests were performed as part of the regulatory submission by Plasmapp Co., Ltd. (South Korea) and its manufacturer Sigma Medical Supplies Corp.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable to the non-clinical performance testing of a sterilization wrap. "Ground truth" in the context of material science and sterilization efficacy is established by validated laboratory testing against specified standards, not expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable as the evaluation involves meeting established engineering and biological standards, not human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable as the device is a physical sterilization wrap, not an AI-assisted diagnostic or decision-support tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the device is a physical sterilization wrap.

    7. The type of ground truth used:

    • The "ground truth" for this device's performance is established by validated laboratory testing against recognized international and national standards (e.g., ISO 14937, ANSI/AAMI/ISO 11140-1, ASTM series, DIN 58953-6). For biocompatibility, it's ISO 10993-5. For chemical indicator performance, it involves observable color changes under controlled sterilization conditions as per ISO 11140-1.

    8. The sample size for the training set:

    • This is not applicable as there is no AI/ML algorithm involved that would require a training set.

    9. How the ground truth for the training set was established:

    • This is not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STERLINK mini sterilizer with STERLOAD™ mini cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

    The STERLINK mini can sterilize *:

    • · Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • · Medical devices with a single stainless-steel lumen with:
    • o An inside diameter of 1.6 mm or larger and a length of 200 mm or shorter

    *The validation testing was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 1.54 lbs.

    Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

    The materials compatible for use in the Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in Chamber mode, are as follows: Aluminum 5052, Aluminum 6061, Stainless Steel 316L, Titanium, Acrylonitrile Butadiene Styrene (ABS), High Density Polyethylene (HDPE), Polypropylene (PP), Polytetrafluoroethylene (PTFE) and Silicone (Hardness 50).

    The maximum load weight that can be placed in the Tyvek® roll is:

    • · 3.97 pounds (1.8 kg) for CHAMBER mode of FPS-15s Plus sterilizer
    • · 1.54 pounds (0.7 kg) for CHAMBER mode of STERLINK mini sterilizer

    The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140 1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.

    The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type: · Sterilization roll, Flat

    Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 106 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

    Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

    Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.

    Terragene Cintape® (CT40) is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

    The self-contained biological indicator and chemical processing indicators are intended for use with:

    • · CHAMBER mode of FPS-15s Plus sterilizer
    • · CHAMBER mode of STERLINK mini sterilizer
    Device Description

    The STERLINK mini sterilizer with STERLOAD™ mini cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments. This system consists of a main device connected pump module and cassette which are called the STERLINK mini and STERLOAD™ mini cassette, respectively. The STERLOAD™ mini cassette contains 58-59.5% (weight concentration) of hydrogen peroxide (H2O2) which is utilized as the sterilant.

    Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed medical devices for up to 1 month post sterilization. The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINKTM sterilizer. The Tyvek® Roll with CI for STERLINKIM Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).

    Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FRLCD or MiniBio Auto-Readers Incubators.

    Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.

    Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper and plastic.

    AI/ML Overview

    The provided text describes acceptance criteria and study results for medical sterilization devices (STERLINK mini Sterilizer with STERLOAD mini Cassette, Tyvek Roll with CI for STERLINK Sterilizer, and Sterilization Process Indicator for STERLINK Sterilizer).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: STERLINK mini Sterilizer with STERLOAD™ mini Cassette

    TestPurposeAcceptable CriteriaReported Device Performance
    Human factors and usability engineeringTo confirm whether the user error affects the risk to user safety and performance of the device when the device is used according to the intended use in accordance with IEC 62366-1.User safety risk and device performance degradation due to user errors should not occur.Pass
    BiocompatibilityTo confirm the biological safety of the sterilized load in accordance with ISO 10993-5.≤ Grade 2Pass
    Software validationTo verify that the software is compatible for the intended use of the device in accordance with IEC 62304.The risk of all identified software hazards should be reduced to an acceptable risk level and the device is appropriate for its intended use.Pass
    Electrical safetyTo confirm the electrical safety of the device in accordance with IEC 60601-1, IEC 61010-1, and IEC 61010-2-040.The device conforms to the relevant standards.Pass
    Electromagnetic compatibility (EMC)To confirm the electromagnetic wave safety of the device in accordance with IEC 60601-1-2 and EN 55011.The device conforms to the relevant standards.Pass
    Resistance validation for biological indicator testTo validate the D-value and total kill endpoint of the biological indicators in accordance with ISO 11138-1.The total kill endpoint time should be shorter than the sterilization half cycle.Pass
    Lumen sterilizationTo confirm the lumen sterilization performance in accordance with ISO 14937.The SAL (Sterility Assurance Level) should be at least 10^-6^.Pass
    Surface sterilizationTo confirm the surface sterilization performance of various materials in accordance with ISO 14937, ISO 11737-1, and ISO11737-2.The SAL (Sterility Assurance Level) should be at least 10^-6^.Pass
    Mated surface sterilizationTo confirm the mated surface sterilization performance of various materials in accordance with ISO 14937, ISO 11737-1, and ISO11737-2.The SAL (Sterility Assurance Level) should be at least 10^-6^.Pass
    Simulated use testTo demonstrate the efficacy of the sterilization validation on medical devices contaminated by organic and inorganic matters with bacterial spores in accordance with ASTM E1837-96 and ISO 11737-1.The microorganisms should not survive after sterilization.Pass
    In-use testTo demonstrate the sterilization validation of the used and clinically pretreated medical devices in accordance with ASTM E1837-96.The microorganisms should not survive after sterilization.Pass
    Sporicidal activity testTo confirm the sporicidal activity of the device in accordance with AOAC Official Method 966.04.All spores inoculated on the carriers must be sterilized.Pass
    Bacteriostasis testTo confirm the bacteriostasis action of sterilized various materials after sterilization process in accordance with ISO 11737-1.Sterilized materials should not exhibit the bacteriostatic effects.Pass
    Material compatibility testTo demonstrate the compatibility of various materials in the sterilization cycle in accordance with ASTM D638, ASTM E8/E8M, ASTM D790, ASTM E290-14, ASTM D256, ASTM E23-18, ASTM E1164, ASTM D3985, and ASTM E1240.The physical properties of the material after sterilization should be similar to those before sterilization.Pass
    Delivery validationTo validate the packaging and performance of the device after exposure to simulated shipping conditions in accordance with ASTM D4169-14.After the distribution cycle of the drop test, significant deformations and defects in the electrical and mechanical performance of the device should not be observed.Pass
    Hydrogen peroxide gas detection testTo measure the concentration of the hydrogen peroxide emitted from the device in accordance with Occupational Safety and Health Administration (OSHA, TWA).< 1 ppm hydrogen peroxide gas detectionPass
    Sterilant preservation testTo verify the shelf life of the sterilant cassette in accordance with the internal test standard.The sterilant should meet the specified weight, concentration, and pH standards during the shelf life.Pass

    Device: Tyvek® Roll with CI for STERLINK™ Sterilizer

    Performance TestPurposeAcceptable CriteriaReported Device Performance
    Sterilization efficacy testTo ensure that the device is suitable for the sterilization process in accordance with ISO 14937.The SAL should be at least 10^-6^.Pass
    CI of sterilization roll validationTo confirm that the shelf life before and after the sterilization of device, and to ensure CI of the device is compatible for the sterilization process in accordance with ANSI/AAMI/ISO 11140-1.- CI should not change completely with temperature, and sterilant exposure time of less than 7 seconds, during the shelf life.- CI should maintain the changed color after real-time aging during shelf life after sterilization.Pass
    Internal pressurization testTo verify the internal pressurization of the sterilized device at the shelf life in accordance with ASTM F1980-16 and ASTM F1140/F1140M-13.- Burst pressure: $ \ge $ 1kPa- Creep time: > 30 sec in creep pressure (> 80% of burst pressure)Pass
    Visual inspection testTo confirm the sterilized device integrity was maintained at shelf life in accordance with ASTM F1980-16 and ASTM F1886/F1886M-16.There should be no defects in the sealing area.Pass
    Dye penetration testTo confirm the packaging of the sterilized device was maintained at shelf life in accordance with ASTM F1980-16 and ASTM F1929-15.There should be no dye leakage.Pass
    Tensile strength of Tyvek®To confirm the physical properties of the sterilized device at shelf life in accordance with ASTM F1980-16, ASTM D5035-11, ASTM D882, ASTM F88, and ASTM D1922-20.The physical properties must meet the criteria specified by the manufacturer.Pass
    Tensile strength of plastic film(Same purpose as above)(Same criteria as above)Pass
    Seal strength(Same purpose as above)(Same criteria as above)Pass
    Tear resistance(Same purpose as above)(Same criteria as above)Pass
    Microbial Barrier TestTo confirm the microbial barrier performance of sterilized devices at shelf life in accordance with ASTM F1980-16 and DIN 58953-6.- Test with liquid inoculum: No microbial growth should appear.- Test with gas inoculum: Only a maximum 15 colonies should grow out of a total of 10 bacterial permeability testers. The number of colonies in each bacterial permeability tester should not be greater than five.Pass
    Residual sterilant on Tyvek® validationTo confirm the time required for the residual sterilant to be reduced to a safe amount after sterilization in accordance with the internal test standard.≤ 0.05 ppmPass
    Biocompatibility testTo confirm the biological safety after sterilization in accordance with ISO 10993-5.≤ Grade 2Pass

    Device: Sterilization Process Indicator for STERLINK™ Sterilizer (BT96, CD42, CT40)

    ItemTestPurposeAcceptable CriteriaReported Device Performance
    Self-Contained Biological Indicator (BT96)Resistance validation for biological indicator testTo validate the D-value and total kill endpoint of biological indicators in accordance with ISO 11138-1.The total kill endpoint time should be shorter than the half cycle sterilant exposure time.Pass
    BI & Test pack validation testTo verify the suitability of the test pack to be used as monitoring of the sterilization process in accordance with the internal test standard.The test pack with SCBI should fail to sterilize in half cycle and succeed in full cycle.Pass
    Chemical Indicator (CD42, CT40)Performance characteristics obtained from resistometerTo evaluate the correct performance of chemical indicators, by testing them at different exposure conditions in accordance with ANSI/AAMI/ISO 11140-1.The CI should change color correctly only when exposed to the temperature and sterilant concentration conditions specified by the manufacturer.Pass
    BiocompatibilityTo demonstrate that the CI does not offset or transfer during and after the sterilization process in accordance with ANSI/AAMI/ISO 11140-1.The CI does not release any substance or bleed when the sterilization process.Pass
    Endpoint stabilityTo demonstrate the stability of the endpoint reaction of CI at the end of their shelf life and specified period after sterilization in accordance with ANSI/AAMI/ISO 11140-1.The endpoint reaction of CI should be maintained until the shelf life and specified period after sterilization.Pass
    Shelf life studyTo verify the shelf life of the CI in accordance with ANSI/AAMI/ISO 11140-1.The endpoint reaction of the real time aged CI should be maintained until shelf life.Pass
    Chemical indicator validationTo confirm the compatibility of the CI for the sterilization process in accordance with the internal test standard.The minimum time required for color change of CI within the sterilization process time should be confirmed.Pass

    Additional Study Information:

    The provided text describes non-clinical tests performed to demonstrate substantial equivalence to predicate devices, not clinical studies involving human patients or complex AI algorithms requiring ground truth from experts. The devices are sterilizers and sterilization indicators, which are typically evaluated through microbiological and engineering performance testing.

    Therefore, many of the specific questions about "test set," "ground truth," "experts," "adjudication," and "AI assistance" are not directly applicable to the type of device and testing described in this document.

    For completeness, here's what can be extracted or inferred:

    1. Sample size used for the test set and the data provenance:

      • Sample size: Not explicitly stated as a single "test set" size. The text mentions "maximum of five (5) lumens per load" for lumen sterilization validation and "a validation load with a total weight of 1.54 lbs." for the sterilizer. For the Tyvek roll microbial barrier test, it mentions "a total of 10 bacterial permeability testers." Specific quantities for other tests are not provided.
      • Data provenance: Not specified. These are non-clinical engineering and microbiological tests, likely conducted in a laboratory setting by the manufacturer (Plasmapp Co., Ltd. and Terragene S.A.). The country of origin of the manufacturer is South Korea for Plasmapp and presumably Argentina for Terragene (as it's a known Argentinian company). The studies are "non-clinical tests," implying they were conducted in a controlled environment, not on human subjects.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. The tests are laboratory-based and use established microbiological and engineering standards (e.g., ISO, ASTM, AOAC). "Ground truth" is determined by scientific measurements and observation of bacterial kill, physical properties, or chemical indicator changes, not by expert consensus on subjective interpretations.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjective assessments (e.g., radiological reads) where multiple experts might disagree. These tests are objective measurements against defined standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. These devices do not involve "human readers" interpreting "cases" with or without "AI assistance."

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no AI algorithm component to these devices described in the document.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for these sterilization and chemical indicator tests is based on:

      • Microbiological kill: Demonstrating the inactivation of a specified population of resistant spores (e.g., Geobacillus stearothermophilus) to a defined Sterility Assurance Level (SAL), typically 10^-6^. This is confirmed through laboratory culture methods.
      • Physical and chemical property measurements: Adherence to established engineering standards (e.g., tensile strength, burst pressure, color change properties of chemical indicators).
      • Safety standards: Compliance with international safety and electromagnetic compatibility standards.
    7. The sample size for the training set: Not applicable. These devices do not involve machine learning or AI models that require a "training set."

    8. How the ground truth for the training set was established: Not applicable, as there is no training set mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212193

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-10-22

    (100 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 10% Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

    Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

    Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.

    Terragene Cintape® CT40 is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

    The self-contained biological indicator and chemical processing indicators are intended for use with the STERLINK™ FPS-15s Plus when operating in chamber mode.

    Device Description

    Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FR, IC10/20FRLCD or MiniBio Auto-Readers Incubators.

    Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.

    Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper and plastic.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for sterilization process indicators (biological and chemical) for use with a specific sterilizer. The study is a non-clinical performance test, and as such, several of the requested sections regarding AI/human reader performance, ground truth establishment, and training sets are not applicable.

    Here's an analysis of the acceptance criteria and study as presented:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for each test are implicitly defined by the "Pass" result, meaning the device met the requirements of the referenced standard or guidance document. The actual quantitative thresholds for "Pass" are not detailed in this summary but would be specified within the referenced standards (e.g., ANSI/AAMI/ISO 11138-1:2017).

    Test PerformedReferenced Standard/GuidanceAcceptance Criteria (Implicit)Reported Device Performance
    Self-Contained Biological Indicator (BT96)
    Viable spore population assayANSI/AAMI/ISO 11138-1:2017Met the requirements of the standardPass
    Resistance characteristics studyANSI/AAMI/ISO 11138-1:2017Met the requirements of the standardPass
    Carrier and primary packaging materials evaluationANSI/AAMI/ISO 11138-1:2017Met the requirements of the standardPass
    Holding time assessmentGuidance for Industry and FDA Staff - Biological Indicator (BI) Premarket Notification [510(k)] SubmissionsMet the requirements of the guidancePass
    Reduced incubation time validation testGuidance for Industry and FDA Staff - Biological Indicator (BI) Premarket Notification [510(k)] SubmissionsMet the requirements of the guidancePass
    Recovery protocols: Recovery medium testANSI/AAMI/ISO 11138-1:2017Met the requirements of the standardPass
    Visual readout stability: Visual inspection testANSI/AAMI/ISO 11138-1:2017Met the requirements of the standardPass
    Shelf life studyANSI/AAMI/ISO 11138-1:2017Met the requirements of the standard (Demonstrated 2 years)2 years
    Resistance Validation for Biological IndicatorISO 11138-1:2017Met the requirements of the standardPass
    BI & Test Pack ValidationGuidance for Industry and FDA Staff - Biological Indicator (BI) Premarket Notification [510(k)] SubmissionsMet the requirements of the guidancePass
    Chemical Indicator (CD42, CT40)
    Performance characteristics obtained from resistometerGuidance for Industry and FDA Staff - Premarket Notification [510(k)] Submissions for Chemical Indicators / ANSI/AAMI/ISO 11140-1:2014Met the requirements of the guidance/standardPass
    BiocompatibilityGuidance for Industry and FDA Staff - Premarket Notification [510(k)] Submissions for Chemical Indicators / ANSI/AAMI/ISO 11140-1:2014Met the requirements of the guidance/standardPass
    Endpoint stabilityANSI/AAMI/ISO 11140-1:2014Met the requirements of the standardPass
    Shelf life studyPremarket Notification [510(k)] Submissions for Chemical IndicatorsMet the requirements of the guidance (Demonstrated 5 years)5 years
    Chemical Indicator ValidationANSI/AAMI/ISO 11140-1:2014Met the requirements of the standardPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test. However, the tests are non-clinical, laboratory-based studies rather than human subject trials. The data provenance would be the laboratory where these tests were conducted, likely in South Korea (Plasmapp Co., Ltd. is based there, and Terragene® S.A. is the manufacturer). These studies are inherently prospective in nature as they involve testing the devices against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is a non-clinical performance study of sterilization indicators, not a study evaluating human interpretation of medical images or diagnoses. The "ground truth" for these tests is defined by the physical or chemical standards and performance requirements outlined in the referenced ANSI/AAMI/ISO standards and FDA guidance documents.

    4. Adjudication Method for the Test Set

    Not applicable. There is no human interpretation or subjective assessment that would require an adjudication method in this type of non-clinical performance testing. The tests yield objective, quantitative or qualitative results (e.g., spore kill, color change, D-value measurements) that are then compared against predefined criteria in the standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC study was not done. This is a non-clinical performance study of sterilization process indicators, not a diagnostic device that would involve human readers interpreting results.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable. The device itself (the biological and chemical indicators) is not an "algorithm" in the typical sense of AI. The performance tests evaluate the physical and chemical characteristics of the indicators in a standalone manner against the specified standards, without human intervention in the measurement of their performance (though human technicians certainly perform the tests).

    7. The Type of Ground Truth Used

    The ground truth used for these performance studies is based on established industry standards and regulatory guidance documents. Specifically:

    • ANSI/AAMI/ISO 11138-1:2017: Sterilization of Health Care Products - Biological Indicators - Part 1: General Requirements.
    • ANSI/AAMI/ISO 11140-1:2014: Sterilization of Health Care Products - Chemical Indicators - Part 1: General Requirements.
    • FDA Guidance for Industry and FDA Staff - Biological Indicator (BI) Premarket Notification [510(k)] Submissions.
    • FDA Guidance for Industry and FDA Staff - Premarket Notification [510(k)] Submissions for Chemical Indicators.

    These standards define the expected performance characteristics (e.g., spore population, resistance, color change properties, stability) that the devices must meet.

    8. The Sample Size for the Training Set

    Not applicable. This is a non-clinical performance study of physical devices, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As this is not an AI model, there is no training set and therefore no ground truth established for one. The "ground truth" for the device's acceptable performance is defined by the rigorous specifications and testing protocols within the aforementioned international standards and FDA guidance documents.